auxin

The auxin commands module.

These commands are used in the following models: DPO5K, DPO5KB, DPO70KC, DPO70KD, DPO70KDX, DPO70KSX, DPO7K, DPO7KC, DSA70KC, DSA70KD, MSO5K, MSO5KB, MSO70KC, MSO70KDX

THIS FILE IS AUTO-GENERATED, IT SHOULD NOT BE MANUALLY MODIFIED.

Please report an issue if one is found.

Commands and Queries
- AUXIn:BANdwidth <NR3>
- AUXIn:BANdwidth?
- AUXIn:COUPling <NR3>
- AUXIn:COUPling?
- AUXIn:OFFSet <NR3>
- AUXIn:OFFSet?
- AUXIn:PROBEFunc:EXTAtten <NR3>
- AUXIn:PROBEFunc:EXTAtten?
- AUXIn:PROBEFunc:EXTDBatten <NR3>
- AUXIn:PROBEFunc:EXTDBatten?
- AUXIn:PROBEFunc:EXTUnits <QString>
- AUXIn:PROBEFunc:EXTUnits?
- AUXIn:PRObe:AUTOZero EXECute
- AUXIn:PRObe:DEGAUSS EXECute
- AUXIn:PRObe:DEGAUSS:STATE?
- AUXIn:PRObe:FORCEDRange <NR3>
- AUXIn:PRObe:FORCEDRange?
- AUXIn:PRObe:GAIN?
- AUXIn:PRObe:ID:SERnumber?
- AUXIn:PRObe:ID:TYPe?
- AUXIn:PRObe:INPUTMode {COMmonmode|DEFault|DIFFerential|A|B}
- AUXIn:PRObe:INPUTMode:AOFFSet <NR3>
- AUXIn:PRObe:INPUTMode:AOFFSet?
- AUXIn:PRObe:INPUTMode:BOFFSet <NR3>
- AUXIn:PRObe:INPUTMode:BOFFSet?
- AUXIn:PRObe:INPUTMode:CMOFFSet <NR3>
- AUXIn:PRObe:INPUTMode:CMOFFSet?
- AUXIn:PRObe:INPUTMode:DMOFFSet <NR3>
- AUXIn:PRObe:INPUTMode:DMOFFSet?
- AUXIn:PRObe:INPUTMode?
- AUXIn:PRObe:RANge {ATTEN1X|ATTEN1_25X|ATTEN1_5X|ATTEN1_75X|ATTEN2X|ATTEN2_5X|ATTEN3X|ATTEN3_5X|ATTEN4X|ATTEN4_5X|ATTEN5X|ATTEN5_5X|ATTEN6X|ATTEN6_5X|ATTEN7X|ATTEN7_5X|ATTEN8X|ATTEN8_5X|ATTEN9X|ATTEN9_5X|ATTEN10X|ATTEN12_5X|ATTEN15X|ATTEN17_5X|ATTEN20X|ATTEN25X|ATTEN30X|ATTEN35X|ATTEN40X|ATTEN45X|ATTEN50X|ATTEN55X|ATTEN60X|ATTEN65X|ATTEN70X|ATTEN75X|ATTEN80X|ATTEN85X|ATTEN90X|ATTEN95X|ATTEN100X|ATTEN125X|ATTEN150X|ATTEN175X|ATTEN200X|ATTEN250X|ATTEN300X|ATTEN350X|ATTEN400X|ATTEN450X|ATTEN500X|ATTEN550X|ATTEN600X|ATTEN650X|ATTEN700X|ATTEN750X|ATTEN800X|ATTEN850X|ATTEN900X|ATTEN950X|ATTEN1000X}
- AUXIn:PRObe:RANge?
- AUXIn:PRObe:RESistance?
- AUXIn:PRObe:SET <QString>
- AUXIn:PRObe:SET?
- AUXIn:PRObe:SIGnal {PASS|BYPass}
- AUXIn:PRObe:SIGnal?
- AUXIn:PRObe:TIPtype {HBWStraightflex|OTHer}
- AUXIn:PRObe:TIPtype?
- AUXIn:PRObe:UNIts?
- AUXIn:VTERm:DUAL:A <NR3>
- AUXIn:VTERm:DUAL:A?
- AUXIn:VTERm:DUAL:B <NR3>
- AUXIn:VTERm:DUAL:B?

Auxin

Bases: SCPICmdRead

The AUXIn command tree.

Usage
  • Using the .query() method will send the AUXIn? query.
  • Using the .verify(value) method will send the AUXIn? query and raise an AssertionError if the returned value does not match value.
Properties
  • .bandwidth: The AUXIn:BANdwidth command.
  • .coupling: The AUXIn:COUPling command.
  • .offset: The AUXIn:OFFSet command.
  • .probefunc: The AUXIn:PROBEFunc command tree.
  • .probe: The AUXIn:PRObe command tree.
  • .vterm: The AUXIn:VTERm command tree.

bandwidth property

bandwidth: AuxinBandwidth

Return the AUXIn:BANdwidth command.

Description
  • This command sets or queries the selectable low-pass bandwidth limit filter of the AUX In input. The bandwidth will change with an appropriate probe attached. This is equivalent to selecting Bandwidth from the Vertical menu. The query form of this command always returns the maximum bandwidth of the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:BANdwidth? query.
  • Using the .verify(value) method will send the AUXIn:BANdwidth? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:BANdwidth value command.
SCPI Syntax
- AUXIn:BANdwidth <NR3>
- AUXIn:BANdwidth?
Info
  • <NR3> is one of the supported bandwidths on the attached probe.

coupling property

coupling: AuxinCoupling

Return the AUXIn:COUPling command.

