power
¶
The power commands module.
These commands are used in the following models: LPD6, MSO4, MSO4B, MSO5, MSO5B, MSO5LP, MSO6, MSO6B
THIS FILE IS AUTO-GENERATED, IT SHOULD NOT BE MANUALLY MODIFIED.
Please report an issue if one is found.
Commands and Queries
- POWer:ADDNew POWER<x>
- POWer:DELete POWER<x>
- POWer:POWer<x>:AUTOSet EXECute
- POWer:POWer<x>:CLRESPONSE:AMP<x>Val <NR3>
- POWer:POWer<x>:CLRESPONSE:AMP<x>Val?
- POWer:POWer<x>:CLRESPONSE:AMPMode {CONSTant|PROFile}
- POWer:POWer<x>:CLRESPONSE:AMPMode?
- POWer:POWer<x>:CLRESPONSE:ANALYSISMethod {SV|FFT}
- POWer:POWer<x>:CLRESPONSE:ANALYSISMethod?
- POWer:POWer<x>:CLRESPONSE:AUTORbw {True|False}
- POWer:POWer<x>:CLRESPONSE:AUTORbw?
- POWer:POWer<x>:CLRESPONSE:CONNECTSTATus?
- POWer:POWer<x>:CLRESPONSE:CONSTAMPlitude <NR3>
- POWer:POWer<x>:CLRESPONSE:CONSTAMPlitude?
- POWer:POWer<x>:CLRESPONSE:FREQ<x>Val <NR3>
- POWer:POWer<x>:CLRESPONSE:FREQ<x>Val?
- POWer:POWer<x>:CLRESPONSE:GENIPADDress <NR2>
- POWer:POWer<x>:CLRESPONSE:GENIPADDress?
- POWer:POWer<x>:CLRESPONSE:GENerator {INTernal|EXTernal}
- POWer:POWer<x>:CLRESPONSE:GENerator?
- POWer:POWer<x>:CLRESPONSE:IMPEDance {FIFTy|HIGHZ}
- POWer:POWer<x>:CLRESPONSE:IMPEDance?
- POWer:POWer<x>:CLRESPONSE:INPUTSOurce CH<x>
- POWer:POWer<x>:CLRESPONSE:INPUTSOurce?
- POWer:POWer<x>:CLRESPONSE:MON {SAVG|SSEQ}
- POWer:POWer<x>:CLRESPONSE:MON?
- POWer:POWer<x>:CLRESPONSE:OUTPUTSOurce CH<x>
- POWer:POWer<x>:CLRESPONSE:OUTPUTSOurce?
- POWer:POWer<x>:CLRESPONSE:PPD <NR3>
- POWer:POWer<x>:CLRESPONSE:PPD?
- POWer:POWer<x>:CLRESPONSE:STARTFREQuency <NR3>
- POWer:POWer<x>:CLRESPONSE:STARTFREQuency?
- POWer:POWer<x>:CLRESPONSE:STOPFREQuency <NR3>
- POWer:POWer<x>:CLRESPONSE:STOPFREQuency?
- POWer:POWer<x>:CLRESPONSE:TESTCONNection EXECute
- POWer:POWer<x>:CYCLEAmp:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLEAmp:INPUTSOurce?
- POWer:POWer<x>:CYCLEBase:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLEBase:INPUTSOurce?
- POWer:POWer<x>:CYCLEMAX:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLEMAX:INPUTSOurce?
- POWer:POWer<x>:CYCLEMin:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLEMin:INPUTSOurce?
- POWer:POWer<x>:CYCLEPKPK:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLEPKPK:INPUTSOurce?
- POWer:POWer<x>:CYCLETop:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLETop:INPUTSOurce?
- POWer:POWer<x>:DIDT:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:DIDT:INPUTSOurce?
- POWer:POWer<x>:DIDT:SOURCEEDGEType {RISE|FALL}
- POWer:POWer<x>:DIDT:SOURCEEDGEType?
- POWer:POWer<x>:DVDT:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:DVDT:INPUTSOurce?
- POWer:POWer<x>:DVDT:SOURCEEDGEType {RISE|FALL}
- POWer:POWer<x>:DVDT:SOURCEEDGEType?
- POWer:POWer<x>:EFFICIENCY:INPUTType {AC|DC}
- POWer:POWer<x>:EFFICIENCY:INPUTType?
- POWer:POWer<x>:EFFICIENCY:IOUT1SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:IOUT1SOUrce?
- POWer:POWer<x>:EFFICIENCY:IOUT2SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:IOUT2SOUrce?
- POWer:POWer<x>:EFFICIENCY:IOUT3SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:IOUT3SOUrce?
- POWer:POWer<x>:EFFICIENCY:ISOUrce
- POWer:POWer<x>:EFFICIENCY:ISOUrce?
- POWer:POWer<x>:EFFICIENCY:NUMOFOutputs {ONE|TWO|THREE}
- POWer:POWer<x>:EFFICIENCY:NUMOFOutputs?
- POWer:POWer<x>:EFFICIENCY:OUTPUT1Type {AC|DC}
- POWer:POWer<x>:EFFICIENCY:OUTPUT1Type?
- POWer:POWer<x>:EFFICIENCY:OUTPUT2Type {AC|DC}
- POWer:POWer<x>:EFFICIENCY:OUTPUT2Type?
- POWer:POWer<x>:EFFICIENCY:OUTPUT3Type {AC|DC}
- POWer:POWer<x>:EFFICIENCY:OUTPUT3Type?
- POWer:POWer<x>:EFFICIENCY:OUTPUTType {AC|DC}
- POWer:POWer<x>:EFFICIENCY:OUTPUTType?
- POWer:POWer<x>:EFFICIENCY:VOUT1SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:VOUT1SOUrce?
- POWer:POWer<x>:EFFICIENCY:VOUT2SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:VOUT2SOUrce?
- POWer:POWer<x>:EFFICIENCY:VOUT3SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:VOUT3SOUrce?
- POWer:POWer<x>:EFFICIENCY:VSOUrce
- POWer:POWer<x>:EFFICIENCY:VSOUrce?
- POWer:POWer<x>:FREQUENCY:EDGe {RISE|FALL}
- POWer:POWer<x>:FREQUENCY:EDGe?
- POWer:POWer<x>:FREQUENCY:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:FREQUENCY:INPUTSOurce?
- POWer:POWer<x>:GATing {NONE|CURSOR|SCREEN|LOGIC}
- POWer:POWer<x>:GATing:GLOBal {ON|OFF|1|0}
- POWer:POWer<x>:GATing:GLOBal?
- POWer:POWer<x>:GATing?
- POWer:POWer<x>:HARMONICS:CLASs {CLASSA|CLASSB|CLASSC|CLASSD}
- POWer:POWer<x>:HARMONICS:CLASs?
- POWer:POWer<x>:HARMONICS:CLFile <QString>
- POWer:POWer<x>:HARMONICS:CLFile?
- POWer:POWer<x>:HARMONICS:CMEThod {RATed|MEASured}
- POWer:POWer<x>:HARMONICS:CMEThod?
- POWer:POWer<x>:HARMONICS:FUNDCURRent <NR1>
- POWer:POWer<x>:HARMONICS:FUNDCURRent?
- POWer:POWer<x>:HARMONICS:HORDer <NR1>
- POWer:POWer<x>:HARMONICS:HORDer?
- POWer:POWer<x>:HARMONICS:HSOURce {CURRent|VOLTage}
- POWer:POWer<x>:HARMONICS:HSOURce?
- POWer:POWer<x>:HARMONICS:IPOWer <NR1>
- POWer:POWer<x>:HARMONICS:IPOWer?
- POWer:POWer<x>:HARMONICS:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:HARMONICS:ISOURce?
- POWer:POWer<x>:HARMONICS:LINEFREQUEncy {Auto|FIFTyhz|SIXTyhz|THREESIXTyhz|FOURHUNDREdhz|SIXFIFTyhz|EIGHTHUNDREdhz|CUSTom}
- POWer:POWer<x>:HARMONICS:LINEFREQUEncy?
- POWer:POWer<x>:HARMONICS:ODDEVen {ALL|ODD|EVEN}
- POWer:POWer<x>:HARMONICS:ODDEVen?
- POWer:POWer<x>:HARMONICS:PFACtor <NR1>
- POWer:POWer<x>:HARMONICS:PFACtor?
- POWer:POWer<x>:HARMONICS:POWERRating {HIGH|LOW}
- POWer:POWer<x>:HARMONICS:POWERRating?
- POWer:POWer<x>:HARMONICS:RCURRent <NR1>
- POWer:POWer<x>:HARMONICS:RCURRent?
- POWer:POWer<x>:HARMONICS:STANDard {NONe|IEC|MIL|AM14|DO160|CUSTOM}
- POWer:POWer<x>:HARMONICS:STANDard?
- POWer:POWer<x>:HARMONICS:STARTFREQUEncy <NR2>
- POWer:POWer<x>:HARMONICS:STARTFREQUEncy?
- POWer:POWer<x>:HARMONICS:UNITs {LOG|LINear}
- POWer:POWer<x>:HARMONICS:UNITs?
- POWer:POWer<x>:HARMONICS:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:HARMONICS:VSOURce?
- POWer:POWer<x>:IMPEDANCE:AMP<x>Val <NR3>
- POWer:POWer<x>:IMPEDANCE:AMP<x>Val?
- POWer:POWer<x>:IMPEDANCE:AMPMode {CONSTant|PROFile}
- POWer:POWer<x>:IMPEDANCE:AMPMode?
- POWer:POWer<x>:IMPEDANCE:ANALYSISMethod {SV|FFT}
- POWer:POWer<x>:IMPEDANCE:ANALYSISMethod?
- POWer:POWer<x>:IMPEDANCE:AUTORbw {True|False}
- POWer:POWer<x>:IMPEDANCE:AUTORbw?
- POWer:POWer<x>:IMPEDANCE:CONNECTSTATus?
- POWer:POWer<x>:IMPEDANCE:CONSTAMPlitude <NR3>
- POWer:POWer<x>:IMPEDANCE:CONSTAMPlitude?
- POWer:POWer<x>:IMPEDANCE:FREQ<x>Val <NR3>
- POWer:POWer<x>:IMPEDANCE:FREQ<x>Val?
- POWer:POWer<x>:IMPEDANCE:GENIPADDress <QString>
- POWer:POWer<x>:IMPEDANCE:GENIPADDress?
- POWer:POWer<x>:IMPEDANCE:GENerator {INTernal|EXTernal}
- POWer:POWer<x>:IMPEDANCE:GENerator?
- POWer:POWer<x>:IMPEDANCE:IMPEDANCE {FIFTy|HIGHZ}
- POWer:POWer<x>:IMPEDANCE:IMPEDANCE?
- POWer:POWer<x>:IMPEDANCE:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:IMPEDANCE:INPUTSOurce?
- POWer:POWer<x>:IMPEDANCE:OUTPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:IMPEDANCE:OUTPUTSOurce?
- POWer:POWer<x>:IMPEDANCE:PPD <NR1>
- POWer:POWer<x>:IMPEDANCE:PPD?
- POWer:POWer<x>:IMPEDANCE:STARTFREQuency <NR3>
- POWer:POWer<x>:IMPEDANCE:STARTFREQuency?
- POWer:POWer<x>:IMPEDANCE:STOPFREQuency <NR3>
- POWer:POWer<x>:IMPEDANCE:STOPFREQuency?
- POWer:POWer<x>:IMPEDANCE:TESTCONNection {EXECute}
- POWer:POWer<x>:INDUCTANCE:EDGESource {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:INDUCTANCE:EDGESource?
- POWer:POWer<x>:INDUCTANCE:ISOUrce
- POWer:POWer<x>:INDUCTANCE:ISOUrce?
- POWer:POWer<x>:INDUCTANCE:VSOURce
- POWer:POWer<x>:INDUCTANCE:VSOURce?
- POWer:POWer<x>:INPUTCAP:ISOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:INPUTCAP:ISOURce?
- POWer:POWer<x>:INPUTCAP:PEAKCURRent <NR3>
- POWer:POWer<x>:INPUTCAP:PEAKCURRent?
