bus

The bus commands module.

These commands are used in the following models: DPO7AX

THIS FILE IS AUTO-GENERATED, IT SHOULD NOT BE MANUALLY MODIFIED.

Please report an issue if one is found.

Commands and Queries
- BUS:ADDNew <QString>
- BUS:B<x>:ARINC429A:BITRate {LOW|HI|CUSTom}
- BUS:B<x>:ARINC429A:BITRate:CUSTom <NR1>
- BUS:B<x>:ARINC429A:BITRate:CUSTom?
- BUS:B<x>:ARINC429A:BITRate?
- BUS:B<x>:ARINC429A:DATAFORmat {DATA|SDIDATA|SDIDATASSM}
- BUS:B<x>:ARINC429A:DATAFORmat?
- BUS:B<x>:ARINC429A:POLARITY {NORMal|INVERTed}
- BUS:B<x>:ARINC429A:POLARITY?
- BUS:B<x>:ARINC429A:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ARINC429A:SOUrce?
- BUS:B<x>:ARINC429A:THRESHold <NR3>
- BUS:B<x>:ARINC429A:THRESHold?
- BUS:B<x>:AUDio:BITDelay <NR1>
- BUS:B<x>:AUDio:BITDelay?
- BUS:B<x>:AUDio:BITOrder {MSB|LSB}
- BUS:B<x>:AUDio:BITOrder?
- BUS:B<x>:AUDio:CLOCk:POLarity {FALL|RISE}
- BUS:B<x>:AUDio:CLOCk:POLarity?
- BUS:B<x>:AUDio:CLOCk:SOUrce {CH<x>|CH<x>_D<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:AUDio:CLOCk:SOUrce?
- BUS:B<x>:AUDio:CLOCk:THReshold <NR3>
- BUS:B<x>:AUDio:CLOCk:THReshold?
- BUS:B<x>:AUDio:DATa:POLarity {HIGH|LOW}
- BUS:B<x>:AUDio:DATa:POLarity?
- BUS:B<x>:AUDio:DATa:SIZe <NR1>
- BUS:B<x>:AUDio:DATa:SIZe?
- BUS:B<x>:AUDio:DATa:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:AUDio:DATa:SOUrce?
- BUS:B<x>:AUDio:DATa:THReshold <NR3>
- BUS:B<x>:AUDio:DATa:THReshold?
- BUS:B<x>:AUDio:DATa:WORDSize <NR1>
- BUS:B<x>:AUDio:DATa:WORDSize?
- BUS:B<x>:AUDio:FRAME:CLOCKBITSPERCHANNEL <NR1>
- BUS:B<x>:AUDio:FRAME:CLOCKBITSPERCHANNEL?
- BUS:B<x>:AUDio:FRAME:SIZe <NR1>
- BUS:B<x>:AUDio:FRAME:SIZe?
- BUS:B<x>:AUDio:TYPe {I2S|LJ|RJ|TDM}
- BUS:B<x>:AUDio:TYPe?
- BUS:B<x>:AUDio:WORDSel:POLarity {NORMal|INVERTed}
- BUS:B<x>:AUDio:WORDSel:POLarity?
- BUS:B<x>:AUDio:WORDSel:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:AUDio:WORDSel:SOUrce?
- BUS:B<x>:AUDio:WORDSel:THReshold <NR3>
- BUS:B<x>:AUDio:WORDSel:THReshold?
- BUS:B<x>:AUTOETHERnet:BITRate <NR1>
- BUS:B<x>:AUTOETHERnet:BITRate?
- BUS:B<x>:AUTOETHERnet:DATAMINUSTHRESHOLD <NR3>
- BUS:B<x>:AUTOETHERnet:DATAMINUSTHRESHOLD?
- BUS:B<x>:AUTOETHERnet:DATAPLUSTHRESHold <NR3>
- BUS:B<x>:AUTOETHERnet:DATAPLUSTHRESHold?
- BUS:B<x>:AUTOETHERnet:LOWDATAMINus <NR3>
- BUS:B<x>:AUTOETHERnet:LOWDATAMINus?
- BUS:B<x>:AUTOETHERnet:LOWDATAPLUS <NR3>
- BUS:B<x>:AUTOETHERnet:LOWDATAPLUS?
- BUS:B<x>:AUTOETHERnet:LOWTHRESHold <NR3>
- BUS:B<x>:AUTOETHERnet:LOWTHRESHold?
- BUS:B<x>:AUTOETHERnet:PACKetview {ON|OFF}
- BUS:B<x>:AUTOETHERnet:PACKetview?
- BUS:B<x>:AUTOETHERnet:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:AUTOETHERnet:SIGNALTYpe?
- BUS:B<x>:AUTOETHERnet:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:AUTOETHERnet:SOUrce:DMINus {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:AUTOETHERnet:SOUrce:DMINus?
- BUS:B<x>:AUTOETHERnet:SOUrce:DPLUs {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:AUTOETHERnet:SOUrce:DPLUs?
- BUS:B<x>:AUTOETHERnet:SOUrce?
- BUS:B<x>:AUTOETHERnet:TBT1s:MTHReshold <NR3>
- BUS:B<x>:AUTOETHERnet:TBT1s:MTHReshold?
- BUS:B<x>:AUTOETHERnet:TENBASET1S:THRESHold <NR3>
- BUS:B<x>:AUTOETHERnet:TENBASET1S:THRESHold?
- BUS:B<x>:AUTOETHERnet:THRESHold <NR3>
- BUS:B<x>:AUTOETHERnet:THRESHold?
- BUS:B<x>:AUTOETHERnet:TYPe {HUNDREDBASET1}
- BUS:B<x>:AUTOETHERnet:TYPe?
- BUS:B<x>:CAN:BITRate {RATE10K|RATE100K|RATE1M|RATE125K|RATE153K|RATE20K|RATE25K|RATE250K|RATE31K|RATE33K|RATE400K|RATE50K|RATE500K|RATE62K|RATE68K|RATE800K|RATE83K|RATE92K|CUSTom}
- BUS:B<x>:CAN:BITRate:VALue <NR3>
- BUS:B<x>:CAN:BITRate:VALue?
- BUS:B<x>:CAN:BITRate?
- BUS:B<x>:CAN:FD:BITRate {RATE1M|RATE2M|RATE3M|RATE4M|RATE5M|RATE6M|RATE7M|RATE8M|RATE9M|RATE10M|RATE11M|RATE12M|RATE13M|RATE14M|RATE15M|RATE16M|CUSTom}
- BUS:B<x>:CAN:FD:BITRate:CUSTom <NR1>
- BUS:B<x>:CAN:FD:BITRate:CUSTom?
- BUS:B<x>:CAN:FD:BITRate?
- BUS:B<x>:CAN:SAMPLEpoint <NR1>
- BUS:B<x>:CAN:SAMPLEpoint?
- BUS:B<x>:CAN:SIGNal {DIFFerential|CANH|CANL|RX|TX}
- BUS:B<x>:CAN:SIGNal?
- BUS:B<x>:CAN:SOUrce {CH<x>|CH<x>_D<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:CAN:SOUrce?
- BUS:B<x>:CAN:STANDard {CAN2X|CAN2X_FDISO_XL|CAN_XL|FDISO|FDNONISO}
- BUS:B<x>:CAN:STANDard?
- BUS:B<x>:CAN:THReshold <NR3>
- BUS:B<x>:CAN:THReshold?
- BUS:B<x>:CAN:XL:BITRate <NR1>
- BUS:B<x>:CAN:XL:BITRate:CUSTom <NR1>
- BUS:B<x>:CAN:XL:BITRate:CUSTom?
- BUS:B<x>:CAN:XL:BITRate?
- BUS:B<x>:CPHY:A:DATA:THRESHold <NR3>
- BUS:B<x>:CPHY:A:DATA:THRESHold?
- BUS:B<x>:CPHY:A:LP:THRESHold <NR3>
- BUS:B<x>:CPHY:A:LP:THRESHold?
- BUS:B<x>:CPHY:A:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:A:SOUrce?
- BUS:B<x>:CPHY:AB:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:AB:SOUrce?
- BUS:B<x>:CPHY:AB:THReshold <NR3>
- BUS:B<x>:CPHY:AB:THReshold?
- BUS:B<x>:CPHY:AGND:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:AGND:SOUrce?
- BUS:B<x>:CPHY:AGND:THReshold <NR3>
- BUS:B<x>:CPHY:AGND:THReshold?
- BUS:B<x>:CPHY:B:DATA:THRESHold <NR3>
- BUS:B<x>:CPHY:B:DATA:THRESHold?
- BUS:B<x>:CPHY:B:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:B:SOUrce?
- BUS:B<x>:CPHY:BC:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:BC:SOUrce?
- BUS:B<x>:CPHY:BC:THReshold <NR3>
- BUS:B<x>:CPHY:BC:THReshold?
- BUS:B<x>:CPHY:BITRate <NR1>
- BUS:B<x>:CPHY:BITRate?
- BUS:B<x>:CPHY:C:DATA:THRESHold <NR3>
- BUS:B<x>:CPHY:C:DATA:THRESHold?
- BUS:B<x>:CPHY:C:LP:THRESHold <NR3>
- BUS:B<x>:CPHY:C:LP:THRESHold?
- BUS:B<x>:CPHY:C:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:C:SOUrce?
- BUS:B<x>:CPHY:CA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:CA:SOUrce?
- BUS:B<x>:CPHY:CA:THReshold <NR3>
- BUS:B<x>:CPHY:CA:THReshold?
- BUS:B<x>:CPHY:CGND:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:CGND:SOUrce?
- BUS:B<x>:CPHY:CGND:THReshold <NR3>
- BUS:B<x>:CPHY:CGND:THReshold?
- BUS:B<x>:CPHY:LP:DIRection {forward|reverse}
- BUS:B<x>:CPHY:LP:DIRection?
- BUS:B<x>:CPHY:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:CPHY:SIGNALTYpe?
- BUS:B<x>:CPHY:SUBTYPe {CSI|DSI|Word|Symbol}
- BUS:B<x>:CPHY:SUBTYPe?
- BUS:B<x>:CXPI:BITRate <NR1>
- BUS:B<x>:CXPI:BITRate?
- BUS:B<x>:CXPI:REC:THReshold <NR3>
- BUS:B<x>:CXPI:REC:THReshold?
- BUS:B<x>:CXPI:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CXPI:SOUrce?
- BUS:B<x>:DISplay:FORMat {HEX|BINARY|MIXEDASCII|MIXEDHEX|ASCII|DECIMAL|MIXED}
- BUS:B<x>:DISplay:FORMat?
- BUS:B<x>:DISplay:LAYout {BUS|BUSANDWAVEFORM}
- BUS:B<x>:DISplay:LAYout?
- BUS:B<x>:DPHY:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:DPHY:CLOCk:SOUrce?
- BUS:B<x>:DPHY:CLOCk:THRESHold <NR3>
- BUS:B<x>:DPHY:CLOCk:THRESHold?
- BUS:B<x>:DPHY:DMINus:DATATHRESHold <NR3>
- BUS:B<x>:DPHY:DMINus:DATATHRESHold?
- BUS:B<x>:DPHY:DMINus:LPTHRESHold <NR3>
- BUS:B<x>:DPHY:DMINus:LPTHRESHold?
- BUS:B<x>:DPHY:DMINus:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:DPHY:DMINus:SOUrce?
- BUS:B<x>:DPHY:DPlus:DATATHRESHold <NR3>
- BUS:B<x>:DPHY:DPlus:DATATHRESHold?
- BUS:B<x>:DPHY:DPlus:LPTHRESHold <NR3>
- BUS:B<x>:DPHY:DPlus:LPTHRESHold?
- BUS:B<x>:DPHY:DPlus:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:DPHY:DPlus:SOUrce?
- BUS:B<x>:DPHY:LP:DIRection {forward|reverse}
- BUS:B<x>:DPHY:LP:DIRection?
- BUS:B<x>:DPHY:PROTocol:TYPe {CSI|DSI}
- BUS:B<x>:DPHY:PROTocol:TYPe?
- BUS:B<x>:DPHY:SIGNal:ENCoding {false|true}
- BUS:B<x>:DPHY:SIGNal:ENCoding?
- BUS:B<x>:ESPI:ALERTVIEW {ON|OFF}
- BUS:B<x>:ESPI:ALERTVIEW?
- BUS:B<x>:ESPI:ALERt:POLarity {HIGH|LOW}
- BUS:B<x>:ESPI:ALERt:POLarity?
- BUS:B<x>:ESPI:ALERt:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ESPI:ALERt:SOUrce?
- BUS:B<x>:ESPI:ALERt:THReshold <NR3>
- BUS:B<x>:ESPI:ALERt:THReshold?
- BUS:B<x>:ESPI:CHIPSELect:POLarity {HIGH|LOW}
- BUS:B<x>:ESPI:CHIPSELect:POLarity?
- BUS:B<x>:ESPI:CHIPSELect:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ESPI:CHIPSELect:SOUrce?
- BUS:B<x>:ESPI:CHIPSELect:THReshold <NR3>
- BUS:B<x>:ESPI:CHIPSELect:THReshold?
- BUS:B<x>:ESPI:CLOCk:POLarity {FALL|RISE}
- BUS:B<x>:ESPI:CLOCk:POLarity?
- BUS:B<x>:ESPI:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ESPI:CLOCk:SOUrce?
- BUS:B<x>:ESPI:CLOCk:THReshold <NR3>
- BUS:B<x>:ESPI:CLOCk:THReshold?
- BUS:B<x>:ESPI:DATAONE:POLarity {HIGH|LOW}
- BUS:B<x>:ESPI:DATAONE:POLarity?
- BUS:B<x>:ESPI:DATAONE:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ESPI:DATAONE:SOUrce?
- BUS:B<x>:ESPI:DATAONE:THReshold <NR3>
- BUS:B<x>:ESPI:DATAONE:THReshold?
- BUS:B<x>:ESPI:DATATWO:POLarity {HIGH|LOW}
- BUS:B<x>:ESPI:DATATWO:POLarity?
- BUS:B<x>:ESPI:DATATWO:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ESPI:DATATWO:SOUrce?
- BUS:B<x>:ESPI:DATATWO:THReshold <NR3>
- BUS:B<x>:ESPI:DATATWO:THReshold?
- BUS:B<x>:ESPI:IOMODe {SINGle|DUAL}
- BUS:B<x>:ESPI:IOMODe?
- BUS:B<x>:ETHERCAT:DATAMINUSTHRESHold <NR3>
- BUS:B<x>:ETHERCAT:DATAMINUSTHRESHold?
- BUS:B<x>:ETHERCAT:DATAPLUSTHRESHold <NR3>
- BUS:B<x>:ETHERCAT:DATAPLUSTHRESHold?
- BUS:B<x>:ETHERCAT:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:ETHERCAT:SIGNALTYpe?
- BUS:B<x>:ETHERCAT:SOUrce:DIFF {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERCAT:SOUrce:DIFF?
- BUS:B<x>:ETHERCAT:SOUrce:DMINus {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERCAT:SOUrce:DMINus?
- BUS:B<x>:ETHERCAT:SOUrce:DPLUs {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERCAT:SOUrce:DPLUs?
- BUS:B<x>:ETHERCAT:THRESHold <NR3>
- BUS:B<x>:ETHERCAT:THRESHold?
- BUS:B<x>:ETHERnet:DATAMINUSTHRESHold <NR3>
- BUS:B<x>:ETHERnet:DATAMINUSTHRESHold?
- BUS:B<x>:ETHERnet:DATAPLUSTHRESHold <NR3>
- BUS:B<x>:ETHERnet:DATAPLUSTHRESHold?
- BUS:B<x>:ETHERnet:IPVFOUR {YES|NO}
- BUS:B<x>:ETHERnet:IPVFOUR?
- BUS:B<x>:ETHERnet:LOWTHRESHold <NR3>
- BUS:B<x>:ETHERnet:LOWTHRESHold?
- BUS:B<x>:ETHERnet:QTAGGING {YES|NO}
- BUS:B<x>:ETHERnet:QTAGGING?
- BUS:B<x>:ETHERnet:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:ETHERnet:SIGNALTYpe?
- BUS:B<x>:ETHERnet:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERnet:SOUrce:DMINus {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERnet:SOUrce:DMINus?
- BUS:B<x>:ETHERnet:SOUrce:DPLUs {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERnet:SOUrce:DPLUs?
- BUS:B<x>:ETHERnet:SOUrce?
- BUS:B<x>:ETHERnet:THRESHold <NR3>
- BUS:B<x>:ETHERnet:THRESHold?
- BUS:B<x>:ETHERnet:TYPe {TENBASET|HUNDREDBASETX}
- BUS:B<x>:ETHERnet:TYPe?
- BUS:B<x>:EUSB:BITRate {HIGH|FULL|LOW}
- BUS:B<x>:EUSB:BITRate?
- BUS:B<x>:EUSB:DATAMINUS:DATA:THRESHold <NR3>
- BUS:B<x>:EUSB:DATAMINUS:DATA:THRESHold?
- BUS:B<x>:EUSB:DATAMINUSTHRESHold <NR3>
- BUS:B<x>:EUSB:DATAMINUSTHRESHold?
- BUS:B<x>:EUSB:DATAPLUS:DATA:THRESHold <NR3>
- BUS:B<x>:EUSB:DATAPLUS:DATA:THRESHold?
- BUS:B<x>:EUSB:DATAPLUSTHRESHold <NR3>
- BUS:B<x>:EUSB:DATAPLUSTHRESHold?
- BUS:B<x>:EUSB:LOWTHRESHold <NR3>
- BUS:B<x>:EUSB:LOWTHRESHold?
- BUS:B<x>:EUSB:OPERating:MODe {NATive|REPEATERHOSt|REPEATERPERIPHERAL}
- BUS:B<x>:EUSB:OPERating:MODe?
- BUS:B<x>:EUSB:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:EUSB:SIGNALTYpe?
- BUS:B<x>:EUSB:SOUrce:DIFF {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:EUSB:SOUrce:DIFF?
- BUS:B<x>:EUSB:SOUrce:DMINus {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:EUSB:SOUrce:DMINus?
- BUS:B<x>:EUSB:SOUrce:DPLUs {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:EUSB:SOUrce:DPLUs?
- BUS:B<x>:EUSB:THRESHold <NR3>
- BUS:B<x>:EUSB:THRESHold?
- BUS:B<x>:FLEXray:BITRate {CUSTOM|RATE2M|RATE5M|RATE10M}
- BUS:B<x>:FLEXray:BITRate:CUSTom <NR1>
- BUS:B<x>:FLEXray:BITRate:CUSTom?
- BUS:B<x>:FLEXray:BITRate?
- BUS:B<x>:FLEXray:CHannel {A|B}
- BUS:B<x>:FLEXray:CHannel?
- BUS:B<x>:FLEXray:LOWTHRESHold <NR3>
- BUS:B<x>:FLEXray:LOWTHRESHold?
- BUS:B<x>:FLEXray:SIGnal {BDIFFBP|BM|TXRX}
- BUS:B<x>:FLEXray:SIGnal?
- BUS:B<x>:FLEXray:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:FLEXray:SOUrce:TXRX {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:FLEXray:SOUrce:TXRX?
- BUS:B<x>:FLEXray:SOUrce?
- BUS:B<x>:FLEXray:THRESHold <NR3>
- BUS:B<x>:FLEXray:THRESHold?
- BUS:B<x>:FLEXray:TXRXTHRESHold <NR3>
- BUS:B<x>:FLEXray:TXRXTHRESHold?
- BUS:B<x>:I2C:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:I2C:CLOCk:SOUrce?
- BUS:B<x>:I2C:CLOCk:THReshold <NR3>
- BUS:B<x>:I2C:CLOCk:THReshold?
- BUS:B<x>:I2C:DATa:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:I2C:DATa:SOUrce?
- BUS:B<x>:I2C:DATa:THReshold <NR3>
- BUS:B<x>:I2C:DATa:THReshold?
- BUS:B<x>:I2C:RWINADDR {0|1}
- BUS:B<x>:I2C:RWINADDR?
- BUS:B<x>:I3C:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:I3C:CLOCk:SOUrce?
- BUS:B<x>:I3C:CLOCk:THReshold <NR3>
- BUS:B<x>:I3C:CLOCk:THReshold?
- BUS:B<x>:I3C:DATa:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:I3C:DATa:SOUrce?
- BUS:B<x>:I3C:DATa:THReshold <NR3>
- BUS:B<x>:I3C:DATa:THReshold?
- BUS:B<x>:I3C:VERSion {VR1<x>}
- BUS:B<x>:I3C:VERSion?
- BUS:B<x>:LABel:COLor <QString>
- BUS:B<x>:LABel:COLor?
- BUS:B<x>:LABel:FONT:BOLD {ON|OFF|1|0}
- BUS:B<x>:LABel:FONT:BOLD?
- BUS:B<x>:LABel:FONT:ITALic {ON|OFF|1|0}
- BUS:B<x>:LABel:FONT:ITALic?
- BUS:B<x>:LABel:FONT:SIZE <NR1>
- BUS:B<x>:LABel:FONT:SIZE?
- BUS:B<x>:LABel:FONT:TYPE <QString>
- BUS:B<x>:LABel:FONT:TYPE?
- BUS:B<x>:LABel:FONT:UNDERline {ON|OFF|1|0}
- BUS:B<x>:LABel:FONT:UNDERline?
- BUS:B<x>:LABel:XPOS <NR3>
- BUS:B<x>:LABel:XPOS?
- BUS:B<x>:LABel:YPOS <NR3>
- BUS:B<x>:LABel:YPOS?
- BUS:B<x>:LABel:name <QString>
- BUS:B<x>:LABel:name?
- BUS:B<x>:LIN:BITRate {RATE10K|RATE1K|RATE19K|RATE2K|RATE4K|RATE9K|CUSTom}
- BUS:B<x>:LIN:BITRate:CUSTom <NR1>
- BUS:B<x>:LIN:BITRate:CUSTom?
- BUS:B<x>:LIN:BITRate?
- BUS:B<x>:LIN:IDFORmat {NOPARity|PARity}
- BUS:B<x>:LIN:IDFORmat?
- BUS:B<x>:LIN:POLarity {INVerted|NORmal}
- BUS:B<x>:LIN:POLarity?
- BUS:B<x>:LIN:SAMPLEpoint <NR1>
- BUS:B<x>:LIN:SAMPLEpoint?
- BUS:B<x>:LIN:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:LIN:SOUrce:THReshold <NR3>
- BUS:B<x>:LIN:SOUrce:THReshold?
- BUS:B<x>:LIN:SOUrce?
- BUS:B<x>:LIN:STANDard {MIXed|V1X|V2X}
- BUS:B<x>:LIN:STANDard?
- BUS:B<x>:MANChester:BITORDer {LSB|MSB}
- BUS:B<x>:MANChester:BITORDer?
- BUS:B<x>:MANChester:BITRate <NR1>
- BUS:B<x>:MANChester:BITRate?
- BUS:B<x>:MANChester:DISplaymode {BITS|PACKET}
- BUS:B<x>:MANChester:DISplaymode?
- BUS:B<x>:MANChester:HEADer:LENGth <NR1>
- BUS:B<x>:MANChester:HEADer:LENGth?
- BUS:B<x>:MANChester:IDLE:BITS <NR1>
- BUS:B<x>:MANChester:IDLE:BITS?
- BUS:B<x>:MANChester:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:MANChester:SOUrce?
- BUS:B<x>:MANChester:START:INDex <NR1>
- BUS:B<x>:MANChester:START:INDex?
- BUS:B<x>:MANChester:SYNC:SIZe <NR1>
- BUS:B<x>:MANChester:SYNC:SIZe?
- BUS:B<x>:MANChester:THReshold <NR3>
- BUS:B<x>:MANChester:THReshold?
- BUS:B<x>:MANChester:TOLerance <NR3>
- BUS:B<x>:MANChester:TOLerance?
- BUS:B<x>:MANChester:TRANstion:ZERo {FALLing|RISing}
- BUS:B<x>:MANChester:TRANstion:ZERo?
- BUS:B<x>:MANChester:TRAiler:LENGth <NR1>
- BUS:B<x>:MANChester:TRAiler:LENGth?
- BUS:B<x>:MANChester:WORD:COUNt <NR1>
- BUS:B<x>:MANChester:WORD:COUNt?
- BUS:B<x>:MANChester:WORDSIZe <NR1>
- BUS:B<x>:MANChester:WORDSIZe?
- BUS:B<x>:MANChester:parity {ODD|EVEN|NONE}
- BUS:B<x>:MANChester:parity?
- BUS:B<x>:MDIO:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:MDIO:CLOCk:SOUrce?
- BUS:B<x>:MDIO:CLOCk:THReshold <NR3>
- BUS:B<x>:MDIO:CLOCk:THReshold?
- BUS:B<x>:MDIO:DATA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:MDIO:DATA:SOUrce?
- BUS:B<x>:MDIO:DATA:THReshold <NR3>
- BUS:B<x>:MDIO:DATA:THReshold?
- BUS:B<x>:MIL1553B:LOWTHRESHold <NR3>
- BUS:B<x>:MIL1553B:LOWTHRESHold?
- BUS:B<x>:MIL1553B:POLarity {NORMal|INVERTed}
- BUS:B<x>:MIL1553B:POLarity?
- BUS:B<x>:MIL1553B:RESPonsetime:MAXimum <NR3>
- BUS:B<x>:MIL1553B:RESPonsetime:MAXimum?
- BUS:B<x>:MIL1553B:RESPonsetime:MINimum <NR3>
- BUS:B<x>:MIL1553B:RESPonsetime:MINimum?
- BUS:B<x>:MIL1553B:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:MIL1553B:SOUrce?
- BUS:B<x>:MIL1553B:THRESHold <NR3>
- BUS:B<x>:MIL1553B:THRESHold?
- BUS:B<x>:NFC:APPLYDEMod {1|0}
- BUS:B<x>:NFC:APPLYDEMod?
- BUS:B<x>:NFC:BITRate <NR1>
- BUS:B<x>:NFC:BITRate?
- BUS:B<x>:NFC:CMD:STARTINDex <NR1>
- BUS:B<x>:NFC:CMD:STARTINDex?
- BUS:B<x>:NFC:CMD:THReshold <NR3>
- BUS:B<x>:NFC:CMD:THReshold?
- BUS:B<x>:NFC:CMD:TRANsition:ZERo {RISing|FALLing}
- BUS:B<x>:NFC:CMD:TRANsition:ZERo?
- BUS:B<x>:NFC:COMMand:POLarity {NORMal|INVerted}
- BUS:B<x>:NFC:COMMand:POLarity?
- BUS:B<x>:NFC:RESPonse:POLarity {NORMal|INVerted}
- BUS:B<x>:NFC:RESPonse:POLarity?
- BUS:B<x>:NFC:RSP:THReshold <NR3>
- BUS:B<x>:NFC:RSP:THReshold?
- BUS:B<x>:NFC:SOUrce {CH<x>}
- BUS:B<x>:NFC:SOUrce?
- BUS:B<x>:NFC:START:INDex <NR1>
- BUS:B<x>:NFC:START:INDex?
- BUS:B<x>:NFC:STD {NFC15693|NFC14443A|NFC14443B|NFCFELica}
- BUS:B<x>:NFC:STD?
- BUS:B<x>:NFC:TOLerance <NR3>
- BUS:B<x>:NFC:TOLerance?
- BUS:B<x>:NFC:TRANsition:ZERo {ge|ieee}
- BUS:B<x>:NFC:TRANsition:ZERo?
- BUS:B<x>:NRZ:BITOrder {LSB|MSB}
- BUS:B<x>:NRZ:BITOrder?
- BUS:B<x>:NRZ:BITRate <NR1>
- BUS:B<x>:NRZ:BITRate?
- BUS:B<x>:NRZ:POLarity {INVerted|NORmal}
- BUS:B<x>:NRZ:POLarity?
- BUS:B<x>:NRZ:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:NRZ:SOUrce?
- BUS:B<x>:NRZ:SPMI:VERsion v<x>
- BUS:B<x>:NRZ:SPMI:VERsion?
- BUS:B<x>:NRZ:THReshold <NR3>
- BUS:B<x>:NRZ:THReshold?
- BUS:B<x>:ONEWIRe:DATA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ONEWIRe:DATA:SOUrce?
- BUS:B<x>:ONEWIRe:DATA:THReshold <NR3>
- BUS:B<x>:ONEWIRe:DATA:THReshold?
- BUS:B<x>:ONEWIRe:MODe {STAndard|OVErdrive}
- BUS:B<x>:ONEWIRe:MODe?
- BUS:B<x>:PARallel:ALLTHResholds <NR3>
- BUS:B<x>:PARallel:ALLTHResholds:APPly
- BUS:B<x>:PARallel:BIT<x>SOUrce {CH<x>|CH<x>_D<x>|MATH<x>|REF<x>|REF<x>_D<x>|NONE}
- BUS:B<x>:PARallel:BIT<x>SOUrce:THReshold <NR3>
- BUS:B<x>:PARallel:CLOCKSOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>|NONE}
- BUS:B<x>:PARallel:CLOCKSOUrce:THReshold <NR3>
- BUS:B<x>:PARallel:CLOCKSOUrce:THReshold?
- BUS:B<x>:PARallel:CLOCKSOUrce?
- BUS:B<x>:PARallel:CLOCk:EDGE {FALLING|RISING|EITHER}
- BUS:B<x>:PARallel:CLOCk:EDGE?
- BUS:B<x>:PARallel:CLOCk:ISCLOCKED {ON|OFF|<NR1>}
- BUS:B<x>:PARallel:CLOCk:ISCLOCKED?
- BUS:B<x>:PCIE:DATAMinus:THRESHold:HIGh <NR3>
- BUS:B<x>:PCIE:DATAMinus:THRESHold:HIGh?
- BUS:B<x>:PCIE:DATAPlus:THRESHold:HIGh <NR3>
- BUS:B<x>:PCIE:DATAPlus:THRESHold:HIGh?
- BUS:B<x>:PCIE:MTHReshold <NR3>
- BUS:B<x>:PCIE:MTHReshold?
- BUS:B<x>:PCIE:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:PCIE:SIGNALTYpe?
- BUS:B<x>:PCIE:SOUrce:DMINus {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:PCIE:SOUrce:DMINus?
- BUS:B<x>:PCIE:SOUrce:DPLUs {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:PCIE:SOUrce:DPLUs?
- BUS:B<x>:PSIFIVe:BITPERiod <NR1>
- BUS:B<x>:PSIFIVe:BITPERiod?
- BUS:B<x>:PSIFIVe:BITRate {RATE125K|RATE189K|RATE83K}
- BUS:B<x>:PSIFIVe:BITRate?
- BUS:B<x>:PSIFIVe:COMM:DIRection {SENSORECU|ECUSENSor}
- BUS:B<x>:PSIFIVe:COMM:DIRection?
- BUS:B<x>:PSIFIVe:DATAA <NR1>
- BUS:B<x>:PSIFIVe:DATAA?
- BUS:B<x>:PSIFIVe:DATAB <NR1>
- BUS:B<x>:PSIFIVe:DATAB?
- BUS:B<x>:PSIFIVe:DATAFORMat {NIBBLe|BYTe}
- BUS:B<x>:PSIFIVe:DATAFORMat?
- BUS:B<x>:PSIFIVe:ECUSOURce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:PSIFIVe:ECUSOURce?
- BUS:B<x>:PSIFIVe:FRAMECONTrol <NR1>
- BUS:B<x>:PSIFIVe:FRAMECONTrol?
- BUS:B<x>:PSIFIVe:MESSaging {OFF|ON}
- BUS:B<x>:PSIFIVe:MESSaging?
- BUS:B<x>:PSIFIVe:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:PSIFIVe:SOUrce?
- BUS:B<x>:PSIFIVe:STATus <NR1>
- BUS:B<x>:PSIFIVe:STATus?
- BUS:B<x>:PSIFIVe:SYNCMODe {PULSEWIDTh|TOOTHGAP}
- BUS:B<x>:PSIFIVe:SYNCMODe?
- BUS:B<x>:PSIFIVe:SYNCTHRESHold <NR3>
- BUS:B<x>:PSIFIVe:SYNCTHRESHold?
- BUS:B<x>:PSIFIVe:THRESHold <NR3>
- BUS:B<x>:PSIFIVe:THRESHold?
- BUS:B<x>:RS232C:BITORDer {LSB|MSB}
- BUS:B<x>:RS232C:BITORDer?
- BUS:B<x>:RS232C:BITRate {CUSTOM|RATE300|RATE1K|RATE2K|RATE9K|RATE19K|RATE38K|RATE115K|RATE921K}
- BUS:B<x>:RS232C:BITRate:CUSTom <NR1>
- BUS:B<x>:RS232C:BITRate:CUSTom?
- BUS:B<x>:RS232C:BITRate?
- BUS:B<x>:RS232C:DATABits {7|8|9}
- BUS:B<x>:RS232C:DATABits?
- BUS:B<x>:RS232C:DELIMiter {NULl|CR|LF|SPace|XFF}
- BUS:B<x>:RS232C:DELIMiter?
- BUS:B<x>:RS232C:DISplaymode {FRame|PACKET}
- BUS:B<x>:RS232C:DISplaymode?
- BUS:B<x>:RS232C:PARity {NONe|EVEN|ODD}
- BUS:B<x>:RS232C:PARity?
- BUS:B<x>:RS232C:POLarity {NORmal|INVERTed}
- BUS:B<x>:RS232C:POLarity?
- BUS:B<x>:RS232C:SOUrce {CH<x>|REF<x>|MATH<x>|REF<x>_D<x>}
- BUS:B<x>:RS232C:SOUrce:THReshold <NR3>
- BUS:B<x>:RS232C:SOUrce:THReshold?
- BUS:B<x>:RS232C:SOUrce?
- BUS:B<x>:S8B10B:BITRate
- BUS:B<x>:S8B10B:BITRate?
- BUS:B<x>:S8B10B:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:S8B10B:SOUrce?
- BUS:B<x>:S8B10B:THReshold <NR3>
- BUS:B<x>:S8B10B:THReshold?
- BUS:B<x>:SDLC:BITRate <NR1>
- BUS:B<x>:SDLC:BITRate?
- BUS:B<x>:SDLC:DATA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SDLC:DATA:SOUrce?
- BUS:B<x>:SDLC:DATA:THReshold <NR3>
- BUS:B<x>:SDLC:DATA:THReshold?
- BUS:B<x>:SDLC:ENCoding {DISCrete|INVert}
- BUS:B<x>:SDLC:ENCoding?
- BUS:B<x>:SDLC:MODulo {8|128}
- BUS:B<x>:SDLC:MODulo?
- BUS:B<x>:SENT:CHANWidth {TWELVEtwelve|FOURTEENten|SIXTEENeight}
- BUS:B<x>:SENT:CHANWidth?
- BUS:B<x>:SENT:NIBBLECount {THREE|FOUR|SIX}
- BUS:B<x>:SENT:NIBBLECount?
- BUS:B<x>:SENT:NUMCHANnel {ONE|TWO}
- BUS:B<x>:SENT:NUMCHANnel?
- BUS:B<x>:SENT:PAUSEPULSe {NO|YES}
- BUS:B<x>:SENT:PAUSEPULSe?
- BUS:B<x>:SENT:POLARITY {INVerted|NORmal}
- BUS:B<x>:SENT:POLARITY?
- BUS:B<x>:SENT:SLOW {NOne|ENHANCED4|ENHANCED8|SHOrt}
- BUS:B<x>:SENT:SLOW?
- BUS:B<x>:SENT:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SENT:SOUrce?
- BUS:B<x>:SENT:THRESHold <NR3>
- BUS:B<x>:SENT:THRESHold?
- BUS:B<x>:SENT:TICKTIME <NR3>
- BUS:B<x>:SENT:TICKTIME?
- BUS:B<x>:SENT:TICKTOLerance <NR3>
- BUS:B<x>:SENT:TICKTOLerance?
- BUS:B<x>:SMBUS:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SMBUS:CLOCk:SOUrce?
- BUS:B<x>:SMBUS:CLOCk:THReshold <NR3>
- BUS:B<x>:SMBUS:CLOCk:THReshold?
- BUS:B<x>:SMBUS:DATA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SMBUS:DATA:SOUrce?
- BUS:B<x>:SMBUS:DATA:THReshold <NR3>
- BUS:B<x>:SMBUS:DATA:THReshold?
- BUS:B<x>:SMBUS:PEC:VALUe {TRUe|FALSe}
- BUS:B<x>:SMBUS:PEC:VALUe?
- BUS:B<x>:SPACEWIRe:BITRate <NR3>
- BUS:B<x>:SPACEWIRe:BITRate?
- BUS:B<x>:SPACEWIRe:DATa:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SPACEWIRe:DATa:SOUrce?
- BUS:B<x>:SPACEWIRe:DATa:THReshold <NR3>
- BUS:B<x>:SPACEWIRe:DATa:THReshold?
- BUS:B<x>:SPACEWIRe:DECode:TYPe {STRObe|DATARate}
- BUS:B<x>:SPACEWIRe:DECode:TYPe?
- BUS:B<x>:SPACEWIRe:STRobe:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SPACEWIRe:STRobe:SOUrce?
- BUS:B<x>:SPACEWIRe:STRobe:THReshold <NR3>
- BUS:B<x>:SPACEWIRe:STRobe:THReshold?
- BUS:B<x>:SPACEWIRe:SYNC {DATA|NULL|AUTO|CUSTom}
- BUS:B<x>:SPACEWIRe:SYNC:COUnt <NR1>
- BUS:B<x>:SPACEWIRe:SYNC:COUnt?
- BUS:B<x>:SPACEWIRe:SYNC:PATTern <NR3>
- BUS:B<x>:SPACEWIRe:SYNC:PATTern?
- BUS:B<x>:SPACEWIRe:SYNC:VALUe <NR3>
- BUS:B<x>:SPACEWIRe:SYNC:VALUe?
- BUS:B<x>:SPACEWIRe:SYNC?
- BUS:B<x>:SPI:BITOrder {LSB|MSB}
- BUS:B<x>:SPI:BITOrder?
- BUS:B<x>:SPI:CLOCk:POLarity {FALL|RISE}
- BUS:B<x>:SPI:CLOCk:POLarity?
- BUS:B<x>:SPI:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SPI:CLOCk:SOUrce?
- BUS:B<x>:SPI:CLOCk:THReshold <NR3>
- BUS:B<x>:SPI:CLOCk:THReshold?
- BUS:B<x>:SPI:DATa:POLarity {HIGH|LOW}
- BUS:B<x>:SPI:DATa:POLarity?
- BUS:B<x>:SPI:DATa:SIZe <NR1>
- BUS:B<x>:SPI:DATa:SIZe?
- BUS:B<x>:SPI:DATa:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SPI:DATa:SOUrce?
- BUS:B<x>:SPI:DATa:THReshold <NR3>
- BUS:B<x>:SPI:DATa:THReshold?
- BUS:B<x>:SPI:FRAMING {IDLE|SS}
- BUS:B<x>:SPI:FRAMING?
- BUS:B<x>:SPI:IDLETime <NR3>
- BUS:B<x>:SPI:IDLETime?
- BUS:B<x>:SPI:MISo:DATa:POLarity {HIGH|LOW}
- BUS:B<x>:SPI:MISo:DATa:POLarity?
- BUS:B<x>:SPI:MISo:INPut {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SPI:MISo:INPut?
- BUS:B<x>:SPI:MISo:THReshold <NR3>
- BUS:B<x>:SPI:MISo:THReshold?
- BUS:B<x>:SPI:MOSi:DATa:POLarity {HIGH|LOW}
- BUS:B<x>:SPI:MOSi:DATa:POLarity?
- BUS:B<x>:SPI:MOSi:INPut {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SPI:MOSi:INPut?
- BUS:B<x>:SPI:MOSi:THReshold <NR3>
- BUS:B<x>:SPI:MOSi:THReshold?
- BUS:B<x>:SPI:NUMBer:INputs {ONE|TWO}
- BUS:B<x>:SPI:NUMBer:INputs?
- BUS:B<x>:SPI:SELect:POLarity {LOW|HIGH}
- BUS:B<x>:SPI:SELect:POLarity?
- BUS:B<x>:SPI:SELect:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SPI:SELect:SOUrce?
- BUS:B<x>:SPI:SELect:THReshold <NR3>
- BUS:B<x>:SPI:SELect:THReshold?
- BUS:B<x>:SPMI:SCLk:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SPMI:SCLk:SOUrce?
- BUS:B<x>:SPMI:SCLk:THReshold <NR3>
- BUS:B<x>:SPMI:SCLk:THReshold?
- BUS:B<x>:SPMI:SDATa:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SPMI:SDATa:SOUrce?
- BUS:B<x>:SPMI:SDATa:THReshold <NR3>
- BUS:B<x>:SPMI:SDATa:THReshold?
- BUS:B<x>:SVID:ALERT:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SVID:ALERT:SOUrce?
- BUS:B<x>:SVID:ALERT:THReshold <NR3>
- BUS:B<x>:SVID:ALERT:THReshold?
- BUS:B<x>:SVID:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SVID:CLOCk:SOUrce?
- BUS:B<x>:SVID:CLOCk:THReshold <NR3>
- BUS:B<x>:SVID:CLOCk:THReshold?
- BUS:B<x>:SVID:DATA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SVID:DATA:SOUrce?
- BUS:B<x>:SVID:DATA:THReshold <NR3>
- BUS:B<x>:SVID:DATA:THReshold?
- BUS:B<x>:TYPe {ARINC429|AUDio|CAN|ETHernet|EUSB|FLEXRAY|I2C|I3C|LIN|MDIO|MIL1553B|PARallel|RS232C|SENT|SPI|SPMI|SVID|USB}
- BUS:B<x>:TYPe?
- BUS:B<x>:USB:BITRate {FULL|HIGH|LOW|USB3O|USB31G1|USB32G1}
- BUS:B<x>:USB:BITRate?
- BUS:B<x>:USB:DATAMINUSTHRESHold <NR3>
- BUS:B<x>:USB:DATAMINUSTHRESHold?
- BUS:B<x>:USB:DATAPLUSTHRESHold <NR3>
- BUS:B<x>:USB:DATAPLUSTHRESHold?
- BUS:B<x>:USB:DISplaymode {ON|OFF}
- BUS:B<x>:USB:DISplaymode?
- BUS:B<x>:USB:LANEONE:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:USB:LANEONE:SOUrce?
- BUS:B<x>:USB:LANEONE:THRESHold <NR3>
- BUS:B<x>:USB:LANEONE:THRESHold?
- BUS:B<x>:USB:LANES {SINGLe|DUAL}
- BUS:B<x>:USB:LANES?
- BUS:B<x>:USB:LANEZERO:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:USB:LANEZERO:SOUrce?
- BUS:B<x>:USB:LANEZERO:THRESHold <NR3>
- BUS:B<x>:USB:LANEZERO:THRESHold?
- BUS:B<x>:USB:LOWTHRESHold <NR3>
- BUS:B<x>:USB:LOWTHRESHold?
- BUS:B<x>:USB:SEMATHTHRESHold <NR3>
- BUS:B<x>:USB:SEMATHTHRESHold?
- BUS:B<x>:USB:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:USB:SIGNALTYpe?
- BUS:B<x>:USB:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:USB:SOUrce:DMINus {CH<x>|CH<x>_D<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:USB:SOUrce:DMINus?
- BUS:B<x>:USB:SOUrce:DPLUs {CH<x>|CH<x>_D<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:USB:SOUrce:DPLUs?
- BUS:B<x>:USB:SOUrce?
- BUS:B<x>:USB:THRESHold <NR3>
- BUS:B<x>:USB:THRESHold?
- BUS:B<x>:USB:TYPe {USB}
- BUS:B<x>:USB:TYPe?
- BUS:DELete <QString>
- BUS:LIST?

Bus

Bases: SCPICmdRead

The BUS command tree.

Usage
  • Using the .query() method will send the BUS? query.
  • Using the .verify(value) method will send the BUS? query and raise an AssertionError if the returned value does not match value.
Properties
  • .addnew: The BUS:ADDNew command.
  • .b: The BUS:B<x> command tree.
  • .delete: The BUS:DELete command.
  • .list: The BUS:LIST command.

addnew property

addnew: BusAddnew

Return the BUS:ADDNew command.

Description
  • This command adds the specified bus. This command creates/adds the bus but does not display it (turn it on). In order to enable bus decoding and see the bus display on screen, send the DISplay:WAVEView<x>:BUS:B<x>:STATE ON command.
Usage
  • Using the .write(value) method will send the BUS:ADDNew value command.
SCPI Syntax
- BUS:ADDNew <QString>
Info
  • <QString> is a quoted string of the form ‘B’ where NR1 is ≥1.

b property

Return the BUS:B<x> command tree.

Usage
  • Using the .query() method will send the BUS:B<x>? query.
  • Using the .verify(value) method will send the BUS:B<x>? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .arinc429a: The BUS:B<x>:ARINC429A command tree.
  • .audio: The BUS:B<x>:AUDio command tree.
  • .autoethernet: The BUS:B<x>:AUTOETHERnet command tree.
  • .can: The BUS:B<x>:CAN command tree.
  • .cphy: The BUS:B<x>:CPHY command tree.
  • .cxpi: The BUS:B<x>:CXPI command tree.
  • .display: The BUS:B<x>:DISplay command tree.
  • .dphy: The BUS:B<x>:DPHY command tree.
  • .espi: The BUS:B<x>:ESPI command tree.
  • .ethercat: The BUS:B<x>:ETHERCAT command tree.
  • .ethernet: The BUS:B<x>:ETHERnet command tree.
  • .eusb: The BUS:B<x>:EUSB command tree.
  • .flexray: The BUS:B<x>:FLEXray command tree.
  • .i2c: The BUS:B<x>:I2C command tree.
  • .i3c: The BUS:B<x>:I3C command tree.
  • .label: The BUS:B<x>:LABel command tree.
  • .lin: The BUS:B<x>:LIN command tree.
  • .manchester: The BUS:B<x>:MANChester command tree.
  • .mdio: The BUS:B<x>:MDIO command tree.
  • .mil1553b: The BUS:B<x>:MIL1553B command tree.
  • .nfc: The BUS:B<x>:NFC command tree.
  • .nrz: The BUS:B<x>:NRZ command tree.
  • .onewire: The BUS:B<x>:ONEWIRe command tree.
  • .parallel: The BUS:B<x>:PARallel command tree.
  • .pcie: The BUS:B<x>:PCIE command tree.
  • .psifive: The BUS:B<x>:PSIFIVe command tree.
  • .rs232c: The BUS:B<x>:RS232C command tree.
  • .s8b10b: The BUS:B<x>:S8B10B command tree.
  • .sdlc: The BUS:B<x>:SDLC command tree.
  • .sent: The BUS:B<x>:SENT command tree.
  • .smbus: The BUS:B<x>:SMBUS command tree.
  • .spacewire: The BUS:B<x>:SPACEWIRe command tree.
  • .spi: The BUS:B<x>:SPI command tree.
  • .spmi: The BUS:B<x>:SPMI command tree.
  • .svid: The BUS:B<x>:SVID command tree.
  • .type: The BUS:B<x>:TYPe command.
  • .usb: The BUS:B<x>:USB command tree.

delete property

delete: BusDelete

Return the BUS:DELete command.

Description
  • This command deletes the specified bus.
Usage
  • Using the .write(value) method will send the BUS:DELete value command.
SCPI Syntax
- BUS:DELete <QString>
Info
  • <QString> specifies the bus to delete and is of the form ‘B‘, where is ≥1.

list property

list: BusList

Return the BUS:LIST command.

Description
  • This query returns a comma separated list of all currently defined buses.
Usage
  • Using the .query() method will send the BUS:LIST? query.
  • Using the .verify(value) method will send the BUS:LIST? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- BUS:LIST?

BusAddnew

Bases: SCPICmdWrite

The BUS:ADDNew command.

Description
  • This command adds the specified bus. This command creates/adds the bus but does not display it (turn it on). In order to enable bus decoding and see the bus display on screen, send the DISplay:WAVEView<x>:BUS:B<x>:STATE ON command.
Usage
  • Using the .write(value) method will send the BUS:ADDNew value command.
SCPI Syntax
- BUS:ADDNew <QString>
Info
  • <QString> is a quoted string of the form ‘B’ where NR1 is ≥1.

BusBItem

Bases: ValidatedDynamicNumberCmd, SCPICmdRead

The BUS:B<x> command tree.

Usage
  • Using the .query() method will send the BUS:B<x>? query.
  • Using the .verify(value) method will send the BUS:B<x>? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .arinc429a: The BUS:B<x>:ARINC429A command tree.
  • .audio: The BUS:B<x>:AUDio command tree.
  • .autoethernet: The BUS:B<x>:AUTOETHERnet command tree.
  • .can: The BUS:B<x>:CAN command tree.
  • .cphy: The BUS:B<x>:CPHY command tree.
  • .cxpi: The BUS:B<x>:CXPI command tree.
  • .display: The BUS:B<x>:DISplay command tree.
  • .dphy: The BUS:B<x>:DPHY command tree.
  • .espi: The BUS:B<x>:ESPI command tree.
  • .ethercat: The BUS:B<x>:ETHERCAT command tree.
  • .ethernet: The BUS:B<x>:ETHERnet command tree.
  • .eusb: The BUS:B<x>:EUSB command tree.
  • .flexray: The BUS:B<x>:FLEXray command tree.
  • .i2c: The BUS:B<x>:I2C command tree.
  • .i3c: The BUS:B<x>:I3C command tree.
  • .label: The BUS:B<x>:LABel command tree.
  • .lin: The BUS:B<x>:LIN command tree.
  • .manchester: The BUS:B<x>:MANChester command tree.
  • .mdio: The BUS:B<x>:MDIO command tree.
  • .mil1553b: The BUS:B<x>:MIL1553B command tree.
  • .nfc: The BUS:B<x>:NFC command tree.
  • .nrz: The BUS:B<x>:NRZ command tree.
  • .onewire: The BUS:B<x>:ONEWIRe command tree.
  • .parallel: The BUS:B<x>:PARallel command tree.
  • .pcie: The BUS:B<x>:PCIE command tree.
  • .psifive: The BUS:B<x>:PSIFIVe command tree.
  • .rs232c: The BUS:B<x>:RS232C command tree.
  • .s8b10b: The BUS:B<x>:S8B10B command tree.
  • .sdlc: The BUS:B<x>:SDLC command tree.
  • .sent: The BUS:B<x>:SENT command tree.
  • .smbus: The BUS:B<x>:SMBUS command tree.
  • .spacewire: The BUS:B<x>:SPACEWIRe command tree.
  • .spi: The BUS:B<x>:SPI command tree.
  • .spmi: The BUS:B<x>:SPMI command tree.
  • .svid: The BUS:B<x>:SVID command tree.
  • .type: The BUS:B<x>:TYPe command.
  • .usb: The BUS:B<x>:USB command tree.

arinc429a property

arinc429a: BusBItemArinc429a

Return the BUS:B<x>:ARINC429A command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ARINC429A? query.
  • Using the .verify(value) method will send the BUS:B<x>:ARINC429A? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .bitrate: The BUS:B<x>:ARINC429A:BITRate command.
  • .dataformat: The BUS:B<x>:ARINC429A:DATAFORmat command.
  • .polarity: The BUS:B<x>:ARINC429A:POLARITY command.
  • .source: The BUS:B<x>:ARINC429A:SOUrce command.
  • .threshold: The BUS:B<x>:ARINC429A:THRESHold command.

audio property

Return the BUS:B<x>:AUDio command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:AUDio? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .bitdelay: The BUS:B<x>:AUDio:BITDelay command.
  • .bitorder: The BUS:B<x>:AUDio:BITOrder command.
  • .clock: The BUS:B<x>:AUDio:CLOCk command tree.
  • .data: The BUS:B<x>:AUDio:DATa command tree.
  • .frame: The BUS:B<x>:AUDio:FRAME command tree.
  • .type: The BUS:B<x>:AUDio:TYPe command.
  • .wordsel: The BUS:B<x>:AUDio:WORDSel command tree.

autoethernet property

autoethernet: BusBItemAutoethernet

Return the BUS:B<x>:AUTOETHERnet command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Sub-properties
  • .bitrate: The BUS:B<x>:AUTOETHERnet:BITRate command.
  • .dataminusthreshold: The BUS:B<x>:AUTOETHERnet:DATAMINUSTHRESHOLD command.
  • .dataplusthreshold: The BUS:B<x>:AUTOETHERnet:DATAPLUSTHRESHold command.
  • .lowdataminus: The BUS:B<x>:AUTOETHERnet:LOWDATAMINus command.
  • .lowdataplus: The BUS:B<x>:AUTOETHERnet:LOWDATAPLUS command.
  • .lowthreshold: The BUS:B<x>:AUTOETHERnet:LOWTHRESHold command.
  • .packetview: The BUS:B<x>:AUTOETHERnet:PACKetview command.
  • .signaltype: The BUS:B<x>:AUTOETHERnet:SIGNALTYpe command.
  • .source: The BUS:B<x>:AUTOETHERnet:SOUrce command.
  • .tbt1s: The BUS:B<x>:AUTOETHERnet:TBT1s command tree.
  • .tenbaset1s: The BUS:B<x>:AUTOETHERnet:TENBASET1S command tree.
  • .threshold: The BUS:B<x>:AUTOETHERnet:THRESHold command.
  • .type: The BUS:B<x>:AUTOETHERnet:TYPe command.

can property

Return the BUS:B<x>:CAN command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CAN? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .bitrate: The BUS:B<x>:CAN:BITRate command.
  • .fd: The BUS:B<x>:CAN:FD command tree.
  • .samplepoint: The BUS:B<x>:CAN:SAMPLEpoint command.
  • .signal: The BUS:B<x>:CAN:SIGNal command.
  • .source: The BUS:B<x>:CAN:SOUrce command.
  • .standard: The BUS:B<x>:CAN:STANDard command.
  • .threshold: The BUS:B<x>:CAN:THReshold command.
  • .xl: The BUS:B<x>:CAN:XL command tree.

cphy property

Return the BUS:B<x>:CPHY command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .a: The BUS:B<x>:CPHY:A command tree.
  • .ab: The BUS:B<x>:CPHY:AB command tree.
  • .agnd: The BUS:B<x>:CPHY:AGND command tree.
  • .b: The BUS:B<x>:CPHY:B command tree.
  • .bc: The BUS:B<x>:CPHY:BC command tree.
  • .bitrate: The BUS:B<x>:CPHY:BITRate command.
  • .c: The BUS:B<x>:CPHY:C command tree.
  • .ca: The BUS:B<x>:CPHY:CA command tree.
  • .cgnd: The BUS:B<x>:CPHY:CGND command tree.
  • .lp: The BUS:B<x>:CPHY:LP command tree.
  • .signaltype: The BUS:B<x>:CPHY:SIGNALTYpe command.
  • .subtype: The BUS:B<x>:CPHY:SUBTYPe command.

cxpi property

Return the BUS:B<x>:CXPI command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CXPI? query.
  • Using the .verify(value) method will send the BUS:B<x>:CXPI? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .bitrate: The BUS:B<x>:CXPI:BITRate command.
  • .rec: The BUS:B<x>:CXPI:REC command tree.
  • .source: The BUS:B<x>:CXPI:SOUrce command.

display property

display: BusBItemDisplay

Return the BUS:B<x>:DISplay command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:DISplay? query.
  • Using the .verify(value) method will send the BUS:B<x>:DISplay? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .format: The BUS:B<x>:DISplay:FORMat command.
  • .layout: The BUS:B<x>:DISplay:LAYout command.

dphy property

Return the BUS:B<x>:DPHY command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:DPHY? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Sub-properties
  • .clock: The BUS:B<x>:DPHY:CLOCk command tree.
  • .dminus: The BUS:B<x>:DPHY:DMINus command tree.
  • .dplus: The BUS:B<x>:DPHY:DPlus command tree.
  • .lp: The BUS:B<x>:DPHY:LP command tree.
  • .protocol: The BUS:B<x>:DPHY:PROTocol command tree.
  • .signal: The BUS:B<x>:DPHY:SIGNal command tree.

espi property

Return the BUS:B<x>:ESPI command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ESPI? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Sub-properties
  • .alertview: The BUS:B<x>:ESPI:ALERTVIEW command.
  • .alert: The BUS:B<x>:ESPI:ALERt command tree.
  • .chipselect: The BUS:B<x>:ESPI:CHIPSELect command tree.
  • .clock: The BUS:B<x>:ESPI:CLOCk command tree.
  • .dataone: The BUS:B<x>:ESPI:DATAONE command tree.
  • .datatwo: The BUS:B<x>:ESPI:DATATWO command tree.
  • .iomode: The BUS:B<x>:ESPI:IOMODe command.

ethercat property

ethercat: BusBItemEthercat

Return the BUS:B<x>:ETHERCAT command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .dataminusthreshold: The BUS:B<x>:ETHERCAT:DATAMINUSTHRESHold command.
  • .dataplusthreshold: The BUS:B<x>:ETHERCAT:DATAPLUSTHRESHold command.
  • .signaltype: The BUS:B<x>:ETHERCAT:SIGNALTYpe command.
  • .source: The BUS:B<x>:ETHERCAT:SOUrce command tree.
  • .threshold: The BUS:B<x>:ETHERCAT:THRESHold command.

ethernet property

ethernet: BusBItemEthernet

Return the BUS:B<x>:ETHERnet command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .dataminusthreshold: The BUS:B<x>:ETHERnet:DATAMINUSTHRESHold command.
  • .dataplusthreshold: The BUS:B<x>:ETHERnet:DATAPLUSTHRESHold command.
  • .ipvfour: The BUS:B<x>:ETHERnet:IPVFOUR command.
  • .lowthreshold: The BUS:B<x>:ETHERnet:LOWTHRESHold command.
  • .qtagging: The BUS:B<x>:ETHERnet:QTAGGING command.
  • .signaltype: The BUS:B<x>:ETHERnet:SIGNALTYpe command.
  • .source: The BUS:B<x>:ETHERnet:SOUrce command.
  • .threshold: The BUS:B<x>:ETHERnet:THRESHold command.
  • .type: The BUS:B<x>:ETHERnet:TYPe command.

eusb property

Return the BUS:B<x>:EUSB command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:EUSB? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .bitrate: The BUS:B<x>:EUSB:BITRate command.
  • .dataminus: The BUS:B<x>:EUSB:DATAMINUS command tree.
  • .dataminusthreshold: The BUS:B<x>:EUSB:DATAMINUSTHRESHold command.
  • .dataplus: The BUS:B<x>:EUSB:DATAPLUS command tree.
  • .dataplusthreshold: The BUS:B<x>:EUSB:DATAPLUSTHRESHold command.
  • .lowthreshold: The BUS:B<x>:EUSB:LOWTHRESHold command.
  • .operating: The BUS:B<x>:EUSB:OPERating command tree.
  • .signaltype: The BUS:B<x>:EUSB:SIGNALTYpe command.
  • .source: The BUS:B<x>:EUSB:SOUrce command tree.
  • .threshold: The BUS:B<x>:EUSB:THRESHold command.

flexray property

flexray: BusBItemFlexray

Return the BUS:B<x>:FLEXray command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .bitrate: The BUS:B<x>:FLEXray:BITRate command.
  • .channel: The BUS:B<x>:FLEXray:CHannel command.
  • .lowthreshold: The BUS:B<x>:FLEXray:LOWTHRESHold command.
  • .signal: The BUS:B<x>:FLEXray:SIGnal command.
  • .source: The BUS:B<x>:FLEXray:SOUrce command.
  • .threshold: The BUS:B<x>:FLEXray:THRESHold command.
  • .txrxthreshold: The BUS:B<x>:FLEXray:TXRXTHRESHold command.

i2c property

Return the BUS:B<x>:I2C command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:I2C? query.
  • Using the .verify(value) method will send the BUS:B<x>:I2C? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .clock: The BUS:B<x>:I2C:CLOCk command tree.
  • .data: The BUS:B<x>:I2C:DATa command tree.
  • .rwinaddr: The BUS:B<x>:I2C:RWINADDR command.

i3c property

Return the BUS:B<x>:I3C command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:I3C? query.
  • Using the .verify(value) method will send the BUS:B<x>:I3C? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> specifies the bus number.
Sub-properties
  • .clock: The BUS:B<x>:I3C:CLOCk command tree.
  • .data: The BUS:B<x>:I3C:DATa command tree.
  • .version: The BUS:B<x>:I3C:VERSion command.

label property

Return the BUS:B<x>:LABel command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:LABel? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .color: The BUS:B<x>:LABel:COLor command.
  • .font: The BUS:B<x>:LABel:FONT command tree.
  • .xpos: The BUS:B<x>:LABel:XPOS command.
  • .ypos: The BUS:B<x>:LABel:YPOS command.
  • .name: The BUS:B<x>:LABel:name command.

lin property

Return the BUS:B<x>:LIN command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:LIN? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .bitrate: The BUS:B<x>:LIN:BITRate command.
  • .idformat: The BUS:B<x>:LIN:IDFORmat command.
  • .polarity: The BUS:B<x>:LIN:POLarity command.
  • .samplepoint: The BUS:B<x>:LIN:SAMPLEpoint command.
  • .source: The BUS:B<x>:LIN:SOUrce command.
  • .standard: The BUS:B<x>:LIN:STANDard command.

