factory

The factory commands module.

These commands are used in the following models: DPO2K, DPO2KB, DPO4K, DPO4KB, DPO5K, DPO5KB, DPO70KC, DPO70KD, DPO70KDX, DPO70KSX, DPO7AX, DPO7K, DPO7KC, DSA70KC, DSA70KD, LPD6, MDO3, MDO3K, MDO4K, MDO4KB, MDO4KC, MSO2, MSO2K, MSO2KB, MSO4, MSO4B, MSO4K, MSO4KB, MSO5, MSO5B, MSO5K, MSO5KB, MSO5LP, MSO6, MSO6B, MSO70KC, MSO70KDX, TekScopePC

THIS FILE IS AUTO-GENERATED, IT SHOULD NOT BE MANUALLY MODIFIED.

Please report an issue if one is found.

Commands and Queries
- FACtory

Factory

Bases: SCPICmdWriteNoArguments

The FACtory command.

Description
  • This command (no query form) resets the instrument to its factory default settings. This command is equivalent to pressing the DEFAULT SETUP button located on the instrument front panel or selecting Default Setup from the File menu. This command Performs the following in addition to what is done for the *RST command: Clears any pending OPC operations. Resets the following IEEE488.2 registers: *ESE 0 (Event Status Enable Register) *SRE 0 (Service Request Enable Register) DESE 255 (Device Event Status Enable Register) *PSC 1 (Power-on Status Clear Flag) Deletes all defined aliases. Enables command headers (:HEADer 1).
Usage
  • Using the .write() method will send the FACtory command.
SCPI Syntax
- FACtory