- POWer:POWer<x>:INPUTCAP:PEAKVOLTage <NR3>
- POWer:POWer<x>:INPUTCAP:PEAKVOLTage?
- POWer:POWer<x>:INPUTCAP:VSOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:INPUTCAP:VSOURce?
- POWer:POWer<x>:INRUSHcurrent:INPUTSOurce
- POWer:POWer<x>:INRUSHcurrent:INPUTSOurce?
- POWer:POWer<x>:INRUSHcurrent:PEAKCURRent <NR3>
- POWer:POWer<x>:INRUSHcurrent:PEAKCURRent?
- POWer:POWer<x>:IVSINTEGRALV:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:IVSINTEGRALV:ISOURce?
- POWer:POWer<x>:IVSINTEGRALV:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:IVSINTEGRALV:VSOURce?
- POWer:POWer<x>:LABel <QString>
- POWer:POWer<x>:LABel?
- POWer:POWer<x>:LINERIPPLE:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:LINERIPPLE:INPUTSOurce?
- POWer:POWer<x>:LINERIPPLE:LFREQuency {FIFty|SIXty|FOURHundred}
- POWer:POWer<x>:LINERIPPLE:LFREQuency?
- POWer:POWer<x>:MAGNETICLOSS:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGNETICLOSS:ISOURce?
- POWer:POWer<x>:MAGNETICLOSS:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGNETICLOSS:VSOURce?
- POWer:POWer<x>:MAGPROPERTY:AREAofcrosssection <NR2>
- POWer:POWer<x>:MAGPROPERTY:AREAofcrosssection?
- POWer:POWer<x>:MAGPROPERTY:EDGESOURce {Current|VOLTAGE}
- POWer:POWer<x>:MAGPROPERTY:EDGESOURce?
- POWer:POWer<x>:MAGPROPERTY:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:ISOURce?
- POWer:POWer<x>:MAGPROPERTY:LENgth <NR2>
- POWer:POWer<x>:MAGPROPERTY:LENgth?
- POWer:POWer<x>:MAGPROPERTY:PRIMARYTURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:PRIMARYTURNs?
- POWer:POWer<x>:MAGPROPERTY:SEC1SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC1SOURce?
- POWer:POWer<x>:MAGPROPERTY:SEC1TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC1TURNs?
- POWer:POWer<x>:MAGPROPERTY:SEC2SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC2SOURce?
- POWer:POWer<x>:MAGPROPERTY:SEC2TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC2TURNs?
- POWer:POWer<x>:MAGPROPERTY:SEC3SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC3SOURce?
- POWer:POWer<x>:MAGPROPERTY:SEC3TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC3TURNs?
- POWer:POWer<x>:MAGPROPERTY:SEC4SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC4SOURce?
- POWer:POWer<x>:MAGPROPERTY:SEC4TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC4TURNs?
- POWer:POWer<x>:MAGPROPERTY:SEC5SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC5SOURce?
- POWer:POWer<x>:MAGPROPERTY:SEC5TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC5TURNs?
- POWer:POWer<x>:MAGPROPERTY:SEC6SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC6SOURce?
- POWer:POWer<x>:MAGPROPERTY:SEC6TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC6TURNs?
- POWer:POWer<x>:MAGPROPERTY:SECPhase <NR3>
- POWer:POWer<x>:MAGPROPERTY:SECPhase?
- POWer:POWer<x>:MAGPROPERTY:SECVolt {True|False}
- POWer:POWer<x>:MAGPROPERTY:SECVolt?
- POWer:POWer<x>:MAGPROPERTY:SECWINDings {None|ONE|TWO|THREE|FOUR|FIVE|SIX}
- POWer:POWer<x>:MAGPROPERTY:SECWINDings?
- POWer:POWer<x>:MAGPROPERTY:UNITs {SI|CGS}
- POWer:POWer<x>:MAGPROPERTY:UNITs?
- POWer:POWer<x>:MAGPROPERTY:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:VSOURce?
- POWer:POWer<x>:NDUTYCYCLE:EDGEType {RISE|FALL|BOTH}
- POWer:POWer<x>:NDUTYCYCLE:EDGEType?
- POWer:POWer<x>:NDUTYCYCLE:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:NDUTYCYCLE:INPUTSOurce?
- POWer:POWer<x>:NPULSEWIDTH:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:NPULSEWIDTH:INPUTSOurce?
- POWer:POWer<x>:PDUTYCYCLE:EDGEType {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:PDUTYCYCLE:EDGEType?
- POWer:POWer<x>:PDUTYCYCLE:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:PDUTYCYCLE:INPUTSOurce?
- POWer:POWer<x>:PERIOD:EDGe {RISE|FALL}
- POWer:POWer<x>:PERIOD:EDGe?
- POWer:POWer<x>:PERIOD:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:PERIOD:INPUTSOurce?
- POWer:POWer<x>:POWERQUALITY:CCYCles {ON|OFF|1|0}
- POWer:POWer<x>:POWERQUALITY:CCYCles?
- POWer:POWer<x>:POWERQUALITY:FREFerence {VOLTage|CURRent}
- POWer:POWer<x>:POWERQUALITY:FREFerence?
- POWer:POWer<x>:POWERQUALITY:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:POWERQUALITY:ISOURce?
- POWer:POWer<x>:POWERQUALITY:STYPe {AC|DC}
- POWer:POWer<x>:POWERQUALITY:STYPe?
- POWer:POWer<x>:POWERQUALITY:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:POWERQUALITY:VSOURce?
- POWer:POWer<x>:PPULSEWIDTH:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:PPULSEWIDTH:INPUTSOurce?
- POWer:POWer<x>:PRESET {EXECute}
- POWer:POWer<x>:PSRR:AMP<x>Val <NR3>
- POWer:POWer<x>:PSRR:AMP<x>Val?
- POWer:POWer<x>:PSRR:AMPMode {CONSTant|PROFile}
- POWer:POWer<x>:PSRR:AMPMode?
- POWer:POWer<x>:PSRR:ANALYSISMethod {SV|FFT}
- POWer:POWer<x>:PSRR:ANALYSISMethod?
- POWer:POWer<x>:PSRR:AUTORbw {True|False}
- POWer:POWer<x>:PSRR:AUTORbw?
- POWer:POWer<x>:PSRR:CONNECTSTATus?
- POWer:POWer<x>:PSRR:CONSTAMPlitude <NR3>
- POWer:POWer<x>:PSRR:CONSTAMPlitude?
- POWer:POWer<x>:PSRR:FREQ<x>Val <NR3>
- POWer:POWer<x>:PSRR:FREQ<x>Val?
- POWer:POWer<x>:PSRR:GENIPADDress <QString>
- POWer:POWer<x>:PSRR:GENIPADDress?
- POWer:POWer<x>:PSRR:GENerator {INTernal|EXTernal}
- POWer:POWer<x>:PSRR:GENerator?
- POWer:POWer<x>:PSRR:IMPEDance {FIFTy|HIGHZ}
- POWer:POWer<x>:PSRR:IMPEDance?
- POWer:POWer<x>:PSRR:INPUTSOurce CH<x>
- POWer:POWer<x>:PSRR:INPUTSOurce?
- POWer:POWer<x>:PSRR:OUTPUTSOurce CH<x>
- POWer:POWer<x>:PSRR:OUTPUTSOurce?
- POWer:POWer<x>:PSRR:PPD <NR3>
- POWer:POWer<x>:PSRR:PPD?
- POWer:POWer<x>:PSRR:STARTFREQuency <NR3>
- POWer:POWer<x>:PSRR:STARTFREQuency?
- POWer:POWer<x>:PSRR:STOPFREQuency <NR3>
- POWer:POWer<x>:PSRR:STOPFREQuency?
- POWer:POWer<x>:PSRR:TESTCONNection {EXECute}
- POWer:POWer<x>:RDSON:DEVICEType {SWITCHING|PNJUNCTION}
- POWer:POWer<x>:RDSON:DEVICEType?
- POWer:POWer<x>:RDSON:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:RDSON:ISOURce?
- POWer:POWer<x>:RDSON:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:RDSON:VSOURce?
- POWer:POWer<x>:REFLevels:ABSolute:FALLHigh <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:FALLHigh?
- POWer:POWer<x>:REFLevels:ABSolute:FALLLow <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:FALLLow?
- POWer:POWer<x>:REFLevels:ABSolute:FALLMid <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:FALLMid?
- POWer:POWer<x>:REFLevels:ABSolute:HYSTeresis <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:HYSTeresis?
- POWer:POWer<x>:REFLevels:ABSolute:RISEHigh <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:RISEHigh?
- POWer:POWer<x>:REFLevels:ABSolute:RISELow <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:RISELow?
- POWer:POWer<x>:REFLevels:ABSolute:RISEMid <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:RISEMid?
- POWer:POWer<x>:REFLevels:ABSolute:TYPE {SAME|UNIQue}
- POWer:POWer<x>:REFLevels:ABSolute:TYPE?
- POWer:POWer<x>:REFLevels:BASETop {AUTO|MINMax|MEANhistogram|MODEhistogram|EYEhistogram}
- POWer:POWer<x>:REFLevels:BASETop?
- POWer:POWer<x>:REFLevels:METHod {PERCent|ABSolute}
- POWer:POWer<x>:REFLevels:METHod?
- POWer:POWer<x>:REFLevels:PERCent:FALLHigh <NR1>
- POWer:POWer<x>:REFLevels:PERCent:FALLHigh?
- POWer:POWer<x>:REFLevels:PERCent:FALLLow <NR1>
- POWer:POWer<x>:REFLevels:PERCent:FALLLow?
- POWer:POWer<x>:REFLevels:PERCent:FALLMid <NR1>
- POWer:POWer<x>:REFLevels:PERCent:FALLMid?
- POWer:POWer<x>:REFLevels:PERCent:HYSTeresis <NR1>
- POWer:POWer<x>:REFLevels:PERCent:HYSTeresis?
- POWer:POWer<x>:REFLevels:PERCent:RISEHigh <NR1>
- POWer:POWer<x>:REFLevels:PERCent:RISEHigh?
- POWer:POWer<x>:REFLevels:PERCent:RISELow <NR1>
- POWer:POWer<x>:REFLevels:PERCent:RISELow?
- POWer:POWer<x>:REFLevels:PERCent:RISEMid <NR1>
- POWer:POWer<x>:REFLevels:PERCent:RISEMid?
- POWer:POWer<x>:REFLevels:PERCent:TYPE {TENNinety|TWENtyeighty|CUSTom}
- POWer:POWer<x>:REFLevels:PERCent:TYPE?
- POWer:POWer<x>:RESUlts:ALLAcqs:MAXimum? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
- POWer:POWer<x>:RESUlts:ALLAcqs:MEAN? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
- POWer:POWer<x>:RESUlts:ALLAcqs:MINimum? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
- POWer:POWer<x>:RESUlts:ALLAcqs:PK2PK? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
- POWer:POWer<x>:RESUlts:ALLAcqs:POPUlation? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
- POWer:POWer<x>:RESUlts:ALLAcqs:STDDev? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
- POWer:POWer<x>:RESUlts:CURRentacq:F1MAG? harmonics
- POWer:POWer<x>:RESUlts:CURRentacq:F3MAG? harmonics
- POWer:POWer<x>:RESUlts:CURRentacq:FREQUENCY? harmonics
- POWer:POWer<x>:RESUlts:CURRentacq:IRMS? harmonics
- POWer:POWer<x>:RESUlts:CURRentacq:MAXimum? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
- POWer:POWer<x>:RESUlts:CURRentacq:MEAN? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
- POWer:POWer<x>:RESUlts:CURRentacq:MINimum? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
- POWer:POWer<x>:RESUlts:CURRentacq:PK2PK ? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
- POWer:POWer<x>:RESUlts:CURRentacq:PK2PK?