Description
  • This command sets or queries the input attenuator coupling setting for the AUX In input. The coupling will change with an appropriate probe attached. This command is equivalent to selecting Coupling from the Vertical menu.
Usage
  • Using the .query() method will send the AUXIn:COUPling? query.
  • Using the .verify(value) method will send the AUXIn:COUPling? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:COUPling value command.
SCPI Syntax
- AUXIn:COUPling <NR3>
- AUXIn:COUPling?
Info
  • <NR3> is one of the supported input attenuator couplings on the attached probe.

offset property

offset: AuxinOffset

Return the AUXIn:OFFSet command.

Description
  • This command sets or queries the vertical offset for the AUX In input. This command is equivalent to selecting Offset from the Vertical menu. This command offsets the vertical acquisition window (moves the level at the vertical center of the acquisition window) for the input. Visualize offset as scrolling the acquisition window towards the top of a large signal for increased offset values, and scrolling towards the bottom for decreased offset values. The resolution of the vertical window sets the offset increment for this control. Offset adjusts only the vertical center of the acquisition window to help determine what data is acquired. The instrument always displays the input signal minus the offset value. The channel reference marker will move to the vertical graticule position given by the negative of the offset value divided by the scale factor, unless that position is off-screen. If the computed coordinate for the reference mark is off-screen, the mark moves to the nearest screen limit and changes from a right-pointing arrow ( →) to an arrow pointing in the appropriate off-screen direction.
Usage
  • Using the .query() method will send the AUXIn:OFFSet? query.
  • Using the .verify(value) method will send the AUXIn:OFFSet? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:OFFSet value command.
SCPI Syntax
- AUXIn:OFFSet <NR3>
- AUXIn:OFFSet?
Info
  • <NR3> is the offset value for the specified channel.

probe property

probe: AuxinProbe

Return the AUXIn:PRObe command tree.

Usage
  • Using the .query() method will send the AUXIn:PRObe? query.
  • Using the .verify(value) method will send the AUXIn:PRObe? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .autozero: The AUXIn:PRObe:AUTOZero command.
  • .degauss: The AUXIn:PRObe:DEGAUSS command.
  • .forcedrange: The AUXIn:PRObe:FORCEDRange command.
  • .gain: The AUXIn:PRObe:GAIN command.
  • .id: The AUXIn:PRObe:ID command tree.
  • .inputmode: The AUXIn:PRObe:INPUTMode command.
  • .range: The AUXIn:PRObe:RANge command.
  • .resistance: The AUXIn:PRObe:RESistance command.
  • .set: The AUXIn:PRObe:SET command.
  • .signal: The AUXIn:PRObe:SIGnal command.
  • .tiptype: The AUXIn:PRObe:TIPtype command.
  • .units: The AUXIn:PRObe:UNIts command.

probefunc property

probefunc: AuxinProbefunc

Return the AUXIn:PROBEFunc command tree.

Usage
  • Using the .query() method will send the AUXIn:PROBEFunc? query.
  • Using the .verify(value) method will send the AUXIn:PROBEFunc? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .extatten: The AUXIn:PROBEFunc:EXTAtten command.
  • .extdbatten: The AUXIn:PROBEFunc:EXTDBatten command.
  • .extunits: The AUXIn:PROBEFunc:EXTUnits command.

vterm property

vterm: AuxinVterm

Return the AUXIn:VTERm command tree.

Usage
  • Using the .query() method will send the AUXIn:VTERm? query.
  • Using the .verify(value) method will send the AUXIn:VTERm? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .dual: The AUXIn:VTERm:DUAL command tree.

AuxinBandwidth

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:BANdwidth command.

Description
  • This command sets or queries the selectable low-pass bandwidth limit filter of the AUX In input. The bandwidth will change with an appropriate probe attached. This is equivalent to selecting Bandwidth from the Vertical menu. The query form of this command always returns the maximum bandwidth of the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:BANdwidth? query.
  • Using the .verify(value) method will send the AUXIn:BANdwidth? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:BANdwidth value command.
SCPI Syntax
- AUXIn:BANdwidth <NR3>
- AUXIn:BANdwidth?
Info
  • <NR3> is one of the supported bandwidths on the attached probe.

AuxinCoupling

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:COUPling command.

Description
  • This command sets or queries the input attenuator coupling setting for the AUX In input. The coupling will change with an appropriate probe attached. This command is equivalent to selecting Coupling from the Vertical menu.
Usage
  • Using the .query() method will send the AUXIn:COUPling? query.
  • Using the .verify(value) method will send the AUXIn:COUPling? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:COUPling value command.
SCPI Syntax
- AUXIn:COUPling <NR3>
- AUXIn:COUPling?
Info
  • <NR3> is one of the supported input attenuator couplings on the attached probe.

AuxinOffset

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:OFFSet command.

Description
  • This command sets or queries the vertical offset for the AUX In input. This command is equivalent to selecting Offset from the Vertical menu. This command offsets the vertical acquisition window (moves the level at the vertical center of the acquisition window) for the input. Visualize offset as scrolling the acquisition window towards the top of a large signal for increased offset values, and scrolling towards the bottom for decreased offset values. The resolution of the vertical window sets the offset increment for this control. Offset adjusts only the vertical center of the acquisition window to help determine what data is acquired. The instrument always displays the input signal minus the offset value. The channel reference marker will move to the vertical graticule position given by the negative of the offset value divided by the scale factor, unless that position is off-screen. If the computed coordinate for the reference mark is off-screen, the mark moves to the nearest screen limit and changes from a right-pointing arrow ( →) to an arrow pointing in the appropriate off-screen direction.
Usage
  • Using the .query() method will send the AUXIn:OFFSet? query.
  • Using the .verify(value) method will send the AUXIn:OFFSet? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:OFFSet value command.
SCPI Syntax
- AUXIn:OFFSet <NR3>
- AUXIn:OFFSet?
Info
  • <NR3> is the offset value for the specified channel.