manchester property

manchester: BusBItemManchester

Return the BUS:B<x>:MANChester command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MANChester? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .bitorder: The BUS:B<x>:MANChester:BITORDer command.
  • .bitrate: The BUS:B<x>:MANChester:BITRate command.
  • .displaymode: The BUS:B<x>:MANChester:DISplaymode command.
  • .header: The BUS:B<x>:MANChester:HEADer command tree.
  • .idle: The BUS:B<x>:MANChester:IDLE command tree.
  • .source: The BUS:B<x>:MANChester:SOUrce command.
  • .start: The BUS:B<x>:MANChester:START command tree.
  • .sync: The BUS:B<x>:MANChester:SYNC command tree.
  • .threshold: The BUS:B<x>:MANChester:THReshold command.
  • .tolerance: The BUS:B<x>:MANChester:TOLerance command.
  • .transtion: The BUS:B<x>:MANChester:TRANstion command tree.
  • .trailer: The BUS:B<x>:MANChester:TRAiler command tree.
  • .word: The BUS:B<x>:MANChester:WORD command tree.
  • .wordsize: The BUS:B<x>:MANChester:WORDSIZe command.
  • .parity: The BUS:B<x>:MANChester:parity command.

mdio property

Return the BUS:B<x>:MDIO command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MDIO? query.
  • Using the .verify(value) method will send the BUS:B<x>:MDIO? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .clock: The BUS:B<x>:MDIO:CLOCk command tree.
  • .data: The BUS:B<x>:MDIO:DATA command tree.

mil1553b property

mil1553b: BusBItemMil1553b

Return the BUS:B<x>:MIL1553B command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MIL1553B? query.
  • Using the .verify(value) method will send the BUS:B<x>:MIL1553B? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .lowthreshold: The BUS:B<x>:MIL1553B:LOWTHRESHold command.
  • .polarity: The BUS:B<x>:MIL1553B:POLarity command.
  • .responsetime: The BUS:B<x>:MIL1553B:RESPonsetime command tree.
  • .source: The BUS:B<x>:MIL1553B:SOUrce command.
  • .threshold: The BUS:B<x>:MIL1553B:THRESHold command.

nfc property

Return the BUS:B<x>:NFC command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NFC? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .applydemod: The BUS:B<x>:NFC:APPLYDEMod command.
  • .bitrate: The BUS:B<x>:NFC:BITRate command.
  • .cmd: The BUS:B<x>:NFC:CMD command tree.
  • .command: The BUS:B<x>:NFC:COMMand command tree.
  • .response: The BUS:B<x>:NFC:RESPonse command tree.
  • .rsp: The BUS:B<x>:NFC:RSP command tree.
  • .source: The BUS:B<x>:NFC:SOUrce command.
  • .start: The BUS:B<x>:NFC:START command tree.
  • .std: The BUS:B<x>:NFC:STD command.
  • .tolerance: The BUS:B<x>:NFC:TOLerance command.
  • .transition: The BUS:B<x>:NFC:TRANsition command tree.

nrz property

Return the BUS:B<x>:NRZ command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NRZ? query.
  • Using the .verify(value) method will send the BUS:B<x>:NRZ? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .bitorder: The BUS:B<x>:NRZ:BITOrder command.
  • .bitrate: The BUS:B<x>:NRZ:BITRate command.
  • .polarity: The BUS:B<x>:NRZ:POLarity command.
  • .source: The BUS:B<x>:NRZ:SOUrce command.
  • .spmi: The BUS:B<x>:NRZ:SPMI command tree.
  • .threshold: The BUS:B<x>:NRZ:THReshold command.

onewire property

onewire: BusBItemOnewire

Return the BUS:B<x>:ONEWIRe command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ONEWIRe? query.
  • Using the .verify(value) method will send the BUS:B<x>:ONEWIRe? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .data: The BUS:B<x>:ONEWIRe:DATA command tree.
  • .mode: The BUS:B<x>:ONEWIRe:MODe command.

parallel property

parallel: BusBItemParallel

Return the BUS:B<x>:PARallel command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PARallel? query.
  • Using the .verify(value) method will send the BUS:B<x>:PARallel? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .allthresholds: The BUS:B<x>:PARallel:ALLTHResholds command.
  • .bitsource: The BUS:B<x>:PARallel:BIT<x>SOUrce command.
  • .clocksource: The BUS:B<x>:PARallel:CLOCKSOUrce command.
  • .clock: The BUS:B<x>:PARallel:CLOCk command tree.

pcie property

Return the BUS:B<x>:PCIE command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PCIE? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .dataminus: The BUS:B<x>:PCIE:DATAMinus command tree.
  • .dataplus: The BUS:B<x>:PCIE:DATAPlus command tree.
  • .mthreshold: The BUS:B<x>:PCIE:MTHReshold command.
  • .signaltype: The BUS:B<x>:PCIE:SIGNALTYpe command.
  • .source: The BUS:B<x>:PCIE:SOUrce command tree.

psifive property

psifive: BusBItemPsifive

Return the BUS:B<x>:PSIFIVe command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .bitperiod: The BUS:B<x>:PSIFIVe:BITPERiod command.
  • .bitrate: The BUS:B<x>:PSIFIVe:BITRate command.
  • .comm: The BUS:B<x>:PSIFIVe:COMM command tree.
  • .dataa: The BUS:B<x>:PSIFIVe:DATAA command.
  • .datab: The BUS:B<x>:PSIFIVe:DATAB command.
  • .dataformat: The BUS:B<x>:PSIFIVe:DATAFORMat command.
  • .ecusource: The BUS:B<x>:PSIFIVe:ECUSOURce command.
  • .framecontrol: The BUS:B<x>:PSIFIVe:FRAMECONTrol command.
  • .messaging: The BUS:B<x>:PSIFIVe:MESSaging command.
  • .source: The BUS:B<x>:PSIFIVe:SOUrce command.
  • .status: The BUS:B<x>:PSIFIVe:STATus command.
  • .syncmode: The BUS:B<x>:PSIFIVe:SYNCMODe command.
  • .syncthreshold: The BUS:B<x>:PSIFIVe:SYNCTHRESHold command.
  • .threshold: The BUS:B<x>:PSIFIVe:THRESHold command.

rs232c property

Return the BUS:B<x>:RS232C command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:RS232C? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .bitorder: The BUS:B<x>:RS232C:BITORDer command.
  • .bitrate: The BUS:B<x>:RS232C:BITRate command.
  • .databits: The BUS:B<x>:RS232C:DATABits command.
  • .delimiter: The BUS:B<x>:RS232C:DELIMiter command.
  • .displaymode: The BUS:B<x>:RS232C:DISplaymode command.
  • .parity: The BUS:B<x>:RS232C:PARity command.
  • .polarity: The BUS:B<x>:RS232C:POLarity command.
  • .source: The BUS:B<x>:RS232C:SOUrce command.

s8b10b property

Return the BUS:B<x>:S8B10B command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:S8B10B? query.
  • Using the .verify(value) method will send the BUS:B<x>:S8B10B? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus. Arguments are available at bit rates up to 1 Tbps.
Sub-properties
  • .bitrate: The BUS:B<x>:S8B10B:BITRate command.
  • .source: The BUS:B<x>:S8B10B:SOUrce command.
  • .threshold: The BUS:B<x>:S8B10B:THReshold command.

sdlc property

Return the BUS:B<x>:SDLC command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SDLC? query.
  • Using the .verify(value) method will send the BUS:B<x>:SDLC? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .bitrate: The BUS:B<x>:SDLC:BITRate command.
  • .data: The BUS:B<x>:SDLC:DATA command tree.
  • .encoding: The BUS:B<x>:SDLC:ENCoding command.
  • .modulo: The BUS:B<x>:SDLC:MODulo command.

sent property

Return the BUS:B<x>:SENT command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SENT? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .chanwidth: The BUS:B<x>:SENT:CHANWidth command.
  • .nibblecount: The BUS:B<x>:SENT:NIBBLECount command.
  • .numchannel: The BUS:B<x>:SENT:NUMCHANnel command.
  • .pausepulse: The BUS:B<x>:SENT:PAUSEPULSe command.
  • .polarity: The BUS:B<x>:SENT:POLARITY command.
  • .slow: The BUS:B<x>:SENT:SLOW command.
  • .source: The BUS:B<x>:SENT:SOUrce command.
  • .threshold: The BUS:B<x>:SENT:THRESHold command.
  • .ticktime: The BUS:B<x>:SENT:TICKTIME command.
  • .ticktolerance: The BUS:B<x>:SENT:TICKTOLerance command.

smbus property

Return the BUS:B<x>:SMBUS command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the bus number.
Sub-properties
  • .clock: The BUS:B<x>:SMBUS:CLOCk command tree.
  • .data: The BUS:B<x>:SMBUS:DATA command tree.
  • .pec: The BUS:B<x>:SMBUS:PEC command tree.

spacewire property

spacewire: BusBItemSpacewire

Return the BUS:B<x>:SPACEWIRe command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the bus number.
Sub-properties
  • .bitrate: The BUS:B<x>:SPACEWIRe:BITRate command.
  • .data: The BUS:B<x>:SPACEWIRe:DATa command tree.
  • .decode: The BUS:B<x>:SPACEWIRe:DECode command tree.
  • .strobe: The BUS:B<x>:SPACEWIRe:STRobe command tree.
  • .sync: The BUS:B<x>:SPACEWIRe:SYNC command.

spi property

Return the BUS:B<x>:SPI command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .bitorder: The BUS:B<x>:SPI:BITOrder command.
  • .clock: The BUS:B<x>:SPI:CLOCk command tree.
  • .data: The BUS:B<x>:SPI:DATa command tree.
  • .framing: The BUS:B<x>:SPI:FRAMING command.
  • .idletime: The BUS:B<x>:SPI:IDLETime command.
  • .miso: The BUS:B<x>:SPI:MISo command tree.
  • .mosi: The BUS:B<x>:SPI:MOSi command tree.
  • .number: The BUS:B<x>:SPI:NUMBer command tree.
  • .select: The BUS:B<x>:SPI:SELect command tree.

spmi property

Return the BUS:B<x>:SPMI command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPMI? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPMI? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .sclk: The BUS:B<x>:SPMI:SCLk command tree.
  • .sdata: The BUS:B<x>:SPMI:SDATa command tree.

svid property

Return the BUS:B<x>:SVID command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SVID? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .alert: The BUS:B<x>:SVID:ALERT command tree.
  • .clock: The BUS:B<x>:SVID:CLOCk command tree.
  • .data: The BUS:B<x>:SVID:DATA command tree.

type property

Return the BUS:B<x>:TYPe command.

Description
  • This command sets or queries the bus type or standard for the specified bus. The bus is specified by x. Arguments for a bus type are only available then the required serial bus option is installed.
Usage
  • Using the .query() method will send the BUS:B<x>:TYPe? query.
  • Using the .verify(value) method will send the BUS:B<x>:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:TYPe value command.
SCPI Syntax
- BUS:B<x>:TYPe {ARINC429|AUDio|CAN|ETHernet|EUSB|FLEXRAY|I2C|I3C|LIN|MDIO|MIL1553B|PARallel|RS232C|SENT|SPI|SPMI|SVID|USB}
- BUS:B<x>:TYPe?
Info
  • B<x> is the number of the bus waveform.
  • ARINC429 specifies the ARINC 429 avionics serial bus.
  • AUDio specifies an audio bus.
  • CAN specifies a Controller Area Network bus.
  • ETHernet specifies the Ethernet bus.
  • EUSB specifies a eUSB bus. Requires option SR-EUSB2.
  • FLEXRAY specifies a FlexRay bus.
  • I2C specifies the Inter-IC bus.
  • I3C specifies the MIPI Improved Inter Integrated Circuit (I3C) bus.
  • LIN specifies a Local Interconnect Network bus.
  • MDIO specifies a MDIO bus.
  • MIL1553B specifies the MIL-STD-1553 avionics serial bus.
  • PARallel specifies a parallel bus.
  • RS232C specifies the RS-232 Serial bus.
  • SENT specifies the Single Edge Nibble Transmission (SENT) automotive serial bus.
  • SPI specifies the Serial Peripheral Interface bus.
  • SPMI Specifies a System Power Management Interface bus.
  • SVID Specifies a Serial VID bus.
  • USB specifies the Universal Serial bus.

usb property

Return the BUS:B<x>:USB command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:USB? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .bitrate: The BUS:B<x>:USB:BITRate command.
  • .dataminusthreshold: The BUS:B<x>:USB:DATAMINUSTHRESHold command.
  • .dataplusthreshold: The BUS:B<x>:USB:DATAPLUSTHRESHold command.
  • .displaymode: The BUS:B<x>:USB:DISplaymode command.
  • .laneone: The BUS:B<x>:USB:LANEONE command tree.
  • .lanes: The BUS:B<x>:USB:LANES command.
  • .lanezero: The BUS:B<x>:USB:LANEZERO command tree.
  • .lowthreshold: The BUS:B<x>:USB:LOWTHRESHold command.
  • .semaththreshold: The BUS:B<x>:USB:SEMATHTHRESHold command.
  • .signaltype: The BUS:B<x>:USB:SIGNALTYpe command.
  • .source: The BUS:B<x>:USB:SOUrce command.
  • .threshold: The BUS:B<x>:USB:THRESHold command.
  • .type: The BUS:B<x>:USB:TYPe command.

BusBItemArinc429a

Bases: SCPICmdRead

The BUS:B<x>:ARINC429A command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ARINC429A? query.
  • Using the .verify(value) method will send the BUS:B<x>:ARINC429A? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .bitrate: The BUS:B<x>:ARINC429A:BITRate command.
  • .dataformat: The BUS:B<x>:ARINC429A:DATAFORmat command.
  • .polarity: The BUS:B<x>:ARINC429A:POLARITY command.
  • .source: The BUS:B<x>:ARINC429A:SOUrce command.
  • .threshold: The BUS:B<x>:ARINC429A:THRESHold command.

bitrate property

Return the BUS:B<x>:ARINC429A:BITRate command.

Description
  • This command sets or queries the ARINC429 bit rate for the specified bus. The bus number is specified by x. If you select Custom, use BUS:B<x>:ARINC429A:BITRate:CUSTom to set the bit rate.
Usage
  • Using the .query() method will send the BUS:B<x>:ARINC429A:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:ARINC429A:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ARINC429A:BITRate value command.
SCPI Syntax
- BUS:B<x>:ARINC429A:BITRate {LOW|HI|CUSTom}
- BUS:B<x>:ARINC429A:BITRate?
Info
  • B<x> is the number of the bus.
  • LOW sets the bit rate to handle low speed signals.
  • HI sets the bit rate to handle high speed signals.
  • CUSTom uses the custom bit rate set by BUS:B<x>:ARINC429A:BITRate:CUSTom.
Sub-properties
  • .custom: The BUS:B<x>:ARINC429A:BITRate:CUSTom command.

dataformat property

Return the BUS:B<x>:ARINC429A:DATAFORmat command.

Description
  • This command sets or queries the format of the DATA field for the specified ARINC429 bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ARINC429A:DATAFORmat? query.
  • Using the .verify(value) method will send the BUS:B<x>:ARINC429A:DATAFORmat? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ARINC429A:DATAFORmat value command.
SCPI Syntax
- BUS:B<x>:ARINC429A:DATAFORmat {DATA|SDIDATA|SDIDATASSM}
- BUS:B<x>:ARINC429A:DATAFORmat?
Info
  • B<x> is the number of the bus.
  • DATA specifies a DATA field width of 19 bits (covering bits 11 through 29 ofthe 32 bit packet).
  • SDIDATA specifies a DATA field width of 21 bits (covering bits 9 through 29 of the 32 bit packet).
  • SDIDATASSM specifies a DATA field width of 23 bits (covering bits 9 through 31 of the 32 bit packet).

polarity property

Return the BUS:B<x>:ARINC429A:POLARITY command.

Description
  • This command sets or queries the source polarity for the specified ARINC429 bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ARINC429A:POLARITY? query.
  • Using the .verify(value) method will send the BUS:B<x>:ARINC429A:POLARITY? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ARINC429A:POLARITY value command.
SCPI Syntax
- BUS:B<x>:ARINC429A:POLARITY {NORMal|INVERTed}
- BUS:B<x>:ARINC429A:POLARITY?
Info
  • B<x> is the number of the bus.
  • NORMal specifies normal polarity.
  • INVERTed specifies inverted polarity.

source property

Return the BUS:B<x>:ARINC429A:SOUrce command.

Description
  • This command sets or queries the source for the specified ARINC429 bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ARINC429A:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ARINC429A:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ARINC429A:SOUrce value command.
SCPI Syntax
- BUS:B<x>:ARINC429A:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ARINC429A:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source waveform for the ARINC429 bus.
  • MATH<x> specifies a math waveform as the source waveform for the ARINC429 bus.
  • REF<x> specifies a reference waveform as the source waveform for the ARINC429 bus.

threshold property

Return the BUS:B<x>:ARINC429A:THRESHold command.

Description
  • This command sets or queries the ARINC429 upper threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ARINC429A:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ARINC429A:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ARINC429A:THRESHold value command.
SCPI Syntax
- BUS:B<x>:ARINC429A:THRESHold <NR3>
- BUS:B<x>:ARINC429A:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the ARINC429 lower threshold for the specified bus.

BusBItemArinc429aBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ARINC429A:BITRate command.

Description
  • This command sets or queries the ARINC429 bit rate for the specified bus. The bus number is specified by x. If you select Custom, use BUS:B<x>:ARINC429A:BITRate:CUSTom to set the bit rate.
Usage
  • Using the .query() method will send the BUS:B<x>:ARINC429A:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:ARINC429A:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ARINC429A:BITRate value command.
SCPI Syntax
- BUS:B<x>:ARINC429A:BITRate {LOW|HI|CUSTom}
- BUS:B<x>:ARINC429A:BITRate?
Info
  • B<x> is the number of the bus.
  • LOW sets the bit rate to handle low speed signals.
  • HI sets the bit rate to handle high speed signals.
  • CUSTom uses the custom bit rate set by BUS:B<x>:ARINC429A:BITRate:CUSTom.
Properties
  • .custom: The BUS:B<x>:ARINC429A:BITRate:CUSTom command.

custom property

Return the BUS:B<x>:ARINC429A:BITRate:CUSTom command.

Description
  • This command sets or queries the ARINC429 custom bit rate for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ARINC429A:BITRate:CUSTom? query.
  • Using the .verify(value) method will send the BUS:B<x>:ARINC429A:BITRate:CUSTom? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ARINC429A:BITRate:CUSTom value command.
SCPI Syntax
- BUS:B<x>:ARINC429A:BITRate:CUSTom <NR1>
- BUS:B<x>:ARINC429A:BITRate:CUSTom?
Info
  • B<x> is the number of the bus.
  • <NR1> is the ARINC429 custom bit rate for the specified bus.

BusBItemArinc429aBitrateCustom

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ARINC429A:BITRate:CUSTom command.

Description
  • This command sets or queries the ARINC429 custom bit rate for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ARINC429A:BITRate:CUSTom? query.
  • Using the .verify(value) method will send the BUS:B<x>:ARINC429A:BITRate:CUSTom? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ARINC429A:BITRate:CUSTom value command.
SCPI Syntax
- BUS:B<x>:ARINC429A:BITRate:CUSTom <NR1>
- BUS:B<x>:ARINC429A:BITRate:CUSTom?
Info
  • B<x> is the number of the bus.
  • <NR1> is the ARINC429 custom bit rate for the specified bus.

BusBItemArinc429aDataformat

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ARINC429A:DATAFORmat command.

Description
  • This command sets or queries the format of the DATA field for the specified ARINC429 bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ARINC429A:DATAFORmat? query.
  • Using the .verify(value) method will send the BUS:B<x>:ARINC429A:DATAFORmat? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ARINC429A:DATAFORmat value command.
SCPI Syntax
- BUS:B<x>:ARINC429A:DATAFORmat {DATA|SDIDATA|SDIDATASSM}
- BUS:B<x>:ARINC429A:DATAFORmat?
Info
  • B<x> is the number of the bus.
  • DATA specifies a DATA field width of 19 bits (covering bits 11 through 29 ofthe 32 bit packet).
  • SDIDATA specifies a DATA field width of 21 bits (covering bits 9 through 29 of the 32 bit packet).
  • SDIDATASSM specifies a DATA field width of 23 bits (covering bits 9 through 31 of the 32 bit packet).

BusBItemArinc429aPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ARINC429A:POLARITY command.

Description
  • This command sets or queries the source polarity for the specified ARINC429 bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ARINC429A:POLARITY? query.
  • Using the .verify(value) method will send the BUS:B<x>:ARINC429A:POLARITY? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ARINC429A:POLARITY value command.
SCPI Syntax
- BUS:B<x>:ARINC429A:POLARITY {NORMal|INVERTed}
- BUS:B<x>:ARINC429A:POLARITY?
Info
  • B<x> is the number of the bus.
  • NORMal specifies normal polarity.
  • INVERTed specifies inverted polarity.

BusBItemArinc429aSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ARINC429A:SOUrce command.

Description
  • This command sets or queries the source for the specified ARINC429 bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ARINC429A:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ARINC429A:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ARINC429A:SOUrce value command.
SCPI Syntax
- BUS:B<x>:ARINC429A:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ARINC429A:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source waveform for the ARINC429 bus.
  • MATH<x> specifies a math waveform as the source waveform for the ARINC429 bus.
  • REF<x> specifies a reference waveform as the source waveform for the ARINC429 bus.

BusBItemArinc429aThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ARINC429A:THRESHold command.

Description
  • This command sets or queries the ARINC429 upper threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ARINC429A:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ARINC429A:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ARINC429A:THRESHold value command.
SCPI Syntax
- BUS:B<x>:ARINC429A:THRESHold <NR3>
- BUS:B<x>:ARINC429A:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the ARINC429 lower threshold for the specified bus.

BusBItemAudio

Bases: SCPICmdRead

The BUS:B<x>:AUDio command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:AUDio? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .bitdelay: The BUS:B<x>:AUDio:BITDelay command.
  • .bitorder: The BUS:B<x>:AUDio:BITOrder command.
  • .clock: The BUS:B<x>:AUDio:CLOCk command tree.
  • .data: The BUS:B<x>:AUDio:DATa command tree.
  • .frame: The BUS:B<x>:AUDio:FRAME command tree.
  • .type: The BUS:B<x>:AUDio:TYPe command.
  • .wordsel: The BUS:B<x>:AUDio:WORDSel command tree.

bitdelay property

Return the BUS:B<x>:AUDio:BITDelay command.

Description
  • This command sets or queries the number of delay bits for the specified AUDIO bus. The bus is specified by x. Note: This command only applies to the TDM Audio type.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:BITDelay? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:BITDelay? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:BITDelay value command.
SCPI Syntax
- BUS:B<x>:AUDio:BITDelay <NR1>
- BUS:B<x>:AUDio:BITDelay?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the number of delay bits.

bitorder property

Return the BUS:B<x>:AUDio:BITOrder command.

Description
  • Specifies the bit order for the specified AUDIO bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:BITOrder? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:BITOrder? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:BITOrder value command.
SCPI Syntax
- BUS:B<x>:AUDio:BITOrder {MSB|LSB}
- BUS:B<x>:AUDio:BITOrder?
Info
  • B<x> is the number of the bus.
  • MSB specifies that the most significant bit will be expected first in the order.
  • LSB specifies that the least significant bit will be expected first in the order.

clock property

Return the BUS:B<x>:AUDio:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .polarity: The BUS:B<x>:AUDio:CLOCk:POLarity command.
  • .source: The BUS:B<x>:AUDio:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:AUDio:CLOCk:THReshold command.

data property

Return the BUS:B<x>:AUDio:DATa command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:DATa? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:DATa? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .polarity: The BUS:B<x>:AUDio:DATa:POLarity command.
  • .size: The BUS:B<x>:AUDio:DATa:SIZe command.
  • .source: The BUS:B<x>:AUDio:DATa:SOUrce command.
  • .threshold: The BUS:B<x>:AUDio:DATa:THReshold command.
  • .wordsize: The BUS:B<x>:AUDio:DATa:WORDSize command.

frame property

Return the BUS:B<x>:AUDio:FRAME command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:FRAME? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:FRAME? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .clockbitsperchannel: The BUS:B<x>:AUDio:FRAME:CLOCKBITSPERCHANNEL command.
  • .size: The BUS:B<x>:AUDio:FRAME:SIZe command.

type property

Return the BUS:B<x>:AUDio:TYPe command.

Description
  • This command sets or queries the audio format (type) for the specified audio bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:TYPe? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:TYPe value command.
SCPI Syntax
- BUS:B<x>:AUDio:TYPe {I2S|LJ|RJ|TDM}
- BUS:B<x>:AUDio:TYPe?
Info
  • B<x> is the number of the bus.
  • I2S specifies the I.
  • LJ specifies the left-justified audio format.
  • RJ specifies the right-justified audio format.
  • TDM specifies the time-division multiplexing audio format.

wordsel property

Return the BUS:B<x>:AUDio:WORDSel command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:WORDSel? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:WORDSel? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .polarity: The BUS:B<x>:AUDio:WORDSel:POLarity command.
  • .source: The BUS:B<x>:AUDio:WORDSel:SOUrce command.
  • .threshold: The BUS:B<x>:AUDio:WORDSel:THReshold command.

BusBItemAudioBitdelay

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUDio:BITDelay command.

Description
  • This command sets or queries the number of delay bits for the specified AUDIO bus. The bus is specified by x. Note: This command only applies to the TDM Audio type.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:BITDelay? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:BITDelay? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:BITDelay value command.
SCPI Syntax
- BUS:B<x>:AUDio:BITDelay <NR1>
- BUS:B<x>:AUDio:BITDelay?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the number of delay bits.

BusBItemAudioBitorder

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUDio:BITOrder command.

Description
  • Specifies the bit order for the specified AUDIO bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:BITOrder? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:BITOrder? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:BITOrder value command.
SCPI Syntax
- BUS:B<x>:AUDio:BITOrder {MSB|LSB}
- BUS:B<x>:AUDio:BITOrder?
Info
  • B<x> is the number of the bus.
  • MSB specifies that the most significant bit will be expected first in the order.
  • LSB specifies that the least significant bit will be expected first in the order.

BusBItemAudioClock

Bases: SCPICmdRead

The BUS:B<x>:AUDio:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .polarity: The BUS:B<x>:AUDio:CLOCk:POLarity command.
  • .source: The BUS:B<x>:AUDio:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:AUDio:CLOCk:THReshold command.

polarity property

Return the BUS:B<x>:AUDio:CLOCk:POLarity command.

Description
  • This command sets or queries the clock source polarity for the specified AUDIO bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:CLOCk:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:CLOCk:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:CLOCk:POLarity value command.
SCPI Syntax
- BUS:B<x>:AUDio:CLOCk:POLarity {FALL|RISE}
- BUS:B<x>:AUDio:CLOCk:POLarity?
Info
  • B<x> is the number of the bus.
  • FALL sets falling edge as the clock polarity.
  • RISE sets rising edge as the clock polarity.

source property

Return the BUS:B<x>:AUDio:CLOCk:SOUrce command.

Description
  • This command sets or queries the clock source waveform for the specified AUDIO bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:AUDio:CLOCk:SOUrce {CH<x>|CH<x>_D<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:AUDio:CLOCk:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the clock source waveform for the audio bus.
  • CH<x>_D<x> specifies a digital channel as the clock source waveform for thespecified audio bus.
  • MATH<x> specifies a math waveform as the clock source waveform for the audiobus.
  • REF<x> specifies a reference waveform as the clock source waveform for theaudio bus.
  • REF<x>_D<x> specifies a digital reference waveform as the clock sourcewaveform for the specified audio bus.

threshold property

Return the BUS:B<x>:AUDio:CLOCk:THReshold command.

Description
  • This command sets or queries the audio clock source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:CLOCk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:CLOCk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:CLOCk:THReshold value command.
SCPI Syntax
- BUS:B<x>:AUDio:CLOCk:THReshold <NR3>
- BUS:B<x>:AUDio:CLOCk:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the audio clock source threshold for the specified bus.

BusBItemAudioClockPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUDio:CLOCk:POLarity command.

Description
  • This command sets or queries the clock source polarity for the specified AUDIO bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:CLOCk:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:CLOCk:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:CLOCk:POLarity value command.
SCPI Syntax
- BUS:B<x>:AUDio:CLOCk:POLarity {FALL|RISE}
- BUS:B<x>:AUDio:CLOCk:POLarity?
Info
  • B<x> is the number of the bus.
  • FALL sets falling edge as the clock polarity.
  • RISE sets rising edge as the clock polarity.

BusBItemAudioClockSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUDio:CLOCk:SOUrce command.

Description
  • This command sets or queries the clock source waveform for the specified AUDIO bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:AUDio:CLOCk:SOUrce {CH<x>|CH<x>_D<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:AUDio:CLOCk:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the clock source waveform for the audio bus.
  • CH<x>_D<x> specifies a digital channel as the clock source waveform for thespecified audio bus.
  • MATH<x> specifies a math waveform as the clock source waveform for the audiobus.
  • REF<x> specifies a reference waveform as the clock source waveform for theaudio bus.
  • REF<x>_D<x> specifies a digital reference waveform as the clock sourcewaveform for the specified audio bus.

BusBItemAudioClockThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUDio:CLOCk:THReshold command.

Description
  • This command sets or queries the audio clock source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:CLOCk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:CLOCk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:CLOCk:THReshold value command.
SCPI Syntax
- BUS:B<x>:AUDio:CLOCk:THReshold <NR3>
- BUS:B<x>:AUDio:CLOCk:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the audio clock source threshold for the specified bus.

BusBItemAudioData

Bases: SCPICmdRead

The BUS:B<x>:AUDio:DATa command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:DATa? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:DATa? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .polarity: The BUS:B<x>:AUDio:DATa:POLarity command.
  • .size: The BUS:B<x>:AUDio:DATa:SIZe command.
  • .source: The BUS:B<x>:AUDio:DATa:SOUrce command.
  • .threshold: The BUS:B<x>:AUDio:DATa:THReshold command.
  • .wordsize: The BUS:B<x>:AUDio:DATa:WORDSize command.

polarity property

Return the BUS:B<x>:AUDio:DATa:POLarity command.

Description
  • This command sets or queries the audio data source polarity for the specified audio bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:DATa:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:DATa:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:DATa:POLarity value command.
SCPI Syntax
- BUS:B<x>:AUDio:DATa:POLarity {HIGH|LOW}
- BUS:B<x>:AUDio:DATa:POLarity?
Info
  • B<x> is the number of the bus.
  • HIGH specifies positive data polarity for the audio bus.
  • LOW specifies negative data polarity for the audio bus.

size property

Return the BUS:B<x>:AUDio:DATa:SIZe command.

Description
  • This command sets or queries the number of bits per channel for the specified audio bus. This command only applies to the TDM Audio type. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:DATa:SIZe? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:DATa:SIZe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:DATa:SIZe value command.
SCPI Syntax
- BUS:B<x>:AUDio:DATa:SIZe <NR1>
- BUS:B<x>:AUDio:DATa:SIZe?
Info
  • B<x> is the number of the bus.
  • NR1 specifies the number of bits per word.

source property

Return the BUS:B<x>:AUDio:DATa:SOUrce command.

Description
  • This command sets or queries the audio data source for the specified audio bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:DATa:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:DATa:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:DATa:SOUrce value command.
SCPI Syntax
- BUS:B<x>:AUDio:DATa:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:AUDio:DATa:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the data source waveform for the audio bus.
  • MATH<x> specifies an math waveform as the data source waveform for the audio bus.
  • REF<x> specifies an reference waveform as the data source waveform for the audio bus.
  • REF<x>_D<x> specifies a digital reference waveform as the data source waveform for the specified audio bus.

threshold property

Return the BUS:B<x>:AUDio:DATa:THReshold command.

Description
  • This command sets or queries the audio data source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:DATa:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:DATa:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:DATa:THReshold value command.
SCPI Syntax
- BUS:B<x>:AUDio:DATa:THReshold <NR3>
- BUS:B<x>:AUDio:DATa:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the audio data source threshold for the specified bus.

wordsize property

Return the BUS:B<x>:AUDio:DATa:WORDSize command.

Description
  • This command sets or queries the audio bits per word for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:DATa:WORDSize? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:DATa:WORDSize? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:DATa:WORDSize value command.
SCPI Syntax
- BUS:B<x>:AUDio:DATa:WORDSize <NR1>
- BUS:B<x>:AUDio:DATa:WORDSize?
Info
  • B<x> is the number of the bus.
  • <NR1> is the audio bits per word for the specified bus.

BusBItemAudioDataPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUDio:DATa:POLarity command.

Description
  • This command sets or queries the audio data source polarity for the specified audio bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:DATa:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:DATa:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:DATa:POLarity value command.
SCPI Syntax
- BUS:B<x>:AUDio:DATa:POLarity {HIGH|LOW}
- BUS:B<x>:AUDio:DATa:POLarity?
Info
  • B<x> is the number of the bus.
  • HIGH specifies positive data polarity for the audio bus.
  • LOW specifies negative data polarity for the audio bus.

BusBItemAudioDataSize

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUDio:DATa:SIZe command.

Description
  • This command sets or queries the number of bits per channel for the specified audio bus. This command only applies to the TDM Audio type. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:DATa:SIZe? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:DATa:SIZe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:DATa:SIZe value command.
SCPI Syntax
- BUS:B<x>:AUDio:DATa:SIZe <NR1>
- BUS:B<x>:AUDio:DATa:SIZe?
Info
  • B<x> is the number of the bus.
  • NR1 specifies the number of bits per word.

BusBItemAudioDataSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUDio:DATa:SOUrce command.

Description
  • This command sets or queries the audio data source for the specified audio bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:DATa:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:DATa:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:DATa:SOUrce value command.
SCPI Syntax
- BUS:B<x>:AUDio:DATa:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:AUDio:DATa:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the data source waveform for the audio bus.
  • MATH<x> specifies an math waveform as the data source waveform for the audio bus.
  • REF<x> specifies an reference waveform as the data source waveform for the audio bus.
  • REF<x>_D<x> specifies a digital reference waveform as the data source waveform for the specified audio bus.

BusBItemAudioDataThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUDio:DATa:THReshold command.

Description
  • This command sets or queries the audio data source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:DATa:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:DATa:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:DATa:THReshold value command.
SCPI Syntax
- BUS:B<x>:AUDio:DATa:THReshold <NR3>
- BUS:B<x>:AUDio:DATa:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the audio data source threshold for the specified bus.

BusBItemAudioDataWordsize

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUDio:DATa:WORDSize command.

Description
  • This command sets or queries the audio bits per word for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:DATa:WORDSize? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:DATa:WORDSize? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:DATa:WORDSize value command.
SCPI Syntax
- BUS:B<x>:AUDio:DATa:WORDSize <NR1>
- BUS:B<x>:AUDio:DATa:WORDSize?
Info
  • B<x> is the number of the bus.
  • <NR1> is the audio bits per word for the specified bus.

BusBItemAudioFrame

Bases: SCPICmdRead

The BUS:B<x>:AUDio:FRAME command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:FRAME? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:FRAME? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .clockbitsperchannel: The BUS:B<x>:AUDio:FRAME:CLOCKBITSPERCHANNEL command.
  • .size: The BUS:B<x>:AUDio:FRAME:SIZe command.

clockbitsperchannel property

Return the BUS:B<x>:AUDio:FRAME:CLOCKBITSPERCHANNEL command.

Description
  • This command sets or queries the audio bits of sync width for the specified bus. The bus is specified by x. Note: This command only applies to the TDM Audio type.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:FRAME:CLOCKBITSPERCHANNEL? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:FRAME:CLOCKBITSPERCHANNEL? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:FRAME:CLOCKBITSPERCHANNEL value command.
SCPI Syntax
- BUS:B<x>:AUDio:FRAME:CLOCKBITSPERCHANNEL <NR1>
- BUS:B<x>:AUDio:FRAME:CLOCKBITSPERCHANNEL?
Info
  • B<x> is the number of the bus.
  • <NR1> is the audio bits of sync width for the specified bus.

size property

Return the BUS:B<x>:AUDio:FRAME:SIZe command.

Description
  • This command sets or queries the number of audio channels in each frame for the specified AUDIO bus. The bus is specified by x. Note: This command only applies to the TDM Audio type.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:FRAME:SIZe? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:FRAME:SIZe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:FRAME:SIZe value command.
SCPI Syntax
- BUS:B<x>:AUDio:FRAME:SIZe <NR1>
- BUS:B<x>:AUDio:FRAME:SIZe?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the number of channels in each frame.

BusBItemAudioFrameClockbitsperchannel

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUDio:FRAME:CLOCKBITSPERCHANNEL command.

Description
  • This command sets or queries the audio bits of sync width for the specified bus. The bus is specified by x. Note: This command only applies to the TDM Audio type.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:FRAME:CLOCKBITSPERCHANNEL? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:FRAME:CLOCKBITSPERCHANNEL? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:FRAME:CLOCKBITSPERCHANNEL value command.
SCPI Syntax
- BUS:B<x>:AUDio:FRAME:CLOCKBITSPERCHANNEL <NR1>
- BUS:B<x>:AUDio:FRAME:CLOCKBITSPERCHANNEL?
Info
  • B<x> is the number of the bus.
  • <NR1> is the audio bits of sync width for the specified bus.

BusBItemAudioFrameSize

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUDio:FRAME:SIZe command.

Description
  • This command sets or queries the number of audio channels in each frame for the specified AUDIO bus. The bus is specified by x. Note: This command only applies to the TDM Audio type.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:FRAME:SIZe? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:FRAME:SIZe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:FRAME:SIZe value command.
SCPI Syntax
- BUS:B<x>:AUDio:FRAME:SIZe <NR1>
- BUS:B<x>:AUDio:FRAME:SIZe?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the number of channels in each frame.

BusBItemAudioType

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUDio:TYPe command.

Description
  • This command sets or queries the audio format (type) for the specified audio bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:TYPe? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:TYPe value command.
SCPI Syntax
- BUS:B<x>:AUDio:TYPe {I2S|LJ|RJ|TDM}
- BUS:B<x>:AUDio:TYPe?
Info
  • B<x> is the number of the bus.
  • I2S specifies the I.
  • LJ specifies the left-justified audio format.
  • RJ specifies the right-justified audio format.
  • TDM specifies the time-division multiplexing audio format.

BusBItemAudioWordsel

Bases: SCPICmdRead

The BUS:B<x>:AUDio:WORDSel command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:WORDSel? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:WORDSel? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .polarity: The BUS:B<x>:AUDio:WORDSel:POLarity command.
  • .source: The BUS:B<x>:AUDio:WORDSel:SOUrce command.
  • .threshold: The BUS:B<x>:AUDio:WORDSel:THReshold command.

polarity property

Return the BUS:B<x>:AUDio:WORDSel:POLarity command.

Description
  • This command sets or queries the word select source polarity for the specified audio bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:WORDSel:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:WORDSel:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:WORDSel:POLarity value command.
SCPI Syntax
- BUS:B<x>:AUDio:WORDSel:POLarity {NORMal|INVERTed}
- BUS:B<x>:AUDio:WORDSel:POLarity?
Info
  • B<x> is the number of the bus.
  • NORMal specifies positive polarity.
  • INVERTed specifies negative polarity.

source property

Return the BUS:B<x>:AUDio:WORDSel:SOUrce command.

Description
  • This command sets or queries the audio word select source waveform for the specified audio bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:WORDSel:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:WORDSel:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:WORDSel:SOUrce value command.
SCPI Syntax
- BUS:B<x>:AUDio:WORDSel:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:AUDio:WORDSel:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the word select source waveform.
  • MATH<x> specifies an math waveform as the word select source waveform.
  • REF<x> specifies an reference waveform as the word select source waveform.
  • REF<x>_D<x> specifies a digital reference waveform as the word select source waveform for the specified audio bus.

threshold property

Return the BUS:B<x>:AUDio:WORDSel:THReshold command.

Description
  • This command sets or queries the audio word select source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:WORDSel:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:WORDSel:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:WORDSel:THReshold value command.
SCPI Syntax
- BUS:B<x>:AUDio:WORDSel:THReshold <NR3>
- BUS:B<x>:AUDio:WORDSel:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the audio word select source threshold for the specified bus.

BusBItemAudioWordselPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUDio:WORDSel:POLarity command.

Description
  • This command sets or queries the word select source polarity for the specified audio bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:WORDSel:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:WORDSel:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:WORDSel:POLarity value command.
SCPI Syntax
- BUS:B<x>:AUDio:WORDSel:POLarity {NORMal|INVERTed}
- BUS:B<x>:AUDio:WORDSel:POLarity?
Info
  • B<x> is the number of the bus.
  • NORMal specifies positive polarity.
  • INVERTed specifies negative polarity.

BusBItemAudioWordselSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUDio:WORDSel:SOUrce command.

Description
  • This command sets or queries the audio word select source waveform for the specified audio bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:WORDSel:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:WORDSel:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:WORDSel:SOUrce value command.
SCPI Syntax
- BUS:B<x>:AUDio:WORDSel:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:AUDio:WORDSel:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the word select source waveform.
  • MATH<x> specifies an math waveform as the word select source waveform.
  • REF<x> specifies an reference waveform as the word select source waveform.
  • REF<x>_D<x> specifies a digital reference waveform as the word select source waveform for the specified audio bus.

BusBItemAudioWordselThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUDio:WORDSel:THReshold command.

Description
  • This command sets or queries the audio word select source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:AUDio:WORDSel:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUDio:WORDSel:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUDio:WORDSel:THReshold value command.
SCPI Syntax
- BUS:B<x>:AUDio:WORDSel:THReshold <NR3>
- BUS:B<x>:AUDio:WORDSel:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the audio word select source threshold for the specified bus.

BusBItemAutoethernet

Bases: SCPICmdRead

The BUS:B<x>:AUTOETHERnet command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Properties
  • .bitrate: The BUS:B<x>:AUTOETHERnet:BITRate command.
  • .dataminusthreshold: The BUS:B<x>:AUTOETHERnet:DATAMINUSTHRESHOLD command.
  • .dataplusthreshold: The BUS:B<x>:AUTOETHERnet:DATAPLUSTHRESHold command.
  • .lowdataminus: The BUS:B<x>:AUTOETHERnet:LOWDATAMINus command.
  • .lowdataplus: The BUS:B<x>:AUTOETHERnet:LOWDATAPLUS command.
  • .lowthreshold: The BUS:B<x>:AUTOETHERnet:LOWTHRESHold command.
  • .packetview: The BUS:B<x>:AUTOETHERnet:PACKetview command.
  • .signaltype: The BUS:B<x>:AUTOETHERnet:SIGNALTYpe command.
  • .source: The BUS:B<x>:AUTOETHERnet:SOUrce command.
  • .tbt1s: The BUS:B<x>:AUTOETHERnet:TBT1s command tree.
  • .tenbaset1s: The BUS:B<x>:AUTOETHERnet:TENBASET1S command tree.
  • .threshold: The BUS:B<x>:AUTOETHERnet:THRESHold command.
  • .type: The BUS:B<x>:AUTOETHERnet:TYPe command.

bitrate property

Return the BUS:B<x>:AUTOETHERnet:BITRate command.

Description
  • This command sets or queries bitrate for Auto Ethernet Bus with speed 10Base-T1S. The bit input for rate argument is NRF and output is NR1.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:BITRate value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:BITRate <NR1>
- BUS:B<x>:AUTOETHERnet:BITRate?
Info
  • B<x> is the Bus number.
  • <NR1> specifies the bit rate. The maximum value is 1 Gbps and the default is 10 Mbps.

dataminusthreshold property

Return the BUS:B<x>:AUTOETHERnet:DATAMINUSTHRESHOLD command.

Description
  • This command sets or queries the AutoEthernet D- source threshold level for the specified bus. This threshold only applies when the AutoEthernet signal type is single ended.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:DATAMINUSTHRESHOLD? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:DATAMINUSTHRESHOLD? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:DATAMINUSTHRESHOLD value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:DATAMINUSTHRESHOLD <NR3>
- BUS:B<x>:AUTOETHERnet:DATAMINUSTHRESHOLD?
Info
  • B<x> is the Bus number.
  • <NR3> specifies the AutoEthernet D- source threshold level for the specified bus, in volts. This threshold only applies when the AutoEthernet signal type is single ended.

dataplusthreshold property

Return the BUS:B<x>:AUTOETHERnet:DATAPLUSTHRESHold command.

Description
  • This command sets or queries the AutoEthernet D+ source threshold level for the specified bus. This threshold only applies when the AutoEthernet signal type is single ended.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:DATAPLUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:DATAPLUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:DATAPLUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:DATAPLUSTHRESHold <NR3>
- BUS:B<x>:AUTOETHERnet:DATAPLUSTHRESHold?
Info
  • B<x> is the Bus number.
  • <NR3> specifies the AutoEthernet D+ source threshold for the specified bus, in volts. This threshold only applies when the AutoEthernet signal type is single ended.

lowdataminus property

Return the BUS:B<x>:AUTOETHERnet:LOWDATAMINus command.

Description
  • This command sets or queries the AutoEthernet D- source low threshold level for the specified bus. This threshold only applies when the AutoEthernet signal type is single ended.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:LOWDATAMINus? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:LOWDATAMINus? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:LOWDATAMINus value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:LOWDATAMINus <NR3>
- BUS:B<x>:AUTOETHERnet:LOWDATAMINus?
Info
  • B<x> is the Bus number.
  • <NR3> specifies the AutoEthernet D- source low threshold level for the specified bus, in volts. This threshold only applies when the AutoEthernet signal type is single ended.

lowdataplus property

Return the BUS:B<x>:AUTOETHERnet:LOWDATAPLUS command.

Description
  • This command sets or queries the AutoEthernet D+ source low threshold level for the specified bus. This threshold only applies when the AutoEthernet signal type is single ended.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:LOWDATAPLUS? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:LOWDATAPLUS? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:LOWDATAPLUS value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:LOWDATAPLUS <NR3>
- BUS:B<x>:AUTOETHERnet:LOWDATAPLUS?
Info
  • B<x> is the Bus number.
  • <NR3> specifies the AutoEthernet D+ source low threshold level for the specified bus, in volts. This threshold only applies when the AutoEthernet signal type is single ended.

lowthreshold property

Return the BUS:B<x>:AUTOETHERnet:LOWTHRESHold command.

Description
  • This command sets or queries the AutoEthernet Data source Low threshold level for the specified bus. This threshold only applies when the AutoEthernet signal type is differential.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:LOWTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:LOWTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:LOWTHRESHold value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:LOWTHRESHold <NR3>
- BUS:B<x>:AUTOETHERnet:LOWTHRESHold?
Info
  • B<x> is the Bus number.
  • <NR3> specifies the AutoEthernet DATA source low threshold level for the specified bus, in volts. This threshold only applies when the AutoEthernet signal type is differential.

packetview property

Return the BUS:B<x>:AUTOETHERnet:PACKetview command.

Description
  • This command sets or queries the Auto Ethernet bus Packet View when speed is selected as 10 Base-T1S.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:PACKetview? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:PACKetview? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:PACKetview value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:PACKetview {ON|OFF}
- BUS:B<x>:AUTOETHERnet:PACKetview?
Info
  • B<x> is the Bus number.
  • ON sets the packet view for the decode to on.
  • OFF sets the packet view for the decode to off.

signaltype property

Return the BUS:B<x>:AUTOETHERnet:SIGNALTYpe command.

Description
  • This command sets or queries the AutoEthernet signal type for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:SIGNALTYpe? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:SIGNALTYpe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:SIGNALTYpe value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:AUTOETHERnet:SIGNALTYpe?
Info
  • B<x> is the Bus number.
  • SINGLE specifies single-ended signals.
  • DIFF specifies differential signals.

source property

Return the BUS:B<x>:AUTOETHERnet:SOUrce command.

Description
  • This command sets or queries the AutoEthernet data (SDATA) source for the specified bus. This command controls the source channel when the signal type is differential.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:SOUrce value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:AUTOETHERnet:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the AutoEthernet data source for differential input, where is the channel number.
  • MATH<x> specifies a math channel as the AutoEthernet data source for differential input, where is the math waveform number.
  • REF<x> specifies a reference waveform as the AutoEthernet data source for differential input, where is the reference waveform number.
Sub-properties
  • .dminus: The BUS:B<x>:AUTOETHERnet:SOUrce:DMINus command.
  • .dplus: The BUS:B<x>:AUTOETHERnet:SOUrce:DPLUs command.

tbt1s property

Return the BUS:B<x>:AUTOETHERnet:TBT1s command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:TBT1s? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:TBT1s? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Sub-properties
  • .mthreshold: The BUS:B<x>:AUTOETHERnet:TBT1s:MTHReshold command.

tenbaset1s property

Return the BUS:B<x>:AUTOETHERnet:TENBASET1S command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:TENBASET1S? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:TENBASET1S? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Sub-properties
  • .threshold: The BUS:B<x>:AUTOETHERnet:TENBASET1S:THRESHold command.

threshold property

Return the BUS:B<x>:AUTOETHERnet:THRESHold command.

Description
  • This command sets or queries the AutoEthernet DATA source High threshold level for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:THRESHold value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:THRESHold <NR3>
- BUS:B<x>:AUTOETHERnet:THRESHold?
Info
  • B<x> is the Bus number.
  • <NR3> specifies the AutoEthernet DATA source High threshold level for the specified bus, in volts.

type property

Return the BUS:B<x>:AUTOETHERnet:TYPe command.

Description
  • This command sets or queries the AutoEthernet standard speed.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:TYPe? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:TYPe value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:TYPe {HUNDREDBASET1}
- BUS:B<x>:AUTOETHERnet:TYPe?
Info
  • B<x> is the Bus number.
  • HUNDREDBASET1 specifies the AutoEthernet speed as 100Base-T1.

BusBItemAutoethernetBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUTOETHERnet:BITRate command.

Description
  • This command sets or queries bitrate for Auto Ethernet Bus with speed 10Base-T1S. The bit input for rate argument is NRF and output is NR1.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:BITRate value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:BITRate <NR1>
- BUS:B<x>:AUTOETHERnet:BITRate?
Info
  • B<x> is the Bus number.
  • <NR1> specifies the bit rate. The maximum value is 1 Gbps and the default is 10 Mbps.

BusBItemAutoethernetDataminusthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUTOETHERnet:DATAMINUSTHRESHOLD command.

Description
  • This command sets or queries the AutoEthernet D- source threshold level for the specified bus. This threshold only applies when the AutoEthernet signal type is single ended.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:DATAMINUSTHRESHOLD? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:DATAMINUSTHRESHOLD? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:DATAMINUSTHRESHOLD value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:DATAMINUSTHRESHOLD <NR3>
- BUS:B<x>:AUTOETHERnet:DATAMINUSTHRESHOLD?
Info
  • B<x> is the Bus number.
  • <NR3> specifies the AutoEthernet D- source threshold level for the specified bus, in volts. This threshold only applies when the AutoEthernet signal type is single ended.

BusBItemAutoethernetDataplusthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUTOETHERnet:DATAPLUSTHRESHold command.

Description
  • This command sets or queries the AutoEthernet D+ source threshold level for the specified bus. This threshold only applies when the AutoEthernet signal type is single ended.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:DATAPLUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:DATAPLUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:DATAPLUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:DATAPLUSTHRESHold <NR3>
- BUS:B<x>:AUTOETHERnet:DATAPLUSTHRESHold?
Info
  • B<x> is the Bus number.
  • <NR3> specifies the AutoEthernet D+ source threshold for the specified bus, in volts. This threshold only applies when the AutoEthernet signal type is single ended.

BusBItemAutoethernetLowdataminus

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUTOETHERnet:LOWDATAMINus command.

Description
  • This command sets or queries the AutoEthernet D- source low threshold level for the specified bus. This threshold only applies when the AutoEthernet signal type is single ended.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:LOWDATAMINus? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:LOWDATAMINus? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:LOWDATAMINus value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:LOWDATAMINus <NR3>
- BUS:B<x>:AUTOETHERnet:LOWDATAMINus?
Info
  • B<x> is the Bus number.
  • <NR3> specifies the AutoEthernet D- source low threshold level for the specified bus, in volts. This threshold only applies when the AutoEthernet signal type is single ended.

BusBItemAutoethernetLowdataplus

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUTOETHERnet:LOWDATAPLUS command.

Description
  • This command sets or queries the AutoEthernet D+ source low threshold level for the specified bus. This threshold only applies when the AutoEthernet signal type is single ended.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:LOWDATAPLUS? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:LOWDATAPLUS? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:LOWDATAPLUS value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:LOWDATAPLUS <NR3>
- BUS:B<x>:AUTOETHERnet:LOWDATAPLUS?
Info
  • B<x> is the Bus number.
  • <NR3> specifies the AutoEthernet D+ source low threshold level for the specified bus, in volts. This threshold only applies when the AutoEthernet signal type is single ended.

BusBItemAutoethernetLowthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUTOETHERnet:LOWTHRESHold command.

Description
  • This command sets or queries the AutoEthernet Data source Low threshold level for the specified bus. This threshold only applies when the AutoEthernet signal type is differential.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:LOWTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:LOWTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:LOWTHRESHold value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:LOWTHRESHold <NR3>
- BUS:B<x>:AUTOETHERnet:LOWTHRESHold?
Info
  • B<x> is the Bus number.
  • <NR3> specifies the AutoEthernet DATA source low threshold level for the specified bus, in volts. This threshold only applies when the AutoEthernet signal type is differential.

BusBItemAutoethernetPacketview

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUTOETHERnet:PACKetview command.

Description
  • This command sets or queries the Auto Ethernet bus Packet View when speed is selected as 10 Base-T1S.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:PACKetview? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:PACKetview? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:PACKetview value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:PACKetview {ON|OFF}
- BUS:B<x>:AUTOETHERnet:PACKetview?
Info
  • B<x> is the Bus number.
  • ON sets the packet view for the decode to on.
  • OFF sets the packet view for the decode to off.

BusBItemAutoethernetSignaltype

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUTOETHERnet:SIGNALTYpe command.

Description
  • This command sets or queries the AutoEthernet signal type for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:SIGNALTYpe? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:SIGNALTYpe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:SIGNALTYpe value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:AUTOETHERnet:SIGNALTYpe?
Info
  • B<x> is the Bus number.
  • SINGLE specifies single-ended signals.
  • DIFF specifies differential signals.

BusBItemAutoethernetSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUTOETHERnet:SOUrce command.

Description
  • This command sets or queries the AutoEthernet data (SDATA) source for the specified bus. This command controls the source channel when the signal type is differential.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:SOUrce value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:AUTOETHERnet:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the AutoEthernet data source for differential input, where is the channel number.
  • MATH<x> specifies a math channel as the AutoEthernet data source for differential input, where is the math waveform number.
  • REF<x> specifies a reference waveform as the AutoEthernet data source for differential input, where is the reference waveform number.
Properties
  • .dminus: The BUS:B<x>:AUTOETHERnet:SOUrce:DMINus command.
  • .dplus: The BUS:B<x>:AUTOETHERnet:SOUrce:DPLUs command.

dminus property

Return the BUS:B<x>:AUTOETHERnet:SOUrce:DMINus command.

Description
  • This command sets or queries the AutoEthernet D- source for the specified bus. This command specifies the source channel to use when the signal type is single ended.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:SOUrce:DMINus? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:SOUrce:DMINus? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:SOUrce:DMINus value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:SOUrce:DMINus {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:AUTOETHERnet:SOUrce:DMINus?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source for the D- signal, where is the channel number.
  • MATH<x> specifies a math channel as the source for the D- signal, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source for the D- signal, where is the reference waveform number.

dplus property

Return the BUS:B<x>:AUTOETHERnet:SOUrce:DPLUs command.

Description
  • This command sets or queries the AutoEthernet D+ source for the specified bus. This command specifies the source channel to use when the signal type is single ended.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:SOUrce:DPLUs? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:SOUrce:DPLUs? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:SOUrce:DPLUs value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:SOUrce:DPLUs {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:AUTOETHERnet:SOUrce:DPLUs?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source for the D+ signal, where is the channel number.
  • MATH<x> specifies a math channel as the source for the D+ signal, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source for the D+ signal, where is the reference waveform number.

BusBItemAutoethernetSourceDminus

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUTOETHERnet:SOUrce:DMINus command.

Description
  • This command sets or queries the AutoEthernet D- source for the specified bus. This command specifies the source channel to use when the signal type is single ended.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:SOUrce:DMINus? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:SOUrce:DMINus? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:SOUrce:DMINus value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:SOUrce:DMINus {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:AUTOETHERnet:SOUrce:DMINus?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source for the D- signal, where is the channel number.
  • MATH<x> specifies a math channel as the source for the D- signal, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source for the D- signal, where is the reference waveform number.

BusBItemAutoethernetSourceDplus

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUTOETHERnet:SOUrce:DPLUs command.

Description
  • This command sets or queries the AutoEthernet D+ source for the specified bus. This command specifies the source channel to use when the signal type is single ended.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:SOUrce:DPLUs? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:SOUrce:DPLUs? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:SOUrce:DPLUs value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:SOUrce:DPLUs {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:AUTOETHERnet:SOUrce:DPLUs?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source for the D+ signal, where is the channel number.
  • MATH<x> specifies a math channel as the source for the D+ signal, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source for the D+ signal, where is the reference waveform number.

BusBItemAutoethernetTbt1s

Bases: SCPICmdRead

The BUS:B<x>:AUTOETHERnet:TBT1s command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:TBT1s? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:TBT1s? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Properties
  • .mthreshold: The BUS:B<x>:AUTOETHERnet:TBT1s:MTHReshold command.

mthreshold property

Return the BUS:B<x>:AUTOETHERnet:TBT1s:MTHReshold command.

Description
  • This command sets or queries Auto Ethernet Bus with speed 10 Base-T1S math source threshold for single-ended for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:TBT1s:MTHReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:TBT1s:MTHReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:TBT1s:MTHReshold value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:TBT1s:MTHReshold <NR3>
- BUS:B<x>:AUTOETHERnet:TBT1s:MTHReshold?
Info
  • B<x> is the Bus number.
  • <NR3> specifies the AutoEthernet D+ source threshold for the specified bus, in volts.

BusBItemAutoethernetTbt1sMthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUTOETHERnet:TBT1s:MTHReshold command.

Description
  • This command sets or queries Auto Ethernet Bus with speed 10 Base-T1S math source threshold for single-ended for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:TBT1s:MTHReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:TBT1s:MTHReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:TBT1s:MTHReshold value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:TBT1s:MTHReshold <NR3>
- BUS:B<x>:AUTOETHERnet:TBT1s:MTHReshold?
Info
  • B<x> is the Bus number.
  • <NR3> specifies the AutoEthernet D+ source threshold for the specified bus, in volts.

BusBItemAutoethernetTenbaset1s

Bases: SCPICmdRead

The BUS:B<x>:AUTOETHERnet:TENBASET1S command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:TENBASET1S? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:TENBASET1S? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Properties
  • .threshold: The BUS:B<x>:AUTOETHERnet:TENBASET1S:THRESHold command.

threshold property

Return the BUS:B<x>:AUTOETHERnet:TENBASET1S:THRESHold command.

Description
  • This command sets or queries AutoEthernet Bus with speed 10 Base-T1S differential source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:TENBASET1S:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:TENBASET1S:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:TENBASET1S:THRESHold value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:TENBASET1S:THRESHold <NR3>
- BUS:B<x>:AUTOETHERnet:TENBASET1S:THRESHold?
Info
  • B<x> is the Bus number.
  • <NR3> specifies the AutoEthernet source threshold for the specified bus, in volts.

BusBItemAutoethernetTenbaset1sThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUTOETHERnet:TENBASET1S:THRESHold command.

Description
  • This command sets or queries AutoEthernet Bus with speed 10 Base-T1S differential source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:TENBASET1S:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:TENBASET1S:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:TENBASET1S:THRESHold value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:TENBASET1S:THRESHold <NR3>
- BUS:B<x>:AUTOETHERnet:TENBASET1S:THRESHold?
Info
  • B<x> is the Bus number.
  • <NR3> specifies the AutoEthernet source threshold for the specified bus, in volts.

BusBItemAutoethernetThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUTOETHERnet:THRESHold command.

Description
  • This command sets or queries the AutoEthernet DATA source High threshold level for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:THRESHold value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:THRESHold <NR3>
- BUS:B<x>:AUTOETHERnet:THRESHold?
Info
  • B<x> is the Bus number.
  • <NR3> specifies the AutoEthernet DATA source High threshold level for the specified bus, in volts.

BusBItemAutoethernetType

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:AUTOETHERnet:TYPe command.

Description
  • This command sets or queries the AutoEthernet standard speed.
Usage
  • Using the .query() method will send the BUS:B<x>:AUTOETHERnet:TYPe? query.
  • Using the .verify(value) method will send the BUS:B<x>:AUTOETHERnet:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:AUTOETHERnet:TYPe value command.
SCPI Syntax
- BUS:B<x>:AUTOETHERnet:TYPe {HUNDREDBASET1}
- BUS:B<x>:AUTOETHERnet:TYPe?
Info
  • B<x> is the Bus number.
  • HUNDREDBASET1 specifies the AutoEthernet speed as 100Base-T1.

