- POWer:CURRENTSOurce {CH<x>|REF<x>}
- POWer:CURRENTSOurce?
- POWer:DISplay {ON|OFF|1|0}
- POWer:DISplay?
- POWer:GATESOurce {CH<x>|REF<x>|NONe}
- POWer:GATESOurce?
- POWer:GATing {OFF|SCREen|CURSor}
- POWer:GATing?
- POWer:HARMonics:DISplay:SELect {ODD|EVEN|ALL}
- POWer:HARMonics:DISplay:SELect?
- POWer:HARMonics:DISplay:TYPe {GRAph|TABle}
- POWer:HARMonics:DISplay:TYPe?
- POWer:HARMonics:FREQRef {VOLTage|CURRent|HARMSOURce|FIXEDFREQuency}
- POWer:HARMonics:FREQRef:FIXEDFREQValue <NR3>
- POWer:HARMonics:FREQRef:FIXEDFREQValue?
- POWer:HARMonics:FREQRef?
- POWer:HARMonics:IEC:CLAss {A|B|C<x>|D}
- POWer:HARMonics:IEC:CLAss?
- POWer:HARMonics:IEC:FILter {ON|OFF|1|0}
- POWer:HARMonics:IEC:FILter?
- POWer:HARMonics:IEC:FUNDamental <NR3>
- POWer:HARMonics:IEC:FUNDamental?
- POWer:HARMonics:IEC:GROUPing {ON|OFF|1|0}
- POWer:HARMonics:IEC:GROUPing?
- POWer:HARMonics:IEC:INPUTPOWer <NR3>
- POWer:HARMonics:IEC:INPUTPOWer?
- POWer:HARMonics:IEC:LINEFREQuency <NR1>
- POWer:HARMonics:IEC:LINEFREQuency?
- POWer:HARMonics:IEC:OBSPERiod <NR3>
- POWer:HARMonics:IEC:OBSPERiod?
- POWer:HARMonics:IEC:POWERFACtor <NR3>
- POWer:HARMonics:IEC:POWERFACtor?
- POWer:HARMonics:MIL:FUNDamental:CALCmethod {MEAS|USER}
- POWer:HARMonics:MIL:FUNDamental:CALCmethod?
- POWer:HARMonics:MIL:FUNDamental:USER:CURrent <NR3>
- POWer:HARMonics:MIL:FUNDamental:USER:CURrent?
- POWer:HARMonics:MIL:LINEFREQuency <NR1>
- POWer:HARMonics:MIL:LINEFREQuency?
- POWer:HARMonics:MIL:POWERLEVel {LOW|HIGH}
- POWer:HARMonics:MIL:POWERLEVel?
- POWer:HARMonics:NR_HARMonics <NR3>
- POWer:HARMonics:NR_HARMonics?
- POWer:HARMonics:RESults:HAR<x>:FREQuency?
- POWer:HARMonics:RESults:HAR<x>:IECMAX?
- POWer:HARMonics:RESults:HAR<x>:LIMit?
- POWer:HARMonics:RESults:HAR<x>:PHASe?
- POWer:HARMonics:RESults:HAR<x>:RMS:ABSolute?
- POWer:HARMonics:RESults:HAR<x>:RMS:PERCent?
- POWer:HARMonics:RESults:HAR<x>:TEST:IEC:CLASSALIMit?
- POWer:HARMonics:RESults:HAR<x>:TEST:IEC:NORMAL?
- POWer:HARMonics:RESults:HAR<x>:TEST:IEC:POHCLIMit?
- POWer:HARMonics:RESults:HAR<x>:TEST:MIL:NORMAL?
- POWer:HARMonics:RESults:IEC:FUNDamental?
- POWer:HARMonics:RESults:IEC:HARM3ALTernate?
- POWer:HARMonics:RESults:IEC:HARM5ALTernate?
- POWer:HARMonics:RESults:IEC:POHC?
- POWer:HARMonics:RESults:IEC:POHL?
- POWer:HARMonics:RESults:IEC:POWERFactor?
- POWer:HARMonics:RESults:IEC:POWer?
- POWer:HARMonics:RESults:PASSFail?
- POWer:HARMonics:RESults:RMS?
- POWer:HARMonics:RESults:SAVe <String>
- POWer:HARMonics:RESults:THDF?
- POWer:HARMonics:RESults:THDR?
- POWer:HARMonics:SOURce {VOLTage|CURRent}
- POWer:HARMonics:SOURce?
- POWer:HARMonics:STANDard {NONe|IEC|MIL}
- POWer:HARMonics:STANDard?
- POWer:INDICators {ON|OFF|1|0}
- POWer:INDICators?
- POWer:MODulation:SOUrce {VOLTage|CURRent}
- POWer:MODulation:SOUrce?
- POWer:MODulation:TYPe {PWIdth|NWIdth|PERIod|PDUty|NDUty|FREQuency}
- POWer:MODulation:TYPe?
- POWer:QUALity:APPpwr?
- POWer:QUALity:DISplay:APPpwr {ON|OFF|1|0}
- POWer:QUALity:DISplay:APPpwr?
- POWer:QUALity:DISplay:FREQuency {ON|OFF|1|0}
- POWer:QUALity:DISplay:FREQuency?
- POWer:QUALity:DISplay:ICRESTfactor {ON|OFF|1|0}
- POWer:QUALity:DISplay:ICRESTfactor?
- POWer:QUALity:DISplay:IRMS {ON|OFF|1|0}
- POWer:QUALity:DISplay:IRMS?
- POWer:QUALity:DISplay:PHASEangle {ON|OFF|1|0}
- POWer:QUALity:DISplay:PHASEangle?
- POWer:QUALity:DISplay:POWERFACtor {ON|OFF|1|0}
- POWer:QUALity:DISplay:POWERFACtor?
- POWer:QUALity:DISplay:REACTpwr {ON|OFF|1|0}
- POWer:QUALity:DISplay:REACTpwr?
- POWer:QUALity:DISplay:TRUEpwr {ON|OFF|1|0}
- POWer:QUALity:DISplay:TRUEpwr?
- POWer:QUALity:DISplay:VCRESTfactor {ON|OFF|1|0}
- POWer:QUALity:DISplay:VCRESTfactor?
- POWer:QUALity:DISplay:VRMS {ON|OFF|1|0}
- POWer:QUALity:DISplay:VRMS?
- POWer:QUALity:FREQREFerence {VOLTage|CURRent}
- POWer:QUALity:FREQREFerence?
- POWer:QUALity:FREQuency?
- POWer:QUALity:ICRESTfactor?
- POWer:QUALity:IRMS?
- POWer:QUALity:PHASEangle?
- POWer:QUALity:POWERFACtor?
- POWer:QUALity:REACTpwr?
- POWer:QUALity:TRUEpwr?
- POWer:QUALity:VCRESTfactor?
- POWer:QUALity:VRMS?
- POWer:REFLevel:ABSolute {SETTODEFaults}
- POWer:REFLevel:ABSolute:HIGH <NR3>; Ranges={D,-1e6,+1E6}
- POWer:REFLevel:ABSolute:HIGH?
- POWer:REFLevel:ABSolute:LOW <NR3>; Ranges={D,-1e6,+1E6}
- POWer:REFLevel:ABSolute:LOW?
- POWer:REFLevel:ABSolute:MID<x> <NR3>; Ranges={D,-1e6,+1E6}
- POWer:REFLevel:ABSolute:MID<x>?
- POWer:REFLevel:HYSTeresis <NR3>
- POWer:REFLevel:HYSTeresis?
- POWer:REFLevel:METHod {ABSolute|PERCent}
- POWer:REFLevel:METHod?
- POWer:REFLevel:PERCent <SETTODEFaults>
- POWer:REFLevel:PERCent:HIGH <NR3>; Ranges={D,0.0,100.0}
- POWer:REFLevel:PERCent:HIGH?
- POWer:REFLevel:PERCent:LOW <NR3>; Ranges={D,0.0,100.0}
- POWer:REFLevel:PERCent:LOW?
- POWer:REFLevel:PERCent:MID<x> <NR3>; Ranges={D,0.0,100.0}
- POWer:REFLevel:PERCent:MID<x>?
- POWer:RIPPle {VERTAUTOset|VERTDEFault}
- POWer:RIPPle:RESults:AMPLitude?
- POWer:RIPPle:RESults:MAX?
- POWer:RIPPle:RESults:MEAN?
- POWer:RIPPle:RESults:MIN?
- POWer:RIPPle:RESults:STDdev?
- POWer:RIPPle:SOUrce {VOLTage|CURRent}
- POWer:SOA:LINear:XMAX <NR3>
- POWer:SOA:LINear:XMAX?
- POWer:SOA:LINear:XMIN <NR3>
- POWer:SOA:LINear:XMIN?
- POWer:SOA:LINear:YMAX <NR3>
- POWer:SOA:LINear:YMAX?
- POWer:SOA:LINear:YMIN <NR3>
- POWer:SOA:LINear:YMIN?
- POWer:SOA:LOG:XMAX <NR3>
- POWer:SOA:LOG:XMAX?
- POWer:SOA:LOG:XMIN <NR3>
- POWer:SOA:LOG:XMIN?
- POWer:SOA:LOG:YMAX <NR3>
- POWer:SOA:LOG:YMAX?
- POWer:SOA:LOG:YMIN <NR3>
- POWer:SOA:LOG:YMIN?
- POWer:SOA:MASK:DEFine <NR3>
- POWer:SOA:MASK:DEFine?
- POWer:SOA:MASK:MAXAmps <NR3>
- POWer:SOA:MASK:MAXAmps?
- POWer:SOA:MASK:MAXVolts <NR3>
- POWer:SOA:MASK:MAXVolts?
- POWer:SOA:MASK:MAXWatts <NR3>
- POWer:SOA:MASK:MAXWatts?
- POWer:SOA:MASK:NR_Pt?
- POWer:SOA:MASK:STATE {OFF|LIMITS|POINTS}
- POWer:SOA:MASK:STATE?
- POWer:SOA:MASK:STOPOnviol {ON|OFF|1|0}
- POWer:SOA:MASK:STOPOnviol?
- POWer:SOA:PLOTTYPe {LOG|LINear}
- POWer:SOA:PLOTTYPe?
- POWer:SOA:RESult:FAILures:QTY?
- POWer:SOA:RESult:NUMACq?
- POWer:SOA:RESult:STATE?
- POWer:STATIstics {RESET}
- POWer:STATIstics:MODe {OFF|ALL}
- POWer:STATIstics:MODe?
- POWer:STATIstics:WEIghting <NR1>;Ranges {L,2,1000}
- POWer:STATIstics:WEIghting?
- POWer:SWLoss:CONDCALCmethod {VOLTage|RDSon|VCEsat}
- POWer:SWLoss:CONDCALCmethod?
- POWer:SWLoss:CONDuction:ENERGY:MAX?
- POWer:SWLoss:CONDuction:ENERGY:MEAN?
- POWer:SWLoss:CONDuction:ENERGY:MIN?
- POWer:SWLoss:CONDuction:POWer:MAX?
- POWer:SWLoss:CONDuction:POWer:MEAN?
- POWer:SWLoss:CONDuction:POWer:MIN?
- POWer:SWLoss:DISplay {ALL|ENERGYLoss|POWERLoss}
- POWer:SWLoss:DISplay?
- POWer:SWLoss:GATe:POLarity {FALL|RISe}
- POWer:SWLoss:GATe:POLarity?
- POWer:SWLoss:GATe:TURNON <NR3>
- POWer:SWLoss:GATe:TURNON?
- POWer:SWLoss:NUMCYCles? <NR3>
- POWer:SWLoss:RDSon <NR3>
- POWer:SWLoss:RDSon?
- POWer:SWLoss:REFLevel:ABSolute:GATEMid <NR3>
- POWer:SWLoss:REFLevel:ABSolute:GATEMid?
- POWer:SWLoss:REFLevel:ABSolute:LOWCurrent <NR3>
- POWer:SWLoss:REFLevel:ABSolute:LOWCurrent?
- POWer:SWLoss:REFLevel:ABSolute:LOWVoltage <NR3>
- POWer:SWLoss:REFLevel:ABSolute:LOWVoltage?
- POWer:SWLoss:REFLevel:PERCent:GATEMid <NR3>
- POWer:SWLoss:REFLevel:PERCent:GATEMid?
- POWer:SWLoss:REFLevel:PERCent:LOWCurrent <NR3>
- POWer:SWLoss:REFLevel:PERCent:LOWCurrent?
- POWer:SWLoss:REFLevel:PERCent:LOWVoltage <NR3>
- POWer:SWLoss:REFLevel:PERCent:LOWVoltage?
- POWer:SWLoss:TOFF:ENERGY:MAX?
- POWer:SWLoss:TOFF:ENERGY:MEAN?
- POWer:SWLoss:TOFF:ENERGY:MIN?
- POWer:SWLoss:TOFF:POWer:MAX?
- POWer:SWLoss:TOFF:POWer:MEAN?
- POWer:SWLoss:TOFF:POWer:MIN?
- POWer:SWLoss:TON:ENERGY:MAX?
- POWer:SWLoss:TON:ENERGY:MEAN?
- POWer:SWLoss:TON:ENERGY:MIN?
- POWer:SWLoss:TON:POWer:MAX?
- POWer:SWLoss:TON:POWer:MEAN?
- POWer:SWLoss:TON:POWer:MIN?
- POWer:SWLoss:TOTal:ENERGY:MAX?
- POWer:SWLoss:TOTal:ENERGY:MEAN?
- POWer:SWLoss:TOTal:ENERGY:MIN?
- POWer:SWLoss:TOTal:POWer:MAX?
- POWer:SWLoss:TOTal:POWer:MEAN?
- POWer:SWLoss:TOTal:POWer:MIN?
- POWer:SWLoss:VCEsat <NR3>
- POWer:SWLoss:VCEsat?
- POWer:TYPe {NONe|QUALity|SWITCHingloss|SOA|HARMonics|RIPPle|MODULationanalysis|DESKew}
- POWer:TYPe?
- POWer:VOLTAGESOurce {CH<x>|REF<x>}
- POWer:VOLTAGESOurce?

Power

Bases: SCPICmdRead

The POWer command tree.

Usage
  • Using the .query() method will send the POWer? query.
  • Using the .verify(value) method will send the POWer? query and raise an AssertionError if the returned value does not match value.
Properties
  • .currentsource: The POWer:CURRENTSOurce command.
  • .display: The POWer:DISplay command.
  • .gatesource: The POWer:GATESOurce command.
  • .gating: The POWer:GATing command.
  • .harmonics: The POWer:HARMonics command tree.
  • .indicators: The POWer:INDICators command.
  • .modulation: The POWer:MODulation command tree.
  • .quality: The POWer:QUALity command tree.
  • .reflevel: The POWer:REFLevel command tree.
  • .ripple: The POWer:RIPPle command.
  • .soa: The POWer:SOA command tree.
  • .statistics: The POWer:STATIstics command.
  • .swloss: The POWer:SWLoss command tree.
  • .type: The POWer:TYPe command.
  • .voltagesource: The POWer:VOLTAGESOurce command.

currentsource property

currentsource: PowerCurrentsource

Return the POWer:CURRENTSOurce command.

Description
  • This command specifies the current source for the power application.
Usage
  • Using the .query() method will send the POWer:CURRENTSOurce? query.
  • Using the .verify(value) method will send the POWer:CURRENTSOurce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:CURRENTSOurce value command.
SCPI Syntax
- POWer:CURRENTSOurce {CH<x>|REF<x>}
- POWer:CURRENTSOurce?
Info
  • CH<x> sets an analog channel as the current source.
  • REF<x> sets a reference waveform as the current source.

display property

display: PowerDisplay

Return the POWer:DISplay command.

Description
  • This command controls whether or not to display the power test results. This is the equivalent to pressing the Test button and then selecting the power application. The same control is provided for each application.
Usage
  • Using the .query() method will send the POWer:DISplay? query.
  • Using the .verify(value) method will send the POWer:DISplay? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:DISplay value command.
SCPI Syntax
- POWer:DISplay {ON|OFF|1|0}
- POWer:DISplay?
Info
  • OFF or 0 turns off the display settings.
  • ON or 1 turns on the display settings.

gatesource property

gatesource: PowerGatesource

Return the POWer:GATESOurce command.

Description
  • This command specifies the gate source for the power application.
Usage
  • Using the .query() method will send the POWer:GATESOurce? query.
  • Using the .verify(value) method will send the POWer:GATESOurce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:GATESOurce value command.
SCPI Syntax
- POWer:GATESOurce {CH<x>|REF<x>|NONe}
- POWer:GATESOurce?
Info
  • CH<x> sets an analog channel as the gate source.
  • REF<x> sets a reference waveform as the gate source.
  • NONe is set when the gate source is not used in the application.

gating property

gating: PowerGating

Return the POWer:GATing command.

Description
  • This command specifies the power application gating.
Usage
  • Using the .query() method will send the POWer:GATing? query.
  • Using the .verify(value) method will send the POWer:GATing? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:GATing value command.
SCPI Syntax
- POWer:GATing {OFF|SCREen|CURSor}
- POWer:GATing?
Info
  • OFF turns off measurement gating (full record).
  • SCREen turns on gating, using the left and right edges of the screen.
  • CURSor limits measurements to the portion of the waveform between the vertical bar cursors, even if they are off screen.

harmonics property

harmonics: PowerHarmonics

Return the POWer:HARMonics command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics? query.
  • Using the .verify(value) method will send the POWer:HARMonics? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .display: The POWer:HARMonics:DISplay command tree.
  • .freqref: The POWer:HARMonics:FREQRef command.
  • .iec: The POWer:HARMonics:IEC command tree.
  • .mil: The POWer:HARMonics:MIL command tree.
  • .nr_harmonics: The POWer:HARMonics:NR_HARMonics command.
  • .results: The POWer:HARMonics:RESults command tree.
  • .source: The POWer:HARMonics:SOURce command.
  • .standard: The POWer:HARMonics:STANDard command.

indicators property

indicators: PowerIndicators

Return the POWer:INDICators command.

Description
  • This command specifies the state of the measurement indicators for the power application.
Usage
  • Using the .query() method will send the POWer:INDICators? query.
  • Using the .verify(value) method will send the POWer:INDICators? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:INDICators value command.
SCPI Syntax
- POWer:INDICators {ON|OFF|1|0}
- POWer:INDICators?
Info
  • OFF or 0 turns off the measurement indicators.
  • ON or 1 turns on the measurement indicators.

modulation property

modulation: PowerModulation

Return the POWer:MODulation command tree.

Usage
  • Using the .query() method will send the POWer:MODulation? query.
  • Using the .verify(value) method will send the POWer:MODulation? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .source: The POWer:MODulation:SOUrce command.
  • .type: The POWer:MODulation:TYPe command.

quality property

quality: PowerQuality

Return the POWer:QUALity command tree.

Usage
  • Using the .query() method will send the POWer:QUALity? query.
  • Using the .verify(value) method will send the POWer:QUALity? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .apppwr: The POWer:QUALity:APPpwr command.
  • .display: The POWer:QUALity:DISplay command tree.
  • .freqreference: The POWer:QUALity:FREQREFerence command.
  • .frequency: The POWer:QUALity:FREQuency command.
  • .icrestfactor: The POWer:QUALity:ICRESTfactor command.
  • .irms: The POWer:QUALity:IRMS command.
  • .phaseangle: The POWer:QUALity:PHASEangle command.
  • .powerfactor: The POWer:QUALity:POWERFACtor command.
  • .reactpwr: The POWer:QUALity:REACTpwr command.
  • .truepwr: The POWer:QUALity:TRUEpwr command.
  • .vcrestfactor: The POWer:QUALity:VCRESTfactor command.
  • .vrms: The POWer:QUALity:VRMS command.

reflevel property

reflevel: PowerReflevel

Return the POWer:REFLevel command tree.

Usage
  • Using the .query() method will send the POWer:REFLevel? query.
  • Using the .verify(value) method will send the POWer:REFLevel? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .absolute: The POWer:REFLevel:ABSolute command.
  • .hysteresis: The POWer:REFLevel:HYSTeresis command.
  • .method: The POWer:REFLevel:METHod command.
  • .percent: The POWer:REFLevel:PERCent command.

ripple property

ripple: PowerRipple

Return the POWer:RIPPle command.

Description
  • This command performs a vertical autoset for ripple measurements or sets the vertical offset to 0.
Usage
  • Using the .write(value) method will send the POWer:RIPPle value command.
SCPI Syntax
- POWer:RIPPle {VERTAUTOset|VERTDEFault}
Info
  • VERTAUTOset automatically scales the source waveform to optimize ripple measurements.
  • VERTDEFault sets the vertical offset of the source waveform to 0 volts (for voltage source) or 0 amperes (for current source).
Sub-properties
  • .results: The POWer:RIPPle:RESults command tree.
  • .source: The POWer:RIPPle:SOUrce command.

soa property

soa: PowerSoa

Return the POWer:SOA command tree.

Usage
  • Using the .query() method will send the POWer:SOA? query.
  • Using the .verify(value) method will send the POWer:SOA? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .linear: The POWer:SOA:LINear command tree.
  • .log: The POWer:SOA:LOG command tree.
  • .mask: The POWer:SOA:MASK command tree.
  • .plottype: The POWer:SOA:PLOTTYPe command.
  • .result: The POWer:SOA:RESult command tree.

statistics property

statistics: PowerStatistics

Return the POWer:STATIstics command.

Description
  • Clears all the accumulated statistics of all measurements. Performs the same function as the MEASUREMENT:STATISTICS command.
Usage
  • Using the .write(value) method will send the POWer:STATIstics value command.
SCPI Syntax
- POWer:STATIstics {RESET}
Info
  • RESET clears the measurement statistics.
Sub-properties
  • .mode: The POWer:STATIstics:MODe command.
  • .weighting: The POWer:STATIstics:WEIghting command.

swloss property

swloss: PowerSwloss

Return the POWer:SWLoss command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss? query.
  • Using the .verify(value) method will send the POWer:SWLoss? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .condcalcmethod: The POWer:SWLoss:CONDCALCmethod command.
  • .conduction: The POWer:SWLoss:CONDuction command tree.
  • .display: The POWer:SWLoss:DISplay command.
  • .gate: The POWer:SWLoss:GATe command tree.
  • .numcycles: The POWer:SWLoss:NUMCYCles command.
  • .rdson: The POWer:SWLoss:RDSon command.
  • .reflevel: The POWer:SWLoss:REFLevel command tree.
  • .toff: The POWer:SWLoss:TOFF command tree.
  • .ton: The POWer:SWLoss:TON command tree.
  • .total: The POWer:SWLoss:TOTal command tree.
  • .vcesat: The POWer:SWLoss:VCEsat command.

type property

type: PowerType

Return the POWer:TYPe command.

Description
  • This command specifies the power application measurement type.
Usage
  • Using the .query() method will send the POWer:TYPe? query.
  • Using the .verify(value) method will send the POWer:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:TYPe value command.
SCPI Syntax
- POWer:TYPe {NONe|QUALity|SWITCHingloss|SOA|HARMonics|RIPPle|MODULationanalysis|DESKew}
- POWer:TYPe?
Info
  • NONe Use to set the measurement type to None.
  • QUALity Use the power quality functions to obtain measurements and statistics about the general power quality in your test circuit.
  • SWITCHingloss Use the switching loss functions to obtain the power loss and energy loss across the acquired waveform, including turn-on loss, turn-off loss, conduction loss, and total loss. Typically, use these functions to characterize losses in power supply switching devices, as they switch on and off.
  • SOA Use the safe operating functions to obtain an X-Y display of the switching device-under-test’s voltage and current. Also use them to perform a mask test of the X-Y signal relative to the graphical X-Y description of the device specification table. The safe operating area is typically the voltage and current values that a semiconductor can operate without damaging itself.
  • HARMonics Use the harmonics functions to obtain the frequency spectrum of the source waveform and associated measurement values. Harmonic measurements can help one perform in-depth troubleshooting of power quality problems.
  • RIPPle Use the ripple functions to obtain measurements and statistics for the AC components of the acquired waveform. Ripples are often found on top of a large DC signal.
  • MODULationanalysis Use the modulation functions to obtain a trend plot of a measurement value across the acquired waveform. This is useful for showing the variations in the modulated switching signal.
  • DESKew Run the deskew procedure to match the delays through the probes. Different probes introduce different delays between the probe tip and the oscilloscope. Many oscilloscope users do not have to worry about this because they use the same type of probe on all channels. Power measurement users, however, frequently use both a voltage probe and a current probe. A current probe typically has a larger delay than a voltage probe, so setting deskew values becomes important.

voltagesource property

voltagesource: PowerVoltagesource

Return the POWer:VOLTAGESOurce command.