- POWer:POWer<x>:RESUlts:CURRentacq:POHCL? harmonics
- POWer:POWer<x>:RESUlts:CURRentacq:POHCM? harmonics
- POWer:POWer<x>:RESUlts:CURRentacq:POHCS? harmonics
- POWer:POWer<x>:RESUlts:CURRentacq:POPUlation? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
- POWer:POWer<x>:RESUlts:CURRentacq:RMS? harmonics
- POWer:POWer<x>:RESUlts:CURRentacq:STATUS? harmonics
- POWer:POWer<x>:RESUlts:CURRentacq:STDDev? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
- POWer:POWer<x>:RESUlts:CURRentacq:THDF? harmonics
- POWer:POWer<x>:RESUlts:CURRentacq:THDR? harmonics
- POWer:POWer<x>:RESUlts:CURRentacq:TRPWR? harmonics
- POWer:POWer<x>:RESUlts:CURRentacq:VRMS? harmonics
- POWer:POWer<x>:SEQSETup Execute
- POWer:POWer<x>:SEQuence {RUN|RERUN}
- POWer:POWer<x>:SEQuence?
- POWer:POWer<x>:SOA:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SOA:ISOURce?
- POWer:POWer<x>:SOA:POINT<x> <NR1>
- POWer:POWer<x>:SOA:POINT<x>?
- POWer:POWer<x>:SOA:RECAllmask
- POWer:POWer<x>:SOA:RECAllmask:FILEName
- POWer:POWer<x>:SOA:RECAllmask:FILEName?
- POWer:POWer<x>:SOA:RECAllmask?
- POWer:POWer<x>:SOA:SAVemask
- POWer:POWer<x>:SOA:SAVemask:AUTOINCrement
- POWer:POWer<x>:SOA:SAVemask:AUTOINCrement?
- POWer:POWer<x>:SOA:SAVemask:FILEName
- POWer:POWer<x>:SOA:SAVemask:FILEName?
- POWer:POWer<x>:SOA:SAVemask:FOLDer
- POWer:POWer<x>:SOA:SAVemask:FOLDer?
- POWer:POWer<x>:SOA:SAVemask?
- POWer:POWer<x>:SOA:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SOA:VSOURce?
- POWer:POWer<x>:SWITCHINGLOSS:DEVICEType {MOSFET|BJT}
- POWer:POWer<x>:SWITCHINGLOSS:DEVICEType?
- POWer:POWer<x>:SWITCHINGLOSS:GATESOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SWITCHINGLOSS:GATESOurce?
- POWer:POWer<x>:SWITCHINGLOSS:ILEVELAbs <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:ILEVELAbs?
- POWer:POWer<x>:SWITCHINGLOSS:ILEVELPct <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:ILEVELPct?
- POWer:POWer<x>:SWITCHINGLOSS:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SWITCHINGLOSS:ISOURce?
- POWer:POWer<x>:SWITCHINGLOSS:LEVELUNIts {PERCent|ABSolute}
- POWer:POWer<x>:SWITCHINGLOSS:LEVELUNIts?
- POWer:POWer<x>:SWITCHINGLOSS:RDSOn <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:RDSOn?
- POWer:POWer<x>:SWITCHINGLOSS:SWLCONFIGType {SMPS|PFC|FLYBACK}
- POWer:POWer<x>:SWITCHINGLOSS:SWLCONFIGType?
- POWer:POWer<x>:SWITCHINGLOSS:VCESat <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:VCESat?
- POWer:POWer<x>:SWITCHINGLOSS:VGLevel <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:VGLevel?
- POWer:POWer<x>:SWITCHINGLOSS:VLEVELAbs <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:VLEVELAbs?
- POWer:POWer<x>:SWITCHINGLOSS:VLEVELPct <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:VLEVELPct?
- POWer:POWer<x>:SWITCHINGLOSS:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SWITCHINGLOSS:VSOURce?
- POWer:POWer<x>:SWITCHINGRIPPLE:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SWITCHINGRIPPLE:INPUTSOurce?
- POWer:POWer<x>:SWITCHINGRIPPLE:LFREQuency <NR1>
- POWer:POWer<x>:SWITCHINGRIPPLE:LFREQuency?
- POWer:POWer<x>:TURNOFFtime:FREQuency <NR3>
- POWer:POWer<x>:TURNOFFtime:FREQuency?
- POWer:POWer<x>:TURNOFFtime:INPUTLEVel <NR3>
- POWer:POWer<x>:TURNOFFtime:INPUTLEVel?
- POWer:POWer<x>:TURNOFFtime:INPUTSOurce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:INPUTSOurce?
- POWer:POWer<x>:TURNOFFtime:MAXTIMe <NR3>
- POWer:POWer<x>:TURNOFFtime:MAXTIMe?
- POWer:POWer<x>:TURNOFFtime:MAXVoltage <NR3>
- POWer:POWer<x>:TURNOFFtime:MAXVoltage?
- POWer:POWer<x>:TURNOFFtime:NUMOUTputs {ONE|TWO|THREE|FOUR|FIVE|SIX|SEVEN}
- POWer:POWer<x>:TURNOFFtime:NUMOUTputs?
- POWer:POWer<x>:TURNOFFtime:OUTPUT1SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT1SOURce?
- POWer:POWer<x>:TURNOFFtime:OUTPUT1VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT1VOLTage?
- POWer:POWer<x>:TURNOFFtime:OUTPUT2SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT2SOURce?
- POWer:POWer<x>:TURNOFFtime:OUTPUT2VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT2VOLTage?
- POWer:POWer<x>:TURNOFFtime:OUTPUT3SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT3SOURce?
- POWer:POWer<x>:TURNOFFtime:OUTPUT3VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT3VOLTage?
- POWer:POWer<x>:TURNOFFtime:OUTPUT4SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT4SOURce?
- POWer:POWer<x>:TURNOFFtime:OUTPUT4VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT4VOLTage?
- POWer:POWer<x>:TURNOFFtime:OUTPUT5SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT5SOURce?
- POWer:POWer<x>:TURNOFFtime:OUTPUT5VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT5VOLTage?
- POWer:POWer<x>:TURNOFFtime:OUTPUT6SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT6SOURce?
- POWer:POWer<x>:TURNOFFtime:OUTPUT6VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT6VOLTage?
- POWer:POWer<x>:TURNOFFtime:OUTPUT7SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT7SOURce?
- POWer:POWer<x>:TURNOFFtime:OUTPUT7VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT7VOLTage?
- POWer:POWer<x>:TURNOFFtime:TYPE {DCDC|ACDC}
- POWer:POWer<x>:TURNOFFtime:TYPE?
- POWer:POWer<x>:TURNONtime:FREQuency <NR3>
- POWer:POWer<x>:TURNONtime:FREQuency?
- POWer:POWer<x>:TURNONtime:INPUTLEVel <NR3>
- POWer:POWer<x>:TURNONtime:INPUTLEVel?
- POWer:POWer<x>:TURNONtime:INPUTSOurce
- POWer:POWer<x>:TURNONtime:INPUTSOurce?
- POWer:POWer<x>:TURNONtime:MAXTIMe <NR3>
- POWer:POWer<x>:TURNONtime:MAXTIMe?
- POWer:POWer<x>:TURNONtime:MAXVoltage <NR3>
- POWer:POWer<x>:TURNONtime:MAXVoltage?
- POWer:POWer<x>:TURNONtime:NUMOUTputs {ONE|TWO|THREE|FOUR|FIVE|SIX|SEVEN}
- POWer:POWer<x>:TURNONtime:NUMOUTputs?
- POWer:POWer<x>:TURNONtime:OUTPUT1SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT1SOURce?
- POWer:POWer<x>:TURNONtime:OUTPUT1VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT1VOLTage?
- POWer:POWer<x>:TURNONtime:OUTPUT2SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT2SOURce?
- POWer:POWer<x>:TURNONtime:OUTPUT2VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT2VOLTage?
- POWer:POWer<x>:TURNONtime:OUTPUT3SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT3SOURce?
- POWer:POWer<x>:TURNONtime:OUTPUT3VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT3VOLTage?
- POWer:POWer<x>:TURNONtime:OUTPUT4SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT4SOURce?
- POWer:POWer<x>:TURNONtime:OUTPUT4VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT4VOLTage?
- POWer:POWer<x>:TURNONtime:OUTPUT5SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT5SOURce?
- POWer:POWer<x>:TURNONtime:OUTPUT5VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT5VOLTage?
- POWer:POWer<x>:TURNONtime:OUTPUT6SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT6SOURce?
- POWer:POWer<x>:TURNONtime:OUTPUT6VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT6VOLTage?
- POWer:POWer<x>:TURNONtime:OUTPUT7SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT7SOURce?
- POWer:POWer<x>:TURNONtime:OUTPUT7VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT7VOLTage?
- POWer:POWer<x>:TURNONtime:TYPE {DCDC|ACDC}
- POWer:POWer<x>:TURNONtime:TYPE?
- POWer:POWer<x>:TYPe P <Measurement Type>
- POWer:POWer<x>:TYPe?
- POWer:POWer<x>:WRAP:DEGrees NR3
- POWer:POWer<x>:WRAP:DEGrees?
- POWer:POWer<x>:WRAP:STATE {ON|OFF}
- POWer:POWer<x>:WRAP:STATE?
Power
¶
Bases: SCPICmdRead
The POWer command tree.
Usage
- Using the
.query()method will send thePOWer?query. - Using the
.verify(value)method will send thePOWer?query and raise an AssertionError if the returned value does not matchvalue.
Properties
.addnew: ThePOWer:ADDNewcommand..delete: ThePOWer:DELetecommand..power: ThePOWer:POWer<x>command tree.
addnew
property
¶
addnew: PowerAddnew
Return the POWer:ADDNew command.
Description
- This command adds the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.write(value)method will send thePOWer:ADDNew valuecommand.
SCPI Syntax
- POWer:ADDNew POWER<x>
delete
property
¶
delete: PowerDelete
Return the POWer:DELete command.
Description
- This command deletes the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.write(value)method will send thePOWer:DELete valuecommand.
SCPI Syntax
- POWer:DELete POWER<x>
power
property
¶
power: dict[int, PowerPowerItem]
Return the POWer:POWer<x> command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>?query. - Using the
.verify(value)method will send thePOWer:POWer<x>?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Sub-properties
.autoset: ThePOWer:POWer<x>:AUTOSetcommand..clresponse: ThePOWer:POWer<x>:CLRESPONSEcommand tree..cycleamp: ThePOWer:POWer<x>:CYCLEAmpcommand tree..cyclebase: ThePOWer:POWer<x>:CYCLEBasecommand tree..cyclemax: ThePOWer:POWer<x>:CYCLEMAXcommand tree..cyclemin: ThePOWer:POWer<x>:CYCLEMincommand tree..cyclepkpk: ThePOWer:POWer<x>:CYCLEPKPKcommand tree..cycletop: ThePOWer:POWer<x>:CYCLETopcommand tree..didt: ThePOWer:POWer<x>:DIDTcommand tree..dvdt: ThePOWer:POWer<x>:DVDTcommand tree..efficiency: ThePOWer:POWer<x>:EFFICIENCYcommand tree..frequency: ThePOWer:POWer<x>:FREQUENCYcommand tree..gating: ThePOWer:POWer<x>:GATingcommand..harmonics: ThePOWer:POWer<x>:HARMONICScommand tree..impedance: ThePOWer:POWer<x>:IMPEDANCEcommand tree..inductance: ThePOWer:POWer<x>:INDUCTANCEcommand tree..inputcap: ThePOWer:POWer<x>:INPUTCAPcommand tree..inrushcurrent: ThePOWer:POWer<x>:INRUSHcurrentcommand tree..ivsintegralv: ThePOWer:POWer<x>:IVSINTEGRALVcommand tree..label: ThePOWer:POWer<x>:LABelcommand..lineripple: ThePOWer:POWer<x>:LINERIPPLEcommand tree..magneticloss: ThePOWer:POWer<x>:MAGNETICLOSScommand tree..magproperty: ThePOWer:POWer<x>:MAGPROPERTYcommand tree..ndutycycle: ThePOWer:POWer<x>:NDUTYCYCLEcommand tree..npulsewidth: ThePOWer:POWer<x>:NPULSEWIDTHcommand tree..pdutycycle: ThePOWer:POWer<x>:PDUTYCYCLEcommand tree..period: ThePOWer:POWer<x>:PERIODcommand tree..powerquality: ThePOWer:POWer<x>:POWERQUALITYcommand tree..ppulsewidth: ThePOWer:POWer<x>:PPULSEWIDTHcommand tree..preset: ThePOWer:POWer<x>:PRESETcommand..psrr: ThePOWer:POWer<x>:PSRRcommand tree..rdson: ThePOWer:POWer<x>:RDSONcommand tree..reflevels: ThePOWer:POWer<x>:REFLevelscommand tree..results: ThePOWer:POWer<x>:RESUltscommand tree..seqsetup: ThePOWer:POWer<x>:SEQSETupcommand..sequence: ThePOWer:POWer<x>:SEQuencecommand..soa: ThePOWer:POWer<x>:SOAcommand tree..switchingloss: ThePOWer:POWer<x>:SWITCHINGLOSScommand tree..switchingripple: ThePOWer:POWer<x>:SWITCHINGRIPPLEcommand tree..turnofftime: ThePOWer:POWer<x>:TURNOFFtimecommand tree..turnontime: ThePOWer:POWer<x>:TURNONtimecommand tree..type: ThePOWer:POWer<x>:TYPecommand..wrap: ThePOWer:POWer<x>:WRAPcommand tree.