BusBItemCan

Bases: SCPICmdRead

The BUS:B<x>:CAN command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CAN? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .bitrate: The BUS:B<x>:CAN:BITRate command.
  • .fd: The BUS:B<x>:CAN:FD command tree.
  • .samplepoint: The BUS:B<x>:CAN:SAMPLEpoint command.
  • .signal: The BUS:B<x>:CAN:SIGNal command.
  • .source: The BUS:B<x>:CAN:SOUrce command.
  • .standard: The BUS:B<x>:CAN:STANDard command.
  • .threshold: The BUS:B<x>:CAN:THReshold command.
  • .xl: The BUS:B<x>:CAN:XL command tree.

bitrate property

Return the BUS:B<x>:CAN:BITRate command.

Description
  • This command sets or queries the CAN bit rate. The bus number is specified by x. If you select Custom, use BUS:B<x>:CAN:BITRate:VALue to set the bit rate.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:BITRate value command.
SCPI Syntax
- BUS:B<x>:CAN:BITRate {RATE10K|RATE100K|RATE1M|RATE125K|RATE153K|RATE20K|RATE25K|RATE250K|RATE31K|RATE33K|RATE400K|RATE50K|RATE500K|RATE62K|RATE68K|RATE800K|RATE83K|RATE92K|CUSTom}
- BUS:B<x>:CAN:BITRate?
Info
  • B<x> is the number of the bus.
  • RATE10K sets the CAN bit rate to 10 k.
  • RATE100K sets the CAN bit rate to 100 k.
  • RATE1M sets the CAN bit rate to 1 Mb.
  • RATE125K sets the CAN bit rate to 125 k.
  • RATE153K sets the CAN bit rate to 153 k.
  • RATE20K sets the CAN bit rate to 20 k.
  • RATE25K sets the CAN bit rate to 25 k.
  • RATE250K sets the CAN bit rate to 250 k.
  • RATE31K sets the CAN bit rate to 31 k.
  • RATE33K sets the CAN bit rate to 33 k.
  • RATE400K sets the CAN bit rate to 400 k.
  • RATE50K sets the CAN bit rate to 50 k.
  • RATE500K sets the CAN bit rate to 500 k.
  • RATE62K sets the CAN bit rate to 62 k.
  • RATE68K sets the CAN bit rate to 68 k.
  • RATE800K sets the CAN bit rate to 800 k.
  • RATE83K sets the CAN bit rate to 83 k.
  • RATE92K sets the CAN bit rate to 92 k.
  • CUSTom sets a custom CAN bit rate.
Sub-properties
  • .value: The BUS:B<x>:CAN:BITRate:VALue command.

fd property

Return the BUS:B<x>:CAN:FD command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CAN:FD? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:FD? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .bitrate: The BUS:B<x>:CAN:FD:BITRate command.

samplepoint property

Return the BUS:B<x>:CAN:SAMPLEpoint command.

Description
  • This command sets or queries the sample point for the specified CAN bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:SAMPLEpoint? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:SAMPLEpoint? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:SAMPLEpoint value command.
SCPI Syntax
- BUS:B<x>:CAN:SAMPLEpoint <NR1>
- BUS:B<x>:CAN:SAMPLEpoint?
Info
  • B<x> is the number of the bus.
  • <NR1> is the sample point, in percent, for the specified CAN bus.

signal property

Return the BUS:B<x>:CAN:SIGNal command.

Description
  • This command sets or queries the signal type for the specified CAN bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:SIGNal? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:SIGNal? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:SIGNal value command.
SCPI Syntax
- BUS:B<x>:CAN:SIGNal {DIFFerential|CANH|CANL|RX|TX}
- BUS:B<x>:CAN:SIGNal?
Info
  • B<x> is the number of the bus.
  • DIFFerential sets the signal type to differential.
  • CANH sets the signal type to CANH.
  • CANL sets the signal type to CANL.
  • RX sets the signal type to RX.
  • TX sets the signal type to TX.

source property

Return the BUS:B<x>:CAN:SOUrce command.

Description
  • This command sets or queries the CAN source channel for the specified CAN bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CAN:SOUrce {CH<x>|CH<x>_D<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:CAN:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • CH<x>_D<x> specifies a digital channel to use as the source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.
  • REF<x>_D<x> specifies a digital reference waveform as the source.

standard property

Return the BUS:B<x>:CAN:STANDard command.

Description
  • This command sets or queries which CAN standard specification to analyze the specified CAN bus with. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:STANDard? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:STANDard? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:STANDard value command.
SCPI Syntax
- BUS:B<x>:CAN:STANDard {CAN2X|CAN2X_FDISO_XL|CAN_XL|FDISO|FDNONISO}
- BUS:B<x>:CAN:STANDard?
Info
  • B<x> is the number of the bus.
  • CAN2X sets the CAN bus standard to CAN 2.0.
  • CAN2X_FDISO_XL sets the CAN bus standard to a combination of BASE CAN, CAN FD and CAN XL.
  • CAN_XL sets the CAN bus standard to CAN XL.
  • FDISO sets the CAN bus standard to ISO CAN FD (11898-1:2015).
  • FDNONISO sets the CAN bus standard to non-ISO CAN FD (Bosch:2012).

threshold property

Return the BUS:B<x>:CAN:THReshold command.

Description
  • This command sets or queries the source channel threshold for the specified CAN bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:THReshold value command.
SCPI Syntax
- BUS:B<x>:CAN:THReshold <NR3>
- BUS:B<x>:CAN:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the source channel threshold for the specified CAN bus.

xl property

Return the BUS:B<x>:CAN:XL command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CAN:XL? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:XL? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .bitrate: The BUS:B<x>:CAN:XL:BITRate command.

BusBItemCanBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CAN:BITRate command.

Description
  • This command sets or queries the CAN bit rate. The bus number is specified by x. If you select Custom, use BUS:B<x>:CAN:BITRate:VALue to set the bit rate.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:BITRate value command.
SCPI Syntax
- BUS:B<x>:CAN:BITRate {RATE10K|RATE100K|RATE1M|RATE125K|RATE153K|RATE20K|RATE25K|RATE250K|RATE31K|RATE33K|RATE400K|RATE50K|RATE500K|RATE62K|RATE68K|RATE800K|RATE83K|RATE92K|CUSTom}
- BUS:B<x>:CAN:BITRate?
Info
  • B<x> is the number of the bus.
  • RATE10K sets the CAN bit rate to 10 k.
  • RATE100K sets the CAN bit rate to 100 k.
  • RATE1M sets the CAN bit rate to 1 Mb.
  • RATE125K sets the CAN bit rate to 125 k.
  • RATE153K sets the CAN bit rate to 153 k.
  • RATE20K sets the CAN bit rate to 20 k.
  • RATE25K sets the CAN bit rate to 25 k.
  • RATE250K sets the CAN bit rate to 250 k.
  • RATE31K sets the CAN bit rate to 31 k.
  • RATE33K sets the CAN bit rate to 33 k.
  • RATE400K sets the CAN bit rate to 400 k.
  • RATE50K sets the CAN bit rate to 50 k.
  • RATE500K sets the CAN bit rate to 500 k.
  • RATE62K sets the CAN bit rate to 62 k.
  • RATE68K sets the CAN bit rate to 68 k.
  • RATE800K sets the CAN bit rate to 800 k.
  • RATE83K sets the CAN bit rate to 83 k.
  • RATE92K sets the CAN bit rate to 92 k.
  • CUSTom sets a custom CAN bit rate.
Properties
  • .value: The BUS:B<x>:CAN:BITRate:VALue command.

value property

Return the BUS:B<x>:CAN:BITRate:VALue command.

Description
  • This command sets or queries CAN bit rate when Custom is selected by BUS:B<x>:CAN:BITRate. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:BITRate:VALue? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:BITRate:VALue? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:BITRate:VALue value command.
SCPI Syntax
- BUS:B<x>:CAN:BITRate:VALue <NR3>
- BUS:B<x>:CAN:BITRate:VALue?
Info
  • B<x> is the number of the bus.
  • <NR3> specifies the CAN bit rate.

BusBItemCanBitrateValue

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CAN:BITRate:VALue command.

Description
  • This command sets or queries CAN bit rate when Custom is selected by BUS:B<x>:CAN:BITRate. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:BITRate:VALue? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:BITRate:VALue? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:BITRate:VALue value command.
SCPI Syntax
- BUS:B<x>:CAN:BITRate:VALue <NR3>
- BUS:B<x>:CAN:BITRate:VALue?
Info
  • B<x> is the number of the bus.
  • <NR3> specifies the CAN bit rate.

BusBItemCanFd

Bases: SCPICmdRead

The BUS:B<x>:CAN:FD command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CAN:FD? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:FD? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .bitrate: The BUS:B<x>:CAN:FD:BITRate command.

bitrate property

Return the BUS:B<x>:CAN:FD:BITRate command.

Description
  • This command sets or queries the increased data phase bit rate used by CAN FD packets on the specified CAN bus. The bus is specified by x. If you select Custom, use BUS:B<x>:CAN:FD:BITRate:CUSTom to set the bit rate.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:FD:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:FD:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:FD:BITRate value command.
SCPI Syntax
- BUS:B<x>:CAN:FD:BITRate {RATE1M|RATE2M|RATE3M|RATE4M|RATE5M|RATE6M|RATE7M|RATE8M|RATE9M|RATE10M|RATE11M|RATE12M|RATE13M|RATE14M|RATE15M|RATE16M|CUSTom}
- BUS:B<x>:CAN:FD:BITRate?
Info
  • B<x> is the number of the bus.
  • RATE1M sets the CAN FD bit rate to 1 Mbps.
  • RATE2M sets the CAN FD bit rate to 2 Mbps.
  • RATE3M sets the CAN FD bit rate to 3 Mbps.
  • RATE4M sets the CAN FD bit rate to 4 Mbps.
  • RATE5M sets the CAN FD bit rate to 5 Mbps.
  • RATE6M sets the CAN FD bit rate to 6 Mbps.
  • RATE7M sets the CAN FD bit rate to 7 Mbps.
  • RATE8M sets the CAN FD bit rate to 8 Mbps.
  • RATE9M sets the CAN FD bit rate to 9 Mbps.
  • RATE10M sets the CAN FD bit rate to 10 Mbps.
  • RATE11M sets the CAN FD bit rate to 11 Mbps.
  • RATE12M sets the CAN FD bit rate to 12 Mbps.
  • RATE13M sets the CAN FD bit rate to 13 Mbps.
  • RATE14M sets the CAN FD bit rate to 14 Mbps.
  • RATE15M sets the CAN FD bit rate to 15 Mbps.
  • RATE16M sets the CAN FD bit rate to 16 Mbps.
  • CUSTom sets a custom CAN FD bit rate.
Sub-properties
  • .custom: The BUS:B<x>:CAN:FD:BITRate:CUSTom command.

BusBItemCanFdBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CAN:FD:BITRate command.

Description
  • This command sets or queries the increased data phase bit rate used by CAN FD packets on the specified CAN bus. The bus is specified by x. If you select Custom, use BUS:B<x>:CAN:FD:BITRate:CUSTom to set the bit rate.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:FD:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:FD:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:FD:BITRate value command.
SCPI Syntax
- BUS:B<x>:CAN:FD:BITRate {RATE1M|RATE2M|RATE3M|RATE4M|RATE5M|RATE6M|RATE7M|RATE8M|RATE9M|RATE10M|RATE11M|RATE12M|RATE13M|RATE14M|RATE15M|RATE16M|CUSTom}
- BUS:B<x>:CAN:FD:BITRate?
Info
  • B<x> is the number of the bus.
  • RATE1M sets the CAN FD bit rate to 1 Mbps.
  • RATE2M sets the CAN FD bit rate to 2 Mbps.
  • RATE3M sets the CAN FD bit rate to 3 Mbps.
  • RATE4M sets the CAN FD bit rate to 4 Mbps.
  • RATE5M sets the CAN FD bit rate to 5 Mbps.
  • RATE6M sets the CAN FD bit rate to 6 Mbps.
  • RATE7M sets the CAN FD bit rate to 7 Mbps.
  • RATE8M sets the CAN FD bit rate to 8 Mbps.
  • RATE9M sets the CAN FD bit rate to 9 Mbps.
  • RATE10M sets the CAN FD bit rate to 10 Mbps.
  • RATE11M sets the CAN FD bit rate to 11 Mbps.
  • RATE12M sets the CAN FD bit rate to 12 Mbps.
  • RATE13M sets the CAN FD bit rate to 13 Mbps.
  • RATE14M sets the CAN FD bit rate to 14 Mbps.
  • RATE15M sets the CAN FD bit rate to 15 Mbps.
  • RATE16M sets the CAN FD bit rate to 16 Mbps.
  • CUSTom sets a custom CAN FD bit rate.
Properties
  • .custom: The BUS:B<x>:CAN:FD:BITRate:CUSTom command.

custom property

Return the BUS:B<x>:CAN:FD:BITRate:CUSTom command.

Description
  • This command sets or queries the custom bit rate for the increased data phase of CAN FD packets on the specified CAN bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:FD:BITRate:CUSTom? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:FD:BITRate:CUSTom? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:FD:BITRate:CUSTom value command.
SCPI Syntax
- BUS:B<x>:CAN:FD:BITRate:CUSTom <NR1>
- BUS:B<x>:CAN:FD:BITRate:CUSTom?
Info
  • B<x> is the number of the bus.
  • <NR1> is the custom FD bit rate for the specified bus.

BusBItemCanFdBitrateCustom

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CAN:FD:BITRate:CUSTom command.

Description
  • This command sets or queries the custom bit rate for the increased data phase of CAN FD packets on the specified CAN bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:FD:BITRate:CUSTom? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:FD:BITRate:CUSTom? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:FD:BITRate:CUSTom value command.
SCPI Syntax
- BUS:B<x>:CAN:FD:BITRate:CUSTom <NR1>
- BUS:B<x>:CAN:FD:BITRate:CUSTom?
Info
  • B<x> is the number of the bus.
  • <NR1> is the custom FD bit rate for the specified bus.

BusBItemCanSamplepoint

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CAN:SAMPLEpoint command.

Description
  • This command sets or queries the sample point for the specified CAN bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:SAMPLEpoint? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:SAMPLEpoint? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:SAMPLEpoint value command.
SCPI Syntax
- BUS:B<x>:CAN:SAMPLEpoint <NR1>
- BUS:B<x>:CAN:SAMPLEpoint?
Info
  • B<x> is the number of the bus.
  • <NR1> is the sample point, in percent, for the specified CAN bus.

BusBItemCanSignal

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CAN:SIGNal command.

Description
  • This command sets or queries the signal type for the specified CAN bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:SIGNal? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:SIGNal? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:SIGNal value command.
SCPI Syntax
- BUS:B<x>:CAN:SIGNal {DIFFerential|CANH|CANL|RX|TX}
- BUS:B<x>:CAN:SIGNal?
Info
  • B<x> is the number of the bus.
  • DIFFerential sets the signal type to differential.
  • CANH sets the signal type to CANH.
  • CANL sets the signal type to CANL.
  • RX sets the signal type to RX.
  • TX sets the signal type to TX.

BusBItemCanSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CAN:SOUrce command.

Description
  • This command sets or queries the CAN source channel for the specified CAN bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CAN:SOUrce {CH<x>|CH<x>_D<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:CAN:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • CH<x>_D<x> specifies a digital channel to use as the source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.
  • REF<x>_D<x> specifies a digital reference waveform as the source.

BusBItemCanStandard

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CAN:STANDard command.

Description
  • This command sets or queries which CAN standard specification to analyze the specified CAN bus with. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:STANDard? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:STANDard? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:STANDard value command.
SCPI Syntax
- BUS:B<x>:CAN:STANDard {CAN2X|CAN2X_FDISO_XL|CAN_XL|FDISO|FDNONISO}
- BUS:B<x>:CAN:STANDard?
Info
  • B<x> is the number of the bus.
  • CAN2X sets the CAN bus standard to CAN 2.0.
  • CAN2X_FDISO_XL sets the CAN bus standard to a combination of BASE CAN, CAN FD and CAN XL.
  • CAN_XL sets the CAN bus standard to CAN XL.
  • FDISO sets the CAN bus standard to ISO CAN FD (11898-1:2015).
  • FDNONISO sets the CAN bus standard to non-ISO CAN FD (Bosch:2012).

BusBItemCanThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CAN:THReshold command.

Description
  • This command sets or queries the source channel threshold for the specified CAN bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:THReshold value command.
SCPI Syntax
- BUS:B<x>:CAN:THReshold <NR3>
- BUS:B<x>:CAN:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the source channel threshold for the specified CAN bus.

BusBItemCanXl

Bases: SCPICmdRead

The BUS:B<x>:CAN:XL command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CAN:XL? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:XL? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .bitrate: The BUS:B<x>:CAN:XL:BITRate command.

bitrate property

Return the BUS:B<x>:CAN:XL:BITRate command.

Description
  • This command sets or queries the bit rate for the CAN XL bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:XL:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:XL:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:XL:BITRate value command.
SCPI Syntax
- BUS:B<x>:CAN:XL:BITRate <NR1>
- BUS:B<x>:CAN:XL:BITRate?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the CAN XL data transfer bit rate. The valid range is from 1000000 to 20000000. The default value is 4000000.
Sub-properties
  • .custom: The BUS:B<x>:CAN:XL:BITRate:CUSTom command.

BusBItemCanXlBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CAN:XL:BITRate command.

Description
  • This command sets or queries the bit rate for the CAN XL bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:XL:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:XL:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:XL:BITRate value command.
SCPI Syntax
- BUS:B<x>:CAN:XL:BITRate <NR1>
- BUS:B<x>:CAN:XL:BITRate?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the CAN XL data transfer bit rate. The valid range is from 1000000 to 20000000. The default value is 4000000.
Properties
  • .custom: The BUS:B<x>:CAN:XL:BITRate:CUSTom command.

custom property

Return the BUS:B<x>:CAN:XL:BITRate:CUSTom command.

Description
  • This command sets or queries the custom bit rate for the CAN XL bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:XL:BITRate:CUSTom? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:XL:BITRate:CUSTom? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:XL:BITRate:CUSTom value command.
SCPI Syntax
- BUS:B<x>:CAN:XL:BITRate:CUSTom <NR1>
- BUS:B<x>:CAN:XL:BITRate:CUSTom?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the CAN XL data transfer custom bit rate. The valid range is from 500000 to 20000000. The default value is 4000000.

BusBItemCanXlBitrateCustom

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CAN:XL:BITRate:CUSTom command.

Description
  • This command sets or queries the custom bit rate for the CAN XL bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CAN:XL:BITRate:CUSTom? query.
  • Using the .verify(value) method will send the BUS:B<x>:CAN:XL:BITRate:CUSTom? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CAN:XL:BITRate:CUSTom value command.
SCPI Syntax
- BUS:B<x>:CAN:XL:BITRate:CUSTom <NR1>
- BUS:B<x>:CAN:XL:BITRate:CUSTom?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the CAN XL data transfer custom bit rate. The valid range is from 500000 to 20000000. The default value is 4000000.

BusBItemCphy

Bases: SCPICmdRead

The BUS:B<x>:CPHY command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .a: The BUS:B<x>:CPHY:A command tree.
  • .ab: The BUS:B<x>:CPHY:AB command tree.
  • .agnd: The BUS:B<x>:CPHY:AGND command tree.
  • .b: The BUS:B<x>:CPHY:B command tree.
  • .bc: The BUS:B<x>:CPHY:BC command tree.
  • .bitrate: The BUS:B<x>:CPHY:BITRate command.
  • .c: The BUS:B<x>:CPHY:C command tree.
  • .ca: The BUS:B<x>:CPHY:CA command tree.
  • .cgnd: The BUS:B<x>:CPHY:CGND command tree.
  • .lp: The BUS:B<x>:CPHY:LP command tree.
  • .signaltype: The BUS:B<x>:CPHY:SIGNALTYpe command.
  • .subtype: The BUS:B<x>:CPHY:SUBTYPe command.

a property

Return the BUS:B<x>:CPHY:A command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:A? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:A? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .data: The BUS:B<x>:CPHY:A:DATA command tree.
  • .lp: The BUS:B<x>:CPHY:A:LP command tree.
  • .source: The BUS:B<x>:CPHY:A:SOUrce command.

ab property

Return the BUS:B<x>:CPHY:AB command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:AB? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:AB? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .source: The BUS:B<x>:CPHY:AB:SOUrce command.
  • .threshold: The BUS:B<x>:CPHY:AB:THReshold command.

agnd property

Return the BUS:B<x>:CPHY:AGND command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:AGND? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:AGND? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .source: The BUS:B<x>:CPHY:AGND:SOUrce command.
  • .threshold: The BUS:B<x>:CPHY:AGND:THReshold command.

b property

Return the BUS:B<x>:CPHY:B command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:B? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:B? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .data: The BUS:B<x>:CPHY:B:DATA command tree.
  • .source: The BUS:B<x>:CPHY:B:SOUrce command.

bc property

Return the BUS:B<x>:CPHY:BC command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:BC? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:BC? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .source: The BUS:B<x>:CPHY:BC:SOUrce command.
  • .threshold: The BUS:B<x>:CPHY:BC:THReshold command.

bitrate property

Return the BUS:B<x>:CPHY:BITRate command.

Description
  • This command sets or queries the bit rate for the specified CPHY bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:BITRate value command.
SCPI Syntax
- BUS:B<x>:CPHY:BITRate <NR1>
- BUS:B<x>:CPHY:BITRate?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the bit rate. The default bit rate is 1 Gbps and varies 4 Mbps ~ 10Gps.

c property

Return the BUS:B<x>:CPHY:C command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:C? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:C? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .data: The BUS:B<x>:CPHY:C:DATA command tree.
  • .lp: The BUS:B<x>:CPHY:C:LP command tree.
  • .source: The BUS:B<x>:CPHY:C:SOUrce command.

ca property

Return the BUS:B<x>:CPHY:CA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:CA? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:CA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .source: The BUS:B<x>:CPHY:CA:SOUrce command.
  • .threshold: The BUS:B<x>:CPHY:CA:THReshold command.

cgnd property

Return the BUS:B<x>:CPHY:CGND command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:CGND? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:CGND? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .source: The BUS:B<x>:CPHY:CGND:SOUrce command.
  • .threshold: The BUS:B<x>:CPHY:CGND:THReshold command.

lp property

Return the BUS:B<x>:CPHY:LP command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:LP? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:LP? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .direction: The BUS:B<x>:CPHY:LP:DIRection command.

signaltype property

Return the BUS:B<x>:CPHY:SIGNALTYpe command.

Description
  • This command sets or queries the signal type for CPHY bus decode. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:SIGNALTYpe? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:SIGNALTYpe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:SIGNALTYpe value command.
SCPI Syntax
- BUS:B<x>:CPHY:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:CPHY:SIGNALTYpe?
Info
  • B<x> is the number of the bus.
  • SINGLE specifies the signal type to single ended.
  • DIFF specifies the signal type to differential. The default type is differential.

subtype property

Return the BUS:B<x>:CPHY:SUBTYPe command.

Description
  • This command sets or queries the sub type for CPHY bus decode. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:SUBTYPe? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:SUBTYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:SUBTYPe value command.
SCPI Syntax
- BUS:B<x>:CPHY:SUBTYPe {CSI|DSI|Word|Symbol}
- BUS:B<x>:CPHY:SUBTYPe?
Info
  • B<x> is the number of the bus.
  • CSI specifies the protocol type to CSI. The default type is CSI.
  • DSI specifies the protocol type to DSI.
  • Word specifies the protocol type to word.
  • Symbol specifies the protocol type to symbol.

BusBItemCphyA

Bases: SCPICmdRead

The BUS:B<x>:CPHY:A command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:A? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:A? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .data: The BUS:B<x>:CPHY:A:DATA command tree.
  • .lp: The BUS:B<x>:CPHY:A:LP command tree.
  • .source: The BUS:B<x>:CPHY:A:SOUrce command.

data property

Return the BUS:B<x>:CPHY:A:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:A:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:A:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .threshold: The BUS:B<x>:CPHY:A:DATA:THRESHold command.

lp property

Return the BUS:B<x>:CPHY:A:LP command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:A:LP? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:A:LP? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .threshold: The BUS:B<x>:CPHY:A:LP:THRESHold command.

source property

Return the BUS:B<x>:CPHY:A:SOUrce command.

Description
  • This command sets or queries the CPHY Single Ended A source for the specified bus line. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:A:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:A:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:A:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CPHY:A:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:A:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies an analog reference waveform as the source.

BusBItemCphyAData

Bases: SCPICmdRead

The BUS:B<x>:CPHY:A:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:A:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:A:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .threshold: The BUS:B<x>:CPHY:A:DATA:THRESHold command.

threshold property

Return the BUS:B<x>:CPHY:A:DATA:THRESHold command.

Description
  • This command sets or queries the CPHY Single Ended A source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:A:DATA:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:A:DATA:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:A:DATA:THRESHold value command.
SCPI Syntax
- BUS:B<x>:CPHY:A:DATA:THRESHold <NR3>
- BUS:B<x>:CPHY:A:DATA:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY Single Ended A threshold in Volts for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyADataThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:A:DATA:THRESHold command.

Description
  • This command sets or queries the CPHY Single Ended A source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:A:DATA:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:A:DATA:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:A:DATA:THRESHold value command.
SCPI Syntax
- BUS:B<x>:CPHY:A:DATA:THRESHold <NR3>
- BUS:B<x>:CPHY:A:DATA:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY Single Ended A threshold in Volts for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyALp

Bases: SCPICmdRead

The BUS:B<x>:CPHY:A:LP command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:A:LP? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:A:LP? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .threshold: The BUS:B<x>:CPHY:A:LP:THRESHold command.

threshold property

Return the BUS:B<x>:CPHY:A:LP:THRESHold command.

Description
  • This command sets or queries the CPHY Single Ended A LP threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:A:LP:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:A:LP:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:A:LP:THRESHold value command.
SCPI Syntax
- BUS:B<x>:CPHY:A:LP:THRESHold <NR3>
- BUS:B<x>:CPHY:A:LP:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY Single Ended A LP threshold in Volts for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyALpThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:A:LP:THRESHold command.

Description
  • This command sets or queries the CPHY Single Ended A LP threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:A:LP:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:A:LP:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:A:LP:THRESHold value command.
SCPI Syntax
- BUS:B<x>:CPHY:A:LP:THRESHold <NR3>
- BUS:B<x>:CPHY:A:LP:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY Single Ended A LP threshold in Volts for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyASource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:A:SOUrce command.

Description
  • This command sets or queries the CPHY Single Ended A source for the specified bus line. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:A:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:A:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:A:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CPHY:A:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:A:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies an analog reference waveform as the source.

BusBItemCphyAb

Bases: SCPICmdRead

The BUS:B<x>:CPHY:AB command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:AB? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:AB? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .source: The BUS:B<x>:CPHY:AB:SOUrce command.
  • .threshold: The BUS:B<x>:CPHY:AB:THReshold command.

source property

Return the BUS:B<x>:CPHY:AB:SOUrce command.

Description
  • This command sets or queries the CPHY differential AB source for the specified bus line. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:AB:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:AB:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:AB:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CPHY:AB:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:AB:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog/digital channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies an analog/digital reference waveform as the source.

threshold property

Return the BUS:B<x>:CPHY:AB:THReshold command.

Description
  • This command sets or queries the CPHY differential AB source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:AB:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:AB:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:AB:THReshold value command.
SCPI Syntax
- BUS:B<x>:CPHY:AB:THReshold <NR3>
- BUS:B<x>:CPHY:AB:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY differential AB threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyAbSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:AB:SOUrce command.

Description
  • This command sets or queries the CPHY differential AB source for the specified bus line. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:AB:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:AB:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:AB:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CPHY:AB:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:AB:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog/digital channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies an analog/digital reference waveform as the source.

BusBItemCphyAbThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:AB:THReshold command.

Description
  • This command sets or queries the CPHY differential AB source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:AB:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:AB:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:AB:THReshold value command.
SCPI Syntax
- BUS:B<x>:CPHY:AB:THReshold <NR3>
- BUS:B<x>:CPHY:AB:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY differential AB threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyAgnd

Bases: SCPICmdRead

The BUS:B<x>:CPHY:AGND command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:AGND? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:AGND? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .source: The BUS:B<x>:CPHY:AGND:SOUrce command.
  • .threshold: The BUS:B<x>:CPHY:AGND:THReshold command.

source property

Return the BUS:B<x>:CPHY:AGND:SOUrce command.

Description
  • This command sets or queries the CPHY differential AGND source for the specified bus line. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:AGND:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:AGND:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:AGND:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CPHY:AGND:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:AGND:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog/digital channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies an analog/digital reference waveform as the source.

threshold property

Return the BUS:B<x>:CPHY:AGND:THReshold command.

Description
  • This command sets or queries the CPHY differential AGND source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:AGND:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:AGND:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:AGND:THReshold value command.
SCPI Syntax
- BUS:B<x>:CPHY:AGND:THReshold <NR3>
- BUS:B<x>:CPHY:AGND:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY differential A GND threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyAgndSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:AGND:SOUrce command.

Description
  • This command sets or queries the CPHY differential AGND source for the specified bus line. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:AGND:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:AGND:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:AGND:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CPHY:AGND:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:AGND:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog/digital channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies an analog/digital reference waveform as the source.

BusBItemCphyAgndThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:AGND:THReshold command.

Description
  • This command sets or queries the CPHY differential AGND source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:AGND:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:AGND:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:AGND:THReshold value command.
SCPI Syntax
- BUS:B<x>:CPHY:AGND:THReshold <NR3>
- BUS:B<x>:CPHY:AGND:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY differential A GND threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyB

Bases: SCPICmdRead

The BUS:B<x>:CPHY:B command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:B? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:B? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .data: The BUS:B<x>:CPHY:B:DATA command tree.
  • .source: The BUS:B<x>:CPHY:B:SOUrce command.

data property

Return the BUS:B<x>:CPHY:B:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:B:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:B:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .threshold: The BUS:B<x>:CPHY:B:DATA:THRESHold command.

source property

Return the BUS:B<x>:CPHY:B:SOUrce command.

Description
  • This command sets or queries the CPHY Single Ended B source for the specified bus line. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:B:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:B:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:B:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CPHY:B:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:B:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies an analog reference waveform as the source.

BusBItemCphyBData

Bases: SCPICmdRead

The BUS:B<x>:CPHY:B:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:B:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:B:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .threshold: The BUS:B<x>:CPHY:B:DATA:THRESHold command.

threshold property

Return the BUS:B<x>:CPHY:B:DATA:THRESHold command.

Description
  • This command sets or queries the CPHY Single Ended B source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:B:DATA:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:B:DATA:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:B:DATA:THRESHold value command.
SCPI Syntax
- BUS:B<x>:CPHY:B:DATA:THRESHold <NR3>
- BUS:B<x>:CPHY:B:DATA:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY Single Ended B threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyBDataThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:B:DATA:THRESHold command.

Description
  • This command sets or queries the CPHY Single Ended B source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:B:DATA:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:B:DATA:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:B:DATA:THRESHold value command.
SCPI Syntax
- BUS:B<x>:CPHY:B:DATA:THRESHold <NR3>
- BUS:B<x>:CPHY:B:DATA:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY Single Ended B threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyBSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:B:SOUrce command.

Description
  • This command sets or queries the CPHY Single Ended B source for the specified bus line. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:B:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:B:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:B:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CPHY:B:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:B:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies an analog reference waveform as the source.

BusBItemCphyBc

Bases: SCPICmdRead

The BUS:B<x>:CPHY:BC command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:BC? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:BC? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .source: The BUS:B<x>:CPHY:BC:SOUrce command.
  • .threshold: The BUS:B<x>:CPHY:BC:THReshold command.

source property

Return the BUS:B<x>:CPHY:BC:SOUrce command.

Description
  • This command sets or queries the CPHY differential BC source for the specified bus line. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:BC:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:BC:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:BC:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CPHY:BC:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:BC:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog/digital channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies an analog/digital reference waveform as the source.

threshold property

Return the BUS:B<x>:CPHY:BC:THReshold command.

Description
  • This command sets or queries the CPHY differential BC source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:BC:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:BC:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:BC:THReshold value command.
SCPI Syntax
- BUS:B<x>:CPHY:BC:THReshold <NR3>
- BUS:B<x>:CPHY:BC:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY differential BC threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyBcSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:BC:SOUrce command.

Description
  • This command sets or queries the CPHY differential BC source for the specified bus line. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:BC:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:BC:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:BC:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CPHY:BC:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:BC:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog/digital channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies an analog/digital reference waveform as the source.

BusBItemCphyBcThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:BC:THReshold command.

Description
  • This command sets or queries the CPHY differential BC source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:BC:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:BC:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:BC:THReshold value command.
SCPI Syntax
- BUS:B<x>:CPHY:BC:THReshold <NR3>
- BUS:B<x>:CPHY:BC:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY differential BC threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:BITRate command.

Description
  • This command sets or queries the bit rate for the specified CPHY bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:BITRate value command.
SCPI Syntax
- BUS:B<x>:CPHY:BITRate <NR1>
- BUS:B<x>:CPHY:BITRate?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the bit rate. The default bit rate is 1 Gbps and varies 4 Mbps ~ 10Gps.

BusBItemCphyC

Bases: SCPICmdRead

The BUS:B<x>:CPHY:C command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:C? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:C? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .data: The BUS:B<x>:CPHY:C:DATA command tree.
  • .lp: The BUS:B<x>:CPHY:C:LP command tree.
  • .source: The BUS:B<x>:CPHY:C:SOUrce command.

data property

Return the BUS:B<x>:CPHY:C:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:C:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:C:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .threshold: The BUS:B<x>:CPHY:C:DATA:THRESHold command.

lp property

Return the BUS:B<x>:CPHY:C:LP command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:C:LP? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:C:LP? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .threshold: The BUS:B<x>:CPHY:C:LP:THRESHold command.

source property

Return the BUS:B<x>:CPHY:C:SOUrce command.

Description
  • This command sets or queries the CPHY Single Ended C source for the specified bus line. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:C:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:C:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:C:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CPHY:C:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:C:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies an analog reference waveform as the source.

BusBItemCphyCData

Bases: SCPICmdRead

The BUS:B<x>:CPHY:C:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:C:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:C:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .threshold: The BUS:B<x>:CPHY:C:DATA:THRESHold command.

threshold property

Return the BUS:B<x>:CPHY:C:DATA:THRESHold command.

Description
  • This command sets or queries the CPHY Single Ended C source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:C:DATA:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:C:DATA:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:C:DATA:THRESHold value command.
SCPI Syntax
- BUS:B<x>:CPHY:C:DATA:THRESHold <NR3>
- BUS:B<x>:CPHY:C:DATA:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY Single Ended C threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyCDataThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:C:DATA:THRESHold command.

Description
  • This command sets or queries the CPHY Single Ended C source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:C:DATA:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:C:DATA:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:C:DATA:THRESHold value command.
SCPI Syntax
- BUS:B<x>:CPHY:C:DATA:THRESHold <NR3>
- BUS:B<x>:CPHY:C:DATA:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY Single Ended C threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyCLp

Bases: SCPICmdRead

The BUS:B<x>:CPHY:C:LP command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:C:LP? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:C:LP? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .threshold: The BUS:B<x>:CPHY:C:LP:THRESHold command.

threshold property

Return the BUS:B<x>:CPHY:C:LP:THRESHold command.

Description
  • This command sets or queries the CPHY Single Ended C LP threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:C:LP:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:C:LP:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:C:LP:THRESHold value command.
SCPI Syntax
- BUS:B<x>:CPHY:C:LP:THRESHold <NR3>
- BUS:B<x>:CPHY:C:LP:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY Single Ended C threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyCLpThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:C:LP:THRESHold command.

Description
  • This command sets or queries the CPHY Single Ended C LP threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:C:LP:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:C:LP:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:C:LP:THRESHold value command.
SCPI Syntax
- BUS:B<x>:CPHY:C:LP:THRESHold <NR3>
- BUS:B<x>:CPHY:C:LP:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY Single Ended C threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyCSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:C:SOUrce command.

Description
  • This command sets or queries the CPHY Single Ended C source for the specified bus line. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:C:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:C:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:C:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CPHY:C:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:C:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies an analog reference waveform as the source.

BusBItemCphyCa

Bases: SCPICmdRead

The BUS:B<x>:CPHY:CA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:CA? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:CA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .source: The BUS:B<x>:CPHY:CA:SOUrce command.
  • .threshold: The BUS:B<x>:CPHY:CA:THReshold command.

source property

Return the BUS:B<x>:CPHY:CA:SOUrce command.

Description
  • This command sets or queries the CPHY differential CA source for the specified bus line. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:CA:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:CA:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:CA:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CPHY:CA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:CA:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies an analog reference waveform as the source.

threshold property

Return the BUS:B<x>:CPHY:CA:THReshold command.

Description
  • This command sets or queries the CPHY differential CA source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:CA:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:CA:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:CA:THReshold value command.
SCPI Syntax
- BUS:B<x>:CPHY:CA:THReshold <NR3>
- BUS:B<x>:CPHY:CA:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY differential CA threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyCaSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:CA:SOUrce command.

Description
  • This command sets or queries the CPHY differential CA source for the specified bus line. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:CA:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:CA:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:CA:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CPHY:CA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:CA:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies an analog reference waveform as the source.

BusBItemCphyCaThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:CA:THReshold command.

Description
  • This command sets or queries the CPHY differential CA source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:CA:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:CA:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:CA:THReshold value command.
SCPI Syntax
- BUS:B<x>:CPHY:CA:THReshold <NR3>
- BUS:B<x>:CPHY:CA:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY differential CA threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyCgnd

Bases: SCPICmdRead

The BUS:B<x>:CPHY:CGND command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:CGND? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:CGND? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .source: The BUS:B<x>:CPHY:CGND:SOUrce command.
  • .threshold: The BUS:B<x>:CPHY:CGND:THReshold command.

source property

Return the BUS:B<x>:CPHY:CGND:SOUrce command.

Description
  • This command sets or queries the CPHY differential CGND source for the specified bus line. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:CGND:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:CGND:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:CGND:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CPHY:CGND:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:CGND:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog/digital channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies an analog/digital reference waveform as the source.

threshold property

Return the BUS:B<x>:CPHY:CGND:THReshold command.

Description
  • This command sets or queries the CPHY differential CGND source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:CGND:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:CGND:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:CGND:THReshold value command.
SCPI Syntax
- BUS:B<x>:CPHY:CGND:THReshold <NR3>
- BUS:B<x>:CPHY:CGND:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY Single Ended C GND threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyCgndSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:CGND:SOUrce command.

Description
  • This command sets or queries the CPHY differential CGND source for the specified bus line. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:CGND:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:CGND:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:CGND:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CPHY:CGND:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CPHY:CGND:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog/digital channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies an analog/digital reference waveform as the source.

BusBItemCphyCgndThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:CGND:THReshold command.

Description
  • This command sets or queries the CPHY differential CGND source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:CGND:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:CGND:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:CGND:THReshold value command.
SCPI Syntax
- BUS:B<x>:CPHY:CGND:THReshold <NR3>
- BUS:B<x>:CPHY:CGND:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the CPHY Single Ended C GND threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemCphyLp

Bases: SCPICmdRead

The BUS:B<x>:CPHY:LP command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:LP? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:LP? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .direction: The BUS:B<x>:CPHY:LP:DIRection command.

direction property

Return the BUS:B<x>:CPHY:LP:DIRection command.

Description
  • This command sets or queries the CPHY bus direction in low power. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:LP:DIRection? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:LP:DIRection? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:LP:DIRection value command.
SCPI Syntax
- BUS:B<x>:CPHY:LP:DIRection {forward|reverse}
- BUS:B<x>:CPHY:LP:DIRection?
Info
  • B<x> is the number of the bus.
  • forward specifies the direction in low power to forward. The default direction is forward.
  • reverse specifies the direction in low power to reverse.

BusBItemCphyLpDirection

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:LP:DIRection command.

Description
  • This command sets or queries the CPHY bus direction in low power. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:LP:DIRection? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:LP:DIRection? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:LP:DIRection value command.
SCPI Syntax
- BUS:B<x>:CPHY:LP:DIRection {forward|reverse}
- BUS:B<x>:CPHY:LP:DIRection?
Info
  • B<x> is the number of the bus.
  • forward specifies the direction in low power to forward. The default direction is forward.
  • reverse specifies the direction in low power to reverse.

BusBItemCphySignaltype

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:SIGNALTYpe command.

Description
  • This command sets or queries the signal type for CPHY bus decode. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:SIGNALTYpe? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:SIGNALTYpe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:SIGNALTYpe value command.
SCPI Syntax
- BUS:B<x>:CPHY:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:CPHY:SIGNALTYpe?
Info
  • B<x> is the number of the bus.
  • SINGLE specifies the signal type to single ended.
  • DIFF specifies the signal type to differential. The default type is differential.

BusBItemCphySubtype

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CPHY:SUBTYPe command.

Description
  • This command sets or queries the sub type for CPHY bus decode. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CPHY:SUBTYPe? query.
  • Using the .verify(value) method will send the BUS:B<x>:CPHY:SUBTYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CPHY:SUBTYPe value command.
SCPI Syntax
- BUS:B<x>:CPHY:SUBTYPe {CSI|DSI|Word|Symbol}
- BUS:B<x>:CPHY:SUBTYPe?
Info
  • B<x> is the number of the bus.
  • CSI specifies the protocol type to CSI. The default type is CSI.
  • DSI specifies the protocol type to DSI.
  • Word specifies the protocol type to word.
  • Symbol specifies the protocol type to symbol.

BusBItemCxpi

Bases: SCPICmdRead

The BUS:B<x>:CXPI command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CXPI? query.
  • Using the .verify(value) method will send the BUS:B<x>:CXPI? query and raise an AssertionError if the returned value does not match value.
Properties
  • .bitrate: The BUS:B<x>:CXPI:BITRate command.
  • .rec: The BUS:B<x>:CXPI:REC command tree.
  • .source: The BUS:B<x>:CXPI:SOUrce command.

bitrate property

Return the BUS:B<x>:CXPI:BITRate command.

Description
  • This command sets or queries the bit rate for the specified CXPI bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CXPI:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:CXPI:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CXPI:BITRate value command.
SCPI Syntax
- BUS:B<x>:CXPI:BITRate <NR1>
- BUS:B<x>:CXPI:BITRate?
Info
  • <NR1> sets the CXPI Source bit rate for the specified bus. The default value is 19.2 kbps and the valid range is 1 bps to 20 kbps.

rec property

Return the BUS:B<x>:CXPI:REC command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CXPI:REC? query.
  • Using the .verify(value) method will send the BUS:B<x>:CXPI:REC? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .threshold: The BUS:B<x>:CXPI:REC:THReshold command.

source property

Return the BUS:B<x>:CXPI:SOUrce command.

Description
  • This command sets or queries the source channel for the specified CXPI bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CXPI:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CXPI:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CXPI:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CXPI:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CXPI:SOUrce?
Info
  • CH<x> specifies an analog channel as the source, where is the channel number.
  • MATH<x> specifies a math channel as the source, where is the math waveform number.
  • REF<x> specifies a digital reference waveform as the source, where is the reference waveform number.

BusBItemCxpiBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CXPI:BITRate command.

Description
  • This command sets or queries the bit rate for the specified CXPI bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CXPI:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:CXPI:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CXPI:BITRate value command.
SCPI Syntax
- BUS:B<x>:CXPI:BITRate <NR1>
- BUS:B<x>:CXPI:BITRate?
Info
  • <NR1> sets the CXPI Source bit rate for the specified bus. The default value is 19.2 kbps and the valid range is 1 bps to 20 kbps.

BusBItemCxpiRec

Bases: SCPICmdRead

The BUS:B<x>:CXPI:REC command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:CXPI:REC? query.
  • Using the .verify(value) method will send the BUS:B<x>:CXPI:REC? query and raise an AssertionError if the returned value does not match value.
Properties
  • .threshold: The BUS:B<x>:CXPI:REC:THReshold command.

threshold property

Return the BUS:B<x>:CXPI:REC:THReshold command.

Description
  • This command sets or queries the source channel recessive threshold for the specified CXPI bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CXPI:REC:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CXPI:REC:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CXPI:REC:THReshold value command.
SCPI Syntax
- BUS:B<x>:CXPI:REC:THReshold <NR3>
- BUS:B<x>:CXPI:REC:THReshold?
Info
  • <NR3> sets the CXPI Source recessive threshold for the specified bus. The default value is 0 V and the valid range is -12 V to +12 V.

BusBItemCxpiRecThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CXPI:REC:THReshold command.

Description
  • This command sets or queries the source channel recessive threshold for the specified CXPI bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CXPI:REC:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:CXPI:REC:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CXPI:REC:THReshold value command.
SCPI Syntax
- BUS:B<x>:CXPI:REC:THReshold <NR3>
- BUS:B<x>:CXPI:REC:THReshold?
Info
  • <NR3> sets the CXPI Source recessive threshold for the specified bus. The default value is 0 V and the valid range is -12 V to +12 V.

BusBItemCxpiSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:CXPI:SOUrce command.

Description
  • This command sets or queries the source channel for the specified CXPI bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:CXPI:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:CXPI:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:CXPI:SOUrce value command.
SCPI Syntax
- BUS:B<x>:CXPI:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:CXPI:SOUrce?
Info
  • CH<x> specifies an analog channel as the source, where is the channel number.
  • MATH<x> specifies a math channel as the source, where is the math waveform number.
  • REF<x> specifies a digital reference waveform as the source, where is the reference waveform number.

BusBItemDisplay

Bases: SCPICmdRead

The BUS:B<x>:DISplay command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:DISplay? query.
  • Using the .verify(value) method will send the BUS:B<x>:DISplay? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .format: The BUS:B<x>:DISplay:FORMat command.
  • .layout: The BUS:B<x>:DISplay:LAYout command.

format property

Return the BUS:B<x>:DISplay:FORMat command.

Description
  • This command sets or queries how the data is represented in the bus form for the specified bus. The bus is specified by x. Note: Different bus types support only a subset of these arguments.
Usage
  • Using the .query() method will send the BUS:B<x>:DISplay:FORMat? query.
  • Using the .verify(value) method will send the BUS:B<x>:DISplay:FORMat? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DISplay:FORMat value command.
SCPI Syntax
- BUS:B<x>:DISplay:FORMat {HEX|BINARY|MIXEDASCII|MIXEDHEX|ASCII|DECIMAL|MIXED}
- BUS:B<x>:DISplay:FORMat?
Info
  • B<x> is the number of the bus.
  • HEX specifies hexadecimal format.
  • BINARY specifies binary format.
  • MIXEDASCII specifies mixed ASCII format.
  • MIXEDHEX specifies mixed hexadecimal format.
  • ASCII specifies ASCII format.
  • DECIMAL specifies decimal format.
  • MIXED specifies mixed format.

layout property

Return the BUS:B<x>:DISplay:LAYout command.

Description
  • This command sets or queries what to display for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DISplay:LAYout? query.
  • Using the .verify(value) method will send the BUS:B<x>:DISplay:LAYout? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DISplay:LAYout value command.
SCPI Syntax
- BUS:B<x>:DISplay:LAYout {BUS|BUSANDWAVEFORM}
- BUS:B<x>:DISplay:LAYout?
Info
  • B<x> is the number of the bus.
  • BUS specifies displaying the bus form only.
  • BUSANDWAVEFORM specifies displaying the bus form and the constituent source waveform(s). This argument is not available for some bus types and some bus configurations.

BusBItemDisplayFormat

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:DISplay:FORMat command.

Description
  • This command sets or queries how the data is represented in the bus form for the specified bus. The bus is specified by x. Note: Different bus types support only a subset of these arguments.
Usage
  • Using the .query() method will send the BUS:B<x>:DISplay:FORMat? query.
  • Using the .verify(value) method will send the BUS:B<x>:DISplay:FORMat? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DISplay:FORMat value command.
SCPI Syntax
- BUS:B<x>:DISplay:FORMat {HEX|BINARY|MIXEDASCII|MIXEDHEX|ASCII|DECIMAL|MIXED}
- BUS:B<x>:DISplay:FORMat?
Info
  • B<x> is the number of the bus.
  • HEX specifies hexadecimal format.
  • BINARY specifies binary format.
  • MIXEDASCII specifies mixed ASCII format.
  • MIXEDHEX specifies mixed hexadecimal format.
  • ASCII specifies ASCII format.
  • DECIMAL specifies decimal format.
  • MIXED specifies mixed format.

BusBItemDisplayLayout

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:DISplay:LAYout command.

Description
  • This command sets or queries what to display for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DISplay:LAYout? query.
  • Using the .verify(value) method will send the BUS:B<x>:DISplay:LAYout? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DISplay:LAYout value command.
SCPI Syntax
- BUS:B<x>:DISplay:LAYout {BUS|BUSANDWAVEFORM}
- BUS:B<x>:DISplay:LAYout?
Info
  • B<x> is the number of the bus.
  • BUS specifies displaying the bus form only.
  • BUSANDWAVEFORM specifies displaying the bus form and the constituent source waveform(s). This argument is not available for some bus types and some bus configurations.

BusBItemDphy

Bases: SCPICmdRead

The BUS:B<x>:DPHY command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:DPHY? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Properties
  • .clock: The BUS:B<x>:DPHY:CLOCk command tree.
  • .dminus: The BUS:B<x>:DPHY:DMINus command tree.
  • .dplus: The BUS:B<x>:DPHY:DPlus command tree.
  • .lp: The BUS:B<x>:DPHY:LP command tree.
  • .protocol: The BUS:B<x>:DPHY:PROTocol command tree.
  • .signal: The BUS:B<x>:DPHY:SIGNal command tree.

clock property

Return the BUS:B<x>:DPHY:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Sub-properties
  • .source: The BUS:B<x>:DPHY:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:DPHY:CLOCk:THRESHold command.

dminus property

Return the BUS:B<x>:DPHY:DMINus command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:DMINus? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:DMINus? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .datathreshold: The BUS:B<x>:DPHY:DMINus:DATATHRESHold command.
  • .lpthreshold: The BUS:B<x>:DPHY:DMINus:LPTHRESHold command.
  • .source: The BUS:B<x>:DPHY:DMINus:SOUrce command.

dplus property

Return the BUS:B<x>:DPHY:DPlus command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:DPlus? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:DPlus? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .datathreshold: The BUS:B<x>:DPHY:DPlus:DATATHRESHold command.
  • .lpthreshold: The BUS:B<x>:DPHY:DPlus:LPTHRESHold command.
  • .source: The BUS:B<x>:DPHY:DPlus:SOUrce command.

lp property

Return the BUS:B<x>:DPHY:LP command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:LP? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:LP? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Sub-properties
  • .direction: The BUS:B<x>:DPHY:LP:DIRection command.

protocol property

Return the BUS:B<x>:DPHY:PROTocol command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:PROTocol? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:PROTocol? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Sub-properties
  • .type: The BUS:B<x>:DPHY:PROTocol:TYPe command.

signal property

Return the BUS:B<x>:DPHY:SIGNal command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:SIGNal? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:SIGNal? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Sub-properties
  • .encoding: The BUS:B<x>:DPHY:SIGNal:ENCoding command.

BusBItemDphyClock

Bases: SCPICmdRead

The BUS:B<x>:DPHY:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Properties
  • .source: The BUS:B<x>:DPHY:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:DPHY:CLOCk:THRESHold command.

source property

Return the BUS:B<x>:DPHY:CLOCk:SOUrce command.

Description
  • This command sets or queries the DPHY Clock source for the specified bus line. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:DPHY:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:DPHY:CLOCk:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source, where is the channel number.
  • MATH<x> specifies a math channel as the source, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source, where is the reference waveform number.

threshold property

Return the BUS:B<x>:DPHY:CLOCk:THRESHold command.

Description
  • This command sets or queries the DPHY Clock source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:CLOCk:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:CLOCk:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:CLOCk:THRESHold value command.
SCPI Syntax
- BUS:B<x>:DPHY:CLOCk:THRESHold <NR3>
- BUS:B<x>:DPHY:CLOCk:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the DPHY clock high threshold for the specified bus. The argument range is -8V to +8V.

BusBItemDphyClockSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:DPHY:CLOCk:SOUrce command.

Description
  • This command sets or queries the DPHY Clock source for the specified bus line. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:DPHY:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:DPHY:CLOCk:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source, where is the channel number.
  • MATH<x> specifies a math channel as the source, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source, where is the reference waveform number.

BusBItemDphyClockThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:DPHY:CLOCk:THRESHold command.

Description
  • This command sets or queries the DPHY Clock source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:CLOCk:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:CLOCk:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:CLOCk:THRESHold value command.
SCPI Syntax
- BUS:B<x>:DPHY:CLOCk:THRESHold <NR3>
- BUS:B<x>:DPHY:CLOCk:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the DPHY clock high threshold for the specified bus. The argument range is -8V to +8V.

BusBItemDphyDminus

Bases: SCPICmdRead

The BUS:B<x>:DPHY:DMINus command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:DMINus? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:DMINus? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .datathreshold: The BUS:B<x>:DPHY:DMINus:DATATHRESHold command.
  • .lpthreshold: The BUS:B<x>:DPHY:DMINus:LPTHRESHold command.
  • .source: The BUS:B<x>:DPHY:DMINus:SOUrce command.

datathreshold property

Return the BUS:B<x>:DPHY:DMINus:DATATHRESHold command.

Description
  • This command sets or queries the DPHY D- source data threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:DMINus:DATATHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:DMINus:DATATHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:DMINus:DATATHRESHold value command.
SCPI Syntax
- BUS:B<x>:DPHY:DMINus:DATATHRESHold <NR3>
- BUS:B<x>:DPHY:DMINus:DATATHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the DPHY D- source data threshold for the specified bus. The argument range is -8V to +8V.

lpthreshold property

Return the BUS:B<x>:DPHY:DMINus:LPTHRESHold command.

Description
  • This command sets or queries the DPHY D- source low power threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:DMINus:LPTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:DMINus:LPTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:DMINus:LPTHRESHold value command.
SCPI Syntax
- BUS:B<x>:DPHY:DMINus:LPTHRESHold <NR3>
- BUS:B<x>:DPHY:DMINus:LPTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the DPHY D- source low power threshold for the specified bus. The argument range is -8V to +8V.

source property

Return the BUS:B<x>:DPHY:DMINus:SOUrce command.

Description
  • This command sets or queries the DPHY D- source for the specified bus line. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:DMINus:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:DMINus:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:DMINus:SOUrce value command.
SCPI Syntax
- BUS:B<x>:DPHY:DMINus:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:DPHY:DMINus:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source, where is the channel number.
  • MATH<x> specifies a math channel as the source, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source, where is the reference waveform number.

BusBItemDphyDminusDatathreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:DPHY:DMINus:DATATHRESHold command.

Description
  • This command sets or queries the DPHY D- source data threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:DMINus:DATATHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:DMINus:DATATHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:DMINus:DATATHRESHold value command.
SCPI Syntax
- BUS:B<x>:DPHY:DMINus:DATATHRESHold <NR3>
- BUS:B<x>:DPHY:DMINus:DATATHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the DPHY D- source data threshold for the specified bus. The argument range is -8V to +8V.

BusBItemDphyDminusLpthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:DPHY:DMINus:LPTHRESHold command.

Description
  • This command sets or queries the DPHY D- source low power threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:DMINus:LPTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:DMINus:LPTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:DMINus:LPTHRESHold value command.
SCPI Syntax
- BUS:B<x>:DPHY:DMINus:LPTHRESHold <NR3>
- BUS:B<x>:DPHY:DMINus:LPTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the DPHY D- source low power threshold for the specified bus. The argument range is -8V to +8V.

BusBItemDphyDminusSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:DPHY:DMINus:SOUrce command.

Description
  • This command sets or queries the DPHY D- source for the specified bus line. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:DMINus:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:DMINus:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:DMINus:SOUrce value command.
SCPI Syntax
- BUS:B<x>:DPHY:DMINus:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:DPHY:DMINus:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source, where is the channel number.
  • MATH<x> specifies a math channel as the source, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source, where is the reference waveform number.

BusBItemDphyDplus

Bases: SCPICmdRead

The BUS:B<x>:DPHY:DPlus command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:DPlus? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:DPlus? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .datathreshold: The BUS:B<x>:DPHY:DPlus:DATATHRESHold command.
  • .lpthreshold: The BUS:B<x>:DPHY:DPlus:LPTHRESHold command.
  • .source: The BUS:B<x>:DPHY:DPlus:SOUrce command.

datathreshold property

Return the BUS:B<x>:DPHY:DPlus:DATATHRESHold command.

Description
  • This command sets or queries the D+ source data threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:DPlus:DATATHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:DPlus:DATATHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:DPlus:DATATHRESHold value command.
SCPI Syntax
- BUS:B<x>:DPHY:DPlus:DATATHRESHold <NR3>
- BUS:B<x>:DPHY:DPlus:DATATHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the DPHY D+ source data threshold for the specified bus. The argument range is -8V to +8V.

lpthreshold property

Return the BUS:B<x>:DPHY:DPlus:LPTHRESHold command.

Description
  • This command sets or queries the DPHY D+ source low power threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:DPlus:LPTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:DPlus:LPTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:DPlus:LPTHRESHold value command.
SCPI Syntax
- BUS:B<x>:DPHY:DPlus:LPTHRESHold <NR3>
- BUS:B<x>:DPHY:DPlus:LPTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the DPHY D+ source low power threshold for the specified bus. The argument range is -8V to +8V.

source property

Return the BUS:B<x>:DPHY:DPlus:SOUrce command.

Description
  • This command sets or queries the DPHY D+ source for the specified bus line. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:DPlus:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:DPlus:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:DPlus:SOUrce value command.
SCPI Syntax
- BUS:B<x>:DPHY:DPlus:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:DPHY:DPlus:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source, where is the channel number.
  • MATH<x> specifies a math channel as the source, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source, where is the reference waveform number.

BusBItemDphyDplusDatathreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:DPHY:DPlus:DATATHRESHold command.

Description
  • This command sets or queries the D+ source data threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:DPlus:DATATHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:DPlus:DATATHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:DPlus:DATATHRESHold value command.
SCPI Syntax
- BUS:B<x>:DPHY:DPlus:DATATHRESHold <NR3>
- BUS:B<x>:DPHY:DPlus:DATATHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the DPHY D+ source data threshold for the specified bus. The argument range is -8V to +8V.

BusBItemDphyDplusLpthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:DPHY:DPlus:LPTHRESHold command.

Description
  • This command sets or queries the DPHY D+ source low power threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:DPlus:LPTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:DPlus:LPTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:DPlus:LPTHRESHold value command.
SCPI Syntax
- BUS:B<x>:DPHY:DPlus:LPTHRESHold <NR3>
- BUS:B<x>:DPHY:DPlus:LPTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the DPHY D+ source low power threshold for the specified bus. The argument range is -8V to +8V.

BusBItemDphyDplusSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:DPHY:DPlus:SOUrce command.

Description
  • This command sets or queries the DPHY D+ source for the specified bus line. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:DPlus:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:DPlus:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:DPlus:SOUrce value command.
SCPI Syntax
- BUS:B<x>:DPHY:DPlus:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:DPHY:DPlus:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source, where is the channel number.
  • MATH<x> specifies a math channel as the source, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source, where is the reference waveform number.

BusBItemDphyLp

Bases: SCPICmdRead

The BUS:B<x>:DPHY:LP command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:LP? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:LP? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Properties
  • .direction: The BUS:B<x>:DPHY:LP:DIRection command.

direction property

Return the BUS:B<x>:DPHY:LP:DIRection command.

Description
  • This command sets or queries the DPHY bus direction in low power. By default lp direction is set to forward. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:LP:DIRection? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:LP:DIRection? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:LP:DIRection value command.
SCPI Syntax
- BUS:B<x>:DPHY:LP:DIRection {forward|reverse}
- BUS:B<x>:DPHY:LP:DIRection?
Info
  • B<x> is the Bus number.
  • forward specifies the direction as forward.
  • reverse specifies the direction as reverse.

BusBItemDphyLpDirection

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:DPHY:LP:DIRection command.

Description
  • This command sets or queries the DPHY bus direction in low power. By default lp direction is set to forward. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:LP:DIRection? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:LP:DIRection? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:LP:DIRection value command.
SCPI Syntax
- BUS:B<x>:DPHY:LP:DIRection {forward|reverse}
- BUS:B<x>:DPHY:LP:DIRection?
Info
  • B<x> is the Bus number.
  • forward specifies the direction as forward.
  • reverse specifies the direction as reverse.

BusBItemDphyProtocol

Bases: SCPICmdRead

The BUS:B<x>:DPHY:PROTocol command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:PROTocol? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:PROTocol? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Properties
  • .type: The BUS:B<x>:DPHY:PROTocol:TYPe command.

type property

Return the BUS:B<x>:DPHY:PROTocol:TYPe command.

Description
  • This command sets or queries the protocol type for DPHY bus decode. The default type is CSI. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:PROTocol:TYPe? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:PROTocol:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:PROTocol:TYPe value command.
SCPI Syntax
- BUS:B<x>:DPHY:PROTocol:TYPe {CSI|DSI}
- BUS:B<x>:DPHY:PROTocol:TYPe?
Info
  • B<x> is the Bus number.
  • CSI specifies the protocol type as CSI.
  • DSI specifies the protocol type as DSI.