AuxinProbe

Bases: SCPICmdRead

The AUXIn:PRObe command tree.

Usage
  • Using the .query() method will send the AUXIn:PRObe? query.
  • Using the .verify(value) method will send the AUXIn:PRObe? query and raise an AssertionError if the returned value does not match value.
Properties
  • .autozero: The AUXIn:PRObe:AUTOZero command.
  • .degauss: The AUXIn:PRObe:DEGAUSS command.
  • .forcedrange: The AUXIn:PRObe:FORCEDRange command.
  • .gain: The AUXIn:PRObe:GAIN command.
  • .id: The AUXIn:PRObe:ID command tree.
  • .inputmode: The AUXIn:PRObe:INPUTMode command.
  • .range: The AUXIn:PRObe:RANge command.
  • .resistance: The AUXIn:PRObe:RESistance command.
  • .set: The AUXIn:PRObe:SET command.
  • .signal: The AUXIn:PRObe:SIGnal command.
  • .tiptype: The AUXIn:PRObe:TIPtype command.
  • .units: The AUXIn:PRObe:UNIts command.

autozero property

Return the AUXIn:PRObe:AUTOZero command.

Description
  • The command sets the probe attached to the AUX In input to autozero.
Usage
  • Using the .write(value) method will send the AUXIn:PRObe:AUTOZero value command.
SCPI Syntax
- AUXIn:PRObe:AUTOZero EXECute
Info
  • EXECute sets the probe to autozero.

degauss property

Return the AUXIn:PRObe:DEGAUSS command.

Description
  • This command starts a degauss cycle of the probe attached to the AUX In input. The degauss cycle will change with an appropriate probe attached.
Usage
  • Using the .write(value) method will send the AUXIn:PRObe:DEGAUSS value command.
SCPI Syntax
- AUXIn:PRObe:DEGAUSS EXECute
Info
  • EXECute starts a probe degauss cycle.
Sub-properties
  • .state: The AUXIn:PRObe:DEGAUSS:STATE command.

forcedrange property

forcedrange: AuxinProbeForcedrange

Return the AUXIn:PRObe:FORCEDRange command.

Description
  • This command sets the probe attached to the AUX In input to the specified range, or the command queries the probe range.
Usage
  • Using the .query() method will send the AUXIn:PRObe:FORCEDRange? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:FORCEDRange? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:FORCEDRange value command.
SCPI Syntax
- AUXIn:PRObe:FORCEDRange <NR3>
- AUXIn:PRObe:FORCEDRange?
Info
  • <NR3> specifies the probe range.

gain property

Return the AUXIn:PRObe:GAIN command.

Description
  • This query-only command returns the gain factor of the probe that is attached to the AUX In input. The ‘gain’ of a probe is the output divided by the input transfer ratio. For example, a common 10x probe has a gain of 0.1.
Usage
  • Using the .query() method will send the AUXIn:PRObe:GAIN? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:GAIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- AUXIn:PRObe:GAIN?

id property

Return the AUXIn:PRObe:ID command tree.

Usage
  • Using the .query() method will send the AUXIn:PRObe:ID? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:ID? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .sernumber: The AUXIn:PRObe:ID:SERnumber command.
  • .type: The AUXIn:PRObe:ID:TYPe command.

inputmode property

Return the AUXIn:PRObe:INPUTMode command.

Description
  • This command sets or queries one of the probe’s four input modes on the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:PRObe:INPUTMode? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:INPUTMode? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:INPUTMode value command.
SCPI Syntax
- AUXIn:PRObe:INPUTMode {COMmonmode|DEFault|DIFFerential|A|B}
- AUXIn:PRObe:INPUTMode?
Info
  • COMmonmode sets the probe to route common-mode signals to the host.
  • DEFault sets the probe to the default mode.
  • DIFFerential sets the probe to route differential signals to the host.
  • A sets the probe to route single-ended A signals to the host.
  • B sets the probe to route single-ended B signals to the host.
Sub-properties
  • .aoffset: The AUXIn:PRObe:INPUTMode:AOFFSet command.
  • .boffset: The AUXIn:PRObe:INPUTMode:BOFFSet command.
  • .cmoffset: The AUXIn:PRObe:INPUTMode:CMOFFSet command.
  • .dmoffset: The AUXIn:PRObe:INPUTMode:DMOFFSet command.

range property

Return the AUXIn:PRObe:RANge command.

Description
  • This command controls or queries the set attenuation range of the probe on the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:PRObe:RANge? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:RANge? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:RANge value command.
SCPI Syntax
- AUXIn:PRObe:RANge {ATTEN1X|ATTEN1_25X|ATTEN1_5X|ATTEN1_75X|ATTEN2X|ATTEN2_5X|ATTEN3X|ATTEN3_5X|ATTEN4X|ATTEN4_5X|ATTEN5X|ATTEN5_5X|ATTEN6X|ATTEN6_5X|ATTEN7X|ATTEN7_5X|ATTEN8X|ATTEN8_5X|ATTEN9X|ATTEN9_5X|ATTEN10X|ATTEN12_5X|ATTEN15X|ATTEN17_5X|ATTEN20X|ATTEN25X|ATTEN30X|ATTEN35X|ATTEN40X|ATTEN45X|ATTEN50X|ATTEN55X|ATTEN60X|ATTEN65X|ATTEN70X|ATTEN75X|ATTEN80X|ATTEN85X|ATTEN90X|ATTEN95X|ATTEN100X|ATTEN125X|ATTEN150X|ATTEN175X|ATTEN200X|ATTEN250X|ATTEN300X|ATTEN350X|ATTEN400X|ATTEN450X|ATTEN500X|ATTEN550X|ATTEN600X|ATTEN650X|ATTEN700X|ATTEN750X|ATTEN800X|ATTEN850X|ATTEN900X|ATTEN950X|ATTEN1000X}
- AUXIn:PRObe:RANge?

resistance property

Return the AUXIn:PRObe:RESistance command.

