autoset
¶
The autoset commands module.
These commands are used in the following models: DPO5K, DPO5KB, DPO70KC, DPO70KD, DPO70KDX, DPO70KSX, DPO7K, DPO7KC, DSA70KC, DSA70KD, MSO5K, MSO5KB, MSO70KC, MSO70KDX
THIS FILE IS AUTO-GENERATED, IT SHOULD NOT BE MANUALLY MODIFIED.
Please report an issue if one is found.
Commands and Queries
- AUTOSet {EXECute|UNDo|VFields|VIDeo|VLines}
- AUTOSet:OVErlay {ON|OFF}
- AUTOSet:PERcent <NR3>
Autoset
¶
Bases: SCPICmdWrite, SCPICmdRead
The AUTOSet command.
Description
- This command (no query format) sets the vertical, horizontal, and trigger controls of the instrument to automatically acquire and display the selected waveform. (To autoset a video waveform, the video trigger must be set to video standard, not custom. Video arguments require video hardware.) This is equivalent to pressing the front panel AUTOSET button. For a detailed description of autoset functionality, see Autoset in the index of the online help for your instrument.
Usage
- Using the
.write(value)method will send theAUTOSet valuecommand.
SCPI Syntax
- AUTOSet {EXECute|UNDo|VFields|VIDeo|VLines}
Info
EXECuteruns the autoset routine; this is equivalent to pressing the front panel AUTOSET button. If the display is set to a PAL, MV, or IRE graticule, this argument forces the graticule display to full mode (frame, grid, and cross hair).UNDoreturns the instrument to the setting prior to executing an autoset.VFieldsautosets the displayed waveform.VIDeoautosets the displayed waveform.VLinesautosets the displayed waveform.
Properties
.overlay: TheAUTOSet:OVErlaycommand..percent: TheAUTOSet:PERcentcommand.
overlay
property
¶
overlay: AutosetOverlay
Return the AUTOSet:OVErlay command.
Description
- Sets or queries the autoset overlay feature. If Autoset Overlay is turned on, Autoset will place all signals center vertically on-screen with the vertical settings on each channel configured so that the amplitude of the signal occupies n divisions = Percent/10 divisions.
Usage
- Using the
.write(value)method will send theAUTOSet:OVErlay valuecommand.
SCPI Syntax
- AUTOSet:OVErlay {ON|OFF}
percent
property
¶
percent: AutosetPercent
Return the AUTOSet:PERcent command.
Description
- Sets or queries the percent of the full vertical scale that the signal should cover after Autoset is run when Overlay is turned on.
Usage
- Using the
.write(value)method will send theAUTOSet:PERcent valuecommand.
SCPI Syntax
- AUTOSet:PERcent <NR3>
AutosetOverlay
¶
Bases: SCPICmdWrite
The AUTOSet:OVErlay command.
Description
- Sets or queries the autoset overlay feature. If Autoset Overlay is turned on, Autoset will place all signals center vertically on-screen with the vertical settings on each channel configured so that the amplitude of the signal occupies n divisions = Percent/10 divisions.
Usage
- Using the
.write(value)method will send theAUTOSet:OVErlay valuecommand.
SCPI Syntax
- AUTOSet:OVErlay {ON|OFF}
AutosetPercent
¶
Bases: SCPICmdWrite
The AUTOSet:PERcent command.
Description
- Sets or queries the percent of the full vertical scale that the signal should cover after Autoset is run when Overlay is turned on.
Usage
- Using the
.write(value)method will send theAUTOSet:PERcent valuecommand.
SCPI Syntax
- AUTOSet:PERcent <NR3>