BusBItemDphyProtocolType

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:DPHY:PROTocol:TYPe command.

Description
  • This command sets or queries the protocol type for DPHY bus decode. The default type is CSI. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:PROTocol:TYPe? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:PROTocol:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:PROTocol:TYPe value command.
SCPI Syntax
- BUS:B<x>:DPHY:PROTocol:TYPe {CSI|DSI}
- BUS:B<x>:DPHY:PROTocol:TYPe?
Info
  • B<x> is the Bus number.
  • CSI specifies the protocol type as CSI.
  • DSI specifies the protocol type as DSI.

BusBItemDphySignal

Bases: SCPICmdRead

The BUS:B<x>:DPHY:SIGNal command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:SIGNal? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:SIGNal? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Properties
  • .encoding: The BUS:B<x>:DPHY:SIGNal:ENCoding command.

encoding property

Return the BUS:B<x>:DPHY:SIGNal:ENCoding command.

Description
  • This command sets or queries the 8b9b encoding for DPHY bus decode. By default 8b9b encoding is set to false. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:SIGNal:ENCoding? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:SIGNal:ENCoding? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:SIGNal:ENCoding value command.
SCPI Syntax
- BUS:B<x>:DPHY:SIGNal:ENCoding {false|true}
- BUS:B<x>:DPHY:SIGNal:ENCoding?
Info
  • B<x> is the Bus number.
  • false specifies the 8b9b encoding disabled.
  • true specifies 8b9b encoding disabled.

BusBItemDphySignalEncoding

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:DPHY:SIGNal:ENCoding command.

Description
  • This command sets or queries the 8b9b encoding for DPHY bus decode. By default 8b9b encoding is set to false. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:DPHY:SIGNal:ENCoding? query.
  • Using the .verify(value) method will send the BUS:B<x>:DPHY:SIGNal:ENCoding? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:DPHY:SIGNal:ENCoding value command.
SCPI Syntax
- BUS:B<x>:DPHY:SIGNal:ENCoding {false|true}
- BUS:B<x>:DPHY:SIGNal:ENCoding?
Info
  • B<x> is the Bus number.
  • false specifies the 8b9b encoding disabled.
  • true specifies 8b9b encoding disabled.

BusBItemEspi

Bases: SCPICmdRead

The BUS:B<x>:ESPI command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ESPI? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Properties
  • .alertview: The BUS:B<x>:ESPI:ALERTVIEW command.
  • .alert: The BUS:B<x>:ESPI:ALERt command tree.
  • .chipselect: The BUS:B<x>:ESPI:CHIPSELect command tree.
  • .clock: The BUS:B<x>:ESPI:CLOCk command tree.
  • .dataone: The BUS:B<x>:ESPI:DATAONE command tree.
  • .datatwo: The BUS:B<x>:ESPI:DATATWO command tree.
  • .iomode: The BUS:B<x>:ESPI:IOMODe command.

alert property

Return the BUS:B<x>:ESPI:ALERt command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:ALERt? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:ALERt? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Sub-properties
  • .polarity: The BUS:B<x>:ESPI:ALERt:POLarity command.
  • .source: The BUS:B<x>:ESPI:ALERt:SOUrce command.
  • .threshold: The BUS:B<x>:ESPI:ALERt:THReshold command.

alertview property

Return the BUS:B<x>:ESPI:ALERTVIEW command.

Description
  • This command sets or queries the ESPI alert view for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:ALERTVIEW? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:ALERTVIEW? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:ALERTVIEW value command.
SCPI Syntax
- BUS:B<x>:ESPI:ALERTVIEW {ON|OFF}
- BUS:B<x>:ESPI:ALERTVIEW?
Info
  • B<x> is the Bus number.
  • ON turns the alert source on.
  • OFF turns the alert source on.

chipselect property

Return the BUS:B<x>:ESPI:CHIPSELect command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:CHIPSELect? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:CHIPSELect? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Sub-properties
  • .polarity: The BUS:B<x>:ESPI:CHIPSELect:POLarity command.
  • .source: The BUS:B<x>:ESPI:CHIPSELect:SOUrce command.
  • .threshold: The BUS:B<x>:ESPI:CHIPSELect:THReshold command.

clock property

Return the BUS:B<x>:ESPI:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Sub-properties
  • .polarity: The BUS:B<x>:ESPI:CLOCk:POLarity command.
  • .source: The BUS:B<x>:ESPI:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:ESPI:CLOCk:THReshold command.

dataone property

Return the BUS:B<x>:ESPI:DATAONE command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:DATAONE? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:DATAONE? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Sub-properties
  • .polarity: The BUS:B<x>:ESPI:DATAONE:POLarity command.
  • .source: The BUS:B<x>:ESPI:DATAONE:SOUrce command.
  • .threshold: The BUS:B<x>:ESPI:DATAONE:THReshold command.

datatwo property

Return the BUS:B<x>:ESPI:DATATWO command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:DATATWO? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:DATATWO? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Sub-properties
  • .polarity: The BUS:B<x>:ESPI:DATATWO:POLarity command.
  • .source: The BUS:B<x>:ESPI:DATATWO:SOUrce command.
  • .threshold: The BUS:B<x>:ESPI:DATATWO:THReshold command.

iomode property

Return the BUS:B<x>:ESPI:IOMODe command.

Description
  • This command sets or queries the ESPI Input/Output mode for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:IOMODe? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:IOMODe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:IOMODe value command.
SCPI Syntax
- BUS:B<x>:ESPI:IOMODe {SINGle|DUAL}
- BUS:B<x>:ESPI:IOMODe?
Info
  • B<x> is the Bus number.
  • SINGle displays the command and response decode in two lanes.
  • DUAL displays the decode in a single data lane.

BusBItemEspiAlert

Bases: SCPICmdRead

The BUS:B<x>:ESPI:ALERt command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:ALERt? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:ALERt? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Properties
  • .polarity: The BUS:B<x>:ESPI:ALERt:POLarity command.
  • .source: The BUS:B<x>:ESPI:ALERt:SOUrce command.
  • .threshold: The BUS:B<x>:ESPI:ALERt:THReshold command.

polarity property

Return the BUS:B<x>:ESPI:ALERt:POLarity command.

Description
  • This command sets or queries the ESPI alert polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:ALERt:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:ALERt:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:ALERt:POLarity value command.
SCPI Syntax
- BUS:B<x>:ESPI:ALERt:POLarity {HIGH|LOW}
- BUS:B<x>:ESPI:ALERt:POLarity?
Info
  • B<x> is the Bus number.
  • HIGH sets the ESPI alert polarity to active high.
  • LOW sets the ESPI alert polarity to active low.

source property

Return the BUS:B<x>:ESPI:ALERt:SOUrce command.

Description
  • This command sets or queries the alert source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:ALERt:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:ALERt:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:ALERt:SOUrce value command.
SCPI Syntax
- BUS:B<x>:ESPI:ALERt:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ESPI:ALERt:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies a reference waveform as the source.

threshold property

Return the BUS:B<x>:ESPI:ALERt:THReshold command.

Description
  • This command sets or queries the alert source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:ALERt:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:ALERt:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:ALERt:THReshold value command.
SCPI Syntax
- BUS:B<x>:ESPI:ALERt:THReshold <NR3>
- BUS:B<x>:ESPI:ALERt:THReshold?
Info
  • B<x> is the Bus number.
  • <NR3> sets the alert threshold for the specified bus. The valid range is -8V to +8V.

BusBItemEspiAlertPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ESPI:ALERt:POLarity command.

Description
  • This command sets or queries the ESPI alert polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:ALERt:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:ALERt:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:ALERt:POLarity value command.
SCPI Syntax
- BUS:B<x>:ESPI:ALERt:POLarity {HIGH|LOW}
- BUS:B<x>:ESPI:ALERt:POLarity?
Info
  • B<x> is the Bus number.
  • HIGH sets the ESPI alert polarity to active high.
  • LOW sets the ESPI alert polarity to active low.

BusBItemEspiAlertSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ESPI:ALERt:SOUrce command.

Description
  • This command sets or queries the alert source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:ALERt:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:ALERt:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:ALERt:SOUrce value command.
SCPI Syntax
- BUS:B<x>:ESPI:ALERt:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ESPI:ALERt:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies a reference waveform as the source.

BusBItemEspiAlertThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ESPI:ALERt:THReshold command.

Description
  • This command sets or queries the alert source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:ALERt:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:ALERt:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:ALERt:THReshold value command.
SCPI Syntax
- BUS:B<x>:ESPI:ALERt:THReshold <NR3>
- BUS:B<x>:ESPI:ALERt:THReshold?
Info
  • B<x> is the Bus number.
  • <NR3> sets the alert threshold for the specified bus. The valid range is -8V to +8V.

BusBItemEspiAlertview

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ESPI:ALERTVIEW command.

Description
  • This command sets or queries the ESPI alert view for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:ALERTVIEW? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:ALERTVIEW? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:ALERTVIEW value command.
SCPI Syntax
- BUS:B<x>:ESPI:ALERTVIEW {ON|OFF}
- BUS:B<x>:ESPI:ALERTVIEW?
Info
  • B<x> is the Bus number.
  • ON turns the alert source on.
  • OFF turns the alert source on.

BusBItemEspiChipselect

Bases: SCPICmdRead

The BUS:B<x>:ESPI:CHIPSELect command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:CHIPSELect? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:CHIPSELect? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Properties
  • .polarity: The BUS:B<x>:ESPI:CHIPSELect:POLarity command.
  • .source: The BUS:B<x>:ESPI:CHIPSELect:SOUrce command.
  • .threshold: The BUS:B<x>:ESPI:CHIPSELect:THReshold command.

polarity property

Return the BUS:B<x>:ESPI:CHIPSELect:POLarity command.

Description
  • This command sets or queries the ESPI chip select polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:CHIPSELect:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:CHIPSELect:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:CHIPSELect:POLarity value command.
SCPI Syntax
- BUS:B<x>:ESPI:CHIPSELect:POLarity {HIGH|LOW}
- BUS:B<x>:ESPI:CHIPSELect:POLarity?
Info
  • B<x> is the Bus number.
  • HIGH sets the ESPI chip select polarity to active high.
  • LOW sets the ESPI chip select polarity to active low.

source property

Return the BUS:B<x>:ESPI:CHIPSELect:SOUrce command.

Description
  • This command sets or queries the chip select source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:CHIPSELect:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:CHIPSELect:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:CHIPSELect:SOUrce value command.
SCPI Syntax
- BUS:B<x>:ESPI:CHIPSELect:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ESPI:CHIPSELect:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies a reference waveform as the source.

threshold property

Return the BUS:B<x>:ESPI:CHIPSELect:THReshold command.

Description
  • This command sets or queries the chip select source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:CHIPSELect:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:CHIPSELect:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:CHIPSELect:THReshold value command.
SCPI Syntax
- BUS:B<x>:ESPI:CHIPSELect:THReshold <NR3>
- BUS:B<x>:ESPI:CHIPSELect:THReshold?
Info
  • B<x> is the Bus number.
  • <NR3> sets the chip select threshold for the specified bus. The valid range is -8V to +8V.

BusBItemEspiChipselectPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ESPI:CHIPSELect:POLarity command.

Description
  • This command sets or queries the ESPI chip select polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:CHIPSELect:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:CHIPSELect:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:CHIPSELect:POLarity value command.
SCPI Syntax
- BUS:B<x>:ESPI:CHIPSELect:POLarity {HIGH|LOW}
- BUS:B<x>:ESPI:CHIPSELect:POLarity?
Info
  • B<x> is the Bus number.
  • HIGH sets the ESPI chip select polarity to active high.
  • LOW sets the ESPI chip select polarity to active low.

BusBItemEspiChipselectSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ESPI:CHIPSELect:SOUrce command.

Description
  • This command sets or queries the chip select source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:CHIPSELect:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:CHIPSELect:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:CHIPSELect:SOUrce value command.
SCPI Syntax
- BUS:B<x>:ESPI:CHIPSELect:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ESPI:CHIPSELect:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies a reference waveform as the source.

BusBItemEspiChipselectThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ESPI:CHIPSELect:THReshold command.

Description
  • This command sets or queries the chip select source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:CHIPSELect:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:CHIPSELect:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:CHIPSELect:THReshold value command.
SCPI Syntax
- BUS:B<x>:ESPI:CHIPSELect:THReshold <NR3>
- BUS:B<x>:ESPI:CHIPSELect:THReshold?
Info
  • B<x> is the Bus number.
  • <NR3> sets the chip select threshold for the specified bus. The valid range is -8V to +8V.

BusBItemEspiClock

Bases: SCPICmdRead

The BUS:B<x>:ESPI:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Properties
  • .polarity: The BUS:B<x>:ESPI:CLOCk:POLarity command.
  • .source: The BUS:B<x>:ESPI:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:ESPI:CLOCk:THReshold command.

polarity property

Return the BUS:B<x>:ESPI:CLOCk:POLarity command.

Description
  • This command sets or queries the ESPI Clock (SCLK) source polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:CLOCk:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:CLOCk:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:CLOCk:POLarity value command.
SCPI Syntax
- BUS:B<x>:ESPI:CLOCk:POLarity {FALL|RISE}
- BUS:B<x>:ESPI:CLOCk:POLarity?
Info
  • B<x> is the Bus number.
  • FALL sets the ESPI clock polarity to fall.
  • RISE sets the ESPI clock polarity to rise.

source property

Return the BUS:B<x>:ESPI:CLOCk:SOUrce command.

Description
  • This command sets or queries the Clock source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:ESPI:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ESPI:CLOCk:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies a reference waveform as the source.

threshold property

Return the BUS:B<x>:ESPI:CLOCk:THReshold command.

Description
  • This command sets or queries the Clock source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:CLOCk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:CLOCk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:CLOCk:THReshold value command.
SCPI Syntax
- BUS:B<x>:ESPI:CLOCk:THReshold <NR3>
- BUS:B<x>:ESPI:CLOCk:THReshold?
Info
  • B<x> is the Bus number.
  • <NR3> sets the clock threshold for the specified bus. The valid range is -8V to +8V.

BusBItemEspiClockPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ESPI:CLOCk:POLarity command.

Description
  • This command sets or queries the ESPI Clock (SCLK) source polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:CLOCk:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:CLOCk:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:CLOCk:POLarity value command.
SCPI Syntax
- BUS:B<x>:ESPI:CLOCk:POLarity {FALL|RISE}
- BUS:B<x>:ESPI:CLOCk:POLarity?
Info
  • B<x> is the Bus number.
  • FALL sets the ESPI clock polarity to fall.
  • RISE sets the ESPI clock polarity to rise.

BusBItemEspiClockSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ESPI:CLOCk:SOUrce command.

Description
  • This command sets or queries the Clock source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:ESPI:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ESPI:CLOCk:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies a reference waveform as the source.

BusBItemEspiClockThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ESPI:CLOCk:THReshold command.

Description
  • This command sets or queries the Clock source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:CLOCk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:CLOCk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:CLOCk:THReshold value command.
SCPI Syntax
- BUS:B<x>:ESPI:CLOCk:THReshold <NR3>
- BUS:B<x>:ESPI:CLOCk:THReshold?
Info
  • B<x> is the Bus number.
  • <NR3> sets the clock threshold for the specified bus. The valid range is -8V to +8V.

BusBItemEspiDataone

Bases: SCPICmdRead

The BUS:B<x>:ESPI:DATAONE command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:DATAONE? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:DATAONE? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Properties
  • .polarity: The BUS:B<x>:ESPI:DATAONE:POLarity command.
  • .source: The BUS:B<x>:ESPI:DATAONE:SOUrce command.
  • .threshold: The BUS:B<x>:ESPI:DATAONE:THReshold command.

polarity property

Return the BUS:B<x>:ESPI:DATAONE:POLarity command.

Description
  • This command sets or queries the ESPI command (single mode)/ IO[0] (dual mode) polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:DATAONE:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:DATAONE:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:DATAONE:POLarity value command.
SCPI Syntax
- BUS:B<x>:ESPI:DATAONE:POLarity {HIGH|LOW}
- BUS:B<x>:ESPI:DATAONE:POLarity?
Info
  • B<x> is the Bus number.
  • HIGH sets the ESPI data polarity to active high.
  • LOW sets the ESPI data polarity to active low.

source property

Return the BUS:B<x>:ESPI:DATAONE:SOUrce command.

Description
  • This command sets or queries the command (single mode)/ IO[0] (dual mode) Data source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:DATAONE:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:DATAONE:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:DATAONE:SOUrce value command.
SCPI Syntax
- BUS:B<x>:ESPI:DATAONE:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ESPI:DATAONE:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies a reference waveform as the source.

threshold property

Return the BUS:B<x>:ESPI:DATAONE:THReshold command.

Description
  • This command sets or queries the command (single mode)/ IO[0] (dual mode) Data source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:DATAONE:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:DATAONE:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:DATAONE:THReshold value command.
SCPI Syntax
- BUS:B<x>:ESPI:DATAONE:THReshold <NR3>
- BUS:B<x>:ESPI:DATAONE:THReshold?
Info
  • B<x> is the Bus number.
  • <NR3> sets the command/data threshold for the specified bus. The valid range is -8V to +8V.

BusBItemEspiDataonePolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ESPI:DATAONE:POLarity command.

Description
  • This command sets or queries the ESPI command (single mode)/ IO[0] (dual mode) polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:DATAONE:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:DATAONE:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:DATAONE:POLarity value command.
SCPI Syntax
- BUS:B<x>:ESPI:DATAONE:POLarity {HIGH|LOW}
- BUS:B<x>:ESPI:DATAONE:POLarity?
Info
  • B<x> is the Bus number.
  • HIGH sets the ESPI data polarity to active high.
  • LOW sets the ESPI data polarity to active low.

BusBItemEspiDataoneSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ESPI:DATAONE:SOUrce command.

Description
  • This command sets or queries the command (single mode)/ IO[0] (dual mode) Data source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:DATAONE:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:DATAONE:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:DATAONE:SOUrce value command.
SCPI Syntax
- BUS:B<x>:ESPI:DATAONE:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ESPI:DATAONE:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies a reference waveform as the source.

BusBItemEspiDataoneThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ESPI:DATAONE:THReshold command.

Description
  • This command sets or queries the command (single mode)/ IO[0] (dual mode) Data source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:DATAONE:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:DATAONE:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:DATAONE:THReshold value command.
SCPI Syntax
- BUS:B<x>:ESPI:DATAONE:THReshold <NR3>
- BUS:B<x>:ESPI:DATAONE:THReshold?
Info
  • B<x> is the Bus number.
  • <NR3> sets the command/data threshold for the specified bus. The valid range is -8V to +8V.

BusBItemEspiDatatwo

Bases: SCPICmdRead

The BUS:B<x>:ESPI:DATATWO command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:DATATWO? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:DATATWO? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Properties
  • .polarity: The BUS:B<x>:ESPI:DATATWO:POLarity command.
  • .source: The BUS:B<x>:ESPI:DATATWO:SOUrce command.
  • .threshold: The BUS:B<x>:ESPI:DATATWO:THReshold command.

polarity property

Return the BUS:B<x>:ESPI:DATATWO:POLarity command.

Description
  • This command sets or queries the ESPI response (single mode)/ IO[1] (dual mode) polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:DATATWO:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:DATATWO:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:DATATWO:POLarity value command.
SCPI Syntax
- BUS:B<x>:ESPI:DATATWO:POLarity {HIGH|LOW}
- BUS:B<x>:ESPI:DATATWO:POLarity?
Info
  • B<x> is the Bus number.
  • HIGH sets the ESPI data polarity to active high.
  • LOW sets the ESPI data polarity to active low.

source property

Return the BUS:B<x>:ESPI:DATATWO:SOUrce command.

Description
  • This command sets or queries the response (single mode)/ IO[1] (dual mode) Data source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:DATATWO:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:DATATWO:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:DATATWO:SOUrce value command.
SCPI Syntax
- BUS:B<x>:ESPI:DATATWO:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ESPI:DATATWO:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies a reference waveform as the source.

threshold property

Return the BUS:B<x>:ESPI:DATATWO:THReshold command.

Description
  • This command sets or queries the response (single mode)/ IO[1] (dual mode) Data source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:DATATWO:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:DATATWO:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:DATATWO:THReshold value command.
SCPI Syntax
- BUS:B<x>:ESPI:DATATWO:THReshold <NR3>
- BUS:B<x>:ESPI:DATATWO:THReshold?
Info
  • B<x> is the Bus number.
  • <NR3> sets the command/data threshold for the specified bus. The valid range is -8V to +8V.

BusBItemEspiDatatwoPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ESPI:DATATWO:POLarity command.

Description
  • This command sets or queries the ESPI response (single mode)/ IO[1] (dual mode) polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:DATATWO:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:DATATWO:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:DATATWO:POLarity value command.
SCPI Syntax
- BUS:B<x>:ESPI:DATATWO:POLarity {HIGH|LOW}
- BUS:B<x>:ESPI:DATATWO:POLarity?
Info
  • B<x> is the Bus number.
  • HIGH sets the ESPI data polarity to active high.
  • LOW sets the ESPI data polarity to active low.

BusBItemEspiDatatwoSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ESPI:DATATWO:SOUrce command.

Description
  • This command sets or queries the response (single mode)/ IO[1] (dual mode) Data source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:DATATWO:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:DATATWO:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:DATATWO:SOUrce value command.
SCPI Syntax
- BUS:B<x>:ESPI:DATATWO:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ESPI:DATATWO:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies a reference waveform as the source.

BusBItemEspiDatatwoThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ESPI:DATATWO:THReshold command.

Description
  • This command sets or queries the response (single mode)/ IO[1] (dual mode) Data source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:DATATWO:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:DATATWO:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:DATATWO:THReshold value command.
SCPI Syntax
- BUS:B<x>:ESPI:DATATWO:THReshold <NR3>
- BUS:B<x>:ESPI:DATATWO:THReshold?
Info
  • B<x> is the Bus number.
  • <NR3> sets the command/data threshold for the specified bus. The valid range is -8V to +8V.

BusBItemEspiIomode

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ESPI:IOMODe command.

Description
  • This command sets or queries the ESPI Input/Output mode for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ESPI:IOMODe? query.
  • Using the .verify(value) method will send the BUS:B<x>:ESPI:IOMODe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ESPI:IOMODe value command.
SCPI Syntax
- BUS:B<x>:ESPI:IOMODe {SINGle|DUAL}
- BUS:B<x>:ESPI:IOMODe?
Info
  • B<x> is the Bus number.
  • SINGle displays the command and response decode in two lanes.
  • DUAL displays the decode in a single data lane.

BusBItemEthercat

Bases: SCPICmdRead

The BUS:B<x>:ETHERCAT command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .dataminusthreshold: The BUS:B<x>:ETHERCAT:DATAMINUSTHRESHold command.
  • .dataplusthreshold: The BUS:B<x>:ETHERCAT:DATAPLUSTHRESHold command.
  • .signaltype: The BUS:B<x>:ETHERCAT:SIGNALTYpe command.
  • .source: The BUS:B<x>:ETHERCAT:SOUrce command tree.
  • .threshold: The BUS:B<x>:ETHERCAT:THRESHold command.

dataminusthreshold property

Return the BUS:B<x>:ETHERCAT:DATAMINUSTHRESHold command.

Description
  • This command sets or queries the DATA Minus source threshold for the specified EtherCAT bus.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT:DATAMINUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT:DATAMINUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERCAT:DATAMINUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:ETHERCAT:DATAMINUSTHRESHold <NR3>
- BUS:B<x>:ETHERCAT:DATAMINUSTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the EtherCAT DataMinus source threshold for the specified bus. The valid range is -8V to +8V. The default value is 0 V.

dataplusthreshold property

Return the BUS:B<x>:ETHERCAT:DATAPLUSTHRESHold command.

Description
  • This command sets or queries the DATA Plus source threshold for the specified EtherCAT bus.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT:DATAPLUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT:DATAPLUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERCAT:DATAPLUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:ETHERCAT:DATAPLUSTHRESHold <NR3>
- BUS:B<x>:ETHERCAT:DATAPLUSTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the EtherCAT DataPinus source threshold for the specified bus. The valid range is -8V to +8V. The default value is 0 V.

signaltype property

Return the BUS:B<x>:ETHERCAT:SIGNALTYpe command.

Description
  • This command sets or queries the signal type for the specified EtherCAT bus.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT:SIGNALTYpe? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT:SIGNALTYpe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERCAT:SIGNALTYpe value command.
SCPI Syntax
- BUS:B<x>:ETHERCAT:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:ETHERCAT:SIGNALTYpe?
Info
  • B<x> is the number of the bus.
  • SINGLE specifies single-ended signals.
  • DIFF specifies differential signals.

source property

Return the BUS:B<x>:ETHERCAT:SOUrce command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT:SOUrce? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .diff: The BUS:B<x>:ETHERCAT:SOUrce:DIFF command.
  • .dminus: The BUS:B<x>:ETHERCAT:SOUrce:DMINus command.
  • .dplus: The BUS:B<x>:ETHERCAT:SOUrce:DPLUs command.

threshold property

Return the BUS:B<x>:ETHERCAT:THRESHold command.

Description
  • This command sets or queries the differential source threshold for the specified EtherCAT bus.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERCAT:THRESHold value command.
SCPI Syntax
- BUS:B<x>:ETHERCAT:THRESHold <NR3>
- BUS:B<x>:ETHERCAT:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> specifies the EtherCAT differential Source threshold for the specified bus. The default value is 0 V. The valid range is -8 V to +8 V.

BusBItemEthercatDataminusthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERCAT:DATAMINUSTHRESHold command.

Description
  • This command sets or queries the DATA Minus source threshold for the specified EtherCAT bus.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT:DATAMINUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT:DATAMINUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERCAT:DATAMINUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:ETHERCAT:DATAMINUSTHRESHold <NR3>
- BUS:B<x>:ETHERCAT:DATAMINUSTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the EtherCAT DataMinus source threshold for the specified bus. The valid range is -8V to +8V. The default value is 0 V.

BusBItemEthercatDataplusthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERCAT:DATAPLUSTHRESHold command.

Description
  • This command sets or queries the DATA Plus source threshold for the specified EtherCAT bus.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT:DATAPLUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT:DATAPLUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERCAT:DATAPLUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:ETHERCAT:DATAPLUSTHRESHold <NR3>
- BUS:B<x>:ETHERCAT:DATAPLUSTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the EtherCAT DataPinus source threshold for the specified bus. The valid range is -8V to +8V. The default value is 0 V.

BusBItemEthercatSignaltype

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERCAT:SIGNALTYpe command.

Description
  • This command sets or queries the signal type for the specified EtherCAT bus.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT:SIGNALTYpe? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT:SIGNALTYpe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERCAT:SIGNALTYpe value command.
SCPI Syntax
- BUS:B<x>:ETHERCAT:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:ETHERCAT:SIGNALTYpe?
Info
  • B<x> is the number of the bus.
  • SINGLE specifies single-ended signals.
  • DIFF specifies differential signals.

BusBItemEthercatSource

Bases: SCPICmdRead

The BUS:B<x>:ETHERCAT:SOUrce command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT:SOUrce? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .diff: The BUS:B<x>:ETHERCAT:SOUrce:DIFF command.
  • .dminus: The BUS:B<x>:ETHERCAT:SOUrce:DMINus command.
  • .dplus: The BUS:B<x>:ETHERCAT:SOUrce:DPLUs command.

diff property

Return the BUS:B<x>:ETHERCAT:SOUrce:DIFF command.

Description
  • This command sets or queries the differential source for the specified EtherCAT bus.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT:SOUrce:DIFF? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT:SOUrce:DIFF? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERCAT:SOUrce:DIFF value command.
SCPI Syntax
- BUS:B<x>:ETHERCAT:SOUrce:DIFF {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERCAT:SOUrce:DIFF?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies to use one of the analog channels as the differential source.
  • MATH<x> specifies to use a math waveform as the differential source.
  • REF<x> specifies to use one of the reference waveforms as the differential source.

dminus property

Return the BUS:B<x>:ETHERCAT:SOUrce:DMINus command.

Description
  • This command sets or queries the DataMinus (SDATAMINUS) source for the specified EtherCAT bus.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT:SOUrce:DMINus? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT:SOUrce:DMINus? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERCAT:SOUrce:DMINus value command.
SCPI Syntax
- BUS:B<x>:ETHERCAT:SOUrce:DMINus {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERCAT:SOUrce:DMINus?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies to use one of the analog channels as the DataMinus source.
  • MATH<x> specifies to use a math waveform as the DataMinus source.
  • REF<x> specifies to use one of the reference waveforms as the DataMinus source.

dplus property

Return the BUS:B<x>:ETHERCAT:SOUrce:DPLUs command.

Description
  • This command sets or queries the DataPlus (SDATAPLUS) source for the specified EtherCAT bus.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT:SOUrce:DPLUs? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT:SOUrce:DPLUs? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERCAT:SOUrce:DPLUs value command.
SCPI Syntax
- BUS:B<x>:ETHERCAT:SOUrce:DPLUs {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERCAT:SOUrce:DPLUs?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies to use one of the analog channels as the DataPlus source.
  • MATH<x> specifies to use a math waveform as the DataPlus source.
  • REF<x> specifies to use one of the reference waveforms as the DataPlus source.

BusBItemEthercatSourceDiff

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERCAT:SOUrce:DIFF command.

Description
  • This command sets or queries the differential source for the specified EtherCAT bus.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT:SOUrce:DIFF? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT:SOUrce:DIFF? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERCAT:SOUrce:DIFF value command.
SCPI Syntax
- BUS:B<x>:ETHERCAT:SOUrce:DIFF {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERCAT:SOUrce:DIFF?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies to use one of the analog channels as the differential source.
  • MATH<x> specifies to use a math waveform as the differential source.
  • REF<x> specifies to use one of the reference waveforms as the differential source.

BusBItemEthercatSourceDminus

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERCAT:SOUrce:DMINus command.

Description
  • This command sets or queries the DataMinus (SDATAMINUS) source for the specified EtherCAT bus.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT:SOUrce:DMINus? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT:SOUrce:DMINus? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERCAT:SOUrce:DMINus value command.
SCPI Syntax
- BUS:B<x>:ETHERCAT:SOUrce:DMINus {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERCAT:SOUrce:DMINus?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies to use one of the analog channels as the DataMinus source.
  • MATH<x> specifies to use a math waveform as the DataMinus source.
  • REF<x> specifies to use one of the reference waveforms as the DataMinus source.

BusBItemEthercatSourceDplus

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERCAT:SOUrce:DPLUs command.

Description
  • This command sets or queries the DataPlus (SDATAPLUS) source for the specified EtherCAT bus.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT:SOUrce:DPLUs? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT:SOUrce:DPLUs? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERCAT:SOUrce:DPLUs value command.
SCPI Syntax
- BUS:B<x>:ETHERCAT:SOUrce:DPLUs {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERCAT:SOUrce:DPLUs?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies to use one of the analog channels as the DataPlus source.
  • MATH<x> specifies to use a math waveform as the DataPlus source.
  • REF<x> specifies to use one of the reference waveforms as the DataPlus source.

BusBItemEthercatThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERCAT:THRESHold command.

Description
  • This command sets or queries the differential source threshold for the specified EtherCAT bus.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERCAT:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERCAT:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERCAT:THRESHold value command.
SCPI Syntax
- BUS:B<x>:ETHERCAT:THRESHold <NR3>
- BUS:B<x>:ETHERCAT:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> specifies the EtherCAT differential Source threshold for the specified bus. The default value is 0 V. The valid range is -8 V to +8 V.

BusBItemEthernet

Bases: SCPICmdRead

The BUS:B<x>:ETHERnet command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .dataminusthreshold: The BUS:B<x>:ETHERnet:DATAMINUSTHRESHold command.
  • .dataplusthreshold: The BUS:B<x>:ETHERnet:DATAPLUSTHRESHold command.
  • .ipvfour: The BUS:B<x>:ETHERnet:IPVFOUR command.
  • .lowthreshold: The BUS:B<x>:ETHERnet:LOWTHRESHold command.
  • .qtagging: The BUS:B<x>:ETHERnet:QTAGGING command.
  • .signaltype: The BUS:B<x>:ETHERnet:SIGNALTYpe command.
  • .source: The BUS:B<x>:ETHERnet:SOUrce command.
  • .threshold: The BUS:B<x>:ETHERnet:THRESHold command.
  • .type: The BUS:B<x>:ETHERnet:TYPe command.

dataminusthreshold property

Return the BUS:B<x>:ETHERnet:DATAMINUSTHRESHold command.

Description
  • This command sets or queries the Ethernet D- source threshold for the specified bus. This threshold only applies when the Ethernet signal type is single ended. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:DATAMINUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:DATAMINUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:DATAMINUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:DATAMINUSTHRESHold <NR3>
- BUS:B<x>:ETHERnet:DATAMINUSTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the Ethernet D- source threshold for the specified bus.

dataplusthreshold property

Return the BUS:B<x>:ETHERnet:DATAPLUSTHRESHold command.

Description
  • This command sets or queries the Ethernet D+ source threshold for the specified bus. This threshold only applies when the Ethernet signal type is single ended. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:DATAPLUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:DATAPLUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:DATAPLUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:DATAPLUSTHRESHold <NR3>
- BUS:B<x>:ETHERnet:DATAPLUSTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the Ethernet D+ source threshold for the specified bus.

ipvfour property

Return the BUS:B<x>:ETHERnet:IPVFOUR command.

Description
  • This command sets or queries whether IPV4 packets are available for triggering on Ethernet. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:IPVFOUR? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:IPVFOUR? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:IPVFOUR value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:IPVFOUR {YES|NO}
- BUS:B<x>:ETHERnet:IPVFOUR?
Info
  • B<x> is the number of the bus.
  • YES specifies that IPV4 packets are available.
  • NO specifies that IPV4 packets are not available.

lowthreshold property

Return the BUS:B<x>:ETHERnet:LOWTHRESHold command.

Description
  • This command sets or queries the Ethernet source Low threshold for the specified bus. This threshold only applies when the Ethernet signal type is differential. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:LOWTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:LOWTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:LOWTHRESHold value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:LOWTHRESHold <NR3>
- BUS:B<x>:ETHERnet:LOWTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the Ethernet source Low threshold for the specified bus.

qtagging property

Return the BUS:B<x>:ETHERnet:QTAGGING command.

Description
  • This command sets or queries whether Q-Tagging packets are available for triggering on Ethernet. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:QTAGGING? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:QTAGGING? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:QTAGGING value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:QTAGGING {YES|NO}
- BUS:B<x>:ETHERnet:QTAGGING?
Info
  • B<x> is the number of the bus.
  • YES specifies that Q-Tagging packets are available.
  • NO specifies that Q-Tagging packets are not available.

signaltype property

Return the BUS:B<x>:ETHERnet:SIGNALTYpe command.

Description
  • This command sets or queries the Ethernet signal type for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:SIGNALTYpe? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:SIGNALTYpe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:SIGNALTYpe value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:ETHERnet:SIGNALTYpe?
Info
  • B<x> is the number of the bus.
  • SINGLE specifies single-ended signals.
  • DIFF specifies differential signals.

source property

Return the BUS:B<x>:ETHERnet:SOUrce command.

Description
  • This command sets or queries the Ethernet data (SDATA) source for the specified bus. This command controls the source channel when the signal type is differential. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:SOUrce value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERnet:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies to use one of the analog channels as the Ethernet data source for differential input.
  • MATH<x> specifies to use a math waveform as the source for Ethernet data differential input.
  • REF<x> specifies to use one of the reference waveforms as the Ethernet data source for differential input.
Sub-properties
  • .dminus: The BUS:B<x>:ETHERnet:SOUrce:DMINus command.
  • .dplus: The BUS:B<x>:ETHERnet:SOUrce:DPLUs command.

threshold property

Return the BUS:B<x>:ETHERnet:THRESHold command.

Description
  • This command sets or queries the Ethernet DATA source High threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:THRESHold value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:THRESHold <NR3>
- BUS:B<x>:ETHERnet:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the Ethernet DATA source High threshold for the specified bus.

type property

Return the BUS:B<x>:ETHERnet:TYPe command.

Description
  • This command specifies the Ethernet standard speed. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:TYPe? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:TYPe value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:TYPe {TENBASET|HUNDREDBASETX}
- BUS:B<x>:ETHERnet:TYPe?
Info
  • B<x> is the number of the bus.
  • TENBASET specifies the Ethernet speed as 10Base-T.
  • HUNDREDBASETX specifies the Ethernet speed as 100Base-T.

BusBItemEthernetDataminusthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERnet:DATAMINUSTHRESHold command.

Description
  • This command sets or queries the Ethernet D- source threshold for the specified bus. This threshold only applies when the Ethernet signal type is single ended. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:DATAMINUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:DATAMINUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:DATAMINUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:DATAMINUSTHRESHold <NR3>
- BUS:B<x>:ETHERnet:DATAMINUSTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the Ethernet D- source threshold for the specified bus.

BusBItemEthernetDataplusthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERnet:DATAPLUSTHRESHold command.

Description
  • This command sets or queries the Ethernet D+ source threshold for the specified bus. This threshold only applies when the Ethernet signal type is single ended. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:DATAPLUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:DATAPLUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:DATAPLUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:DATAPLUSTHRESHold <NR3>
- BUS:B<x>:ETHERnet:DATAPLUSTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the Ethernet D+ source threshold for the specified bus.

BusBItemEthernetIpvfour

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERnet:IPVFOUR command.

Description
  • This command sets or queries whether IPV4 packets are available for triggering on Ethernet. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:IPVFOUR? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:IPVFOUR? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:IPVFOUR value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:IPVFOUR {YES|NO}
- BUS:B<x>:ETHERnet:IPVFOUR?
Info
  • B<x> is the number of the bus.
  • YES specifies that IPV4 packets are available.
  • NO specifies that IPV4 packets are not available.

BusBItemEthernetLowthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERnet:LOWTHRESHold command.

Description
  • This command sets or queries the Ethernet source Low threshold for the specified bus. This threshold only applies when the Ethernet signal type is differential. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:LOWTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:LOWTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:LOWTHRESHold value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:LOWTHRESHold <NR3>
- BUS:B<x>:ETHERnet:LOWTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the Ethernet source Low threshold for the specified bus.

BusBItemEthernetQtagging

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERnet:QTAGGING command.

Description
  • This command sets or queries whether Q-Tagging packets are available for triggering on Ethernet. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:QTAGGING? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:QTAGGING? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:QTAGGING value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:QTAGGING {YES|NO}
- BUS:B<x>:ETHERnet:QTAGGING?
Info
  • B<x> is the number of the bus.
  • YES specifies that Q-Tagging packets are available.
  • NO specifies that Q-Tagging packets are not available.

BusBItemEthernetSignaltype

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERnet:SIGNALTYpe command.

Description
  • This command sets or queries the Ethernet signal type for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:SIGNALTYpe? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:SIGNALTYpe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:SIGNALTYpe value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:ETHERnet:SIGNALTYpe?
Info
  • B<x> is the number of the bus.
  • SINGLE specifies single-ended signals.
  • DIFF specifies differential signals.

BusBItemEthernetSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERnet:SOUrce command.

Description
  • This command sets or queries the Ethernet data (SDATA) source for the specified bus. This command controls the source channel when the signal type is differential. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:SOUrce value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERnet:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies to use one of the analog channels as the Ethernet data source for differential input.
  • MATH<x> specifies to use a math waveform as the source for Ethernet data differential input.
  • REF<x> specifies to use one of the reference waveforms as the Ethernet data source for differential input.
Properties
  • .dminus: The BUS:B<x>:ETHERnet:SOUrce:DMINus command.
  • .dplus: The BUS:B<x>:ETHERnet:SOUrce:DPLUs command.

dminus property

Return the BUS:B<x>:ETHERnet:SOUrce:DMINus command.

Description
  • This command sets or queries the Ethernet D- source for the specified bus. this command specifies the source channel to use when the signal type is single ended. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:SOUrce:DMINus? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:SOUrce:DMINus? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:SOUrce:DMINus value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:SOUrce:DMINus {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERnet:SOUrce:DMINus?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.

dplus property

Return the BUS:B<x>:ETHERnet:SOUrce:DPLUs command.

Description
  • This command sets or queries the Ethernet D+ source for the specified bus. this command specifies the source channel to use when the signal type is single ended. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:SOUrce:DPLUs? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:SOUrce:DPLUs? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:SOUrce:DPLUs value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:SOUrce:DPLUs {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERnet:SOUrce:DPLUs?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.

BusBItemEthernetSourceDminus

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERnet:SOUrce:DMINus command.

Description
  • This command sets or queries the Ethernet D- source for the specified bus. this command specifies the source channel to use when the signal type is single ended. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:SOUrce:DMINus? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:SOUrce:DMINus? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:SOUrce:DMINus value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:SOUrce:DMINus {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERnet:SOUrce:DMINus?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.

BusBItemEthernetSourceDplus

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERnet:SOUrce:DPLUs command.

Description
  • This command sets or queries the Ethernet D+ source for the specified bus. this command specifies the source channel to use when the signal type is single ended. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:SOUrce:DPLUs? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:SOUrce:DPLUs? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:SOUrce:DPLUs value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:SOUrce:DPLUs {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ETHERnet:SOUrce:DPLUs?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.

BusBItemEthernetThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERnet:THRESHold command.

Description
  • This command sets or queries the Ethernet DATA source High threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:THRESHold value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:THRESHold <NR3>
- BUS:B<x>:ETHERnet:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the Ethernet DATA source High threshold for the specified bus.

BusBItemEthernetType

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ETHERnet:TYPe command.

Description
  • This command specifies the Ethernet standard speed. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ETHERnet:TYPe? query.
  • Using the .verify(value) method will send the BUS:B<x>:ETHERnet:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ETHERnet:TYPe value command.
SCPI Syntax
- BUS:B<x>:ETHERnet:TYPe {TENBASET|HUNDREDBASETX}
- BUS:B<x>:ETHERnet:TYPe?
Info
  • B<x> is the number of the bus.
  • TENBASET specifies the Ethernet speed as 10Base-T.
  • HUNDREDBASETX specifies the Ethernet speed as 100Base-T.

BusBItemEusb

Bases: SCPICmdRead

The BUS:B<x>:EUSB command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:EUSB? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .bitrate: The BUS:B<x>:EUSB:BITRate command.
  • .dataminus: The BUS:B<x>:EUSB:DATAMINUS command tree.
  • .dataminusthreshold: The BUS:B<x>:EUSB:DATAMINUSTHRESHold command.
  • .dataplus: The BUS:B<x>:EUSB:DATAPLUS command tree.
  • .dataplusthreshold: The BUS:B<x>:EUSB:DATAPLUSTHRESHold command.
  • .lowthreshold: The BUS:B<x>:EUSB:LOWTHRESHold command.
  • .operating: The BUS:B<x>:EUSB:OPERating command tree.
  • .signaltype: The BUS:B<x>:EUSB:SIGNALTYpe command.
  • .source: The BUS:B<x>:EUSB:SOUrce command tree.
  • .threshold: The BUS:B<x>:EUSB:THRESHold command.

bitrate property

Return the BUS:B<x>:EUSB:BITRate command.

Description
  • This command sets or queries the eUSB data rate for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:BITRate value command.
SCPI Syntax
- BUS:B<x>:EUSB:BITRate {HIGH|FULL|LOW}
- BUS:B<x>:EUSB:BITRate?
Info
  • B<x> is the number of the bus.
  • HIGH sets the bit rate to Bit Rate to 480 Mbps.
  • FULL sets the bit rate to Bit Rate to 12 Mbps.
  • LOW sets the bit rate to Bit Rate to 1.5 Mbps.

dataminus property

Return the BUS:B<x>:EUSB:DATAMINUS command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:DATAMINUS? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:DATAMINUS? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .data: The BUS:B<x>:EUSB:DATAMINUS:DATA command tree.

dataminusthreshold property

dataminusthreshold: BusBItemEusbDataminusthreshold

Return the BUS:B<x>:EUSB:DATAMINUSTHRESHold command.

Description
  • This command sets or queries the eUSB D- Input Source Data Threshold for Data line decode for specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:DATAMINUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:DATAMINUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:DATAMINUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:EUSB:DATAMINUSTHRESHold <NR3>
- BUS:B<x>:EUSB:DATAMINUSTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the eUSB DATA Minus source threshold. The argument range is -8 V to +8 V.

dataplus property

Return the BUS:B<x>:EUSB:DATAPLUS command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:DATAPLUS? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:DATAPLUS? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .data: The BUS:B<x>:EUSB:DATAPLUS:DATA command tree.

dataplusthreshold property

dataplusthreshold: BusBItemEusbDataplusthreshold

Return the BUS:B<x>:EUSB:DATAPLUSTHRESHold command.

Description
  • This command sets or queries the eUSB DATA Plus source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:DATAPLUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:DATAPLUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:DATAPLUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:EUSB:DATAPLUSTHRESHold <NR3>
- BUS:B<x>:EUSB:DATAPLUSTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the eUSB DATA Plus source threshold. The argument range is -8 V to +8 V.

lowthreshold property

Return the BUS:B<x>:EUSB:LOWTHRESHold command.

Description
  • This command sets or queries the eUSB source Low threshold for the specified bus when signal type is Differential.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:LOWTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:LOWTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:LOWTHRESHold value command.
SCPI Syntax
- BUS:B<x>:EUSB:LOWTHRESHold <NR3>
- BUS:B<x>:EUSB:LOWTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the eUSB Strobe threshold for the specified bus. The argument range is -8 V to +8 V.

operating property

Return the BUS:B<x>:EUSB:OPERating command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:OPERating? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:OPERating? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .mode: The BUS:B<x>:EUSB:OPERating:MODe command.

signaltype property

Return the BUS:B<x>:EUSB:SIGNALTYpe command.

Description
  • This command sets or queries the eUSB signal type for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:SIGNALTYpe? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:SIGNALTYpe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:SIGNALTYpe value command.
SCPI Syntax
- BUS:B<x>:EUSB:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:EUSB:SIGNALTYpe?
Info
  • B<x> is the number of the bus.
  • SINGLE sets the signal type to single.
  • DIFF sets the signal type to differential.

source property

Return the BUS:B<x>:EUSB:SOUrce command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:SOUrce? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .diff: The BUS:B<x>:EUSB:SOUrce:DIFF command.
  • .dminus: The BUS:B<x>:EUSB:SOUrce:DMINus command.
  • .dplus: The BUS:B<x>:EUSB:SOUrce:DPLUs command.

threshold property

Return the BUS:B<x>:EUSB:THRESHold command.

Description
  • This command sets or queries the eUSB source High threshold for the specified bus when signal type is Diff.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:THRESHold value command.
SCPI Syntax
- BUS:B<x>:EUSB:THRESHold <NR3>
- BUS:B<x>:EUSB:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the eUSB Strobe threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemEusbBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:EUSB:BITRate command.

Description
  • This command sets or queries the eUSB data rate for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:BITRate value command.
SCPI Syntax
- BUS:B<x>:EUSB:BITRate {HIGH|FULL|LOW}
- BUS:B<x>:EUSB:BITRate?
Info
  • B<x> is the number of the bus.
  • HIGH sets the bit rate to Bit Rate to 480 Mbps.
  • FULL sets the bit rate to Bit Rate to 12 Mbps.
  • LOW sets the bit rate to Bit Rate to 1.5 Mbps.

BusBItemEusbDataminus

Bases: SCPICmdRead

The BUS:B<x>:EUSB:DATAMINUS command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:DATAMINUS? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:DATAMINUS? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .data: The BUS:B<x>:EUSB:DATAMINUS:DATA command tree.

data property

Return the BUS:B<x>:EUSB:DATAMINUS:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:DATAMINUS:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:DATAMINUS:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .threshold: The BUS:B<x>:EUSB:DATAMINUS:DATA:THRESHold command.

BusBItemEusbDataminusData

Bases: SCPICmdRead

The BUS:B<x>:EUSB:DATAMINUS:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:DATAMINUS:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:DATAMINUS:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .threshold: The BUS:B<x>:EUSB:DATAMINUS:DATA:THRESHold command.

threshold property

Return the BUS:B<x>:EUSB:DATAMINUS:DATA:THRESHold command.

Description
  • This command sets or queries the eUSB D- Input Source Data Threshold for Data line decode for specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:DATAMINUS:DATA:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:DATAMINUS:DATA:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:DATAMINUS:DATA:THRESHold value command.
SCPI Syntax
- BUS:B<x>:EUSB:DATAMINUS:DATA:THRESHold <NR3>
- BUS:B<x>:EUSB:DATAMINUS:DATA:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the EUSB Strobe threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemEusbDataminusDataThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:EUSB:DATAMINUS:DATA:THRESHold command.

Description
  • This command sets or queries the eUSB D- Input Source Data Threshold for Data line decode for specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:DATAMINUS:DATA:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:DATAMINUS:DATA:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:DATAMINUS:DATA:THRESHold value command.
SCPI Syntax
- BUS:B<x>:EUSB:DATAMINUS:DATA:THRESHold <NR3>
- BUS:B<x>:EUSB:DATAMINUS:DATA:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the EUSB Strobe threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemEusbDataminusthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:EUSB:DATAMINUSTHRESHold command.

Description
  • This command sets or queries the eUSB D- Input Source Data Threshold for Data line decode for specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:DATAMINUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:DATAMINUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:DATAMINUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:EUSB:DATAMINUSTHRESHold <NR3>
- BUS:B<x>:EUSB:DATAMINUSTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the eUSB DATA Minus source threshold. The argument range is -8 V to +8 V.

BusBItemEusbDataplus

Bases: SCPICmdRead

The BUS:B<x>:EUSB:DATAPLUS command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:DATAPLUS? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:DATAPLUS? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .data: The BUS:B<x>:EUSB:DATAPLUS:DATA command tree.

data property

Return the BUS:B<x>:EUSB:DATAPLUS:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:DATAPLUS:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:DATAPLUS:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .threshold: The BUS:B<x>:EUSB:DATAPLUS:DATA:THRESHold command.

BusBItemEusbDataplusData

Bases: SCPICmdRead

The BUS:B<x>:EUSB:DATAPLUS:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:DATAPLUS:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:DATAPLUS:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .threshold: The BUS:B<x>:EUSB:DATAPLUS:DATA:THRESHold command.

threshold property

Return the BUS:B<x>:EUSB:DATAPLUS:DATA:THRESHold command.

Description
  • This command sets or queries the eUSB D+ Input Source Data Threshold for Data line decode for specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:DATAPLUS:DATA:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:DATAPLUS:DATA:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:DATAPLUS:DATA:THRESHold value command.
SCPI Syntax
- BUS:B<x>:EUSB:DATAPLUS:DATA:THRESHold <NR3>
- BUS:B<x>:EUSB:DATAPLUS:DATA:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the eUSB Strobe threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemEusbDataplusDataThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:EUSB:DATAPLUS:DATA:THRESHold command.

Description
  • This command sets or queries the eUSB D+ Input Source Data Threshold for Data line decode for specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:DATAPLUS:DATA:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:DATAPLUS:DATA:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:DATAPLUS:DATA:THRESHold value command.
SCPI Syntax
- BUS:B<x>:EUSB:DATAPLUS:DATA:THRESHold <NR3>
- BUS:B<x>:EUSB:DATAPLUS:DATA:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the eUSB Strobe threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemEusbDataplusthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:EUSB:DATAPLUSTHRESHold command.

Description
  • This command sets or queries the eUSB DATA Plus source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:DATAPLUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:DATAPLUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:DATAPLUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:EUSB:DATAPLUSTHRESHold <NR3>
- BUS:B<x>:EUSB:DATAPLUSTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the eUSB DATA Plus source threshold. The argument range is -8 V to +8 V.

BusBItemEusbLowthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:EUSB:LOWTHRESHold command.

Description
  • This command sets or queries the eUSB source Low threshold for the specified bus when signal type is Differential.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:LOWTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:LOWTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:LOWTHRESHold value command.
SCPI Syntax
- BUS:B<x>:EUSB:LOWTHRESHold <NR3>
- BUS:B<x>:EUSB:LOWTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the eUSB Strobe threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemEusbOperating

Bases: SCPICmdRead

The BUS:B<x>:EUSB:OPERating command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:OPERating? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:OPERating? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .mode: The BUS:B<x>:EUSB:OPERating:MODe command.

mode property

Return the BUS:B<x>:EUSB:OPERating:MODe command.

Description
  • This command sets or queries the eUSB mode for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:OPERating:MODe? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:OPERating:MODe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:OPERating:MODe value command.
SCPI Syntax
- BUS:B<x>:EUSB:OPERating:MODe {NATive|REPEATERHOSt|REPEATERPERIPHERAL}
- BUS:B<x>:EUSB:OPERating:MODe?
Info
  • B<x> is the number of the bus.
  • NATive sets the operating mode as native.
  • REPEATERHOSt sets the operating mode as repeater host.
  • REPEATERPERIPHERAL sets the operating mode as repeater peripheral.

BusBItemEusbOperatingMode

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:EUSB:OPERating:MODe command.

Description
  • This command sets or queries the eUSB mode for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:OPERating:MODe? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:OPERating:MODe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:OPERating:MODe value command.
SCPI Syntax
- BUS:B<x>:EUSB:OPERating:MODe {NATive|REPEATERHOSt|REPEATERPERIPHERAL}
- BUS:B<x>:EUSB:OPERating:MODe?
Info
  • B<x> is the number of the bus.
  • NATive sets the operating mode as native.
  • REPEATERHOSt sets the operating mode as repeater host.
  • REPEATERPERIPHERAL sets the operating mode as repeater peripheral.

BusBItemEusbSignaltype

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:EUSB:SIGNALTYpe command.

Description
  • This command sets or queries the eUSB signal type for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:SIGNALTYpe? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:SIGNALTYpe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:SIGNALTYpe value command.
SCPI Syntax
- BUS:B<x>:EUSB:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:EUSB:SIGNALTYpe?
Info
  • B<x> is the number of the bus.
  • SINGLE sets the signal type to single.
  • DIFF sets the signal type to differential.

BusBItemEusbSource

Bases: SCPICmdRead

The BUS:B<x>:EUSB:SOUrce command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:SOUrce? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .diff: The BUS:B<x>:EUSB:SOUrce:DIFF command.
  • .dminus: The BUS:B<x>:EUSB:SOUrce:DMINus command.
  • .dplus: The BUS:B<x>:EUSB:SOUrce:DPLUs command.

diff property

Return the BUS:B<x>:EUSB:SOUrce:DIFF command.

Description
  • This command sets or queries the eUSB Diff source for the specified bus when signal type is Diff.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:SOUrce:DIFF? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:SOUrce:DIFF? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:SOUrce:DIFF value command.
SCPI Syntax
- BUS:B<x>:EUSB:SOUrce:DIFF {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:EUSB:SOUrce:DIFF?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.

dminus property

Return the BUS:B<x>:EUSB:SOUrce:DMINus command.

Description
  • This command sets or queries the eUSB DataMinus (SDATAMINUS) source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:SOUrce:DMINus? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:SOUrce:DMINus? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:SOUrce:DMINus value command.
SCPI Syntax
- BUS:B<x>:EUSB:SOUrce:DMINus {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:EUSB:SOUrce:DMINus?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.

dplus property

Return the BUS:B<x>:EUSB:SOUrce:DPLUs command.

Description
  • This command sets or queries the eUSB dataPlus (SDATAPLUS) source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:SOUrce:DPLUs? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:SOUrce:DPLUs? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:SOUrce:DPLUs value command.
SCPI Syntax
- BUS:B<x>:EUSB:SOUrce:DPLUs {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:EUSB:SOUrce:DPLUs?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.

BusBItemEusbSourceDiff

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:EUSB:SOUrce:DIFF command.

Description
  • This command sets or queries the eUSB Diff source for the specified bus when signal type is Diff.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:SOUrce:DIFF? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:SOUrce:DIFF? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:SOUrce:DIFF value command.
SCPI Syntax
- BUS:B<x>:EUSB:SOUrce:DIFF {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:EUSB:SOUrce:DIFF?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.

BusBItemEusbSourceDminus

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:EUSB:SOUrce:DMINus command.

Description
  • This command sets or queries the eUSB DataMinus (SDATAMINUS) source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:SOUrce:DMINus? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:SOUrce:DMINus? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:SOUrce:DMINus value command.
SCPI Syntax
- BUS:B<x>:EUSB:SOUrce:DMINus {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:EUSB:SOUrce:DMINus?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.

BusBItemEusbSourceDplus

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:EUSB:SOUrce:DPLUs command.

Description
  • This command sets or queries the eUSB dataPlus (SDATAPLUS) source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:SOUrce:DPLUs? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:SOUrce:DPLUs? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:SOUrce:DPLUs value command.
SCPI Syntax
- BUS:B<x>:EUSB:SOUrce:DPLUs {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:EUSB:SOUrce:DPLUs?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.

BusBItemEusbThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:EUSB:THRESHold command.

Description
  • This command sets or queries the eUSB source High threshold for the specified bus when signal type is Diff.
Usage
  • Using the .query() method will send the BUS:B<x>:EUSB:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:EUSB:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:EUSB:THRESHold value command.
SCPI Syntax
- BUS:B<x>:EUSB:THRESHold <NR3>
- BUS:B<x>:EUSB:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the eUSB Strobe threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemFlexray

Bases: SCPICmdRead

The BUS:B<x>:FLEXray command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .bitrate: The BUS:B<x>:FLEXray:BITRate command.
  • .channel: The BUS:B<x>:FLEXray:CHannel command.
  • .lowthreshold: The BUS:B<x>:FLEXray:LOWTHRESHold command.
  • .signal: The BUS:B<x>:FLEXray:SIGnal command.
  • .source: The BUS:B<x>:FLEXray:SOUrce command.
  • .threshold: The BUS:B<x>:FLEXray:THRESHold command.
  • .txrxthreshold: The BUS:B<x>:FLEXray:TXRXTHRESHold command.

bitrate property

Return the BUS:B<x>:FLEXray:BITRate command.

Description
  • This command sets or queries the FlexRay bus bit rate. The bus is specified by x. If you select Custom, use BUS:B<x>:FLEXray:BITRate:CUSTom to set the bit rate.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:BITRate value command.
SCPI Syntax
- BUS:B<x>:FLEXray:BITRate {CUSTOM|RATE2M|RATE5M|RATE10M}
- BUS:B<x>:FLEXray:BITRate?
Info
  • B<x> is the number of the bus.
  • CUSTOM sets a custom bit rate.
  • RATE2M sets the bit rate to 2 Mb.
  • RATE5M sets the bit rate to 5 Mb.
  • RATE10M sets the bit rate to 10 Mb.
Sub-properties
  • .custom: The BUS:B<x>:FLEXray:BITRate:CUSTom command.

channel property

Return the BUS:B<x>:FLEXray:CHannel command.

Description
  • This command sets or queries the FlexRay channel type for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:CHannel? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:CHannel? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:CHannel value command.
SCPI Syntax
- BUS:B<x>:FLEXray:CHannel {A|B}
- BUS:B<x>:FLEXray:CHannel?
Info
  • B<x> is the number of the bus.
  • A specifies the A channel.
  • B specifies the B channel.

lowthreshold property

Return the BUS:B<x>:FLEXray:LOWTHRESHold command.

Description
  • This command sets or queries the FlexRay data source low threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:LOWTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:LOWTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:LOWTHRESHold value command.
SCPI Syntax
- BUS:B<x>:FLEXray:LOWTHRESHold <NR3>
- BUS:B<x>:FLEXray:LOWTHRESHold?
Info
  • <NR3> is the FlexRay data source low threshold for the specified bus.
  • B<x> is the number of the bus.

signal property

Return the BUS:B<x>:FLEXray:SIGnal command.

Description
  • This command sets or queries the FlexRay signal type for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:SIGnal? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:SIGnal? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:SIGnal value command.
SCPI Syntax
- BUS:B<x>:FLEXray:SIGnal {BDIFFBP|BM|TXRX}
- BUS:B<x>:FLEXray:SIGnal?
Info
  • B<x> is the number of the bus.
  • BDIFFBP sets the FlexRay signal type to BDIFFBP.
  • BM sets the FlexRay signal type to BM.
  • TXRX sets the FlexRay signal type to TXRX.

source property

Return the BUS:B<x>:FLEXray:SOUrce command.

Description
  • This command sets or queries the Flexray bus data source for the specified bus when the signal type is BDIFFBP or BM. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:SOUrce value command.
SCPI Syntax
- BUS:B<x>:FLEXray:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:FLEXray:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.
Sub-properties
  • .txrx: The BUS:B<x>:FLEXray:SOUrce:TXRX command.

threshold property

Return the BUS:B<x>:FLEXray:THRESHold command.

Description
  • This command sets or queries the FlexRay data source high threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:THRESHold value command.
SCPI Syntax
- BUS:B<x>:FLEXray:THRESHold <NR3>
- BUS:B<x>:FLEXray:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the FlexRay data source high threshold for the specified bus.

txrxthreshold property

Return the BUS:B<x>:FLEXray:TXRXTHRESHold command.