Description
  • This query-only command returns the resistance factor of the probe that is attached to the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:PRObe:RESistance? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:RESistance? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- AUXIn:PRObe:RESistance?

set_ property

Return the AUXIn:PRObe:SET command.

Description
  • This command sets or queries aspects of probe accessory user interfaces, for example probe attenuation factors. The available arguments for this command will vary depending on the accessory you attach to the instrument auxiliary trigger input. For the P7260 probe, you can select between two attenuation factors using either this GPIB command or the push switch on the probe. The probe enables the relevant path and adjusts the settings based on the characteristics of the path in use. The probe signal path selection is not kept in persistent storage. The probe will lose the selection if you reboot the instrument or remove the probe. Also, the instrument does not store the selection in the save/recall setup operation.
Usage
  • Using the .query() method will send the AUXIn:PRObe:SET? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:SET? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:SET value command.
SCPI Syntax
- AUXIn:PRObe:SET <QString>
- AUXIn:PRObe:SET?
Info
  • <QString> is a quoted string representing a settable aspect of the attached accessory.
  • ATTENUATION 5X sets the P7260 probe to ±0.75 V dynamic range with 6 GHz bandwidth and 5X attenuation.
  • ATTENUATION 25X sets the P7260 probe to ±3 V dynamic range with 6 GHz bandwidth and 25X attenuation.
  • VTERMsource AUTO sets the P7380SMA probe voltage termination source to auto.
  • VTERMsource INTernal sets the P7380SMA probe voltage termination source to internal.
  • VTERMsource EXTernal sets the P7380SMA probe voltage termination source to external.

signal property

Return the AUXIn:PRObe:SIGnal command.

Description
  • This command sets or queries aspects of probe accessory user interfaces. The available arguments for this command will vary depending on the accessory you attach to the instrument.
Usage
  • Using the .query() method will send the AUXIn:PRObe:SIGnal? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:SIGnal? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:SIGnal value command.
SCPI Syntax
- AUXIn:PRObe:SIGnal {PASS|BYPass}
- AUXIn:PRObe:SIGnal?
Info
  • PASS opens a relay passing your signal to the instrument.
  • BYPass closes a relay preventing your signal from reaching the instrument. During probe degauss, the signal should be bypassed.

tiptype property

Return the AUXIn:PRObe:TIPtype command.

Description
  • This command sets or queries the type of probe tip being used.
Usage
  • Using the .query() method will send the AUXIn:PRObe:TIPtype? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:TIPtype? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:TIPtype value command.
SCPI Syntax
- AUXIn:PRObe:TIPtype {HBWStraightflex|OTHer}
- AUXIn:PRObe:TIPtype?
Info
  • HBWStraightflex lets the instrument know you are using a high bandwidth straight-flex probe tip.
  • OTHer lets the instrument know you are not using a high bandwidth straight-flex probe tip.

units property

Return the AUXIn:PRObe:UNIts command.

Description
  • This query-only command returns a string describing the units of measure for the probe attached to the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:PRObe:UNIts? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:UNIts? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- AUXIn:PRObe:UNIts?

AuxinProbeAutozero

Bases: SCPICmdWrite

The AUXIn:PRObe:AUTOZero command.

Description
  • The command sets the probe attached to the AUX In input to autozero.
Usage
  • Using the .write(value) method will send the AUXIn:PRObe:AUTOZero value command.
SCPI Syntax
- AUXIn:PRObe:AUTOZero EXECute
Info
  • EXECute sets the probe to autozero.

AuxinProbeDegauss

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:PRObe:DEGAUSS command.

Description
  • This command starts a degauss cycle of the probe attached to the AUX In input. The degauss cycle will change with an appropriate probe attached.
Usage
  • Using the .write(value) method will send the AUXIn:PRObe:DEGAUSS value command.
SCPI Syntax
- AUXIn:PRObe:DEGAUSS EXECute
Info
  • EXECute starts a probe degauss cycle.
Properties
  • .state: The AUXIn:PRObe:DEGAUSS:STATE command.

state property

Return the AUXIn:PRObe:DEGAUSS:STATE command.

Description
  • This command queries whether the probe attached to the AUX In input is degaussed.
Usage
  • Using the .query() method will send the AUXIn:PRObe:DEGAUSS:STATE? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:DEGAUSS:STATE? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- AUXIn:PRObe:DEGAUSS:STATE?

AuxinProbeDegaussState

Bases: SCPICmdRead

The AUXIn:PRObe:DEGAUSS:STATE command.

Description
  • This command queries whether the probe attached to the AUX In input is degaussed.
Usage
  • Using the .query() method will send the AUXIn:PRObe:DEGAUSS:STATE? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:DEGAUSS:STATE? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- AUXIn:PRObe:DEGAUSS:STATE?

AuxinProbeForcedrange

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:PRObe:FORCEDRange command.

Description
  • This command sets the probe attached to the AUX In input to the specified range, or the command queries the probe range.
Usage
  • Using the .query() method will send the AUXIn:PRObe:FORCEDRange? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:FORCEDRange? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:FORCEDRange value command.
SCPI Syntax
- AUXIn:PRObe:FORCEDRange <NR3>
- AUXIn:PRObe:FORCEDRange?
Info
  • <NR3> specifies the probe range.

AuxinProbeGain

Bases: SCPICmdRead

The AUXIn:PRObe:GAIN command.