PowerAddnew
¶
Bases: SCPICmdWrite
The POWer:ADDNew command.
Description
- This command adds the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.write(value)method will send thePOWer:ADDNew valuecommand.
SCPI Syntax
- POWer:ADDNew POWER<x>
PowerDelete
¶
Bases: SCPICmdWrite
The POWer:DELete command.
Description
- This command deletes the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.write(value)method will send thePOWer:DELete valuecommand.
SCPI Syntax
- POWer:DELete POWER<x>
PowerPowerItem
¶
Bases: ValidatedDynamicNumberCmd, SCPICmdRead
The POWer:POWer<x> command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>?query. - Using the
.verify(value)method will send thePOWer:POWer<x>?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Properties
.autoset: ThePOWer:POWer<x>:AUTOSetcommand..clresponse: ThePOWer:POWer<x>:CLRESPONSEcommand tree..cycleamp: ThePOWer:POWer<x>:CYCLEAmpcommand tree..cyclebase: ThePOWer:POWer<x>:CYCLEBasecommand tree..cyclemax: ThePOWer:POWer<x>:CYCLEMAXcommand tree..cyclemin: ThePOWer:POWer<x>:CYCLEMincommand tree..cyclepkpk: ThePOWer:POWer<x>:CYCLEPKPKcommand tree..cycletop: ThePOWer:POWer<x>:CYCLETopcommand tree..didt: ThePOWer:POWer<x>:DIDTcommand tree..dvdt: ThePOWer:POWer<x>:DVDTcommand tree..efficiency: ThePOWer:POWer<x>:EFFICIENCYcommand tree..frequency: ThePOWer:POWer<x>:FREQUENCYcommand tree..gating: ThePOWer:POWer<x>:GATingcommand..harmonics: ThePOWer:POWer<x>:HARMONICScommand tree..impedance: ThePOWer:POWer<x>:IMPEDANCEcommand tree..inductance: ThePOWer:POWer<x>:INDUCTANCEcommand tree..inputcap: ThePOWer:POWer<x>:INPUTCAPcommand tree..inrushcurrent: ThePOWer:POWer<x>:INRUSHcurrentcommand tree..ivsintegralv: ThePOWer:POWer<x>:IVSINTEGRALVcommand tree..label: ThePOWer:POWer<x>:LABelcommand..lineripple: ThePOWer:POWer<x>:LINERIPPLEcommand tree..magneticloss: ThePOWer:POWer<x>:MAGNETICLOSScommand tree..magproperty: ThePOWer:POWer<x>:MAGPROPERTYcommand tree..ndutycycle: ThePOWer:POWer<x>:NDUTYCYCLEcommand tree..npulsewidth: ThePOWer:POWer<x>:NPULSEWIDTHcommand tree..pdutycycle: ThePOWer:POWer<x>:PDUTYCYCLEcommand tree..period: ThePOWer:POWer<x>:PERIODcommand tree..powerquality: ThePOWer:POWer<x>:POWERQUALITYcommand tree..ppulsewidth: ThePOWer:POWer<x>:PPULSEWIDTHcommand tree..preset: ThePOWer:POWer<x>:PRESETcommand..psrr: ThePOWer:POWer<x>:PSRRcommand tree..rdson: ThePOWer:POWer<x>:RDSONcommand tree..reflevels: ThePOWer:POWer<x>:REFLevelscommand tree..results: ThePOWer:POWer<x>:RESUltscommand tree..seqsetup: ThePOWer:POWer<x>:SEQSETupcommand..sequence: ThePOWer:POWer<x>:SEQuencecommand..soa: ThePOWer:POWer<x>:SOAcommand tree..switchingloss: ThePOWer:POWer<x>:SWITCHINGLOSScommand tree..switchingripple: ThePOWer:POWer<x>:SWITCHINGRIPPLEcommand tree..turnofftime: ThePOWer:POWer<x>:TURNOFFtimecommand tree..turnontime: ThePOWer:POWer<x>:TURNONtimecommand tree..type: ThePOWer:POWer<x>:TYPecommand..wrap: ThePOWer:POWer<x>:WRAPcommand tree.
autoset
property
¶
autoset: PowerPowerItemAutoset
Return the POWer:POWer<x>:AUTOSet command.
Description
- This command executes power autoset for the specified power measurement number.
Usage
- Using the
.write(value)method will send thePOWer:POWer<x>:AUTOSet valuecommand.
SCPI Syntax
- POWer:POWer<x>:AUTOSet EXECute
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
clresponse
property
¶
clresponse: PowerPowerItemClresponse
Return the POWer:POWer<x>:CLRESPONSE command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>sets the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Sub-properties
.ampval: ThePOWer:POWer<x>:CLRESPONSE:AMP<x>Valcommand..ampmode: ThePOWer:POWer<x>:CLRESPONSE:AMPModecommand..analysismethod: ThePOWer:POWer<x>:CLRESPONSE:ANALYSISMethodcommand..autorbw: ThePOWer:POWer<x>:CLRESPONSE:AUTORbwcommand..connectstatus: ThePOWer:POWer<x>:CLRESPONSE:CONNECTSTATuscommand..constamplitude: ThePOWer:POWer<x>:CLRESPONSE:CONSTAMPlitudecommand..freqval: ThePOWer:POWer<x>:CLRESPONSE:FREQ<x>Valcommand..genipaddress: ThePOWer:POWer<x>:CLRESPONSE:GENIPADDresscommand..generator: ThePOWer:POWer<x>:CLRESPONSE:GENeratorcommand..impedance: ThePOWer:POWer<x>:CLRESPONSE:IMPEDancecommand..inputsource: ThePOWer:POWer<x>:CLRESPONSE:INPUTSOurcecommand..mon: ThePOWer:POWer<x>:CLRESPONSE:MONcommand..outputsource: ThePOWer:POWer<x>:CLRESPONSE:OUTPUTSOurcecommand..ppd: ThePOWer:POWer<x>:CLRESPONSE:PPDcommand..startfrequency: ThePOWer:POWer<x>:CLRESPONSE:STARTFREQuencycommand..stopfrequency: ThePOWer:POWer<x>:CLRESPONSE:STOPFREQuencycommand..testconnection: ThePOWer:POWer<x>:CLRESPONSE:TESTCONNectioncommand.
cycleamp
property
¶
cycleamp: PowerPowerItemCycleamp
Return the POWer:POWer<x>:CYCLEAmp command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEAmp?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEAmp?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Sub-properties
.inputsource: ThePOWer:POWer<x>:CYCLEAmp:INPUTSOurcecommand.
cyclebase
property
¶
cyclebase: PowerPowerItemCyclebase
Return the POWer:POWer<x>:CYCLEBase command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEBase?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEBase?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Sub-properties
.inputsource: ThePOWer:POWer<x>:CYCLEBase:INPUTSOurcecommand.
cyclemax
property
¶
cyclemax: PowerPowerItemCyclemax
Return the POWer:POWer<x>:CYCLEMAX command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEMAX?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEMAX?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Sub-properties
.inputsource: ThePOWer:POWer<x>:CYCLEMAX:INPUTSOurcecommand.
cyclemin
property
¶
cyclemin: PowerPowerItemCyclemin
Return the POWer:POWer<x>:CYCLEMin command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEMin?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEMin?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Sub-properties
.inputsource: ThePOWer:POWer<x>:CYCLEMin:INPUTSOurcecommand.
cyclepkpk
property
¶
cyclepkpk: PowerPowerItemCyclepkpk
Return the POWer:POWer<x>:CYCLEPKPK command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEPKPK?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEPKPK?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Sub-properties
.inputsource: ThePOWer:POWer<x>:CYCLEPKPK:INPUTSOurcecommand.
cycletop
property
¶
cycletop: PowerPowerItemCycletop
Return the POWer:POWer<x>:CYCLETop command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLETop?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLETop?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Sub-properties
.inputsource: ThePOWer:POWer<x>:CYCLETop:INPUTSOurcecommand.
didt
property
¶
didt: PowerPowerItemDidt
Return the POWer:POWer<x>:DIDT command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:DIDT?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:DIDT?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Sub-properties
.inputsource: ThePOWer:POWer<x>:DIDT:INPUTSOurcecommand..sourceedgetype: ThePOWer:POWer<x>:DIDT:SOURCEEDGETypecommand.
dvdt
property
¶
dvdt: PowerPowerItemDvdt
Return the POWer:POWer<x>:DVDT command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:DVDT?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:DVDT?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Sub-properties
.inputsource: ThePOWer:POWer<x>:DVDT:INPUTSOurcecommand..sourceedgetype: ThePOWer:POWer<x>:DVDT:SOURCEEDGETypecommand.
efficiency
property
¶
efficiency: PowerPowerItemEfficiency
Return the POWer:POWer<x>:EFFICIENCY command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown on the power measurement badge of the UI.
Sub-properties
.inputtype: ThePOWer:POWer<x>:EFFICIENCY:INPUTTypecommand..iout1source: ThePOWer:POWer<x>:EFFICIENCY:IOUT1SOUrcecommand..iout2source: ThePOWer:POWer<x>:EFFICIENCY:IOUT2SOUrcecommand..iout3source: ThePOWer:POWer<x>:EFFICIENCY:IOUT3SOUrcecommand..isource: ThePOWer:POWer<x>:EFFICIENCY:ISOUrcecommand..numofoutputs: ThePOWer:POWer<x>:EFFICIENCY:NUMOFOutputscommand..output1type: ThePOWer:POWer<x>:EFFICIENCY:OUTPUT1Typecommand..output2type: ThePOWer:POWer<x>:EFFICIENCY:OUTPUT2Typecommand..output3type: ThePOWer:POWer<x>:EFFICIENCY:OUTPUT3Typecommand..outputtype: ThePOWer:POWer<x>:EFFICIENCY:OUTPUTTypecommand..vout1source: ThePOWer:POWer<x>:EFFICIENCY:VOUT1SOUrcecommand..vout2source: ThePOWer:POWer<x>:EFFICIENCY:VOUT2SOUrcecommand..vout3source: ThePOWer:POWer<x>:EFFICIENCY:VOUT3SOUrcecommand..vsource: ThePOWer:POWer<x>:EFFICIENCY:VSOUrcecommand.
frequency
property
¶
frequency: PowerPowerItemFrequency
Return the POWer:POWer<x>:FREQUENCY command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:FREQUENCY?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:FREQUENCY?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Sub-properties
.edge: ThePOWer:POWer<x>:FREQUENCY:EDGecommand..inputsource: ThePOWer:POWer<x>:FREQUENCY:INPUTSOurcecommand.
gating
property
¶
gating: PowerPowerItemGating
Return the POWer:POWer<x>:GATing command.
Description
- This command sets or queries the gating type for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:GATing?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:GATing?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:GATing valuecommand.
SCPI Syntax
- POWer:POWer<x>:GATing {NONE|CURSOR|SCREEN|LOGIC}
- POWer:POWer<x>:GATing?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.NONEmakes measurement across the entire waveform record.CURSORmakes measurements on that portion of the waveform between the cursors. Selecting Cursors opens cursors on the measurement source. Set the cursors so that the waveform area of interest is in between the cursors.SCREENtakes measurements on that portion of the waveform shown in the display. When Zoom is on, the display is the zoom window.LOGICtakes measurements only when the logical state of a specified waveform is true.