Description
  • This command specifies the voltage source for the power application.
Usage
  • Using the .query() method will send the POWer:VOLTAGESOurce? query.
  • Using the .verify(value) method will send the POWer:VOLTAGESOurce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:VOLTAGESOurce value command.
SCPI Syntax
- POWer:VOLTAGESOurce {CH<x>|REF<x>}
- POWer:VOLTAGESOurce?
Info
  • CH<x> sets the analog channel as the voltage source.
  • REF<x> sets the reference waveform as the voltage source.

PowerCurrentsource

Bases: SCPICmdWrite, SCPICmdRead

The POWer:CURRENTSOurce command.

Description
  • This command specifies the current source for the power application.
Usage
  • Using the .query() method will send the POWer:CURRENTSOurce? query.
  • Using the .verify(value) method will send the POWer:CURRENTSOurce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:CURRENTSOurce value command.
SCPI Syntax
- POWer:CURRENTSOurce {CH<x>|REF<x>}
- POWer:CURRENTSOurce?
Info
  • CH<x> sets an analog channel as the current source.
  • REF<x> sets a reference waveform as the current source.

PowerDisplay

Bases: SCPICmdWrite, SCPICmdRead

The POWer:DISplay command.

Description
  • This command controls whether or not to display the power test results. This is the equivalent to pressing the Test button and then selecting the power application. The same control is provided for each application.
Usage
  • Using the .query() method will send the POWer:DISplay? query.
  • Using the .verify(value) method will send the POWer:DISplay? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:DISplay value command.
SCPI Syntax
- POWer:DISplay {ON|OFF|1|0}
- POWer:DISplay?
Info
  • OFF or 0 turns off the display settings.
  • ON or 1 turns on the display settings.

PowerGatesource

Bases: SCPICmdWrite, SCPICmdRead

The POWer:GATESOurce command.

Description
  • This command specifies the gate source for the power application.
Usage
  • Using the .query() method will send the POWer:GATESOurce? query.
  • Using the .verify(value) method will send the POWer:GATESOurce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:GATESOurce value command.
SCPI Syntax
- POWer:GATESOurce {CH<x>|REF<x>|NONe}
- POWer:GATESOurce?
Info
  • CH<x> sets an analog channel as the gate source.
  • REF<x> sets a reference waveform as the gate source.
  • NONe is set when the gate source is not used in the application.

PowerGating

Bases: SCPICmdWrite, SCPICmdRead

The POWer:GATing command.

Description
  • This command specifies the power application gating.
Usage
  • Using the .query() method will send the POWer:GATing? query.
  • Using the .verify(value) method will send the POWer:GATing? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:GATing value command.
SCPI Syntax
- POWer:GATing {OFF|SCREen|CURSor}
- POWer:GATing?
Info
  • OFF turns off measurement gating (full record).
  • SCREen turns on gating, using the left and right edges of the screen.
  • CURSor limits measurements to the portion of the waveform between the vertical bar cursors, even if they are off screen.

PowerHarmonics

Bases: SCPICmdRead

The POWer:HARMonics command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics? query.
  • Using the .verify(value) method will send the POWer:HARMonics? query and raise an AssertionError if the returned value does not match value.
Properties
  • .display: The POWer:HARMonics:DISplay command tree.
  • .freqref: The POWer:HARMonics:FREQRef command.
  • .iec: The POWer:HARMonics:IEC command tree.
  • .mil: The POWer:HARMonics:MIL command tree.
  • .nr_harmonics: The POWer:HARMonics:NR_HARMonics command.
  • .results: The POWer:HARMonics:RESults command tree.
  • .source: The POWer:HARMonics:SOURce command.
  • .standard: The POWer:HARMonics:STANDard command.

display property

Return the POWer:HARMonics:DISplay command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:DISplay? query.
  • Using the .verify(value) method will send the POWer:HARMonics:DISplay? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .select: The POWer:HARMonics:DISplay:SELect command.
  • .type: The POWer:HARMonics:DISplay:TYPe command.

freqref property

Return the POWer:HARMonics:FREQRef command.

Description
  • This command specifies the frequency reference used when the harmonic standard is None.
Usage
  • Using the .query() method will send the POWer:HARMonics:FREQRef? query.
  • Using the .verify(value) method will send the POWer:HARMonics:FREQRef? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:FREQRef value command.
SCPI Syntax
- POWer:HARMonics:FREQRef {VOLTage|CURRent|HARMSOURce|FIXEDFREQuency}
- POWer:HARMonics:FREQRef?
Info
  • VOLTage to use a voltage waveform as the frequency reference.
  • CURRent to use a current waveform as the frequency reference.
  • HARMSOURce to use a harmonic source waveform as the frequency reference.
  • FIXEDFREQuency to use a fixed frequency value instead of a waveform for the frequency reference.
Sub-properties
  • .fixedfreqvalue: The POWer:HARMonics:FREQRef:FIXEDFREQValue command.

iec property

Return the POWer:HARMonics:IEC command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:IEC? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .class: The POWer:HARMonics:IEC:CLAss command.
  • .filter: The POWer:HARMonics:IEC:FILter command.
  • .fundamental: The POWer:HARMonics:IEC:FUNDamental command.
  • .grouping: The POWer:HARMonics:IEC:GROUPing command.
  • .inputpower: The POWer:HARMonics:IEC:INPUTPOWer command.
  • .linefrequency: The POWer:HARMonics:IEC:LINEFREQuency command.
  • .obsperiod: The POWer:HARMonics:IEC:OBSPERiod command.
  • .powerfactor: The POWer:HARMonics:IEC:POWERFACtor command.

mil property

Return the POWer:HARMonics:MIL command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:MIL? query.
  • Using the .verify(value) method will send the POWer:HARMonics:MIL? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .fundamental: The POWer:HARMonics:MIL:FUNDamental command tree.
  • .linefrequency: The POWer:HARMonics:MIL:LINEFREQuency command.
  • .powerlevel: The POWer:HARMonics:MIL:POWERLEVel command.

nr_harmonics property

Return the POWer:HARMonics:NR_HARMonics command.

Description
  • This command specifies the number of harmonics (value ranging from 20 to 400) when the harmonics standard is NONe.
Usage
  • Using the .query() method will send the POWer:HARMonics:NR_HARMonics? query.
  • Using the .verify(value) method will send the POWer:HARMonics:NR_HARMonics? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:NR_HARMonics value command.
SCPI Syntax
- POWer:HARMonics:NR_HARMonics <NR3>
- POWer:HARMonics:NR_HARMonics?
Info
  • <NR3> is a floating point number that specifies the number of harmonics.

results property

Return the POWer:HARMonics:RESults command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:RESults? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .har: The POWer:HARMonics:RESults:HAR<x> command tree.
  • .iec: The POWer:HARMonics:RESults:IEC command tree.
  • .passfail: The POWer:HARMonics:RESults:PASSFail command.
  • .rms: The POWer:HARMonics:RESults:RMS command.
  • .save: The POWer:HARMonics:RESults:SAVe command.
  • .thdf: The POWer:HARMonics:RESults:THDF command.
  • .thdr: The POWer:HARMonics:RESults:THDR command.

source property

Return the POWer:HARMonics:SOURce command.

Description
  • This command specifies the source waveform for harmonics tests. The voltage source waveform is specified using the POWER:VOLTAGESOURCE command and the current waveform is specified using the POWER:CURRENTSOURCE command.
Usage
  • Using the .query() method will send the POWer:HARMonics:SOURce? query.
  • Using the .verify(value) method will send the POWer:HARMonics:SOURce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:SOURce value command.
SCPI Syntax
- POWer:HARMonics:SOURce {VOLTage|CURRent}
- POWer:HARMonics:SOURce?
Info
  • VOLTage specifies voltage source waveform for harmonic tests.
  • CURRent specifies current source waveform for harmonic tests.

standard property

Return the POWer:HARMonics:STANDard command.

Description
  • This command specifies the standard for harmonics tests.
Usage
  • Using the .query() method will send the POWer:HARMonics:STANDard? query.
  • Using the .verify(value) method will send the POWer:HARMonics:STANDard? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:STANDard value command.
SCPI Syntax
- POWer:HARMonics:STANDard {NONe|IEC|MIL}
- POWer:HARMonics:STANDard?
Info
  • NONe sets no standard for harmonic tests.
  • IEC sets IEC 610003-2 standard for harmonic tests.
  • MIL sets MIL1399 standard for harmonic tests.

PowerHarmonicsDisplay

Bases: SCPICmdRead

The POWer:HARMonics:DISplay command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:DISplay? query.
  • Using the .verify(value) method will send the POWer:HARMonics:DISplay? query and raise an AssertionError if the returned value does not match value.
Properties
  • .select: The POWer:HARMonics:DISplay:SELect command.
  • .type: The POWer:HARMonics:DISplay:TYPe command.

select property

Return the POWer:HARMonics:DISplay:SELect command.

Description
  • This command specifies the harmonics to be displayed when the harmonics standard is NONe.
Usage
  • Using the .query() method will send the POWer:HARMonics:DISplay:SELect? query.
  • Using the .verify(value) method will send the POWer:HARMonics:DISplay:SELect? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:DISplay:SELect value command.
SCPI Syntax
- POWer:HARMonics:DISplay:SELect {ODD|EVEN|ALL}
- POWer:HARMonics:DISplay:SELect?
Info
  • ODD to display only odd harmonics.
  • EVEN to display only even harmonics.
  • ALL to display both odd and even harmonics.

type property

Return the POWer:HARMonics:DISplay:TYPe command.

Description
  • This command specifies the display type for harmonics tests.
Usage
  • Using the .query() method will send the POWer:HARMonics:DISplay:TYPe? query.
  • Using the .verify(value) method will send the POWer:HARMonics:DISplay:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:DISplay:TYPe value command.
SCPI Syntax
- POWer:HARMonics:DISplay:TYPe {GRAph|TABle}
- POWer:HARMonics:DISplay:TYPe?
Info
  • GRAph displays harmonic tests results in graphical format.
  • TABle displays harmonic tests results in tabular format.

PowerHarmonicsDisplaySelect

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:DISplay:SELect command.

Description
  • This command specifies the harmonics to be displayed when the harmonics standard is NONe.
Usage
  • Using the .query() method will send the POWer:HARMonics:DISplay:SELect? query.
  • Using the .verify(value) method will send the POWer:HARMonics:DISplay:SELect? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:DISplay:SELect value command.
SCPI Syntax
- POWer:HARMonics:DISplay:SELect {ODD|EVEN|ALL}
- POWer:HARMonics:DISplay:SELect?
Info
  • ODD to display only odd harmonics.
  • EVEN to display only even harmonics.
  • ALL to display both odd and even harmonics.

PowerHarmonicsDisplayType

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:DISplay:TYPe command.

Description
  • This command specifies the display type for harmonics tests.
Usage
  • Using the .query() method will send the POWer:HARMonics:DISplay:TYPe? query.
  • Using the .verify(value) method will send the POWer:HARMonics:DISplay:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:DISplay:TYPe value command.
SCPI Syntax
- POWer:HARMonics:DISplay:TYPe {GRAph|TABle}
- POWer:HARMonics:DISplay:TYPe?
Info
  • GRAph displays harmonic tests results in graphical format.
  • TABle displays harmonic tests results in tabular format.

PowerHarmonicsFreqref

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:FREQRef command.

Description
  • This command specifies the frequency reference used when the harmonic standard is None.
Usage
  • Using the .query() method will send the POWer:HARMonics:FREQRef? query.
  • Using the .verify(value) method will send the POWer:HARMonics:FREQRef? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:FREQRef value command.
SCPI Syntax
- POWer:HARMonics:FREQRef {VOLTage|CURRent|HARMSOURce|FIXEDFREQuency}
- POWer:HARMonics:FREQRef?
Info
  • VOLTage to use a voltage waveform as the frequency reference.
  • CURRent to use a current waveform as the frequency reference.
  • HARMSOURce to use a harmonic source waveform as the frequency reference.
  • FIXEDFREQuency to use a fixed frequency value instead of a waveform for the frequency reference.
Properties
  • .fixedfreqvalue: The POWer:HARMonics:FREQRef:FIXEDFREQValue command.

fixedfreqvalue property

Return the POWer:HARMonics:FREQRef:FIXEDFREQValue command.

Description
  • This command specifies the frequency value when the :FREQRef selection is FIXEDFREQuency.
Usage
  • Using the .query() method will send the POWer:HARMonics:FREQRef:FIXEDFREQValue? query.
  • Using the .verify(value) method will send the POWer:HARMonics:FREQRef:FIXEDFREQValue? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:FREQRef:FIXEDFREQValue value command.
SCPI Syntax
- POWer:HARMonics:FREQRef:FIXEDFREQValue <NR3>
- POWer:HARMonics:FREQRef:FIXEDFREQValue?
Info
  • <NR3> is a floating point number that specifies the fixed frequency value.

PowerHarmonicsFreqrefFixedfreqvalue

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:FREQRef:FIXEDFREQValue command.

Description
  • This command specifies the frequency value when the :FREQRef selection is FIXEDFREQuency.
Usage
  • Using the .query() method will send the POWer:HARMonics:FREQRef:FIXEDFREQValue? query.
  • Using the .verify(value) method will send the POWer:HARMonics:FREQRef:FIXEDFREQValue? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:FREQRef:FIXEDFREQValue value command.
SCPI Syntax
- POWer:HARMonics:FREQRef:FIXEDFREQValue <NR3>
- POWer:HARMonics:FREQRef:FIXEDFREQValue?
Info
  • <NR3> is a floating point number that specifies the fixed frequency value.

PowerHarmonicsIec

Bases: SCPICmdRead

The POWer:HARMonics:IEC command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:IEC? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC? query and raise an AssertionError if the returned value does not match value.
Properties
  • .class: The POWer:HARMonics:IEC:CLAss command.
  • .filter: The POWer:HARMonics:IEC:FILter command.
  • .fundamental: The POWer:HARMonics:IEC:FUNDamental command.
  • .grouping: The POWer:HARMonics:IEC:GROUPing command.
  • .inputpower: The POWer:HARMonics:IEC:INPUTPOWer command.
  • .linefrequency: The POWer:HARMonics:IEC:LINEFREQuency command.
  • .obsperiod: The POWer:HARMonics:IEC:OBSPERiod command.
  • .powerfactor: The POWer:HARMonics:IEC:POWERFACtor command.

class_ property

Return the POWer:HARMonics:IEC:CLAss command.

Description
  • This command specifies the equipment class for IEC harmonics.
Usage
  • Using the .query() method will send the POWer:HARMonics:IEC:CLAss? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC:CLAss? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:IEC:CLAss value command.
SCPI Syntax
- POWer:HARMonics:IEC:CLAss {A|B|C<x>|D}
- POWer:HARMonics:IEC:CLAss?
Info
  • A specifies Class A Equipment.
  • B specifies Class B Equipment.
  • C1 specifies Class C Equipment that use Table 1 limits of the IEC standard.
  • C2 specifies Class C Equipment that use Table 2 limits of the IEC standard.
  • C3 specifies Class C Equipment that use Table 3 limits of the IEC standard.
  • D specifies Class D Equipment.

filter property

Return the POWer:HARMonics:IEC:FILter command.

Description
  • This command specifies the enabled state for filtering of IEC harmonics.
Usage
  • Using the .query() method will send the POWer:HARMonics:IEC:FILter? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC:FILter? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:IEC:FILter value command.
SCPI Syntax
- POWer:HARMonics:IEC:FILter {ON|OFF|1|0}
- POWer:HARMonics:IEC:FILter?
Info
  • ON or 1 enables filtering of IEC harmonics.
  • OFF or 0 disables filtering of IEC harmonics.

fundamental property

Return the POWer:HARMonics:IEC:FUNDamental command.

Description
  • This command specifies the rated fundamental current for IEC harmonics. Valid values ranges from 0 to 16 in increments of 0.1. The unit is ampere.
Usage
  • Using the .query() method will send the POWer:HARMonics:IEC:FUNDamental? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC:FUNDamental? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:IEC:FUNDamental value command.
SCPI Syntax
- POWer:HARMonics:IEC:FUNDamental <NR3>
- POWer:HARMonics:IEC:FUNDamental?
Info
  • <NR3> is a floating point number that specifies the fundamental current in amperes.

grouping property

Return the POWer:HARMonics:IEC:GROUPing command.

Description
  • This command specifies the enabled state for grouping of IEC harmonics.
Usage
  • Using the .query() method will send the POWer:HARMonics:IEC:GROUPing? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC:GROUPing? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:IEC:GROUPing value command.
SCPI Syntax
- POWer:HARMonics:IEC:GROUPing {ON|OFF|1|0}
- POWer:HARMonics:IEC:GROUPing?
Info
  • ON or 1 enables grouping of IEC harmonics.
  • OFF or 0 disables grouping of IEC harmonics.

inputpower property

Return the POWer:HARMonics:IEC:INPUTPOWer command.

Description
  • This command specifies the class D rated input power for IEC harmonics.
Usage
  • Using the .query() method will send the POWer:HARMonics:IEC:INPUTPOWer? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC:INPUTPOWer? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:IEC:INPUTPOWer value command.
SCPI Syntax
- POWer:HARMonics:IEC:INPUTPOWer <NR3>
- POWer:HARMonics:IEC:INPUTPOWer?
Info
  • <NR3> is a floating point number that specifies the class D input power. Valid values ranges from 0 to 600 in increments of 10. The unit of measure is watt.

linefrequency property

Return the POWer:HARMonics:IEC:LINEFREQuency command.

Description
  • This command specifies the line frequency for the IEC standard.
Usage
  • Using the .query() method will send the POWer:HARMonics:IEC:LINEFREQuency? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC:LINEFREQuency? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:IEC:LINEFREQuency value command.
SCPI Syntax
- POWer:HARMonics:IEC:LINEFREQuency <NR1>
- POWer:HARMonics:IEC:LINEFREQuency?
Info
  • <NR1> is an unsigned integer that specifies the line frequency. The valid values are 50 and 60.

obsperiod property

Return the POWer:HARMonics:IEC:OBSPERiod command.

Description
  • This command specifies the IEC observation period.
Usage
  • Using the .query() method will send the POWer:HARMonics:IEC:OBSPERiod? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC:OBSPERiod? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:IEC:OBSPERiod value command.
SCPI Syntax
- POWer:HARMonics:IEC:OBSPERiod <NR3>
- POWer:HARMonics:IEC:OBSPERiod?
Info
  • <NR3> is a floating point number that specifies the IEC observation period. Valid values ranges from 0.2 to 10 s.

powerfactor property

Return the POWer:HARMonics:IEC:POWERFACtor command.

Description
  • This command specifies the rated power factor for IEC harmonics.
Usage
  • Using the .query() method will send the POWer:HARMonics:IEC:POWERFACtor? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC:POWERFACtor? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:IEC:POWERFACtor value command.
SCPI Syntax
- POWer:HARMonics:IEC:POWERFACtor <NR3>
- POWer:HARMonics:IEC:POWERFACtor?
Info
  • <NR3> is a floating point number that specifies the power factor. Valid values ranges from 0 to 1 in increments of 0.1.

PowerHarmonicsIecClass

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:IEC:CLAss command.

Description
  • This command specifies the equipment class for IEC harmonics.
Usage
  • Using the .query() method will send the POWer:HARMonics:IEC:CLAss? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC:CLAss? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:IEC:CLAss value command.
SCPI Syntax
- POWer:HARMonics:IEC:CLAss {A|B|C<x>|D}
- POWer:HARMonics:IEC:CLAss?
Info
  • A specifies Class A Equipment.
  • B specifies Class B Equipment.
  • C1 specifies Class C Equipment that use Table 1 limits of the IEC standard.
  • C2 specifies Class C Equipment that use Table 2 limits of the IEC standard.
  • C3 specifies Class C Equipment that use Table 3 limits of the IEC standard.
  • D specifies Class D Equipment.

PowerHarmonicsIecFilter

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:IEC:FILter command.

Description
  • This command specifies the enabled state for filtering of IEC harmonics.
Usage
  • Using the .query() method will send the POWer:HARMonics:IEC:FILter? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC:FILter? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:IEC:FILter value command.
SCPI Syntax
- POWer:HARMonics:IEC:FILter {ON|OFF|1|0}
- POWer:HARMonics:IEC:FILter?
Info
  • ON or 1 enables filtering of IEC harmonics.
  • OFF or 0 disables filtering of IEC harmonics.

PowerHarmonicsIecFundamental

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:IEC:FUNDamental command.

Description
  • This command specifies the rated fundamental current for IEC harmonics. Valid values ranges from 0 to 16 in increments of 0.1. The unit is ampere.
Usage
  • Using the .query() method will send the POWer:HARMonics:IEC:FUNDamental? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC:FUNDamental? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:IEC:FUNDamental value command.
SCPI Syntax
- POWer:HARMonics:IEC:FUNDamental <NR3>
- POWer:HARMonics:IEC:FUNDamental?
Info
  • <NR3> is a floating point number that specifies the fundamental current in amperes.

PowerHarmonicsIecGrouping

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:IEC:GROUPing command.

Description
  • This command specifies the enabled state for grouping of IEC harmonics.
Usage
  • Using the .query() method will send the POWer:HARMonics:IEC:GROUPing? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC:GROUPing? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:IEC:GROUPing value command.
SCPI Syntax
- POWer:HARMonics:IEC:GROUPing {ON|OFF|1|0}
- POWer:HARMonics:IEC:GROUPing?
Info
  • ON or 1 enables grouping of IEC harmonics.
  • OFF or 0 disables grouping of IEC harmonics.

PowerHarmonicsIecInputpower

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:IEC:INPUTPOWer command.

Description
  • This command specifies the class D rated input power for IEC harmonics.
Usage
  • Using the .query() method will send the POWer:HARMonics:IEC:INPUTPOWer? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC:INPUTPOWer? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:IEC:INPUTPOWer value command.
SCPI Syntax
- POWer:HARMonics:IEC:INPUTPOWer <NR3>
- POWer:HARMonics:IEC:INPUTPOWer?
Info
  • <NR3> is a floating point number that specifies the class D input power. Valid values ranges from 0 to 600 in increments of 10. The unit of measure is watt.

PowerHarmonicsIecLinefrequency

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:IEC:LINEFREQuency command.

Description
  • This command specifies the line frequency for the IEC standard.
Usage
  • Using the .query() method will send the POWer:HARMonics:IEC:LINEFREQuency? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC:LINEFREQuency? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:IEC:LINEFREQuency value command.
SCPI Syntax
- POWer:HARMonics:IEC:LINEFREQuency <NR1>
- POWer:HARMonics:IEC:LINEFREQuency?
Info
  • <NR1> is an unsigned integer that specifies the line frequency. The valid values are 50 and 60.

PowerHarmonicsIecObsperiod

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:IEC:OBSPERiod command.

Description
  • This command specifies the IEC observation period.
Usage
  • Using the .query() method will send the POWer:HARMonics:IEC:OBSPERiod? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC:OBSPERiod? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:IEC:OBSPERiod value command.
SCPI Syntax
- POWer:HARMonics:IEC:OBSPERiod <NR3>
- POWer:HARMonics:IEC:OBSPERiod?
Info
  • <NR3> is a floating point number that specifies the IEC observation period. Valid values ranges from 0.2 to 10 s.

PowerHarmonicsIecPowerfactor

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:IEC:POWERFACtor command.

Description
  • This command specifies the rated power factor for IEC harmonics.
Usage
  • Using the .query() method will send the POWer:HARMonics:IEC:POWERFACtor? query.
  • Using the .verify(value) method will send the POWer:HARMonics:IEC:POWERFACtor? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:IEC:POWERFACtor value command.
SCPI Syntax
- POWer:HARMonics:IEC:POWERFACtor <NR3>
- POWer:HARMonics:IEC:POWERFACtor?
Info
  • <NR3> is a floating point number that specifies the power factor. Valid values ranges from 0 to 1 in increments of 0.1.