Description
  • This query-only command returns the gain factor of the probe that is attached to the AUX In input. The ‘gain’ of a probe is the output divided by the input transfer ratio. For example, a common 10x probe has a gain of 0.1.
Usage
  • Using the .query() method will send the AUXIn:PRObe:GAIN? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:GAIN? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- AUXIn:PRObe:GAIN?

AuxinProbeId

Bases: SCPICmdRead

The AUXIn:PRObe:ID command tree.

Usage
  • Using the .query() method will send the AUXIn:PRObe:ID? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:ID? query and raise an AssertionError if the returned value does not match value.
Properties
  • .sernumber: The AUXIn:PRObe:ID:SERnumber command.
  • .type: The AUXIn:PRObe:ID:TYPe command.

sernumber property

Return the AUXIn:PRObe:ID:SERnumber command.

Description
  • This query-only command returns the probe serial number.
Usage
  • Using the .query() method will send the AUXIn:PRObe:ID:SERnumber? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:ID:SERnumber? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- AUXIn:PRObe:ID:SERnumber?

type property

Return the AUXIn:PRObe:ID:TYPe command.

Description
  • This query-only command returns the probe type.
Usage
  • Using the .query() method will send the AUXIn:PRObe:ID:TYPe? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:ID:TYPe? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- AUXIn:PRObe:ID:TYPe?

AuxinProbeIdSernumber

Bases: SCPICmdRead

The AUXIn:PRObe:ID:SERnumber command.

Description
  • This query-only command returns the probe serial number.
Usage
  • Using the .query() method will send the AUXIn:PRObe:ID:SERnumber? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:ID:SERnumber? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- AUXIn:PRObe:ID:SERnumber?

AuxinProbeIdType

Bases: SCPICmdRead

The AUXIn:PRObe:ID:TYPe command.

Description
  • This query-only command returns the probe type.
Usage
  • Using the .query() method will send the AUXIn:PRObe:ID:TYPe? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:ID:TYPe? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- AUXIn:PRObe:ID:TYPe?

AuxinProbeInputmode

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:PRObe:INPUTMode command.

Description
  • This command sets or queries one of the probe’s four input modes on the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:PRObe:INPUTMode? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:INPUTMode? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:INPUTMode value command.
SCPI Syntax
- AUXIn:PRObe:INPUTMode {COMmonmode|DEFault|DIFFerential|A|B}
- AUXIn:PRObe:INPUTMode?
Info
  • COMmonmode sets the probe to route common-mode signals to the host.
  • DEFault sets the probe to the default mode.
  • DIFFerential sets the probe to route differential signals to the host.
  • A sets the probe to route single-ended A signals to the host.
  • B sets the probe to route single-ended B signals to the host.
Properties
  • .aoffset: The AUXIn:PRObe:INPUTMode:AOFFSet command.
  • .boffset: The AUXIn:PRObe:INPUTMode:BOFFSet command.
  • .cmoffset: The AUXIn:PRObe:INPUTMode:CMOFFSet command.
  • .dmoffset: The AUXIn:PRObe:INPUTMode:DMOFFSet command.

aoffset property

Return the AUXIn:PRObe:INPUTMode:AOFFSet command.

Description
  • This command sets or queries the requested A mode offset control of the probe that is attached to the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:PRObe:INPUTMode:AOFFSet? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:INPUTMode:AOFFSet? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:INPUTMode:AOFFSet value command.
SCPI Syntax
- AUXIn:PRObe:INPUTMode:AOFFSet <NR3>
- AUXIn:PRObe:INPUTMode:AOFFSet?
Info
  • <NR3> specifies the A mode offset control.

boffset property

Return the AUXIn:PRObe:INPUTMode:BOFFSet command.

Description
  • This command sets or queries the requested B mode offset control of the probe that is attached to the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:PRObe:INPUTMode:BOFFSet? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:INPUTMode:BOFFSet? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:INPUTMode:BOFFSet value command.
SCPI Syntax
- AUXIn:PRObe:INPUTMode:BOFFSet <NR3>
- AUXIn:PRObe:INPUTMode:BOFFSet?
Info
  • <NR3> specifies the B mode offset control.

cmoffset property

Return the AUXIn:PRObe:INPUTMode:CMOFFSet command.

Description
  • This command sets or queries the requested common mode offset control of the probe that is attached to the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:PRObe:INPUTMode:CMOFFSet? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:INPUTMode:CMOFFSet? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:INPUTMode:CMOFFSet value command.
SCPI Syntax
- AUXIn:PRObe:INPUTMode:CMOFFSet <NR3>
- AUXIn:PRObe:INPUTMode:CMOFFSet?
Info
  • <NR3> specifies the common mode offset control.

dmoffset property

Return the AUXIn:PRObe:INPUTMode:DMOFFSet command.

Description
  • This command sets or queries the requested differential mode offset control of the probe that is attached to the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:PRObe:INPUTMode:DMOFFSet? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:INPUTMode:DMOFFSet? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:INPUTMode:DMOFFSet value command.
SCPI Syntax
- AUXIn:PRObe:INPUTMode:DMOFFSet <NR3>
- AUXIn:PRObe:INPUTMode:DMOFFSet?
Info
  • <NR3> specifies the differential mode offset control.

AuxinProbeInputmodeAoffset

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:PRObe:INPUTMode:AOFFSet command.

Description
  • This command sets or queries the requested A mode offset control of the probe that is attached to the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:PRObe:INPUTMode:AOFFSet? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:INPUTMode:AOFFSet? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:INPUTMode:AOFFSet value command.
SCPI Syntax
- AUXIn:PRObe:INPUTMode:AOFFSet <NR3>
- AUXIn:PRObe:INPUTMode:AOFFSet?
Info
  • <NR3> specifies the A mode offset control.