Sub-properties
.global: ThePOWer:POWer<x>:GATing:GLOBalcommand.
harmonics
property
¶
harmonics: PowerPowerItemHarmonics
Return the POWer:POWer<x>:HARMONICS command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS?query and raise an AssertionError if the returned value does not matchvalue.
Sub-properties
.class: ThePOWer:POWer<x>:HARMONICS:CLASscommand..clfile: ThePOWer:POWer<x>:HARMONICS:CLFilecommand..cmethod: ThePOWer:POWer<x>:HARMONICS:CMEThodcommand..fundcurrent: ThePOWer:POWer<x>:HARMONICS:FUNDCURRentcommand..horder: ThePOWer:POWer<x>:HARMONICS:HORDercommand..hsource: ThePOWer:POWer<x>:HARMONICS:HSOURcecommand..ipower: ThePOWer:POWer<x>:HARMONICS:IPOWercommand..isource: ThePOWer:POWer<x>:HARMONICS:ISOURcecommand..linefrequency: ThePOWer:POWer<x>:HARMONICS:LINEFREQUEncycommand..oddeven: ThePOWer:POWer<x>:HARMONICS:ODDEVencommand..pfactor: ThePOWer:POWer<x>:HARMONICS:PFACtorcommand..powerrating: ThePOWer:POWer<x>:HARMONICS:POWERRatingcommand..rcurrent: ThePOWer:POWer<x>:HARMONICS:RCURRentcommand..standard: ThePOWer:POWer<x>:HARMONICS:STANDardcommand..startfrequency: ThePOWer:POWer<x>:HARMONICS:STARTFREQUEncycommand..units: ThePOWer:POWer<x>:HARMONICS:UNITscommand..vsource: ThePOWer:POWer<x>:HARMONICS:VSOURcecommand.
impedance
property
¶
impedance: PowerPowerItemImpedance
Return the POWer:POWer<x>:IMPEDANCE command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the number of the Impedance power measurement.
Sub-properties
.ampval: ThePOWer:POWer<x>:IMPEDANCE:AMP<x>Valcommand..ampmode: ThePOWer:POWer<x>:IMPEDANCE:AMPModecommand..analysismethod: ThePOWer:POWer<x>:IMPEDANCE:ANALYSISMethodcommand..autorbw: ThePOWer:POWer<x>:IMPEDANCE:AUTORbwcommand..connectstatus: ThePOWer:POWer<x>:IMPEDANCE:CONNECTSTATuscommand..constamplitude: ThePOWer:POWer<x>:IMPEDANCE:CONSTAMPlitudecommand..freqval: ThePOWer:POWer<x>:IMPEDANCE:FREQ<x>Valcommand..genipaddress: ThePOWer:POWer<x>:IMPEDANCE:GENIPADDresscommand..generator: ThePOWer:POWer<x>:IMPEDANCE:GENeratorcommand..impedance: ThePOWer:POWer<x>:IMPEDANCE:IMPEDANCEcommand..inputsource: ThePOWer:POWer<x>:IMPEDANCE:INPUTSOurcecommand..outputsource: ThePOWer:POWer<x>:IMPEDANCE:OUTPUTSOurcecommand..ppd: ThePOWer:POWer<x>:IMPEDANCE:PPDcommand..startfrequency: ThePOWer:POWer<x>:IMPEDANCE:STARTFREQuencycommand..stopfrequency: ThePOWer:POWer<x>:IMPEDANCE:STOPFREQuencycommand..testconnection: ThePOWer:POWer<x>:IMPEDANCE:TESTCONNectioncommand.
inductance
property
¶
inductance: PowerPowerItemInductance
Return the POWer:POWer<x>:INDUCTANCE command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INDUCTANCE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INDUCTANCE?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Sub-properties
.edgesource: ThePOWer:POWer<x>:INDUCTANCE:EDGESourcecommand..isource: ThePOWer:POWer<x>:INDUCTANCE:ISOUrcecommand..vsource: ThePOWer:POWer<x>:INDUCTANCE:VSOURcecommand.
inputcap
property
¶
inputcap: PowerPowerItemInputcap
Return the POWer:POWer<x>:INPUTCAP command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INPUTCAP?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INPUTCAP?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Sub-properties
.isource: ThePOWer:POWer<x>:INPUTCAP:ISOURcecommand..peakcurrent: ThePOWer:POWer<x>:INPUTCAP:PEAKCURRentcommand..peakvoltage: ThePOWer:POWer<x>:INPUTCAP:PEAKVOLTagecommand..vsource: ThePOWer:POWer<x>:INPUTCAP:VSOURcecommand.
inrushcurrent
property
¶
inrushcurrent: PowerPowerItemInrushcurrent
Return the POWer:POWer<x>:INRUSHcurrent command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INRUSHcurrent?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INRUSHcurrent?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Sub-properties
.inputsource: ThePOWer:POWer<x>:INRUSHcurrent:INPUTSOurcecommand..peakcurrent: ThePOWer:POWer<x>:INRUSHcurrent:PEAKCURRentcommand.
ivsintegralv
property
¶
ivsintegralv: PowerPowerItemIvsintegralv
Return the POWer:POWer<x>:IVSINTEGRALV command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IVSINTEGRALV?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IVSINTEGRALV?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Sub-properties
.isource: ThePOWer:POWer<x>:IVSINTEGRALV:ISOURcecommand..vsource: ThePOWer:POWer<x>:IVSINTEGRALV:VSOURcecommand.
label
property
¶
label: PowerPowerItemLabel
Return the POWer:POWer<x>:LABel command.
Description
- This command sets or queries the label for the specified power measurement. As the label can contain non 7-bit ASCII text, it is stored in Percent Encoding format. The power measurement badge is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:LABel?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:LABel?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:LABel valuecommand.
SCPI Syntax
- POWer:POWer<x>:LABel <QString>
- POWer:POWer<x>:LABel?
lineripple
property
¶
lineripple: PowerPowerItemLineripple
Return the POWer:POWer<x>:LINERIPPLE command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:LINERIPPLE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:LINERIPPLE?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Sub-properties
.inputsource: ThePOWer:POWer<x>:LINERIPPLE:INPUTSOurcecommand..lfrequency: ThePOWer:POWer<x>:LINERIPPLE:LFREQuencycommand.
magneticloss
property
¶
magneticloss: PowerPowerItemMagneticloss
Return the POWer:POWer<x>:MAGNETICLOSS command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGNETICLOSS?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGNETICLOSS?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Sub-properties
.isource: ThePOWer:POWer<x>:MAGNETICLOSS:ISOURcecommand..vsource: ThePOWer:POWer<x>:MAGNETICLOSS:VSOURcecommand.
magproperty
property
¶
magproperty: PowerPowerItemMagproperty
Return the POWer:POWer<x>:MAGPROPERTY command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the magnetic property power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Sub-properties
.areaofcrosssection: ThePOWer:POWer<x>:MAGPROPERTY:AREAofcrosssectioncommand..edgesource: ThePOWer:POWer<x>:MAGPROPERTY:EDGESOURcecommand..isource: ThePOWer:POWer<x>:MAGPROPERTY:ISOURcecommand..length: ThePOWer:POWer<x>:MAGPROPERTY:LENgthcommand..primaryturns: ThePOWer:POWer<x>:MAGPROPERTY:PRIMARYTURNscommand..sec1source: ThePOWer:POWer<x>:MAGPROPERTY:SEC1SOURcecommand..sec1turns: ThePOWer:POWer<x>:MAGPROPERTY:SEC1TURNscommand..sec2source: ThePOWer:POWer<x>:MAGPROPERTY:SEC2SOURcecommand..sec2turns: ThePOWer:POWer<x>:MAGPROPERTY:SEC2TURNscommand..sec3source: ThePOWer:POWer<x>:MAGPROPERTY:SEC3SOURcecommand..sec3turns: ThePOWer:POWer<x>:MAGPROPERTY:SEC3TURNscommand..sec4source: ThePOWer:POWer<x>:MAGPROPERTY:SEC4SOURcecommand..sec4turns: ThePOWer:POWer<x>:MAGPROPERTY:SEC4TURNscommand..sec5source: ThePOWer:POWer<x>:MAGPROPERTY:SEC5SOURcecommand..sec5turns: ThePOWer:POWer<x>:MAGPROPERTY:SEC5TURNscommand..sec6source: ThePOWer:POWer<x>:MAGPROPERTY:SEC6SOURcecommand..sec6turns: ThePOWer:POWer<x>:MAGPROPERTY:SEC6TURNscommand..secphase: ThePOWer:POWer<x>:MAGPROPERTY:SECPhasecommand..secvolt: ThePOWer:POWer<x>:MAGPROPERTY:SECVoltcommand..secwindings: ThePOWer:POWer<x>:MAGPROPERTY:SECWINDingscommand..units: ThePOWer:POWer<x>:MAGPROPERTY:UNITscommand..vsource: ThePOWer:POWer<x>:MAGPROPERTY:VSOURcecommand.
ndutycycle
property
¶
ndutycycle: PowerPowerItemNdutycycle
Return the POWer:POWer<x>:NDUTYCYCLE command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:NDUTYCYCLE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:NDUTYCYCLE?query and raise an AssertionError if the returned value does not matchvalue.
Sub-properties
.edgetype: ThePOWer:POWer<x>:NDUTYCYCLE:EDGETypecommand..inputsource: ThePOWer:POWer<x>:NDUTYCYCLE:INPUTSOurcecommand.
npulsewidth
property
¶
npulsewidth: PowerPowerItemNpulsewidth
Return the POWer:POWer<x>:NPULSEWIDTH command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:NPULSEWIDTH?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:NPULSEWIDTH?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Sub-properties
.inputsource: ThePOWer:POWer<x>:NPULSEWIDTH:INPUTSOurcecommand.
pdutycycle
property
¶
pdutycycle: PowerPowerItemPdutycycle
Return the POWer:POWer<x>:PDUTYCYCLE command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PDUTYCYCLE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PDUTYCYCLE?query and raise an AssertionError if the returned value does not matchvalue.
Sub-properties
.edgetype: ThePOWer:POWer<x>:PDUTYCYCLE:EDGETypecommand..inputsource: ThePOWer:POWer<x>:PDUTYCYCLE:INPUTSOurcecommand.
period
property
¶
period: PowerPowerItemPeriod
Return the POWer:POWer<x>:PERIOD command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PERIOD?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PERIOD?query and raise an AssertionError if the returned value does not matchvalue.
Sub-properties
.edge: ThePOWer:POWer<x>:PERIOD:EDGecommand..inputsource: ThePOWer:POWer<x>:PERIOD:INPUTSOurcecommand.
powerquality
property
¶
powerquality: PowerPowerItemPowerquality
Return the POWer:POWer<x>:POWERQUALITY command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:POWERQUALITY?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:POWERQUALITY?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Sub-properties
.ccycles: ThePOWer:POWer<x>:POWERQUALITY:CCYClescommand..freference: ThePOWer:POWer<x>:POWERQUALITY:FREFerencecommand..isource: ThePOWer:POWer<x>:POWERQUALITY:ISOURcecommand..stype: ThePOWer:POWer<x>:POWERQUALITY:STYPecommand..vsource: ThePOWer:POWer<x>:POWERQUALITY:VSOURcecommand.
ppulsewidth
property
¶
ppulsewidth: PowerPowerItemPpulsewidth
Return the POWer:POWer<x>:PPULSEWIDTH command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PPULSEWIDTH?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PPULSEWIDTH?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Sub-properties
.inputsource: ThePOWer:POWer<x>:PPULSEWIDTH:INPUTSOurcecommand.
preset
property
¶
preset: PowerPowerItemPreset
Return the POWer:POWer<x>:PRESET command.
Description
- This command runs a power preset action for the specified power measurement number.
Usage
- Using the
.write(value)method will send thePOWer:POWer<x>:PRESET valuecommand.