Description
  • This command sets or queries the FlexRay data source TxRx threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:TXRXTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:TXRXTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:TXRXTHRESHold value command.
SCPI Syntax
- BUS:B<x>:FLEXray:TXRXTHRESHold <NR3>
- BUS:B<x>:FLEXray:TXRXTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the TxRx threshold.

BusBItemFlexrayBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:FLEXray:BITRate command.

Description
  • This command sets or queries the FlexRay bus bit rate. The bus is specified by x. If you select Custom, use BUS:B<x>:FLEXray:BITRate:CUSTom to set the bit rate.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:BITRate value command.
SCPI Syntax
- BUS:B<x>:FLEXray:BITRate {CUSTOM|RATE2M|RATE5M|RATE10M}
- BUS:B<x>:FLEXray:BITRate?
Info
  • B<x> is the number of the bus.
  • CUSTOM sets a custom bit rate.
  • RATE2M sets the bit rate to 2 Mb.
  • RATE5M sets the bit rate to 5 Mb.
  • RATE10M sets the bit rate to 10 Mb.
Properties
  • .custom: The BUS:B<x>:FLEXray:BITRate:CUSTom command.

custom property

Return the BUS:B<x>:FLEXray:BITRate:CUSTom command.

Description
  • This command sets or queries the FlexRay custom bit rate for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:BITRate:CUSTom? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:BITRate:CUSTom? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:BITRate:CUSTom value command.
SCPI Syntax
- BUS:B<x>:FLEXray:BITRate:CUSTom <NR1>
- BUS:B<x>:FLEXray:BITRate:CUSTom?
Info
  • B<x> is the number of the bus.
  • <NR1> is the FlexRay custom bit rate for the specified bus.

BusBItemFlexrayBitrateCustom

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:FLEXray:BITRate:CUSTom command.

Description
  • This command sets or queries the FlexRay custom bit rate for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:BITRate:CUSTom? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:BITRate:CUSTom? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:BITRate:CUSTom value command.
SCPI Syntax
- BUS:B<x>:FLEXray:BITRate:CUSTom <NR1>
- BUS:B<x>:FLEXray:BITRate:CUSTom?
Info
  • B<x> is the number of the bus.
  • <NR1> is the FlexRay custom bit rate for the specified bus.

BusBItemFlexrayChannel

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:FLEXray:CHannel command.

Description
  • This command sets or queries the FlexRay channel type for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:CHannel? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:CHannel? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:CHannel value command.
SCPI Syntax
- BUS:B<x>:FLEXray:CHannel {A|B}
- BUS:B<x>:FLEXray:CHannel?
Info
  • B<x> is the number of the bus.
  • A specifies the A channel.
  • B specifies the B channel.

BusBItemFlexrayLowthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:FLEXray:LOWTHRESHold command.

Description
  • This command sets or queries the FlexRay data source low threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:LOWTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:LOWTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:LOWTHRESHold value command.
SCPI Syntax
- BUS:B<x>:FLEXray:LOWTHRESHold <NR3>
- BUS:B<x>:FLEXray:LOWTHRESHold?
Info
  • <NR3> is the FlexRay data source low threshold for the specified bus.
  • B<x> is the number of the bus.

BusBItemFlexraySignal

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:FLEXray:SIGnal command.

Description
  • This command sets or queries the FlexRay signal type for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:SIGnal? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:SIGnal? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:SIGnal value command.
SCPI Syntax
- BUS:B<x>:FLEXray:SIGnal {BDIFFBP|BM|TXRX}
- BUS:B<x>:FLEXray:SIGnal?
Info
  • B<x> is the number of the bus.
  • BDIFFBP sets the FlexRay signal type to BDIFFBP.
  • BM sets the FlexRay signal type to BM.
  • TXRX sets the FlexRay signal type to TXRX.

BusBItemFlexraySource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:FLEXray:SOUrce command.

Description
  • This command sets or queries the Flexray bus data source for the specified bus when the signal type is BDIFFBP or BM. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:SOUrce value command.
SCPI Syntax
- BUS:B<x>:FLEXray:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:FLEXray:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.
Properties
  • .txrx: The BUS:B<x>:FLEXray:SOUrce:TXRX command.

txrx property

Return the BUS:B<x>:FLEXray:SOUrce:TXRX command.

Description
  • This command sets or queries the FlexRay TxRx data source for the specified bus when the signal type is TXRX. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:SOUrce:TXRX? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:SOUrce:TXRX? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:SOUrce:TXRX value command.
SCPI Syntax
- BUS:B<x>:FLEXray:SOUrce:TXRX {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:FLEXray:SOUrce:TXRX?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.
  • REF<x>_D<x> specifies a digital reference waveform as the source.

BusBItemFlexraySourceTxrx

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:FLEXray:SOUrce:TXRX command.

Description
  • This command sets or queries the FlexRay TxRx data source for the specified bus when the signal type is TXRX. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:SOUrce:TXRX? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:SOUrce:TXRX? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:SOUrce:TXRX value command.
SCPI Syntax
- BUS:B<x>:FLEXray:SOUrce:TXRX {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:FLEXray:SOUrce:TXRX?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.
  • REF<x>_D<x> specifies a digital reference waveform as the source.

BusBItemFlexrayThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:FLEXray:THRESHold command.

Description
  • This command sets or queries the FlexRay data source high threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:THRESHold value command.
SCPI Syntax
- BUS:B<x>:FLEXray:THRESHold <NR3>
- BUS:B<x>:FLEXray:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the FlexRay data source high threshold for the specified bus.

BusBItemFlexrayTxrxthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:FLEXray:TXRXTHRESHold command.

Description
  • This command sets or queries the FlexRay data source TxRx threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:FLEXray:TXRXTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:FLEXray:TXRXTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:FLEXray:TXRXTHRESHold value command.
SCPI Syntax
- BUS:B<x>:FLEXray:TXRXTHRESHold <NR3>
- BUS:B<x>:FLEXray:TXRXTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the TxRx threshold.

BusBItemI2c

Bases: SCPICmdRead

The BUS:B<x>:I2C command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:I2C? query.
  • Using the .verify(value) method will send the BUS:B<x>:I2C? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .clock: The BUS:B<x>:I2C:CLOCk command tree.
  • .data: The BUS:B<x>:I2C:DATa command tree.
  • .rwinaddr: The BUS:B<x>:I2C:RWINADDR command.

clock property

Return the BUS:B<x>:I2C:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:I2C:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:I2C:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .source: The BUS:B<x>:I2C:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:I2C:CLOCk:THReshold command.

data property

Return the BUS:B<x>:I2C:DATa command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:I2C:DATa? query.
  • Using the .verify(value) method will send the BUS:B<x>:I2C:DATa? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Sub-properties
  • .source: The BUS:B<x>:I2C:DATa:SOUrce command.
  • .threshold: The BUS:B<x>:I2C:DATa:THReshold command.

rwinaddr property

Return the BUS:B<x>:I2C:RWINADDR command.

Description
  • This command sets or queries the manner in which seven-bit I2C addresses are represented in the busform display of the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:I2C:RWINADDR? query.
  • Using the .verify(value) method will send the BUS:B<x>:I2C:RWINADDR? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I2C:RWINADDR value command.
SCPI Syntax
- BUS:B<x>:I2C:RWINADDR {0|1}
- BUS:B<x>:I2C:RWINADDR?
Info
  • B<x> is the number of the bus.
  • 0 displays seven-bit slave addresses as integers in the range of 0 to 127, with the state of the R/W* bit from the LSB of the slave address byte. For example, the slave address byte of 0b10100101 is displayed as the value 0x52 R.
  • 1 displays the entire slave address byte as a number, with the R/W* signal as its LSB (bit 0) and the slave address in bits 7..1. For example, the slave address byte of 0b10100101 is displayed as the value 0xA5 R.

BusBItemI2cClock

Bases: SCPICmdRead

The BUS:B<x>:I2C:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:I2C:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:I2C:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .source: The BUS:B<x>:I2C:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:I2C:CLOCk:THReshold command.

source property

Return the BUS:B<x>:I2C:CLOCk:SOUrce command.

Description
  • This command sets or queries the I2C clock (SCLK) source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:I2C:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:I2C:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I2C:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:I2C:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:I2C:CLOCk:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use as the I2C SCLK source.
  • MATH<x> specifies a math waveform to use as the I2C SCLK source.
  • REF<x> specifies a reference waveform to use as the I2C SCLK source.
  • REF<x>_D<x> specifies a digital reference waveform as the clock source waveform for the specified I2C bus.

threshold property

Return the BUS:B<x>:I2C:CLOCk:THReshold command.

Description
  • This command sets or queries the I2C Clock (SCLK) source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:I2C:CLOCk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:I2C:CLOCk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I2C:CLOCk:THReshold value command.
SCPI Syntax
- BUS:B<x>:I2C:CLOCk:THReshold <NR3>
- BUS:B<x>:I2C:CLOCk:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the I2C Clock (SCLK) source threshold for the specified bus.

BusBItemI2cClockSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:I2C:CLOCk:SOUrce command.

Description
  • This command sets or queries the I2C clock (SCLK) source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:I2C:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:I2C:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I2C:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:I2C:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:I2C:CLOCk:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use as the I2C SCLK source.
  • MATH<x> specifies a math waveform to use as the I2C SCLK source.
  • REF<x> specifies a reference waveform to use as the I2C SCLK source.
  • REF<x>_D<x> specifies a digital reference waveform as the clock source waveform for the specified I2C bus.

BusBItemI2cClockThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:I2C:CLOCk:THReshold command.

Description
  • This command sets or queries the I2C Clock (SCLK) source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:I2C:CLOCk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:I2C:CLOCk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I2C:CLOCk:THReshold value command.
SCPI Syntax
- BUS:B<x>:I2C:CLOCk:THReshold <NR3>
- BUS:B<x>:I2C:CLOCk:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the I2C Clock (SCLK) source threshold for the specified bus.

BusBItemI2cData

Bases: SCPICmdRead

The BUS:B<x>:I2C:DATa command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:I2C:DATa? query.
  • Using the .verify(value) method will send the BUS:B<x>:I2C:DATa? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the Bus number.
Properties
  • .source: The BUS:B<x>:I2C:DATa:SOUrce command.
  • .threshold: The BUS:B<x>:I2C:DATa:THReshold command.

source property

Return the BUS:B<x>:I2C:DATa:SOUrce command.

Description
  • This command sets or queries the I2C data (SDA) source for the specified I2C bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:I2C:DATa:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:I2C:DATa:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I2C:DATa:SOUrce value command.
SCPI Syntax
- BUS:B<x>:I2C:DATa:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:I2C:DATa:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel to use as the I2C SDA source.
  • MATH<x> specifies a math waveform to use as the I2C SDA source.
  • REF<x> specifies a reference waveform to use as the I2C SDA source.
  • REF<x>_D<x> specifies a digital reference waveform as the data source waveform for the specified I2C bus.

threshold property

Return the BUS:B<x>:I2C:DATa:THReshold command.

Description
  • This command sets or queries the I2C Data (SDA) source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:I2C:DATa:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:I2C:DATa:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I2C:DATa:THReshold value command.
SCPI Syntax
- BUS:B<x>:I2C:DATa:THReshold <NR3>
- BUS:B<x>:I2C:DATa:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the I2C Data (SDA) source threshold for the specified bus.

BusBItemI2cDataSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:I2C:DATa:SOUrce command.

Description
  • This command sets or queries the I2C data (SDA) source for the specified I2C bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:I2C:DATa:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:I2C:DATa:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I2C:DATa:SOUrce value command.
SCPI Syntax
- BUS:B<x>:I2C:DATa:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:I2C:DATa:SOUrce?
Info
  • B<x> is the Bus number.
  • CH<x> specifies an analog channel to use as the I2C SDA source.
  • MATH<x> specifies a math waveform to use as the I2C SDA source.
  • REF<x> specifies a reference waveform to use as the I2C SDA source.
  • REF<x>_D<x> specifies a digital reference waveform as the data source waveform for the specified I2C bus.

BusBItemI2cDataThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:I2C:DATa:THReshold command.

Description
  • This command sets or queries the I2C Data (SDA) source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:I2C:DATa:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:I2C:DATa:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I2C:DATa:THReshold value command.
SCPI Syntax
- BUS:B<x>:I2C:DATa:THReshold <NR3>
- BUS:B<x>:I2C:DATa:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the I2C Data (SDA) source threshold for the specified bus.

BusBItemI2cRwinaddr

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:I2C:RWINADDR command.

Description
  • This command sets or queries the manner in which seven-bit I2C addresses are represented in the busform display of the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:I2C:RWINADDR? query.
  • Using the .verify(value) method will send the BUS:B<x>:I2C:RWINADDR? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I2C:RWINADDR value command.
SCPI Syntax
- BUS:B<x>:I2C:RWINADDR {0|1}
- BUS:B<x>:I2C:RWINADDR?
Info
  • B<x> is the number of the bus.
  • 0 displays seven-bit slave addresses as integers in the range of 0 to 127, with the state of the R/W* bit from the LSB of the slave address byte. For example, the slave address byte of 0b10100101 is displayed as the value 0x52 R.
  • 1 displays the entire slave address byte as a number, with the R/W* signal as its LSB (bit 0) and the slave address in bits 7..1. For example, the slave address byte of 0b10100101 is displayed as the value 0xA5 R.

BusBItemI3c

Bases: SCPICmdRead

The BUS:B<x>:I3C command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:I3C? query.
  • Using the .verify(value) method will send the BUS:B<x>:I3C? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> specifies the bus number.
Properties
  • .clock: The BUS:B<x>:I3C:CLOCk command tree.
  • .data: The BUS:B<x>:I3C:DATa command tree.
  • .version: The BUS:B<x>:I3C:VERSion command.

clock property

Return the BUS:B<x>:I3C:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:I3C:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:I3C:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> specifies the bus number.
Sub-properties
  • .source: The BUS:B<x>:I3C:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:I3C:CLOCk:THReshold command.

data property

Return the BUS:B<x>:I3C:DATa command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:I3C:DATa? query.
  • Using the .verify(value) method will send the BUS:B<x>:I3C:DATa? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> specifies the bus number.
Sub-properties
  • .source: The BUS:B<x>:I3C:DATa:SOUrce command.
  • .threshold: The BUS:B<x>:I3C:DATa:THReshold command.

version property

Return the BUS:B<x>:I3C:VERSion command.

Description
  • This command sets or queries the version for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:I3C:VERSion? query.
  • Using the .verify(value) method will send the BUS:B<x>:I3C:VERSion? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I3C:VERSion value command.
SCPI Syntax
- BUS:B<x>:I3C:VERSion {VR1<x>}
- BUS:B<x>:I3C:VERSion?
Info
  • B<x> is the number of the bus.
  • VR10 sets the version to Version 1.0. This is the default version.
  • VR11 sets the version to Version 1.1.

BusBItemI3cClock

Bases: SCPICmdRead

The BUS:B<x>:I3C:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:I3C:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:I3C:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> specifies the bus number.
Properties
  • .source: The BUS:B<x>:I3C:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:I3C:CLOCk:THReshold command.

source property

Return the BUS:B<x>:I3C:CLOCk:SOUrce command.

Description
  • This command sets or queries the I3C clock (SCLK) source for the specified bus. Requires purchase and installation of option SRI3C.
Usage
  • Using the .query() method will send the BUS:B<x>:I3C:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:I3C:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I3C:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:I3C:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:I3C:CLOCk:SOUrce?
Info
  • B<x> specifies the bus number.
  • CH<x> specifies the analog channel to use as the I3C SCLK source.
  • MATH<x> specifies the math waveform to use as the I3C SCLK source.
  • REF<x> specifies the reference waveform to use as the I3C SCLK source.
  • REF<x>_D<x> specifies the digital reference waveform to use as the I3C SCLK source.

threshold property

Return the BUS:B<x>:I3C:CLOCk:THReshold command.

Description
  • This command sets or queries the I3C clock (SCLK) source threshold level for the specified bus. Requires purchase and installation of option SRI3C.
Usage
  • Using the .query() method will send the BUS:B<x>:I3C:CLOCk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:I3C:CLOCk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I3C:CLOCk:THReshold value command.
SCPI Syntax
- BUS:B<x>:I3C:CLOCk:THReshold <NR3>
- BUS:B<x>:I3C:CLOCk:THReshold?
Info
  • B<x> specifies the bus number.
  • <NR3> is the threshold value for I3C Clock (SCLK) source of the specified bus.

BusBItemI3cClockSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:I3C:CLOCk:SOUrce command.

Description
  • This command sets or queries the I3C clock (SCLK) source for the specified bus. Requires purchase and installation of option SRI3C.
Usage
  • Using the .query() method will send the BUS:B<x>:I3C:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:I3C:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I3C:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:I3C:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:I3C:CLOCk:SOUrce?
Info
  • B<x> specifies the bus number.
  • CH<x> specifies the analog channel to use as the I3C SCLK source.
  • MATH<x> specifies the math waveform to use as the I3C SCLK source.
  • REF<x> specifies the reference waveform to use as the I3C SCLK source.
  • REF<x>_D<x> specifies the digital reference waveform to use as the I3C SCLK source.

BusBItemI3cClockThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:I3C:CLOCk:THReshold command.

Description
  • This command sets or queries the I3C clock (SCLK) source threshold level for the specified bus. Requires purchase and installation of option SRI3C.
Usage
  • Using the .query() method will send the BUS:B<x>:I3C:CLOCk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:I3C:CLOCk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I3C:CLOCk:THReshold value command.
SCPI Syntax
- BUS:B<x>:I3C:CLOCk:THReshold <NR3>
- BUS:B<x>:I3C:CLOCk:THReshold?
Info
  • B<x> specifies the bus number.
  • <NR3> is the threshold value for I3C Clock (SCLK) source of the specified bus.

BusBItemI3cData

Bases: SCPICmdRead

The BUS:B<x>:I3C:DATa command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:I3C:DATa? query.
  • Using the .verify(value) method will send the BUS:B<x>:I3C:DATa? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> specifies the bus number.
Properties
  • .source: The BUS:B<x>:I3C:DATa:SOUrce command.
  • .threshold: The BUS:B<x>:I3C:DATa:THReshold command.

source property

Return the BUS:B<x>:I3C:DATa:SOUrce command.

Description
  • This command sets or queries the I3C data clock (SDA) source for the specified bus. Requires purchase and installation of option SRI3C.
Usage
  • Using the .query() method will send the BUS:B<x>:I3C:DATa:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:I3C:DATa:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I3C:DATa:SOUrce value command.
SCPI Syntax
- BUS:B<x>:I3C:DATa:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:I3C:DATa:SOUrce?
Info
  • B<x> specifies the bus number.
  • CH<x> specifies the analog channel to use as the I3C SDA source.
  • MATH<x> specifies the math waveform to use as the I3C SDA source.
  • REF<x> specifies the reference waveform to use as the I3C SDA source.
  • REF<x>_D<x> specifies the digital reference waveform to use as the I3C SDA source.

threshold property

Return the BUS:B<x>:I3C:DATa:THReshold command.

Description
  • This command sets or queries the I3C clock (SDA) data threshold level for the specified bus. Requires purchase and installation of option SRI3C.
Usage
  • Using the .query() method will send the BUS:B<x>:I3C:DATa:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:I3C:DATa:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I3C:DATa:THReshold value command.
SCPI Syntax
- BUS:B<x>:I3C:DATa:THReshold <NR3>
- BUS:B<x>:I3C:DATa:THReshold?
Info
  • B<x> specifies the bus number.
  • <NR3> is the threshold value for I3C Data Clock (SDA) source of the specified bus.

BusBItemI3cDataSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:I3C:DATa:SOUrce command.

Description
  • This command sets or queries the I3C data clock (SDA) source for the specified bus. Requires purchase and installation of option SRI3C.
Usage
  • Using the .query() method will send the BUS:B<x>:I3C:DATa:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:I3C:DATa:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I3C:DATa:SOUrce value command.
SCPI Syntax
- BUS:B<x>:I3C:DATa:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:I3C:DATa:SOUrce?
Info
  • B<x> specifies the bus number.
  • CH<x> specifies the analog channel to use as the I3C SDA source.
  • MATH<x> specifies the math waveform to use as the I3C SDA source.
  • REF<x> specifies the reference waveform to use as the I3C SDA source.
  • REF<x>_D<x> specifies the digital reference waveform to use as the I3C SDA source.

BusBItemI3cDataThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:I3C:DATa:THReshold command.

Description
  • This command sets or queries the I3C clock (SDA) data threshold level for the specified bus. Requires purchase and installation of option SRI3C.
Usage
  • Using the .query() method will send the BUS:B<x>:I3C:DATa:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:I3C:DATa:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I3C:DATa:THReshold value command.
SCPI Syntax
- BUS:B<x>:I3C:DATa:THReshold <NR3>
- BUS:B<x>:I3C:DATa:THReshold?
Info
  • B<x> specifies the bus number.
  • <NR3> is the threshold value for I3C Data Clock (SDA) source of the specified bus.

BusBItemI3cVersion

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:I3C:VERSion command.

Description
  • This command sets or queries the version for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:I3C:VERSion? query.
  • Using the .verify(value) method will send the BUS:B<x>:I3C:VERSion? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:I3C:VERSion value command.
SCPI Syntax
- BUS:B<x>:I3C:VERSion {VR1<x>}
- BUS:B<x>:I3C:VERSion?
Info
  • B<x> is the number of the bus.
  • VR10 sets the version to Version 1.0. This is the default version.
  • VR11 sets the version to Version 1.1.

BusBItemLabel

Bases: SCPICmdRead

The BUS:B<x>:LABel command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:LABel? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .color: The BUS:B<x>:LABel:COLor command.
  • .font: The BUS:B<x>:LABel:FONT command tree.
  • .xpos: The BUS:B<x>:LABel:XPOS command.
  • .ypos: The BUS:B<x>:LABel:YPOS command.
  • .name: The BUS:B<x>:LABel:name command.

color property

Return the BUS:B<x>:LABel:COLor command.

Description
  • This command sets or queries the color of the specified bus label. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:COLor? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:COLor? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:COLor value command.
SCPI Syntax
- BUS:B<x>:LABel:COLor <QString>
- BUS:B<x>:LABel:COLor?
Info
  • B<x> is the number of the bus.
  • <QString> is the bus label color. To return the color to the default color, send an empty string as in this example: :BUS:B1:LABEL:COLOR ‘’.

font property

Return the BUS:B<x>:LABel:FONT command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:LABel:FONT? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:FONT? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .bold: The BUS:B<x>:LABel:FONT:BOLD command.
  • .italic: The BUS:B<x>:LABel:FONT:ITALic command.
  • .size: The BUS:B<x>:LABel:FONT:SIZE command.
  • .type: The BUS:B<x>:LABel:FONT:TYPE command.
  • .underline: The BUS:B<x>:LABel:FONT:UNDERline command.

name property

Return the BUS:B<x>:LABel:name command.

Description
  • This command sets or queries the label for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:name? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:name? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:name value command.
SCPI Syntax
- BUS:B<x>:LABel:name <QString>
- BUS:B<x>:LABel:name?
Info
  • B<x> is the number of the bus.
  • <QString> is an alphanumeric string of text enclosed in quotes. The text string is limited to 30 characters. It contains the text label information for the bus.

xpos property

Return the BUS:B<x>:LABel:XPOS command.

Description
  • This command sets or queries the x-position of the specified bus label. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:XPOS? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:XPOS? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:XPOS value command.
SCPI Syntax
- BUS:B<x>:LABel:XPOS <NR3>
- BUS:B<x>:LABel:XPOS?
Info
  • B<x> is the number of the bus.
  • <NR3> is the x-position, in pixels relative to the left edge of the screen of the specified bus label.

ypos property

Return the BUS:B<x>:LABel:YPOS command.

Description
  • This command sets or queries the y-position of the specified bus label. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:YPOS? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:YPOS? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:YPOS value command.
SCPI Syntax
- BUS:B<x>:LABel:YPOS <NR3>
- BUS:B<x>:LABel:YPOS?
Info
  • B<x> is the number of the bus.
  • <NR3> is the y-position, in pixels relative to the left edge of the screen of the specified bus label.

BusBItemLabelColor

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:LABel:COLor command.

Description
  • This command sets or queries the color of the specified bus label. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:COLor? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:COLor? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:COLor value command.
SCPI Syntax
- BUS:B<x>:LABel:COLor <QString>
- BUS:B<x>:LABel:COLor?
Info
  • B<x> is the number of the bus.
  • <QString> is the bus label color. To return the color to the default color, send an empty string as in this example: :BUS:B1:LABEL:COLOR ‘’.

BusBItemLabelFont

Bases: SCPICmdRead

The BUS:B<x>:LABel:FONT command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:LABel:FONT? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:FONT? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .bold: The BUS:B<x>:LABel:FONT:BOLD command.
  • .italic: The BUS:B<x>:LABel:FONT:ITALic command.
  • .size: The BUS:B<x>:LABel:FONT:SIZE command.
  • .type: The BUS:B<x>:LABel:FONT:TYPE command.
  • .underline: The BUS:B<x>:LABel:FONT:UNDERline command.

bold property

Return the BUS:B<x>:LABel:FONT:BOLD command.

Description
  • This command sets or queries the bold state of the specified bus label. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:FONT:BOLD? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:FONT:BOLD? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:FONT:BOLD value command.
SCPI Syntax
- BUS:B<x>:LABel:FONT:BOLD {ON|OFF|1|0}
- BUS:B<x>:LABel:FONT:BOLD?
Info
  • B<x> is the number of the bus.
  • ON displays the label in bold font.
  • OFF does not display the label in bold font.
  • 1 displays the label in bold font.
  • 0 does not display the label in bold font.

italic property

Return the BUS:B<x>:LABel:FONT:ITALic command.

Description
  • This command sets or queries the italic state of the specified bus label. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:FONT:ITALic? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:FONT:ITALic? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:FONT:ITALic value command.
SCPI Syntax
- BUS:B<x>:LABel:FONT:ITALic {ON|OFF|1|0}
- BUS:B<x>:LABel:FONT:ITALic?
Info
  • B<x> is the number of the bus.
  • ON displays the label in italic font.
  • OFF does not display the label in italic font.
  • 1 displays the label in italic font.
  • 0 does not display the label in italic font.

size property

Return the BUS:B<x>:LABel:FONT:SIZE command.

Description
  • This command sets or queries the font size of the specified bus label. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:FONT:SIZE? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:FONT:SIZE? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:FONT:SIZE value command.
SCPI Syntax
- BUS:B<x>:LABel:FONT:SIZE <NR1>
- BUS:B<x>:LABel:FONT:SIZE?
Info
  • B<x> is the number of the bus.
  • <NR1> is the font size.

type property

Return the BUS:B<x>:LABel:FONT:TYPE command.

Description
  • This command sets or queries the font type of the specified bus label, such as Arial or Times New Roman. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:FONT:TYPE? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:FONT:TYPE? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:FONT:TYPE value command.
SCPI Syntax
- BUS:B<x>:LABel:FONT:TYPE <QString>
- BUS:B<x>:LABel:FONT:TYPE?
Info
  • B<x> is the number of the bus.
  • <QString> is the specified font type. Available fonts include: DejaVu Sans, DejaVu Sans Mono, DejaVu Serif, Frutiger LT Std, Monospace, Sans Serif, Serif, Ubuntu, Ubuntu Condensed, and Ubuntu Mono.

underline property

Return the BUS:B<x>:LABel:FONT:UNDERline command.

Description
  • This command sets or queries the underline state of the specified bus label. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:FONT:UNDERline? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:FONT:UNDERline? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:FONT:UNDERline value command.
SCPI Syntax
- BUS:B<x>:LABel:FONT:UNDERline {ON|OFF|1|0}
- BUS:B<x>:LABel:FONT:UNDERline?
Info
  • B<x> is the number of the bus.
  • ON displays the label in underlined font.
  • OFF does not display the label in underlined font.
  • 1 displays the label in underlined font.
  • 0 does not display the label in underlined font.

BusBItemLabelFontBold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:LABel:FONT:BOLD command.

Description
  • This command sets or queries the bold state of the specified bus label. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:FONT:BOLD? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:FONT:BOLD? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:FONT:BOLD value command.
SCPI Syntax
- BUS:B<x>:LABel:FONT:BOLD {ON|OFF|1|0}
- BUS:B<x>:LABel:FONT:BOLD?
Info
  • B<x> is the number of the bus.
  • ON displays the label in bold font.
  • OFF does not display the label in bold font.
  • 1 displays the label in bold font.
  • 0 does not display the label in bold font.

BusBItemLabelFontItalic

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:LABel:FONT:ITALic command.

Description
  • This command sets or queries the italic state of the specified bus label. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:FONT:ITALic? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:FONT:ITALic? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:FONT:ITALic value command.
SCPI Syntax
- BUS:B<x>:LABel:FONT:ITALic {ON|OFF|1|0}
- BUS:B<x>:LABel:FONT:ITALic?
Info
  • B<x> is the number of the bus.
  • ON displays the label in italic font.
  • OFF does not display the label in italic font.
  • 1 displays the label in italic font.
  • 0 does not display the label in italic font.

BusBItemLabelFontSize

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:LABel:FONT:SIZE command.

Description
  • This command sets or queries the font size of the specified bus label. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:FONT:SIZE? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:FONT:SIZE? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:FONT:SIZE value command.
SCPI Syntax
- BUS:B<x>:LABel:FONT:SIZE <NR1>
- BUS:B<x>:LABel:FONT:SIZE?
Info
  • B<x> is the number of the bus.
  • <NR1> is the font size.

BusBItemLabelFontType

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:LABel:FONT:TYPE command.

Description
  • This command sets or queries the font type of the specified bus label, such as Arial or Times New Roman. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:FONT:TYPE? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:FONT:TYPE? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:FONT:TYPE value command.
SCPI Syntax
- BUS:B<x>:LABel:FONT:TYPE <QString>
- BUS:B<x>:LABel:FONT:TYPE?
Info
  • B<x> is the number of the bus.
  • <QString> is the specified font type. Available fonts include: DejaVu Sans, DejaVu Sans Mono, DejaVu Serif, Frutiger LT Std, Monospace, Sans Serif, Serif, Ubuntu, Ubuntu Condensed, and Ubuntu Mono.

BusBItemLabelFontUnderline

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:LABel:FONT:UNDERline command.

Description
  • This command sets or queries the underline state of the specified bus label. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:FONT:UNDERline? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:FONT:UNDERline? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:FONT:UNDERline value command.
SCPI Syntax
- BUS:B<x>:LABel:FONT:UNDERline {ON|OFF|1|0}
- BUS:B<x>:LABel:FONT:UNDERline?
Info
  • B<x> is the number of the bus.
  • ON displays the label in underlined font.
  • OFF does not display the label in underlined font.
  • 1 displays the label in underlined font.
  • 0 does not display the label in underlined font.

BusBItemLabelName

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:LABel:name command.

Description
  • This command sets or queries the label for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:name? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:name? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:name value command.
SCPI Syntax
- BUS:B<x>:LABel:name <QString>
- BUS:B<x>:LABel:name?
Info
  • B<x> is the number of the bus.
  • <QString> is an alphanumeric string of text enclosed in quotes. The text string is limited to 30 characters. It contains the text label information for the bus.

BusBItemLabelXpos

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:LABel:XPOS command.

Description
  • This command sets or queries the x-position of the specified bus label. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:XPOS? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:XPOS? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:XPOS value command.
SCPI Syntax
- BUS:B<x>:LABel:XPOS <NR3>
- BUS:B<x>:LABel:XPOS?
Info
  • B<x> is the number of the bus.
  • <NR3> is the x-position, in pixels relative to the left edge of the screen of the specified bus label.

BusBItemLabelYpos

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:LABel:YPOS command.

Description
  • This command sets or queries the y-position of the specified bus label. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LABel:YPOS? query.
  • Using the .verify(value) method will send the BUS:B<x>:LABel:YPOS? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LABel:YPOS value command.
SCPI Syntax
- BUS:B<x>:LABel:YPOS <NR3>
- BUS:B<x>:LABel:YPOS?
Info
  • B<x> is the number of the bus.
  • <NR3> is the y-position, in pixels relative to the left edge of the screen of the specified bus label.

BusBItemLin

Bases: SCPICmdRead

The BUS:B<x>:LIN command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:LIN? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .bitrate: The BUS:B<x>:LIN:BITRate command.
  • .idformat: The BUS:B<x>:LIN:IDFORmat command.
  • .polarity: The BUS:B<x>:LIN:POLarity command.
  • .samplepoint: The BUS:B<x>:LIN:SAMPLEpoint command.
  • .source: The BUS:B<x>:LIN:SOUrce command.
  • .standard: The BUS:B<x>:LIN:STANDard command.

bitrate property

Return the BUS:B<x>:LIN:BITRate command.

Description
  • This command sets or queries the LIN bus bit rate. The bus number is specified by x. If you select Custom, use BUS:B<x>:LIN:BITRate:CUSTom to set the bit rate.
Usage
  • Using the .query() method will send the BUS:B<x>:LIN:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LIN:BITRate value command.
SCPI Syntax
- BUS:B<x>:LIN:BITRate {RATE10K|RATE1K|RATE19K|RATE2K|RATE4K|RATE9K|CUSTom}
- BUS:B<x>:LIN:BITRate?
Info
  • B<x> is the number of the bus.
  • RATE10K sets the bit rate to 10 kb.
  • RATE1K sets the bit rate to 1 kb.
  • RATE19K sets the bit rate to 19 kb.
  • RATE2K sets the bit rate to 2 kb.
  • RATE4K sets the bit rate to 4 kb.
  • RATE9K sets the bit rate to 9 kb.
  • CUSTom sets a custom bit rate.
Sub-properties
  • .custom: The BUS:B<x>:LIN:BITRate:CUSTom command.

idformat property

Return the BUS:B<x>:LIN:IDFORmat command.

Description
  • This command sets or queries LIN bus identifier format for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LIN:IDFORmat? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN:IDFORmat? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LIN:IDFORmat value command.
SCPI Syntax
- BUS:B<x>:LIN:IDFORmat {NOPARity|PARity}
- BUS:B<x>:LIN:IDFORmat?
Info
  • B<x> is the number of the bus.
  • NOPARity specifies an id format that includes parity.
  • PARity specifies an id format that separates parity.

polarity property

Return the BUS:B<x>:LIN:POLarity command.

Description
  • This command sets or queries the LIN source polarity for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LIN:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LIN:POLarity value command.
SCPI Syntax
- BUS:B<x>:LIN:POLarity {INVerted|NORmal}
- BUS:B<x>:LIN:POLarity?
Info
  • B<x> is the number of the bus.
  • INVerted specifies inverted polarity.
  • NORmal specifies normal polarity.

samplepoint property

Return the BUS:B<x>:LIN:SAMPLEpoint command.

Description
  • Specifies the LIN sample point, for the specified LIN bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LIN:SAMPLEpoint? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN:SAMPLEpoint? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LIN:SAMPLEpoint value command.
SCPI Syntax
- BUS:B<x>:LIN:SAMPLEpoint <NR1>
- BUS:B<x>:LIN:SAMPLEpoint?
Info
  • B<x> is the number of the bus.
  • <NR1> is a percentage that represents the point at which to sample during each bit period.

source property

Return the BUS:B<x>:LIN:SOUrce command.

Description
  • This command sets or queries the LIN data source for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LIN:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LIN:SOUrce value command.
SCPI Syntax
- BUS:B<x>:LIN:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:LIN:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.
  • REF<x>_D<x> specifies a digital reference waveform as the source.
Sub-properties
  • .threshold: The BUS:B<x>:LIN:SOUrce:THReshold command.

standard property

Return the BUS:B<x>:LIN:STANDard command.

Description
  • This command sets or queries the LIN bus standard for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LIN:STANDard? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN:STANDard? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LIN:STANDard value command.
SCPI Syntax
- BUS:B<x>:LIN:STANDard {MIXed|V1X|V2X}
- BUS:B<x>:LIN:STANDard?
Info
  • B<x> is the number of the bus.
  • MIXed specifies both versions 1.x and 2.x of the LIN standard.
  • V1X specifies version 1.x of the LIN standard.
  • V2X specifies version 2.x of the LIN standard.

BusBItemLinBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:LIN:BITRate command.

Description
  • This command sets or queries the LIN bus bit rate. The bus number is specified by x. If you select Custom, use BUS:B<x>:LIN:BITRate:CUSTom to set the bit rate.
Usage
  • Using the .query() method will send the BUS:B<x>:LIN:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LIN:BITRate value command.
SCPI Syntax
- BUS:B<x>:LIN:BITRate {RATE10K|RATE1K|RATE19K|RATE2K|RATE4K|RATE9K|CUSTom}
- BUS:B<x>:LIN:BITRate?
Info
  • B<x> is the number of the bus.
  • RATE10K sets the bit rate to 10 kb.
  • RATE1K sets the bit rate to 1 kb.
  • RATE19K sets the bit rate to 19 kb.
  • RATE2K sets the bit rate to 2 kb.
  • RATE4K sets the bit rate to 4 kb.
  • RATE9K sets the bit rate to 9 kb.
  • CUSTom sets a custom bit rate.
Properties
  • .custom: The BUS:B<x>:LIN:BITRate:CUSTom command.

custom property

Return the BUS:B<x>:LIN:BITRate:CUSTom command.

Description
  • This command sets or queries LIN custom bit rate for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LIN:BITRate:CUSTom? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN:BITRate:CUSTom? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LIN:BITRate:CUSTom value command.
SCPI Syntax
- BUS:B<x>:LIN:BITRate:CUSTom <NR1>
- BUS:B<x>:LIN:BITRate:CUSTom?
Info
  • B<x> is the number of the bus.
  • <NR1> is the LIN custom bit rate for the specified bus.

BusBItemLinBitrateCustom

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:LIN:BITRate:CUSTom command.

Description
  • This command sets or queries LIN custom bit rate for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LIN:BITRate:CUSTom? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN:BITRate:CUSTom? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LIN:BITRate:CUSTom value command.
SCPI Syntax
- BUS:B<x>:LIN:BITRate:CUSTom <NR1>
- BUS:B<x>:LIN:BITRate:CUSTom?
Info
  • B<x> is the number of the bus.
  • <NR1> is the LIN custom bit rate for the specified bus.

BusBItemLinIdformat

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:LIN:IDFORmat command.

Description
  • This command sets or queries LIN bus identifier format for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LIN:IDFORmat? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN:IDFORmat? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LIN:IDFORmat value command.
SCPI Syntax
- BUS:B<x>:LIN:IDFORmat {NOPARity|PARity}
- BUS:B<x>:LIN:IDFORmat?
Info
  • B<x> is the number of the bus.
  • NOPARity specifies an id format that includes parity.
  • PARity specifies an id format that separates parity.

BusBItemLinPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:LIN:POLarity command.

Description
  • This command sets or queries the LIN source polarity for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LIN:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LIN:POLarity value command.
SCPI Syntax
- BUS:B<x>:LIN:POLarity {INVerted|NORmal}
- BUS:B<x>:LIN:POLarity?
Info
  • B<x> is the number of the bus.
  • INVerted specifies inverted polarity.
  • NORmal specifies normal polarity.

BusBItemLinSamplepoint

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:LIN:SAMPLEpoint command.

Description
  • Specifies the LIN sample point, for the specified LIN bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LIN:SAMPLEpoint? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN:SAMPLEpoint? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LIN:SAMPLEpoint value command.
SCPI Syntax
- BUS:B<x>:LIN:SAMPLEpoint <NR1>
- BUS:B<x>:LIN:SAMPLEpoint?
Info
  • B<x> is the number of the bus.
  • <NR1> is a percentage that represents the point at which to sample during each bit period.

BusBItemLinSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:LIN:SOUrce command.

Description
  • This command sets or queries the LIN data source for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LIN:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LIN:SOUrce value command.
SCPI Syntax
- BUS:B<x>:LIN:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:LIN:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use source.
  • MATH<x> specifies a math waveform to use as the source.
  • REF<x> specifies a reference waveform to use as the source.
  • REF<x>_D<x> specifies a digital reference waveform as the source.
Properties
  • .threshold: The BUS:B<x>:LIN:SOUrce:THReshold command.

threshold property

Return the BUS:B<x>:LIN:SOUrce:THReshold command.

Description
  • This command sets or queries the LIN source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LIN:SOUrce:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN:SOUrce:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LIN:SOUrce:THReshold value command.
SCPI Syntax
- BUS:B<x>:LIN:SOUrce:THReshold <NR3>
- BUS:B<x>:LIN:SOUrce:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> the LIN source threshold for the specified bus.

BusBItemLinSourceThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:LIN:SOUrce:THReshold command.

Description
  • This command sets or queries the LIN source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LIN:SOUrce:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN:SOUrce:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LIN:SOUrce:THReshold value command.
SCPI Syntax
- BUS:B<x>:LIN:SOUrce:THReshold <NR3>
- BUS:B<x>:LIN:SOUrce:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> the LIN source threshold for the specified bus.

BusBItemLinStandard

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:LIN:STANDard command.

Description
  • This command sets or queries the LIN bus standard for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:LIN:STANDard? query.
  • Using the .verify(value) method will send the BUS:B<x>:LIN:STANDard? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:LIN:STANDard value command.
SCPI Syntax
- BUS:B<x>:LIN:STANDard {MIXed|V1X|V2X}
- BUS:B<x>:LIN:STANDard?
Info
  • B<x> is the number of the bus.
  • MIXed specifies both versions 1.x and 2.x of the LIN standard.
  • V1X specifies version 1.x of the LIN standard.
  • V2X specifies version 2.x of the LIN standard.

BusBItemManchester

Bases: SCPICmdRead

The BUS:B<x>:MANChester command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MANChester? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .bitorder: The BUS:B<x>:MANChester:BITORDer command.
  • .bitrate: The BUS:B<x>:MANChester:BITRate command.
  • .displaymode: The BUS:B<x>:MANChester:DISplaymode command.
  • .header: The BUS:B<x>:MANChester:HEADer command tree.
  • .idle: The BUS:B<x>:MANChester:IDLE command tree.
  • .source: The BUS:B<x>:MANChester:SOUrce command.
  • .start: The BUS:B<x>:MANChester:START command tree.
  • .sync: The BUS:B<x>:MANChester:SYNC command tree.
  • .threshold: The BUS:B<x>:MANChester:THReshold command.
  • .tolerance: The BUS:B<x>:MANChester:TOLerance command.
  • .transtion: The BUS:B<x>:MANChester:TRANstion command tree.
  • .trailer: The BUS:B<x>:MANChester:TRAiler command tree.
  • .word: The BUS:B<x>:MANChester:WORD command tree.
  • .wordsize: The BUS:B<x>:MANChester:WORDSIZe command.
  • .parity: The BUS:B<x>:MANChester:parity command.

bitorder property

Return the BUS:B<x>:MANChester:BITORDer command.

Description
  • This command sets or queries the Manchester bus Bit Order. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:BITORDer? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:BITORDer? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:BITORDer value command.
SCPI Syntax
- BUS:B<x>:MANChester:BITORDer {LSB|MSB}
- BUS:B<x>:MANChester:BITORDer?
Info
  • B<x> is the number of the bus.
  • LSB arranges the bits in LSB format, least significant bits first.
  • MSB arranges the bits in MSB format, most significant bits first.

bitrate property

Return the BUS:B<x>:MANChester:BITRate command.

Description
  • This command sets or queries the Manchester bus bit rate. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:BITRate value command.
SCPI Syntax
- BUS:B<x>:MANChester:BITRate <NR1>
- BUS:B<x>:MANChester:BITRate?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the bit rate up to 1 Gbps.

displaymode property

Return the BUS:B<x>:MANChester:DISplaymode command.

Description
  • This command sets or queries the Manchester bus Packet View. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:DISplaymode? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:DISplaymode? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:DISplaymode value command.
SCPI Syntax
- BUS:B<x>:MANChester:DISplaymode {BITS|PACKET}
- BUS:B<x>:MANChester:DISplaymode?
Info
  • B<x> is the number of the bus.
  • BITS sets the PacketView to off and the data to be seen as single bits formats.
  • PACKET sets the PacketView to on and the data to be seen in the form of fields.

header property

Return the BUS:B<x>:MANChester:HEADer command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:HEADer? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:HEADer? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .length: The BUS:B<x>:MANChester:HEADer:LENGth command.

idle property

Return the BUS:B<x>:MANChester:IDLE command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:IDLE? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:IDLE? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .bits: The BUS:B<x>:MANChester:IDLE:BITS command.

parity property

Return the BUS:B<x>:MANChester:parity command.

Description
  • This command sets or queries the Manchester bus Parity. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:parity? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:parity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:parity value command.
SCPI Syntax
- BUS:B<x>:MANChester:parity {ODD|EVEN|NONE}
- BUS:B<x>:MANChester:parity?
Info
  • ODD sets the number of 1’s to odd.
  • EVEN sets the number of 1’s to even.
  • NONE specifies that Parity is not considered.

source property

Return the BUS:B<x>:MANChester:SOUrce command.

Description
  • This command sets or queries the Manchester source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:SOUrce value command.
SCPI Syntax
- BUS:B<x>:MANChester:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:MANChester:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a reference waveform as the source.

start property

Return the BUS:B<x>:MANChester:START command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:START? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:START? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .index: The BUS:B<x>:MANChester:START:INDex command.

sync property

Return the BUS:B<x>:MANChester:SYNC command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:SYNC? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:SYNC? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .size: The BUS:B<x>:MANChester:SYNC:SIZe command.

threshold property

Return the BUS:B<x>:MANChester:THReshold command.

Description
  • This command sets or queries the Manchester threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:THReshold value command.
SCPI Syntax
- BUS:B<x>:MANChester:THReshold <NR3>
- BUS:B<x>:MANChester:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> sets the Manchester Strobe threshold for the specified bus in Volts. The argument range is -8V to +8V.

tolerance property

Return the BUS:B<x>:MANChester:TOLerance command.

Description
  • This command sets or queries the Tolerance bus parameter. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:TOLerance? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:TOLerance? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:TOLerance value command.
SCPI Syntax
- BUS:B<x>:MANChester:TOLerance <NR3>
- BUS:B<x>:MANChester:TOLerance?
Info
  • B<x> is the number of the bus.
  • <NR3> sets the Tolerance bus parameter.

trailer property

Return the BUS:B<x>:MANChester:TRAiler command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:TRAiler? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:TRAiler? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .length: The BUS:B<x>:MANChester:TRAiler:LENGth command.

transtion property

Return the BUS:B<x>:MANChester:TRANstion command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:TRANstion? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:TRANstion? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .zero: The BUS:B<x>:MANChester:TRANstion:ZERo command.

word property

Return the BUS:B<x>:MANChester:WORD command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:WORD? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:WORD? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .count: The BUS:B<x>:MANChester:WORD:COUNt command.

wordsize property

Return the BUS:B<x>:MANChester:WORDSIZe command.

Description
  • This command sets or queries the Manchester word size in bits. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:WORDSIZe? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:WORDSIZe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:WORDSIZe value command.
SCPI Syntax
- BUS:B<x>:MANChester:WORDSIZe <NR1>
- BUS:B<x>:MANChester:WORDSIZe?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the word size in bits.

BusBItemManchesterBitorder

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MANChester:BITORDer command.

Description
  • This command sets or queries the Manchester bus Bit Order. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:BITORDer? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:BITORDer? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:BITORDer value command.
SCPI Syntax
- BUS:B<x>:MANChester:BITORDer {LSB|MSB}
- BUS:B<x>:MANChester:BITORDer?
Info
  • B<x> is the number of the bus.
  • LSB arranges the bits in LSB format, least significant bits first.
  • MSB arranges the bits in MSB format, most significant bits first.

BusBItemManchesterBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MANChester:BITRate command.

Description
  • This command sets or queries the Manchester bus bit rate. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:BITRate value command.
SCPI Syntax
- BUS:B<x>:MANChester:BITRate <NR1>
- BUS:B<x>:MANChester:BITRate?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the bit rate up to 1 Gbps.

BusBItemManchesterDisplaymode

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MANChester:DISplaymode command.

Description
  • This command sets or queries the Manchester bus Packet View. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:DISplaymode? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:DISplaymode? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:DISplaymode value command.
SCPI Syntax
- BUS:B<x>:MANChester:DISplaymode {BITS|PACKET}
- BUS:B<x>:MANChester:DISplaymode?
Info
  • B<x> is the number of the bus.
  • BITS sets the PacketView to off and the data to be seen as single bits formats.
  • PACKET sets the PacketView to on and the data to be seen in the form of fields.

BusBItemManchesterHeader

Bases: SCPICmdRead

The BUS:B<x>:MANChester:HEADer command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:HEADer? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:HEADer? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .length: The BUS:B<x>:MANChester:HEADer:LENGth command.

length property

Return the BUS:B<x>:MANChester:HEADer:LENGth command.

Description
  • This command sets or queries the Manchester header length in bits. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:HEADer:LENGth? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:HEADer:LENGth? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:HEADer:LENGth value command.
SCPI Syntax
- BUS:B<x>:MANChester:HEADer:LENGth <NR1>
- BUS:B<x>:MANChester:HEADer:LENGth?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the header length in bits.

BusBItemManchesterHeaderLength

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MANChester:HEADer:LENGth command.

Description
  • This command sets or queries the Manchester header length in bits. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:HEADer:LENGth? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:HEADer:LENGth? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:HEADer:LENGth value command.
SCPI Syntax
- BUS:B<x>:MANChester:HEADer:LENGth <NR1>
- BUS:B<x>:MANChester:HEADer:LENGth?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the header length in bits.

BusBItemManchesterIdle

Bases: SCPICmdRead

The BUS:B<x>:MANChester:IDLE command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:IDLE? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:IDLE? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .bits: The BUS:B<x>:MANChester:IDLE:BITS command.

bits property

Return the BUS:B<x>:MANChester:IDLE:BITS command.

Description
  • This command sets or queries the Manchester idle bit size in bits. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:IDLE:BITS? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:IDLE:BITS? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:IDLE:BITS value command.
SCPI Syntax
- BUS:B<x>:MANChester:IDLE:BITS <NR1>
- BUS:B<x>:MANChester:IDLE:BITS?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the idle bit size.

BusBItemManchesterIdleBits

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MANChester:IDLE:BITS command.

Description
  • This command sets or queries the Manchester idle bit size in bits. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:IDLE:BITS? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:IDLE:BITS? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:IDLE:BITS value command.
SCPI Syntax
- BUS:B<x>:MANChester:IDLE:BITS <NR1>
- BUS:B<x>:MANChester:IDLE:BITS?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the idle bit size.

BusBItemManchesterParity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MANChester:parity command.

Description
  • This command sets or queries the Manchester bus Parity. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:parity? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:parity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:parity value command.
SCPI Syntax
- BUS:B<x>:MANChester:parity {ODD|EVEN|NONE}
- BUS:B<x>:MANChester:parity?
Info
  • ODD sets the number of 1’s to odd.
  • EVEN sets the number of 1’s to even.
  • NONE specifies that Parity is not considered.

BusBItemManchesterSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MANChester:SOUrce command.

Description
  • This command sets or queries the Manchester source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:SOUrce value command.
SCPI Syntax
- BUS:B<x>:MANChester:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:MANChester:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a reference waveform as the source.

BusBItemManchesterStart

Bases: SCPICmdRead

The BUS:B<x>:MANChester:START command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:START? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:START? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .index: The BUS:B<x>:MANChester:START:INDex command.

index property

Return the BUS:B<x>:MANChester:START:INDex command.

Description
  • This command sets or queries the Manchester start Index in bits. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:START:INDex? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:START:INDex? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:START:INDex value command.
SCPI Syntax
- BUS:B<x>:MANChester:START:INDex <NR1>
- BUS:B<x>:MANChester:START:INDex?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the start Index in bits.

BusBItemManchesterStartIndex

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MANChester:START:INDex command.

Description
  • This command sets or queries the Manchester start Index in bits. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:START:INDex? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:START:INDex? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:START:INDex value command.
SCPI Syntax
- BUS:B<x>:MANChester:START:INDex <NR1>
- BUS:B<x>:MANChester:START:INDex?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the start Index in bits.

BusBItemManchesterSync

Bases: SCPICmdRead

The BUS:B<x>:MANChester:SYNC command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:SYNC? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:SYNC? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .size: The BUS:B<x>:MANChester:SYNC:SIZe command.

size property

Return the BUS:B<x>:MANChester:SYNC:SIZe command.

Description
  • This command sets or queries the Manchester sync Bit Size in bits. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:SYNC:SIZe? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:SYNC:SIZe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:SYNC:SIZe value command.
SCPI Syntax
- BUS:B<x>:MANChester:SYNC:SIZe <NR1>
- BUS:B<x>:MANChester:SYNC:SIZe?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the Bit Size in bits.

BusBItemManchesterSyncSize

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MANChester:SYNC:SIZe command.

Description
  • This command sets or queries the Manchester sync Bit Size in bits. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:SYNC:SIZe? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:SYNC:SIZe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:SYNC:SIZe value command.
SCPI Syntax
- BUS:B<x>:MANChester:SYNC:SIZe <NR1>
- BUS:B<x>:MANChester:SYNC:SIZe?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the Bit Size in bits.

BusBItemManchesterThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MANChester:THReshold command.

Description
  • This command sets or queries the Manchester threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:THReshold value command.
SCPI Syntax
- BUS:B<x>:MANChester:THReshold <NR3>
- BUS:B<x>:MANChester:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> sets the Manchester Strobe threshold for the specified bus in Volts. The argument range is -8V to +8V.

BusBItemManchesterTolerance

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MANChester:TOLerance command.

Description
  • This command sets or queries the Tolerance bus parameter. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:TOLerance? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:TOLerance? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:TOLerance value command.
SCPI Syntax
- BUS:B<x>:MANChester:TOLerance <NR3>
- BUS:B<x>:MANChester:TOLerance?
Info
  • B<x> is the number of the bus.
  • <NR3> sets the Tolerance bus parameter.

BusBItemManchesterTrailer

Bases: SCPICmdRead

The BUS:B<x>:MANChester:TRAiler command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:TRAiler? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:TRAiler? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .length: The BUS:B<x>:MANChester:TRAiler:LENGth command.

length property

Return the BUS:B<x>:MANChester:TRAiler:LENGth command.

Description
  • This command sets or queries the Manchester trailer length in bits. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:TRAiler:LENGth? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:TRAiler:LENGth? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:TRAiler:LENGth value command.
SCPI Syntax
- BUS:B<x>:MANChester:TRAiler:LENGth <NR1>
- BUS:B<x>:MANChester:TRAiler:LENGth?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the trailer length in bits.

BusBItemManchesterTrailerLength

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MANChester:TRAiler:LENGth command.

Description
  • This command sets or queries the Manchester trailer length in bits. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:TRAiler:LENGth? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:TRAiler:LENGth? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:TRAiler:LENGth value command.
SCPI Syntax
- BUS:B<x>:MANChester:TRAiler:LENGth <NR1>
- BUS:B<x>:MANChester:TRAiler:LENGth?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the trailer length in bits.

BusBItemManchesterTranstion

Bases: SCPICmdRead

The BUS:B<x>:MANChester:TRANstion command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:TRANstion? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:TRANstion? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .zero: The BUS:B<x>:MANChester:TRANstion:ZERo command.

zero property

Return the BUS:B<x>:MANChester:TRANstion:ZERo command.

Description
  • This command sets or queries the Manchester bus for zero falling or rising. Manchester bit are defined by transition in the middle of the bit. Depending on the transition, conventions are defined. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:TRANstion:ZERo? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:TRANstion:ZERo? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:TRANstion:ZERo value command.
SCPI Syntax
- BUS:B<x>:MANChester:TRANstion:ZERo {FALLing|RISing}
- BUS:B<x>:MANChester:TRANstion:ZERo?
Info
  • B<x> is the number of the bus.
  • FALLing sets Falling as 0.
  • RISing sets Rising as 0.

BusBItemManchesterTranstionZero

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MANChester:TRANstion:ZERo command.

Description
  • This command sets or queries the Manchester bus for zero falling or rising. Manchester bit are defined by transition in the middle of the bit. Depending on the transition, conventions are defined. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:TRANstion:ZERo? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:TRANstion:ZERo? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:TRANstion:ZERo value command.
SCPI Syntax
- BUS:B<x>:MANChester:TRANstion:ZERo {FALLing|RISing}
- BUS:B<x>:MANChester:TRANstion:ZERo?
Info
  • B<x> is the number of the bus.
  • FALLing sets Falling as 0.
  • RISing sets Rising as 0.

BusBItemManchesterWord

Bases: SCPICmdRead

The BUS:B<x>:MANChester:WORD command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:WORD? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:WORD? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .count: The BUS:B<x>:MANChester:WORD:COUNt command.

count property

Return the BUS:B<x>:MANChester:WORD:COUNt command.

Description
  • This command sets or queries the Manchester word count in bits. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:WORD:COUNt? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:WORD:COUNt? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:WORD:COUNt value command.
SCPI Syntax
- BUS:B<x>:MANChester:WORD:COUNt <NR1>
- BUS:B<x>:MANChester:WORD:COUNt?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the word count in bits.

BusBItemManchesterWordCount

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MANChester:WORD:COUNt command.

Description
  • This command sets or queries the Manchester word count in bits. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:WORD:COUNt? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:WORD:COUNt? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:WORD:COUNt value command.
SCPI Syntax
- BUS:B<x>:MANChester:WORD:COUNt <NR1>
- BUS:B<x>:MANChester:WORD:COUNt?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the word count in bits.

BusBItemManchesterWordsize

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MANChester:WORDSIZe command.

Description
  • This command sets or queries the Manchester word size in bits. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MANChester:WORDSIZe? query.
  • Using the .verify(value) method will send the BUS:B<x>:MANChester:WORDSIZe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MANChester:WORDSIZe value command.
SCPI Syntax
- BUS:B<x>:MANChester:WORDSIZe <NR1>
- BUS:B<x>:MANChester:WORDSIZe?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the word size in bits.

BusBItemMdio

Bases: SCPICmdRead

The BUS:B<x>:MDIO command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MDIO? query.
  • Using the .verify(value) method will send the BUS:B<x>:MDIO? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .clock: The BUS:B<x>:MDIO:CLOCk command tree.
  • .data: The BUS:B<x>:MDIO:DATA command tree.

clock property

Return the BUS:B<x>:MDIO:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MDIO:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:MDIO:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .source: The BUS:B<x>:MDIO:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:MDIO:CLOCk:THReshold command.

data property

Return the BUS:B<x>:MDIO:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MDIO:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:MDIO:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .source: The BUS:B<x>:MDIO:DATA:SOUrce command.
  • .threshold: The BUS:B<x>:MDIO:DATA:THReshold command.

BusBItemMdioClock

Bases: SCPICmdRead

The BUS:B<x>:MDIO:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MDIO:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:MDIO:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .source: The BUS:B<x>:MDIO:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:MDIO:CLOCk:THReshold command.

source property

Return the BUS:B<x>:MDIO:CLOCk:SOUrce command.

Description
  • This command sets or queries the MDIO Clock source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MDIO:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:MDIO:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MDIO:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:MDIO:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:MDIO:CLOCk:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the Clock source.
  • MATH<x> specifies a math waveform as the Clock source.
  • REF<x> specifies a reference waveform as the Clock source.

threshold property

Return the BUS:B<x>:MDIO:CLOCk:THReshold command.

Description
  • This command sets or queries the MDIO clock source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MDIO:CLOCk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:MDIO:CLOCk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MDIO:CLOCk:THReshold value command.
SCPI Syntax
- BUS:B<x>:MDIO:CLOCk:THReshold <NR3>
- BUS:B<x>:MDIO:CLOCk:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the clock source threshold for the specified bus. The argument range is -8V to +8V.

BusBItemMdioClockSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MDIO:CLOCk:SOUrce command.

Description
  • This command sets or queries the MDIO Clock source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MDIO:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:MDIO:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MDIO:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:MDIO:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:MDIO:CLOCk:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the Clock source.
  • MATH<x> specifies a math waveform as the Clock source.
  • REF<x> specifies a reference waveform as the Clock source.

BusBItemMdioClockThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MDIO:CLOCk:THReshold command.

Description
  • This command sets or queries the MDIO clock source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MDIO:CLOCk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:MDIO:CLOCk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MDIO:CLOCk:THReshold value command.
SCPI Syntax
- BUS:B<x>:MDIO:CLOCk:THReshold <NR3>
- BUS:B<x>:MDIO:CLOCk:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the clock source threshold for the specified bus. The argument range is -8V to +8V.

BusBItemMdioData

Bases: SCPICmdRead

The BUS:B<x>:MDIO:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MDIO:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:MDIO:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .source: The BUS:B<x>:MDIO:DATA:SOUrce command.
  • .threshold: The BUS:B<x>:MDIO:DATA:THReshold command.

source property

Return the BUS:B<x>:MDIO:DATA:SOUrce command.

Description
  • This command sets or queries the data for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MDIO:DATA:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:MDIO:DATA:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MDIO:DATA:SOUrce value command.
SCPI Syntax
- BUS:B<x>:MDIO:DATA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:MDIO:DATA:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the data source.
  • MATH<x> specifies a math waveform as the data source.
  • REF<x> specifies a reference waveform as the data source.

threshold property

Return the BUS:B<x>:MDIO:DATA:THReshold command.

Description
  • This command sets or queries the MDIO Data source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MDIO:DATA:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:MDIO:DATA:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MDIO:DATA:THReshold value command.
SCPI Syntax
- BUS:B<x>:MDIO:DATA:THReshold <NR3>
- BUS:B<x>:MDIO:DATA:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the data source threshold for the specified bus. The argument range is -8V to +8V.

BusBItemMdioDataSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MDIO:DATA:SOUrce command.

Description
  • This command sets or queries the data for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MDIO:DATA:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:MDIO:DATA:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MDIO:DATA:SOUrce value command.
SCPI Syntax
- BUS:B<x>:MDIO:DATA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:MDIO:DATA:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the data source.
  • MATH<x> specifies a math waveform as the data source.
  • REF<x> specifies a reference waveform as the data source.

BusBItemMdioDataThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MDIO:DATA:THReshold command.

Description
  • This command sets or queries the MDIO Data source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MDIO:DATA:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:MDIO:DATA:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MDIO:DATA:THReshold value command.
SCPI Syntax
- BUS:B<x>:MDIO:DATA:THReshold <NR3>
- BUS:B<x>:MDIO:DATA:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the data source threshold for the specified bus. The argument range is -8V to +8V.

BusBItemMil1553b

Bases: SCPICmdRead

The BUS:B<x>:MIL1553B command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MIL1553B? query.
  • Using the .verify(value) method will send the BUS:B<x>:MIL1553B? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .lowthreshold: The BUS:B<x>:MIL1553B:LOWTHRESHold command.
  • .polarity: The BUS:B<x>:MIL1553B:POLarity command.
  • .responsetime: The BUS:B<x>:MIL1553B:RESPonsetime command tree.
  • .source: The BUS:B<x>:MIL1553B:SOUrce command.
  • .threshold: The BUS:B<x>:MIL1553B:THRESHold command.

lowthreshold property

Return the BUS:B<x>:MIL1553B:LOWTHRESHold command.

Description
  • This command sets or queries the MIL-STD-1553 lower threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MIL1553B:LOWTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:MIL1553B:LOWTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MIL1553B:LOWTHRESHold value command.
SCPI Syntax
- BUS:B<x>:MIL1553B:LOWTHRESHold <NR3>
- BUS:B<x>:MIL1553B:LOWTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the MIL-STD-1553 lower threshold for the specified bus.

polarity property

Return the BUS:B<x>:MIL1553B:POLarity command.

Description
  • This command sets or queries the source polarity for the specified MIL-STD-1553 bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MIL1553B:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:MIL1553B:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MIL1553B:POLarity value command.
SCPI Syntax
- BUS:B<x>:MIL1553B:POLarity {NORMal|INVERTed}
- BUS:B<x>:MIL1553B:POLarity?
Info
  • B<x> is the number of the bus.
  • NORMal specifies normal polarity.
  • INVERTed specifies inverted polarity.

responsetime property

Return the BUS:B<x>:MIL1553B:RESPonsetime command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MIL1553B:RESPonsetime? query.
  • Using the .verify(value) method will send the BUS:B<x>:MIL1553B:RESPonsetime? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .maximum: The BUS:B<x>:MIL1553B:RESPonsetime:MAXimum command.
  • .minimum: The BUS:B<x>:MIL1553B:RESPonsetime:MINimum command.

source property

Return the BUS:B<x>:MIL1553B:SOUrce command.

Description
  • This command sets or queries the source for the specified MIL-STD-1553 bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MIL1553B:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:MIL1553B:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MIL1553B:SOUrce value command.
SCPI Syntax
- BUS:B<x>:MIL1553B:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:MIL1553B:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source waveform for the MIL-STD-1553 bus.
  • MATH<x> specifies a math waveform as the source waveform for the MIL-STD-1553 bus.
  • REF<x> specifies a reference waveform as the source waveform for the MIL-STD-1553 bus.

threshold property

Return the BUS:B<x>:MIL1553B:THRESHold command.

Description
  • This command sets or queries the MIL-STD-1553 upper threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MIL1553B:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:MIL1553B:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MIL1553B:THRESHold value command.
SCPI Syntax
- BUS:B<x>:MIL1553B:THRESHold <NR3>
- BUS:B<x>:MIL1553B:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the MIL-STD-1553 upper threshold for the specified bus.

BusBItemMil1553bLowthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MIL1553B:LOWTHRESHold command.

Description
  • This command sets or queries the MIL-STD-1553 lower threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MIL1553B:LOWTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:MIL1553B:LOWTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MIL1553B:LOWTHRESHold value command.
SCPI Syntax
- BUS:B<x>:MIL1553B:LOWTHRESHold <NR3>
- BUS:B<x>:MIL1553B:LOWTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the MIL-STD-1553 lower threshold for the specified bus.

BusBItemMil1553bPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MIL1553B:POLarity command.

Description
  • This command sets or queries the source polarity for the specified MIL-STD-1553 bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MIL1553B:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:MIL1553B:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MIL1553B:POLarity value command.
SCPI Syntax
- BUS:B<x>:MIL1553B:POLarity {NORMal|INVERTed}
- BUS:B<x>:MIL1553B:POLarity?
Info
  • B<x> is the number of the bus.
  • NORMal specifies normal polarity.
  • INVERTed specifies inverted polarity.

BusBItemMil1553bResponsetime

Bases: SCPICmdRead

The BUS:B<x>:MIL1553B:RESPonsetime command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:MIL1553B:RESPonsetime? query.
  • Using the .verify(value) method will send the BUS:B<x>:MIL1553B:RESPonsetime? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .maximum: The BUS:B<x>:MIL1553B:RESPonsetime:MAXimum command.
  • .minimum: The BUS:B<x>:MIL1553B:RESPonsetime:MINimum command.

maximum property

Return the BUS:B<x>:MIL1553B:RESPonsetime:MAXimum command.

Description
  • This command sets or queries the maximum response time to a valid command issued for the specified MIL-STD-1553 bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MIL1553B:RESPonsetime:MAXimum? query.
  • Using the .verify(value) method will send the BUS:B<x>:MIL1553B:RESPonsetime:MAXimum? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MIL1553B:RESPonsetime:MAXimum value command.
SCPI Syntax
- BUS:B<x>:MIL1553B:RESPonsetime:MAXimum <NR3>
- BUS:B<x>:MIL1553B:RESPonsetime:MAXimum?
Info
  • B<x> is the number of the bus.
  • <NR3> is a floating point number that specifies the maximum response time, in seconds.

minimum property

Return the BUS:B<x>:MIL1553B:RESPonsetime:MINimum command.

Description
  • This command sets or queries the minimum response time to a valid command issued for the specified MIL-STD-1553 bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MIL1553B:RESPonsetime:MINimum? query.
  • Using the .verify(value) method will send the BUS:B<x>:MIL1553B:RESPonsetime:MINimum? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MIL1553B:RESPonsetime:MINimum value command.
SCPI Syntax
- BUS:B<x>:MIL1553B:RESPonsetime:MINimum <NR3>
- BUS:B<x>:MIL1553B:RESPonsetime:MINimum?
Info
  • B<x> is the Bus number.
  • <NR3> is a floating point number that specifies the minimum response time, in seconds.

BusBItemMil1553bResponsetimeMaximum

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MIL1553B:RESPonsetime:MAXimum command.

Description
  • This command sets or queries the maximum response time to a valid command issued for the specified MIL-STD-1553 bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MIL1553B:RESPonsetime:MAXimum? query.
  • Using the .verify(value) method will send the BUS:B<x>:MIL1553B:RESPonsetime:MAXimum? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MIL1553B:RESPonsetime:MAXimum value command.
SCPI Syntax
- BUS:B<x>:MIL1553B:RESPonsetime:MAXimum <NR3>
- BUS:B<x>:MIL1553B:RESPonsetime:MAXimum?
Info
  • B<x> is the number of the bus.
  • <NR3> is a floating point number that specifies the maximum response time, in seconds.

BusBItemMil1553bResponsetimeMinimum

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MIL1553B:RESPonsetime:MINimum command.

Description
  • This command sets or queries the minimum response time to a valid command issued for the specified MIL-STD-1553 bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MIL1553B:RESPonsetime:MINimum? query.
  • Using the .verify(value) method will send the BUS:B<x>:MIL1553B:RESPonsetime:MINimum? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MIL1553B:RESPonsetime:MINimum value command.
SCPI Syntax
- BUS:B<x>:MIL1553B:RESPonsetime:MINimum <NR3>
- BUS:B<x>:MIL1553B:RESPonsetime:MINimum?
Info
  • B<x> is the Bus number.
  • <NR3> is a floating point number that specifies the minimum response time, in seconds.

BusBItemMil1553bSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MIL1553B:SOUrce command.

Description
  • This command sets or queries the source for the specified MIL-STD-1553 bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MIL1553B:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:MIL1553B:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MIL1553B:SOUrce value command.
SCPI Syntax
- BUS:B<x>:MIL1553B:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:MIL1553B:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source waveform for the MIL-STD-1553 bus.
  • MATH<x> specifies a math waveform as the source waveform for the MIL-STD-1553 bus.
  • REF<x> specifies a reference waveform as the source waveform for the MIL-STD-1553 bus.

BusBItemMil1553bThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:MIL1553B:THRESHold command.

Description
  • This command sets or queries the MIL-STD-1553 upper threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:MIL1553B:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:MIL1553B:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:MIL1553B:THRESHold value command.
SCPI Syntax
- BUS:B<x>:MIL1553B:THRESHold <NR3>
- BUS:B<x>:MIL1553B:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the MIL-STD-1553 upper threshold for the specified bus.

BusBItemNfc

Bases: SCPICmdRead

The BUS:B<x>:NFC command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NFC? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC? query and raise an AssertionError if the returned value does not match value.
Properties
  • .applydemod: The BUS:B<x>:NFC:APPLYDEMod command.
  • .bitrate: The BUS:B<x>:NFC:BITRate command.
  • .cmd: The BUS:B<x>:NFC:CMD command tree.
  • .command: The BUS:B<x>:NFC:COMMand command tree.
  • .response: The BUS:B<x>:NFC:RESPonse command tree.
  • .rsp: The BUS:B<x>:NFC:RSP command tree.
  • .source: The BUS:B<x>:NFC:SOUrce command.
  • .start: The BUS:B<x>:NFC:START command tree.
  • .std: The BUS:B<x>:NFC:STD command.
  • .tolerance: The BUS:B<x>:NFC:TOLerance command.
  • .transition: The BUS:B<x>:NFC:TRANsition command tree.

applydemod property

Return the BUS:B<x>:NFC:APPLYDEMod command.

Description
  • This command sets or queries if the load modulation is applied on the response for the specified bus. The bus number is specified by .
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:APPLYDEMod? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:APPLYDEMod? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:APPLYDEMod value command.
SCPI Syntax
- BUS:B<x>:NFC:APPLYDEMod {1|0}
- BUS:B<x>:NFC:APPLYDEMod?
Info
  • 1 sets the demodulation application as true. This is the default value.
  • 0 sets the demodulation application as false.

bitrate property

Return the BUS:B<x>:NFC:BITRate command.

Description
  • This command sets or queries the bit rate for the specified NFC bus on the response lane. The bus number is specified by .
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:BITRate value command.
SCPI Syntax
- BUS:B<x>:NFC:BITRate <NR1>
- BUS:B<x>:NFC:BITRate?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the bit rate. The default bit rate value is 106000 bps. The minimum value is 1 and the maximum is 1000000000.

cmd property

Return the BUS:B<x>:NFC:CMD command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NFC:CMD? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:CMD? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .startindex: The BUS:B<x>:NFC:CMD:STARTINDex command.
  • .threshold: The BUS:B<x>:NFC:CMD:THReshold command.
  • .transition: The BUS:B<x>:NFC:CMD:TRANsition command tree.

command property

Return the BUS:B<x>:NFC:COMMand command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NFC:COMMand? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:COMMand? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .polarity: The BUS:B<x>:NFC:COMMand:POLarity command.

response property

Return the BUS:B<x>:NFC:RESPonse command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NFC:RESPonse? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:RESPonse? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .polarity: The BUS:B<x>:NFC:RESPonse:POLarity command.

rsp property

Return the BUS:B<x>:NFC:RSP command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NFC:RSP? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:RSP? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .threshold: The BUS:B<x>:NFC:RSP:THReshold command.

source property

Return the BUS:B<x>:NFC:SOUrce command.

Description
  • This command sets or queries the source for the specified NFC bus.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:SOUrce value command.
SCPI Syntax
- BUS:B<x>:NFC:SOUrce {CH<x>}
- BUS:B<x>:NFC:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.

start property

Return the BUS:B<x>:NFC:START command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NFC:START? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:START? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .index: The BUS:B<x>:NFC:START:INDex command.

std property

Return the BUS:B<x>:NFC:STD command.

Description
  • This command sets or queries the standard for the specified NFC bus.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:STD? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:STD? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:STD value command.
SCPI Syntax
- BUS:B<x>:NFC:STD {NFC15693|NFC14443A|NFC14443B|NFCFELica}
- BUS:B<x>:NFC:STD?
Info
  • B<x> is the number of the bus.
  • NFC15693 sets the standard to 15693.
  • NFC14443A sets the standard to 14443A.
  • NFC14443B sets the standard to 14443B.
  • NFCFELica sets the standard to Felica.

tolerance property

Return the BUS:B<x>:NFC:TOLerance command.

Description
  • This command sets or queries the tolerance for the specified NFC bus.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:TOLerance? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:TOLerance? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:TOLerance value command.
SCPI Syntax
- BUS:B<x>:NFC:TOLerance <NR3>
- BUS:B<x>:NFC:TOLerance?
Info
  • B<x> is the number of the bus.
  • <NR3> sets the tolerance. The default tolerance value is 10%. The minimum value is 1% and the maximum is 50%.

transition property

Return the BUS:B<x>:NFC:TRANsition command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NFC:TRANsition? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:TRANsition? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .zero: The BUS:B<x>:NFC:TRANsition:ZERo command.

BusBItemNfcApplydemod

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NFC:APPLYDEMod command.

Description
  • This command sets or queries if the load modulation is applied on the response for the specified bus. The bus number is specified by .
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:APPLYDEMod? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:APPLYDEMod? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:APPLYDEMod value command.
SCPI Syntax
- BUS:B<x>:NFC:APPLYDEMod {1|0}
- BUS:B<x>:NFC:APPLYDEMod?
Info
  • 1 sets the demodulation application as true. This is the default value.
  • 0 sets the demodulation application as false.

BusBItemNfcBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NFC:BITRate command.

Description
  • This command sets or queries the bit rate for the specified NFC bus on the response lane. The bus number is specified by .
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:BITRate value command.
SCPI Syntax
- BUS:B<x>:NFC:BITRate <NR1>
- BUS:B<x>:NFC:BITRate?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the bit rate. The default bit rate value is 106000 bps. The minimum value is 1 and the maximum is 1000000000.

BusBItemNfcCmd

Bases: SCPICmdRead

The BUS:B<x>:NFC:CMD command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NFC:CMD? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:CMD? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .startindex: The BUS:B<x>:NFC:CMD:STARTINDex command.
  • .threshold: The BUS:B<x>:NFC:CMD:THReshold command.
  • .transition: The BUS:B<x>:NFC:CMD:TRANsition command tree.

startindex property

Return the BUS:B<x>:NFC:CMD:STARTINDex command.

Description
  • This command sets or queries the Manchester start index for NFC command signal type for the specified bus
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:CMD:STARTINDex? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:CMD:STARTINDex? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:CMD:STARTINDex value command.
SCPI Syntax
- BUS:B<x>:NFC:CMD:STARTINDex <NR1>
- BUS:B<x>:NFC:CMD:STARTINDex?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the Manchester start index. The default value is 1. The minimum value is 1 and the maximum is 8.

threshold property

Return the BUS:B<x>:NFC:CMD:THReshold command.

Description
  • This command sets or queries the command threshold for the specified NFC bus.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:CMD:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:CMD:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:CMD:THReshold value command.
SCPI Syntax
- BUS:B<x>:NFC:CMD:THReshold <NR3>
- BUS:B<x>:NFC:CMD:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> sets the threshold. The default bit rate value is 1.0 V. The minimum value is -8.0 V and the maximum is 8.0 V.

transition property

Return the BUS:B<x>:NFC:CMD:TRANsition command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NFC:CMD:TRANsition? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:CMD:TRANsition? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .zero: The BUS:B<x>:NFC:CMD:TRANsition:ZERo command.

BusBItemNfcCmdStartindex

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NFC:CMD:STARTINDex command.

Description
  • This command sets or queries the Manchester start index for NFC command signal type for the specified bus
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:CMD:STARTINDex? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:CMD:STARTINDex? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:CMD:STARTINDex value command.
SCPI Syntax
- BUS:B<x>:NFC:CMD:STARTINDex <NR1>
- BUS:B<x>:NFC:CMD:STARTINDex?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the Manchester start index. The default value is 1. The minimum value is 1 and the maximum is 8.

BusBItemNfcCmdThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NFC:CMD:THReshold command.

Description
  • This command sets or queries the command threshold for the specified NFC bus.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:CMD:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:CMD:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:CMD:THReshold value command.
SCPI Syntax
- BUS:B<x>:NFC:CMD:THReshold <NR3>
- BUS:B<x>:NFC:CMD:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> sets the threshold. The default bit rate value is 1.0 V. The minimum value is -8.0 V and the maximum is 8.0 V.

BusBItemNfcCmdTransition

Bases: SCPICmdRead

The BUS:B<x>:NFC:CMD:TRANsition command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NFC:CMD:TRANsition? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:CMD:TRANsition? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .zero: The BUS:B<x>:NFC:CMD:TRANsition:ZERo command.

zero property

Return the BUS:B<x>:NFC:CMD:TRANsition:ZERo command.

Description
  • This command sets or queries the Manchester type for NFC command signal type for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:CMD:TRANsition:ZERo? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:CMD:TRANsition:ZERo? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:CMD:TRANsition:ZERo value command.
SCPI Syntax
- BUS:B<x>:NFC:CMD:TRANsition:ZERo {RISing|FALLing}
- BUS:B<x>:NFC:CMD:TRANsition:ZERo?
Info
  • B<x> is the number of the bus.
  • RISing sets the Manchester type to G.E thomas. This is the default value. FALLing sets the Manchester type to IEEE.

BusBItemNfcCmdTransitionZero

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NFC:CMD:TRANsition:ZERo command.

Description
  • This command sets or queries the Manchester type for NFC command signal type for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:CMD:TRANsition:ZERo? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:CMD:TRANsition:ZERo? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:CMD:TRANsition:ZERo value command.
SCPI Syntax
- BUS:B<x>:NFC:CMD:TRANsition:ZERo {RISing|FALLing}
- BUS:B<x>:NFC:CMD:TRANsition:ZERo?
Info
  • B<x> is the number of the bus.
  • RISing sets the Manchester type to G.E thomas. This is the default value. FALLing sets the Manchester type to IEEE.

BusBItemNfcCommand

Bases: SCPICmdRead

The BUS:B<x>:NFC:COMMand command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NFC:COMMand? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:COMMand? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .polarity: The BUS:B<x>:NFC:COMMand:POLarity command.

polarity property

Return the BUS:B<x>:NFC:COMMand:POLarity command.

Description
  • This command sets or queries the polarity for NFC Type B command signal type for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:COMMand:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:COMMand:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:COMMand:POLarity value command.
SCPI Syntax
- BUS:B<x>:NFC:COMMand:POLarity {NORMal|INVerted}
- BUS:B<x>:NFC:COMMand:POLarity?
Info
  • B<x> is the number of the bus.
  • NORMal sets the polarity of command to Normal.
  • INVerted sets the polarity of command to Inverted.

BusBItemNfcCommandPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NFC:COMMand:POLarity command.

Description
  • This command sets or queries the polarity for NFC Type B command signal type for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:COMMand:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:COMMand:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:COMMand:POLarity value command.
SCPI Syntax
- BUS:B<x>:NFC:COMMand:POLarity {NORMal|INVerted}
- BUS:B<x>:NFC:COMMand:POLarity?
Info
  • B<x> is the number of the bus.
  • NORMal sets the polarity of command to Normal.
  • INVerted sets the polarity of command to Inverted.

BusBItemNfcResponse

Bases: SCPICmdRead

The BUS:B<x>:NFC:RESPonse command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NFC:RESPonse? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:RESPonse? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .polarity: The BUS:B<x>:NFC:RESPonse:POLarity command.

polarity property

Return the BUS:B<x>:NFC:RESPonse:POLarity command.

Description
  • This command sets or queries the polarity for NFC Type B response signal type for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:RESPonse:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:RESPonse:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:RESPonse:POLarity value command.
SCPI Syntax
- BUS:B<x>:NFC:RESPonse:POLarity {NORMal|INVerted}
- BUS:B<x>:NFC:RESPonse:POLarity?
Info
  • B<x> is the number of the bus.
  • NORMal sets the polarity of command to Normal.
  • INVerted sets the polarity of command to Inverted.

BusBItemNfcResponsePolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NFC:RESPonse:POLarity command.

Description
  • This command sets or queries the polarity for NFC Type B response signal type for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:RESPonse:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:RESPonse:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:RESPonse:POLarity value command.
SCPI Syntax
- BUS:B<x>:NFC:RESPonse:POLarity {NORMal|INVerted}
- BUS:B<x>:NFC:RESPonse:POLarity?
Info
  • B<x> is the number of the bus.
  • NORMal sets the polarity of command to Normal.
  • INVerted sets the polarity of command to Inverted.

BusBItemNfcRsp

Bases: SCPICmdRead

The BUS:B<x>:NFC:RSP command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NFC:RSP? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:RSP? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .threshold: The BUS:B<x>:NFC:RSP:THReshold command.

threshold property

Return the BUS:B<x>:NFC:RSP:THReshold command.

Description
  • This command sets or queries the response threshold for the specified NFC bus.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:RSP:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:RSP:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:RSP:THReshold value command.
SCPI Syntax
- BUS:B<x>:NFC:RSP:THReshold <NR3>
- BUS:B<x>:NFC:RSP:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> sets the threshold. The default value is 1.0 V. The minimum value is -8.0 V and the maximum is 8.0 V.

BusBItemNfcRspThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NFC:RSP:THReshold command.

Description
  • This command sets or queries the response threshold for the specified NFC bus.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:RSP:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:RSP:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:RSP:THReshold value command.
SCPI Syntax
- BUS:B<x>:NFC:RSP:THReshold <NR3>
- BUS:B<x>:NFC:RSP:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> sets the threshold. The default value is 1.0 V. The minimum value is -8.0 V and the maximum is 8.0 V.

BusBItemNfcSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NFC:SOUrce command.

Description
  • This command sets or queries the source for the specified NFC bus.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:SOUrce value command.
SCPI Syntax
- BUS:B<x>:NFC:SOUrce {CH<x>}
- BUS:B<x>:NFC:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.

BusBItemNfcStart

Bases: SCPICmdRead

The BUS:B<x>:NFC:START command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NFC:START? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:START? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .index: The BUS:B<x>:NFC:START:INDex command.

index property

Return the BUS:B<x>:NFC:START:INDex command.

Description
  • This command sets or queries the start index for the specified NFC bus on the response lane.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:START:INDex? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:START:INDex? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:START:INDex value command.
SCPI Syntax
- BUS:B<x>:NFC:START:INDex <NR1>
- BUS:B<x>:NFC:START:INDex?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the index. The default value is 1. The minimum value is 1 and the maximum is 8.

BusBItemNfcStartIndex

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NFC:START:INDex command.

Description
  • This command sets or queries the start index for the specified NFC bus on the response lane.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:START:INDex? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:START:INDex? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:START:INDex value command.
SCPI Syntax
- BUS:B<x>:NFC:START:INDex <NR1>
- BUS:B<x>:NFC:START:INDex?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the index. The default value is 1. The minimum value is 1 and the maximum is 8.

BusBItemNfcStd

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NFC:STD command.

Description
  • This command sets or queries the standard for the specified NFC bus.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:STD? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:STD? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:STD value command.
SCPI Syntax
- BUS:B<x>:NFC:STD {NFC15693|NFC14443A|NFC14443B|NFCFELica}
- BUS:B<x>:NFC:STD?
Info
  • B<x> is the number of the bus.
  • NFC15693 sets the standard to 15693.
  • NFC14443A sets the standard to 14443A.
  • NFC14443B sets the standard to 14443B.
  • NFCFELica sets the standard to Felica.

BusBItemNfcTolerance

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NFC:TOLerance command.

Description
  • This command sets or queries the tolerance for the specified NFC bus.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:TOLerance? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:TOLerance? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:TOLerance value command.
SCPI Syntax
- BUS:B<x>:NFC:TOLerance <NR3>
- BUS:B<x>:NFC:TOLerance?
Info
  • B<x> is the number of the bus.
  • <NR3> sets the tolerance. The default tolerance value is 10%. The minimum value is 1% and the maximum is 50%.

BusBItemNfcTransition

Bases: SCPICmdRead

The BUS:B<x>:NFC:TRANsition command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NFC:TRANsition? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:TRANsition? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .zero: The BUS:B<x>:NFC:TRANsition:ZERo command.

zero property

Return the BUS:B<x>:NFC:TRANsition:ZERo command.

Description
  • The command sets or queries the Manchester standard convention for the specified NFC bus on the response lane.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:TRANsition:ZERo? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:TRANsition:ZERo? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:TRANsition:ZERo value command.
SCPI Syntax
- BUS:B<x>:NFC:TRANsition:ZERo {ge|ieee}
- BUS:B<x>:NFC:TRANsition:ZERo?
Info
  • B<x> is the number of the bus.
  • ge sets the transition type to Manchester G.E thomas convention.
  • ieee sets the transition type to Manchester IEEE convention.

BusBItemNfcTransitionZero

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NFC:TRANsition:ZERo command.

Description
  • The command sets or queries the Manchester standard convention for the specified NFC bus on the response lane.
Usage
  • Using the .query() method will send the BUS:B<x>:NFC:TRANsition:ZERo? query.
  • Using the .verify(value) method will send the BUS:B<x>:NFC:TRANsition:ZERo? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NFC:TRANsition:ZERo value command.
SCPI Syntax
- BUS:B<x>:NFC:TRANsition:ZERo {ge|ieee}
- BUS:B<x>:NFC:TRANsition:ZERo?
Info
  • B<x> is the number of the bus.
  • ge sets the transition type to Manchester G.E thomas convention.
  • ieee sets the transition type to Manchester IEEE convention.

BusBItemNrz

Bases: SCPICmdRead

The BUS:B<x>:NRZ command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NRZ? query.
  • Using the .verify(value) method will send the BUS:B<x>:NRZ? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .bitorder: The BUS:B<x>:NRZ:BITOrder command.
  • .bitrate: The BUS:B<x>:NRZ:BITRate command.
  • .polarity: The BUS:B<x>:NRZ:POLarity command.
  • .source: The BUS:B<x>:NRZ:SOUrce command.
  • .spmi: The BUS:B<x>:NRZ:SPMI command tree.
  • .threshold: The BUS:B<x>:NRZ:THReshold command.

bitorder property

Return the BUS:B<x>:NRZ:BITOrder command.

Description
  • This command sets or queries the NRZ bit order for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:NRZ:BITOrder? query.
  • Using the .verify(value) method will send the BUS:B<x>:NRZ:BITOrder? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NRZ:BITOrder value command.
SCPI Syntax
- BUS:B<x>:NRZ:BITOrder {LSB|MSB}
- BUS:B<x>:NRZ:BITOrder?
Info
  • B<x> is the number of the bus.
  • LSB specifies that each bit becomes the recovered value’s new LSB, after shifting previously recovered bits one place to the left. The decoding happens right to left.
  • MSB specifies that each successive bit from the bus’s data line becomes the new MSB of the recovered value, shifting any previously recovered bits one place to the right. The decoding happens left to right.

bitrate property

Return the BUS:B<x>:NRZ:BITRate command.

Description
  • This command sets or queries the NRZ bus bit rate. The bus number is specified by .
Usage
  • Using the .query() method will send the BUS:B<x>:NRZ:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:NRZ:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NRZ:BITRate value command.
SCPI Syntax
- BUS:B<x>:NRZ:BITRate <NR1>
- BUS:B<x>:NRZ:BITRate?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the bit rate up to 1 G.

polarity property

Return the BUS:B<x>:NRZ:POLarity command.

Description
  • This command sets or queries the NRZ source polarity for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:NRZ:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:NRZ:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NRZ:POLarity value command.
SCPI Syntax
- BUS:B<x>:NRZ:POLarity {INVerted|NORmal}
- BUS:B<x>:NRZ:POLarity?
Info
  • B<x> is the number of the bus.
  • INVerted specifies inverted polarity.
  • NORmal specifies normal polarity.

source property

Return the BUS:B<x>:NRZ:SOUrce command.

Description
  • This command sets or queries the NRZ source for the specified bus. This command specifies the source channel. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:NRZ:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:NRZ:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NRZ:SOUrce value command.
SCPI Syntax
- BUS:B<x>:NRZ:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:NRZ:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a digital reference waveform as the source.

spmi property

Return the BUS:B<x>:NRZ:SPMI command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NRZ:SPMI? query.
  • Using the .verify(value) method will send the BUS:B<x>:NRZ:SPMI? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .version: The BUS:B<x>:NRZ:SPMI:VERsion command.

threshold property

Return the BUS:B<x>:NRZ:THReshold command.

Description
  • This command sets or queries the NRZ threshold for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:NRZ:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:NRZ:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NRZ:THReshold value command.
SCPI Syntax
- BUS:B<x>:NRZ:THReshold <NR3>
- BUS:B<x>:NRZ:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the NRZ Strobe threshold for the specified bus in volts. The valid range is -8 V to +8 V.

BusBItemNrzBitorder

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NRZ:BITOrder command.

Description
  • This command sets or queries the NRZ bit order for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:NRZ:BITOrder? query.
  • Using the .verify(value) method will send the BUS:B<x>:NRZ:BITOrder? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NRZ:BITOrder value command.
SCPI Syntax
- BUS:B<x>:NRZ:BITOrder {LSB|MSB}
- BUS:B<x>:NRZ:BITOrder?
Info
  • B<x> is the number of the bus.
  • LSB specifies that each bit becomes the recovered value’s new LSB, after shifting previously recovered bits one place to the left. The decoding happens right to left.
  • MSB specifies that each successive bit from the bus’s data line becomes the new MSB of the recovered value, shifting any previously recovered bits one place to the right. The decoding happens left to right.

BusBItemNrzBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NRZ:BITRate command.

Description
  • This command sets or queries the NRZ bus bit rate. The bus number is specified by .
Usage
  • Using the .query() method will send the BUS:B<x>:NRZ:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:NRZ:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NRZ:BITRate value command.
SCPI Syntax
- BUS:B<x>:NRZ:BITRate <NR1>
- BUS:B<x>:NRZ:BITRate?
Info
  • B<x> is the number of the bus.
  • <NR1> sets the bit rate up to 1 G.

BusBItemNrzPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NRZ:POLarity command.

Description
  • This command sets or queries the NRZ source polarity for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:NRZ:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:NRZ:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NRZ:POLarity value command.
SCPI Syntax
- BUS:B<x>:NRZ:POLarity {INVerted|NORmal}
- BUS:B<x>:NRZ:POLarity?
Info
  • B<x> is the number of the bus.
  • INVerted specifies inverted polarity.
  • NORmal specifies normal polarity.

BusBItemNrzSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NRZ:SOUrce command.

Description
  • This command sets or queries the NRZ source for the specified bus. This command specifies the source channel. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:NRZ:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:NRZ:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NRZ:SOUrce value command.
SCPI Syntax
- BUS:B<x>:NRZ:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:NRZ:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a digital reference waveform as the source.

BusBItemNrzSpmi

Bases: SCPICmdRead

The BUS:B<x>:NRZ:SPMI command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:NRZ:SPMI? query.
  • Using the .verify(value) method will send the BUS:B<x>:NRZ:SPMI? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .version: The BUS:B<x>:NRZ:SPMI:VERsion command.

version property

Return the BUS:B<x>:NRZ:SPMI:VERsion command.

Description
  • This command sets or queries the Version for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:NRZ:SPMI:VERsion? query.
  • Using the .verify(value) method will send the BUS:B<x>:NRZ:SPMI:VERsion? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NRZ:SPMI:VERsion value command.
SCPI Syntax
- BUS:B<x>:NRZ:SPMI:VERsion v<x>
- BUS:B<x>:NRZ:SPMI:VERsion?
Info
  • B<x> is the number of the bus.
  • v<x> specifies version the version number. Valid version are 1 or 2.

BusBItemNrzSpmiVersion

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NRZ:SPMI:VERsion command.

Description
  • This command sets or queries the Version for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:NRZ:SPMI:VERsion? query.
  • Using the .verify(value) method will send the BUS:B<x>:NRZ:SPMI:VERsion? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NRZ:SPMI:VERsion value command.
SCPI Syntax
- BUS:B<x>:NRZ:SPMI:VERsion v<x>
- BUS:B<x>:NRZ:SPMI:VERsion?
Info
  • B<x> is the number of the bus.
  • v<x> specifies version the version number. Valid version are 1 or 2.

BusBItemNrzThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:NRZ:THReshold command.

Description
  • This command sets or queries the NRZ threshold for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:NRZ:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:NRZ:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:NRZ:THReshold value command.
SCPI Syntax
- BUS:B<x>:NRZ:THReshold <NR3>
- BUS:B<x>:NRZ:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the NRZ Strobe threshold for the specified bus in volts. The valid range is -8 V to +8 V.

BusBItemOnewire

Bases: SCPICmdRead

The BUS:B<x>:ONEWIRe command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ONEWIRe? query.
  • Using the .verify(value) method will send the BUS:B<x>:ONEWIRe? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .data: The BUS:B<x>:ONEWIRe:DATA command tree.
  • .mode: The BUS:B<x>:ONEWIRe:MODe command.

data property

Return the BUS:B<x>:ONEWIRe:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ONEWIRe:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:ONEWIRe:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .source: The BUS:B<x>:ONEWIRe:DATA:SOUrce command.
  • .threshold: The BUS:B<x>:ONEWIRe:DATA:THReshold command.

mode property

Return the BUS:B<x>:ONEWIRe:MODe command.

Description
  • This command sets or queries the mode for the specified ONEWIRe bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ONEWIRe:MODe? query.
  • Using the .verify(value) method will send the BUS:B<x>:ONEWIRe:MODe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ONEWIRe:MODe value command.
SCPI Syntax
- BUS:B<x>:ONEWIRe:MODe {STAndard|OVErdrive}
- BUS:B<x>:ONEWIRe:MODe?
Info
  • B<x> is the number of the bus.
  • STAndard specifies the mode as standard. Standard is the default mode whose value is 15.4 kbs.
  • OVErdrive specifies the mode as overdrive.

BusBItemOnewireData

Bases: SCPICmdRead

The BUS:B<x>:ONEWIRe:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:ONEWIRe:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:ONEWIRe:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .source: The BUS:B<x>:ONEWIRe:DATA:SOUrce command.
  • .threshold: The BUS:B<x>:ONEWIRe:DATA:THReshold command.

source property

Return the BUS:B<x>:ONEWIRe:DATA:SOUrce command.

Description
  • This command sets or queries the ONEWIRe source for the specified bus. This command specifies the source channel. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ONEWIRe:DATA:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ONEWIRe:DATA:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ONEWIRe:DATA:SOUrce value command.
SCPI Syntax
- BUS:B<x>:ONEWIRe:DATA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ONEWIRe:DATA:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a digital reference waveform as the source.

threshold property

Return the BUS:B<x>:ONEWIRe:DATA:THReshold command.

Description
  • This command sets or queries the ONEWIRe data source threshold for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ONEWIRe:DATA:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ONEWIRe:DATA:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ONEWIRe:DATA:THReshold value command.
SCPI Syntax
- BUS:B<x>:ONEWIRe:DATA:THReshold <NR3>
- BUS:B<x>:ONEWIRe:DATA:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the ONEWIRe Strobe threshold for the specified bus in volts. The valid range is -8 V to +8 V. The default value is 1.25 V.

BusBItemOnewireDataSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ONEWIRe:DATA:SOUrce command.

Description
  • This command sets or queries the ONEWIRe source for the specified bus. This command specifies the source channel. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ONEWIRe:DATA:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:ONEWIRe:DATA:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ONEWIRe:DATA:SOUrce value command.
SCPI Syntax
- BUS:B<x>:ONEWIRe:DATA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:ONEWIRe:DATA:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a digital reference waveform as the source.

BusBItemOnewireDataThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ONEWIRe:DATA:THReshold command.

Description
  • This command sets or queries the ONEWIRe data source threshold for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ONEWIRe:DATA:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:ONEWIRe:DATA:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ONEWIRe:DATA:THReshold value command.
SCPI Syntax
- BUS:B<x>:ONEWIRe:DATA:THReshold <NR3>
- BUS:B<x>:ONEWIRe:DATA:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the ONEWIRe Strobe threshold for the specified bus in volts. The valid range is -8 V to +8 V. The default value is 1.25 V.

BusBItemOnewireMode

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:ONEWIRe:MODe command.

Description
  • This command sets or queries the mode for the specified ONEWIRe bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:ONEWIRe:MODe? query.
  • Using the .verify(value) method will send the BUS:B<x>:ONEWIRe:MODe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:ONEWIRe:MODe value command.
SCPI Syntax
- BUS:B<x>:ONEWIRe:MODe {STAndard|OVErdrive}
- BUS:B<x>:ONEWIRe:MODe?
Info
  • B<x> is the number of the bus.
  • STAndard specifies the mode as standard. Standard is the default mode whose value is 15.4 kbs.
  • OVErdrive specifies the mode as overdrive.

BusBItemParallel

Bases: SCPICmdRead

The BUS:B<x>:PARallel command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PARallel? query.
  • Using the .verify(value) method will send the BUS:B<x>:PARallel? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .allthresholds: The BUS:B<x>:PARallel:ALLTHResholds command.
  • .bitsource: The BUS:B<x>:PARallel:BIT<x>SOUrce command.
  • .clocksource: The BUS:B<x>:PARallel:CLOCKSOUrce command.
  • .clock: The BUS:B<x>:PARallel:CLOCk command tree.

allthresholds property

Return the BUS:B<x>:PARallel:ALLTHResholds command.

Description
  • This command sets or queries a threshold value for sources for the parallel bus. Use the BUS:B<x>:PARallel:ALLTHResholds:APPly command to set the thresholds to this value. The bus is specified by x.
Usage
  • Using the .write(value) method will send the BUS:B<x>:PARallel:ALLTHResholds value command.
SCPI Syntax
- BUS:B<x>:PARallel:ALLTHResholds <NR3>
Info
  • B<x> is the number of the bus.
  • <NR3> is the source threshold.
Sub-properties
  • .apply: The BUS:B<x>:PARallel:ALLTHResholds:APPly command.

bitsource property

Return the BUS:B<x>:PARallel:BIT<x>SOUrce command.

Description
  • This command sets or queries the specified bit source for specified parallel bus.
Usage
  • Using the .write(value) method will send the BUS:B<x>:PARallel:BIT<x>SOUrce value command.
SCPI Syntax
- BUS:B<x>:PARallel:BIT<x>SOUrce {CH<x>|CH<x>_D<x>|MATH<x>|REF<x>|REF<x>_D<x>|NONE}
Info
  • B<x> is the number of the bus.
  • BIT<x> is the number of the bit and is an integer in the range of 1 to 64.
  • CH<x> is the specified bit source.
  • CH<x>_D<x> is the specified bit source.
  • MATH<x> is the specified bit source.
  • REF<x> is the specified bit source.
  • REF<x>_D<x> specifies a digital reference waveform as the bit source waveform for the specified parallel bus.
  • NONE disables the bit source.
Sub-properties
  • .threshold: The BUS:B<x>:PARallel:BIT<x>SOUrce:THReshold command.

clock property

Return the BUS:B<x>:PARallel:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PARallel:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:PARallel:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .edge: The BUS:B<x>:PARallel:CLOCk:EDGE command.
  • .isclocked: The BUS:B<x>:PARallel:CLOCk:ISCLOCKED command.

clocksource property

Return the BUS:B<x>:PARallel:CLOCKSOUrce command.

Description
  • This command sets or queries the Parallel clock bit source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:PARallel:CLOCKSOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:PARallel:CLOCKSOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PARallel:CLOCKSOUrce value command.
SCPI Syntax
- BUS:B<x>:PARallel:CLOCKSOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>|NONE}
- BUS:B<x>:PARallel:CLOCKSOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog FlexChannel to use as the bus clock source.
  • MATH<x> specifies the math channel to use as the bus clock source.
  • REF<x> specifies the reference channel to use as the bus clock source.
  • REF<x>_D<x> specifies a digital reference waveform as the clock source waveform for the specified parallel bus.
  • NONE specifies the reference channel to use as the bus clock source.
Sub-properties
  • .threshold: The BUS:B<x>:PARallel:CLOCKSOUrce:THReshold command.

BusBItemParallelAllthresholds

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PARallel:ALLTHResholds command.

Description
  • This command sets or queries a threshold value for sources for the parallel bus. Use the BUS:B<x>:PARallel:ALLTHResholds:APPly command to set the thresholds to this value. The bus is specified by x.
Usage
  • Using the .write(value) method will send the BUS:B<x>:PARallel:ALLTHResholds value command.
SCPI Syntax
- BUS:B<x>:PARallel:ALLTHResholds <NR3>
Info
  • B<x> is the number of the bus.
  • <NR3> is the source threshold.
Properties
  • .apply: The BUS:B<x>:PARallel:ALLTHResholds:APPly command.

apply property

Return the BUS:B<x>:PARallel:ALLTHResholds:APPly command.

Description
  • This command sets all of the data source thresholds to the value set by BUS:B<x>:PARallel:ALLTHResholds for the parallel bus. The bus is specified by x.
Usage
  • Using the .write() method will send the BUS:B<x>:PARallel:ALLTHResholds:APPly command.
SCPI Syntax
- BUS:B<x>:PARallel:ALLTHResholds:APPly
Info
  • B<x> is the Bus number.

BusBItemParallelAllthresholdsApply

Bases: SCPICmdWriteNoArguments

The BUS:B<x>:PARallel:ALLTHResholds:APPly command.

Description
  • This command sets all of the data source thresholds to the value set by BUS:B<x>:PARallel:ALLTHResholds for the parallel bus. The bus is specified by x.
Usage
  • Using the .write() method will send the BUS:B<x>:PARallel:ALLTHResholds:APPly command.
SCPI Syntax
- BUS:B<x>:PARallel:ALLTHResholds:APPly
Info
  • B<x> is the Bus number.

BusBItemParallelBitsourceItem

Bases: ValidatedDynamicNumberCmd, SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PARallel:BIT<x>SOUrce command.

Description
  • This command sets or queries the specified bit source for specified parallel bus.
Usage
  • Using the .write(value) method will send the BUS:B<x>:PARallel:BIT<x>SOUrce value command.
SCPI Syntax
- BUS:B<x>:PARallel:BIT<x>SOUrce {CH<x>|CH<x>_D<x>|MATH<x>|REF<x>|REF<x>_D<x>|NONE}
Info
  • B<x> is the number of the bus.
  • BIT<x> is the number of the bit and is an integer in the range of 1 to 64.
  • CH<x> is the specified bit source.
  • CH<x>_D<x> is the specified bit source.
  • MATH<x> is the specified bit source.
  • REF<x> is the specified bit source.
  • REF<x>_D<x> specifies a digital reference waveform as the bit source waveform for the specified parallel bus.
  • NONE disables the bit source.
Properties
  • .threshold: The BUS:B<x>:PARallel:BIT<x>SOUrce:THReshold command.

threshold property

Return the BUS:B<x>:PARallel:BIT<x>SOUrce:THReshold command.

Description
  • This command sets or queries the specified bit source threshold for the specified parallel bus.
Usage
  • Using the .write(value) method will send the BUS:B<x>:PARallel:BIT<x>SOUrce:THReshold value command.
SCPI Syntax
- BUS:B<x>:PARallel:BIT<x>SOUrce:THReshold <NR3>
Info
  • B<x> is the number of the bus.
  • BIT<x> is the number of the bit and is an integer in the range of 1 to 64.
  • <NR3> is the specified bit source threshold for the specified parallel bus.

BusBItemParallelBitsourceItemThreshold

Bases: SCPICmdWrite

The BUS:B<x>:PARallel:BIT<x>SOUrce:THReshold command.

Description
  • This command sets or queries the specified bit source threshold for the specified parallel bus.
Usage
  • Using the .write(value) method will send the BUS:B<x>:PARallel:BIT<x>SOUrce:THReshold value command.
SCPI Syntax
- BUS:B<x>:PARallel:BIT<x>SOUrce:THReshold <NR3>
Info
  • B<x> is the number of the bus.
  • BIT<x> is the number of the bit and is an integer in the range of 1 to 64.
  • <NR3> is the specified bit source threshold for the specified parallel bus.

BusBItemParallelClock

Bases: SCPICmdRead

The BUS:B<x>:PARallel:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PARallel:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:PARallel:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .edge: The BUS:B<x>:PARallel:CLOCk:EDGE command.
  • .isclocked: The BUS:B<x>:PARallel:CLOCk:ISCLOCKED command.

edge property

Return the BUS:B<x>:PARallel:CLOCk:EDGE command.

Description
  • This command sets or queries the clock edge for the parallel bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:PARallel:CLOCk:EDGE? query.
  • Using the .verify(value) method will send the BUS:B<x>:PARallel:CLOCk:EDGE? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PARallel:CLOCk:EDGE value command.
SCPI Syntax
- BUS:B<x>:PARallel:CLOCk:EDGE {FALLING|RISING|EITHER}
- BUS:B<x>:PARallel:CLOCk:EDGE?
Info
  • B<x> is the number of the bus.
  • FALLING decodes on the falling edge of the clocked parallel bus signal.
  • RISING decodes on the rising edge of the clocked parallel bus signal.
  • EITHER decodes on the rising or falling edge of the clocked parallel bus signal.

isclocked property

Return the BUS:B<x>:PARallel:CLOCk:ISCLOCKED command.

Description
  • This command determines whether the bus operates in a clocked or asynchronous fashion. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:PARallel:CLOCk:ISCLOCKED? query.
  • Using the .verify(value) method will send the BUS:B<x>:PARallel:CLOCk:ISCLOCKED? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PARallel:CLOCk:ISCLOCKED value command.
SCPI Syntax
- BUS:B<x>:PARallel:CLOCk:ISCLOCKED {ON|OFF|<NR1>}
- BUS:B<x>:PARallel:CLOCk:ISCLOCKED?
Info
  • B<x> is the number of the bus.
  • OFF argument specifies an asynchronous bus.
  • ON argument specifies a clocked bus.
  • <NR1> = 0 specifies an asynchronous bus; any other value specifies a clocked bus.

BusBItemParallelClockEdge

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PARallel:CLOCk:EDGE command.

Description
  • This command sets or queries the clock edge for the parallel bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:PARallel:CLOCk:EDGE? query.
  • Using the .verify(value) method will send the BUS:B<x>:PARallel:CLOCk:EDGE? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PARallel:CLOCk:EDGE value command.
SCPI Syntax
- BUS:B<x>:PARallel:CLOCk:EDGE {FALLING|RISING|EITHER}
- BUS:B<x>:PARallel:CLOCk:EDGE?
Info
  • B<x> is the number of the bus.
  • FALLING decodes on the falling edge of the clocked parallel bus signal.
  • RISING decodes on the rising edge of the clocked parallel bus signal.
  • EITHER decodes on the rising or falling edge of the clocked parallel bus signal.

BusBItemParallelClockIsclocked

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PARallel:CLOCk:ISCLOCKED command.

Description
  • This command determines whether the bus operates in a clocked or asynchronous fashion. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:PARallel:CLOCk:ISCLOCKED? query.
  • Using the .verify(value) method will send the BUS:B<x>:PARallel:CLOCk:ISCLOCKED? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PARallel:CLOCk:ISCLOCKED value command.
SCPI Syntax
- BUS:B<x>:PARallel:CLOCk:ISCLOCKED {ON|OFF|<NR1>}
- BUS:B<x>:PARallel:CLOCk:ISCLOCKED?
Info
  • B<x> is the number of the bus.
  • OFF argument specifies an asynchronous bus.
  • ON argument specifies a clocked bus.
  • <NR1> = 0 specifies an asynchronous bus; any other value specifies a clocked bus.

BusBItemParallelClocksource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PARallel:CLOCKSOUrce command.

Description
  • This command sets or queries the Parallel clock bit source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:PARallel:CLOCKSOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:PARallel:CLOCKSOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PARallel:CLOCKSOUrce value command.
SCPI Syntax
- BUS:B<x>:PARallel:CLOCKSOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>|NONE}
- BUS:B<x>:PARallel:CLOCKSOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog FlexChannel to use as the bus clock source.
  • MATH<x> specifies the math channel to use as the bus clock source.
  • REF<x> specifies the reference channel to use as the bus clock source.
  • REF<x>_D<x> specifies a digital reference waveform as the clock source waveform for the specified parallel bus.
  • NONE specifies the reference channel to use as the bus clock source.
Properties
  • .threshold: The BUS:B<x>:PARallel:CLOCKSOUrce:THReshold command.

threshold property

Return the BUS:B<x>:PARallel:CLOCKSOUrce:THReshold command.

Description
  • This command sets or queries the clock source threshold for the parallel bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:PARallel:CLOCKSOUrce:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:PARallel:CLOCKSOUrce:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PARallel:CLOCKSOUrce:THReshold value command.
SCPI Syntax
- BUS:B<x>:PARallel:CLOCKSOUrce:THReshold <NR3>
- BUS:B<x>:PARallel:CLOCKSOUrce:THReshold?
Info
  • B<x> is the Bus number.
  • <NR3> is the clock bit source threshold for the parallel bus.

BusBItemParallelClocksourceThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PARallel:CLOCKSOUrce:THReshold command.

Description
  • This command sets or queries the clock source threshold for the parallel bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:PARallel:CLOCKSOUrce:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:PARallel:CLOCKSOUrce:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PARallel:CLOCKSOUrce:THReshold value command.
SCPI Syntax
- BUS:B<x>:PARallel:CLOCKSOUrce:THReshold <NR3>
- BUS:B<x>:PARallel:CLOCKSOUrce:THReshold?
Info
  • B<x> is the Bus number.
  • <NR3> is the clock bit source threshold for the parallel bus.

BusBItemPcie

Bases: SCPICmdRead

The BUS:B<x>:PCIE command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PCIE? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .dataminus: The BUS:B<x>:PCIE:DATAMinus command tree.
  • .dataplus: The BUS:B<x>:PCIE:DATAPlus command tree.
  • .mthreshold: The BUS:B<x>:PCIE:MTHReshold command.
  • .signaltype: The BUS:B<x>:PCIE:SIGNALTYpe command.
  • .source: The BUS:B<x>:PCIE:SOUrce command tree.

dataminus property

Return the BUS:B<x>:PCIE:DATAMinus command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:DATAMinus? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:DATAMinus? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .threshold: The BUS:B<x>:PCIE:DATAMinus:THRESHold command tree.

dataplus property

Return the BUS:B<x>:PCIE:DATAPlus command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:DATAPlus? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:DATAPlus? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .threshold: The BUS:B<x>:PCIE:DATAPlus:THRESHold command tree.

mthreshold property

Return the BUS:B<x>:PCIE:MTHReshold command.

Description
  • This command sets or queries PCIe math source threshold for single ended for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:MTHReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:MTHReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PCIE:MTHReshold value command.
SCPI Syntax
- BUS:B<x>:PCIE:MTHReshold <NR3>
- BUS:B<x>:PCIE:MTHReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the specified PCIe math source threshold for the specified bus. The valid range is -8.0 V to 8.0 V.

signaltype property

Return the BUS:B<x>:PCIE:SIGNALTYpe command.

Description
  • This command sets or queries the PCIe signal type for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:SIGNALTYpe? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:SIGNALTYpe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PCIE:SIGNALTYpe value command.
SCPI Syntax
- BUS:B<x>:PCIE:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:PCIE:SIGNALTYpe?
Info
  • B<x> is the number of the bus.
  • SINGLE specifies the Signal Type for PCIE bus as single.
  • DIFF specifies the Signal Type for PCIE bus as differential.

source property

Return the BUS:B<x>:PCIE:SOUrce command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:SOUrce? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .dminus: The BUS:B<x>:PCIE:SOUrce:DMINus command.
  • .dplus: The BUS:B<x>:PCIE:SOUrce:DPLUs command.

BusBItemPcieDataminus

Bases: SCPICmdRead

The BUS:B<x>:PCIE:DATAMinus command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:DATAMinus? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:DATAMinus? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .threshold: The BUS:B<x>:PCIE:DATAMinus:THRESHold command tree.

threshold property

Return the BUS:B<x>:PCIE:DATAMinus:THRESHold command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:DATAMinus:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:DATAMinus:THRESHold? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .high: The BUS:B<x>:PCIE:DATAMinus:THRESHold:HIGh command.

BusBItemPcieDataminusThreshold

Bases: SCPICmdRead

The BUS:B<x>:PCIE:DATAMinus:THRESHold command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:DATAMinus:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:DATAMinus:THRESHold? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .high: The BUS:B<x>:PCIE:DATAMinus:THRESHold:HIGh command.

high property

Return the BUS:B<x>:PCIE:DATAMinus:THRESHold:HIGh command.

Description
  • This command sets or queries the PCIe DATA Minus source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:DATAMinus:THRESHold:HIGh? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:DATAMinus:THRESHold:HIGh? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PCIE:DATAMinus:THRESHold:HIGh value command.
SCPI Syntax
- BUS:B<x>:PCIE:DATAMinus:THRESHold:HIGh <NR3>
- BUS:B<x>:PCIE:DATAMinus:THRESHold:HIGh?
Info
  • B<x> is the number of the bus.
  • <NR3> is the specified PCIe DATA Minus source threshold for the specified bus. The valid range is -8.0 V to 8.0 V.

BusBItemPcieDataminusThresholdHigh

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PCIE:DATAMinus:THRESHold:HIGh command.

Description
  • This command sets or queries the PCIe DATA Minus source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:DATAMinus:THRESHold:HIGh? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:DATAMinus:THRESHold:HIGh? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PCIE:DATAMinus:THRESHold:HIGh value command.
SCPI Syntax
- BUS:B<x>:PCIE:DATAMinus:THRESHold:HIGh <NR3>
- BUS:B<x>:PCIE:DATAMinus:THRESHold:HIGh?
Info
  • B<x> is the number of the bus.
  • <NR3> is the specified PCIe DATA Minus source threshold for the specified bus. The valid range is -8.0 V to 8.0 V.

BusBItemPcieDataplus

Bases: SCPICmdRead

The BUS:B<x>:PCIE:DATAPlus command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:DATAPlus? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:DATAPlus? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .threshold: The BUS:B<x>:PCIE:DATAPlus:THRESHold command tree.

threshold property

Return the BUS:B<x>:PCIE:DATAPlus:THRESHold command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:DATAPlus:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:DATAPlus:THRESHold? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .high: The BUS:B<x>:PCIE:DATAPlus:THRESHold:HIGh command.

BusBItemPcieDataplusThreshold

Bases: SCPICmdRead

The BUS:B<x>:PCIE:DATAPlus:THRESHold command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:DATAPlus:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:DATAPlus:THRESHold? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .high: The BUS:B<x>:PCIE:DATAPlus:THRESHold:HIGh command.

high property

Return the BUS:B<x>:PCIE:DATAPlus:THRESHold:HIGh command.

Description
  • This command sets or queries the PCIe DATA Plus source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:DATAPlus:THRESHold:HIGh? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:DATAPlus:THRESHold:HIGh? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PCIE:DATAPlus:THRESHold:HIGh value command.
SCPI Syntax
- BUS:B<x>:PCIE:DATAPlus:THRESHold:HIGh <NR3>
- BUS:B<x>:PCIE:DATAPlus:THRESHold:HIGh?
Info
  • B<x> is the number of the bus.
  • <NR3> is the specified PCIe DATA Plus source threshold for the specified bus. The valid range is -8.0 V to 8.0 V.

BusBItemPcieDataplusThresholdHigh

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PCIE:DATAPlus:THRESHold:HIGh command.

Description
  • This command sets or queries the PCIe DATA Plus source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:DATAPlus:THRESHold:HIGh? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:DATAPlus:THRESHold:HIGh? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PCIE:DATAPlus:THRESHold:HIGh value command.
SCPI Syntax
- BUS:B<x>:PCIE:DATAPlus:THRESHold:HIGh <NR3>
- BUS:B<x>:PCIE:DATAPlus:THRESHold:HIGh?
Info
  • B<x> is the number of the bus.
  • <NR3> is the specified PCIe DATA Plus source threshold for the specified bus. The valid range is -8.0 V to 8.0 V.

BusBItemPcieMthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PCIE:MTHReshold command.

Description
  • This command sets or queries PCIe math source threshold for single ended for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:MTHReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:MTHReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PCIE:MTHReshold value command.
SCPI Syntax
- BUS:B<x>:PCIE:MTHReshold <NR3>
- BUS:B<x>:PCIE:MTHReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the specified PCIe math source threshold for the specified bus. The valid range is -8.0 V to 8.0 V.

BusBItemPcieSignaltype

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PCIE:SIGNALTYpe command.

Description
  • This command sets or queries the PCIe signal type for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:SIGNALTYpe? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:SIGNALTYpe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PCIE:SIGNALTYpe value command.
SCPI Syntax
- BUS:B<x>:PCIE:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:PCIE:SIGNALTYpe?
Info
  • B<x> is the number of the bus.
  • SINGLE specifies the Signal Type for PCIE bus as single.
  • DIFF specifies the Signal Type for PCIE bus as differential.

BusBItemPcieSource

Bases: SCPICmdRead

The BUS:B<x>:PCIE:SOUrce command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:SOUrce? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .dminus: The BUS:B<x>:PCIE:SOUrce:DMINus command.
  • .dplus: The BUS:B<x>:PCIE:SOUrce:DPLUs command.

dminus property

Return the BUS:B<x>:PCIE:SOUrce:DMINus command.

Description
  • This command sets or queries the PCIe DataMinus (SDATAMINUS) source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:SOUrce:DMINus? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:SOUrce:DMINus? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PCIE:SOUrce:DMINus value command.
SCPI Syntax
- BUS:B<x>:PCIE:SOUrce:DMINus {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:PCIE:SOUrce:DMINus?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a digital reference waveform as the source.

dplus property

Return the BUS:B<x>:PCIE:SOUrce:DPLUs command.

Description
  • This command sets or queries the PCIe dataPlus (SDATAPLUS) source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:SOUrce:DPLUs? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:SOUrce:DPLUs? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PCIE:SOUrce:DPLUs value command.
SCPI Syntax
- BUS:B<x>:PCIE:SOUrce:DPLUs {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:PCIE:SOUrce:DPLUs?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a digital reference waveform as the source.

BusBItemPcieSourceDminus

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PCIE:SOUrce:DMINus command.

Description
  • This command sets or queries the PCIe DataMinus (SDATAMINUS) source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:SOUrce:DMINus? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:SOUrce:DMINus? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PCIE:SOUrce:DMINus value command.
SCPI Syntax
- BUS:B<x>:PCIE:SOUrce:DMINus {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:PCIE:SOUrce:DMINus?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a digital reference waveform as the source.

BusBItemPcieSourceDplus

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PCIE:SOUrce:DPLUs command.

Description
  • This command sets or queries the PCIe dataPlus (SDATAPLUS) source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PCIE:SOUrce:DPLUs? query.
  • Using the .verify(value) method will send the BUS:B<x>:PCIE:SOUrce:DPLUs? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PCIE:SOUrce:DPLUs value command.
SCPI Syntax
- BUS:B<x>:PCIE:SOUrce:DPLUs {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:PCIE:SOUrce:DPLUs?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a digital reference waveform as the source.

BusBItemPsifive

Bases: SCPICmdRead

The BUS:B<x>:PSIFIVe command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .bitperiod: The BUS:B<x>:PSIFIVe:BITPERiod command.
  • .bitrate: The BUS:B<x>:PSIFIVe:BITRate command.
  • .comm: The BUS:B<x>:PSIFIVe:COMM command tree.
  • .dataa: The BUS:B<x>:PSIFIVe:DATAA command.
  • .datab: The BUS:B<x>:PSIFIVe:DATAB command.
  • .dataformat: The BUS:B<x>:PSIFIVe:DATAFORMat command.
  • .ecusource: The BUS:B<x>:PSIFIVe:ECUSOURce command.
  • .framecontrol: The BUS:B<x>:PSIFIVe:FRAMECONTrol command.
  • .messaging: The BUS:B<x>:PSIFIVe:MESSaging command.
  • .source: The BUS:B<x>:PSIFIVe:SOUrce command.
  • .status: The BUS:B<x>:PSIFIVe:STATus command.
  • .syncmode: The BUS:B<x>:PSIFIVe:SYNCMODe command.
  • .syncthreshold: The BUS:B<x>:PSIFIVe:SYNCTHRESHold command.
  • .threshold: The BUS:B<x>:PSIFIVe:THRESHold command.

bitperiod property

Return the BUS:B<x>:PSIFIVe:BITPERiod command.

Description
  • This command sets or queries the PSI5 Bit period bus parameter.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:BITPERiod? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:BITPERiod? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:BITPERiod value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:BITPERiod <NR1>
- BUS:B<x>:PSIFIVe:BITPERiod?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the bit period. The default bit period is 60 Micro seconds.

bitrate property

Return the BUS:B<x>:PSIFIVe:BITRate command.