AuxinProbeInputmodeBoffset

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:PRObe:INPUTMode:BOFFSet command.

Description
  • This command sets or queries the requested B mode offset control of the probe that is attached to the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:PRObe:INPUTMode:BOFFSet? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:INPUTMode:BOFFSet? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:INPUTMode:BOFFSet value command.
SCPI Syntax
- AUXIn:PRObe:INPUTMode:BOFFSet <NR3>
- AUXIn:PRObe:INPUTMode:BOFFSet?
Info
  • <NR3> specifies the B mode offset control.

AuxinProbeInputmodeCmoffset

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:PRObe:INPUTMode:CMOFFSet command.

Description
  • This command sets or queries the requested common mode offset control of the probe that is attached to the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:PRObe:INPUTMode:CMOFFSet? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:INPUTMode:CMOFFSet? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:INPUTMode:CMOFFSet value command.
SCPI Syntax
- AUXIn:PRObe:INPUTMode:CMOFFSet <NR3>
- AUXIn:PRObe:INPUTMode:CMOFFSet?
Info
  • <NR3> specifies the common mode offset control.

AuxinProbeInputmodeDmoffset

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:PRObe:INPUTMode:DMOFFSet command.

Description
  • This command sets or queries the requested differential mode offset control of the probe that is attached to the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:PRObe:INPUTMode:DMOFFSet? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:INPUTMode:DMOFFSet? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:INPUTMode:DMOFFSet value command.
SCPI Syntax
- AUXIn:PRObe:INPUTMode:DMOFFSet <NR3>
- AUXIn:PRObe:INPUTMode:DMOFFSet?
Info
  • <NR3> specifies the differential mode offset control.

AuxinProbeRange

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:PRObe:RANge command.

Description
  • This command controls or queries the set attenuation range of the probe on the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:PRObe:RANge? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:RANge? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:RANge value command.
SCPI Syntax
- AUXIn:PRObe:RANge {ATTEN1X|ATTEN1_25X|ATTEN1_5X|ATTEN1_75X|ATTEN2X|ATTEN2_5X|ATTEN3X|ATTEN3_5X|ATTEN4X|ATTEN4_5X|ATTEN5X|ATTEN5_5X|ATTEN6X|ATTEN6_5X|ATTEN7X|ATTEN7_5X|ATTEN8X|ATTEN8_5X|ATTEN9X|ATTEN9_5X|ATTEN10X|ATTEN12_5X|ATTEN15X|ATTEN17_5X|ATTEN20X|ATTEN25X|ATTEN30X|ATTEN35X|ATTEN40X|ATTEN45X|ATTEN50X|ATTEN55X|ATTEN60X|ATTEN65X|ATTEN70X|ATTEN75X|ATTEN80X|ATTEN85X|ATTEN90X|ATTEN95X|ATTEN100X|ATTEN125X|ATTEN150X|ATTEN175X|ATTEN200X|ATTEN250X|ATTEN300X|ATTEN350X|ATTEN400X|ATTEN450X|ATTEN500X|ATTEN550X|ATTEN600X|ATTEN650X|ATTEN700X|ATTEN750X|ATTEN800X|ATTEN850X|ATTEN900X|ATTEN950X|ATTEN1000X}
- AUXIn:PRObe:RANge?

AuxinProbeResistance

Bases: SCPICmdRead

The AUXIn:PRObe:RESistance command.

Description
  • This query-only command returns the resistance factor of the probe that is attached to the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:PRObe:RESistance? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:RESistance? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- AUXIn:PRObe:RESistance?

AuxinProbeSet

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:PRObe:SET command.

Description
  • This command sets or queries aspects of probe accessory user interfaces, for example probe attenuation factors. The available arguments for this command will vary depending on the accessory you attach to the instrument auxiliary trigger input. For the P7260 probe, you can select between two attenuation factors using either this GPIB command or the push switch on the probe. The probe enables the relevant path and adjusts the settings based on the characteristics of the path in use. The probe signal path selection is not kept in persistent storage. The probe will lose the selection if you reboot the instrument or remove the probe. Also, the instrument does not store the selection in the save/recall setup operation.
Usage
  • Using the .query() method will send the AUXIn:PRObe:SET? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:SET? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:SET value command.
SCPI Syntax
- AUXIn:PRObe:SET <QString>
- AUXIn:PRObe:SET?
Info
  • <QString> is a quoted string representing a settable aspect of the attached accessory.
  • ATTENUATION 5X sets the P7260 probe to ±0.75 V dynamic range with 6 GHz bandwidth and 5X attenuation.
  • ATTENUATION 25X sets the P7260 probe to ±3 V dynamic range with 6 GHz bandwidth and 25X attenuation.
  • VTERMsource AUTO sets the P7380SMA probe voltage termination source to auto.
  • VTERMsource INTernal sets the P7380SMA probe voltage termination source to internal.
  • VTERMsource EXTernal sets the P7380SMA probe voltage termination source to external.

AuxinProbeSignal

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:PRObe:SIGnal command.

Description
  • This command sets or queries aspects of probe accessory user interfaces. The available arguments for this command will vary depending on the accessory you attach to the instrument.
Usage
  • Using the .query() method will send the AUXIn:PRObe:SIGnal? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:SIGnal? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:SIGnal value command.
SCPI Syntax
- AUXIn:PRObe:SIGnal {PASS|BYPass}
- AUXIn:PRObe:SIGnal?
Info
  • PASS opens a relay passing your signal to the instrument.
  • BYPass closes a relay preventing your signal from reaching the instrument. During probe degauss, the signal should be bypassed.

AuxinProbeTiptype

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:PRObe:TIPtype command.