SCPI Syntax
- POWer:POWer<x>:PRESET {EXECute}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.EXECuteruns the power preset action.
psrr
property
¶
psrr: PowerPowerItemPsrr
Return the POWer:POWer<x>:PSRR command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Sub-properties
.ampval: ThePOWer:POWer<x>:PSRR:AMP<x>Valcommand..ampmode: ThePOWer:POWer<x>:PSRR:AMPModecommand..analysismethod: ThePOWer:POWer<x>:PSRR:ANALYSISMethodcommand..autorbw: ThePOWer:POWer<x>:PSRR:AUTORbwcommand..connectstatus: ThePOWer:POWer<x>:PSRR:CONNECTSTATuscommand..constamplitude: ThePOWer:POWer<x>:PSRR:CONSTAMPlitudecommand..freqval: ThePOWer:POWer<x>:PSRR:FREQ<x>Valcommand..genipaddress: ThePOWer:POWer<x>:PSRR:GENIPADDresscommand..generator: ThePOWer:POWer<x>:PSRR:GENeratorcommand..impedance: ThePOWer:POWer<x>:PSRR:IMPEDancecommand..inputsource: ThePOWer:POWer<x>:PSRR:INPUTSOurcecommand..outputsource: ThePOWer:POWer<x>:PSRR:OUTPUTSOurcecommand..ppd: ThePOWer:POWer<x>:PSRR:PPDcommand..startfrequency: ThePOWer:POWer<x>:PSRR:STARTFREQuencycommand..stopfrequency: ThePOWer:POWer<x>:PSRR:STOPFREQuencycommand..testconnection: ThePOWer:POWer<x>:PSRR:TESTCONNectioncommand.
rdson
property
¶
rdson: PowerPowerItemRdson
Return the POWer:POWer<x>:RDSON command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:RDSON?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:RDSON?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Sub-properties
.devicetype: ThePOWer:POWer<x>:RDSON:DEVICETypecommand..isource: ThePOWer:POWer<x>:RDSON:ISOURcecommand..vsource: ThePOWer:POWer<x>:RDSON:VSOURcecommand.
reflevels
property
¶
reflevels: PowerPowerItemReflevels
Return the POWer:POWer<x>:REFLevels command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Sub-properties
.absolute: ThePOWer:POWer<x>:REFLevels:ABSolutecommand tree..basetop: ThePOWer:POWer<x>:REFLevels:BASETopcommand..method: ThePOWer:POWer<x>:REFLevels:METHodcommand..percent: ThePOWer:POWer<x>:REFLevels:PERCentcommand tree.
results
property
¶
results: PowerPowerItemResults
Return the POWer:POWer<x>:RESUlts command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:RESUlts?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:RESUlts?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Sub-properties
.allacqs: ThePOWer:POWer<x>:RESUlts:ALLAcqscommand tree..currentacq: ThePOWer:POWer<x>:RESUlts:CURRentacqcommand tree.
seqsetup
property
¶
seqsetup: PowerPowerItemSeqsetup
Return the POWer:POWer<x>:SEQSETup command.
Description
- This command sets up the instrument’s horizontal, vertical, and trigger parameters to optimize for taking the specified power measurement.
Usage
- Using the
.write(value)method will send thePOWer:POWer<x>:SEQSETup valuecommand.
SCPI Syntax
- POWer:POWer<x>:SEQSETup Execute
Info
POWer<x>is the Power measurement identifier number. The number must be for a power measurement that requires a single sequence acquisition.Executesets the measurement to run an acquisition and acquire data for the specified single sequence power measurement.
sequence
property
¶
sequence: PowerPowerItemSequence
Return the POWer:POWer<x>:SEQuence command.
Description
- This command sets or queries the run state of a single sequence power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SEQuence?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SEQuence?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SEQuence valuecommand.
SCPI Syntax
- POWer:POWer<x>:SEQuence {RUN|RERUN}
- POWer:POWer<x>:SEQuence?
Info
POWer<x>is the Power measurement identifier number. The number must be for a power measurement that requires a single sequence acquisition.RUNsets the measurement to run an acquisition and acquire data for the specified single sequence power measurement.RERUNsets the measurement to rerun an acquisition and acquire data for the specified single sequence power measurement.
soa
property
¶
soa: PowerPowerItemSoa
Return the POWer:POWer<x>:SOA command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Sub-properties
.isource: ThePOWer:POWer<x>:SOA:ISOURcecommand..point: ThePOWer:POWer<x>:SOA:POINT<x>command..recallmask: ThePOWer:POWer<x>:SOA:RECAllmaskcommand..savemask: ThePOWer:POWer<x>:SOA:SAVemaskcommand..vsource: ThePOWer:POWer<x>:SOA:VSOURcecommand.
switchingloss
property
¶
switchingloss: PowerPowerItemSwitchingloss
Return the POWer:POWer<x>:SWITCHINGLOSS command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS?query and raise an AssertionError if the returned value does not matchvalue.
Sub-properties
.devicetype: ThePOWer:POWer<x>:SWITCHINGLOSS:DEVICETypecommand..gatesource: ThePOWer:POWer<x>:SWITCHINGLOSS:GATESOurcecommand..ilevelabs: ThePOWer:POWer<x>:SWITCHINGLOSS:ILEVELAbscommand..ilevelpct: ThePOWer:POWer<x>:SWITCHINGLOSS:ILEVELPctcommand..isource: ThePOWer:POWer<x>:SWITCHINGLOSS:ISOURcecommand..levelunits: ThePOWer:POWer<x>:SWITCHINGLOSS:LEVELUNItscommand..rdson: ThePOWer:POWer<x>:SWITCHINGLOSS:RDSOncommand..swlconfigtype: ThePOWer:POWer<x>:SWITCHINGLOSS:SWLCONFIGTypecommand..vcesat: ThePOWer:POWer<x>:SWITCHINGLOSS:VCESatcommand..vglevel: ThePOWer:POWer<x>:SWITCHINGLOSS:VGLevelcommand..vlevelabs: ThePOWer:POWer<x>:SWITCHINGLOSS:VLEVELAbscommand..vlevelpct: ThePOWer:POWer<x>:SWITCHINGLOSS:VLEVELPctcommand..vsource: ThePOWer:POWer<x>:SWITCHINGLOSS:VSOURcecommand.
switchingripple
property
¶
switchingripple: PowerPowerItemSwitchingripple
Return the POWer:POWer<x>:SWITCHINGRIPPLE command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGRIPPLE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGRIPPLE?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Sub-properties
.inputsource: ThePOWer:POWer<x>:SWITCHINGRIPPLE:INPUTSOurcecommand..lfrequency: ThePOWer:POWer<x>:SWITCHINGRIPPLE:LFREQuencycommand.
turnofftime
property
¶
turnofftime: PowerPowerItemTurnofftime
Return the POWer:POWer<x>:TURNOFFtime command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Sub-properties
.frequency: ThePOWer:POWer<x>:TURNOFFtime:FREQuencycommand..inputlevel: ThePOWer:POWer<x>:TURNOFFtime:INPUTLEVelcommand..inputsource: ThePOWer:POWer<x>:TURNOFFtime:INPUTSOurcecommand..maxtime: ThePOWer:POWer<x>:TURNOFFtime:MAXTIMecommand..maxvoltage: ThePOWer:POWer<x>:TURNOFFtime:MAXVoltagecommand..numoutputs: ThePOWer:POWer<x>:TURNOFFtime:NUMOUTputscommand..output1source: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT1SOURcecommand..output1voltage: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT1VOLTagecommand..output2source: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT2SOURcecommand..output2voltage: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT2VOLTagecommand..output3source: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT3SOURcecommand..output3voltage: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT3VOLTagecommand..output4source: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT4SOURcecommand..output4voltage: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT4VOLTagecommand..output5source: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT5SOURcecommand..output5voltage: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT5VOLTagecommand..output6source: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT6SOURcecommand..output6voltage: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT6VOLTagecommand..output7source: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT7SOURcecommand..output7voltage: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT7VOLTagecommand..type: ThePOWer:POWer<x>:TURNOFFtime:TYPEcommand.
turnontime
property
¶
turnontime: PowerPowerItemTurnontime
Return the POWer:POWer<x>:TURNONtime command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Sub-properties
.frequency: ThePOWer:POWer<x>:TURNONtime:FREQuencycommand..inputlevel: ThePOWer:POWer<x>:TURNONtime:INPUTLEVelcommand..inputsource: ThePOWer:POWer<x>:TURNONtime:INPUTSOurcecommand..maxtime: ThePOWer:POWer<x>:TURNONtime:MAXTIMecommand..maxvoltage: ThePOWer:POWer<x>:TURNONtime:MAXVoltagecommand..numoutputs: ThePOWer:POWer<x>:TURNONtime:NUMOUTputscommand..output1source: ThePOWer:POWer<x>:TURNONtime:OUTPUT1SOURcecommand..output1voltage: ThePOWer:POWer<x>:TURNONtime:OUTPUT1VOLTagecommand..output2source: ThePOWer:POWer<x>:TURNONtime:OUTPUT2SOURcecommand..output2voltage: ThePOWer:POWer<x>:TURNONtime:OUTPUT2VOLTagecommand..output3source: ThePOWer:POWer<x>:TURNONtime:OUTPUT3SOURcecommand..output3voltage: ThePOWer:POWer<x>:TURNONtime:OUTPUT3VOLTagecommand..output4source: ThePOWer:POWer<x>:TURNONtime:OUTPUT4SOURcecommand..output4voltage: ThePOWer:POWer<x>:TURNONtime:OUTPUT4VOLTagecommand..output5source: ThePOWer:POWer<x>:TURNONtime:OUTPUT5SOURcecommand..output5voltage: ThePOWer:POWer<x>:TURNONtime:OUTPUT5VOLTagecommand..output6source: ThePOWer:POWer<x>:TURNONtime:OUTPUT6SOURcecommand..output6voltage: ThePOWer:POWer<x>:TURNONtime:OUTPUT6VOLTagecommand..output7source: ThePOWer:POWer<x>:TURNONtime:OUTPUT7SOURcecommand..output7voltage: ThePOWer:POWer<x>:TURNONtime:OUTPUT7VOLTagecommand..type: ThePOWer:POWer<x>:TURNONtime:TYPEcommand.
type
property
¶
type: PowerPowerItemType
Return the POWer:POWer<x>:TYPe command.
Description
- This command sets or queries the measurement type of the specified power measurement number. If the measurement number does not exist, this command creates a new power measurement, assigns the specified measurement number to the new measurement, and then assigns the measurement type to the new measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TYPe?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TYPe?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TYPe valuecommand.
SCPI Syntax
- POWer:POWer<x>:TYPe P <Measurement Type>
- POWer:POWer<x>:TYPe?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<Measurement Type> = CYCLEAmp | CYCLEBase | CYCLEMAx |CYCLEMIn | CYCLEPKpk | CYCLETop | DIDT | DVDT | EFFICIENCY| FREQuency | HARMonics |IMPEDANCE| INDUCTANCE | INPUTCAP| INRUSHcurrent| | IVSINTEGRALV | LINERIpple | MAGNETICLOSS |MAGPROPERTY | NDUTYCycle | NPULSEWidth | PDUTYCycle | PERIod| POWERQUALity | PPULSEWidth | RDSON | SOA | SWITCHINGLOss |SWITCHINGRIpple | TURNOFFtime | TURNONtime| CLRESPONSE | PSRR
wrap
property
¶
wrap: PowerPowerItemWrap
Return the POWer:POWer<x>:WRAP command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:WRAP?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:WRAP?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Sub-properties
.degrees: ThePOWer:POWer<x>:WRAP:DEGreescommand..state: ThePOWer:POWer<x>:WRAP:STATEcommand.
PowerPowerItemAutoset
¶
Bases: SCPICmdWrite
The POWer:POWer<x>:AUTOSet command.
Description
- This command executes power autoset for the specified power measurement number.
Usage
- Using the
.write(value)method will send thePOWer:POWer<x>:AUTOSet valuecommand.