Description
  • This command sets or queries the PSI5 bitrate.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:BITRate value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:BITRate {RATE125K|RATE189K|RATE83K}
- BUS:B<x>:PSIFIVe:BITRate?
Info
  • B<x> is the number of the bus.
  • RATE125K sets the bit rate to 125 kb.
  • RATE189K sets the bit rate to 189 kb.
  • RATE83K sets the bit rate to 83 kb.

comm property

Return the BUS:B<x>:PSIFIVe:COMM command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:COMM? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:COMM? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .direction: The BUS:B<x>:PSIFIVe:COMM:DIRection command.

dataa property

Return the BUS:B<x>:PSIFIVe:DATAA command.

Description
  • This command sets or queries the PSI5 frame mandatory data region A.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:DATAA? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:DATAA? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:DATAA value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:DATAA <NR1>
- BUS:B<x>:PSIFIVe:DATAA?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the dataA value in bits. The default dataA value is 10 bits, otherwise it ranges between 10 to 24 bits.

datab property

Return the BUS:B<x>:PSIFIVe:DATAB command.

Description
  • This command sets or queries the PSI5 frame optional bits of data region B.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:DATAB? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:DATAB? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:DATAB value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:DATAB <NR1>
- BUS:B<x>:PSIFIVe:DATAB?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the dataA value in bits. The default dataB value is 0 bits, otherwise it ranges between 0 to 12 bits.

dataformat property

Return the BUS:B<x>:PSIFIVe:DATAFORMat command.

Description
  • This command sets or queries the data format in PSI5 Frame2 packet.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:DATAFORMat? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:DATAFORMat? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:DATAFORMat value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:DATAFORMat {NIBBLe|BYTe}
- BUS:B<x>:PSIFIVe:DATAFORMat?
Info
  • B<x> is the number of the bus.
  • NIBBLe specifies the data format as Nibble.
  • BYTe specifies the data format as Byte.

ecusource property

Return the BUS:B<x>:PSIFIVe:ECUSOURce command.

Description
  • This command sets or queries the ECU to sensor source channel for the specified Bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:ECUSOURce? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:ECUSOURce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:ECUSOURce value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:ECUSOURce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:PSIFIVe:ECUSOURce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a digital reference waveform as the source.

framecontrol property

Return the BUS:B<x>:PSIFIVe:FRAMECONTrol command.

Description
  • This command sets or queries the PSI5 frame optional control bits.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:FRAMECONTrol? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:FRAMECONTrol? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:FRAMECONTrol value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:FRAMECONTrol <NR1>
- BUS:B<x>:PSIFIVe:FRAMECONTrol?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the frame control value in bits. The default frame control value is 0 bits, otherwise it ranges between 0 to 4 bits.

messaging property

Return the BUS:B<x>:PSIFIVe:MESSaging command.

Description
  • This command sets or queries the PSI5 optional messaging bits.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:MESSaging? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:MESSaging? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:MESSaging value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:MESSaging {OFF|ON}
- BUS:B<x>:PSIFIVe:MESSaging?
Info
  • B<x> is the number of the bus.
  • OFF specifies the messaging bits as the default value of 0.
  • ON specifies the messaging bits as 2.

source property

Return the BUS:B<x>:PSIFIVe:SOUrce command.

Description
  • This command sets or queries serial channel on or off.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:SOUrce value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:PSIFIVe:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a digital reference waveform as the source.

status property

Return the BUS:B<x>:PSIFIVe:STATus command.

Description
  • This command sets or queries the optional status bits of PSI5.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:STATus? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:STATus? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:STATus value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:STATus <NR1>
- BUS:B<x>:PSIFIVe:STATus?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the status value in bits. The default status value is 0 bits, otherwise it ranges between 0 to 2 bits.

syncmode property

Return the BUS:B<x>:PSIFIVe:SYNCMODe command.

Description
  • This command sets or queries the PSI5 Sync Mode.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:SYNCMODe? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:SYNCMODe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:SYNCMODe value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:SYNCMODe {PULSEWIDTh|TOOTHGAP}
- BUS:B<x>:PSIFIVe:SYNCMODe?
Info
  • B<x> is the number of the bus.
  • PULSEWIDTh specifies the Sync Mode as Pulse Width.
  • TOOTHGAP specifies the Sync Mode as Tooth Gap.

syncthreshold property

Return the BUS:B<x>:PSIFIVe:SYNCTHRESHold command.

Description
  • This command sets or queries the PSI5 threshold for the ECU To Sensor specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:SYNCTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:SYNCTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:SYNCTHRESHold value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:SYNCTHRESHold <NR3>
- BUS:B<x>:PSIFIVe:SYNCTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> specifies the PSI5 Strobe threshold for the specified bus. The threshold range is -8 V to +8 V.

threshold property

Return the BUS:B<x>:PSIFIVe:THRESHold command.

Description
  • This command sets or queries the PSI5 threshold for the Sensor To ECU specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:THRESHold value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:THRESHold <NR3>
- BUS:B<x>:PSIFIVe:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> specifies the PSI5 Strobe threshold for the specified bus. The threshold range is -8 V to +8 V.

BusBItemPsifiveBitperiod

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PSIFIVe:BITPERiod command.

Description
  • This command sets or queries the PSI5 Bit period bus parameter.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:BITPERiod? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:BITPERiod? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:BITPERiod value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:BITPERiod <NR1>
- BUS:B<x>:PSIFIVe:BITPERiod?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the bit period. The default bit period is 60 Micro seconds.

BusBItemPsifiveBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PSIFIVe:BITRate command.

Description
  • This command sets or queries the PSI5 bitrate.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:BITRate value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:BITRate {RATE125K|RATE189K|RATE83K}
- BUS:B<x>:PSIFIVe:BITRate?
Info
  • B<x> is the number of the bus.
  • RATE125K sets the bit rate to 125 kb.
  • RATE189K sets the bit rate to 189 kb.
  • RATE83K sets the bit rate to 83 kb.

BusBItemPsifiveComm

Bases: SCPICmdRead

The BUS:B<x>:PSIFIVe:COMM command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:COMM? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:COMM? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .direction: The BUS:B<x>:PSIFIVe:COMM:DIRection command.

direction property

Return the BUS:B<x>:PSIFIVe:COMM:DIRection command.

Description
  • This command sets or queries the PSI5 bus communication direction. Communication direction by default is set to Sensor to ECU.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:COMM:DIRection? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:COMM:DIRection? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:COMM:DIRection value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:COMM:DIRection {SENSORECU|ECUSENSor}
- BUS:B<x>:PSIFIVe:COMM:DIRection?
Info
  • B<x> is the number of the bus.
  • SENSORECU specifies the communication direction to Sensor to ECU. This is the default value.
  • ECUSENSor specifies the communication direction to ECU to Sensor.

BusBItemPsifiveCommDirection

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PSIFIVe:COMM:DIRection command.

Description
  • This command sets or queries the PSI5 bus communication direction. Communication direction by default is set to Sensor to ECU.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:COMM:DIRection? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:COMM:DIRection? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:COMM:DIRection value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:COMM:DIRection {SENSORECU|ECUSENSor}
- BUS:B<x>:PSIFIVe:COMM:DIRection?
Info
  • B<x> is the number of the bus.
  • SENSORECU specifies the communication direction to Sensor to ECU. This is the default value.
  • ECUSENSor specifies the communication direction to ECU to Sensor.

BusBItemPsifiveDataa

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PSIFIVe:DATAA command.

Description
  • This command sets or queries the PSI5 frame mandatory data region A.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:DATAA? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:DATAA? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:DATAA value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:DATAA <NR1>
- BUS:B<x>:PSIFIVe:DATAA?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the dataA value in bits. The default dataA value is 10 bits, otherwise it ranges between 10 to 24 bits.

BusBItemPsifiveDatab

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PSIFIVe:DATAB command.

Description
  • This command sets or queries the PSI5 frame optional bits of data region B.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:DATAB? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:DATAB? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:DATAB value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:DATAB <NR1>
- BUS:B<x>:PSIFIVe:DATAB?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the dataA value in bits. The default dataB value is 0 bits, otherwise it ranges between 0 to 12 bits.

BusBItemPsifiveDataformat

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PSIFIVe:DATAFORMat command.

Description
  • This command sets or queries the data format in PSI5 Frame2 packet.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:DATAFORMat? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:DATAFORMat? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:DATAFORMat value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:DATAFORMat {NIBBLe|BYTe}
- BUS:B<x>:PSIFIVe:DATAFORMat?
Info
  • B<x> is the number of the bus.
  • NIBBLe specifies the data format as Nibble.
  • BYTe specifies the data format as Byte.

BusBItemPsifiveEcusource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PSIFIVe:ECUSOURce command.

Description
  • This command sets or queries the ECU to sensor source channel for the specified Bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:ECUSOURce? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:ECUSOURce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:ECUSOURce value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:ECUSOURce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:PSIFIVe:ECUSOURce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a digital reference waveform as the source.

BusBItemPsifiveFramecontrol

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PSIFIVe:FRAMECONTrol command.

Description
  • This command sets or queries the PSI5 frame optional control bits.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:FRAMECONTrol? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:FRAMECONTrol? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:FRAMECONTrol value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:FRAMECONTrol <NR1>
- BUS:B<x>:PSIFIVe:FRAMECONTrol?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the frame control value in bits. The default frame control value is 0 bits, otherwise it ranges between 0 to 4 bits.

BusBItemPsifiveMessaging

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PSIFIVe:MESSaging command.

Description
  • This command sets or queries the PSI5 optional messaging bits.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:MESSaging? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:MESSaging? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:MESSaging value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:MESSaging {OFF|ON}
- BUS:B<x>:PSIFIVe:MESSaging?
Info
  • B<x> is the number of the bus.
  • OFF specifies the messaging bits as the default value of 0.
  • ON specifies the messaging bits as 2.

BusBItemPsifiveSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PSIFIVe:SOUrce command.

Description
  • This command sets or queries serial channel on or off.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:SOUrce value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:PSIFIVe:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a digital reference waveform as the source.

BusBItemPsifiveStatus

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PSIFIVe:STATus command.

Description
  • This command sets or queries the optional status bits of PSI5.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:STATus? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:STATus? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:STATus value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:STATus <NR1>
- BUS:B<x>:PSIFIVe:STATus?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the status value in bits. The default status value is 0 bits, otherwise it ranges between 0 to 2 bits.

BusBItemPsifiveSyncmode

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PSIFIVe:SYNCMODe command.

Description
  • This command sets or queries the PSI5 Sync Mode.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:SYNCMODe? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:SYNCMODe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:SYNCMODe value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:SYNCMODe {PULSEWIDTh|TOOTHGAP}
- BUS:B<x>:PSIFIVe:SYNCMODe?
Info
  • B<x> is the number of the bus.
  • PULSEWIDTh specifies the Sync Mode as Pulse Width.
  • TOOTHGAP specifies the Sync Mode as Tooth Gap.

BusBItemPsifiveSyncthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PSIFIVe:SYNCTHRESHold command.

Description
  • This command sets or queries the PSI5 threshold for the ECU To Sensor specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:SYNCTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:SYNCTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:SYNCTHRESHold value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:SYNCTHRESHold <NR3>
- BUS:B<x>:PSIFIVe:SYNCTHRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> specifies the PSI5 Strobe threshold for the specified bus. The threshold range is -8 V to +8 V.

BusBItemPsifiveThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:PSIFIVe:THRESHold command.

Description
  • This command sets or queries the PSI5 threshold for the Sensor To ECU specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:PSIFIVe:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:PSIFIVe:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:PSIFIVe:THRESHold value command.
SCPI Syntax
- BUS:B<x>:PSIFIVe:THRESHold <NR3>
- BUS:B<x>:PSIFIVe:THRESHold?
Info
  • B<x> is the number of the bus.
  • <NR3> specifies the PSI5 Strobe threshold for the specified bus. The threshold range is -8 V to +8 V.

BusBItemRs232c

Bases: SCPICmdRead

The BUS:B<x>:RS232C command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:RS232C? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .bitorder: The BUS:B<x>:RS232C:BITORDer command.
  • .bitrate: The BUS:B<x>:RS232C:BITRate command.
  • .databits: The BUS:B<x>:RS232C:DATABits command.
  • .delimiter: The BUS:B<x>:RS232C:DELIMiter command.
  • .displaymode: The BUS:B<x>:RS232C:DISplaymode command.
  • .parity: The BUS:B<x>:RS232C:PARity command.
  • .polarity: The BUS:B<x>:RS232C:POLarity command.
  • .source: The BUS:B<x>:RS232C:SOUrce command.

bitorder property

Return the BUS:B<x>:RS232C:BITORDer command.

Description
  • This command sets or queries the bit order for the specified RS-232 bus.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:BITORDer? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:BITORDer? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:BITORDer value command.
SCPI Syntax
- BUS:B<x>:RS232C:BITORDer {LSB|MSB}
- BUS:B<x>:RS232C:BITORDer?
Info
  • B<x> is the number of the bus.
  • LSB sets the bit order to LSB. This is the default value.
  • MSB sets the bit order to MSB.

bitrate property

Return the BUS:B<x>:RS232C:BITRate command.

Description
  • This command sets or queries the RS-232C bit rate for bus, where the bus number is specified by x. If you select Custom, use BUS:B<x>:RS232C:BITRate:CUSTom to set the bit rate.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:BITRate value command.
SCPI Syntax
- BUS:B<x>:RS232C:BITRate {CUSTOM|RATE300|RATE1K|RATE2K|RATE9K|RATE19K|RATE38K|RATE115K|RATE921K}
- BUS:B<x>:RS232C:BITRate?
Info
  • B<x> is the number of the bus.
  • CUSTOM sets a custom bit rate.
  • RATE300 sets the bit rate to 300 bits-per-second.
  • RATE1K sets the bit rate to 1000 bits-per-second.
  • RATE2K sets the bit rate to 2000 bits-per-second.
  • RATE9K sets the bit rate to 9000 bits-per-second.
  • RATE19K sets the bit rate to 19000 bits-per-second.
  • RATE38K sets the bit rate to 38000 bits-per-second.
  • RATE115K sets the bit rate to 115000 bits-per-second.
  • RATE921K sets the bit rate to 921000 bits-per-second.
Sub-properties
  • .custom: The BUS:B<x>:RS232C:BITRate:CUSTom command.

databits property

Return the BUS:B<x>:RS232C:DATABits command.

Description
  • This command sets or queries the RS-232C data width for bus, where the bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:DATABits? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:DATABits? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:DATABits value command.
SCPI Syntax
- BUS:B<x>:RS232C:DATABits {7|8|9}
- BUS:B<x>:RS232C:DATABits?
Info
  • B<x> is the number of the bus.
  • 7 specifies the number of bits as 7 in the RS-232C data frame.
  • 8 specifies the number of bits as 8 in the RS-232C data frame.
  • 9 specifies the number of bits as 9 in the RS-232C data frame.

delimiter property

Return the BUS:B<x>:RS232C:DELIMiter command.

Description
  • This command sets or queries the RS-232C string delimiter on bus , where the bus number is specified by x. This command only applies when Packet view is turned On.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:DELIMiter? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:DELIMiter? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:DELIMiter value command.
SCPI Syntax
- BUS:B<x>:RS232C:DELIMiter {NULl|CR|LF|SPace|XFF}
- BUS:B<x>:RS232C:DELIMiter?
Info
  • B<x> is the number of the bus.
  • NULl specifies NULL (0x00) delimiting value for a packet.
  • CR specifies CR (0x0D) delimiting value for a packet.
  • LF specifies LF (0x0A) delimiting value for a packet.
  • XFF specifies XFF (0xFF) delimiting value for a packet.
  • SPace specifies SPace delimiting value for a packet.

displaymode property

Return the BUS:B<x>:RS232C:DISplaymode command.

Description
  • This command sets or queries the RS-232C display mode for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:DISplaymode? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:DISplaymode? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:DISplaymode value command.
SCPI Syntax
- BUS:B<x>:RS232C:DISplaymode {FRame|PACKET}
- BUS:B<x>:RS232C:DISplaymode?
Info
  • B<x> is the number of the bus.
  • FRame displays each frame as a single entity.
  • PACKET displays a group of frames terminated with a single frame defined by the BUS:B<x>:RS232C:DELImiter command.

parity property

Return the BUS:B<x>:RS232C:PARity command.

Description
  • This command sets or queries the RS-232C parity for bus , where the bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:PARity? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:PARity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:PARity value command.
SCPI Syntax
- BUS:B<x>:RS232C:PARity {NONe|EVEN|ODD}
- BUS:B<x>:RS232C:PARity?
Info
  • B<x> is the number of the bus.
  • NONe specifies no parity.
  • EVEN specifies even parity.
  • ODD specifies odd parity.

polarity property

Return the BUS:B<x>:RS232C:POLarity command.

Description
  • This command sets or queries the RS-232C source polarity for bus , where the bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:POLarity value command.
SCPI Syntax
- BUS:B<x>:RS232C:POLarity {NORmal|INVERTed}
- BUS:B<x>:RS232C:POLarity?
Info
  • B<x> is the number of the bus.
  • NORmal sets the RS-232C bus polarity to positive.
  • INVERTed sets the RS-232C bus polarity to negative.

source property

Return the BUS:B<x>:RS232C:SOUrce command.

Description
  • This command sets or queries the RS-232C source for bus , where the bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:SOUrce value command.
SCPI Syntax
- BUS:B<x>:RS232C:SOUrce {CH<x>|REF<x>|MATH<x>|REF<x>_D<x>}
- BUS:B<x>:RS232C:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use as the RS-232C source.
  • MATH<x> specifies a math channel to use for the RS-232C source.
  • REF<x> specifies a reference channel to use for the RS-232C source.
  • REF<x>_D<x> specifies a digital reference waveform as the source waveform for the specified RS-232C bus.
Sub-properties
  • .threshold: The BUS:B<x>:RS232C:SOUrce:THReshold command.

BusBItemRs232cBitorder

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:RS232C:BITORDer command.

Description
  • This command sets or queries the bit order for the specified RS-232 bus.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:BITORDer? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:BITORDer? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:BITORDer value command.
SCPI Syntax
- BUS:B<x>:RS232C:BITORDer {LSB|MSB}
- BUS:B<x>:RS232C:BITORDer?
Info
  • B<x> is the number of the bus.
  • LSB sets the bit order to LSB. This is the default value.
  • MSB sets the bit order to MSB.

BusBItemRs232cBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:RS232C:BITRate command.

Description
  • This command sets or queries the RS-232C bit rate for bus, where the bus number is specified by x. If you select Custom, use BUS:B<x>:RS232C:BITRate:CUSTom to set the bit rate.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:BITRate value command.
SCPI Syntax
- BUS:B<x>:RS232C:BITRate {CUSTOM|RATE300|RATE1K|RATE2K|RATE9K|RATE19K|RATE38K|RATE115K|RATE921K}
- BUS:B<x>:RS232C:BITRate?
Info
  • B<x> is the number of the bus.
  • CUSTOM sets a custom bit rate.
  • RATE300 sets the bit rate to 300 bits-per-second.
  • RATE1K sets the bit rate to 1000 bits-per-second.
  • RATE2K sets the bit rate to 2000 bits-per-second.
  • RATE9K sets the bit rate to 9000 bits-per-second.
  • RATE19K sets the bit rate to 19000 bits-per-second.
  • RATE38K sets the bit rate to 38000 bits-per-second.
  • RATE115K sets the bit rate to 115000 bits-per-second.
  • RATE921K sets the bit rate to 921000 bits-per-second.
Properties
  • .custom: The BUS:B<x>:RS232C:BITRate:CUSTom command.

custom property

Return the BUS:B<x>:RS232C:BITRate:CUSTom command.

Description
  • This command sets or queries the RS-232C custom bit rate for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:BITRate:CUSTom? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:BITRate:CUSTom? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:BITRate:CUSTom value command.
SCPI Syntax
- BUS:B<x>:RS232C:BITRate:CUSTom <NR1>
- BUS:B<x>:RS232C:BITRate:CUSTom?
Info
  • B<x> is the number of the bus.
  • <NR1> is the custom bit rate.

BusBItemRs232cBitrateCustom

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:RS232C:BITRate:CUSTom command.

Description
  • This command sets or queries the RS-232C custom bit rate for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:BITRate:CUSTom? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:BITRate:CUSTom? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:BITRate:CUSTom value command.
SCPI Syntax
- BUS:B<x>:RS232C:BITRate:CUSTom <NR1>
- BUS:B<x>:RS232C:BITRate:CUSTom?
Info
  • B<x> is the number of the bus.
  • <NR1> is the custom bit rate.

BusBItemRs232cDatabits

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:RS232C:DATABits command.

Description
  • This command sets or queries the RS-232C data width for bus, where the bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:DATABits? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:DATABits? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:DATABits value command.
SCPI Syntax
- BUS:B<x>:RS232C:DATABits {7|8|9}
- BUS:B<x>:RS232C:DATABits?
Info
  • B<x> is the number of the bus.
  • 7 specifies the number of bits as 7 in the RS-232C data frame.
  • 8 specifies the number of bits as 8 in the RS-232C data frame.
  • 9 specifies the number of bits as 9 in the RS-232C data frame.

BusBItemRs232cDelimiter

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:RS232C:DELIMiter command.

Description
  • This command sets or queries the RS-232C string delimiter on bus , where the bus number is specified by x. This command only applies when Packet view is turned On.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:DELIMiter? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:DELIMiter? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:DELIMiter value command.
SCPI Syntax
- BUS:B<x>:RS232C:DELIMiter {NULl|CR|LF|SPace|XFF}
- BUS:B<x>:RS232C:DELIMiter?
Info
  • B<x> is the number of the bus.
  • NULl specifies NULL (0x00) delimiting value for a packet.
  • CR specifies CR (0x0D) delimiting value for a packet.
  • LF specifies LF (0x0A) delimiting value for a packet.
  • XFF specifies XFF (0xFF) delimiting value for a packet.
  • SPace specifies SPace delimiting value for a packet.

BusBItemRs232cDisplaymode

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:RS232C:DISplaymode command.

Description
  • This command sets or queries the RS-232C display mode for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:DISplaymode? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:DISplaymode? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:DISplaymode value command.
SCPI Syntax
- BUS:B<x>:RS232C:DISplaymode {FRame|PACKET}
- BUS:B<x>:RS232C:DISplaymode?
Info
  • B<x> is the number of the bus.
  • FRame displays each frame as a single entity.
  • PACKET displays a group of frames terminated with a single frame defined by the BUS:B<x>:RS232C:DELImiter command.

BusBItemRs232cParity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:RS232C:PARity command.

Description
  • This command sets or queries the RS-232C parity for bus , where the bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:PARity? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:PARity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:PARity value command.
SCPI Syntax
- BUS:B<x>:RS232C:PARity {NONe|EVEN|ODD}
- BUS:B<x>:RS232C:PARity?
Info
  • B<x> is the number of the bus.
  • NONe specifies no parity.
  • EVEN specifies even parity.
  • ODD specifies odd parity.

BusBItemRs232cPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:RS232C:POLarity command.

Description
  • This command sets or queries the RS-232C source polarity for bus , where the bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:POLarity value command.
SCPI Syntax
- BUS:B<x>:RS232C:POLarity {NORmal|INVERTed}
- BUS:B<x>:RS232C:POLarity?
Info
  • B<x> is the number of the bus.
  • NORmal sets the RS-232C bus polarity to positive.
  • INVERTed sets the RS-232C bus polarity to negative.

BusBItemRs232cSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:RS232C:SOUrce command.

Description
  • This command sets or queries the RS-232C source for bus , where the bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:SOUrce value command.
SCPI Syntax
- BUS:B<x>:RS232C:SOUrce {CH<x>|REF<x>|MATH<x>|REF<x>_D<x>}
- BUS:B<x>:RS232C:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel to use as the RS-232C source.
  • MATH<x> specifies a math channel to use for the RS-232C source.
  • REF<x> specifies a reference channel to use for the RS-232C source.
  • REF<x>_D<x> specifies a digital reference waveform as the source waveform for the specified RS-232C bus.
Properties
  • .threshold: The BUS:B<x>:RS232C:SOUrce:THReshold command.

threshold property

Return the BUS:B<x>:RS232C:SOUrce:THReshold command.

Description
  • This command sets or queries the RS-232C source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:SOUrce:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:SOUrce:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:SOUrce:THReshold value command.
SCPI Syntax
- BUS:B<x>:RS232C:SOUrce:THReshold <NR3>
- BUS:B<x>:RS232C:SOUrce:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the RS-232C source threshold for the specified bus.

BusBItemRs232cSourceThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:RS232C:SOUrce:THReshold command.

Description
  • This command sets or queries the RS-232C source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:RS232C:SOUrce:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:RS232C:SOUrce:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:RS232C:SOUrce:THReshold value command.
SCPI Syntax
- BUS:B<x>:RS232C:SOUrce:THReshold <NR3>
- BUS:B<x>:RS232C:SOUrce:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the RS-232C source threshold for the specified bus.

BusBItemS8b10b

Bases: SCPICmdRead

The BUS:B<x>:S8B10B command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:S8B10B? query.
  • Using the .verify(value) method will send the BUS:B<x>:S8B10B? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus. Arguments are available at bit rates up to 1 Tbps.
Properties
  • .bitrate: The BUS:B<x>:S8B10B:BITRate command.
  • .source: The BUS:B<x>:S8B10B:SOUrce command.
  • .threshold: The BUS:B<x>:S8B10B:THReshold command.

bitrate property

Return the BUS:B<x>:S8B10B:BITRate command.

Description
  • This command sets or queries the 8B10b bit rate for the specified bus, where the bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:S8B10B:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:S8B10B:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write() method will send the BUS:B<x>:S8B10B:BITRate command.
SCPI Syntax
- BUS:B<x>:S8B10B:BITRate
- BUS:B<x>:S8B10B:BITRate?
Info
  • B<x> is the number of the bus. Arguments are available at bit rates up to 1 Tbps.

source property

Return the BUS:B<x>:S8B10B:SOUrce command.

Description
  • This command sets or queries the 8B10b source for the specified bus, where the bus is specified by x. This command specifies the source channel.
Usage
  • Using the .query() method will send the BUS:B<x>:S8B10B:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:S8B10B:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:S8B10B:SOUrce value command.
SCPI Syntax
- BUS:B<x>:S8B10B:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:S8B10B:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a digital reference waveform as the source.

threshold property

Return the BUS:B<x>:S8B10B:THReshold command.

Description
  • This command sets or queries the 8B10b threshold for the specified bus, where the bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:S8B10B:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:S8B10B:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:S8B10B:THReshold value command.
SCPI Syntax
- BUS:B<x>:S8B10B:THReshold <NR3>
- BUS:B<x>:S8B10B:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the 8B10b Strobe threshold for the specified bus in volts. The valid range is -8V to +8V.

BusBItemS8b10bBitrate

Bases: SCPICmdWriteNoArguments, SCPICmdRead

The BUS:B<x>:S8B10B:BITRate command.

Description
  • This command sets or queries the 8B10b bit rate for the specified bus, where the bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:S8B10B:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:S8B10B:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write() method will send the BUS:B<x>:S8B10B:BITRate command.
SCPI Syntax
- BUS:B<x>:S8B10B:BITRate
- BUS:B<x>:S8B10B:BITRate?
Info
  • B<x> is the number of the bus. Arguments are available at bit rates up to 1 Tbps.

BusBItemS8b10bSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:S8B10B:SOUrce command.

Description
  • This command sets or queries the 8B10b source for the specified bus, where the bus is specified by x. This command specifies the source channel.
Usage
  • Using the .query() method will send the BUS:B<x>:S8B10B:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:S8B10B:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:S8B10B:SOUrce value command.
SCPI Syntax
- BUS:B<x>:S8B10B:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:S8B10B:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a digital reference waveform as the source.

BusBItemS8b10bThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:S8B10B:THReshold command.

Description
  • This command sets or queries the 8B10b threshold for the specified bus, where the bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:S8B10B:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:S8B10B:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:S8B10B:THReshold value command.
SCPI Syntax
- BUS:B<x>:S8B10B:THReshold <NR3>
- BUS:B<x>:S8B10B:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the 8B10b Strobe threshold for the specified bus in volts. The valid range is -8V to +8V.

BusBItemSdlc

Bases: SCPICmdRead

The BUS:B<x>:SDLC command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SDLC? query.
  • Using the .verify(value) method will send the BUS:B<x>:SDLC? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .bitrate: The BUS:B<x>:SDLC:BITRate command.
  • .data: The BUS:B<x>:SDLC:DATA command tree.
  • .encoding: The BUS:B<x>:SDLC:ENCoding command.
  • .modulo: The BUS:B<x>:SDLC:MODulo command.

bitrate property

Return the BUS:B<x>:SDLC:BITRate command.

Description
  • This command sets or queries the bit rate for the specified SDLC bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SDLC:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:SDLC:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SDLC:BITRate value command.
SCPI Syntax
- BUS:B<x>:SDLC:BITRate <NR1>
- BUS:B<x>:SDLC:BITRate?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the bit rate. The default bit rate is 10 kbs and varies 300 ~ 1000000000.

data property

Return the BUS:B<x>:SDLC:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SDLC:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:SDLC:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Sub-properties
  • .source: The BUS:B<x>:SDLC:DATA:SOUrce command.
  • .threshold: The BUS:B<x>:SDLC:DATA:THReshold command.

encoding property

Return the BUS:B<x>:SDLC:ENCoding command.

Description
  • This command sets or queries the SDLC Bus Encoding. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SDLC:ENCoding? query.
  • Using the .verify(value) method will send the BUS:B<x>:SDLC:ENCoding? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SDLC:ENCoding value command.
SCPI Syntax
- BUS:B<x>:SDLC:ENCoding {DISCrete|INVert}
- BUS:B<x>:SDLC:ENCoding?
Info
  • B<x> is the number of the bus waveform.
  • DISCrete specifies the encoding mechanism is Discrete Transmission (NRZ).
  • INVert specifies that encoding mechanism is Invert On Zero i.e. NRZI.

modulo property

Return the BUS:B<x>:SDLC:MODulo command.

Description
  • This command sets or queries the SDLC Bus Modulo. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SDLC:MODulo? query.
  • Using the .verify(value) method will send the BUS:B<x>:SDLC:MODulo? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SDLC:MODulo value command.
SCPI Syntax
- BUS:B<x>:SDLC:MODulo {8|128}
- BUS:B<x>:SDLC:MODulo?
Info
  • B<x> is the number of the bus waveform.
  • 8 specifies the C-Field size is 8 bit in SDLC frame.
  • 128 specifies the C-Field size is 16 bit in SDLC frame.

BusBItemSdlcBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SDLC:BITRate command.

Description
  • This command sets or queries the bit rate for the specified SDLC bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SDLC:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:SDLC:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SDLC:BITRate value command.
SCPI Syntax
- BUS:B<x>:SDLC:BITRate <NR1>
- BUS:B<x>:SDLC:BITRate?
Info
  • B<x> is the number of the bus.
  • <NR1> specifies the bit rate. The default bit rate is 10 kbs and varies 300 ~ 1000000000.

BusBItemSdlcData

Bases: SCPICmdRead

The BUS:B<x>:SDLC:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SDLC:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:SDLC:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus.
Properties
  • .source: The BUS:B<x>:SDLC:DATA:SOUrce command.
  • .threshold: The BUS:B<x>:SDLC:DATA:THReshold command.

source property

Return the BUS:B<x>:SDLC:DATA:SOUrce command.

Description
  • This command sets or queries the source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SDLC:DATA:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SDLC:DATA:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SDLC:DATA:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SDLC:DATA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SDLC:DATA:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a digital reference waveform as the source.

threshold property

Return the BUS:B<x>:SDLC:DATA:THReshold command.

Description
  • This command sets or queries the SDLC data source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SDLC:DATA:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SDLC:DATA:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SDLC:DATA:THReshold value command.
SCPI Syntax
- BUS:B<x>:SDLC:DATA:THReshold <NR3>
- BUS:B<x>:SDLC:DATA:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the SDLC Strobe threshold for the specified bus. The valid range is -8V to +8V.

BusBItemSdlcDataSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SDLC:DATA:SOUrce command.

Description
  • This command sets or queries the source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SDLC:DATA:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SDLC:DATA:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SDLC:DATA:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SDLC:DATA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SDLC:DATA:SOUrce?
Info
  • B<x> is the number of the bus.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math waveform as the source.
  • REF<x> specifies a digital reference waveform as the source.

BusBItemSdlcDataThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SDLC:DATA:THReshold command.

Description
  • This command sets or queries the SDLC data source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SDLC:DATA:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SDLC:DATA:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SDLC:DATA:THReshold value command.
SCPI Syntax
- BUS:B<x>:SDLC:DATA:THReshold <NR3>
- BUS:B<x>:SDLC:DATA:THReshold?
Info
  • B<x> is the number of the bus.
  • <NR3> is the SDLC Strobe threshold for the specified bus. The valid range is -8V to +8V.

BusBItemSdlcEncoding

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SDLC:ENCoding command.

Description
  • This command sets or queries the SDLC Bus Encoding. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SDLC:ENCoding? query.
  • Using the .verify(value) method will send the BUS:B<x>:SDLC:ENCoding? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SDLC:ENCoding value command.
SCPI Syntax
- BUS:B<x>:SDLC:ENCoding {DISCrete|INVert}
- BUS:B<x>:SDLC:ENCoding?
Info
  • B<x> is the number of the bus waveform.
  • DISCrete specifies the encoding mechanism is Discrete Transmission (NRZ).
  • INVert specifies that encoding mechanism is Invert On Zero i.e. NRZI.

BusBItemSdlcModulo

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SDLC:MODulo command.

Description
  • This command sets or queries the SDLC Bus Modulo. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SDLC:MODulo? query.
  • Using the .verify(value) method will send the BUS:B<x>:SDLC:MODulo? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SDLC:MODulo value command.
SCPI Syntax
- BUS:B<x>:SDLC:MODulo {8|128}
- BUS:B<x>:SDLC:MODulo?
Info
  • B<x> is the number of the bus waveform.
  • 8 specifies the C-Field size is 8 bit in SDLC frame.
  • 128 specifies the C-Field size is 16 bit in SDLC frame.

BusBItemSent

Bases: SCPICmdRead

The BUS:B<x>:SENT command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SENT? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .chanwidth: The BUS:B<x>:SENT:CHANWidth command.
  • .nibblecount: The BUS:B<x>:SENT:NIBBLECount command.
  • .numchannel: The BUS:B<x>:SENT:NUMCHANnel command.
  • .pausepulse: The BUS:B<x>:SENT:PAUSEPULSe command.
  • .polarity: The BUS:B<x>:SENT:POLARITY command.
  • .slow: The BUS:B<x>:SENT:SLOW command.
  • .source: The BUS:B<x>:SENT:SOUrce command.
  • .threshold: The BUS:B<x>:SENT:THRESHold command.
  • .ticktime: The BUS:B<x>:SENT:TICKTIME command.
  • .ticktolerance: The BUS:B<x>:SENT:TICKTOLerance command.

chanwidth property

Return the BUS:B<x>:SENT:CHANWidth command.

Description
  • This command sets or queries SENT fast channel bit widths for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:CHANWidth? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:CHANWidth? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:CHANWidth value command.
SCPI Syntax
- BUS:B<x>:SENT:CHANWidth {TWELVEtwelve|FOURTEENten|SIXTEENeight}
- BUS:B<x>:SENT:CHANWidth?
Info
  • B<x> is the number of the bus waveform.
  • TWELVEtwelve sets both Fast Channel 1 and Fast Channel 2 to12 bits wide.
  • FOURTEENten sets Fast Channel 1 to 14 bits and Fast Channel 2 to 10 bits.
  • SIXTEENeight sets Fast Channel 1 to 16 bits and Fast Channel 2 to eight bits.

nibblecount property

Return the BUS:B<x>:SENT:NIBBLECount command.

Description
  • This command sets or queries SENT data nibbles for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:NIBBLECount? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:NIBBLECount? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:NIBBLECount value command.
SCPI Syntax
- BUS:B<x>:SENT:NIBBLECount {THREE|FOUR|SIX}
- BUS:B<x>:SENT:NIBBLECount?
Info
  • B<x> is the number of the bus waveform.
  • THREE specifies regular data with 3 nibbles.
  • FOUR specifies regular data with 4nibbles.
  • SIX specifies regular data with 6nibbles.

numchannel property

Return the BUS:B<x>:SENT:NUMCHANnel command.

Description
  • This command sets or queries SENT fast data channels for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:NUMCHANnel? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:NUMCHANnel? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:NUMCHANnel value command.
SCPI Syntax
- BUS:B<x>:SENT:NUMCHANnel {ONE|TWO}
- BUS:B<x>:SENT:NUMCHANnel?
Info
  • B<x> is the number of the bus waveform.
  • ONE specifies a SENT bus with one fast channel.
  • TWO specifies a SENT bus with two fast channels.

pausepulse property

Return the BUS:B<x>:SENT:PAUSEPULSe command.

Description
  • This command sets or queries SENT pause pulse for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:PAUSEPULSe? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:PAUSEPULSe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:PAUSEPULSe value command.
SCPI Syntax
- BUS:B<x>:SENT:PAUSEPULSe {NO|YES}
- BUS:B<x>:SENT:PAUSEPULSe?
Info
  • B<x> is the number of the bus waveform.
  • NO specifies no pause pulse.
  • YES specifies a pause pulse is used.

polarity property

Return the BUS:B<x>:SENT:POLARITY command.

Description
  • This command sets or queries SENT Idle State signal polarity for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:POLARITY? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:POLARITY? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:POLARITY value command.
SCPI Syntax
- BUS:B<x>:SENT:POLARITY {INVerted|NORmal}
- BUS:B<x>:SENT:POLARITY?
Info
  • B<x> is the number of the bus waveform.
  • INVerted specifies inverted polarity.
  • NORmal specifies normal polarity.

slow property

Return the BUS:B<x>:SENT:SLOW command.

Description
  • This command sets or queries the SENT slow channel configuration for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:SLOW? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:SLOW? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:SLOW value command.
SCPI Syntax
- BUS:B<x>:SENT:SLOW {NOne|ENHANCED4|ENHANCED8|SHOrt}
- BUS:B<x>:SENT:SLOW?
Info
  • B<x> is the number of the bus waveform.
  • NOne specifies no slow channel configured.
  • ENHANCED4 specifies Enhanced 4 slow channel configuration.
  • ENHANCED8 specifies Enhanced 8 slow channel configuration.
  • SHOrt specifies short slow channel configuration.

source property

Return the BUS:B<x>:SENT:SOUrce command.

Description
  • This command sets or queries the SENT DATA source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SENT:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SENT:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the clock source waveform for the audio bus.
  • MATH<x> specifies a math waveform as the clock source waveform for the audio bus.
  • REF<x> specifies a reference waveform as the clock source waveform for the audio bus.
  • REF<x>_D<x> specifies a digital reference waveform as the clock source waveform for the specified audio bus.

threshold property

Return the BUS:B<x>:SENT:THRESHold command.

Description
  • This command sets or queries the SENT DATA source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:THRESHold value command.
SCPI Syntax
- BUS:B<x>:SENT:THRESHold <NR3>
- BUS:B<x>:SENT:THRESHold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> sets the data source threshold value in volts.

ticktime property

Return the BUS:B<x>:SENT:TICKTIME command.

Description
  • This command sets or queries the SENT bus Clock Tick parameter for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:TICKTIME? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:TICKTIME? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:TICKTIME value command.
SCPI Syntax
- BUS:B<x>:SENT:TICKTIME <NR3>
- BUS:B<x>:SENT:TICKTIME?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SENT clock tick time, in seconds.

ticktolerance property

Return the BUS:B<x>:SENT:TICKTOLerance command.

Description
  • This command sets or queries the SENT bus Tick Tolerance percent parameter for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:TICKTOLerance? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:TICKTOLerance? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:TICKTOLerance value command.
SCPI Syntax
- BUS:B<x>:SENT:TICKTOLerance <NR3>
- BUS:B<x>:SENT:TICKTOLerance?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the tick tolerance percentage.

BusBItemSentChanwidth

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SENT:CHANWidth command.

Description
  • This command sets or queries SENT fast channel bit widths for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:CHANWidth? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:CHANWidth? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:CHANWidth value command.
SCPI Syntax
- BUS:B<x>:SENT:CHANWidth {TWELVEtwelve|FOURTEENten|SIXTEENeight}
- BUS:B<x>:SENT:CHANWidth?
Info
  • B<x> is the number of the bus waveform.
  • TWELVEtwelve sets both Fast Channel 1 and Fast Channel 2 to12 bits wide.
  • FOURTEENten sets Fast Channel 1 to 14 bits and Fast Channel 2 to 10 bits.
  • SIXTEENeight sets Fast Channel 1 to 16 bits and Fast Channel 2 to eight bits.

BusBItemSentNibblecount

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SENT:NIBBLECount command.

Description
  • This command sets or queries SENT data nibbles for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:NIBBLECount? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:NIBBLECount? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:NIBBLECount value command.
SCPI Syntax
- BUS:B<x>:SENT:NIBBLECount {THREE|FOUR|SIX}
- BUS:B<x>:SENT:NIBBLECount?
Info
  • B<x> is the number of the bus waveform.
  • THREE specifies regular data with 3 nibbles.
  • FOUR specifies regular data with 4nibbles.
  • SIX specifies regular data with 6nibbles.

BusBItemSentNumchannel

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SENT:NUMCHANnel command.

Description
  • This command sets or queries SENT fast data channels for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:NUMCHANnel? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:NUMCHANnel? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:NUMCHANnel value command.
SCPI Syntax
- BUS:B<x>:SENT:NUMCHANnel {ONE|TWO}
- BUS:B<x>:SENT:NUMCHANnel?
Info
  • B<x> is the number of the bus waveform.
  • ONE specifies a SENT bus with one fast channel.
  • TWO specifies a SENT bus with two fast channels.

BusBItemSentPausepulse

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SENT:PAUSEPULSe command.

Description
  • This command sets or queries SENT pause pulse for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:PAUSEPULSe? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:PAUSEPULSe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:PAUSEPULSe value command.
SCPI Syntax
- BUS:B<x>:SENT:PAUSEPULSe {NO|YES}
- BUS:B<x>:SENT:PAUSEPULSe?
Info
  • B<x> is the number of the bus waveform.
  • NO specifies no pause pulse.
  • YES specifies a pause pulse is used.

BusBItemSentPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SENT:POLARITY command.

Description
  • This command sets or queries SENT Idle State signal polarity for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:POLARITY? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:POLARITY? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:POLARITY value command.
SCPI Syntax
- BUS:B<x>:SENT:POLARITY {INVerted|NORmal}
- BUS:B<x>:SENT:POLARITY?
Info
  • B<x> is the number of the bus waveform.
  • INVerted specifies inverted polarity.
  • NORmal specifies normal polarity.

BusBItemSentSlow

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SENT:SLOW command.

Description
  • This command sets or queries the SENT slow channel configuration for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:SLOW? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:SLOW? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:SLOW value command.
SCPI Syntax
- BUS:B<x>:SENT:SLOW {NOne|ENHANCED4|ENHANCED8|SHOrt}
- BUS:B<x>:SENT:SLOW?
Info
  • B<x> is the number of the bus waveform.
  • NOne specifies no slow channel configured.
  • ENHANCED4 specifies Enhanced 4 slow channel configuration.
  • ENHANCED8 specifies Enhanced 8 slow channel configuration.
  • SHOrt specifies short slow channel configuration.

BusBItemSentSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SENT:SOUrce command.

Description
  • This command sets or queries the SENT DATA source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SENT:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SENT:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the clock source waveform for the audio bus.
  • MATH<x> specifies a math waveform as the clock source waveform for the audio bus.
  • REF<x> specifies a reference waveform as the clock source waveform for the audio bus.
  • REF<x>_D<x> specifies a digital reference waveform as the clock source waveform for the specified audio bus.

BusBItemSentThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SENT:THRESHold command.

Description
  • This command sets or queries the SENT DATA source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:THRESHold value command.
SCPI Syntax
- BUS:B<x>:SENT:THRESHold <NR3>
- BUS:B<x>:SENT:THRESHold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> sets the data source threshold value in volts.

BusBItemSentTicktime

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SENT:TICKTIME command.

Description
  • This command sets or queries the SENT bus Clock Tick parameter for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:TICKTIME? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:TICKTIME? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:TICKTIME value command.
SCPI Syntax
- BUS:B<x>:SENT:TICKTIME <NR3>
- BUS:B<x>:SENT:TICKTIME?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SENT clock tick time, in seconds.

BusBItemSentTicktolerance

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SENT:TICKTOLerance command.

Description
  • This command sets or queries the SENT bus Tick Tolerance percent parameter for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SENT:TICKTOLerance? query.
  • Using the .verify(value) method will send the BUS:B<x>:SENT:TICKTOLerance? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SENT:TICKTOLerance value command.
SCPI Syntax
- BUS:B<x>:SENT:TICKTOLerance <NR3>
- BUS:B<x>:SENT:TICKTOLerance?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the tick tolerance percentage.

BusBItemSmbus

Bases: SCPICmdRead

The BUS:B<x>:SMBUS command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the bus number.
Properties
  • .clock: The BUS:B<x>:SMBUS:CLOCk command tree.
  • .data: The BUS:B<x>:SMBUS:DATA command tree.
  • .pec: The BUS:B<x>:SMBUS:PEC command tree.

clock property

Return the BUS:B<x>:SMBUS:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the bus number.
Sub-properties
  • .source: The BUS:B<x>:SMBUS:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:SMBUS:CLOCk:THReshold command.

data property

Return the BUS:B<x>:SMBUS:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the bus number.
Sub-properties
  • .source: The BUS:B<x>:SMBUS:DATA:SOUrce command.
  • .threshold: The BUS:B<x>:SMBUS:DATA:THReshold command.

pec property

Return the BUS:B<x>:SMBUS:PEC command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS:PEC? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS:PEC? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the bus number.
Sub-properties
  • .value: The BUS:B<x>:SMBUS:PEC:VALUe command.

BusBItemSmbusClock

Bases: SCPICmdRead

The BUS:B<x>:SMBUS:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the bus number.
Properties
  • .source: The BUS:B<x>:SMBUS:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:SMBUS:CLOCk:THReshold command.

source property

Return the BUS:B<x>:SMBUS:CLOCk:SOUrce command.

Description
  • This command sets or queries the clock source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SMBUS:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SMBUS:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SMBUS:CLOCk:SOUrce?
Info
  • B<x> is the bus number.
  • CH<x> specifies an analog channel as the source for the data signal, where is the channel number.
  • MATH<x> specifies a math channel as the source for the data signal, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source for the data signal, where is the reference waveform number.

threshold property

Return the BUS:B<x>:SMBUS:CLOCk:THReshold command.

Description
  • This command sets or queries the SMBUS clock source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS:CLOCk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS:CLOCk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SMBUS:CLOCk:THReshold value command.
SCPI Syntax
- BUS:B<x>:SMBUS:CLOCk:THReshold <NR3>
- BUS:B<x>:SMBUS:CLOCk:THReshold?
Info
  • B<x> is the bus number.
  • <NR3> specifies the SMBUS clock threshold for the specified bus. The valid range is -8V to +8V.

BusBItemSmbusClockSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SMBUS:CLOCk:SOUrce command.

Description
  • This command sets or queries the clock source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SMBUS:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SMBUS:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SMBUS:CLOCk:SOUrce?
Info
  • B<x> is the bus number.
  • CH<x> specifies an analog channel as the source for the data signal, where is the channel number.
  • MATH<x> specifies a math channel as the source for the data signal, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source for the data signal, where is the reference waveform number.

BusBItemSmbusClockThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SMBUS:CLOCk:THReshold command.

Description
  • This command sets or queries the SMBUS clock source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS:CLOCk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS:CLOCk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SMBUS:CLOCk:THReshold value command.
SCPI Syntax
- BUS:B<x>:SMBUS:CLOCk:THReshold <NR3>
- BUS:B<x>:SMBUS:CLOCk:THReshold?
Info
  • B<x> is the bus number.
  • <NR3> specifies the SMBUS clock threshold for the specified bus. The valid range is -8V to +8V.

BusBItemSmbusData

Bases: SCPICmdRead

The BUS:B<x>:SMBUS:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the bus number.
Properties
  • .source: The BUS:B<x>:SMBUS:DATA:SOUrce command.
  • .threshold: The BUS:B<x>:SMBUS:DATA:THReshold command.

source property

Return the BUS:B<x>:SMBUS:DATA:SOUrce command.

Description
  • This command sets or queries the data source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS:DATA:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS:DATA:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SMBUS:DATA:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SMBUS:DATA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SMBUS:DATA:SOUrce?
Info
  • B<x> is the bus number.
  • CH<x> specifies an analog channel as the source for the data signal, where is the channel number.
  • MATH<x> specifies a math channel as the source for the data signal, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source for the data signal, where is the reference waveform number.

threshold property

Return the BUS:B<x>:SMBUS:DATA:THReshold command.

Description
  • This command sets or queries the SMBUS data source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS:DATA:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS:DATA:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SMBUS:DATA:THReshold value command.
SCPI Syntax
- BUS:B<x>:SMBUS:DATA:THReshold <NR3>
- BUS:B<x>:SMBUS:DATA:THReshold?
Info
  • B<x> is the bus number.
  • <NR3> specifies the SMBUS data threshold for the specified bus. The valid range is -8V to +8V.

BusBItemSmbusDataSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SMBUS:DATA:SOUrce command.

Description
  • This command sets or queries the data source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS:DATA:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS:DATA:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SMBUS:DATA:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SMBUS:DATA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SMBUS:DATA:SOUrce?
Info
  • B<x> is the bus number.
  • CH<x> specifies an analog channel as the source for the data signal, where is the channel number.
  • MATH<x> specifies a math channel as the source for the data signal, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source for the data signal, where is the reference waveform number.

BusBItemSmbusDataThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SMBUS:DATA:THReshold command.

Description
  • This command sets or queries the SMBUS data source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS:DATA:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS:DATA:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SMBUS:DATA:THReshold value command.
SCPI Syntax
- BUS:B<x>:SMBUS:DATA:THReshold <NR3>
- BUS:B<x>:SMBUS:DATA:THReshold?
Info
  • B<x> is the bus number.
  • <NR3> specifies the SMBUS data threshold for the specified bus. The valid range is -8V to +8V.

BusBItemSmbusPec

Bases: SCPICmdRead

The BUS:B<x>:SMBUS:PEC command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS:PEC? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS:PEC? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the bus number.
Properties
  • .value: The BUS:B<x>:SMBUS:PEC:VALUe command.

value property

Return the BUS:B<x>:SMBUS:PEC:VALUe command.

Description
  • This command sets or queries the SMBus PEC selection for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS:PEC:VALUe? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS:PEC:VALUe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SMBUS:PEC:VALUe value command.
SCPI Syntax
- BUS:B<x>:SMBUS:PEC:VALUe {TRUe|FALSe}
- BUS:B<x>:SMBUS:PEC:VALUe?
Info
  • B<x> is the bus number.
  • TRUe specifies the SMBus PEC selection as true.
  • FALSe specifies the SMBus PEC selection as false.

BusBItemSmbusPecValue

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SMBUS:PEC:VALUe command.

Description
  • This command sets or queries the SMBus PEC selection for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SMBUS:PEC:VALUe? query.
  • Using the .verify(value) method will send the BUS:B<x>:SMBUS:PEC:VALUe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SMBUS:PEC:VALUe value command.
SCPI Syntax
- BUS:B<x>:SMBUS:PEC:VALUe {TRUe|FALSe}
- BUS:B<x>:SMBUS:PEC:VALUe?
Info
  • B<x> is the bus number.
  • TRUe specifies the SMBus PEC selection as true.
  • FALSe specifies the SMBus PEC selection as false.

BusBItemSpacewire

Bases: SCPICmdRead

The BUS:B<x>:SPACEWIRe command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the bus number.
Properties
  • .bitrate: The BUS:B<x>:SPACEWIRe:BITRate command.
  • .data: The BUS:B<x>:SPACEWIRe:DATa command tree.
  • .decode: The BUS:B<x>:SPACEWIRe:DECode command tree.
  • .strobe: The BUS:B<x>:SPACEWIRe:STRobe command tree.
  • .sync: The BUS:B<x>:SPACEWIRe:SYNC command.

bitrate property

Return the BUS:B<x>:SPACEWIRe:BITRate command.

Description
  • This command sets or queries the SpaceWire bit rate.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:BITRate value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:BITRate <NR3>
- BUS:B<x>:SPACEWIRe:BITRate?
Info
  • B<x> is the bus number.
  • <NR3> specifies the SpaceWire bit rate for the specified bus. The valid bit rate range is 2 Mbps to 200 Mbps. The default value is 10 Mbps.

data property

Return the BUS:B<x>:SPACEWIRe:DATa command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:DATa? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:DATa? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the bus number.
Sub-properties
  • .source: The BUS:B<x>:SPACEWIRe:DATa:SOUrce command.
  • .threshold: The BUS:B<x>:SPACEWIRe:DATa:THReshold command.

decode property

Return the BUS:B<x>:SPACEWIRe:DECode command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:DECode? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:DECode? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the bus number.
Sub-properties
  • .type: The BUS:B<x>:SPACEWIRe:DECode:TYPe command.

strobe property

Return the BUS:B<x>:SPACEWIRe:STRobe command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:STRobe? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:STRobe? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the bus number.
Sub-properties
  • .source: The BUS:B<x>:SPACEWIRe:STRobe:SOUrce command.
  • .threshold: The BUS:B<x>:SPACEWIRe:STRobe:THReshold command.

sync property

Return the BUS:B<x>:SPACEWIRe:SYNC command.

Description
  • This command sets or queries sync for SpaceWire decoding.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:SYNC? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:SYNC? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:SYNC value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:SYNC {DATA|NULL|AUTO|CUSTom}
- BUS:B<x>:SPACEWIRe:SYNC?
Info
  • B<x> is the bus number.
  • DATA specifies sync as data.
  • NULL specifies sync as null.
  • AUTO specifies sync as auto.
  • CUSTom specifies sync as custom.
Sub-properties
  • .count: The BUS:B<x>:SPACEWIRe:SYNC:COUnt command.
  • .pattern: The BUS:B<x>:SPACEWIRe:SYNC:PATTern command.
  • .value: The BUS:B<x>:SPACEWIRe:SYNC:VALUe command.

BusBItemSpacewireBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPACEWIRe:BITRate command.

Description
  • This command sets or queries the SpaceWire bit rate.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:BITRate value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:BITRate <NR3>
- BUS:B<x>:SPACEWIRe:BITRate?
Info
  • B<x> is the bus number.
  • <NR3> specifies the SpaceWire bit rate for the specified bus. The valid bit rate range is 2 Mbps to 200 Mbps. The default value is 10 Mbps.

BusBItemSpacewireData

Bases: SCPICmdRead

The BUS:B<x>:SPACEWIRe:DATa command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:DATa? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:DATa? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the bus number.
Properties
  • .source: The BUS:B<x>:SPACEWIRe:DATa:SOUrce command.
  • .threshold: The BUS:B<x>:SPACEWIRe:DATa:THReshold command.

source property

Return the BUS:B<x>:SPACEWIRe:DATa:SOUrce command.

Description
  • This command sets or queries the source of the SpaceWire Data signal for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:DATa:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:DATa:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:DATa:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:DATa:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SPACEWIRe:DATa:SOUrce?
Info
  • B<x> is the bus number.
  • CH<x> specifies an analog channel as the source for the data signal, where is the channel number.
  • MATH<x> specifies a math channel as the source for the data signal, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source for the data signal, where is the reference waveform number.

threshold property

Return the BUS:B<x>:SPACEWIRe:DATa:THReshold command.

Description
  • This command sets or queries the threshold of the SpaceWire Data signal for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:DATa:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:DATa:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:DATa:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:DATa:THReshold <NR3>
- BUS:B<x>:SPACEWIRe:DATa:THReshold?
Info
  • B<x> is the bus number.
  • <NR3> specifies the SpaceWire Data threshold level for the specified bus, in volts.

BusBItemSpacewireDataSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPACEWIRe:DATa:SOUrce command.

Description
  • This command sets or queries the source of the SpaceWire Data signal for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:DATa:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:DATa:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:DATa:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:DATa:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SPACEWIRe:DATa:SOUrce?
Info
  • B<x> is the bus number.
  • CH<x> specifies an analog channel as the source for the data signal, where is the channel number.
  • MATH<x> specifies a math channel as the source for the data signal, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source for the data signal, where is the reference waveform number.

BusBItemSpacewireDataThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPACEWIRe:DATa:THReshold command.

Description
  • This command sets or queries the threshold of the SpaceWire Data signal for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:DATa:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:DATa:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:DATa:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:DATa:THReshold <NR3>
- BUS:B<x>:SPACEWIRe:DATa:THReshold?
Info
  • B<x> is the bus number.
  • <NR3> specifies the SpaceWire Data threshold level for the specified bus, in volts.

BusBItemSpacewireDecode

Bases: SCPICmdRead

The BUS:B<x>:SPACEWIRe:DECode command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:DECode? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:DECode? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the bus number.
Properties
  • .type: The BUS:B<x>:SPACEWIRe:DECode:TYPe command.

type property

Return the BUS:B<x>:SPACEWIRe:DECode:TYPe command.

Description
  • This command sets or queries the decode type for SpaceWire bus decode.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:DECode:TYPe? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:DECode:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:DECode:TYPe value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:DECode:TYPe {STRObe|DATARate}
- BUS:B<x>:SPACEWIRe:DECode:TYPe?
Info
  • B<x> is the bus number.
  • STRObe specifies the decode type as strobe.
  • DATARate specifies the decode type as data rate.

BusBItemSpacewireDecodeType

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPACEWIRe:DECode:TYPe command.

Description
  • This command sets or queries the decode type for SpaceWire bus decode.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:DECode:TYPe? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:DECode:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:DECode:TYPe value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:DECode:TYPe {STRObe|DATARate}
- BUS:B<x>:SPACEWIRe:DECode:TYPe?
Info
  • B<x> is the bus number.
  • STRObe specifies the decode type as strobe.
  • DATARate specifies the decode type as data rate.

BusBItemSpacewireStrobe

Bases: SCPICmdRead

The BUS:B<x>:SPACEWIRe:STRobe command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:STRobe? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:STRobe? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the bus number.
Properties
  • .source: The BUS:B<x>:SPACEWIRe:STRobe:SOUrce command.
  • .threshold: The BUS:B<x>:SPACEWIRe:STRobe:THReshold command.

source property

Return the BUS:B<x>:SPACEWIRe:STRobe:SOUrce command.

Description
  • This command sets or queries the source of the SpaceWire Strobe signal for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:STRobe:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:STRobe:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:STRobe:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:STRobe:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SPACEWIRe:STRobe:SOUrce?
Info
  • B<x> is the bus number.
  • CH<x> specifies an analog channel as the source for the strobe signal, where is the channel number.
  • MATH<x> specifies a math channel as the source for the strobe signal, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source for the strobe signal, where is the reference waveform number.

threshold property

Return the BUS:B<x>:SPACEWIRe:STRobe:THReshold command.

Description
  • This command sets or queries the threshold level of the SpaceWire Strobe signal for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:STRobe:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:STRobe:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:STRobe:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:STRobe:THReshold <NR3>
- BUS:B<x>:SPACEWIRe:STRobe:THReshold?
Info
  • B<x> is the bus number.
  • <NR3> specifies the SpaceWire Strobe signal threshold level for the specified bus, in volts.

BusBItemSpacewireStrobeSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPACEWIRe:STRobe:SOUrce command.

Description
  • This command sets or queries the source of the SpaceWire Strobe signal for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:STRobe:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:STRobe:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:STRobe:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:STRobe:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SPACEWIRe:STRobe:SOUrce?
Info
  • B<x> is the bus number.
  • CH<x> specifies an analog channel as the source for the strobe signal, where is the channel number.
  • MATH<x> specifies a math channel as the source for the strobe signal, where is the math waveform number.
  • REF<x> specifies a reference waveform as the source for the strobe signal, where is the reference waveform number.

BusBItemSpacewireStrobeThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPACEWIRe:STRobe:THReshold command.

Description
  • This command sets or queries the threshold level of the SpaceWire Strobe signal for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:STRobe:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:STRobe:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:STRobe:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:STRobe:THReshold <NR3>
- BUS:B<x>:SPACEWIRe:STRobe:THReshold?
Info
  • B<x> is the bus number.
  • <NR3> specifies the SpaceWire Strobe signal threshold level for the specified bus, in volts.

BusBItemSpacewireSync

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPACEWIRe:SYNC command.

Description
  • This command sets or queries sync for SpaceWire decoding.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:SYNC? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:SYNC? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:SYNC value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:SYNC {DATA|NULL|AUTO|CUSTom}
- BUS:B<x>:SPACEWIRe:SYNC?
Info
  • B<x> is the bus number.
  • DATA specifies sync as data.
  • NULL specifies sync as null.
  • AUTO specifies sync as auto.
  • CUSTom specifies sync as custom.
Properties
  • .count: The BUS:B<x>:SPACEWIRe:SYNC:COUnt command.
  • .pattern: The BUS:B<x>:SPACEWIRe:SYNC:PATTern command.
  • .value: The BUS:B<x>:SPACEWIRe:SYNC:VALUe command.

count property

Return the BUS:B<x>:SPACEWIRe:SYNC:COUnt command.