PowerHarmonicsMil

Bases: SCPICmdRead

The POWer:HARMonics:MIL command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:MIL? query.
  • Using the .verify(value) method will send the POWer:HARMonics:MIL? query and raise an AssertionError if the returned value does not match value.
Properties
  • .fundamental: The POWer:HARMonics:MIL:FUNDamental command tree.
  • .linefrequency: The POWer:HARMonics:MIL:LINEFREQuency command.
  • .powerlevel: The POWer:HARMonics:MIL:POWERLEVel command.

fundamental property

Return the POWer:HARMonics:MIL:FUNDamental command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:MIL:FUNDamental? query.
  • Using the .verify(value) method will send the POWer:HARMonics:MIL:FUNDamental? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .calcmethod: The POWer:HARMonics:MIL:FUNDamental:CALCmethod command.
  • .user: The POWer:HARMonics:MIL:FUNDamental:USER command tree.

linefrequency property

Return the POWer:HARMonics:MIL:LINEFREQuency command.

Description
  • This command specifies the line frequency for MIL-STD-1399 Section 300A harmonics tests. Valid values are 60 or 400 Hz.
Usage
  • Using the .query() method will send the POWer:HARMonics:MIL:LINEFREQuency? query.
  • Using the .verify(value) method will send the POWer:HARMonics:MIL:LINEFREQuency? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:MIL:LINEFREQuency value command.
SCPI Syntax
- POWer:HARMonics:MIL:LINEFREQuency <NR1>
- POWer:HARMonics:MIL:LINEFREQuency?
Info
  • <NR1> is an unsigned integer that specifies the line frequency for the MIL standard.

powerlevel property

Return the POWer:HARMonics:MIL:POWERLEVel command.

Description
  • This command specifies the power level for calculating limits for MIL-STD-1399 Section 300A harmonics tests.
Usage
  • Using the .query() method will send the POWer:HARMonics:MIL:POWERLEVel? query.
  • Using the .verify(value) method will send the POWer:HARMonics:MIL:POWERLEVel? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:MIL:POWERLEVel value command.
SCPI Syntax
- POWer:HARMonics:MIL:POWERLEVel {LOW|HIGH}
- POWer:HARMonics:MIL:POWERLEVel?
Info
  • LOW specifies low power level for MIL-STD-1399 harmonics tests.
  • HIGH specifies high power level for MIL-STD-1399 harmonics tests.

PowerHarmonicsMilFundamental

Bases: SCPICmdRead

The POWer:HARMonics:MIL:FUNDamental command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:MIL:FUNDamental? query.
  • Using the .verify(value) method will send the POWer:HARMonics:MIL:FUNDamental? query and raise an AssertionError if the returned value does not match value.
Properties
  • .calcmethod: The POWer:HARMonics:MIL:FUNDamental:CALCmethod command.
  • .user: The POWer:HARMonics:MIL:FUNDamental:USER command tree.

calcmethod property

Return the POWer:HARMonics:MIL:FUNDamental:CALCmethod command.

Description
  • This command specifies the measurement method for the MIL harmonics fundamental current for use in calculating limits.
Usage
  • Using the .query() method will send the POWer:HARMonics:MIL:FUNDamental:CALCmethod? query.
  • Using the .verify(value) method will send the POWer:HARMonics:MIL:FUNDamental:CALCmethod? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:MIL:FUNDamental:CALCmethod value command.
SCPI Syntax
- POWer:HARMonics:MIL:FUNDamental:CALCmethod {MEAS|USER}
- POWer:HARMonics:MIL:FUNDamental:CALCmethod?
Info
  • MEAS specifies that the value of the fundamental current used in calculating limits is measured.
  • USER specifies that the value of the fundamental current used in calculated limits is user defined.

user property

Return the POWer:HARMonics:MIL:FUNDamental:USER command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:MIL:FUNDamental:USER? query.
  • Using the .verify(value) method will send the POWer:HARMonics:MIL:FUNDamental:USER? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .current: The POWer:HARMonics:MIL:FUNDamental:USER:CURrent command.

PowerHarmonicsMilFundamentalCalcmethod

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:MIL:FUNDamental:CALCmethod command.

Description
  • This command specifies the measurement method for the MIL harmonics fundamental current for use in calculating limits.
Usage
  • Using the .query() method will send the POWer:HARMonics:MIL:FUNDamental:CALCmethod? query.
  • Using the .verify(value) method will send the POWer:HARMonics:MIL:FUNDamental:CALCmethod? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:MIL:FUNDamental:CALCmethod value command.
SCPI Syntax
- POWer:HARMonics:MIL:FUNDamental:CALCmethod {MEAS|USER}
- POWer:HARMonics:MIL:FUNDamental:CALCmethod?
Info
  • MEAS specifies that the value of the fundamental current used in calculating limits is measured.
  • USER specifies that the value of the fundamental current used in calculated limits is user defined.

PowerHarmonicsMilFundamentalUser

Bases: SCPICmdRead

The POWer:HARMonics:MIL:FUNDamental:USER command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:MIL:FUNDamental:USER? query.
  • Using the .verify(value) method will send the POWer:HARMonics:MIL:FUNDamental:USER? query and raise an AssertionError if the returned value does not match value.
Properties
  • .current: The POWer:HARMonics:MIL:FUNDamental:USER:CURrent command.

current property

Return the POWer:HARMonics:MIL:FUNDamental:USER:CURrent command.

Description
  • This command specifies RMS amperes for USER CALCmethod.
Usage
  • Using the .query() method will send the POWer:HARMonics:MIL:FUNDamental:USER:CURrent? query.
  • Using the .verify(value) method will send the POWer:HARMonics:MIL:FUNDamental:USER:CURrent? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:MIL:FUNDamental:USER:CURrent value command.
SCPI Syntax
- POWer:HARMonics:MIL:FUNDamental:USER:CURrent <NR3>
- POWer:HARMonics:MIL:FUNDamental:USER:CURrent?
Info
  • <NR3> is a floating point number that specifies the current in amperes for USER CALCmethod.

PowerHarmonicsMilFundamentalUserCurrent

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:MIL:FUNDamental:USER:CURrent command.

Description
  • This command specifies RMS amperes for USER CALCmethod.
Usage
  • Using the .query() method will send the POWer:HARMonics:MIL:FUNDamental:USER:CURrent? query.
  • Using the .verify(value) method will send the POWer:HARMonics:MIL:FUNDamental:USER:CURrent? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:MIL:FUNDamental:USER:CURrent value command.
SCPI Syntax
- POWer:HARMonics:MIL:FUNDamental:USER:CURrent <NR3>
- POWer:HARMonics:MIL:FUNDamental:USER:CURrent?
Info
  • <NR3> is a floating point number that specifies the current in amperes for USER CALCmethod.

PowerHarmonicsMilLinefrequency

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:MIL:LINEFREQuency command.

Description
  • This command specifies the line frequency for MIL-STD-1399 Section 300A harmonics tests. Valid values are 60 or 400 Hz.
Usage
  • Using the .query() method will send the POWer:HARMonics:MIL:LINEFREQuency? query.
  • Using the .verify(value) method will send the POWer:HARMonics:MIL:LINEFREQuency? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:MIL:LINEFREQuency value command.
SCPI Syntax
- POWer:HARMonics:MIL:LINEFREQuency <NR1>
- POWer:HARMonics:MIL:LINEFREQuency?
Info
  • <NR1> is an unsigned integer that specifies the line frequency for the MIL standard.

PowerHarmonicsMilPowerlevel

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:MIL:POWERLEVel command.

Description
  • This command specifies the power level for calculating limits for MIL-STD-1399 Section 300A harmonics tests.
Usage
  • Using the .query() method will send the POWer:HARMonics:MIL:POWERLEVel? query.
  • Using the .verify(value) method will send the POWer:HARMonics:MIL:POWERLEVel? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:MIL:POWERLEVel value command.
SCPI Syntax
- POWer:HARMonics:MIL:POWERLEVel {LOW|HIGH}
- POWer:HARMonics:MIL:POWERLEVel?
Info
  • LOW specifies low power level for MIL-STD-1399 harmonics tests.
  • HIGH specifies high power level for MIL-STD-1399 harmonics tests.

PowerHarmonicsNrHarmonics

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:NR_HARMonics command.

Description
  • This command specifies the number of harmonics (value ranging from 20 to 400) when the harmonics standard is NONe.
Usage
  • Using the .query() method will send the POWer:HARMonics:NR_HARMonics? query.
  • Using the .verify(value) method will send the POWer:HARMonics:NR_HARMonics? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:NR_HARMonics value command.
SCPI Syntax
- POWer:HARMonics:NR_HARMonics <NR3>
- POWer:HARMonics:NR_HARMonics?
Info
  • <NR3> is a floating point number that specifies the number of harmonics.

PowerHarmonicsResults

Bases: SCPICmdRead

The POWer:HARMonics:RESults command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:RESults? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults? query and raise an AssertionError if the returned value does not match value.
Properties
  • .har: The POWer:HARMonics:RESults:HAR<x> command tree.
  • .iec: The POWer:HARMonics:RESults:IEC command tree.
  • .passfail: The POWer:HARMonics:RESults:PASSFail command.
  • .rms: The POWer:HARMonics:RESults:RMS command.
  • .save: The POWer:HARMonics:RESults:SAVe command.
  • .thdf: The POWer:HARMonics:RESults:THDF command.
  • .thdr: The POWer:HARMonics:RESults:THDR command.

har property

Return the POWer:HARMonics:RESults:HAR<x> command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .frequency: The POWer:HARMonics:RESults:HAR<x>:FREQuency command.
  • .iecmax: The POWer:HARMonics:RESults:HAR<x>:IECMAX command.
  • .limit: The POWer:HARMonics:RESults:HAR<x>:LIMit command.
  • .phase: The POWer:HARMonics:RESults:HAR<x>:PHASe command.
  • .rms: The POWer:HARMonics:RESults:HAR<x>:RMS command tree.
  • .test: The POWer:HARMonics:RESults:HAR<x>:TEST command tree.

iec property

Return the POWer:HARMonics:RESults:IEC command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:IEC? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:IEC? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .fundamental: The POWer:HARMonics:RESults:IEC:FUNDamental command.
  • .harm3alternate: The POWer:HARMonics:RESults:IEC:HARM3ALTernate command.
  • .harm5alternate: The POWer:HARMonics:RESults:IEC:HARM5ALTernate command.
  • .pohc: The POWer:HARMonics:RESults:IEC:POHC command.
  • .pohl: The POWer:HARMonics:RESults:IEC:POHL command.
  • .powerfactor: The POWer:HARMonics:RESults:IEC:POWERFactor command.
  • .power: The POWer:HARMonics:RESults:IEC:POWer command.

passfail property

Return the POWer:HARMonics:RESults:PASSFail command.

Description
  • Returns the overall harmonics test result: PASS, FAIL or NA.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:PASSFail? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:PASSFail? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:PASSFail?

rms property

Return the POWer:HARMonics:RESults:RMS command.

Description
  • Returns the root mean square value of the harmonics source waveform.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:RMS? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:RMS? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:RMS?

save property

Return the POWer:HARMonics:RESults:SAVe command.

Description
  • Saves the harmonic results to the specified file in CSV format.
Usage
  • Using the .write(value) method will send the POWer:HARMonics:RESults:SAVe value command.
SCPI Syntax
- POWer:HARMonics:RESults:SAVe <String>

thdf property

Return the POWer:HARMonics:RESults:THDF command.

Description
  • Returns the Total Harmonic Distortion (THD) in percentage, measured as a ratio to the RMS value of the fundamental component of the source waveform.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:THDF? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:THDF? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:THDF?

thdr property

Return the POWer:HARMonics:RESults:THDR command.

Description
  • Returns the Total Harmonic Distortion (THD) in percentage, measured as a ratio to the RMS value of the source waveform.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:THDR? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:THDR? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:THDR?

PowerHarmonicsResultsHarItem

Bases: ValidatedDynamicNumberCmd, SCPICmdRead

The POWer:HARMonics:RESults:HAR<x> command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>? query and raise an AssertionError if the returned value does not match value.
Properties
  • .frequency: The POWer:HARMonics:RESults:HAR<x>:FREQuency command.
  • .iecmax: The POWer:HARMonics:RESults:HAR<x>:IECMAX command.
  • .limit: The POWer:HARMonics:RESults:HAR<x>:LIMit command.
  • .phase: The POWer:HARMonics:RESults:HAR<x>:PHASe command.
  • .rms: The POWer:HARMonics:RESults:HAR<x>:RMS command tree.
  • .test: The POWer:HARMonics:RESults:HAR<x>:TEST command tree.

frequency property

Return the POWer:HARMonics:RESults:HAR<x>:FREQuency command.

Description
  • Returns the frequency of the harmonic.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:FREQuency? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:FREQuency? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:FREQuency?

iecmax property

Return the POWer:HARMonics:RESults:HAR<x>:IECMAX command.

Description
  • The IEC Standard specifies harmonics measurements to be computed in time windows, with each time window being nominally 200 ms. This returns the maximum of the RMS magnitude of the harmonic, computed across successive 200 ms time windows within an observation period entered by the user.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:IECMAX? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:IECMAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:IECMAX?

limit property

Return the POWer:HARMonics:RESults:HAR<x>:LIMit command.

Description
  • The IEC and MIL standards specify a limit for each harmonic magnitude. Returns the limit in absolute units, or as a percentage of the fundamental as specified by the standard. IEC Class C (Table 2) and MIL standards specify the limit as a percentage of the fundamental.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:LIMit? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:LIMit? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:LIMit?

phase property

Return the POWer:HARMonics:RESults:HAR<x>:PHASe command.

Description
  • Returns the phase of the harmonic in degrees. The phase is measured relative to the zero-crossing of the reference waveform. When there is no reference waveform, the phase is relative to the fundamental component.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:PHASe? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:PHASe? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:PHASe?

rms property

Return the POWer:HARMonics:RESults:HAR<x>:RMS command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:RMS? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:RMS? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .absolute: The POWer:HARMonics:RESults:HAR<x>:RMS:ABSolute command.
  • .percent: The POWer:HARMonics:RESults:HAR<x>:RMS:PERCent command.

test property

Return the POWer:HARMonics:RESults:HAR<x>:TEST command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:TEST? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:TEST? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .iec: The POWer:HARMonics:RESults:HAR<x>:TEST:IEC command tree.
  • .mil: The POWer:HARMonics:RESults:HAR<x>:TEST:MIL command tree.

PowerHarmonicsResultsHarItemFrequency

Bases: SCPICmdRead

The POWer:HARMonics:RESults:HAR<x>:FREQuency command.

Description
  • Returns the frequency of the harmonic.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:FREQuency? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:FREQuency? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:FREQuency?

PowerHarmonicsResultsHarItemIecmax

Bases: SCPICmdRead

The POWer:HARMonics:RESults:HAR<x>:IECMAX command.

Description
  • The IEC Standard specifies harmonics measurements to be computed in time windows, with each time window being nominally 200 ms. This returns the maximum of the RMS magnitude of the harmonic, computed across successive 200 ms time windows within an observation period entered by the user.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:IECMAX? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:IECMAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:IECMAX?

PowerHarmonicsResultsHarItemLimit

Bases: SCPICmdRead

The POWer:HARMonics:RESults:HAR<x>:LIMit command.

Description
  • The IEC and MIL standards specify a limit for each harmonic magnitude. Returns the limit in absolute units, or as a percentage of the fundamental as specified by the standard. IEC Class C (Table 2) and MIL standards specify the limit as a percentage of the fundamental.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:LIMit? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:LIMit? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:LIMit?

PowerHarmonicsResultsHarItemPhase

Bases: SCPICmdRead

The POWer:HARMonics:RESults:HAR<x>:PHASe command.

Description
  • Returns the phase of the harmonic in degrees. The phase is measured relative to the zero-crossing of the reference waveform. When there is no reference waveform, the phase is relative to the fundamental component.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:PHASe? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:PHASe? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:PHASe?

PowerHarmonicsResultsHarItemRms

Bases: SCPICmdRead

The POWer:HARMonics:RESults:HAR<x>:RMS command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:RMS? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:RMS? query and raise an AssertionError if the returned value does not match value.
Properties
  • .absolute: The POWer:HARMonics:RESults:HAR<x>:RMS:ABSolute command.
  • .percent: The POWer:HARMonics:RESults:HAR<x>:RMS:PERCent command.

absolute property

Return the POWer:HARMonics:RESults:HAR<x>:RMS:ABSolute command.

Description
  • Returns the RMS magnitude of the harmonic in absolute units.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:RMS:ABSolute? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:RMS:ABSolute? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:RMS:ABSolute?

percent property

Return the POWer:HARMonics:RESults:HAR<x>:RMS:PERCent command.

Description
  • Returns the RMS magnitude of the harmonic expressed as a percentage of the fundamental.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:RMS:PERCent? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:RMS:PERCent? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:RMS:PERCent?

PowerHarmonicsResultsHarItemRmsAbsolute

Bases: SCPICmdRead

The POWer:HARMonics:RESults:HAR<x>:RMS:ABSolute command.

Description
  • Returns the RMS magnitude of the harmonic in absolute units.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:RMS:ABSolute? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:RMS:ABSolute? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:RMS:ABSolute?

PowerHarmonicsResultsHarItemRmsPercent

Bases: SCPICmdRead

The POWer:HARMonics:RESults:HAR<x>:RMS:PERCent command.

Description
  • Returns the RMS magnitude of the harmonic expressed as a percentage of the fundamental.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:RMS:PERCent? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:RMS:PERCent? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:RMS:PERCent?

PowerHarmonicsResultsHarItemTest

Bases: SCPICmdRead

The POWer:HARMonics:RESults:HAR<x>:TEST command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:TEST? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:TEST? query and raise an AssertionError if the returned value does not match value.
Properties
  • .iec: The POWer:HARMonics:RESults:HAR<x>:TEST:IEC command tree.
  • .mil: The POWer:HARMonics:RESults:HAR<x>:TEST:MIL command tree.

iec property

Return the POWer:HARMonics:RESults:HAR<x>:TEST:IEC command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:TEST:IEC? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:TEST:IEC? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .classalimit: The POWer:HARMonics:RESults:HAR<x>:TEST:IEC:CLASSALIMit command.
  • .normal: The POWer:HARMonics:RESults:HAR<x>:TEST:IEC:NORMAL command.
  • .pohclimit: The POWer:HARMonics:RESults:HAR<x>:TEST:IEC:POHCLIMit command.

mil property

Return the POWer:HARMonics:RESults:HAR<x>:TEST:MIL command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:TEST:MIL? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:TEST:MIL? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .normal: The POWer:HARMonics:RESults:HAR<x>:TEST:MIL:NORMAL command.

PowerHarmonicsResultsHarItemTestIec

Bases: SCPICmdRead

The POWer:HARMonics:RESults:HAR<x>:TEST:IEC command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:TEST:IEC? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:TEST:IEC? query and raise an AssertionError if the returned value does not match value.
Properties
  • .classalimit: The POWer:HARMonics:RESults:HAR<x>:TEST:IEC:CLASSALIMit command.
  • .normal: The POWer:HARMonics:RESults:HAR<x>:TEST:IEC:NORMAL command.
  • .pohclimit: The POWer:HARMonics:RESults:HAR<x>:TEST:IEC:POHCLIMit command.

classalimit property

Return the POWer:HARMonics:RESults:HAR<x>:TEST:IEC:CLASSALIMit command.

Description
  • Specifies if the IEC Class A higher harmonic limit (and conditions) are met.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:TEST:IEC:CLASSALIMit? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:TEST:IEC:CLASSALIMit? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:TEST:IEC:CLASSALIMit?

normal property

Return the POWer:HARMonics:RESults:HAR<x>:TEST:IEC:NORMAL command.

Description
  • Specifies if the Normal IEC harmonic limits are met.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:TEST:IEC:NORMAL? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:TEST:IEC:NORMAL? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:TEST:IEC:NORMAL?

pohclimit property

Return the POWer:HARMonics:RESults:HAR<x>:TEST:IEC:POHCLIMit command.

Description
  • Specifies if the higher harmonic limit (and conditions) for the 21st and higher order odd harmonics are met.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:TEST:IEC:POHCLIMit? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:TEST:IEC:POHCLIMit? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:TEST:IEC:POHCLIMit?

PowerHarmonicsResultsHarItemTestIecClassalimit

Bases: SCPICmdRead

The POWer:HARMonics:RESults:HAR<x>:TEST:IEC:CLASSALIMit command.

Description
  • Specifies if the IEC Class A higher harmonic limit (and conditions) are met.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:TEST:IEC:CLASSALIMit? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:TEST:IEC:CLASSALIMit? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:TEST:IEC:CLASSALIMit?

PowerHarmonicsResultsHarItemTestIecNormal

Bases: SCPICmdRead

The POWer:HARMonics:RESults:HAR<x>:TEST:IEC:NORMAL command.

Description
  • Specifies if the Normal IEC harmonic limits are met.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:TEST:IEC:NORMAL? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:TEST:IEC:NORMAL? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:TEST:IEC:NORMAL?

PowerHarmonicsResultsHarItemTestIecPohclimit

Bases: SCPICmdRead

The POWer:HARMonics:RESults:HAR<x>:TEST:IEC:POHCLIMit command.

Description
  • Specifies if the higher harmonic limit (and conditions) for the 21st and higher order odd harmonics are met.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:TEST:IEC:POHCLIMit? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:TEST:IEC:POHCLIMit? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:TEST:IEC:POHCLIMit?

PowerHarmonicsResultsHarItemTestMil

Bases: SCPICmdRead

The POWer:HARMonics:RESults:HAR<x>:TEST:MIL command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:TEST:MIL? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:TEST:MIL? query and raise an AssertionError if the returned value does not match value.
Properties
  • .normal: The POWer:HARMonics:RESults:HAR<x>:TEST:MIL:NORMAL command.

normal property

Return the POWer:HARMonics:RESults:HAR<x>:TEST:MIL:NORMAL command.

Description
  • Returns the test result for the specified harmonic for the MIL-STD-1399 Section 300A testing standard. This query is analogous to that for the IEC 61000-3-2 standard POWER:HARMONICS:RESULTS:HAR1400:TEST:IEC:NORMAL command.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:TEST:MIL:NORMAL? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:TEST:MIL:NORMAL? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:TEST:MIL:NORMAL?

PowerHarmonicsResultsHarItemTestMilNormal

Bases: SCPICmdRead

The POWer:HARMonics:RESults:HAR<x>:TEST:MIL:NORMAL command.

Description
  • Returns the test result for the specified harmonic for the MIL-STD-1399 Section 300A testing standard. This query is analogous to that for the IEC 61000-3-2 standard POWER:HARMONICS:RESULTS:HAR1400:TEST:IEC:NORMAL command.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:HAR<x>:TEST:MIL:NORMAL? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:HAR<x>:TEST:MIL:NORMAL? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:HAR<x>:TEST:MIL:NORMAL?

PowerHarmonicsResultsIec

Bases: SCPICmdRead

The POWer:HARMonics:RESults:IEC command tree.

Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:IEC? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:IEC? query and raise an AssertionError if the returned value does not match value.
Properties
  • .fundamental: The POWer:HARMonics:RESults:IEC:FUNDamental command.
  • .harm3alternate: The POWer:HARMonics:RESults:IEC:HARM3ALTernate command.
  • .harm5alternate: The POWer:HARMonics:RESults:IEC:HARM5ALTernate command.
  • .pohc: The POWer:HARMonics:RESults:IEC:POHC command.
  • .pohl: The POWer:HARMonics:RESults:IEC:POHL command.
  • .powerfactor: The POWer:HARMonics:RESults:IEC:POWERFactor command.
  • .power: The POWer:HARMonics:RESults:IEC:POWer command.

fundamental property

Return the POWer:HARMonics:RESults:IEC:FUNDamental command.

Description
  • Returns the IEC fundamental current used in calculating limits.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:IEC:FUNDamental? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:IEC:FUNDamental? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:IEC:FUNDamental?

harm3alternate property

Return the POWer:HARMonics:RESults:IEC:HARM3ALTernate command.

Description
  • Returns the IEC harmonics test result for the 3rd harmonic.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:IEC:HARM3ALTernate? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:IEC:HARM3ALTernate? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:IEC:HARM3ALTernate?

harm5alternate property

Return the POWer:HARMonics:RESults:IEC:HARM5ALTernate command.

Description
  • Returns the overall harmonics test result for the 5th harmonic.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:IEC:HARM5ALTernate? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:IEC:HARM5ALTernate? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:IEC:HARM5ALTernate?

pohc property

Return the POWer:HARMonics:RESults:IEC:POHC command.

Description
  • Returns the IEC POHC measurement.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:IEC:POHC? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:IEC:POHC? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:IEC:POHC?

pohl property

Return the POWer:HARMonics:RESults:IEC:POHL command.