SCPI Syntax
- POWer:POWer<x>:AUTOSet EXECute
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
PowerPowerItemClresponse
¶
Bases: SCPICmdRead
The POWer:POWer<x>:CLRESPONSE command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>sets the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Properties
.ampval: ThePOWer:POWer<x>:CLRESPONSE:AMP<x>Valcommand..ampmode: ThePOWer:POWer<x>:CLRESPONSE:AMPModecommand..analysismethod: ThePOWer:POWer<x>:CLRESPONSE:ANALYSISMethodcommand..autorbw: ThePOWer:POWer<x>:CLRESPONSE:AUTORbwcommand..connectstatus: ThePOWer:POWer<x>:CLRESPONSE:CONNECTSTATuscommand..constamplitude: ThePOWer:POWer<x>:CLRESPONSE:CONSTAMPlitudecommand..freqval: ThePOWer:POWer<x>:CLRESPONSE:FREQ<x>Valcommand..genipaddress: ThePOWer:POWer<x>:CLRESPONSE:GENIPADDresscommand..generator: ThePOWer:POWer<x>:CLRESPONSE:GENeratorcommand..impedance: ThePOWer:POWer<x>:CLRESPONSE:IMPEDancecommand..inputsource: ThePOWer:POWer<x>:CLRESPONSE:INPUTSOurcecommand..mon: ThePOWer:POWer<x>:CLRESPONSE:MONcommand..outputsource: ThePOWer:POWer<x>:CLRESPONSE:OUTPUTSOurcecommand..ppd: ThePOWer:POWer<x>:CLRESPONSE:PPDcommand..startfrequency: ThePOWer:POWer<x>:CLRESPONSE:STARTFREQuencycommand..stopfrequency: ThePOWer:POWer<x>:CLRESPONSE:STOPFREQuencycommand..testconnection: ThePOWer:POWer<x>:CLRESPONSE:TESTCONNectioncommand.
ampmode
property
¶
ampmode: PowerPowerItemClresponseAmpmode
Return the POWer:POWer<x>:CLRESPONSE:AMPMode command.
Description
- This command sets or queries the amplitude mode for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:AMPMode?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:AMPMode?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:AMPMode valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:AMPMode {CONSTant|PROFile}
- POWer:POWer<x>:CLRESPONSE:AMPMode?
Info
POWer<x>sets the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.CONSTantsets the amplitude mode to output a constant amplitude signal from the DUT stimulus generator for all frequency bands.PROFileenables configuring the generator to set amplitude values for each frequency band.
ampval
property
¶
ampval: dict[int, PowerPowerItemClresponseAmpvalItem]
Return the POWer:POWer<x>:CLRESPONSE:AMP<x>Val command.
Description
- This command sets or queries the generator amplitude value of the specified configuration step for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:AMP<x>Val?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:AMP<x>Val?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:AMP<x>Val valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:AMP<x>Val <NR3>
- POWer:POWer<x>:CLRESPONSE:AMP<x>Val?
Info
POWer<x>sets the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.AMP<x>sets the configuration step number, in the range of 1 to 10. Values outside this range will report an error.<NR3>sets the generator amplitude for the specified configuration step, in the range of -100 V to 100 V.
analysismethod
property
¶
analysismethod: PowerPowerItemClresponseAnalysismethod
Return the POWer:POWer<x>:CLRESPONSE:ANALYSISMethod command.
Description
- This command sets or queries the Analysis Method for Control Loop Response measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:ANALYSISMethod?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:ANALYSISMethod?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:ANALYSISMethod valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:ANALYSISMethod {SV|FFT}
- POWer:POWer<x>:CLRESPONSE:ANALYSISMethod?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.SVsets the Analysis Method as Spectrum View.FFTsets the Analysis Method as FFT.
autorbw
property
¶
autorbw: PowerPowerItemClresponseAutorbw
Return the POWer:POWer<x>:CLRESPONSE:AUTORbw command.
Description
- This command enables Auto RBW computation.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:AUTORbw?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:AUTORbw?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:AUTORbw valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:AUTORbw {True|False}
- POWer:POWer<x>:CLRESPONSE:AUTORbw?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.Trueenables Auto RBW computation.Falsedisables Auto RBW computation.
connectstatus
property
¶
connectstatus: PowerPowerItemClresponseConnectstatus
Return the POWer:POWer<x>:CLRESPONSE:CONNECTSTATus command.
Description
- Queries connection status to the external generator used with the specified Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:CONNECTSTATus?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:CONNECTSTATus?query and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:CONNECTSTATus?
Info
POWer<x>is the number of the power measurement.
constamplitude
property
¶
constamplitude: PowerPowerItemClresponseConstamplitude
Return the POWer:POWer<x>:CLRESPONSE:CONSTAMPlitude command.
Description
- This command sets or queries the constant amplitude voltage for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:CONSTAMPlitude?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:CONSTAMPlitude?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:CONSTAMPlitude valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:CONSTAMPlitude <NR3>
- POWer:POWer<x>:CLRESPONSE:CONSTAMPlitude?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>is the constant amplitude voltage value for the measurement, in the range of -100 V to 100 V.
freqval
property
¶
freqval: dict[int, PowerPowerItemClresponseFreqvalItem]
Return the POWer:POWer<x>:CLRESPONSE:FREQ<x>Val command.
Description
- This command sets or queries the generator frequency value of the specified configuration step for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:FREQ<x>Val?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:FREQ<x>Val?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:FREQ<x>Val valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:FREQ<x>Val <NR3>
- POWer:POWer<x>:CLRESPONSE:FREQ<x>Val?
Info
POWer<x>sets the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.FREQ<x>sets the configuration step number, in the range of 1 to 11. Values outside this range will report an error.<NR3>sets the frequency of the specified configuration step number, in the range of 10 Hz to 50 MHz.
generator
property
¶
generator: PowerPowerItemClresponseGenerator
Return the POWer:POWer<x>:CLRESPONSE:GENerator command.
Description
- Sets or queries the generator source for the specified Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:GENerator?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:GENerator?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:GENerator valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:GENerator {INTernal|EXTernal}
- POWer:POWer<x>:CLRESPONSE:GENerator?
Info
POWer<x>is the number of the power measurement.INTernalsets the internal generator as the source for the Control Loop Response power measurement.EXTernalsets the external generator as the source for the Control Loop Response power measurement.
genipaddress
property
¶
genipaddress: PowerPowerItemClresponseGenipaddress
Return the POWer:POWer<x>:CLRESPONSE:GENIPADDress command.
Description
- Sets or queries the IP address of the external generator to be used with the specified Control Loop Response measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:GENIPADDress?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:GENIPADDress?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:GENIPADDress valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:GENIPADDress <NR2>
- POWer:POWer<x>:CLRESPONSE:GENIPADDress?
Info
POWer<x>is the number of the power measurement.<NR2>is the IP address of the generator.
impedance
property
¶
impedance: PowerPowerItemClresponseImpedance
Return the POWer:POWer<x>:CLRESPONSE:IMPEDance command.
Description
- This command sets or queries the vertical termination impedance for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:IMPEDance?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:IMPEDance?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:IMPEDance valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:IMPEDance {FIFTy|HIGHZ}
- POWer:POWer<x>:CLRESPONSE:IMPEDance?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.FIFTysets the impedance to be 50 Ω.HIGHZsets the impedance to be 1 MΩ.
inputsource
property
¶
inputsource: PowerPowerItemClresponseInputsource
Return the POWer:POWer<x>:CLRESPONSE:INPUTSOurce command.
Description
- This command sets or queries the input source for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:INPUTSOurce CH<x>
- POWer:POWer<x>:CLRESPONSE:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.CH<x>sets the channel to use for the output signal source.
mon
property
¶
Return the POWer:POWer<x>:CLRESPONSE:MON command.
Description
- This command sets or returns the measure on for the control loop response, PSRR or impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:MON?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:MON?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:MON valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:MON {SAVG|SSEQ}
- POWer:POWer<x>:CLRESPONSE:MON?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.SAVGspecifies the measure analysis method for the Control Loop Response/PSRR or Impedance measurement to SAVG.SSEQspecifies the measure analysis method for the Control Loop Response/PSRR or Impedance measurement to SSEQ.
outputsource
property
¶
outputsource: PowerPowerItemClresponseOutputsource
Return the POWer:POWer<x>:CLRESPONSE:OUTPUTSOurce command.
Description
- This command sets or queries the output source for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:OUTPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:OUTPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:OUTPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:OUTPUTSOurce CH<x>
- POWer:POWer<x>:CLRESPONSE:OUTPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.CH<x>sets the channel to use for the output signal source.
ppd
property
¶
Return the POWer:POWer<x>:CLRESPONSE:PPD command.
Description
- This command sets or queries the points per decade (PPD) value for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:PPD?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:PPD?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:PPD valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:PPD <NR3>
- POWer:POWer<x>:CLRESPONSE:PPD?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>is the PPD value for the measurement, in the range of 10 to 100 points.
startfrequency
property
¶
startfrequency: PowerPowerItemClresponseStartfrequency
Return the POWer:POWer<x>:CLRESPONSE:STARTFREQuency command.
Description
- This command sets or queries the start frequency value for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:STARTFREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:STARTFREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:STARTFREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:STARTFREQuency <NR3>
- POWer:POWer<x>:CLRESPONSE:STARTFREQuency?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>is the starting frequency for the measurement, in the range of 10 Hz to 50 MHz.
stopfrequency
property
¶
stopfrequency: PowerPowerItemClresponseStopfrequency
Return the POWer:POWer<x>:CLRESPONSE:STOPFREQuency command.
Description
- This command sets or queries the stop frequency value for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:STOPFREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:STOPFREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:STOPFREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:STOPFREQuency <NR3>
- POWer:POWer<x>:CLRESPONSE:STOPFREQuency?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>is the stop frequency for the measurement, in the range of 10 Hz to 50 MHz.
testconnection
property
¶
testconnection: PowerPowerItemClresponseTestconnection
Return the POWer:POWer<x>:CLRESPONSE:TESTCONNection command.
Description
- This command tests the connection to the external generator used with the specified Control Loop Response power measurement.
Usage
- Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:TESTCONNection valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:TESTCONNection EXECute
Info
POWer<x>is the number of the PSRR power measurement.EXECuteruns the test connection function.
PowerPowerItemClresponseAmpmode
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CLRESPONSE:AMPMode command.
Description
- This command sets or queries the amplitude mode for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:AMPMode?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:AMPMode?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:AMPMode valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:AMPMode {CONSTant|PROFile}
- POWer:POWer<x>:CLRESPONSE:AMPMode?
Info
POWer<x>sets the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.CONSTantsets the amplitude mode to output a constant amplitude signal from the DUT stimulus generator for all frequency bands.PROFileenables configuring the generator to set amplitude values for each frequency band.
PowerPowerItemClresponseAmpvalItem
¶
Bases: ValidatedDynamicNumberCmd, SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CLRESPONSE:AMP<x>Val command.
Description
- This command sets or queries the generator amplitude value of the specified configuration step for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:AMP<x>Val?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:AMP<x>Val?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:AMP<x>Val valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:AMP<x>Val <NR3>
- POWer:POWer<x>:CLRESPONSE:AMP<x>Val?
Info
POWer<x>sets the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.AMP<x>sets the configuration step number, in the range of 1 to 10. Values outside this range will report an error.<NR3>sets the generator amplitude for the specified configuration step, in the range of -100 V to 100 V.
PowerPowerItemClresponseAnalysismethod
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CLRESPONSE:ANALYSISMethod command.
Description
- This command sets or queries the Analysis Method for Control Loop Response measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:ANALYSISMethod?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:ANALYSISMethod?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:ANALYSISMethod valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:ANALYSISMethod {SV|FFT}
- POWer:POWer<x>:CLRESPONSE:ANALYSISMethod?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.SVsets the Analysis Method as Spectrum View.FFTsets the Analysis Method as FFT.
PowerPowerItemClresponseAutorbw
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CLRESPONSE:AUTORbw command.
Description
- This command enables Auto RBW computation.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:AUTORbw?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:AUTORbw?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:AUTORbw valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:AUTORbw {True|False}
- POWer:POWer<x>:CLRESPONSE:AUTORbw?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.Trueenables Auto RBW computation.Falsedisables Auto RBW computation.