Description
  • This command sets or queries the length of the data string in bytes to be used for a SpaceWire sync bytes.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:SYNC:COUnt? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:SYNC:COUnt? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:SYNC:COUnt value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:SYNC:COUnt <NR1>
- BUS:B<x>:SPACEWIRe:SYNC:COUnt?
Info
  • B<x> is the bus number.
  • <NR1> specifies the length of the data string in bytes. The mininum is 2 and the maximum is 10. The default is 2 bytes.

pattern property

Return the BUS:B<x>:SPACEWIRe:SYNC:PATTern command.

Description
  • This command sets or queries sync pattern for SpaceWire decoding.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:SYNC:PATTern? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:SYNC:PATTern? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:SYNC:PATTern value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:SYNC:PATTern <NR3>
- BUS:B<x>:SPACEWIRe:SYNC:PATTern?
Info
  • B<x> is the bus number.
  • <NR3> specifies the sync pattern.

value property

Return the BUS:B<x>:SPACEWIRe:SYNC:VALUe command.

Description
  • This command sets or queries sync value for sync option data.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:SYNC:VALUe? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:SYNC:VALUe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:SYNC:VALUe value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:SYNC:VALUe <NR3>
- BUS:B<x>:SPACEWIRe:SYNC:VALUe?
Info
  • B<x> is the bus number.
  • <NR3> specifies the sync value for sync option data.

BusBItemSpacewireSyncCount

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPACEWIRe:SYNC:COUnt command.

Description
  • This command sets or queries the length of the data string in bytes to be used for a SpaceWire sync bytes.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:SYNC:COUnt? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:SYNC:COUnt? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:SYNC:COUnt value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:SYNC:COUnt <NR1>
- BUS:B<x>:SPACEWIRe:SYNC:COUnt?
Info
  • B<x> is the bus number.
  • <NR1> specifies the length of the data string in bytes. The mininum is 2 and the maximum is 10. The default is 2 bytes.

BusBItemSpacewireSyncPattern

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPACEWIRe:SYNC:PATTern command.

Description
  • This command sets or queries sync pattern for SpaceWire decoding.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:SYNC:PATTern? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:SYNC:PATTern? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:SYNC:PATTern value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:SYNC:PATTern <NR3>
- BUS:B<x>:SPACEWIRe:SYNC:PATTern?
Info
  • B<x> is the bus number.
  • <NR3> specifies the sync pattern.

BusBItemSpacewireSyncValue

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPACEWIRe:SYNC:VALUe command.

Description
  • This command sets or queries sync value for sync option data.
Usage
  • Using the .query() method will send the BUS:B<x>:SPACEWIRe:SYNC:VALUe? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPACEWIRe:SYNC:VALUe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPACEWIRe:SYNC:VALUe value command.
SCPI Syntax
- BUS:B<x>:SPACEWIRe:SYNC:VALUe <NR3>
- BUS:B<x>:SPACEWIRe:SYNC:VALUe?
Info
  • B<x> is the bus number.
  • <NR3> specifies the sync value for sync option data.

BusBItemSpi

Bases: SCPICmdRead

The BUS:B<x>:SPI command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .bitorder: The BUS:B<x>:SPI:BITOrder command.
  • .clock: The BUS:B<x>:SPI:CLOCk command tree.
  • .data: The BUS:B<x>:SPI:DATa command tree.
  • .framing: The BUS:B<x>:SPI:FRAMING command.
  • .idletime: The BUS:B<x>:SPI:IDLETime command.
  • .miso: The BUS:B<x>:SPI:MISo command tree.
  • .mosi: The BUS:B<x>:SPI:MOSi command tree.
  • .number: The BUS:B<x>:SPI:NUMBer command tree.
  • .select: The BUS:B<x>:SPI:SELect command tree.

bitorder property

Return the BUS:B<x>:SPI:BITOrder command.

Description
  • This command sets or queries the SPI bit order for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:BITOrder? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:BITOrder? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:BITOrder value command.
SCPI Syntax
- BUS:B<x>:SPI:BITOrder {LSB|MSB}
- BUS:B<x>:SPI:BITOrder?
Info
  • B<x> is the number of the bus waveform.
  • LSB specifies that each bit becomes the recovered value’s new LSB, after shifting previously recovered bits one place to the left. The decoding happens right to left.
  • MSB specifies that each successive bit from the bus’s data line becomes the new MSB of the recovered value, shifting any previously recovered bits one place to the right. The decoding happens left to right.

clock property

Return the BUS:B<x>:SPI:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .polarity: The BUS:B<x>:SPI:CLOCk:POLarity command.
  • .source: The BUS:B<x>:SPI:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:SPI:CLOCk:THReshold command.

data property

Return the BUS:B<x>:SPI:DATa command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI:DATa? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:DATa? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .polarity: The BUS:B<x>:SPI:DATa:POLarity command.
  • .size: The BUS:B<x>:SPI:DATa:SIZe command.
  • .source: The BUS:B<x>:SPI:DATa:SOUrce command.
  • .threshold: The BUS:B<x>:SPI:DATa:THReshold command.

framing property

Return the BUS:B<x>:SPI:FRAMING command.

Description
  • This command sets or queries the SPI framing setting for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:FRAMING? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:FRAMING? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:FRAMING value command.
SCPI Syntax
- BUS:B<x>:SPI:FRAMING {IDLE|SS}
- BUS:B<x>:SPI:FRAMING?
Info
  • B<x> is the number of the bus waveform.
  • IDLE specifies IDLE SPI framing.
  • SS specifies SS SPI framing.

idletime property

Return the BUS:B<x>:SPI:IDLETime command.

Description
  • This command sets or queries the SPI idle time for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:IDLETime? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:IDLETime? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:IDLETime value command.
SCPI Syntax
- BUS:B<x>:SPI:IDLETime <NR3>
- BUS:B<x>:SPI:IDLETime?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> specifies the SPI idle time.

miso property

Return the BUS:B<x>:SPI:MISo command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MISo? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MISo? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .data: The BUS:B<x>:SPI:MISo:DATa command tree.
  • .input: The BUS:B<x>:SPI:MISo:INPut command.
  • .threshold: The BUS:B<x>:SPI:MISo:THReshold command.

mosi property

Return the BUS:B<x>:SPI:MOSi command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MOSi? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MOSi? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .data: The BUS:B<x>:SPI:MOSi:DATa command tree.
  • .input: The BUS:B<x>:SPI:MOSi:INPut command.
  • .threshold: The BUS:B<x>:SPI:MOSi:THReshold command.

number property

Return the BUS:B<x>:SPI:NUMBer command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI:NUMBer? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:NUMBer? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .inputs: The BUS:B<x>:SPI:NUMBer:INputs command.

select property

Return the BUS:B<x>:SPI:SELect command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI:SELect? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:SELect? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .polarity: The BUS:B<x>:SPI:SELect:POLarity command.
  • .source: The BUS:B<x>:SPI:SELect:SOUrce command.
  • .threshold: The BUS:B<x>:SPI:SELect:THReshold command.

BusBItemSpiBitorder

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:BITOrder command.

Description
  • This command sets or queries the SPI bit order for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:BITOrder? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:BITOrder? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:BITOrder value command.
SCPI Syntax
- BUS:B<x>:SPI:BITOrder {LSB|MSB}
- BUS:B<x>:SPI:BITOrder?
Info
  • B<x> is the number of the bus waveform.
  • LSB specifies that each bit becomes the recovered value’s new LSB, after shifting previously recovered bits one place to the left. The decoding happens right to left.
  • MSB specifies that each successive bit from the bus’s data line becomes the new MSB of the recovered value, shifting any previously recovered bits one place to the right. The decoding happens left to right.

BusBItemSpiClock

Bases: SCPICmdRead

The BUS:B<x>:SPI:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .polarity: The BUS:B<x>:SPI:CLOCk:POLarity command.
  • .source: The BUS:B<x>:SPI:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:SPI:CLOCk:THReshold command.

polarity property

Return the BUS:B<x>:SPI:CLOCk:POLarity command.

Description
  • This command sets or queries the SPI clock (SCLK) source polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:CLOCk:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:CLOCk:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:CLOCk:POLarity value command.
SCPI Syntax
- BUS:B<x>:SPI:CLOCk:POLarity {FALL|RISE}
- BUS:B<x>:SPI:CLOCk:POLarity?
Info
  • B<x> is the number of the bus waveform.
  • FALL sets the clock to the falling edge of the signal.
  • RISE sets the clock to the rising edge of the signal.

source property

Return the BUS:B<x>:SPI:CLOCk:SOUrce command.

Description
  • This command sets or queries the SPI clock (SCLK) source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SPI:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SPI:CLOCk:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> designates an analog channel as the bus SPI clock source.
  • MATH<x> designates a math waveform as the clock source.
  • REF<x> designates a reference waveform as the clock source.
  • REF<x>_D<x> specifies a digital reference waveform as the clock source waveform for the specified SPI bus.

threshold property

Return the BUS:B<x>:SPI:CLOCk:THReshold command.

Description
  • This command sets or queries the SPI Clock (SCLK) source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:CLOCk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:CLOCk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:CLOCk:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPI:CLOCk:THReshold <NR3>
- BUS:B<x>:SPI:CLOCk:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SPI Clock (SCLK) source threshold for the specified bus.

BusBItemSpiClockPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:CLOCk:POLarity command.

Description
  • This command sets or queries the SPI clock (SCLK) source polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:CLOCk:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:CLOCk:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:CLOCk:POLarity value command.
SCPI Syntax
- BUS:B<x>:SPI:CLOCk:POLarity {FALL|RISE}
- BUS:B<x>:SPI:CLOCk:POLarity?
Info
  • B<x> is the number of the bus waveform.
  • FALL sets the clock to the falling edge of the signal.
  • RISE sets the clock to the rising edge of the signal.

BusBItemSpiClockSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:CLOCk:SOUrce command.

Description
  • This command sets or queries the SPI clock (SCLK) source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SPI:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SPI:CLOCk:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> designates an analog channel as the bus SPI clock source.
  • MATH<x> designates a math waveform as the clock source.
  • REF<x> designates a reference waveform as the clock source.
  • REF<x>_D<x> specifies a digital reference waveform as the clock source waveform for the specified SPI bus.

BusBItemSpiClockThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:CLOCk:THReshold command.

Description
  • This command sets or queries the SPI Clock (SCLK) source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:CLOCk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:CLOCk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:CLOCk:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPI:CLOCk:THReshold <NR3>
- BUS:B<x>:SPI:CLOCk:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SPI Clock (SCLK) source threshold for the specified bus.

BusBItemSpiData

Bases: SCPICmdRead

The BUS:B<x>:SPI:DATa command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI:DATa? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:DATa? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .polarity: The BUS:B<x>:SPI:DATa:POLarity command.
  • .size: The BUS:B<x>:SPI:DATa:SIZe command.
  • .source: The BUS:B<x>:SPI:DATa:SOUrce command.
  • .threshold: The BUS:B<x>:SPI:DATa:THReshold command.

polarity property

Return the BUS:B<x>:SPI:DATa:POLarity command.

Description
  • This command sets or queries the SPI Data (Data) source polarity for the bus number specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:DATa:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:DATa:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:DATa:POLarity value command.
SCPI Syntax
- BUS:B<x>:SPI:DATa:POLarity {HIGH|LOW}
- BUS:B<x>:SPI:DATa:POLarity?
Info
  • B<x> is the number of the bus waveform.
  • HIGH sets the SPI data polarity to active high.
  • LOW sets the SPI data polarity to active low.

size property

Return the BUS:B<x>:SPI:DATa:SIZe command.

Description
  • This command sets or queries the number of bits per word for the specified SPI bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:DATa:SIZe? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:DATa:SIZe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:DATa:SIZe value command.
SCPI Syntax
- BUS:B<x>:SPI:DATa:SIZe <NR1>
- BUS:B<x>:SPI:DATa:SIZe?
Info
  • B<x> is the number of the bus waveform.
  • <NR1> is the data size for the specified bus. The minimum value is 2 and maximum is 32.

source property

Return the BUS:B<x>:SPI:DATa:SOUrce command.

Description
  • This command sets or queries the SPI Data (Data) source for the bus number specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:DATa:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:DATa:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:DATa:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SPI:DATa:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SPI:DATa:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> designates an analog channel as the data source for the specified SPI bus.
  • MATH<x> designates a math waveform as the data source.
  • REF<x> designates a reference waveform as the data source.
  • REF<x>_D<x> specifies a digital reference waveform as the data source waveform for the specified SPI bus.

threshold property

Return the BUS:B<x>:SPI:DATa:THReshold command.

Description
  • This command sets or queries the SPI Data (Data) source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:DATa:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:DATa:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:DATa:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPI:DATa:THReshold <NR3>
- BUS:B<x>:SPI:DATa:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SPI Data (SDA) source threshold for the specified bus.

BusBItemSpiDataPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:DATa:POLarity command.

Description
  • This command sets or queries the SPI Data (Data) source polarity for the bus number specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:DATa:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:DATa:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:DATa:POLarity value command.
SCPI Syntax
- BUS:B<x>:SPI:DATa:POLarity {HIGH|LOW}
- BUS:B<x>:SPI:DATa:POLarity?
Info
  • B<x> is the number of the bus waveform.
  • HIGH sets the SPI data polarity to active high.
  • LOW sets the SPI data polarity to active low.

BusBItemSpiDataSize

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:DATa:SIZe command.

Description
  • This command sets or queries the number of bits per word for the specified SPI bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:DATa:SIZe? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:DATa:SIZe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:DATa:SIZe value command.
SCPI Syntax
- BUS:B<x>:SPI:DATa:SIZe <NR1>
- BUS:B<x>:SPI:DATa:SIZe?
Info
  • B<x> is the number of the bus waveform.
  • <NR1> is the data size for the specified bus. The minimum value is 2 and maximum is 32.

BusBItemSpiDataSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:DATa:SOUrce command.

Description
  • This command sets or queries the SPI Data (Data) source for the bus number specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:DATa:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:DATa:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:DATa:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SPI:DATa:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SPI:DATa:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> designates an analog channel as the data source for the specified SPI bus.
  • MATH<x> designates a math waveform as the data source.
  • REF<x> designates a reference waveform as the data source.
  • REF<x>_D<x> specifies a digital reference waveform as the data source waveform for the specified SPI bus.

BusBItemSpiDataThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:DATa:THReshold command.

Description
  • This command sets or queries the SPI Data (Data) source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:DATa:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:DATa:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:DATa:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPI:DATa:THReshold <NR3>
- BUS:B<x>:SPI:DATa:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SPI Data (SDA) source threshold for the specified bus.

BusBItemSpiFraming

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:FRAMING command.

Description
  • This command sets or queries the SPI framing setting for the specified bus. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:FRAMING? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:FRAMING? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:FRAMING value command.
SCPI Syntax
- BUS:B<x>:SPI:FRAMING {IDLE|SS}
- BUS:B<x>:SPI:FRAMING?
Info
  • B<x> is the number of the bus waveform.
  • IDLE specifies IDLE SPI framing.
  • SS specifies SS SPI framing.

BusBItemSpiIdletime

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:IDLETime command.

Description
  • This command sets or queries the SPI idle time for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:IDLETime? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:IDLETime? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:IDLETime value command.
SCPI Syntax
- BUS:B<x>:SPI:IDLETime <NR3>
- BUS:B<x>:SPI:IDLETime?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> specifies the SPI idle time.

BusBItemSpiMiso

Bases: SCPICmdRead

The BUS:B<x>:SPI:MISo command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MISo? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MISo? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .data: The BUS:B<x>:SPI:MISo:DATa command tree.
  • .input: The BUS:B<x>:SPI:MISo:INPut command.
  • .threshold: The BUS:B<x>:SPI:MISo:THReshold command.

data property

Return the BUS:B<x>:SPI:MISo:DATa command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MISo:DATa? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MISo:DATa? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .polarity: The BUS:B<x>:SPI:MISo:DATa:POLarity command.

input property

Return the BUS:B<x>:SPI:MISo:INPut command.

Description
  • This command sets or queries the SPI MISo source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MISo:INPut? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MISo:INPut? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:MISo:INPut value command.
SCPI Syntax
- BUS:B<x>:SPI:MISo:INPut {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SPI:MISo:INPut?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> designates an analog channel as the source.
  • MATH<x> designates a math waveform as the source.
  • REF<x> designates a reference waveform as the source.

threshold property

Return the BUS:B<x>:SPI:MISo:THReshold command.

Description
  • This command sets or queries the SPI MISo Data source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MISo:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MISo:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:MISo:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPI:MISo:THReshold <NR3>
- BUS:B<x>:SPI:MISo:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SPI MISo Data source threshold for the specified bus.

BusBItemSpiMisoData

Bases: SCPICmdRead

The BUS:B<x>:SPI:MISo:DATa command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MISo:DATa? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MISo:DATa? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .polarity: The BUS:B<x>:SPI:MISo:DATa:POLarity command.

polarity property

Return the BUS:B<x>:SPI:MISo:DATa:POLarity command.

Description
  • This command sets or queries the SPI MISo Data source polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MISo:DATa:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MISo:DATa:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:MISo:DATa:POLarity value command.
SCPI Syntax
- BUS:B<x>:SPI:MISo:DATa:POLarity {HIGH|LOW}
- BUS:B<x>:SPI:MISo:DATa:POLarity?
Info
  • B<x> is the number of the bus waveform.
  • HIGH sets the polarity to high.
  • LOW sets the polarity to low.

BusBItemSpiMisoDataPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:MISo:DATa:POLarity command.

Description
  • This command sets or queries the SPI MISo Data source polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MISo:DATa:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MISo:DATa:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:MISo:DATa:POLarity value command.
SCPI Syntax
- BUS:B<x>:SPI:MISo:DATa:POLarity {HIGH|LOW}
- BUS:B<x>:SPI:MISo:DATa:POLarity?
Info
  • B<x> is the number of the bus waveform.
  • HIGH sets the polarity to high.
  • LOW sets the polarity to low.

BusBItemSpiMisoInput

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:MISo:INPut command.

Description
  • This command sets or queries the SPI MISo source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MISo:INPut? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MISo:INPut? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:MISo:INPut value command.
SCPI Syntax
- BUS:B<x>:SPI:MISo:INPut {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SPI:MISo:INPut?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> designates an analog channel as the source.
  • MATH<x> designates a math waveform as the source.
  • REF<x> designates a reference waveform as the source.

BusBItemSpiMisoThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:MISo:THReshold command.

Description
  • This command sets or queries the SPI MISo Data source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MISo:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MISo:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:MISo:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPI:MISo:THReshold <NR3>
- BUS:B<x>:SPI:MISo:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SPI MISo Data source threshold for the specified bus.

BusBItemSpiMosi

Bases: SCPICmdRead

The BUS:B<x>:SPI:MOSi command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MOSi? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MOSi? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .data: The BUS:B<x>:SPI:MOSi:DATa command tree.
  • .input: The BUS:B<x>:SPI:MOSi:INPut command.
  • .threshold: The BUS:B<x>:SPI:MOSi:THReshold command.

data property

Return the BUS:B<x>:SPI:MOSi:DATa command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MOSi:DATa? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MOSi:DATa? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .polarity: The BUS:B<x>:SPI:MOSi:DATa:POLarity command.

input property

Return the BUS:B<x>:SPI:MOSi:INPut command.

Description
  • This command sets or queries the SPI MOSI source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MOSi:INPut? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MOSi:INPut? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:MOSi:INPut value command.
SCPI Syntax
- BUS:B<x>:SPI:MOSi:INPut {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SPI:MOSi:INPut?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> designates an analog channel as the source.
  • MATH<x> designates a math waveform as the source.
  • REF<x> designates a reference waveform as the source.

threshold property

Return the BUS:B<x>:SPI:MOSi:THReshold command.

Description
  • This command sets or queries the SPI MOSI source thresold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MOSi:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MOSi:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:MOSi:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPI:MOSi:THReshold <NR3>
- BUS:B<x>:SPI:MOSi:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SPI MOSI Data source threshold for the specified bus.

BusBItemSpiMosiData

Bases: SCPICmdRead

The BUS:B<x>:SPI:MOSi:DATa command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MOSi:DATa? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MOSi:DATa? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .polarity: The BUS:B<x>:SPI:MOSi:DATa:POLarity command.

polarity property

Return the BUS:B<x>:SPI:MOSi:DATa:POLarity command.

Description
  • This command sets or queries the SPI MOSI source polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MOSi:DATa:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MOSi:DATa:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:MOSi:DATa:POLarity value command.
SCPI Syntax
- BUS:B<x>:SPI:MOSi:DATa:POLarity {HIGH|LOW}
- BUS:B<x>:SPI:MOSi:DATa:POLarity?
Info
  • B<x> is the number of the bus waveform.
  • HIGH sets the polarity to high.
  • LOW sets the polarity to low.

BusBItemSpiMosiDataPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:MOSi:DATa:POLarity command.

Description
  • This command sets or queries the SPI MOSI source polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MOSi:DATa:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MOSi:DATa:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:MOSi:DATa:POLarity value command.
SCPI Syntax
- BUS:B<x>:SPI:MOSi:DATa:POLarity {HIGH|LOW}
- BUS:B<x>:SPI:MOSi:DATa:POLarity?
Info
  • B<x> is the number of the bus waveform.
  • HIGH sets the polarity to high.
  • LOW sets the polarity to low.

BusBItemSpiMosiInput

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:MOSi:INPut command.

Description
  • This command sets or queries the SPI MOSI source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MOSi:INPut? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MOSi:INPut? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:MOSi:INPut value command.
SCPI Syntax
- BUS:B<x>:SPI:MOSi:INPut {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SPI:MOSi:INPut?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> designates an analog channel as the source.
  • MATH<x> designates a math waveform as the source.
  • REF<x> designates a reference waveform as the source.

BusBItemSpiMosiThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:MOSi:THReshold command.

Description
  • This command sets or queries the SPI MOSI source thresold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:MOSi:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:MOSi:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:MOSi:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPI:MOSi:THReshold <NR3>
- BUS:B<x>:SPI:MOSi:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SPI MOSI Data source threshold for the specified bus.

BusBItemSpiNumber

Bases: SCPICmdRead

The BUS:B<x>:SPI:NUMBer command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI:NUMBer? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:NUMBer? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .inputs: The BUS:B<x>:SPI:NUMBer:INputs command.

inputs property

Return the BUS:B<x>:SPI:NUMBer:INputs command.

Description
  • This command sets or queries the number of inputs for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:NUMBer:INputs? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:NUMBer:INputs? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:NUMBer:INputs value command.
SCPI Syntax
- BUS:B<x>:SPI:NUMBer:INputs {ONE|TWO}
- BUS:B<x>:SPI:NUMBer:INputs?
Info
  • B<x> is the number of the bus waveform.
  • ONE sets the Data Inputs as one.
  • TWO sets the Data Inputs as two.

BusBItemSpiNumberInputs

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:NUMBer:INputs command.

Description
  • This command sets or queries the number of inputs for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:NUMBer:INputs? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:NUMBer:INputs? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:NUMBer:INputs value command.
SCPI Syntax
- BUS:B<x>:SPI:NUMBer:INputs {ONE|TWO}
- BUS:B<x>:SPI:NUMBer:INputs?
Info
  • B<x> is the number of the bus waveform.
  • ONE sets the Data Inputs as one.
  • TWO sets the Data Inputs as two.

BusBItemSpiSelect

Bases: SCPICmdRead

The BUS:B<x>:SPI:SELect command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPI:SELect? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:SELect? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .polarity: The BUS:B<x>:SPI:SELect:POLarity command.
  • .source: The BUS:B<x>:SPI:SELect:SOUrce command.
  • .threshold: The BUS:B<x>:SPI:SELect:THReshold command.

polarity property

Return the BUS:B<x>:SPI:SELect:POLarity command.

Description
  • This command sets or queries the SPI Slave Select (SS) polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:SELect:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:SELect:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:SELect:POLarity value command.
SCPI Syntax
- BUS:B<x>:SPI:SELect:POLarity {LOW|HIGH}
- BUS:B<x>:SPI:SELect:POLarity?
Info
  • B<x> is the number of the bus waveform.
  • LOW sets an active low polarity.
  • HIGH sets an active high polarity.

source property

Return the BUS:B<x>:SPI:SELect:SOUrce command.

Description
  • This command sets or queries the SPI Slave Select (SS) source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:SELect:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:SELect:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:SELect:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SPI:SELect:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SPI:SELect:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> designates an analog channel as the buses’ SPI Slave Select source.
  • MATH<x> designates a math waveform as the Slave Select source.
  • REF<x> designates a reference waveform as the Slave Select source.
  • REF<x>_D<x> specifies a digital reference waveform as the clock source waveform for the specified SPI bus.

threshold property

Return the BUS:B<x>:SPI:SELect:THReshold command.

Description
  • This command sets or queries the SPI Select (SS) source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:SELect:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:SELect:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:SELect:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPI:SELect:THReshold <NR3>
- BUS:B<x>:SPI:SELect:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SPI Select (SS) source threshold for the specified bus.

BusBItemSpiSelectPolarity

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:SELect:POLarity command.

Description
  • This command sets or queries the SPI Slave Select (SS) polarity for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:SELect:POLarity? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:SELect:POLarity? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:SELect:POLarity value command.
SCPI Syntax
- BUS:B<x>:SPI:SELect:POLarity {LOW|HIGH}
- BUS:B<x>:SPI:SELect:POLarity?
Info
  • B<x> is the number of the bus waveform.
  • LOW sets an active low polarity.
  • HIGH sets an active high polarity.

BusBItemSpiSelectSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:SELect:SOUrce command.

Description
  • This command sets or queries the SPI Slave Select (SS) source for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:SELect:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:SELect:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:SELect:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SPI:SELect:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SPI:SELect:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> designates an analog channel as the buses’ SPI Slave Select source.
  • MATH<x> designates a math waveform as the Slave Select source.
  • REF<x> designates a reference waveform as the Slave Select source.
  • REF<x>_D<x> specifies a digital reference waveform as the clock source waveform for the specified SPI bus.

BusBItemSpiSelectThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPI:SELect:THReshold command.

Description
  • This command sets or queries the SPI Select (SS) source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SPI:SELect:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPI:SELect:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPI:SELect:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPI:SELect:THReshold <NR3>
- BUS:B<x>:SPI:SELect:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SPI Select (SS) source threshold for the specified bus.

BusBItemSpmi

Bases: SCPICmdRead

The BUS:B<x>:SPMI command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPMI? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPMI? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .sclk: The BUS:B<x>:SPMI:SCLk command tree.
  • .sdata: The BUS:B<x>:SPMI:SDATa command tree.

sclk property

Return the BUS:B<x>:SPMI:SCLk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPMI:SCLk? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPMI:SCLk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .source: The BUS:B<x>:SPMI:SCLk:SOUrce command.
  • .threshold: The BUS:B<x>:SPMI:SCLk:THReshold command.

sdata property

Return the BUS:B<x>:SPMI:SDATa command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPMI:SDATa? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPMI:SDATa? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .source: The BUS:B<x>:SPMI:SDATa:SOUrce command.
  • .threshold: The BUS:B<x>:SPMI:SDATa:THReshold command.

BusBItemSpmiSclk

Bases: SCPICmdRead

The BUS:B<x>:SPMI:SCLk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPMI:SCLk? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPMI:SCLk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .source: The BUS:B<x>:SPMI:SCLk:SOUrce command.
  • .threshold: The BUS:B<x>:SPMI:SCLk:THReshold command.

source property

Return the BUS:B<x>:SPMI:SCLk:SOUrce command.

Description
  • This command sets or queries the SPMI Clock (SCLK) source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SPMI:SCLk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPMI:SCLk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPMI:SCLk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SPMI:SCLk:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SPMI:SCLk:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the clock source waveform for the SPMI bus.
  • MATH<x> specifies a math waveform as the clock source waveform for the specified SPMI bus.
  • REF<x> specifies a reference waveform as the clock source waveform for the specified SPMI bus.
  • REF<x>_D<x> specifies a digital reference waveform and bit as the clock source waveform for the specified SPMI bus.

threshold property

Return the BUS:B<x>:SPMI:SCLk:THReshold command.

Description
  • This command sets or queries the SPMI Clock (SCLK) source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SPMI:SCLk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPMI:SCLk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPMI:SCLk:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPMI:SCLk:THReshold <NR3>
- BUS:B<x>:SPMI:SCLk:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the clock (SCLK) source threshold value for the specified SPMI bus.

BusBItemSpmiSclkSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPMI:SCLk:SOUrce command.

Description
  • This command sets or queries the SPMI Clock (SCLK) source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SPMI:SCLk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPMI:SCLk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPMI:SCLk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SPMI:SCLk:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SPMI:SCLk:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the clock source waveform for the SPMI bus.
  • MATH<x> specifies a math waveform as the clock source waveform for the specified SPMI bus.
  • REF<x> specifies a reference waveform as the clock source waveform for the specified SPMI bus.
  • REF<x>_D<x> specifies a digital reference waveform and bit as the clock source waveform for the specified SPMI bus.

BusBItemSpmiSclkThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPMI:SCLk:THReshold command.

Description
  • This command sets or queries the SPMI Clock (SCLK) source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SPMI:SCLk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPMI:SCLk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPMI:SCLk:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPMI:SCLk:THReshold <NR3>
- BUS:B<x>:SPMI:SCLk:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the clock (SCLK) source threshold value for the specified SPMI bus.

BusBItemSpmiSdata

Bases: SCPICmdRead

The BUS:B<x>:SPMI:SDATa command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SPMI:SDATa? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPMI:SDATa? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .source: The BUS:B<x>:SPMI:SDATa:SOUrce command.
  • .threshold: The BUS:B<x>:SPMI:SDATa:THReshold command.

source property

Return the BUS:B<x>:SPMI:SDATa:SOUrce command.

Description
  • This command sets or queries the SPMI Data (SDATA) source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SPMI:SDATa:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPMI:SDATa:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPMI:SDATa:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SPMI:SDATa:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SPMI:SDATa:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the data source waveform for the SPMI bus.
  • MATH<x> specifies a math waveform as the data source waveform for the specified SPMI bus.
  • REF<x> specifies a reference waveform as the data source waveform for the specified SPMI bus.
  • REF<x>_D<x> specifies a digital reference waveform and bit as the data source waveform for the specified SPMI bus.

threshold property

Return the BUS:B<x>:SPMI:SDATa:THReshold command.

Description
  • This command sets or queries the SPMI Data (SDATA) source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SPMI:SDATa:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPMI:SDATa:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPMI:SDATa:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPMI:SDATa:THReshold <NR3>
- BUS:B<x>:SPMI:SDATa:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the data (SDATA) source threshold value for the specified SPMI bus.

BusBItemSpmiSdataSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPMI:SDATa:SOUrce command.

Description
  • This command sets or queries the SPMI Data (SDATA) source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SPMI:SDATa:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPMI:SDATa:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPMI:SDATa:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SPMI:SDATa:SOUrce {CH<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:SPMI:SDATa:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the data source waveform for the SPMI bus.
  • MATH<x> specifies a math waveform as the data source waveform for the specified SPMI bus.
  • REF<x> specifies a reference waveform as the data source waveform for the specified SPMI bus.
  • REF<x>_D<x> specifies a digital reference waveform and bit as the data source waveform for the specified SPMI bus.

BusBItemSpmiSdataThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SPMI:SDATa:THReshold command.

Description
  • This command sets or queries the SPMI Data (SDATA) source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:SPMI:SDATa:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SPMI:SDATa:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SPMI:SDATa:THReshold value command.
SCPI Syntax
- BUS:B<x>:SPMI:SDATa:THReshold <NR3>
- BUS:B<x>:SPMI:SDATa:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the data (SDATA) source threshold value for the specified SPMI bus.

BusBItemSvid

Bases: SCPICmdRead

The BUS:B<x>:SVID command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SVID? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .alert: The BUS:B<x>:SVID:ALERT command tree.
  • .clock: The BUS:B<x>:SVID:CLOCk command tree.
  • .data: The BUS:B<x>:SVID:DATA command tree.

alert property

Return the BUS:B<x>:SVID:ALERT command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SVID:ALERT? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:ALERT? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .source: The BUS:B<x>:SVID:ALERT:SOUrce command.
  • .threshold: The BUS:B<x>:SVID:ALERT:THReshold command.

clock property

Return the BUS:B<x>:SVID:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SVID:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .source: The BUS:B<x>:SVID:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:SVID:CLOCk:THReshold command.

data property

Return the BUS:B<x>:SVID:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SVID:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .source: The BUS:B<x>:SVID:DATA:SOUrce command.
  • .threshold: The BUS:B<x>:SVID:DATA:THReshold command.

BusBItemSvidAlert

Bases: SCPICmdRead

The BUS:B<x>:SVID:ALERT command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SVID:ALERT? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:ALERT? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .source: The BUS:B<x>:SVID:ALERT:SOUrce command.
  • .threshold: The BUS:B<x>:SVID:ALERT:THReshold command.

source property

Return the BUS:B<x>:SVID:ALERT:SOUrce command.

Description
  • This command sets or queries alert source channel for the specified SVID bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SVID:ALERT:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:ALERT:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SVID:ALERT:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SVID:ALERT:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SVID:ALERT:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the data source.
  • MATH<x> specifies a math channel as the data source.
  • REF<x> specifies a digital reference waveform as the data source.

threshold property

Return the BUS:B<x>:SVID:ALERT:THReshold command.

Description
  • This command sets or queries the alert threshold for the specified SVID bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SVID:ALERT:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:ALERT:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SVID:ALERT:THReshold value command.
SCPI Syntax
- BUS:B<x>:SVID:ALERT:THReshold <NR3>
- BUS:B<x>:SVID:ALERT:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SVID Strobe threshold for the specified bus. The argument range is -8V to +8V.

BusBItemSvidAlertSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SVID:ALERT:SOUrce command.

Description
  • This command sets or queries alert source channel for the specified SVID bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SVID:ALERT:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:ALERT:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SVID:ALERT:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SVID:ALERT:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SVID:ALERT:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the data source.
  • MATH<x> specifies a math channel as the data source.
  • REF<x> specifies a digital reference waveform as the data source.

BusBItemSvidAlertThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SVID:ALERT:THReshold command.

Description
  • This command sets or queries the alert threshold for the specified SVID bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SVID:ALERT:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:ALERT:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SVID:ALERT:THReshold value command.
SCPI Syntax
- BUS:B<x>:SVID:ALERT:THReshold <NR3>
- BUS:B<x>:SVID:ALERT:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SVID Strobe threshold for the specified bus. The argument range is -8V to +8V.

BusBItemSvidClock

Bases: SCPICmdRead

The BUS:B<x>:SVID:CLOCk command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SVID:CLOCk? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:CLOCk? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .source: The BUS:B<x>:SVID:CLOCk:SOUrce command.
  • .threshold: The BUS:B<x>:SVID:CLOCk:THReshold command.

source property

Return the BUS:B<x>:SVID:CLOCk:SOUrce command.

Description
  • This command sets or queries clock source channel for the specified SVID bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SVID:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SVID:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SVID:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SVID:CLOCk:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies a digital reference waveform as the source.

threshold property

Return the BUS:B<x>:SVID:CLOCk:THReshold command.

Description
  • This command sets or queries the clock threshold for the specified SVID bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SVID:CLOCk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:CLOCk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SVID:CLOCk:THReshold value command.
SCPI Syntax
- BUS:B<x>:SVID:CLOCk:THReshold <NR3>
- BUS:B<x>:SVID:CLOCk:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SVID Strobe threshold for the specified bus. The argument range is -8V to +8V.

BusBItemSvidClockSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SVID:CLOCk:SOUrce command.

Description
  • This command sets or queries clock source channel for the specified SVID bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SVID:CLOCk:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:CLOCk:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SVID:CLOCk:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SVID:CLOCk:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SVID:CLOCk:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies a digital reference waveform as the source.

BusBItemSvidClockThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SVID:CLOCk:THReshold command.

Description
  • This command sets or queries the clock threshold for the specified SVID bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SVID:CLOCk:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:CLOCk:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SVID:CLOCk:THReshold value command.
SCPI Syntax
- BUS:B<x>:SVID:CLOCk:THReshold <NR3>
- BUS:B<x>:SVID:CLOCk:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SVID Strobe threshold for the specified bus. The argument range is -8V to +8V.

BusBItemSvidData

Bases: SCPICmdRead

The BUS:B<x>:SVID:DATA command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:SVID:DATA? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:DATA? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .source: The BUS:B<x>:SVID:DATA:SOUrce command.
  • .threshold: The BUS:B<x>:SVID:DATA:THReshold command.

source property

Return the BUS:B<x>:SVID:DATA:SOUrce command.

Description
  • This command sets or queries data source channel for the specified SVID bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SVID:DATA:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:DATA:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SVID:DATA:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SVID:DATA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SVID:DATA:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies a digital reference waveform as the source.

threshold property

Return the BUS:B<x>:SVID:DATA:THReshold command.

Description
  • This command sets or queries the data threshold for the specified SVID bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SVID:DATA:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:DATA:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SVID:DATA:THReshold value command.
SCPI Syntax
- BUS:B<x>:SVID:DATA:THReshold <NR3>
- BUS:B<x>:SVID:DATA:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SVID Strobe threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemSvidDataSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SVID:DATA:SOUrce command.

Description
  • This command sets or queries data source channel for the specified SVID bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SVID:DATA:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:DATA:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SVID:DATA:SOUrce value command.
SCPI Syntax
- BUS:B<x>:SVID:DATA:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:SVID:DATA:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the source.
  • MATH<x> specifies a math channel as the source.
  • REF<x> specifies a digital reference waveform as the source.

BusBItemSvidDataThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:SVID:DATA:THReshold command.

Description
  • This command sets or queries the data threshold for the specified SVID bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:SVID:DATA:THReshold? query.
  • Using the .verify(value) method will send the BUS:B<x>:SVID:DATA:THReshold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:SVID:DATA:THReshold value command.
SCPI Syntax
- BUS:B<x>:SVID:DATA:THReshold <NR3>
- BUS:B<x>:SVID:DATA:THReshold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the SVID Strobe threshold for the specified bus. The argument range is -8 V to +8 V.

BusBItemType

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:TYPe command.

Description
  • This command sets or queries the bus type or standard for the specified bus. The bus is specified by x. Arguments for a bus type are only available then the required serial bus option is installed.
Usage
  • Using the .query() method will send the BUS:B<x>:TYPe? query.
  • Using the .verify(value) method will send the BUS:B<x>:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:TYPe value command.
SCPI Syntax
- BUS:B<x>:TYPe {ARINC429|AUDio|CAN|ETHernet|EUSB|FLEXRAY|I2C|I3C|LIN|MDIO|MIL1553B|PARallel|RS232C|SENT|SPI|SPMI|SVID|USB}
- BUS:B<x>:TYPe?
Info
  • B<x> is the number of the bus waveform.
  • ARINC429 specifies the ARINC 429 avionics serial bus.
  • AUDio specifies an audio bus.
  • CAN specifies a Controller Area Network bus.
  • ETHernet specifies the Ethernet bus.
  • EUSB specifies a eUSB bus. Requires option SR-EUSB2.
  • FLEXRAY specifies a FlexRay bus.
  • I2C specifies the Inter-IC bus.
  • I3C specifies the MIPI Improved Inter Integrated Circuit (I3C) bus.
  • LIN specifies a Local Interconnect Network bus.
  • MDIO specifies a MDIO bus.
  • MIL1553B specifies the MIL-STD-1553 avionics serial bus.
  • PARallel specifies a parallel bus.
  • RS232C specifies the RS-232 Serial bus.
  • SENT specifies the Single Edge Nibble Transmission (SENT) automotive serial bus.
  • SPI specifies the Serial Peripheral Interface bus.
  • SPMI Specifies a System Power Management Interface bus.
  • SVID Specifies a Serial VID bus.
  • USB specifies the Universal Serial bus.

BusBItemUsb

Bases: SCPICmdRead

The BUS:B<x>:USB command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:USB? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .bitrate: The BUS:B<x>:USB:BITRate command.
  • .dataminusthreshold: The BUS:B<x>:USB:DATAMINUSTHRESHold command.
  • .dataplusthreshold: The BUS:B<x>:USB:DATAPLUSTHRESHold command.
  • .displaymode: The BUS:B<x>:USB:DISplaymode command.
  • .laneone: The BUS:B<x>:USB:LANEONE command tree.
  • .lanes: The BUS:B<x>:USB:LANES command.
  • .lanezero: The BUS:B<x>:USB:LANEZERO command tree.
  • .lowthreshold: The BUS:B<x>:USB:LOWTHRESHold command.
  • .semaththreshold: The BUS:B<x>:USB:SEMATHTHRESHold command.
  • .signaltype: The BUS:B<x>:USB:SIGNALTYpe command.
  • .source: The BUS:B<x>:USB:SOUrce command.
  • .threshold: The BUS:B<x>:USB:THRESHold command.
  • .type: The BUS:B<x>:USB:TYPe command.

bitrate property

Return the BUS:B<x>:USB:BITRate command.

Description
  • This command sets or queries the USB data rate for bus , where the bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:BITRate value command.
SCPI Syntax
- BUS:B<x>:USB:BITRate {FULL|HIGH|LOW|USB3O|USB31G1|USB32G1}
- BUS:B<x>:USB:BITRate?
Info
  • B<x> is the number of the bus waveform.
  • FULL indicates the bit rate is 12 Mbps.
  • HIGH indicates the bit rate is 480 Mbps.
  • LOW indicates the bit rate is 1.5 Mbps.
  • USB3O indicates the bit rate is USB 3.0 (5 Gbps). Requires option SR-USB3.
  • USB31G1 indicates the bit rate is USB 3.1 Gen 1 (5 Gbps). Requires option SR-USB3.
  • USB32G1 indicates the bit rate is USB 3.2 Gen 1 (5 Gbps). Requires option SR-USB3.

dataminusthreshold property

dataminusthreshold: BusBItemUsbDataminusthreshold

Return the BUS:B<x>:USB:DATAMINUSTHRESHold command.

Description
  • This command sets or queries the USB D- source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:DATAMINUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:DATAMINUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:DATAMINUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:USB:DATAMINUSTHRESHold <NR3>
- BUS:B<x>:USB:DATAMINUSTHRESHold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the Minus threshold.

dataplusthreshold property

dataplusthreshold: BusBItemUsbDataplusthreshold

Return the BUS:B<x>:USB:DATAPLUSTHRESHold command.

Description
  • This command sets or queries the USB D+ source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:DATAPLUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:DATAPLUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:DATAPLUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:USB:DATAPLUSTHRESHold <NR3>
- BUS:B<x>:USB:DATAPLUSTHRESHold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the Plus threshold.

displaymode property

Return the BUS:B<x>:USB:DISplaymode command.

Description
  • This command sets or queries the USB bus packet view. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:DISplaymode? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:DISplaymode? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:DISplaymode value command.
SCPI Syntax
- BUS:B<x>:USB:DISplaymode {ON|OFF}
- BUS:B<x>:USB:DISplaymode?
Info
  • B<x> is the number of the bus waveform.
  • ON sets the packet view for the decode to on.
  • OFF sets the packet view for the decode to off.

laneone property

Return the BUS:B<x>:USB:LANEONE command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:USB:LANEONE? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:LANEONE? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .source: The BUS:B<x>:USB:LANEONE:SOUrce command.
  • .threshold: The BUS:B<x>:USB:LANEONE:THRESHold command.

lanes property

Return the BUS:B<x>:USB:LANES command.

Description
  • This command sets or queries the number of lanes for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:LANES? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:LANES? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:LANES value command.
SCPI Syntax
- BUS:B<x>:USB:LANES {SINGLe|DUAL}
- BUS:B<x>:USB:LANES?
Info
  • B<x> is the number of the bus waveform.
  • SINGLe specifies the number of lanes as single.
  • DUAL specifies the number of lanes as dual.

lanezero property

Return the BUS:B<x>:USB:LANEZERO command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:USB:LANEZERO? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:LANEZERO? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Sub-properties
  • .source: The BUS:B<x>:USB:LANEZERO:SOUrce command.
  • .threshold: The BUS:B<x>:USB:LANEZERO:THRESHold command.

lowthreshold property

Return the BUS:B<x>:USB:LOWTHRESHold command.

Description
  • This command sets or queries the USB data source threshold for the specified bus when the signal type is differential. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:LOWTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:LOWTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:LOWTHRESHold value command.
SCPI Syntax
- BUS:B<x>:USB:LOWTHRESHold <NR3>
- BUS:B<x>:USB:LOWTHRESHold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the Low threshold.

semaththreshold property

semaththreshold: BusBItemUsbSemaththreshold

Return the BUS:B<x>:USB:SEMATHTHRESHold command.

Description
  • This command sets or queries the single ended source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:SEMATHTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:SEMATHTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:SEMATHTHRESHold value command.
SCPI Syntax
- BUS:B<x>:USB:SEMATHTHRESHold <NR3>
- BUS:B<x>:USB:SEMATHTHRESHold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> specifies the single ended source threshold for the specified bus. The valid range is -8.0 to 8.0.

signaltype property

Return the BUS:B<x>:USB:SIGNALTYpe command.

Description
  • This command sets or queries the USB signal type for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:SIGNALTYpe? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:SIGNALTYpe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:SIGNALTYpe value command.
SCPI Syntax
- BUS:B<x>:USB:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:USB:SIGNALTYpe?
Info
  • B<x> is the number of the bus waveform.
  • SINGLE specifies single-ended signals.
  • DIFF specifies differential signals.

source property

Return the BUS:B<x>:USB:SOUrce command.

Description
  • This command sets or queries the USB data source when the signal type is differential for bus . The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:SOUrce value command.
SCPI Syntax
- BUS:B<x>:USB:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:USB:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the data source for the specified USB bus.
  • MATH<x> specifies a math channel as the data source for the specified USB bus.
  • REF<x> specifies a reference waveform as the data source.
Sub-properties
  • .dminus: The BUS:B<x>:USB:SOUrce:DMINus command.
  • .dplus: The BUS:B<x>:USB:SOUrce:DPLUs command.

threshold property

Return the BUS:B<x>:USB:THRESHold command.

Description
  • This command sets or queries the USB DATA source High threshold for the specified bus when the signal source is differential. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:THRESHold value command.
SCPI Syntax
- BUS:B<x>:USB:THRESHold <NR3>
- BUS:B<x>:USB:THRESHold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the USB DATA source High threshold for the specified bus.

type property

Return the BUS:B<x>:USB:TYPe command.

Description
  • This command sets or queries the bus type or standard for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:TYPe? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:TYPe value command.
SCPI Syntax
- BUS:B<x>:USB:TYPe {USB}
- BUS:B<x>:USB:TYPe?
Info
  • B<x> is the number of the bus waveform.
  • USB specifies the bus type as USB.

BusBItemUsbBitrate

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:USB:BITRate command.

Description
  • This command sets or queries the USB data rate for bus , where the bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:BITRate? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:BITRate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:BITRate value command.
SCPI Syntax
- BUS:B<x>:USB:BITRate {FULL|HIGH|LOW|USB3O|USB31G1|USB32G1}
- BUS:B<x>:USB:BITRate?
Info
  • B<x> is the number of the bus waveform.
  • FULL indicates the bit rate is 12 Mbps.
  • HIGH indicates the bit rate is 480 Mbps.
  • LOW indicates the bit rate is 1.5 Mbps.
  • USB3O indicates the bit rate is USB 3.0 (5 Gbps). Requires option SR-USB3.
  • USB31G1 indicates the bit rate is USB 3.1 Gen 1 (5 Gbps). Requires option SR-USB3.
  • USB32G1 indicates the bit rate is USB 3.2 Gen 1 (5 Gbps). Requires option SR-USB3.

BusBItemUsbDataminusthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:USB:DATAMINUSTHRESHold command.

Description
  • This command sets or queries the USB D- source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:DATAMINUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:DATAMINUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:DATAMINUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:USB:DATAMINUSTHRESHold <NR3>
- BUS:B<x>:USB:DATAMINUSTHRESHold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the Minus threshold.

BusBItemUsbDataplusthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:USB:DATAPLUSTHRESHold command.

Description
  • This command sets or queries the USB D+ source threshold for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:DATAPLUSTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:DATAPLUSTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:DATAPLUSTHRESHold value command.
SCPI Syntax
- BUS:B<x>:USB:DATAPLUSTHRESHold <NR3>
- BUS:B<x>:USB:DATAPLUSTHRESHold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the Plus threshold.

BusBItemUsbDisplaymode

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:USB:DISplaymode command.

Description
  • This command sets or queries the USB bus packet view. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:DISplaymode? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:DISplaymode? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:DISplaymode value command.
SCPI Syntax
- BUS:B<x>:USB:DISplaymode {ON|OFF}
- BUS:B<x>:USB:DISplaymode?
Info
  • B<x> is the number of the bus waveform.
  • ON sets the packet view for the decode to on.
  • OFF sets the packet view for the decode to off.

BusBItemUsbLaneone

Bases: SCPICmdRead

The BUS:B<x>:USB:LANEONE command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:USB:LANEONE? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:LANEONE? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .source: The BUS:B<x>:USB:LANEONE:SOUrce command.
  • .threshold: The BUS:B<x>:USB:LANEONE:THRESHold command.

source property

Return the BUS:B<x>:USB:LANEONE:SOUrce command.

Description
  • This command sets or queries the USB data (SDATA) of the Lane 1 source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:LANEONE:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:LANEONE:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:LANEONE:SOUrce value command.
SCPI Syntax
- BUS:B<x>:USB:LANEONE:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:USB:LANEONE:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the data source for the specified USB bus.
  • MATH<x> specifies a math channel as the data source for the specified USB bus.
  • REF<x> specifies a reference waveform as the data source.

threshold property

Return the BUS:B<x>:USB:LANEONE:THRESHold command.

Description
  • This command sets or queries the USB data (SDATA) of the Lane 1 threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:LANEONE:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:LANEONE:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:LANEONE:THRESHold value command.
SCPI Syntax
- BUS:B<x>:USB:LANEONE:THRESHold <NR3>
- BUS:B<x>:USB:LANEONE:THRESHold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> specifies the USB Lane 1 threshold level for the specified bus. The valid range is -8.0 to 8.0.

BusBItemUsbLaneoneSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:USB:LANEONE:SOUrce command.

Description
  • This command sets or queries the USB data (SDATA) of the Lane 1 source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:LANEONE:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:LANEONE:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:LANEONE:SOUrce value command.
SCPI Syntax
- BUS:B<x>:USB:LANEONE:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:USB:LANEONE:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the data source for the specified USB bus.
  • MATH<x> specifies a math channel as the data source for the specified USB bus.
  • REF<x> specifies a reference waveform as the data source.

BusBItemUsbLaneoneThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:USB:LANEONE:THRESHold command.

Description
  • This command sets or queries the USB data (SDATA) of the Lane 1 threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:LANEONE:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:LANEONE:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:LANEONE:THRESHold value command.
SCPI Syntax
- BUS:B<x>:USB:LANEONE:THRESHold <NR3>
- BUS:B<x>:USB:LANEONE:THRESHold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> specifies the USB Lane 1 threshold level for the specified bus. The valid range is -8.0 to 8.0.

BusBItemUsbLanes

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:USB:LANES command.

Description
  • This command sets or queries the number of lanes for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:LANES? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:LANES? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:LANES value command.
SCPI Syntax
- BUS:B<x>:USB:LANES {SINGLe|DUAL}
- BUS:B<x>:USB:LANES?
Info
  • B<x> is the number of the bus waveform.
  • SINGLe specifies the number of lanes as single.
  • DUAL specifies the number of lanes as dual.

BusBItemUsbLanezero

Bases: SCPICmdRead

The BUS:B<x>:USB:LANEZERO command tree.

Usage
  • Using the .query() method will send the BUS:B<x>:USB:LANEZERO? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:LANEZERO? query and raise an AssertionError if the returned value does not match value.
Info
  • B<x> is the number of the bus waveform.
Properties
  • .source: The BUS:B<x>:USB:LANEZERO:SOUrce command.
  • .threshold: The BUS:B<x>:USB:LANEZERO:THRESHold command.

source property

Return the BUS:B<x>:USB:LANEZERO:SOUrce command.

Description
  • This command sets or queries the USB data (SDATA) of the Lane 0 source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:LANEZERO:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:LANEZERO:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:LANEZERO:SOUrce value command.
SCPI Syntax
- BUS:B<x>:USB:LANEZERO:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:USB:LANEZERO:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the data source for the specified USB bus.
  • MATH<x> specifies a math channel as the data source for the specified USB bus.
  • REF<x> specifies a reference waveform as the data source.

threshold property

Return the BUS:B<x>:USB:LANEZERO:THRESHold command.

Description
  • This command sets or queries the USB data (SDATA) of the Lane 0 threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:LANEZERO:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:LANEZERO:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:LANEZERO:THRESHold value command.
SCPI Syntax
- BUS:B<x>:USB:LANEZERO:THRESHold <NR3>
- BUS:B<x>:USB:LANEZERO:THRESHold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> specifies the USB Lane 0 threshold level for the specified bus. The valid range is -8.0 to 8.0.

BusBItemUsbLanezeroSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:USB:LANEZERO:SOUrce command.

Description
  • This command sets or queries the USB data (SDATA) of the Lane 0 source for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:LANEZERO:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:LANEZERO:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:LANEZERO:SOUrce value command.
SCPI Syntax
- BUS:B<x>:USB:LANEZERO:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:USB:LANEZERO:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the data source for the specified USB bus.
  • MATH<x> specifies a math channel as the data source for the specified USB bus.
  • REF<x> specifies a reference waveform as the data source.

BusBItemUsbLanezeroThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:USB:LANEZERO:THRESHold command.

Description
  • This command sets or queries the USB data (SDATA) of the Lane 0 threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:LANEZERO:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:LANEZERO:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:LANEZERO:THRESHold value command.
SCPI Syntax
- BUS:B<x>:USB:LANEZERO:THRESHold <NR3>
- BUS:B<x>:USB:LANEZERO:THRESHold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> specifies the USB Lane 0 threshold level for the specified bus. The valid range is -8.0 to 8.0.

BusBItemUsbLowthreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:USB:LOWTHRESHold command.

Description
  • This command sets or queries the USB data source threshold for the specified bus when the signal type is differential. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:LOWTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:LOWTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:LOWTHRESHold value command.
SCPI Syntax
- BUS:B<x>:USB:LOWTHRESHold <NR3>
- BUS:B<x>:USB:LOWTHRESHold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the Low threshold.

BusBItemUsbSemaththreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:USB:SEMATHTHRESHold command.

Description
  • This command sets or queries the single ended source threshold for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:SEMATHTHRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:SEMATHTHRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:SEMATHTHRESHold value command.
SCPI Syntax
- BUS:B<x>:USB:SEMATHTHRESHold <NR3>
- BUS:B<x>:USB:SEMATHTHRESHold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> specifies the single ended source threshold for the specified bus. The valid range is -8.0 to 8.0.

BusBItemUsbSignaltype

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:USB:SIGNALTYpe command.

Description
  • This command sets or queries the USB signal type for the specified bus. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:SIGNALTYpe? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:SIGNALTYpe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:SIGNALTYpe value command.
SCPI Syntax
- BUS:B<x>:USB:SIGNALTYpe {SINGLE|DIFF}
- BUS:B<x>:USB:SIGNALTYpe?
Info
  • B<x> is the number of the bus waveform.
  • SINGLE specifies single-ended signals.
  • DIFF specifies differential signals.

BusBItemUsbSource

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:USB:SOUrce command.

Description
  • This command sets or queries the USB data source when the signal type is differential for bus . The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:SOUrce? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:SOUrce? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:SOUrce value command.
SCPI Syntax
- BUS:B<x>:USB:SOUrce {CH<x>|MATH<x>|REF<x>}
- BUS:B<x>:USB:SOUrce?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the data source for the specified USB bus.
  • MATH<x> specifies a math channel as the data source for the specified USB bus.
  • REF<x> specifies a reference waveform as the data source.
Properties
  • .dminus: The BUS:B<x>:USB:SOUrce:DMINus command.
  • .dplus: The BUS:B<x>:USB:SOUrce:DPLUs command.

dminus property

Return the BUS:B<x>:USB:SOUrce:DMINus command.

Description
  • This command sets or queries the USB D- (SDATAMINUS) source for bus when the signal type is single ended. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:SOUrce:DMINus? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:SOUrce:DMINus? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:SOUrce:DMINus value command.
SCPI Syntax
- BUS:B<x>:USB:SOUrce:DMINus {CH<x>|CH<x>_D<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:USB:SOUrce:DMINus?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the D- source for the specified USB bus.
  • CH<x>_D<x> specifies a digital channel as the D- source for the specified USB bus.
  • MATH<x> specifies a math channel as the D- source for the specified USB bus.
  • REF<x> specifies a reference waveform as the source.
  • REF<x>_D<x> specifies a digital reference waveform as the clock source waveform for the specified USB bus.

dplus property

Return the BUS:B<x>:USB:SOUrce:DPLUs command.

Description
  • This command sets or queries the USB dataPlus (SDATAPLUS) source for the specified bus when the signal type is single ended. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:SOUrce:DPLUs? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:SOUrce:DPLUs? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:SOUrce:DPLUs value command.
SCPI Syntax
- BUS:B<x>:USB:SOUrce:DPLUs {CH<x>|CH<x>_D<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:USB:SOUrce:DPLUs?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the D+ source for the specified USB bus.
  • CH<x>_D<x> specifies a digital channel as the D+ source for the specified USB bus.
  • MATH<x> specifies a math channel as the D+ source for the specified USB bus.
  • REF<x> specifies a reference waveform as the source.
  • REF<x>_D<x> specifies a digital reference waveform as the clock source waveform for the specified USB bus.

BusBItemUsbSourceDminus

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:USB:SOUrce:DMINus command.

Description
  • This command sets or queries the USB D- (SDATAMINUS) source for bus when the signal type is single ended. The bus number is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:SOUrce:DMINus? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:SOUrce:DMINus? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:SOUrce:DMINus value command.
SCPI Syntax
- BUS:B<x>:USB:SOUrce:DMINus {CH<x>|CH<x>_D<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:USB:SOUrce:DMINus?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the D- source for the specified USB bus.
  • CH<x>_D<x> specifies a digital channel as the D- source for the specified USB bus.
  • MATH<x> specifies a math channel as the D- source for the specified USB bus.
  • REF<x> specifies a reference waveform as the source.
  • REF<x>_D<x> specifies a digital reference waveform as the clock source waveform for the specified USB bus.

BusBItemUsbSourceDplus

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:USB:SOUrce:DPLUs command.

Description
  • This command sets or queries the USB dataPlus (SDATAPLUS) source for the specified bus when the signal type is single ended. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:SOUrce:DPLUs? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:SOUrce:DPLUs? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:SOUrce:DPLUs value command.
SCPI Syntax
- BUS:B<x>:USB:SOUrce:DPLUs {CH<x>|CH<x>_D<x>|MATH<x>|REF<x>|REF<x>_D<x>}
- BUS:B<x>:USB:SOUrce:DPLUs?
Info
  • B<x> is the number of the bus waveform.
  • CH<x> specifies an analog channel as the D+ source for the specified USB bus.
  • CH<x>_D<x> specifies a digital channel as the D+ source for the specified USB bus.
  • MATH<x> specifies a math channel as the D+ source for the specified USB bus.
  • REF<x> specifies a reference waveform as the source.
  • REF<x>_D<x> specifies a digital reference waveform as the clock source waveform for the specified USB bus.

BusBItemUsbThreshold

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:USB:THRESHold command.

Description
  • This command sets or queries the USB DATA source High threshold for the specified bus when the signal source is differential. The bus is specified by x.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:THRESHold? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:THRESHold? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:THRESHold value command.
SCPI Syntax
- BUS:B<x>:USB:THRESHold <NR3>
- BUS:B<x>:USB:THRESHold?
Info
  • B<x> is the number of the bus waveform.
  • <NR3> is the USB DATA source High threshold for the specified bus.

BusBItemUsbType

Bases: SCPICmdWrite, SCPICmdRead

The BUS:B<x>:USB:TYPe command.

Description
  • This command sets or queries the bus type or standard for the specified bus.
Usage
  • Using the .query() method will send the BUS:B<x>:USB:TYPe? query.
  • Using the .verify(value) method will send the BUS:B<x>:USB:TYPe? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the BUS:B<x>:USB:TYPe value command.
SCPI Syntax
- BUS:B<x>:USB:TYPe {USB}
- BUS:B<x>:USB:TYPe?
Info
  • B<x> is the number of the bus waveform.
  • USB specifies the bus type as USB.

BusDelete

Bases: SCPICmdWrite

The BUS:DELete command.

Description
  • This command deletes the specified bus.
Usage
  • Using the .write(value) method will send the BUS:DELete value command.
SCPI Syntax
- BUS:DELete <QString>
Info
  • <QString> specifies the bus to delete and is of the form ‘B‘, where is ≥1.

BusList

Bases: SCPICmdRead

The BUS:LIST command.

Description
  • This query returns a comma separated list of all currently defined buses.
Usage
  • Using the .query() method will send the BUS:LIST? query.
  • Using the .verify(value) method will send the BUS:LIST? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax