calibration

The calibration commands module.

These commands are used in the following models: DPO7AX, LPD6, MSO4, MSO4B, MSO5, MSO5B, MSO5LP, MSO6, MSO6B

THIS FILE IS AUTO-GENERATED, IT SHOULD NOT BE MANUALLY MODIFIED.

Please report an issue if one is found.

Commands and Queries
- *CAL?

Cal

Bases: SCPICmdRead

The *CAL command.

Description
  • This query-only command starts signal path calibration (SPC) and returns the status upon completion. Note: When running SPC through the remote interface, calibration status cannot be obtained until after the SPC completes. SPC takes approximately 15 minutes per channel which means a total of 2 hours on an 8-channel model. Any remote command that performs an action on the instrument is also disabled until the SPC is complete.
Usage
  • Using the .query() method will send the *CAL? query.
  • Using the .verify(value) method will send the *CAL? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- *CAL?