Description
  • Returns the IEC POHL measurement.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:IEC:POHL? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:IEC:POHL? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:IEC:POHL?

power property

Return the POWer:HARMonics:RESults:IEC:POWer command.

Description
  • Returns the IEC input power that is used to calculate limits.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:IEC:POWer? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:IEC:POWer? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:IEC:POWer?

powerfactor property

Return the POWer:HARMonics:RESults:IEC:POWERFactor command.

Description
  • Returns the IEC power factor measurement.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:IEC:POWERFactor? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:IEC:POWERFactor? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:IEC:POWERFactor?

PowerHarmonicsResultsIecFundamental

Bases: SCPICmdRead

The POWer:HARMonics:RESults:IEC:FUNDamental command.

Description
  • Returns the IEC fundamental current used in calculating limits.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:IEC:FUNDamental? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:IEC:FUNDamental? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:IEC:FUNDamental?

PowerHarmonicsResultsIecHarm3alternate

Bases: SCPICmdRead

The POWer:HARMonics:RESults:IEC:HARM3ALTernate command.

Description
  • Returns the IEC harmonics test result for the 3rd harmonic.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:IEC:HARM3ALTernate? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:IEC:HARM3ALTernate? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:IEC:HARM3ALTernate?

PowerHarmonicsResultsIecHarm5alternate

Bases: SCPICmdRead

The POWer:HARMonics:RESults:IEC:HARM5ALTernate command.

Description
  • Returns the overall harmonics test result for the 5th harmonic.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:IEC:HARM5ALTernate? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:IEC:HARM5ALTernate? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:IEC:HARM5ALTernate?

PowerHarmonicsResultsIecPohc

Bases: SCPICmdRead

The POWer:HARMonics:RESults:IEC:POHC command.

Description
  • Returns the IEC POHC measurement.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:IEC:POHC? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:IEC:POHC? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:IEC:POHC?

PowerHarmonicsResultsIecPohl

Bases: SCPICmdRead

The POWer:HARMonics:RESults:IEC:POHL command.

Description
  • Returns the IEC POHL measurement.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:IEC:POHL? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:IEC:POHL? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:IEC:POHL?

PowerHarmonicsResultsIecPower

Bases: SCPICmdRead

The POWer:HARMonics:RESults:IEC:POWer command.

Description
  • Returns the IEC input power that is used to calculate limits.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:IEC:POWer? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:IEC:POWer? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:IEC:POWer?

PowerHarmonicsResultsIecPowerfactor

Bases: SCPICmdRead

The POWer:HARMonics:RESults:IEC:POWERFactor command.

Description
  • Returns the IEC power factor measurement.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:IEC:POWERFactor? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:IEC:POWERFactor? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:IEC:POWERFactor?

PowerHarmonicsResultsPassfail

Bases: SCPICmdRead

The POWer:HARMonics:RESults:PASSFail command.

Description
  • Returns the overall harmonics test result: PASS, FAIL or NA.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:PASSFail? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:PASSFail? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:PASSFail?

PowerHarmonicsResultsRms

Bases: SCPICmdRead

The POWer:HARMonics:RESults:RMS command.

Description
  • Returns the root mean square value of the harmonics source waveform.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:RMS? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:RMS? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:RMS?

PowerHarmonicsResultsSave

Bases: SCPICmdWrite

The POWer:HARMonics:RESults:SAVe command.

Description
  • Saves the harmonic results to the specified file in CSV format.
Usage
  • Using the .write(value) method will send the POWer:HARMonics:RESults:SAVe value command.
SCPI Syntax
- POWer:HARMonics:RESults:SAVe <String>

PowerHarmonicsResultsThdf

Bases: SCPICmdRead

The POWer:HARMonics:RESults:THDF command.

Description
  • Returns the Total Harmonic Distortion (THD) in percentage, measured as a ratio to the RMS value of the fundamental component of the source waveform.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:THDF? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:THDF? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:THDF?

PowerHarmonicsResultsThdr

Bases: SCPICmdRead

The POWer:HARMonics:RESults:THDR command.

Description
  • Returns the Total Harmonic Distortion (THD) in percentage, measured as a ratio to the RMS value of the source waveform.
Usage
  • Using the .query() method will send the POWer:HARMonics:RESults:THDR? query.
  • Using the .verify(value) method will send the POWer:HARMonics:RESults:THDR? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:HARMonics:RESults:THDR?

PowerHarmonicsSource

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:SOURce command.

Description
  • This command specifies the source waveform for harmonics tests. The voltage source waveform is specified using the POWER:VOLTAGESOURCE command and the current waveform is specified using the POWER:CURRENTSOURCE command.
Usage
  • Using the .query() method will send the POWer:HARMonics:SOURce? query.
  • Using the .verify(value) method will send the POWer:HARMonics:SOURce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:SOURce value command.
SCPI Syntax
- POWer:HARMonics:SOURce {VOLTage|CURRent}
- POWer:HARMonics:SOURce?
Info
  • VOLTage specifies voltage source waveform for harmonic tests.
  • CURRent specifies current source waveform for harmonic tests.

PowerHarmonicsStandard

Bases: SCPICmdWrite, SCPICmdRead

The POWer:HARMonics:STANDard command.

Description
  • This command specifies the standard for harmonics tests.
Usage
  • Using the .query() method will send the POWer:HARMonics:STANDard? query.
  • Using the .verify(value) method will send the POWer:HARMonics:STANDard? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:HARMonics:STANDard value command.
SCPI Syntax
- POWer:HARMonics:STANDard {NONe|IEC|MIL}
- POWer:HARMonics:STANDard?
Info
  • NONe sets no standard for harmonic tests.
  • IEC sets IEC 610003-2 standard for harmonic tests.
  • MIL sets MIL1399 standard for harmonic tests.

PowerIndicators

Bases: SCPICmdWrite, SCPICmdRead

The POWer:INDICators command.

Description
  • This command specifies the state of the measurement indicators for the power application.
Usage
  • Using the .query() method will send the POWer:INDICators? query.
  • Using the .verify(value) method will send the POWer:INDICators? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:INDICators value command.
SCPI Syntax
- POWer:INDICators {ON|OFF|1|0}
- POWer:INDICators?
Info
  • OFF or 0 turns off the measurement indicators.
  • ON or 1 turns on the measurement indicators.

PowerModulation

Bases: SCPICmdRead

The POWer:MODulation command tree.

Usage
  • Using the .query() method will send the POWer:MODulation? query.
  • Using the .verify(value) method will send the POWer:MODulation? query and raise an AssertionError if the returned value does not match value.
Properties
  • .source: The POWer:MODulation:SOUrce command.
  • .type: The POWer:MODulation:TYPe command.

source property

Return the POWer:MODulation:SOUrce command.

Description
  • This command specifies the source waveform for modulation tests. The voltage source waveform is specified using the POWER:VOLTAGESOURCE command and the current waveform is specified using the POWER:CURRENTSOURCE command.
Usage
  • Using the .query() method will send the POWer:MODulation:SOUrce? query.
  • Using the .verify(value) method will send the POWer:MODulation:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:MODulation:SOUrce value command.
SCPI Syntax
- POWer:MODulation:SOUrce {VOLTage|CURRent}
- POWer:MODulation:SOUrce?
Info
  • VOLTage specifies voltage source waveform for modulation tests.
  • CURRent specifies current source waveform for modulation tests.

type property

Return the POWer:MODulation:TYPe command.

Description
  • This command specifies the modulation type.
Usage
  • Using the .query() method will send the POWer:MODulation:TYPe? query.
  • Using the .verify(value) method will send the POWer:MODulation:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:MODulation:TYPe value command.
SCPI Syntax
- POWer:MODulation:TYPe {PWIdth|NWIdth|PERIod|PDUty|NDUty|FREQuency}
- POWer:MODulation:TYPe?
Info
  • PWIdth (positive width) is the distance (time) between the middle reference (default = 50%) amplitude points of a positive pulse. The measurement is made on all the cycles in the waveform or gated region.
  • NWIdth (negative width) measurement is the distance (time) between the middle reference (default = 50%) amplitude points of a negative pulse. The measurement is made on all the cycles in the waveform or gated region.
  • PERIod is the time required to complete the first cycle in a waveform or the gated region. The time is measured between the mid reference (default being 50%) amplitude points of the waveform.
  • PDUty (positive duty cycle) is the ratio of the positive pulse width to the signal period, expressed as a percentage. It is measured on all the cycles in the waveform or gated region.
  • Positive Duty Cycle = ((Positive Width)/Period) × 100%
  • NDUty (negative duty cycle) is the ratio of the negative pulse width to the signal period, expressed as a percentage. The duty cycle is measured on all the cycles in the waveform or gated region.
  • Negative Duty Cycle = ((Negative Width) / Period) × 100%
  • FREQuency measures all the cycles in the waveform or gated region. Frequency is the reciprocal of the period and is measured in hertz (Hz), where 1 Hz = 1 cycle per second.
  • Frequency = 1 / Period

PowerModulationSource

Bases: SCPICmdWrite, SCPICmdRead

The POWer:MODulation:SOUrce command.

Description
  • This command specifies the source waveform for modulation tests. The voltage source waveform is specified using the POWER:VOLTAGESOURCE command and the current waveform is specified using the POWER:CURRENTSOURCE command.
Usage
  • Using the .query() method will send the POWer:MODulation:SOUrce? query.
  • Using the .verify(value) method will send the POWer:MODulation:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:MODulation:SOUrce value command.
SCPI Syntax
- POWer:MODulation:SOUrce {VOLTage|CURRent}
- POWer:MODulation:SOUrce?
Info
  • VOLTage specifies voltage source waveform for modulation tests.
  • CURRent specifies current source waveform for modulation tests.

PowerModulationType

Bases: SCPICmdWrite, SCPICmdRead

The POWer:MODulation:TYPe command.

Description
  • This command specifies the modulation type.
Usage
  • Using the .query() method will send the POWer:MODulation:TYPe? query.
  • Using the .verify(value) method will send the POWer:MODulation:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:MODulation:TYPe value command.
SCPI Syntax
- POWer:MODulation:TYPe {PWIdth|NWIdth|PERIod|PDUty|NDUty|FREQuency}
- POWer:MODulation:TYPe?
Info
  • PWIdth (positive width) is the distance (time) between the middle reference (default = 50%) amplitude points of a positive pulse. The measurement is made on all the cycles in the waveform or gated region.
  • NWIdth (negative width) measurement is the distance (time) between the middle reference (default = 50%) amplitude points of a negative pulse. The measurement is made on all the cycles in the waveform or gated region.
  • PERIod is the time required to complete the first cycle in a waveform or the gated region. The time is measured between the mid reference (default being 50%) amplitude points of the waveform.
  • PDUty (positive duty cycle) is the ratio of the positive pulse width to the signal period, expressed as a percentage. It is measured on all the cycles in the waveform or gated region.
  • Positive Duty Cycle = ((Positive Width)/Period) × 100%
  • NDUty (negative duty cycle) is the ratio of the negative pulse width to the signal period, expressed as a percentage. The duty cycle is measured on all the cycles in the waveform or gated region.
  • Negative Duty Cycle = ((Negative Width) / Period) × 100%
  • FREQuency measures all the cycles in the waveform or gated region. Frequency is the reciprocal of the period and is measured in hertz (Hz), where 1 Hz = 1 cycle per second.
  • Frequency = 1 / Period

PowerQuality

Bases: SCPICmdRead

The POWer:QUALity command tree.

Usage
  • Using the .query() method will send the POWer:QUALity? query.
  • Using the .verify(value) method will send the POWer:QUALity? query and raise an AssertionError if the returned value does not match value.
Properties
  • .apppwr: The POWer:QUALity:APPpwr command.
  • .display: The POWer:QUALity:DISplay command tree.
  • .freqreference: The POWer:QUALity:FREQREFerence command.
  • .frequency: The POWer:QUALity:FREQuency command.
  • .icrestfactor: The POWer:QUALity:ICRESTfactor command.
  • .irms: The POWer:QUALity:IRMS command.
  • .phaseangle: The POWer:QUALity:PHASEangle command.
  • .powerfactor: The POWer:QUALity:POWERFACtor command.
  • .reactpwr: The POWer:QUALity:REACTpwr command.
  • .truepwr: The POWer:QUALity:TRUEpwr command.
  • .vcrestfactor: The POWer:QUALity:VCRESTfactor command.
  • .vrms: The POWer:QUALity:VRMS command.

apppwr property

Return the POWer:QUALity:APPpwr command.

Description
  • Returns the apparent power measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:APPpwr? query.
  • Using the .verify(value) method will send the POWer:QUALity:APPpwr? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:APPpwr?

display property

Return the POWer:QUALity:DISplay command tree.

Usage
  • Using the .query() method will send the POWer:QUALity:DISplay? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .apppwr: The POWer:QUALity:DISplay:APPpwr command.
  • .frequency: The POWer:QUALity:DISplay:FREQuency command.
  • .icrestfactor: The POWer:QUALity:DISplay:ICRESTfactor command.
  • .irms: The POWer:QUALity:DISplay:IRMS command.
  • .phaseangle: The POWer:QUALity:DISplay:PHASEangle command.
  • .powerfactor: The POWer:QUALity:DISplay:POWERFACtor command.
  • .reactpwr: The POWer:QUALity:DISplay:REACTpwr command.
  • .truepwr: The POWer:QUALity:DISplay:TRUEpwr command.
  • .vcrestfactor: The POWer:QUALity:DISplay:VCRESTfactor command.
  • .vrms: The POWer:QUALity:DISplay:VRMS command.

freqreference property

Return the POWer:QUALity:FREQREFerence command.

Description
  • This command specifies the power quality frequency reference.
Usage
  • Using the .query() method will send the POWer:QUALity:FREQREFerence? query.
  • Using the .verify(value) method will send the POWer:QUALity:FREQREFerence? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:FREQREFerence value command.
SCPI Syntax
- POWer:QUALity:FREQREFerence {VOLTage|CURRent}
- POWer:QUALity:FREQREFerence?
Info
  • VOLTage sets voltage as the power quality frequency reference source.
  • CURRent sets current as the power quality frequency reference source.

frequency property

Return the POWer:QUALity:FREQuency command.

Description
  • Returns the frequency measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:FREQuency? query.
  • Using the .verify(value) method will send the POWer:QUALity:FREQuency? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:FREQuency?

icrestfactor property

Return the POWer:QUALity:ICRESTfactor command.

Description
  • Returns the current crest factor measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:ICRESTfactor? query.
  • Using the .verify(value) method will send the POWer:QUALity:ICRESTfactor? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:ICRESTfactor?

irms property

Return the POWer:QUALity:IRMS command.

Description
  • Returns the RMS current measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:IRMS? query.
  • Using the .verify(value) method will send the POWer:QUALity:IRMS? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:IRMS?

phaseangle property

Return the POWer:QUALity:PHASEangle command.

Description
  • Returns the phase angle measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:PHASEangle? query.
  • Using the .verify(value) method will send the POWer:QUALity:PHASEangle? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:PHASEangle?

powerfactor property

Return the POWer:QUALity:POWERFACtor command.

Description
  • Returns the power factor measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:POWERFACtor? query.
  • Using the .verify(value) method will send the POWer:QUALity:POWERFACtor? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:POWERFACtor?

reactpwr property

Return the POWer:QUALity:REACTpwr command.

Description
  • Returns the reactive power measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:REACTpwr? query.
  • Using the .verify(value) method will send the POWer:QUALity:REACTpwr? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:REACTpwr?

truepwr property

Return the POWer:QUALity:TRUEpwr command.

Description
  • Returns the true power measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:TRUEpwr? query.
  • Using the .verify(value) method will send the POWer:QUALity:TRUEpwr? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:TRUEpwr?

vcrestfactor property

Return the POWer:QUALity:VCRESTfactor command.

Description
  • This query returns the measurement for the voltage crest factor.
Usage
  • Using the .query() method will send the POWer:QUALity:VCRESTfactor? query.
  • Using the .verify(value) method will send the POWer:QUALity:VCRESTfactor? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:VCRESTfactor?

vrms property

Return the POWer:QUALity:VRMS command.

Description
  • Returns the RMS voltage measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:VRMS? query.
  • Using the .verify(value) method will send the POWer:QUALity:VRMS? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:VRMS?

PowerQualityApppwr

Bases: SCPICmdRead

The POWer:QUALity:APPpwr command.

Description
  • Returns the apparent power measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:APPpwr? query.
  • Using the .verify(value) method will send the POWer:QUALity:APPpwr? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:APPpwr?

PowerQualityDisplay

Bases: SCPICmdRead

The POWer:QUALity:DISplay command tree.

Usage
  • Using the .query() method will send the POWer:QUALity:DISplay? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay? query and raise an AssertionError if the returned value does not match value.
Properties
  • .apppwr: The POWer:QUALity:DISplay:APPpwr command.
  • .frequency: The POWer:QUALity:DISplay:FREQuency command.
  • .icrestfactor: The POWer:QUALity:DISplay:ICRESTfactor command.
  • .irms: The POWer:QUALity:DISplay:IRMS command.
  • .phaseangle: The POWer:QUALity:DISplay:PHASEangle command.
  • .powerfactor: The POWer:QUALity:DISplay:POWERFACtor command.
  • .reactpwr: The POWer:QUALity:DISplay:REACTpwr command.
  • .truepwr: The POWer:QUALity:DISplay:TRUEpwr command.
  • .vcrestfactor: The POWer:QUALity:DISplay:VCRESTfactor command.
  • .vrms: The POWer:QUALity:DISplay:VRMS command.

apppwr property

Return the POWer:QUALity:DISplay:APPpwr command.

Description
  • This command specifies the display state for the apparent power readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:APPpwr? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:APPpwr? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:APPpwr value command.
SCPI Syntax
- POWer:QUALity:DISplay:APPpwr {ON|OFF|1|0}
- POWer:QUALity:DISplay:APPpwr?
Info
  • OFF or 0 turns off the apparent power display.
  • ON or 1 turns on the apparent power display.

frequency property

Return the POWer:QUALity:DISplay:FREQuency command.

Description
  • This command specifies the display state for the frequency readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:FREQuency? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:FREQuency? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:FREQuency value command.
SCPI Syntax
- POWer:QUALity:DISplay:FREQuency {ON|OFF|1|0}
- POWer:QUALity:DISplay:FREQuency?
Info
  • OFF or 0 turns off the frequency display.
  • ON or 1 turns on the frequency display.

icrestfactor property

Return the POWer:QUALity:DISplay:ICRESTfactor command.

Description
  • This command specifies the display state for the current crest factor readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:ICRESTfactor? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:ICRESTfactor? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:ICRESTfactor value command.
SCPI Syntax
- POWer:QUALity:DISplay:ICRESTfactor {ON|OFF|1|0}
- POWer:QUALity:DISplay:ICRESTfactor?
Info
  • OFF or 0 turns off the current crest factor display.
  • ON or 1 turns on the current crest factor display.

irms property

Return the POWer:QUALity:DISplay:IRMS command.

Description
  • This command specifies the display state for the rms current readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:IRMS? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:IRMS? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:IRMS value command.
SCPI Syntax
- POWer:QUALity:DISplay:IRMS {ON|OFF|1|0}
- POWer:QUALity:DISplay:IRMS?
Info
  • OFF or 0 turns off the rms current display.
  • ON or 1 turns on the rms current display.

phaseangle property

Return the POWer:QUALity:DISplay:PHASEangle command.

Description
  • This command specifies the display state for the phase angle readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:PHASEangle? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:PHASEangle? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:PHASEangle value command.
SCPI Syntax
- POWer:QUALity:DISplay:PHASEangle {ON|OFF|1|0}
- POWer:QUALity:DISplay:PHASEangle?
Info
  • OFF or 0 turns off the phase angle display.
  • ON or 1 turns on the phase angle display.

powerfactor property

Return the POWer:QUALity:DISplay:POWERFACtor command.

Description
  • This command specifies the display state for the power factor readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:POWERFACtor? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:POWERFACtor? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:POWERFACtor value command.
SCPI Syntax
- POWer:QUALity:DISplay:POWERFACtor {ON|OFF|1|0}
- POWer:QUALity:DISplay:POWERFACtor?
Info
  • OFF or 0 turns off the power factor display.
  • ON or 1 turns on the power factor display.

reactpwr property

Return the POWer:QUALity:DISplay:REACTpwr command.

Description
  • This command specifies the display state for the reactor power readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:REACTpwr? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:REACTpwr? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:REACTpwr value command.
SCPI Syntax
- POWer:QUALity:DISplay:REACTpwr {ON|OFF|1|0}
- POWer:QUALity:DISplay:REACTpwr?
Info
  • OFF or 0 turns off the reactor power display.
  • ON or 1 turns on the reactor power display.

truepwr property

Return the POWer:QUALity:DISplay:TRUEpwr command.

Description
  • This command specifies the display state for the true power readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:TRUEpwr? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:TRUEpwr? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:TRUEpwr value command.
SCPI Syntax
- POWer:QUALity:DISplay:TRUEpwr {ON|OFF|1|0}
- POWer:QUALity:DISplay:TRUEpwr?
Info
  • OFF or 0 turns off the true power display.
  • ON or 1 turns on the true power display.

vcrestfactor property

Return the POWer:QUALity:DISplay:VCRESTfactor command.

Description
  • This command specifies the display state for the voltage crest factor readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:VCRESTfactor? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:VCRESTfactor? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:VCRESTfactor value command.
SCPI Syntax
- POWer:QUALity:DISplay:VCRESTfactor {ON|OFF|1|0}
- POWer:QUALity:DISplay:VCRESTfactor?
Info
  • OFF or 0 turns off the voltage crest factor display.
  • ON or 1 turns on the voltage crest factor display.

vrms property

Return the POWer:QUALity:DISplay:VRMS command.

Description
  • This command specifies the display state for the rms voltage readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:VRMS? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:VRMS? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:VRMS value command.
SCPI Syntax
- POWer:QUALity:DISplay:VRMS {ON|OFF|1|0}
- POWer:QUALity:DISplay:VRMS?
Info
  • OFF or 0 turns off the rms voltage display.
  • ON or 1 turns on the rms voltage display.

PowerQualityDisplayApppwr

Bases: SCPICmdWrite, SCPICmdRead

The POWer:QUALity:DISplay:APPpwr command.

Description
  • This command specifies the display state for the apparent power readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:APPpwr? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:APPpwr? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:APPpwr value command.
SCPI Syntax
- POWer:QUALity:DISplay:APPpwr {ON|OFF|1|0}
- POWer:QUALity:DISplay:APPpwr?
Info
  • OFF or 0 turns off the apparent power display.
  • ON or 1 turns on the apparent power display.

PowerQualityDisplayFrequency

Bases: SCPICmdWrite, SCPICmdRead

The POWer:QUALity:DISplay:FREQuency command.

Description
  • This command specifies the display state for the frequency readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:FREQuency? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:FREQuency? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:FREQuency value command.
SCPI Syntax
- POWer:QUALity:DISplay:FREQuency {ON|OFF|1|0}
- POWer:QUALity:DISplay:FREQuency?
Info
  • OFF or 0 turns off the frequency display.
  • ON or 1 turns on the frequency display.

PowerQualityDisplayIcrestfactor

Bases: SCPICmdWrite, SCPICmdRead

The POWer:QUALity:DISplay:ICRESTfactor command.

Description
  • This command specifies the display state for the current crest factor readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:ICRESTfactor? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:ICRESTfactor? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:ICRESTfactor value command.
SCPI Syntax
- POWer:QUALity:DISplay:ICRESTfactor {ON|OFF|1|0}
- POWer:QUALity:DISplay:ICRESTfactor?
Info
  • OFF or 0 turns off the current crest factor display.
  • ON or 1 turns on the current crest factor display.

PowerQualityDisplayIrms

Bases: SCPICmdWrite, SCPICmdRead

The POWer:QUALity:DISplay:IRMS command.

Description
  • This command specifies the display state for the rms current readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:IRMS? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:IRMS? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:IRMS value command.
SCPI Syntax
- POWer:QUALity:DISplay:IRMS {ON|OFF|1|0}
- POWer:QUALity:DISplay:IRMS?
Info
  • OFF or 0 turns off the rms current display.
  • ON or 1 turns on the rms current display.

PowerQualityDisplayPhaseangle

Bases: SCPICmdWrite, SCPICmdRead

The POWer:QUALity:DISplay:PHASEangle command.