Description
  • This command sets or queries the type of probe tip being used.
Usage
  • Using the .query() method will send the AUXIn:PRObe:TIPtype? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:TIPtype? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PRObe:TIPtype value command.
SCPI Syntax
- AUXIn:PRObe:TIPtype {HBWStraightflex|OTHer}
- AUXIn:PRObe:TIPtype?
Info
  • HBWStraightflex lets the instrument know you are using a high bandwidth straight-flex probe tip.
  • OTHer lets the instrument know you are not using a high bandwidth straight-flex probe tip.

AuxinProbeUnits

Bases: SCPICmdRead

The AUXIn:PRObe:UNIts command.

Description
  • This query-only command returns a string describing the units of measure for the probe attached to the AUX In input.
Usage
  • Using the .query() method will send the AUXIn:PRObe:UNIts? query.
  • Using the .verify(value) method will send the AUXIn:PRObe:UNIts? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- AUXIn:PRObe:UNIts?

AuxinProbefunc

Bases: SCPICmdRead

The AUXIn:PROBEFunc command tree.

Usage
  • Using the .query() method will send the AUXIn:PROBEFunc? query.
  • Using the .verify(value) method will send the AUXIn:PROBEFunc? query and raise an AssertionError if the returned value does not match value.
Properties
  • .extatten: The AUXIn:PROBEFunc:EXTAtten command.
  • .extdbatten: The AUXIn:PROBEFunc:EXTDBatten command.
  • .extunits: The AUXIn:PROBEFunc:EXTUnits command.

extatten property

Return the AUXIn:PROBEFunc:EXTAtten command.

Description
  • This command sets the input-output ratio, of external attenuation or gain, between the signal and the AUX In input. The query form of this command returns the user-specified attenuation. Note that this command deals with an attenuation factor, not a gain factor, unlike CHX:PROBE (This command returns a value independent of the external attenuation). For example, if you specify a 20X attenuation factor, the commands return the following values (assuming that a 1x probe is presently attached, since the external attenuation is used in combination with the probe attenuation): AUXIN:PROBE:EXTA? 20.00E+0 AUXIN:PROBE? 1.0E+0 This command is equivalent to selecting Attenuation from the Vertical menu, and then either viewing or setting Ext Atten.
Usage
  • Using the .query() method will send the AUXIn:PROBEFunc:EXTAtten? query.
  • Using the .verify(value) method will send the AUXIn:PROBEFunc:EXTAtten? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PROBEFunc:EXTAtten value command.
SCPI Syntax
- AUXIn:PROBEFunc:EXTAtten <NR3>
- AUXIn:PROBEFunc:EXTAtten?
Info
  • <NR3> is the attenuation value, which is specified as a multiplier in the range from 1.00E-10 to 1.00E+10.

extdbatten property

Return the AUXIn:PROBEFunc:EXTDBatten command.

Description
  • This command sets the input-output ratio (expressed in decibel units) of external attenuation or gain between the signal and the instrument AUX In input. The query form of this command returns the user-specified attenuation in decibels: 1X = 0 dB, 10X = 20 dB, 100X = 40 dB, and similar things. This command is equivalent to selecting Attenuation from the Vertical menu, and then either viewing or setting Ext Att(dB).
Usage
  • Using the .query() method will send the AUXIn:PROBEFunc:EXTDBatten? query.
  • Using the .verify(value) method will send the AUXIn:PROBEFunc:EXTDBatten? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PROBEFunc:EXTDBatten value command.
SCPI Syntax
- AUXIn:PROBEFunc:EXTDBatten <NR3>
- AUXIn:PROBEFunc:EXTDBatten?
Info
  • <NR3> is the attenuation value, which is specified in the range from -200.00 dB to 200.00 dB.

extunits property

Return the AUXIn:PROBEFunc:EXTUnits command.

Description
  • This command sets the unit of measurement for the external attenuator of the AUX In input. There is also a corresponding query that returns the user-specified unit of measurement for the external attenuator. Unless these units are set to the factory default string value of ‘None’, they become the attenuated units of measurement for the input. It is assumed that the probe connected to the input is of the correct type to receive the output of the user’s external transducer or network.
Usage
  • Using the .query() method will send the AUXIn:PROBEFunc:EXTUnits? query.
  • Using the .verify(value) method will send the AUXIn:PROBEFunc:EXTUnits? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PROBEFunc:EXTUnits value command.
SCPI Syntax
- AUXIn:PROBEFunc:EXTUnits <QString>
- AUXIn:PROBEFunc:EXTUnits?
Info
  • <QString> can contain a string of up to eight characters to indicate the attenuation unit of measurement for the AUX In input. However, most instrument attenuators only display the first two characters.

AuxinProbefuncExtatten

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:PROBEFunc:EXTAtten command.

Description
  • This command sets the input-output ratio, of external attenuation or gain, between the signal and the AUX In input. The query form of this command returns the user-specified attenuation. Note that this command deals with an attenuation factor, not a gain factor, unlike CHX:PROBE (This command returns a value independent of the external attenuation). For example, if you specify a 20X attenuation factor, the commands return the following values (assuming that a 1x probe is presently attached, since the external attenuation is used in combination with the probe attenuation): AUXIN:PROBE:EXTA? 20.00E+0 AUXIN:PROBE? 1.0E+0 This command is equivalent to selecting Attenuation from the Vertical menu, and then either viewing or setting Ext Atten.
Usage
  • Using the .query() method will send the AUXIn:PROBEFunc:EXTAtten? query.
  • Using the .verify(value) method will send the AUXIn:PROBEFunc:EXTAtten? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PROBEFunc:EXTAtten value command.
SCPI Syntax
- AUXIn:PROBEFunc:EXTAtten <NR3>
- AUXIn:PROBEFunc:EXTAtten?
Info
  • <NR3> is the attenuation value, which is specified as a multiplier in the range from 1.00E-10 to 1.00E+10.