Description
  • This command specifies the display state for the phase angle readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:PHASEangle? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:PHASEangle? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:PHASEangle value command.
SCPI Syntax
- POWer:QUALity:DISplay:PHASEangle {ON|OFF|1|0}
- POWer:QUALity:DISplay:PHASEangle?
Info
  • OFF or 0 turns off the phase angle display.
  • ON or 1 turns on the phase angle display.

PowerQualityDisplayPowerfactor

Bases: SCPICmdWrite, SCPICmdRead

The POWer:QUALity:DISplay:POWERFACtor command.

Description
  • This command specifies the display state for the power factor readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:POWERFACtor? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:POWERFACtor? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:POWERFACtor value command.
SCPI Syntax
- POWer:QUALity:DISplay:POWERFACtor {ON|OFF|1|0}
- POWer:QUALity:DISplay:POWERFACtor?
Info
  • OFF or 0 turns off the power factor display.
  • ON or 1 turns on the power factor display.

PowerQualityDisplayReactpwr

Bases: SCPICmdWrite, SCPICmdRead

The POWer:QUALity:DISplay:REACTpwr command.

Description
  • This command specifies the display state for the reactor power readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:REACTpwr? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:REACTpwr? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:REACTpwr value command.
SCPI Syntax
- POWer:QUALity:DISplay:REACTpwr {ON|OFF|1|0}
- POWer:QUALity:DISplay:REACTpwr?
Info
  • OFF or 0 turns off the reactor power display.
  • ON or 1 turns on the reactor power display.

PowerQualityDisplayTruepwr

Bases: SCPICmdWrite, SCPICmdRead

The POWer:QUALity:DISplay:TRUEpwr command.

Description
  • This command specifies the display state for the true power readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:TRUEpwr? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:TRUEpwr? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:TRUEpwr value command.
SCPI Syntax
- POWer:QUALity:DISplay:TRUEpwr {ON|OFF|1|0}
- POWer:QUALity:DISplay:TRUEpwr?
Info
  • OFF or 0 turns off the true power display.
  • ON or 1 turns on the true power display.

PowerQualityDisplayVcrestfactor

Bases: SCPICmdWrite, SCPICmdRead

The POWer:QUALity:DISplay:VCRESTfactor command.

Description
  • This command specifies the display state for the voltage crest factor readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:VCRESTfactor? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:VCRESTfactor? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:VCRESTfactor value command.
SCPI Syntax
- POWer:QUALity:DISplay:VCRESTfactor {ON|OFF|1|0}
- POWer:QUALity:DISplay:VCRESTfactor?
Info
  • OFF or 0 turns off the voltage crest factor display.
  • ON or 1 turns on the voltage crest factor display.

PowerQualityDisplayVrms

Bases: SCPICmdWrite, SCPICmdRead

The POWer:QUALity:DISplay:VRMS command.

Description
  • This command specifies the display state for the rms voltage readout.
Usage
  • Using the .query() method will send the POWer:QUALity:DISplay:VRMS? query.
  • Using the .verify(value) method will send the POWer:QUALity:DISplay:VRMS? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:DISplay:VRMS value command.
SCPI Syntax
- POWer:QUALity:DISplay:VRMS {ON|OFF|1|0}
- POWer:QUALity:DISplay:VRMS?
Info
  • OFF or 0 turns off the rms voltage display.
  • ON or 1 turns on the rms voltage display.

PowerQualityFreqreference

Bases: SCPICmdWrite, SCPICmdRead

The POWer:QUALity:FREQREFerence command.

Description
  • This command specifies the power quality frequency reference.
Usage
  • Using the .query() method will send the POWer:QUALity:FREQREFerence? query.
  • Using the .verify(value) method will send the POWer:QUALity:FREQREFerence? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:QUALity:FREQREFerence value command.
SCPI Syntax
- POWer:QUALity:FREQREFerence {VOLTage|CURRent}
- POWer:QUALity:FREQREFerence?
Info
  • VOLTage sets voltage as the power quality frequency reference source.
  • CURRent sets current as the power quality frequency reference source.

PowerQualityFrequency

Bases: SCPICmdRead

The POWer:QUALity:FREQuency command.

Description
  • Returns the frequency measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:FREQuency? query.
  • Using the .verify(value) method will send the POWer:QUALity:FREQuency? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:FREQuency?

PowerQualityIcrestfactor

Bases: SCPICmdRead

The POWer:QUALity:ICRESTfactor command.

Description
  • Returns the current crest factor measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:ICRESTfactor? query.
  • Using the .verify(value) method will send the POWer:QUALity:ICRESTfactor? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:ICRESTfactor?

PowerQualityIrms

Bases: SCPICmdRead

The POWer:QUALity:IRMS command.

Description
  • Returns the RMS current measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:IRMS? query.
  • Using the .verify(value) method will send the POWer:QUALity:IRMS? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:IRMS?

PowerQualityPhaseangle

Bases: SCPICmdRead

The POWer:QUALity:PHASEangle command.

Description
  • Returns the phase angle measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:PHASEangle? query.
  • Using the .verify(value) method will send the POWer:QUALity:PHASEangle? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:PHASEangle?

PowerQualityPowerfactor

Bases: SCPICmdRead

The POWer:QUALity:POWERFACtor command.

Description
  • Returns the power factor measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:POWERFACtor? query.
  • Using the .verify(value) method will send the POWer:QUALity:POWERFACtor? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:POWERFACtor?

PowerQualityReactpwr

Bases: SCPICmdRead

The POWer:QUALity:REACTpwr command.

Description
  • Returns the reactive power measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:REACTpwr? query.
  • Using the .verify(value) method will send the POWer:QUALity:REACTpwr? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:REACTpwr?

PowerQualityTruepwr

Bases: SCPICmdRead

The POWer:QUALity:TRUEpwr command.

Description
  • Returns the true power measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:TRUEpwr? query.
  • Using the .verify(value) method will send the POWer:QUALity:TRUEpwr? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:TRUEpwr?

PowerQualityVcrestfactor

Bases: SCPICmdRead

The POWer:QUALity:VCRESTfactor command.

Description
  • This query returns the measurement for the voltage crest factor.
Usage
  • Using the .query() method will send the POWer:QUALity:VCRESTfactor? query.
  • Using the .verify(value) method will send the POWer:QUALity:VCRESTfactor? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:VCRESTfactor?

PowerQualityVrms

Bases: SCPICmdRead

The POWer:QUALity:VRMS command.

Description
  • Returns the RMS voltage measurement.
Usage
  • Using the .query() method will send the POWer:QUALity:VRMS? query.
  • Using the .verify(value) method will send the POWer:QUALity:VRMS? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:QUALity:VRMS?

PowerReflevel

Bases: SCPICmdRead

The POWer:REFLevel command tree.

Usage
  • Using the .query() method will send the POWer:REFLevel? query.
  • Using the .verify(value) method will send the POWer:REFLevel? query and raise an AssertionError if the returned value does not match value.
Properties
  • .absolute: The POWer:REFLevel:ABSolute command.
  • .hysteresis: The POWer:REFLevel:HYSTeresis command.
  • .method: The POWer:REFLevel:METHod command.
  • .percent: The POWer:REFLevel:PERCent command.

absolute property

Return the POWer:REFLevel:ABSolute command.

Description
  • This command sets the reference levels to be used for power measurements their default unit values.
Usage
  • Using the .write(value) method will send the POWer:REFLevel:ABSolute value command.
SCPI Syntax
- POWer:REFLevel:ABSolute {SETTODEFaults}
Info
  • SETTODEFaults sets the reference levels to their default values.
Sub-properties
  • .high: The POWer:REFLevel:ABSolute:HIGH command.
  • .low: The POWer:REFLevel:ABSolute:LOW command.
  • .mid: The POWer:REFLevel:ABSolute:MID<x> command.

hysteresis property

Return the POWer:REFLevel:HYSTeresis command.

Description
  • This command specifies the reference level hysteresis value to be used for power measurements.
Usage
  • Using the .query() method will send the POWer:REFLevel:HYSTeresis? query.
  • Using the .verify(value) method will send the POWer:REFLevel:HYSTeresis? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:REFLevel:HYSTeresis value command.
SCPI Syntax
- POWer:REFLevel:HYSTeresis <NR3>
- POWer:REFLevel:HYSTeresis?
Info
  • <NR3> is a floating point number that specifies the hysteresis value.

method property

Return the POWer:REFLevel:METHod command.

Description
  • This command specifies the method used to calculate the 0% and 100% reference level to be used for power measurements.
Usage
  • Using the .query() method will send the POWer:REFLevel:METHod? query.
  • Using the .verify(value) method will send the POWer:REFLevel:METHod? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:REFLevel:METHod value command.
SCPI Syntax
- POWer:REFLevel:METHod {ABSolute|PERCent}
- POWer:REFLevel:METHod?
Info
  • ABSolute specifies that the reference levels are set explicitly using the MEASUrement:REFLevel:ABSolute commands. This method is useful when precise values are required.
  • PERCent specifies that the reference levels are calculated as a percent of the signal amplitude. The percentages are defined using the MEASUrement:REFLevel:PERCent commands.

percent property

Return the POWer:REFLevel:PERCent command.

Description
  • This command sets the reference levels to be used for power measurements to the default percentage values.
Usage
  • Using the .write(value) method will send the POWer:REFLevel:PERCent value command.
SCPI Syntax
- POWer:REFLevel:PERCent <SETTODEFaults>
Info
  • SETTODEFaults sets the reference levels to their default percentage values.
Sub-properties
  • .high: The POWer:REFLevel:PERCent:HIGH command.
  • .low: The POWer:REFLevel:PERCent:LOW command.
  • .mid: The POWer:REFLevel:PERCent:MID<x> command.

PowerReflevelAbsolute

Bases: SCPICmdWrite, SCPICmdRead

The POWer:REFLevel:ABSolute command.

Description
  • This command sets the reference levels to be used for power measurements their default unit values.
Usage
  • Using the .write(value) method will send the POWer:REFLevel:ABSolute value command.
SCPI Syntax
- POWer:REFLevel:ABSolute {SETTODEFaults}
Info
  • SETTODEFaults sets the reference levels to their default values.
Properties
  • .high: The POWer:REFLevel:ABSolute:HIGH command.
  • .low: The POWer:REFLevel:ABSolute:LOW command.
  • .mid: The POWer:REFLevel:ABSolute:MID<x> command.

high property

Return the POWer:REFLevel:ABSolute:HIGH command.

Description
  • This command specifies the top reference level to be used for power measurements.
Usage
  • Using the .query() method will send the POWer:REFLevel:ABSolute:HIGH? query.
  • Using the .verify(value) method will send the POWer:REFLevel:ABSolute:HIGH? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:REFLevel:ABSolute:HIGH value command.
SCPI Syntax
- POWer:REFLevel:ABSolute:HIGH <NR3>; Ranges={D,-1e6,+1E6}
- POWer:REFLevel:ABSolute:HIGH?
Info
  • <NR3> is a floating point number that specifies the absolute high value in volts. Default value is 0.0E+0.

low property

Return the POWer:REFLevel:ABSolute:LOW command.

Description
  • This command specifies the low reference level to be used for power measurements.
Usage
  • Using the .query() method will send the POWer:REFLevel:ABSolute:LOW? query.
  • Using the .verify(value) method will send the POWer:REFLevel:ABSolute:LOW? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:REFLevel:ABSolute:LOW value command.
SCPI Syntax
- POWer:REFLevel:ABSolute:LOW <NR3>; Ranges={D,-1e6,+1E6}
- POWer:REFLevel:ABSolute:LOW?
Info
  • <NR3> is a floating point number that specifies the absolute low value in volts. Default value is 0.0E+0.

mid property

Return the POWer:REFLevel:ABSolute:MID<x> command.

Description
  • This command specifies the mid reference level to be used for power measurements. MID1 is used on the user’s voltage waveform. MID2 is used on the user’s current waveform. MID3 is used on the user’s gate waveform. (MID3 is specific to the power application.)
Usage
  • Using the .query() method will send the POWer:REFLevel:ABSolute:MID<x>? query.
  • Using the .verify(value) method will send the POWer:REFLevel:ABSolute:MID<x>? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:REFLevel:ABSolute:MID<x> value command.
SCPI Syntax
- POWer:REFLevel:ABSolute:MID<x> <NR3>; Ranges={D,-1e6,+1E6}
- POWer:REFLevel:ABSolute:MID<x>?
Info
  • <NR3> is a floating point number that specifies the absolute mid reference value.

PowerReflevelAbsoluteHigh

Bases: SCPICmdWrite, SCPICmdRead

The POWer:REFLevel:ABSolute:HIGH command.

Description
  • This command specifies the top reference level to be used for power measurements.
Usage
  • Using the .query() method will send the POWer:REFLevel:ABSolute:HIGH? query.
  • Using the .verify(value) method will send the POWer:REFLevel:ABSolute:HIGH? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:REFLevel:ABSolute:HIGH value command.
SCPI Syntax
- POWer:REFLevel:ABSolute:HIGH <NR3>; Ranges={D,-1e6,+1E6}
- POWer:REFLevel:ABSolute:HIGH?
Info
  • <NR3> is a floating point number that specifies the absolute high value in volts. Default value is 0.0E+0.

PowerReflevelAbsoluteLow

Bases: SCPICmdWrite, SCPICmdRead

The POWer:REFLevel:ABSolute:LOW command.

Description
  • This command specifies the low reference level to be used for power measurements.
Usage
  • Using the .query() method will send the POWer:REFLevel:ABSolute:LOW? query.
  • Using the .verify(value) method will send the POWer:REFLevel:ABSolute:LOW? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:REFLevel:ABSolute:LOW value command.
SCPI Syntax
- POWer:REFLevel:ABSolute:LOW <NR3>; Ranges={D,-1e6,+1E6}
- POWer:REFLevel:ABSolute:LOW?
Info
  • <NR3> is a floating point number that specifies the absolute low value in volts. Default value is 0.0E+0.

PowerReflevelAbsoluteMidItem

Bases: ValidatedDynamicNumberCmd, SCPICmdWrite, SCPICmdRead

The POWer:REFLevel:ABSolute:MID<x> command.

Description
  • This command specifies the mid reference level to be used for power measurements. MID1 is used on the user’s voltage waveform. MID2 is used on the user’s current waveform. MID3 is used on the user’s gate waveform. (MID3 is specific to the power application.)
Usage
  • Using the .query() method will send the POWer:REFLevel:ABSolute:MID<x>? query.
  • Using the .verify(value) method will send the POWer:REFLevel:ABSolute:MID<x>? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:REFLevel:ABSolute:MID<x> value command.
SCPI Syntax
- POWer:REFLevel:ABSolute:MID<x> <NR3>; Ranges={D,-1e6,+1E6}
- POWer:REFLevel:ABSolute:MID<x>?
Info
  • <NR3> is a floating point number that specifies the absolute mid reference value.

PowerReflevelHysteresis

Bases: SCPICmdWrite, SCPICmdRead

The POWer:REFLevel:HYSTeresis command.

Description
  • This command specifies the reference level hysteresis value to be used for power measurements.
Usage
  • Using the .query() method will send the POWer:REFLevel:HYSTeresis? query.
  • Using the .verify(value) method will send the POWer:REFLevel:HYSTeresis? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:REFLevel:HYSTeresis value command.
SCPI Syntax
- POWer:REFLevel:HYSTeresis <NR3>
- POWer:REFLevel:HYSTeresis?
Info
  • <NR3> is a floating point number that specifies the hysteresis value.

PowerReflevelMethod

Bases: SCPICmdWrite, SCPICmdRead

The POWer:REFLevel:METHod command.

Description
  • This command specifies the method used to calculate the 0% and 100% reference level to be used for power measurements.
Usage
  • Using the .query() method will send the POWer:REFLevel:METHod? query.
  • Using the .verify(value) method will send the POWer:REFLevel:METHod? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:REFLevel:METHod value command.
SCPI Syntax
- POWer:REFLevel:METHod {ABSolute|PERCent}
- POWer:REFLevel:METHod?
Info
  • ABSolute specifies that the reference levels are set explicitly using the MEASUrement:REFLevel:ABSolute commands. This method is useful when precise values are required.
  • PERCent specifies that the reference levels are calculated as a percent of the signal amplitude. The percentages are defined using the MEASUrement:REFLevel:PERCent commands.

PowerReflevelPercent

Bases: SCPICmdWrite, SCPICmdRead

The POWer:REFLevel:PERCent command.

Description
  • This command sets the reference levels to be used for power measurements to the default percentage values.
Usage
  • Using the .write(value) method will send the POWer:REFLevel:PERCent value command.
SCPI Syntax
- POWer:REFLevel:PERCent <SETTODEFaults>
Info
  • SETTODEFaults sets the reference levels to their default percentage values.
Properties
  • .high: The POWer:REFLevel:PERCent:HIGH command.
  • .low: The POWer:REFLevel:PERCent:LOW command.
  • .mid: The POWer:REFLevel:PERCent:MID<x> command.

high property

Return the POWer:REFLevel:PERCent:HIGH command.

Description
  • This command specifies the top reference percent level to be used for power measurements.
Usage
  • Using the .query() method will send the POWer:REFLevel:PERCent:HIGH? query.
  • Using the .verify(value) method will send the POWer:REFLevel:PERCent:HIGH? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:REFLevel:PERCent:HIGH value command.
SCPI Syntax
- POWer:REFLevel:PERCent:HIGH <NR3>; Ranges={D,0.0,100.0}
- POWer:REFLevel:PERCent:HIGH?
Info
  • <NR3> is a floating point number that specifies the high value in percent.

low property

Return the POWer:REFLevel:PERCent:LOW command.

Description
  • This command specifies the low reference percent level to be used for power measurements.
Usage
  • Using the .query() method will send the POWer:REFLevel:PERCent:LOW? query.
  • Using the .verify(value) method will send the POWer:REFLevel:PERCent:LOW? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:REFLevel:PERCent:LOW value command.
SCPI Syntax
- POWer:REFLevel:PERCent:LOW <NR3>; Ranges={D,0.0,100.0}
- POWer:REFLevel:PERCent:LOW?
Info
  • <NR3> is a floating point number that specifies the low value in percentage.

mid property

Return the POWer:REFLevel:PERCent:MID<x> command.

Description
  • This command specifies one of 3 mid reference percentage levels to be used for power measurements. MID1 is used on the user’s voltage waveform. MID2 is used on the user’s current waveform. MID3 is used on the user’s gate waveform. (MID3 is specific to power applications.)
Usage
  • Using the .query() method will send the POWer:REFLevel:PERCent:MID<x>? query.
  • Using the .verify(value) method will send the POWer:REFLevel:PERCent:MID<x>? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:REFLevel:PERCent:MID<x> value command.
SCPI Syntax
- POWer:REFLevel:PERCent:MID<x> <NR3>; Ranges={D,0.0,100.0}
- POWer:REFLevel:PERCent:MID<x>?
Info
  • <NR3> is a floating point number that specifies the mid value in percentage.

PowerReflevelPercentHigh

Bases: SCPICmdWrite, SCPICmdRead

The POWer:REFLevel:PERCent:HIGH command.

Description
  • This command specifies the top reference percent level to be used for power measurements.
Usage
  • Using the .query() method will send the POWer:REFLevel:PERCent:HIGH? query.
  • Using the .verify(value) method will send the POWer:REFLevel:PERCent:HIGH? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:REFLevel:PERCent:HIGH value command.
SCPI Syntax
- POWer:REFLevel:PERCent:HIGH <NR3>; Ranges={D,0.0,100.0}
- POWer:REFLevel:PERCent:HIGH?
Info
  • <NR3> is a floating point number that specifies the high value in percent.

PowerReflevelPercentLow

Bases: SCPICmdWrite, SCPICmdRead

The POWer:REFLevel:PERCent:LOW command.

Description
  • This command specifies the low reference percent level to be used for power measurements.
Usage
  • Using the .query() method will send the POWer:REFLevel:PERCent:LOW? query.
  • Using the .verify(value) method will send the POWer:REFLevel:PERCent:LOW? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:REFLevel:PERCent:LOW value command.
SCPI Syntax
- POWer:REFLevel:PERCent:LOW <NR3>; Ranges={D,0.0,100.0}
- POWer:REFLevel:PERCent:LOW?
Info
  • <NR3> is a floating point number that specifies the low value in percentage.

PowerReflevelPercentMidItem

Bases: ValidatedDynamicNumberCmd, SCPICmdWrite, SCPICmdRead

The POWer:REFLevel:PERCent:MID<x> command.

Description
  • This command specifies one of 3 mid reference percentage levels to be used for power measurements. MID1 is used on the user’s voltage waveform. MID2 is used on the user’s current waveform. MID3 is used on the user’s gate waveform. (MID3 is specific to power applications.)
Usage
  • Using the .query() method will send the POWer:REFLevel:PERCent:MID<x>? query.
  • Using the .verify(value) method will send the POWer:REFLevel:PERCent:MID<x>? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:REFLevel:PERCent:MID<x> value command.
SCPI Syntax
- POWer:REFLevel:PERCent:MID<x> <NR3>; Ranges={D,0.0,100.0}
- POWer:REFLevel:PERCent:MID<x>?
Info
  • <NR3> is a floating point number that specifies the mid value in percentage.

PowerRipple

Bases: SCPICmdWrite, SCPICmdRead

The POWer:RIPPle command.

Description
  • This command performs a vertical autoset for ripple measurements or sets the vertical offset to 0.
Usage
  • Using the .write(value) method will send the POWer:RIPPle value command.
SCPI Syntax
- POWer:RIPPle {VERTAUTOset|VERTDEFault}
Info
  • VERTAUTOset automatically scales the source waveform to optimize ripple measurements.
  • VERTDEFault sets the vertical offset of the source waveform to 0 volts (for voltage source) or 0 amperes (for current source).
Properties
  • .results: The POWer:RIPPle:RESults command tree.
  • .source: The POWer:RIPPle:SOUrce command.

results property

Return the POWer:RIPPle:RESults command tree.

Usage
  • Using the .query() method will send the POWer:RIPPle:RESults? query.
  • Using the .verify(value) method will send the POWer:RIPPle:RESults? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .amplitude: The POWer:RIPPle:RESults:AMPLitude command.
  • .max: The POWer:RIPPle:RESults:MAX command.
  • .mean: The POWer:RIPPle:RESults:MEAN command.
  • .min: The POWer:RIPPle:RESults:MIN command.
  • .stddev: The POWer:RIPPle:RESults:STDdev command.

source property

Return the POWer:RIPPle:SOUrce command.

Description
  • This command specifies the source waveform for ripple tests. The voltage source waveform is specified using the POWER:VOLTAGESOURCE command and the current waveform is specified using the POWER:CURRENTSOURCE command.
Usage
  • Using the .write(value) method will send the POWer:RIPPle:SOUrce value command.
SCPI Syntax
- POWer:RIPPle:SOUrce {VOLTage|CURRent}
Info
  • VOLTage specifies voltage source waveform for ripple tests.
  • CURRent specifies current source waveform for ripple tests.

PowerRippleResults

Bases: SCPICmdRead

The POWer:RIPPle:RESults command tree.

Usage
  • Using the .query() method will send the POWer:RIPPle:RESults? query.
  • Using the .verify(value) method will send the POWer:RIPPle:RESults? query and raise an AssertionError if the returned value does not match value.
Properties
  • .amplitude: The POWer:RIPPle:RESults:AMPLitude command.
  • .max: The POWer:RIPPle:RESults:MAX command.
  • .mean: The POWer:RIPPle:RESults:MEAN command.
  • .min: The POWer:RIPPle:RESults:MIN command.
  • .stddev: The POWer:RIPPle:RESults:STDdev command.

amplitude property

Return the POWer:RIPPle:RESults:AMPLitude command.

Description
  • Returns the peak-to-peak ripple measurement.
Usage
  • Using the .query() method will send the POWer:RIPPle:RESults:AMPLitude? query.
  • Using the .verify(value) method will send the POWer:RIPPle:RESults:AMPLitude? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:RIPPle:RESults:AMPLitude?

max property

Return the POWer:RIPPle:RESults:MAX command.

Description
  • Returns the maximum of the peak-to-peak ripple measurements.
Usage
  • Using the .query() method will send the POWer:RIPPle:RESults:MAX? query.
  • Using the .verify(value) method will send the POWer:RIPPle:RESults:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:RIPPle:RESults:MAX?

mean property

Return the POWer:RIPPle:RESults:MEAN command.