AuxinProbefuncExtdbatten

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:PROBEFunc:EXTDBatten command.

Description
  • This command sets the input-output ratio (expressed in decibel units) of external attenuation or gain between the signal and the instrument AUX In input. The query form of this command returns the user-specified attenuation in decibels: 1X = 0 dB, 10X = 20 dB, 100X = 40 dB, and similar things. This command is equivalent to selecting Attenuation from the Vertical menu, and then either viewing or setting Ext Att(dB).
Usage
  • Using the .query() method will send the AUXIn:PROBEFunc:EXTDBatten? query.
  • Using the .verify(value) method will send the AUXIn:PROBEFunc:EXTDBatten? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PROBEFunc:EXTDBatten value command.
SCPI Syntax
- AUXIn:PROBEFunc:EXTDBatten <NR3>
- AUXIn:PROBEFunc:EXTDBatten?
Info
  • <NR3> is the attenuation value, which is specified in the range from -200.00 dB to 200.00 dB.

AuxinProbefuncExtunits

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:PROBEFunc:EXTUnits command.

Description
  • This command sets the unit of measurement for the external attenuator of the AUX In input. There is also a corresponding query that returns the user-specified unit of measurement for the external attenuator. Unless these units are set to the factory default string value of ‘None’, they become the attenuated units of measurement for the input. It is assumed that the probe connected to the input is of the correct type to receive the output of the user’s external transducer or network.
Usage
  • Using the .query() method will send the AUXIn:PROBEFunc:EXTUnits? query.
  • Using the .verify(value) method will send the AUXIn:PROBEFunc:EXTUnits? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:PROBEFunc:EXTUnits value command.
SCPI Syntax
- AUXIn:PROBEFunc:EXTUnits <QString>
- AUXIn:PROBEFunc:EXTUnits?
Info
  • <QString> can contain a string of up to eight characters to indicate the attenuation unit of measurement for the AUX In input. However, most instrument attenuators only display the first two characters.

AuxinVterm

Bases: SCPICmdRead

The AUXIn:VTERm command tree.

Usage
  • Using the .query() method will send the AUXIn:VTERm? query.
  • Using the .verify(value) method will send the AUXIn:VTERm? query and raise an AssertionError if the returned value does not match value.
Properties
  • .dual: The AUXIn:VTERm:DUAL command tree.

dual property

Return the AUXIn:VTERm:DUAL command tree.

Usage
  • Using the .query() method will send the AUXIn:VTERm:DUAL? query.
  • Using the .verify(value) method will send the AUXIn:VTERm:DUAL? query and raise an AssertionError if the returned value does not match value.
Sub-properties
  • .a: The AUXIn:VTERm:DUAL:A command.
  • .b: The AUXIn:VTERm:DUAL:B command.

AuxinVtermDual

Bases: SCPICmdRead

The AUXIn:VTERm:DUAL command tree.

Usage
  • Using the .query() method will send the AUXIn:VTERm:DUAL? query.
  • Using the .verify(value) method will send the AUXIn:VTERm:DUAL? query and raise an AssertionError if the returned value does not match value.
Properties
  • .a: The AUXIn:VTERm:DUAL:A command.
  • .b: The AUXIn:VTERm:DUAL:B command.

a property

Return the AUXIn:VTERm:DUAL:A command.

Description
  • This command sets or queries the termination voltage for probes with dual inputs that support settable termination voltage.
Usage
  • Using the .query() method will send the AUXIn:VTERm:DUAL:A? query.
  • Using the .verify(value) method will send the AUXIn:VTERm:DUAL:A? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:VTERm:DUAL:A value command.
SCPI Syntax
- AUXIn:VTERm:DUAL:A <NR3>
- AUXIn:VTERm:DUAL:A?
Info
  • <NR3> specifies the termination voltage.

b property

Return the AUXIn:VTERm:DUAL:B command.

Description
  • This command sets or queries the termination voltage for probes with dual inputs that support settable termination voltage.
Usage
  • Using the .query() method will send the AUXIn:VTERm:DUAL:B? query.
  • Using the .verify(value) method will send the AUXIn:VTERm:DUAL:B? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:VTERm:DUAL:B value command.
SCPI Syntax
- AUXIn:VTERm:DUAL:B <NR3>
- AUXIn:VTERm:DUAL:B?
Info
  • <NR3> specifies the termination voltage.

AuxinVtermDualA

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:VTERm:DUAL:A command.

Description
  • This command sets or queries the termination voltage for probes with dual inputs that support settable termination voltage.
Usage
  • Using the .query() method will send the AUXIn:VTERm:DUAL:A? query.
  • Using the .verify(value) method will send the AUXIn:VTERm:DUAL:A? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:VTERm:DUAL:A value command.
SCPI Syntax
- AUXIn:VTERm:DUAL:A <NR3>
- AUXIn:VTERm:DUAL:A?
Info
  • <NR3> specifies the termination voltage.

AuxinVtermDualB

Bases: SCPICmdWrite, SCPICmdRead

The AUXIn:VTERm:DUAL:B command.

Description
  • This command sets or queries the termination voltage for probes with dual inputs that support settable termination voltage.
Usage
  • Using the .query() method will send the AUXIn:VTERm:DUAL:B? query.
  • Using the .verify(value) method will send the AUXIn:VTERm:DUAL:B? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the AUXIn:VTERm:DUAL:B value command.
SCPI Syntax
- AUXIn:VTERm:DUAL:B <NR3>
- AUXIn:VTERm:DUAL:B?
Info
  • <NR3> specifies the termination voltage.