Description
  • Returns the mean of the peak-to-peak ripple measurements.
Usage
  • Using the .query() method will send the POWer:RIPPle:RESults:MEAN? query.
  • Using the .verify(value) method will send the POWer:RIPPle:RESults:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:RIPPle:RESults:MEAN?

min property

Return the POWer:RIPPle:RESults:MIN command.

Description
  • Returns the minimum of the peak-to-peak ripple measurement.
Usage
  • Using the .query() method will send the POWer:RIPPle:RESults:MIN? query.
  • Using the .verify(value) method will send the POWer:RIPPle:RESults:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:RIPPle:RESults:MIN?

stddev property

Return the POWer:RIPPle:RESults:STDdev command.

Description
  • Returns the standard deviation of the peak-to-peak ripple measurements.
Usage
  • Using the .query() method will send the POWer:RIPPle:RESults:STDdev? query.
  • Using the .verify(value) method will send the POWer:RIPPle:RESults:STDdev? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:RIPPle:RESults:STDdev?

PowerRippleResultsAmplitude

Bases: SCPICmdRead

The POWer:RIPPle:RESults:AMPLitude command.

Description
  • Returns the peak-to-peak ripple measurement.
Usage
  • Using the .query() method will send the POWer:RIPPle:RESults:AMPLitude? query.
  • Using the .verify(value) method will send the POWer:RIPPle:RESults:AMPLitude? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:RIPPle:RESults:AMPLitude?

PowerRippleResultsMax

Bases: SCPICmdRead

The POWer:RIPPle:RESults:MAX command.

Description
  • Returns the maximum of the peak-to-peak ripple measurements.
Usage
  • Using the .query() method will send the POWer:RIPPle:RESults:MAX? query.
  • Using the .verify(value) method will send the POWer:RIPPle:RESults:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:RIPPle:RESults:MAX?

PowerRippleResultsMean

Bases: SCPICmdRead

The POWer:RIPPle:RESults:MEAN command.

Description
  • Returns the mean of the peak-to-peak ripple measurements.
Usage
  • Using the .query() method will send the POWer:RIPPle:RESults:MEAN? query.
  • Using the .verify(value) method will send the POWer:RIPPle:RESults:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:RIPPle:RESults:MEAN?

PowerRippleResultsMin

Bases: SCPICmdRead

The POWer:RIPPle:RESults:MIN command.

Description
  • Returns the minimum of the peak-to-peak ripple measurement.
Usage
  • Using the .query() method will send the POWer:RIPPle:RESults:MIN? query.
  • Using the .verify(value) method will send the POWer:RIPPle:RESults:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:RIPPle:RESults:MIN?

PowerRippleResultsStddev

Bases: SCPICmdRead

The POWer:RIPPle:RESults:STDdev command.

Description
  • Returns the standard deviation of the peak-to-peak ripple measurements.
Usage
  • Using the .query() method will send the POWer:RIPPle:RESults:STDdev? query.
  • Using the .verify(value) method will send the POWer:RIPPle:RESults:STDdev? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:RIPPle:RESults:STDdev?

PowerRippleSource

Bases: SCPICmdWrite

The POWer:RIPPle:SOUrce command.

Description
  • This command specifies the source waveform for ripple tests. The voltage source waveform is specified using the POWER:VOLTAGESOURCE command and the current waveform is specified using the POWER:CURRENTSOURCE command.
Usage
  • Using the .write(value) method will send the POWer:RIPPle:SOUrce value command.
SCPI Syntax
- POWer:RIPPle:SOUrce {VOLTage|CURRent}
Info
  • VOLTage specifies voltage source waveform for ripple tests.
  • CURRent specifies current source waveform for ripple tests.

PowerSoa

Bases: SCPICmdRead

The POWer:SOA command tree.

Usage
  • Using the .query() method will send the POWer:SOA? query.
  • Using the .verify(value) method will send the POWer:SOA? query and raise an AssertionError if the returned value does not match value.
Properties
  • .linear: The POWer:SOA:LINear command tree.
  • .log: The POWer:SOA:LOG command tree.
  • .mask: The POWer:SOA:MASK command tree.
  • .plottype: The POWer:SOA:PLOTTYPe command.
  • .result: The POWer:SOA:RESult command tree.

linear property

Return the POWer:SOA:LINear command tree.

Usage
  • Using the .query() method will send the POWer:SOA:LINear? query.
  • Using the .verify(value) method will send the POWer:SOA:LINear? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .xmax: The POWer:SOA:LINear:XMAX command.
  • .xmin: The POWer:SOA:LINear:XMIN command.
  • .ymax: The POWer:SOA:LINear:YMAX command.
  • .ymin: The POWer:SOA:LINear:YMIN command.

log property

Return the POWer:SOA:LOG command tree.

Usage
  • Using the .query() method will send the POWer:SOA:LOG? query.
  • Using the .verify(value) method will send the POWer:SOA:LOG? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .xmax: The POWer:SOA:LOG:XMAX command.
  • .xmin: The POWer:SOA:LOG:XMIN command.
  • .ymax: The POWer:SOA:LOG:YMAX command.
  • .ymin: The POWer:SOA:LOG:YMIN command.

mask property

Return the POWer:SOA:MASK command tree.

Usage
  • Using the .query() method will send the POWer:SOA:MASK? query.
  • Using the .verify(value) method will send the POWer:SOA:MASK? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .define: The POWer:SOA:MASK:DEFine command.
  • .maxamps: The POWer:SOA:MASK:MAXAmps command.
  • .maxvolts: The POWer:SOA:MASK:MAXVolts command.
  • .maxwatts: The POWer:SOA:MASK:MAXWatts command.
  • .nr_pt: The POWer:SOA:MASK:NR_Pt command.
  • .state: The POWer:SOA:MASK:STATE command.
  • .stoponviol: The POWer:SOA:MASK:STOPOnviol command.

plottype property

plottype: PowerSoaPlottype

Return the POWer:SOA:PLOTTYPe command.

Description
  • This command specifies the Safe Operating Area (SOA) plot type.
Usage
  • Using the .query() method will send the POWer:SOA:PLOTTYPe? query.
  • Using the .verify(value) method will send the POWer:SOA:PLOTTYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:PLOTTYPe value command.
SCPI Syntax
- POWer:SOA:PLOTTYPe {LOG|LINear}
- POWer:SOA:PLOTTYPe?
Info
  • LOG for logarithmic SOA plot type.
  • LINear for linear SOA plot type.

result property

Return the POWer:SOA:RESult command tree.

Usage
  • Using the .query() method will send the POWer:SOA:RESult? query.
  • Using the .verify(value) method will send the POWer:SOA:RESult? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .failures: The POWer:SOA:RESult:FAILures command tree.
  • .numacq: The POWer:SOA:RESult:NUMACq command.
  • .state: The POWer:SOA:RESult:STATE command.

PowerSoaLinear

Bases: SCPICmdRead

The POWer:SOA:LINear command tree.

Usage
  • Using the .query() method will send the POWer:SOA:LINear? query.
  • Using the .verify(value) method will send the POWer:SOA:LINear? query and raise an AssertionError if the returned value does not match value.
Properties
  • .xmax: The POWer:SOA:LINear:XMAX command.
  • .xmin: The POWer:SOA:LINear:XMIN command.
  • .ymax: The POWer:SOA:LINear:YMAX command.
  • .ymin: The POWer:SOA:LINear:YMIN command.

xmax property

Return the POWer:SOA:LINear:XMAX command.

Description
  • This command specifies the user XMAX value for use in linear SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:LINear:XMAX? query.
  • Using the .verify(value) method will send the POWer:SOA:LINear:XMAX? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:LINear:XMAX value command.
SCPI Syntax
- POWer:SOA:LINear:XMAX <NR3>
- POWer:SOA:LINear:XMAX?
Info
  • <NR3> is a floating point number that specifies the XMAX value used for linear SOA calculations.

xmin property

Return the POWer:SOA:LINear:XMIN command.

Description
  • This command specifies the user XMIN value for use in linear SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:LINear:XMIN? query.
  • Using the .verify(value) method will send the POWer:SOA:LINear:XMIN? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:LINear:XMIN value command.
SCPI Syntax
- POWer:SOA:LINear:XMIN <NR3>
- POWer:SOA:LINear:XMIN?
Info
  • <NR3> is a floating point number that specifies the XMIN value used for linear SOA calculations.

ymax property

Return the POWer:SOA:LINear:YMAX command.

Description
  • This command specifies the user YMAX value for use in linear SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:LINear:YMAX? query.
  • Using the .verify(value) method will send the POWer:SOA:LINear:YMAX? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:LINear:YMAX value command.
SCPI Syntax
- POWer:SOA:LINear:YMAX <NR3>
- POWer:SOA:LINear:YMAX?
Info
  • <NR3> is a floating point number that specifies the YMAX value used for linear SOA calculations.

ymin property

Return the POWer:SOA:LINear:YMIN command.

Description
  • This command specifies the user YMIN value for use in linear SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:LINear:YMIN? query.
  • Using the .verify(value) method will send the POWer:SOA:LINear:YMIN? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:LINear:YMIN value command.
SCPI Syntax
- POWer:SOA:LINear:YMIN <NR3>
- POWer:SOA:LINear:YMIN?
Info
  • <NR3> is a floating point number that specifies the YMIN value used for linear SOA calculations.

PowerSoaLinearXmax

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SOA:LINear:XMAX command.

Description
  • This command specifies the user XMAX value for use in linear SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:LINear:XMAX? query.
  • Using the .verify(value) method will send the POWer:SOA:LINear:XMAX? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:LINear:XMAX value command.
SCPI Syntax
- POWer:SOA:LINear:XMAX <NR3>
- POWer:SOA:LINear:XMAX?
Info
  • <NR3> is a floating point number that specifies the XMAX value used for linear SOA calculations.

PowerSoaLinearXmin

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SOA:LINear:XMIN command.

Description
  • This command specifies the user XMIN value for use in linear SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:LINear:XMIN? query.
  • Using the .verify(value) method will send the POWer:SOA:LINear:XMIN? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:LINear:XMIN value command.
SCPI Syntax
- POWer:SOA:LINear:XMIN <NR3>
- POWer:SOA:LINear:XMIN?
Info
  • <NR3> is a floating point number that specifies the XMIN value used for linear SOA calculations.

PowerSoaLinearYmax

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SOA:LINear:YMAX command.

Description
  • This command specifies the user YMAX value for use in linear SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:LINear:YMAX? query.
  • Using the .verify(value) method will send the POWer:SOA:LINear:YMAX? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:LINear:YMAX value command.
SCPI Syntax
- POWer:SOA:LINear:YMAX <NR3>
- POWer:SOA:LINear:YMAX?
Info
  • <NR3> is a floating point number that specifies the YMAX value used for linear SOA calculations.

PowerSoaLinearYmin

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SOA:LINear:YMIN command.

Description
  • This command specifies the user YMIN value for use in linear SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:LINear:YMIN? query.
  • Using the .verify(value) method will send the POWer:SOA:LINear:YMIN? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:LINear:YMIN value command.
SCPI Syntax
- POWer:SOA:LINear:YMIN <NR3>
- POWer:SOA:LINear:YMIN?
Info
  • <NR3> is a floating point number that specifies the YMIN value used for linear SOA calculations.

PowerSoaLog

Bases: SCPICmdRead

The POWer:SOA:LOG command tree.

Usage
  • Using the .query() method will send the POWer:SOA:LOG? query.
  • Using the .verify(value) method will send the POWer:SOA:LOG? query and raise an AssertionError if the returned value does not match value.
Properties
  • .xmax: The POWer:SOA:LOG:XMAX command.
  • .xmin: The POWer:SOA:LOG:XMIN command.
  • .ymax: The POWer:SOA:LOG:YMAX command.
  • .ymin: The POWer:SOA:LOG:YMIN command.

xmax property

Return the POWer:SOA:LOG:XMAX command.

Description
  • This command specifies the user XMAX value for use in Log SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:LOG:XMAX? query.
  • Using the .verify(value) method will send the POWer:SOA:LOG:XMAX? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:LOG:XMAX value command.
SCPI Syntax
- POWer:SOA:LOG:XMAX <NR3>
- POWer:SOA:LOG:XMAX?
Info
  • <NR3> is a floating point number that specifies the XMAX value used for log SOA calculations.

xmin property

Return the POWer:SOA:LOG:XMIN command.

Description
  • This command specifies the user XMIN value for use in Log SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:LOG:XMIN? query.
  • Using the .verify(value) method will send the POWer:SOA:LOG:XMIN? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:LOG:XMIN value command.
SCPI Syntax
- POWer:SOA:LOG:XMIN <NR3>
- POWer:SOA:LOG:XMIN?
Info
  • <NR3> is a floating point number that specifies the XMIN value used for log SOA calculations.

ymax property

Return the POWer:SOA:LOG:YMAX command.

Description
  • This command specifies the user YMAX value for use in Log SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:LOG:YMAX? query.
  • Using the .verify(value) method will send the POWer:SOA:LOG:YMAX? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:LOG:YMAX value command.
SCPI Syntax
- POWer:SOA:LOG:YMAX <NR3>
- POWer:SOA:LOG:YMAX?
Info
  • <NR3> is a floating point number that specifies the YMAX value used for log SOA calculations.

ymin property

Return the POWer:SOA:LOG:YMIN command.

Description
  • This command specifies the user YMIN value for use in Log SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:LOG:YMIN? query.
  • Using the .verify(value) method will send the POWer:SOA:LOG:YMIN? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:LOG:YMIN value command.
SCPI Syntax
- POWer:SOA:LOG:YMIN <NR3>
- POWer:SOA:LOG:YMIN?
Info
  • <NR3> is a floating point number that specifies the YMIN value used for log SOA calculations.

PowerSoaLogXmax

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SOA:LOG:XMAX command.

Description
  • This command specifies the user XMAX value for use in Log SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:LOG:XMAX? query.
  • Using the .verify(value) method will send the POWer:SOA:LOG:XMAX? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:LOG:XMAX value command.
SCPI Syntax
- POWer:SOA:LOG:XMAX <NR3>
- POWer:SOA:LOG:XMAX?
Info
  • <NR3> is a floating point number that specifies the XMAX value used for log SOA calculations.

PowerSoaLogXmin

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SOA:LOG:XMIN command.

Description
  • This command specifies the user XMIN value for use in Log SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:LOG:XMIN? query.
  • Using the .verify(value) method will send the POWer:SOA:LOG:XMIN? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:LOG:XMIN value command.
SCPI Syntax
- POWer:SOA:LOG:XMIN <NR3>
- POWer:SOA:LOG:XMIN?
Info
  • <NR3> is a floating point number that specifies the XMIN value used for log SOA calculations.

PowerSoaLogYmax

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SOA:LOG:YMAX command.

Description
  • This command specifies the user YMAX value for use in Log SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:LOG:YMAX? query.
  • Using the .verify(value) method will send the POWer:SOA:LOG:YMAX? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:LOG:YMAX value command.
SCPI Syntax
- POWer:SOA:LOG:YMAX <NR3>
- POWer:SOA:LOG:YMAX?
Info
  • <NR3> is a floating point number that specifies the YMAX value used for log SOA calculations.

PowerSoaLogYmin

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SOA:LOG:YMIN command.

Description
  • This command specifies the user YMIN value for use in Log SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:LOG:YMIN? query.
  • Using the .verify(value) method will send the POWer:SOA:LOG:YMIN? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:LOG:YMIN value command.
SCPI Syntax
- POWer:SOA:LOG:YMIN <NR3>
- POWer:SOA:LOG:YMIN?
Info
  • <NR3> is a floating point number that specifies the YMIN value used for log SOA calculations.

PowerSoaMask

Bases: SCPICmdRead

The POWer:SOA:MASK command tree.

Usage
  • Using the .query() method will send the POWer:SOA:MASK? query.
  • Using the .verify(value) method will send the POWer:SOA:MASK? query and raise an AssertionError if the returned value does not match value.
Properties
  • .define: The POWer:SOA:MASK:DEFine command.
  • .maxamps: The POWer:SOA:MASK:MAXAmps command.
  • .maxvolts: The POWer:SOA:MASK:MAXVolts command.
  • .maxwatts: The POWer:SOA:MASK:MAXWatts command.
  • .nr_pt: The POWer:SOA:MASK:NR_Pt command.
  • .state: The POWer:SOA:MASK:STATE command.
  • .stoponviol: The POWer:SOA:MASK:STOPOnviol command.

define property

Return the POWer:SOA:MASK:DEFine command.

Description
  • This command specifies the X (volts) and Y (Amps) coordinates of the current SOA mask. You can specify the number of points from 2 to 10, minimum being 2. Successive X values must be ≥ the preceding X values. The number of XY points sent determines the value of NR_PT.
Usage
  • Using the .query() method will send the POWer:SOA:MASK:DEFine? query.
  • Using the .verify(value) method will send the POWer:SOA:MASK:DEFine? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:MASK:DEFine value command.
SCPI Syntax
- POWer:SOA:MASK:DEFine <NR3>
- POWer:SOA:MASK:DEFine?
Info
  • <NR3> is a floating point number that represents the SOA mask coordinates.

maxamps property

Return the POWer:SOA:MASK:MAXAmps command.

Description
  • This command specifies the maximum current applied to SOA mask testing.
Usage
  • Using the .query() method will send the POWer:SOA:MASK:MAXAmps? query.
  • Using the .verify(value) method will send the POWer:SOA:MASK:MAXAmps? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:MASK:MAXAmps value command.
SCPI Syntax
- POWer:SOA:MASK:MAXAmps <NR3>
- POWer:SOA:MASK:MAXAmps?
Info
  • <NR3> is a floating point number that specifies the maximum current applied to SOA mask testing.

maxvolts property

Return the POWer:SOA:MASK:MAXVolts command.

Description
  • This command specifies the maximum voltage applied to SOA mask testing.
Usage
  • Using the .query() method will send the POWer:SOA:MASK:MAXVolts? query.
  • Using the .verify(value) method will send the POWer:SOA:MASK:MAXVolts? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:MASK:MAXVolts value command.
SCPI Syntax
- POWer:SOA:MASK:MAXVolts <NR3>
- POWer:SOA:MASK:MAXVolts?
Info
  • <NR3> is a floating point number that specifies the maximum voltage applied to SOA mask testing.

maxwatts property

Return the POWer:SOA:MASK:MAXWatts command.

Description
  • This command specifies the maximum power applied to SOA mask testing.
Usage
  • Using the .query() method will send the POWer:SOA:MASK:MAXWatts? query.
  • Using the .verify(value) method will send the POWer:SOA:MASK:MAXWatts? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:MASK:MAXWatts value command.
SCPI Syntax
- POWer:SOA:MASK:MAXWatts <NR3>
- POWer:SOA:MASK:MAXWatts?
Info
  • <NR3> is a floating point number that specifies the maximum power applied to SOA mask testing.

nr_pt property

Return the POWer:SOA:MASK:NR_Pt command.

Description
  • Returns the number of mask points defined.
Usage
  • Using the .query() method will send the POWer:SOA:MASK:NR_Pt? query.
  • Using the .verify(value) method will send the POWer:SOA:MASK:NR_Pt? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SOA:MASK:NR_Pt?

state property

Return the POWer:SOA:MASK:STATE command.

Description
  • This command specifies the state of the mask for SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:MASK:STATE? query.
  • Using the .verify(value) method will send the POWer:SOA:MASK:STATE? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:MASK:STATE value command.
SCPI Syntax
- POWer:SOA:MASK:STATE {OFF|LIMITS|POINTS}
- POWer:SOA:MASK:STATE?
Info
  • OFF disables mask testing.
  • LIMITS enables mask testing based on limits specified using POWER:SOA:MASK:MAXAMPS, POWER:SOA:MASK:MAXVOLTS, and POWER:SOA:MASK:MAXWATTS commands.
  • POINTS enables mask testing based on masks points defined.

stoponviol property

Return the POWer:SOA:MASK:STOPOnviol command.

Description
  • This command specifies the enabled state of the mask stop on violation condition.
Usage
  • Using the .query() method will send the POWer:SOA:MASK:STOPOnviol? query.
  • Using the .verify(value) method will send the POWer:SOA:MASK:STOPOnviol? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:MASK:STOPOnviol value command.
SCPI Syntax
- POWer:SOA:MASK:STOPOnviol {ON|OFF|1|0}
- POWer:SOA:MASK:STOPOnviol?
Info
  • OFF or 0 enables mask stop on violations.
  • ON or 1 disables mask stop on violations.

PowerSoaMaskDefine

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SOA:MASK:DEFine command.

Description
  • This command specifies the X (volts) and Y (Amps) coordinates of the current SOA mask. You can specify the number of points from 2 to 10, minimum being 2. Successive X values must be ≥ the preceding X values. The number of XY points sent determines the value of NR_PT.
Usage
  • Using the .query() method will send the POWer:SOA:MASK:DEFine? query.
  • Using the .verify(value) method will send the POWer:SOA:MASK:DEFine? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:MASK:DEFine value command.
SCPI Syntax
- POWer:SOA:MASK:DEFine <NR3>
- POWer:SOA:MASK:DEFine?
Info
  • <NR3> is a floating point number that represents the SOA mask coordinates.

PowerSoaMaskMaxamps

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SOA:MASK:MAXAmps command.

Description
  • This command specifies the maximum current applied to SOA mask testing.
Usage
  • Using the .query() method will send the POWer:SOA:MASK:MAXAmps? query.
  • Using the .verify(value) method will send the POWer:SOA:MASK:MAXAmps? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:MASK:MAXAmps value command.
SCPI Syntax
- POWer:SOA:MASK:MAXAmps <NR3>
- POWer:SOA:MASK:MAXAmps?
Info
  • <NR3> is a floating point number that specifies the maximum current applied to SOA mask testing.

PowerSoaMaskMaxvolts

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SOA:MASK:MAXVolts command.

Description
  • This command specifies the maximum voltage applied to SOA mask testing.
Usage
  • Using the .query() method will send the POWer:SOA:MASK:MAXVolts? query.
  • Using the .verify(value) method will send the POWer:SOA:MASK:MAXVolts? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:MASK:MAXVolts value command.
SCPI Syntax
- POWer:SOA:MASK:MAXVolts <NR3>
- POWer:SOA:MASK:MAXVolts?
Info
  • <NR3> is a floating point number that specifies the maximum voltage applied to SOA mask testing.

PowerSoaMaskMaxwatts

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SOA:MASK:MAXWatts command.

Description
  • This command specifies the maximum power applied to SOA mask testing.
Usage
  • Using the .query() method will send the POWer:SOA:MASK:MAXWatts? query.
  • Using the .verify(value) method will send the POWer:SOA:MASK:MAXWatts? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:MASK:MAXWatts value command.
SCPI Syntax
- POWer:SOA:MASK:MAXWatts <NR3>
- POWer:SOA:MASK:MAXWatts?
Info
  • <NR3> is a floating point number that specifies the maximum power applied to SOA mask testing.

PowerSoaMaskNrPt

Bases: SCPICmdRead

The POWer:SOA:MASK:NR_Pt command.

Description
  • Returns the number of mask points defined.
Usage
  • Using the .query() method will send the POWer:SOA:MASK:NR_Pt? query.
  • Using the .verify(value) method will send the POWer:SOA:MASK:NR_Pt? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SOA:MASK:NR_Pt?

PowerSoaMaskState

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SOA:MASK:STATE command.

Description
  • This command specifies the state of the mask for SOA calculations.
Usage
  • Using the .query() method will send the POWer:SOA:MASK:STATE? query.
  • Using the .verify(value) method will send the POWer:SOA:MASK:STATE? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:MASK:STATE value command.
SCPI Syntax
- POWer:SOA:MASK:STATE {OFF|LIMITS|POINTS}
- POWer:SOA:MASK:STATE?
Info
  • OFF disables mask testing.
  • LIMITS enables mask testing based on limits specified using POWER:SOA:MASK:MAXAMPS, POWER:SOA:MASK:MAXVOLTS, and POWER:SOA:MASK:MAXWATTS commands.
  • POINTS enables mask testing based on masks points defined.

PowerSoaMaskStoponviol

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SOA:MASK:STOPOnviol command.

Description
  • This command specifies the enabled state of the mask stop on violation condition.
Usage
  • Using the .query() method will send the POWer:SOA:MASK:STOPOnviol? query.
  • Using the .verify(value) method will send the POWer:SOA:MASK:STOPOnviol? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:MASK:STOPOnviol value command.
SCPI Syntax
- POWer:SOA:MASK:STOPOnviol {ON|OFF|1|0}
- POWer:SOA:MASK:STOPOnviol?
Info
  • OFF or 0 enables mask stop on violations.
  • ON or 1 disables mask stop on violations.

PowerSoaPlottype

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SOA:PLOTTYPe command.

Description
  • This command specifies the Safe Operating Area (SOA) plot type.
Usage
  • Using the .query() method will send the POWer:SOA:PLOTTYPe? query.
  • Using the .verify(value) method will send the POWer:SOA:PLOTTYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SOA:PLOTTYPe value command.
SCPI Syntax
- POWer:SOA:PLOTTYPe {LOG|LINear}
- POWer:SOA:PLOTTYPe?
Info
  • LOG for logarithmic SOA plot type.
  • LINear for linear SOA plot type.

PowerSoaResult

Bases: SCPICmdRead

The POWer:SOA:RESult command tree.

Usage
  • Using the .query() method will send the POWer:SOA:RESult? query.
  • Using the .verify(value) method will send the POWer:SOA:RESult? query and raise an AssertionError if the returned value does not match value.
Properties
  • .failures: The POWer:SOA:RESult:FAILures command tree.
  • .numacq: The POWer:SOA:RESult:NUMACq command.
  • .state: The POWer:SOA:RESult:STATE command.

failures property

Return the POWer:SOA:RESult:FAILures command tree.

Usage
  • Using the .query() method will send the POWer:SOA:RESult:FAILures? query.
  • Using the .verify(value) method will send the POWer:SOA:RESult:FAILures? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .qty: The POWer:SOA:RESult:FAILures:QTY command.

numacq property

Return the POWer:SOA:RESult:NUMACq command.

Description
  • Returns the number of acquisitions in the test.
Usage
  • Using the .query() method will send the POWer:SOA:RESult:NUMACq? query.
  • Using the .verify(value) method will send the POWer:SOA:RESult:NUMACq? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SOA:RESult:NUMACq?

state property

Return the POWer:SOA:RESult:STATE command.

Description
  • Returns the pass/fail state of the SOA test.
Usage
  • Using the .query() method will send the POWer:SOA:RESult:STATE? query.
  • Using the .verify(value) method will send the POWer:SOA:RESult:STATE? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SOA:RESult:STATE?

PowerSoaResultFailures

Bases: SCPICmdRead

The POWer:SOA:RESult:FAILures command tree.

Usage
  • Using the .query() method will send the POWer:SOA:RESult:FAILures? query.
  • Using the .verify(value) method will send the POWer:SOA:RESult:FAILures? query and raise an AssertionError if the returned value does not match value.
Properties
  • .qty: The POWer:SOA:RESult:FAILures:QTY command.

qty property

Return the POWer:SOA:RESult:FAILures:QTY command.

Description
  • Returns the number of failures in the test.
Usage
  • Using the .query() method will send the POWer:SOA:RESult:FAILures:QTY? query.
  • Using the .verify(value) method will send the POWer:SOA:RESult:FAILures:QTY? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SOA:RESult:FAILures:QTY?

PowerSoaResultFailuresQty

Bases: SCPICmdRead

The POWer:SOA:RESult:FAILures:QTY command.

Description
  • Returns the number of failures in the test.
Usage
  • Using the .query() method will send the POWer:SOA:RESult:FAILures:QTY? query.
  • Using the .verify(value) method will send the POWer:SOA:RESult:FAILures:QTY? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SOA:RESult:FAILures:QTY?

PowerSoaResultNumacq

Bases: SCPICmdRead

The POWer:SOA:RESult:NUMACq command.

Description
  • Returns the number of acquisitions in the test.
Usage
  • Using the .query() method will send the POWer:SOA:RESult:NUMACq? query.
  • Using the .verify(value) method will send the POWer:SOA:RESult:NUMACq? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SOA:RESult:NUMACq?

PowerSoaResultState

Bases: SCPICmdRead

The POWer:SOA:RESult:STATE command.

Description
  • Returns the pass/fail state of the SOA test.
Usage
  • Using the .query() method will send the POWer:SOA:RESult:STATE? query.
  • Using the .verify(value) method will send the POWer:SOA:RESult:STATE? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SOA:RESult:STATE?

PowerStatistics

Bases: SCPICmdWrite, SCPICmdRead

The POWer:STATIstics command.

Description
  • Clears all the accumulated statistics of all measurements. Performs the same function as the MEASUREMENT:STATISTICS command.
Usage
  • Using the .write(value) method will send the POWer:STATIstics value command.
SCPI Syntax
- POWer:STATIstics {RESET}
Info
  • RESET clears the measurement statistics.
Properties
  • .mode: The POWer:STATIstics:MODe command.
  • .weighting: The POWer:STATIstics:WEIghting command.

mode property

Return the POWer:STATIstics:MODe command.

Description
  • Enables or disables the display of the measurement statistics. Performs the same function as the MEASUREMENT:STATISTICS:MODE command.
Usage
  • Using the .query() method will send the POWer:STATIstics:MODe? query.
  • Using the .verify(value) method will send the POWer:STATIstics:MODe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:STATIstics:MODe value command.
SCPI Syntax
- POWer:STATIstics:MODe {OFF|ALL}
- POWer:STATIstics:MODe?
Info
  • ALL turns on measurement statistics display.
  • OFF turns all measurements statistics off.

weighting property

Return the POWer:STATIstics:WEIghting command.

Description
  • Sets the number of samples which are included for the statistics computations for mean and the standard deviation. Performs the same function as the MEASUREMENT:STATISTICS:WEIGHTING command.
Usage
  • Using the .query() method will send the POWer:STATIstics:WEIghting? query.
  • Using the .verify(value) method will send the POWer:STATIstics:WEIghting? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:STATIstics:WEIghting value command.
SCPI Syntax
- POWer:STATIstics:WEIghting <NR1>;Ranges {L,2,1000}
- POWer:STATIstics:WEIghting?
Info
  • <NR1> is the number of samples used for the mean and standard deviation statistical accumulations.

PowerStatisticsMode

Bases: SCPICmdWrite, SCPICmdRead

The POWer:STATIstics:MODe command.

Description
  • Enables or disables the display of the measurement statistics. Performs the same function as the MEASUREMENT:STATISTICS:MODE command.
Usage
  • Using the .query() method will send the POWer:STATIstics:MODe? query.
  • Using the .verify(value) method will send the POWer:STATIstics:MODe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:STATIstics:MODe value command.
SCPI Syntax
- POWer:STATIstics:MODe {OFF|ALL}
- POWer:STATIstics:MODe?
Info
  • ALL turns on measurement statistics display.
  • OFF turns all measurements statistics off.

PowerStatisticsWeighting

Bases: SCPICmdWrite, SCPICmdRead

The POWer:STATIstics:WEIghting command.

Description
  • Sets the number of samples which are included for the statistics computations for mean and the standard deviation. Performs the same function as the MEASUREMENT:STATISTICS:WEIGHTING command.
Usage
  • Using the .query() method will send the POWer:STATIstics:WEIghting? query.
  • Using the .verify(value) method will send the POWer:STATIstics:WEIghting? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:STATIstics:WEIghting value command.
SCPI Syntax
- POWer:STATIstics:WEIghting <NR1>;Ranges {L,2,1000}
- POWer:STATIstics:WEIghting?
Info
  • <NR1> is the number of samples used for the mean and standard deviation statistical accumulations.

PowerSwloss

Bases: SCPICmdRead

The POWer:SWLoss command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss? query.
  • Using the .verify(value) method will send the POWer:SWLoss? query and raise an AssertionError if the returned value does not match value.
Properties
  • .condcalcmethod: The POWer:SWLoss:CONDCALCmethod command.
  • .conduction: The POWer:SWLoss:CONDuction command tree.
  • .display: The POWer:SWLoss:DISplay command.
  • .gate: The POWer:SWLoss:GATe command tree.
  • .numcycles: The POWer:SWLoss:NUMCYCles command.
  • .rdson: The POWer:SWLoss:RDSon command.
  • .reflevel: The POWer:SWLoss:REFLevel command tree.
  • .toff: The POWer:SWLoss:TOFF command tree.
  • .ton: The POWer:SWLoss:TON command tree.
  • .total: The POWer:SWLoss:TOTal command tree.
  • .vcesat: The POWer:SWLoss:VCEsat command.

condcalcmethod property

condcalcmethod: PowerSwlossCondcalcmethod

Return the POWer:SWLoss:CONDCALCmethod command.

Description
  • This command specifies the power application switching loss conduction calculation method.
Usage
  • Using the .query() method will send the POWer:SWLoss:CONDCALCmethod? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDCALCmethod? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:CONDCALCmethod value command.
SCPI Syntax
- POWer:SWLoss:CONDCALCmethod {VOLTage|RDSon|VCEsat}
- POWer:SWLoss:CONDCALCmethod?
Info
  • VOLTage sets voltage as the conduction calculation method.
  • RDSon sets RDSon as the conduction calculation method.
  • VCEsat sets VCEsat as the conduction calculation method.

conduction property

Return the POWer:SWLoss:CONDuction command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .energy: The POWer:SWLoss:CONDuction:ENERGY command tree.
  • .power: The POWer:SWLoss:CONDuction:POWer command tree.

display property

Return the POWer:SWLoss:DISplay command.

Description
  • This command specifies the display selection for switching loss results: All measurements, energy loss measurements or power loss measurements.
Usage
  • Using the .query() method will send the POWer:SWLoss:DISplay? query.
  • Using the .verify(value) method will send the POWer:SWLoss:DISplay? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:DISplay value command.
SCPI Syntax
- POWer:SWLoss:DISplay {ALL|ENERGYLoss|POWERLoss}
- POWer:SWLoss:DISplay?
Info
  • ALL displays both energy and power loss measurements in the results.
  • ENERGYLoss displays only energy loss measurements in the results.
  • POWERLoss displays only power loss measurements in the results.

gate property

Return the POWer:SWLoss:GATe command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:GATe? query.
  • Using the .verify(value) method will send the POWer:SWLoss:GATe? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .polarity: The POWer:SWLoss:GATe:POLarity command.
  • .turnon: The POWer:SWLoss:GATe:TURNON command.

numcycles property

Return the POWer:SWLoss:NUMCYCles command.

Description
  • Returns the number of cycles counted for the switching loss calculation.
Usage
  • Using the .query(argument) method will send the POWer:SWLoss:NUMCYCles? argument query.
  • Using the .verify(argument, value) method will send the POWer:SWLoss:NUMCYCles? argument query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:NUMCYCles? <NR3>

rdson property

Return the POWer:SWLoss:RDSon command.

Description
  • This command specifies the user RDSON value for use in switching loss calculations when the conduction calculation method is RDSON.
Usage
  • Using the .query() method will send the POWer:SWLoss:RDSon? query.
  • Using the .verify(value) method will send the POWer:SWLoss:RDSon? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:RDSon value command.
SCPI Syntax
- POWer:SWLoss:RDSon <NR3>
- POWer:SWLoss:RDSon?
Info
  • <NR3> is a floating point number that specifies the RDSON switching loss calculation.

reflevel property

Return the POWer:SWLoss:REFLevel command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .absolute: The POWer:SWLoss:REFLevel:ABSolute command tree.
  • .percent: The POWer:SWLoss:REFLevel:PERCent command tree.

toff property

Return the POWer:SWLoss:TOFF command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .energy: The POWer:SWLoss:TOFF:ENERGY command tree.
  • .power: The POWer:SWLoss:TOFF:POWer command tree.

ton property

Return the POWer:SWLoss:TON command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TON? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .energy: The POWer:SWLoss:TON:ENERGY command tree.
  • .power: The POWer:SWLoss:TON:POWer command tree.

total property

Return the POWer:SWLoss:TOTal command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .energy: The POWer:SWLoss:TOTal:ENERGY command tree.
  • .power: The POWer:SWLoss:TOTal:POWer command tree.

vcesat property

Return the POWer:SWLoss:VCEsat command.

Description
  • This command specifies VCESAT value for use in switching loss calculations when the conduction calculation method is VCESAT.
Usage
  • Using the .query() method will send the POWer:SWLoss:VCEsat? query.
  • Using the .verify(value) method will send the POWer:SWLoss:VCEsat? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:VCEsat value command.
SCPI Syntax
- POWer:SWLoss:VCEsat <NR3>
- POWer:SWLoss:VCEsat?
Info
  • <NR3> is a floating point number that specifies the VCEsat switching loss calculation.

PowerSwlossCondcalcmethod

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SWLoss:CONDCALCmethod command.

Description
  • This command specifies the power application switching loss conduction calculation method.
Usage
  • Using the .query() method will send the POWer:SWLoss:CONDCALCmethod? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDCALCmethod? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:CONDCALCmethod value command.
SCPI Syntax
- POWer:SWLoss:CONDCALCmethod {VOLTage|RDSon|VCEsat}
- POWer:SWLoss:CONDCALCmethod?
Info
  • VOLTage sets voltage as the conduction calculation method.
  • RDSon sets RDSon as the conduction calculation method.
  • VCEsat sets VCEsat as the conduction calculation method.

PowerSwlossConduction

Bases: SCPICmdRead

The POWer:SWLoss:CONDuction command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction? query and raise an AssertionError if the returned value does not match value.
Properties
  • .energy: The POWer:SWLoss:CONDuction:ENERGY command tree.
  • .power: The POWer:SWLoss:CONDuction:POWer command tree.

energy property

Return the POWer:SWLoss:CONDuction:ENERGY command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction:ENERGY? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction:ENERGY? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .max: The POWer:SWLoss:CONDuction:ENERGY:MAX command.
  • .mean: The POWer:SWLoss:CONDuction:ENERGY:MEAN command.
  • .min: The POWer:SWLoss:CONDuction:ENERGY:MIN command.

power property

Return the POWer:SWLoss:CONDuction:POWer command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction:POWer? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction:POWer? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .max: The POWer:SWLoss:CONDuction:POWer:MAX command.
  • .mean: The POWer:SWLoss:CONDuction:POWer:MEAN command.
  • .min: The POWer:SWLoss:CONDuction:POWer:MIN command.

PowerSwlossConductionEnergy

Bases: SCPICmdRead

The POWer:SWLoss:CONDuction:ENERGY command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction:ENERGY? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction:ENERGY? query and raise an AssertionError if the returned value does not match value.
Properties
  • .max: The POWer:SWLoss:CONDuction:ENERGY:MAX command.
  • .mean: The POWer:SWLoss:CONDuction:ENERGY:MEAN command.
  • .min: The POWer:SWLoss:CONDuction:ENERGY:MIN command.

max property

Return the POWer:SWLoss:CONDuction:ENERGY:MAX command.

Description
  • Returns the maximum conduction energy for switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction:ENERGY:MAX? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction:ENERGY:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:CONDuction:ENERGY:MAX?

mean property

Return the POWer:SWLoss:CONDuction:ENERGY:MEAN command.

Description
  • Returns the mean conduction energy in the conduction period for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction:ENERGY:MEAN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction:ENERGY:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:CONDuction:ENERGY:MEAN?

min property

Return the POWer:SWLoss:CONDuction:ENERGY:MIN command.

Description
  • Returns the minimum conduction energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction:ENERGY:MIN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction:ENERGY:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:CONDuction:ENERGY:MIN?

PowerSwlossConductionEnergyMax

Bases: SCPICmdRead

The POWer:SWLoss:CONDuction:ENERGY:MAX command.

Description
  • Returns the maximum conduction energy for switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction:ENERGY:MAX? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction:ENERGY:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:CONDuction:ENERGY:MAX?

PowerSwlossConductionEnergyMean

Bases: SCPICmdRead

The POWer:SWLoss:CONDuction:ENERGY:MEAN command.

Description
  • Returns the mean conduction energy in the conduction period for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction:ENERGY:MEAN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction:ENERGY:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:CONDuction:ENERGY:MEAN?

PowerSwlossConductionEnergyMin

Bases: SCPICmdRead

The POWer:SWLoss:CONDuction:ENERGY:MIN command.

Description
  • Returns the minimum conduction energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction:ENERGY:MIN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction:ENERGY:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:CONDuction:ENERGY:MIN?

PowerSwlossConductionPower

Bases: SCPICmdRead

The POWer:SWLoss:CONDuction:POWer command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction:POWer? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction:POWer? query and raise an AssertionError if the returned value does not match value.
Properties
  • .max: The POWer:SWLoss:CONDuction:POWer:MAX command.
  • .mean: The POWer:SWLoss:CONDuction:POWer:MEAN command.
  • .min: The POWer:SWLoss:CONDuction:POWer:MIN command.

max property

Return the POWer:SWLoss:CONDuction:POWer:MAX command.

Description
  • Returns the maximum conduction power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction:POWer:MAX? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction:POWer:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:CONDuction:POWer:MAX?

mean property

Return the POWer:SWLoss:CONDuction:POWer:MEAN command.

Description
  • Returns the mean conduction power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction:POWer:MEAN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction:POWer:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:CONDuction:POWer:MEAN?

min property

Return the POWer:SWLoss:CONDuction:POWer:MIN command.

Description
  • Returns the minimum conduction power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction:POWer:MIN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction:POWer:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:CONDuction:POWer:MIN?

PowerSwlossConductionPowerMax

Bases: SCPICmdRead

The POWer:SWLoss:CONDuction:POWer:MAX command.

Description
  • Returns the maximum conduction power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction:POWer:MAX? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction:POWer:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:CONDuction:POWer:MAX?

PowerSwlossConductionPowerMean

Bases: SCPICmdRead

The POWer:SWLoss:CONDuction:POWer:MEAN command.

Description
  • Returns the mean conduction power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction:POWer:MEAN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction:POWer:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:CONDuction:POWer:MEAN?

PowerSwlossConductionPowerMin

Bases: SCPICmdRead

The POWer:SWLoss:CONDuction:POWer:MIN command.

Description
  • Returns the minimum conduction power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:CONDuction:POWer:MIN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:CONDuction:POWer:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:CONDuction:POWer:MIN?

PowerSwlossDisplay

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SWLoss:DISplay command.

Description
  • This command specifies the display selection for switching loss results: All measurements, energy loss measurements or power loss measurements.
Usage
  • Using the .query() method will send the POWer:SWLoss:DISplay? query.
  • Using the .verify(value) method will send the POWer:SWLoss:DISplay? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:DISplay value command.
SCPI Syntax
- POWer:SWLoss:DISplay {ALL|ENERGYLoss|POWERLoss}
- POWer:SWLoss:DISplay?
Info
  • ALL displays both energy and power loss measurements in the results.
  • ENERGYLoss displays only energy loss measurements in the results.
  • POWERLoss displays only power loss measurements in the results.

PowerSwlossGate

Bases: SCPICmdRead

The POWer:SWLoss:GATe command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:GATe? query.
  • Using the .verify(value) method will send the POWer:SWLoss:GATe? query and raise an AssertionError if the returned value does not match value.
Properties
  • .polarity: The POWer:SWLoss:GATe:POLarity command.
  • .turnon: The POWer:SWLoss:GATe:TURNON command.

polarity property

Return the POWer:SWLoss:GATe:POLarity command.

Description
  • This command specifies the switching loss gate polarity.
Usage
  • Using the .query() method will send the POWer:SWLoss:GATe:POLarity? query.
  • Using the .verify(value) method will send the POWer:SWLoss:GATe:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:GATe:POLarity value command.
SCPI Syntax
- POWer:SWLoss:GATe:POLarity {FALL|RISe}
- POWer:SWLoss:GATe:POLarity?
Info
  • FALL sets falling edge as the switching loss gate polarity.
  • RISe sets rising edge as the switching loss gate polarity.

turnon property

Return the POWer:SWLoss:GATe:TURNON command.

Description
  • This command specifies the gate turn on level for switching loss power measurements.
Usage
  • Using the .query() method will send the POWer:SWLoss:GATe:TURNON? query.
  • Using the .verify(value) method will send the POWer:SWLoss:GATe:TURNON? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:GATe:TURNON value command.
SCPI Syntax
- POWer:SWLoss:GATe:TURNON <NR3>
- POWer:SWLoss:GATe:TURNON?

PowerSwlossGatePolarity

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SWLoss:GATe:POLarity command.

Description
  • This command specifies the switching loss gate polarity.
Usage
  • Using the .query() method will send the POWer:SWLoss:GATe:POLarity? query.
  • Using the .verify(value) method will send the POWer:SWLoss:GATe:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:GATe:POLarity value command.
SCPI Syntax
- POWer:SWLoss:GATe:POLarity {FALL|RISe}
- POWer:SWLoss:GATe:POLarity?
Info
  • FALL sets falling edge as the switching loss gate polarity.
  • RISe sets rising edge as the switching loss gate polarity.

PowerSwlossGateTurnon

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SWLoss:GATe:TURNON command.

Description
  • This command specifies the gate turn on level for switching loss power measurements.
Usage
  • Using the .query() method will send the POWer:SWLoss:GATe:TURNON? query.
  • Using the .verify(value) method will send the POWer:SWLoss:GATe:TURNON? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:GATe:TURNON value command.
SCPI Syntax
- POWer:SWLoss:GATe:TURNON <NR3>
- POWer:SWLoss:GATe:TURNON?

PowerSwlossNumcycles

Bases: SCPICmdReadWithArguments

The POWer:SWLoss:NUMCYCles command.

Description
  • Returns the number of cycles counted for the switching loss calculation.
Usage
  • Using the .query(argument) method will send the POWer:SWLoss:NUMCYCles? argument query.
  • Using the .verify(argument, value) method will send the POWer:SWLoss:NUMCYCles? argument query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:NUMCYCles? <NR3>

PowerSwlossRdson

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SWLoss:RDSon command.

Description
  • This command specifies the user RDSON value for use in switching loss calculations when the conduction calculation method is RDSON.
Usage
  • Using the .query() method will send the POWer:SWLoss:RDSon? query.
  • Using the .verify(value) method will send the POWer:SWLoss:RDSon? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:RDSon value command.
SCPI Syntax
- POWer:SWLoss:RDSon <NR3>
- POWer:SWLoss:RDSon?
Info
  • <NR3> is a floating point number that specifies the RDSON switching loss calculation.

PowerSwlossReflevel

Bases: SCPICmdRead

The POWer:SWLoss:REFLevel command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel? query and raise an AssertionError if the returned value does not match value.
Properties
  • .absolute: The POWer:SWLoss:REFLevel:ABSolute command tree.
  • .percent: The POWer:SWLoss:REFLevel:PERCent command tree.

absolute property

Return the POWer:SWLoss:REFLevel:ABSolute command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel:ABSolute? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel:ABSolute? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .gatemid: The POWer:SWLoss:REFLevel:ABSolute:GATEMid command.
  • .lowcurrent: The POWer:SWLoss:REFLevel:ABSolute:LOWCurrent command.
  • .lowvoltage: The POWer:SWLoss:REFLevel:ABSolute:LOWVoltage command.

percent property

Return the POWer:SWLoss:REFLevel:PERCent command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel:PERCent? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel:PERCent? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .gatemid: The POWer:SWLoss:REFLevel:PERCent:GATEMid command.
  • .lowcurrent: The POWer:SWLoss:REFLevel:PERCent:LOWCurrent command.
  • .lowvoltage: The POWer:SWLoss:REFLevel:PERCent:LOWVoltage command.

PowerSwlossReflevelAbsolute

Bases: SCPICmdRead

The POWer:SWLoss:REFLevel:ABSolute command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel:ABSolute? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel:ABSolute? query and raise an AssertionError if the returned value does not match value.
Properties
  • .gatemid: The POWer:SWLoss:REFLevel:ABSolute:GATEMid command.
  • .lowcurrent: The POWer:SWLoss:REFLevel:ABSolute:LOWCurrent command.
  • .lowvoltage: The POWer:SWLoss:REFLevel:ABSolute:LOWVoltage command.

gatemid property

Return the POWer:SWLoss:REFLevel:ABSolute:GATEMid command.

Description
  • This command specifies the mid voltage reference level used in switching loss power measurements in volts.
Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel:ABSolute:GATEMid? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel:ABSolute:GATEMid? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:REFLevel:ABSolute:GATEMid value command.
SCPI Syntax
- POWer:SWLoss:REFLevel:ABSolute:GATEMid <NR3>
- POWer:SWLoss:REFLevel:ABSolute:GATEMid?
Info
  • <NR3> is a floating point number that specifies the mid voltage reference level in volts.

lowcurrent property

Return the POWer:SWLoss:REFLevel:ABSolute:LOWCurrent command.

Description
  • This command specifies the low current reference level used in switching loss power measurements in amperes.
Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel:ABSolute:LOWCurrent? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel:ABSolute:LOWCurrent? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:REFLevel:ABSolute:LOWCurrent value command.
SCPI Syntax
- POWer:SWLoss:REFLevel:ABSolute:LOWCurrent <NR3>
- POWer:SWLoss:REFLevel:ABSolute:LOWCurrent?
Info
  • <NR3> is a floating point number that specifies the low voltage current level in amperes.

lowvoltage property

Return the POWer:SWLoss:REFLevel:ABSolute:LOWVoltage command.

Description
  • This command specifies the low voltage reference level used in switching loss power measurements in volts.
Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel:ABSolute:LOWVoltage? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel:ABSolute:LOWVoltage? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:REFLevel:ABSolute:LOWVoltage value command.
SCPI Syntax
- POWer:SWLoss:REFLevel:ABSolute:LOWVoltage <NR3>
- POWer:SWLoss:REFLevel:ABSolute:LOWVoltage?
Info
  • <NR3> is a floating point number that specifies the low voltage reference level in volts.

PowerSwlossReflevelAbsoluteGatemid

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SWLoss:REFLevel:ABSolute:GATEMid command.

Description
  • This command specifies the mid voltage reference level used in switching loss power measurements in volts.
Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel:ABSolute:GATEMid? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel:ABSolute:GATEMid? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:REFLevel:ABSolute:GATEMid value command.
SCPI Syntax
- POWer:SWLoss:REFLevel:ABSolute:GATEMid <NR3>
- POWer:SWLoss:REFLevel:ABSolute:GATEMid?
Info
  • <NR3> is a floating point number that specifies the mid voltage reference level in volts.

PowerSwlossReflevelAbsoluteLowcurrent

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SWLoss:REFLevel:ABSolute:LOWCurrent command.

Description
  • This command specifies the low current reference level used in switching loss power measurements in amperes.
Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel:ABSolute:LOWCurrent? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel:ABSolute:LOWCurrent? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:REFLevel:ABSolute:LOWCurrent value command.
SCPI Syntax
- POWer:SWLoss:REFLevel:ABSolute:LOWCurrent <NR3>
- POWer:SWLoss:REFLevel:ABSolute:LOWCurrent?
Info
  • <NR3> is a floating point number that specifies the low voltage current level in amperes.

PowerSwlossReflevelAbsoluteLowvoltage

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SWLoss:REFLevel:ABSolute:LOWVoltage command.

Description
  • This command specifies the low voltage reference level used in switching loss power measurements in volts.
Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel:ABSolute:LOWVoltage? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel:ABSolute:LOWVoltage? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:REFLevel:ABSolute:LOWVoltage value command.
SCPI Syntax
- POWer:SWLoss:REFLevel:ABSolute:LOWVoltage <NR3>
- POWer:SWLoss:REFLevel:ABSolute:LOWVoltage?
Info
  • <NR3> is a floating point number that specifies the low voltage reference level in volts.

PowerSwlossReflevelPercent

Bases: SCPICmdRead

The POWer:SWLoss:REFLevel:PERCent command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel:PERCent? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel:PERCent? query and raise an AssertionError if the returned value does not match value.
Properties
  • .gatemid: The POWer:SWLoss:REFLevel:PERCent:GATEMid command.
  • .lowcurrent: The POWer:SWLoss:REFLevel:PERCent:LOWCurrent command.
  • .lowvoltage: The POWer:SWLoss:REFLevel:PERCent:LOWVoltage command.

gatemid property

Return the POWer:SWLoss:REFLevel:PERCent:GATEMid command.

Description
  • This command specifies the mid voltage reference level used in switching loss power measurements in percent.
Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel:PERCent:GATEMid? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel:PERCent:GATEMid? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:REFLevel:PERCent:GATEMid value command.
SCPI Syntax
- POWer:SWLoss:REFLevel:PERCent:GATEMid <NR3>
- POWer:SWLoss:REFLevel:PERCent:GATEMid?
Info
  • <NR3> is a floating point number that specifies the mid voltage reference level in volts.

lowcurrent property

Return the POWer:SWLoss:REFLevel:PERCent:LOWCurrent command.

Description
  • This command specifies the low current reference level used in switching loss power measurements in percent.
Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel:PERCent:LOWCurrent? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel:PERCent:LOWCurrent? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:REFLevel:PERCent:LOWCurrent value command.
SCPI Syntax
- POWer:SWLoss:REFLevel:PERCent:LOWCurrent <NR3>
- POWer:SWLoss:REFLevel:PERCent:LOWCurrent?
Info
  • <NR3> is a floating point number that specifies the low voltage reference level percent.

lowvoltage property

Return the POWer:SWLoss:REFLevel:PERCent:LOWVoltage command.

Description
  • This command specifies the low voltage reference level used in switching loss power measurements in percent.
Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel:PERCent:LOWVoltage? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel:PERCent:LOWVoltage? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:REFLevel:PERCent:LOWVoltage value command.
SCPI Syntax
- POWer:SWLoss:REFLevel:PERCent:LOWVoltage <NR3>
- POWer:SWLoss:REFLevel:PERCent:LOWVoltage?
Info
  • <NR3> is a floating point number that specifies the low voltage reference level in percent.

PowerSwlossReflevelPercentGatemid

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SWLoss:REFLevel:PERCent:GATEMid command.

Description
  • This command specifies the mid voltage reference level used in switching loss power measurements in percent.
Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel:PERCent:GATEMid? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel:PERCent:GATEMid? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:REFLevel:PERCent:GATEMid value command.
SCPI Syntax
- POWer:SWLoss:REFLevel:PERCent:GATEMid <NR3>
- POWer:SWLoss:REFLevel:PERCent:GATEMid?
Info
  • <NR3> is a floating point number that specifies the mid voltage reference level in volts.

PowerSwlossReflevelPercentLowcurrent

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SWLoss:REFLevel:PERCent:LOWCurrent command.

Description
  • This command specifies the low current reference level used in switching loss power measurements in percent.
Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel:PERCent:LOWCurrent? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel:PERCent:LOWCurrent? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:REFLevel:PERCent:LOWCurrent value command.
SCPI Syntax
- POWer:SWLoss:REFLevel:PERCent:LOWCurrent <NR3>
- POWer:SWLoss:REFLevel:PERCent:LOWCurrent?
Info
  • <NR3> is a floating point number that specifies the low voltage reference level percent.

PowerSwlossReflevelPercentLowvoltage

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SWLoss:REFLevel:PERCent:LOWVoltage command.

Description
  • This command specifies the low voltage reference level used in switching loss power measurements in percent.
Usage
  • Using the .query() method will send the POWer:SWLoss:REFLevel:PERCent:LOWVoltage? query.
  • Using the .verify(value) method will send the POWer:SWLoss:REFLevel:PERCent:LOWVoltage? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:REFLevel:PERCent:LOWVoltage value command.
SCPI Syntax
- POWer:SWLoss:REFLevel:PERCent:LOWVoltage <NR3>
- POWer:SWLoss:REFLevel:PERCent:LOWVoltage?
Info
  • <NR3> is a floating point number that specifies the low voltage reference level in percent.

PowerSwlossToff

Bases: SCPICmdRead

The POWer:SWLoss:TOFF command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF? query and raise an AssertionError if the returned value does not match value.
Properties
  • .energy: The POWer:SWLoss:TOFF:ENERGY command tree.
  • .power: The POWer:SWLoss:TOFF:POWer command tree.

energy property

Return the POWer:SWLoss:TOFF:ENERGY command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF:ENERGY? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF:ENERGY? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .max: The POWer:SWLoss:TOFF:ENERGY:MAX command.
  • .mean: The POWer:SWLoss:TOFF:ENERGY:MEAN command.
  • .min: The POWer:SWLoss:TOFF:ENERGY:MIN command.

power property

Return the POWer:SWLoss:TOFF:POWer command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF:POWer? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF:POWer? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .max: The POWer:SWLoss:TOFF:POWer:MAX command.
  • .mean: The POWer:SWLoss:TOFF:POWer:MEAN command.
  • .min: The POWer:SWLoss:TOFF:POWer:MIN command.

PowerSwlossToffEnergy

Bases: SCPICmdRead

The POWer:SWLoss:TOFF:ENERGY command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF:ENERGY? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF:ENERGY? query and raise an AssertionError if the returned value does not match value.
Properties
  • .max: The POWer:SWLoss:TOFF:ENERGY:MAX command.
  • .mean: The POWer:SWLoss:TOFF:ENERGY:MEAN command.
  • .min: The POWer:SWLoss:TOFF:ENERGY:MIN command.

max property

Return the POWer:SWLoss:TOFF:ENERGY:MAX command.

Description
  • Returns the maximum Toff energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF:ENERGY:MAX? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF:ENERGY:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOFF:ENERGY:MAX?

mean property

Return the POWer:SWLoss:TOFF:ENERGY:MEAN command.

Description
  • Returns the mean Toff energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF:ENERGY:MEAN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF:ENERGY:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOFF:ENERGY:MEAN?

min property

Return the POWer:SWLoss:TOFF:ENERGY:MIN command.

Description
  • Returns the minimum Toff energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF:ENERGY:MIN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF:ENERGY:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOFF:ENERGY:MIN?

PowerSwlossToffEnergyMax

Bases: SCPICmdRead

The POWer:SWLoss:TOFF:ENERGY:MAX command.

Description
  • Returns the maximum Toff energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF:ENERGY:MAX? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF:ENERGY:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOFF:ENERGY:MAX?

PowerSwlossToffEnergyMean

Bases: SCPICmdRead

The POWer:SWLoss:TOFF:ENERGY:MEAN command.

Description
  • Returns the mean Toff energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF:ENERGY:MEAN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF:ENERGY:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOFF:ENERGY:MEAN?

PowerSwlossToffEnergyMin

Bases: SCPICmdRead

The POWer:SWLoss:TOFF:ENERGY:MIN command.

Description
  • Returns the minimum Toff energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF:ENERGY:MIN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF:ENERGY:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOFF:ENERGY:MIN?

PowerSwlossToffPower

Bases: SCPICmdRead

The POWer:SWLoss:TOFF:POWer command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF:POWer? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF:POWer? query and raise an AssertionError if the returned value does not match value.
Properties
  • .max: The POWer:SWLoss:TOFF:POWer:MAX command.
  • .mean: The POWer:SWLoss:TOFF:POWer:MEAN command.
  • .min: The POWer:SWLoss:TOFF:POWer:MIN command.

max property

Return the POWer:SWLoss:TOFF:POWer:MAX command.

Description
  • Returns the maximum Toff power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF:POWer:MAX? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF:POWer:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOFF:POWer:MAX?

mean property

Return the POWer:SWLoss:TOFF:POWer:MEAN command.

Description
  • Returns the mean Toff power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF:POWer:MEAN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF:POWer:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOFF:POWer:MEAN?

min property

Return the POWer:SWLoss:TOFF:POWer:MIN command.

Description
  • Returns the minimum Toff power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF:POWer:MIN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF:POWer:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOFF:POWer:MIN?

PowerSwlossToffPowerMax

Bases: SCPICmdRead

The POWer:SWLoss:TOFF:POWer:MAX command.

Description
  • Returns the maximum Toff power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF:POWer:MAX? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF:POWer:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOFF:POWer:MAX?

PowerSwlossToffPowerMean

Bases: SCPICmdRead

The POWer:SWLoss:TOFF:POWer:MEAN command.

Description
  • Returns the mean Toff power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF:POWer:MEAN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF:POWer:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOFF:POWer:MEAN?

PowerSwlossToffPowerMin

Bases: SCPICmdRead

The POWer:SWLoss:TOFF:POWer:MIN command.

Description
  • Returns the minimum Toff power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOFF:POWer:MIN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOFF:POWer:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOFF:POWer:MIN?

PowerSwlossTon

Bases: SCPICmdRead

The POWer:SWLoss:TON command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TON? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON? query and raise an AssertionError if the returned value does not match value.
Properties
  • .energy: The POWer:SWLoss:TON:ENERGY command tree.
  • .power: The POWer:SWLoss:TON:POWer command tree.

energy property

Return the POWer:SWLoss:TON:ENERGY command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TON:ENERGY? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON:ENERGY? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .max: The POWer:SWLoss:TON:ENERGY:MAX command.
  • .mean: The POWer:SWLoss:TON:ENERGY:MEAN command.
  • .min: The POWer:SWLoss:TON:ENERGY:MIN command.

power property

Return the POWer:SWLoss:TON:POWer command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TON:POWer? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON:POWer? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .max: The POWer:SWLoss:TON:POWer:MAX command.
  • .mean: The POWer:SWLoss:TON:POWer:MEAN command.
  • .min: The POWer:SWLoss:TON:POWer:MIN command.

PowerSwlossTonEnergy

Bases: SCPICmdRead

The POWer:SWLoss:TON:ENERGY command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TON:ENERGY? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON:ENERGY? query and raise an AssertionError if the returned value does not match value.
Properties
  • .max: The POWer:SWLoss:TON:ENERGY:MAX command.
  • .mean: The POWer:SWLoss:TON:ENERGY:MEAN command.
  • .min: The POWer:SWLoss:TON:ENERGY:MIN command.

max property

Return the POWer:SWLoss:TON:ENERGY:MAX command.

Description
  • Returns the maximum Ton energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TON:ENERGY:MAX? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON:ENERGY:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TON:ENERGY:MAX?

mean property

Return the POWer:SWLoss:TON:ENERGY:MEAN command.

Description
  • Returns the mean Ton energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TON:ENERGY:MEAN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON:ENERGY:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TON:ENERGY:MEAN?

min property

Return the POWer:SWLoss:TON:ENERGY:MIN command.

Description
  • Returns the minimum Ton energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TON:ENERGY:MIN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON:ENERGY:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TON:ENERGY:MIN?

PowerSwlossTonEnergyMax

Bases: SCPICmdRead

The POWer:SWLoss:TON:ENERGY:MAX command.

Description
  • Returns the maximum Ton energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TON:ENERGY:MAX? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON:ENERGY:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TON:ENERGY:MAX?

PowerSwlossTonEnergyMean

Bases: SCPICmdRead

The POWer:SWLoss:TON:ENERGY:MEAN command.

Description
  • Returns the mean Ton energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TON:ENERGY:MEAN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON:ENERGY:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TON:ENERGY:MEAN?

PowerSwlossTonEnergyMin

Bases: SCPICmdRead

The POWer:SWLoss:TON:ENERGY:MIN command.

Description
  • Returns the minimum Ton energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TON:ENERGY:MIN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON:ENERGY:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TON:ENERGY:MIN?

PowerSwlossTonPower

Bases: SCPICmdRead

The POWer:SWLoss:TON:POWer command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TON:POWer? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON:POWer? query and raise an AssertionError if the returned value does not match value.
Properties
  • .max: The POWer:SWLoss:TON:POWer:MAX command.
  • .mean: The POWer:SWLoss:TON:POWer:MEAN command.
  • .min: The POWer:SWLoss:TON:POWer:MIN command.

max property

Return the POWer:SWLoss:TON:POWer:MAX command.

Description
  • Returns the maximum Ton power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TON:POWer:MAX? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON:POWer:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TON:POWer:MAX?

mean property

Return the POWer:SWLoss:TON:POWer:MEAN command.

Description
  • Returns the mean Ton power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TON:POWer:MEAN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON:POWer:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TON:POWer:MEAN?

min property

Return the POWer:SWLoss:TON:POWer:MIN command.

Description
  • Returns the minimum Ton power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TON:POWer:MIN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON:POWer:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TON:POWer:MIN?

PowerSwlossTonPowerMax

Bases: SCPICmdRead

The POWer:SWLoss:TON:POWer:MAX command.

Description
  • Returns the maximum Ton power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TON:POWer:MAX? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON:POWer:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TON:POWer:MAX?

PowerSwlossTonPowerMean

Bases: SCPICmdRead

The POWer:SWLoss:TON:POWer:MEAN command.

Description
  • Returns the mean Ton power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TON:POWer:MEAN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON:POWer:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TON:POWer:MEAN?

PowerSwlossTonPowerMin

Bases: SCPICmdRead

The POWer:SWLoss:TON:POWer:MIN command.

Description
  • Returns the minimum Ton power for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TON:POWer:MIN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TON:POWer:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TON:POWer:MIN?

PowerSwlossTotal

Bases: SCPICmdRead

The POWer:SWLoss:TOTal command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal? query and raise an AssertionError if the returned value does not match value.
Properties
  • .energy: The POWer:SWLoss:TOTal:ENERGY command tree.
  • .power: The POWer:SWLoss:TOTal:POWer command tree.

energy property

Return the POWer:SWLoss:TOTal:ENERGY command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal:ENERGY? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal:ENERGY? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .max: The POWer:SWLoss:TOTal:ENERGY:MAX command.
  • .mean: The POWer:SWLoss:TOTal:ENERGY:MEAN command.
  • .min: The POWer:SWLoss:TOTal:ENERGY:MIN command.

power property

Return the POWer:SWLoss:TOTal:POWer command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal:POWer? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal:POWer? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .max: The POWer:SWLoss:TOTal:POWer:MAX command.
  • .mean: The POWer:SWLoss:TOTal:POWer:MEAN command.
  • .min: The POWer:SWLoss:TOTal:POWer:MIN command.

PowerSwlossTotalEnergy

Bases: SCPICmdRead

The POWer:SWLoss:TOTal:ENERGY command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal:ENERGY? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal:ENERGY? query and raise an AssertionError if the returned value does not match value.
Properties
  • .max: The POWer:SWLoss:TOTal:ENERGY:MAX command.
  • .mean: The POWer:SWLoss:TOTal:ENERGY:MEAN command.
  • .min: The POWer:SWLoss:TOTal:ENERGY:MIN command.

max property

Return the POWer:SWLoss:TOTal:ENERGY:MAX command.

Description
  • Returns the maximum total energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal:ENERGY:MAX? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal:ENERGY:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOTal:ENERGY:MAX?

mean property

Return the POWer:SWLoss:TOTal:ENERGY:MEAN command.

Description
  • Returns the mean total energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal:ENERGY:MEAN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal:ENERGY:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOTal:ENERGY:MEAN?

min property

Return the POWer:SWLoss:TOTal:ENERGY:MIN command.

Description
  • Returns the minimum total energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal:ENERGY:MIN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal:ENERGY:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOTal:ENERGY:MIN?

PowerSwlossTotalEnergyMax

Bases: SCPICmdRead

The POWer:SWLoss:TOTal:ENERGY:MAX command.

Description
  • Returns the maximum total energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal:ENERGY:MAX? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal:ENERGY:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOTal:ENERGY:MAX?

PowerSwlossTotalEnergyMean

Bases: SCPICmdRead

The POWer:SWLoss:TOTal:ENERGY:MEAN command.

Description
  • Returns the mean total energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal:ENERGY:MEAN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal:ENERGY:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOTal:ENERGY:MEAN?

PowerSwlossTotalEnergyMin

Bases: SCPICmdRead

The POWer:SWLoss:TOTal:ENERGY:MIN command.

Description
  • Returns the minimum total energy for the switching loss calculation.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal:ENERGY:MIN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal:ENERGY:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOTal:ENERGY:MIN?

PowerSwlossTotalPower

Bases: SCPICmdRead

The POWer:SWLoss:TOTal:POWer command tree.

Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal:POWer? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal:POWer? query and raise an AssertionError if the returned value does not match value.
Properties
  • .max: The POWer:SWLoss:TOTal:POWer:MAX command.
  • .mean: The POWer:SWLoss:TOTal:POWer:MEAN command.
  • .min: The POWer:SWLoss:TOTal:POWer:MIN command.

max property

Return the POWer:SWLoss:TOTal:POWer:MAX command.

Description
  • Returns the maximum total power loss.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal:POWer:MAX? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal:POWer:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOTal:POWer:MAX?

mean property

Return the POWer:SWLoss:TOTal:POWer:MEAN command.

Description
  • Returns the mean total power loss.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal:POWer:MEAN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal:POWer:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOTal:POWer:MEAN?

min property

Return the POWer:SWLoss:TOTal:POWer:MIN command.

Description
  • Returns the minimum total power loss.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal:POWer:MIN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal:POWer:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOTal:POWer:MIN?

PowerSwlossTotalPowerMax

Bases: SCPICmdRead

The POWer:SWLoss:TOTal:POWer:MAX command.

Description
  • Returns the maximum total power loss.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal:POWer:MAX? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal:POWer:MAX? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOTal:POWer:MAX?

PowerSwlossTotalPowerMean

Bases: SCPICmdRead

The POWer:SWLoss:TOTal:POWer:MEAN command.

Description
  • Returns the mean total power loss.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal:POWer:MEAN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal:POWer:MEAN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOTal:POWer:MEAN?

PowerSwlossTotalPowerMin

Bases: SCPICmdRead

The POWer:SWLoss:TOTal:POWer:MIN command.

Description
  • Returns the minimum total power loss.
Usage
  • Using the .query() method will send the POWer:SWLoss:TOTal:POWer:MIN? query.
  • Using the .verify(value) method will send the POWer:SWLoss:TOTal:POWer:MIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- POWer:SWLoss:TOTal:POWer:MIN?

PowerSwlossVcesat

Bases: SCPICmdWrite, SCPICmdRead

The POWer:SWLoss:VCEsat command.

Description
  • This command specifies VCESAT value for use in switching loss calculations when the conduction calculation method is VCESAT.
Usage
  • Using the .query() method will send the POWer:SWLoss:VCEsat? query.
  • Using the .verify(value) method will send the POWer:SWLoss:VCEsat? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:SWLoss:VCEsat value command.
SCPI Syntax
- POWer:SWLoss:VCEsat <NR3>
- POWer:SWLoss:VCEsat?
Info
  • <NR3> is a floating point number that specifies the VCEsat switching loss calculation.

PowerType

Bases: SCPICmdWrite, SCPICmdRead

The POWer:TYPe command.

Description
  • This command specifies the power application measurement type.
Usage
  • Using the .query() method will send the POWer:TYPe? query.
  • Using the .verify(value) method will send the POWer:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:TYPe value command.
SCPI Syntax
- POWer:TYPe {NONe|QUALity|SWITCHingloss|SOA|HARMonics|RIPPle|MODULationanalysis|DESKew}
- POWer:TYPe?
Info
  • NONe Use to set the measurement type to None.
  • QUALity Use the power quality functions to obtain measurements and statistics about the general power quality in your test circuit.
  • SWITCHingloss Use the switching loss functions to obtain the power loss and energy loss across the acquired waveform, including turn-on loss, turn-off loss, conduction loss, and total loss. Typically, use these functions to characterize losses in power supply switching devices, as they switch on and off.
  • SOA Use the safe operating functions to obtain an X-Y display of the switching device-under-test’s voltage and current. Also use them to perform a mask test of the X-Y signal relative to the graphical X-Y description of the device specification table. The safe operating area is typically the voltage and current values that a semiconductor can operate without damaging itself.
  • HARMonics Use the harmonics functions to obtain the frequency spectrum of the source waveform and associated measurement values. Harmonic measurements can help one perform in-depth troubleshooting of power quality problems.
  • RIPPle Use the ripple functions to obtain measurements and statistics for the AC components of the acquired waveform. Ripples are often found on top of a large DC signal.
  • MODULationanalysis Use the modulation functions to obtain a trend plot of a measurement value across the acquired waveform. This is useful for showing the variations in the modulated switching signal.
  • DESKew Run the deskew procedure to match the delays through the probes. Different probes introduce different delays between the probe tip and the oscilloscope. Many oscilloscope users do not have to worry about this because they use the same type of probe on all channels. Power measurement users, however, frequently use both a voltage probe and a current probe. A current probe typically has a larger delay than a voltage probe, so setting deskew values becomes important.

PowerVoltagesource

Bases: SCPICmdWrite, SCPICmdRead

The POWer:VOLTAGESOurce command.

Description
  • This command specifies the voltage source for the power application.
Usage
  • Using the .query() method will send the POWer:VOLTAGESOurce? query.
  • Using the .verify(value) method will send the POWer:VOLTAGESOurce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the POWer:VOLTAGESOurce value command.
SCPI Syntax
- POWer:VOLTAGESOurce {CH<x>|REF<x>}
- POWer:VOLTAGESOurce?
Info
  • CH<x> sets the analog channel as the voltage source.
  • REF<x> sets the reference waveform as the voltage source.