PowerPowerItemClresponseConnectstatus
¶
Bases: SCPICmdRead
The POWer:POWer<x>:CLRESPONSE:CONNECTSTATus command.
Description
- Queries connection status to the external generator used with the specified Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:CONNECTSTATus?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:CONNECTSTATus?query and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:CONNECTSTATus?
Info
POWer<x>is the number of the power measurement.
PowerPowerItemClresponseConstamplitude
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CLRESPONSE:CONSTAMPlitude command.
Description
- This command sets or queries the constant amplitude voltage for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:CONSTAMPlitude?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:CONSTAMPlitude?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:CONSTAMPlitude valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:CONSTAMPlitude <NR3>
- POWer:POWer<x>:CLRESPONSE:CONSTAMPlitude?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>is the constant amplitude voltage value for the measurement, in the range of -100 V to 100 V.
PowerPowerItemClresponseFreqvalItem
¶
Bases: ValidatedDynamicNumberCmd, SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CLRESPONSE:FREQ<x>Val command.
Description
- This command sets or queries the generator frequency value of the specified configuration step for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:FREQ<x>Val?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:FREQ<x>Val?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:FREQ<x>Val valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:FREQ<x>Val <NR3>
- POWer:POWer<x>:CLRESPONSE:FREQ<x>Val?
Info
POWer<x>sets the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.FREQ<x>sets the configuration step number, in the range of 1 to 11. Values outside this range will report an error.<NR3>sets the frequency of the specified configuration step number, in the range of 10 Hz to 50 MHz.
PowerPowerItemClresponseGenerator
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CLRESPONSE:GENerator command.
Description
- Sets or queries the generator source for the specified Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:GENerator?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:GENerator?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:GENerator valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:GENerator {INTernal|EXTernal}
- POWer:POWer<x>:CLRESPONSE:GENerator?
Info
POWer<x>is the number of the power measurement.INTernalsets the internal generator as the source for the Control Loop Response power measurement.EXTernalsets the external generator as the source for the Control Loop Response power measurement.
PowerPowerItemClresponseGenipaddress
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CLRESPONSE:GENIPADDress command.
Description
- Sets or queries the IP address of the external generator to be used with the specified Control Loop Response measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:GENIPADDress?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:GENIPADDress?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:GENIPADDress valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:GENIPADDress <NR2>
- POWer:POWer<x>:CLRESPONSE:GENIPADDress?
Info
POWer<x>is the number of the power measurement.<NR2>is the IP address of the generator.
PowerPowerItemClresponseImpedance
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CLRESPONSE:IMPEDance command.
Description
- This command sets or queries the vertical termination impedance for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:IMPEDance?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:IMPEDance?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:IMPEDance valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:IMPEDance {FIFTy|HIGHZ}
- POWer:POWer<x>:CLRESPONSE:IMPEDance?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.FIFTysets the impedance to be 50 Ω.HIGHZsets the impedance to be 1 MΩ.
PowerPowerItemClresponseInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CLRESPONSE:INPUTSOurce command.
Description
- This command sets or queries the input source for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:INPUTSOurce CH<x>
- POWer:POWer<x>:CLRESPONSE:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.CH<x>sets the channel to use for the output signal source.
PowerPowerItemClresponseMon
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CLRESPONSE:MON command.
Description
- This command sets or returns the measure on for the control loop response, PSRR or impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:MON?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:MON?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:MON valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:MON {SAVG|SSEQ}
- POWer:POWer<x>:CLRESPONSE:MON?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.SAVGspecifies the measure analysis method for the Control Loop Response/PSRR or Impedance measurement to SAVG.SSEQspecifies the measure analysis method for the Control Loop Response/PSRR or Impedance measurement to SSEQ.
PowerPowerItemClresponseOutputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CLRESPONSE:OUTPUTSOurce command.
Description
- This command sets or queries the output source for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:OUTPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:OUTPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:OUTPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:OUTPUTSOurce CH<x>
- POWer:POWer<x>:CLRESPONSE:OUTPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.CH<x>sets the channel to use for the output signal source.
PowerPowerItemClresponsePpd
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CLRESPONSE:PPD command.
Description
- This command sets or queries the points per decade (PPD) value for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:PPD?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:PPD?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:PPD valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:PPD <NR3>
- POWer:POWer<x>:CLRESPONSE:PPD?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>is the PPD value for the measurement, in the range of 10 to 100 points.
PowerPowerItemClresponseStartfrequency
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CLRESPONSE:STARTFREQuency command.
Description
- This command sets or queries the start frequency value for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:STARTFREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:STARTFREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:STARTFREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:STARTFREQuency <NR3>
- POWer:POWer<x>:CLRESPONSE:STARTFREQuency?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>is the starting frequency for the measurement, in the range of 10 Hz to 50 MHz.
PowerPowerItemClresponseStopfrequency
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CLRESPONSE:STOPFREQuency command.
Description
- This command sets or queries the stop frequency value for the Control Loop Response power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CLRESPONSE:STOPFREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CLRESPONSE:STOPFREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:STOPFREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:STOPFREQuency <NR3>
- POWer:POWer<x>:CLRESPONSE:STOPFREQuency?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>is the stop frequency for the measurement, in the range of 10 Hz to 50 MHz.
PowerPowerItemClresponseTestconnection
¶
Bases: SCPICmdWrite
The POWer:POWer<x>:CLRESPONSE:TESTCONNection command.
Description
- This command tests the connection to the external generator used with the specified Control Loop Response power measurement.
Usage
- Using the
.write(value)method will send thePOWer:POWer<x>:CLRESPONSE:TESTCONNection valuecommand.
SCPI Syntax
- POWer:POWer<x>:CLRESPONSE:TESTCONNection EXECute
Info
POWer<x>is the number of the PSRR power measurement.EXECuteruns the test connection function.
PowerPowerItemCycleamp
¶
Bases: SCPICmdRead
The POWer:POWer<x>:CYCLEAmp command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEAmp?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEAmp?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Properties
.inputsource: ThePOWer:POWer<x>:CYCLEAmp:INPUTSOurcecommand.
inputsource
property
¶
inputsource: PowerPowerItemCycleampInputsource
Return the POWer:POWer<x>:CYCLEAmp:INPUTSOurce command.
Description
- This command sets or queries the input source for cycle amplitude measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEAmp:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEAmp:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CYCLEAmp:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:CYCLEAmp:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLEAmp:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x> = Achannel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x> = Amath waveform specifier;is ≥1. REF<x> = Areference waveform specifier;is ≥1.
PowerPowerItemCycleampInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CYCLEAmp:INPUTSOurce command.
Description
- This command sets or queries the input source for cycle amplitude measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEAmp:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEAmp:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CYCLEAmp:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:CYCLEAmp:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLEAmp:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x> = Achannel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x> = Amath waveform specifier;is ≥1. REF<x> = Areference waveform specifier;is ≥1.
PowerPowerItemCyclebase
¶
Bases: SCPICmdRead
The POWer:POWer<x>:CYCLEBase command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEBase?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEBase?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Properties
.inputsource: ThePOWer:POWer<x>:CYCLEBase:INPUTSOurcecommand.
inputsource
property
¶
inputsource: PowerPowerItemCyclebaseInputsource
Return the POWer:POWer<x>:CYCLEBase:INPUTSOurce command.
Description
- This command sets or queries the input source for cycle base measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEBase:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEBase:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CYCLEBase:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:CYCLEBase:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLEBase:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemCyclebaseInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CYCLEBase:INPUTSOurce command.
Description
- This command sets or queries the input source for cycle base measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEBase:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEBase:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CYCLEBase:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:CYCLEBase:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLEBase:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemCyclemax
¶
Bases: SCPICmdRead
The POWer:POWer<x>:CYCLEMAX command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEMAX?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEMAX?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Properties
.inputsource: ThePOWer:POWer<x>:CYCLEMAX:INPUTSOurcecommand.
inputsource
property
¶
inputsource: PowerPowerItemCyclemaxInputsource
Return the POWer:POWer<x>:CYCLEMAX:INPUTSOurce command.
Description
- This command sets or queries the input source for cycle maximum measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEMAX:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEMAX:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CYCLEMAX:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:CYCLEMAX:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLEMAX:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemCyclemaxInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CYCLEMAX:INPUTSOurce command.
Description
- This command sets or queries the input source for cycle maximum measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEMAX:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEMAX:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CYCLEMAX:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:CYCLEMAX:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLEMAX:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemCyclemin
¶
Bases: SCPICmdRead
The POWer:POWer<x>:CYCLEMin command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEMin?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEMin?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Properties
.inputsource: ThePOWer:POWer<x>:CYCLEMin:INPUTSOurcecommand.
inputsource
property
¶
inputsource: PowerPowerItemCycleminInputsource
Return the POWer:POWer<x>:CYCLEMin:INPUTSOurce command.
Description
- This command sets or queries the input source for cycle minimum measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEMin:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEMin:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CYCLEMin:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:CYCLEMin:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLEMin:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemCycleminInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CYCLEMin:INPUTSOurce command.
Description
- This command sets or queries the input source for cycle minimum measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEMin:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEMin:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CYCLEMin:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:CYCLEMin:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLEMin:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemCyclepkpk
¶
Bases: SCPICmdRead
The POWer:POWer<x>:CYCLEPKPK command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEPKPK?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEPKPK?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Properties
.inputsource: ThePOWer:POWer<x>:CYCLEPKPK:INPUTSOurcecommand.
inputsource
property
¶
inputsource: PowerPowerItemCyclepkpkInputsource
Return the POWer:POWer<x>:CYCLEPKPK:INPUTSOurce command.
Description
- This command sets or queries the input source for cycle peak-to-peak measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEPKPK:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEPKPK:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CYCLEPKPK:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:CYCLEPKPK:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLEPKPK:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemCyclepkpkInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CYCLEPKPK:INPUTSOurce command.
Description
- This command sets or queries the input source for cycle peak-to-peak measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLEPKPK:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLEPKPK:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CYCLEPKPK:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:CYCLEPKPK:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLEPKPK:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemCycletop
¶
Bases: SCPICmdRead
The POWer:POWer<x>:CYCLETop command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLETop?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLETop?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Properties
.inputsource: ThePOWer:POWer<x>:CYCLETop:INPUTSOurcecommand.
inputsource
property
¶
inputsource: PowerPowerItemCycletopInputsource
Return the POWer:POWer<x>:CYCLETop:INPUTSOurce command.
Description
- This command sets or queries the input source for cycle top measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLETop:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLETop:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CYCLETop:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:CYCLETop:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLETop:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemCycletopInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:CYCLETop:INPUTSOurce command.
Description
- This command sets or queries the input source for cycle top measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:CYCLETop:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:CYCLETop:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:CYCLETop:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:CYCLETop:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:CYCLETop:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemDidt
¶
Bases: SCPICmdRead
The POWer:POWer<x>:DIDT command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:DIDT?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:DIDT?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Properties
.inputsource: ThePOWer:POWer<x>:DIDT:INPUTSOurcecommand..sourceedgetype: ThePOWer:POWer<x>:DIDT:SOURCEEDGETypecommand.
inputsource
property
¶
inputsource: PowerPowerItemDidtInputsource
Return the POWer:POWer<x>:DIDT:INPUTSOurce command.
Description
- This command sets or queries the input source for di/dt measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:DIDT:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:DIDT:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:DIDT:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:DIDT:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:DIDT:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
sourceedgetype
property
¶
sourceedgetype: PowerPowerItemDidtSourceedgetype
Return the POWer:POWer<x>:DIDT:SOURCEEDGEType command.
Description
- This command sets or queries the edge type for di/dt measurement of the specified
power measurement number.
specifies the number of the power measurement badge.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:DIDT:SOURCEEDGEType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:DIDT:SOURCEEDGEType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:DIDT:SOURCEEDGEType valuecommand.
SCPI Syntax
- POWer:POWer<x>:DIDT:SOURCEEDGEType {RISE|FALL}
- POWer:POWer<x>:DIDT:SOURCEEDGEType?
PowerPowerItemDidtInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:DIDT:INPUTSOurce command.
Description
- This command sets or queries the input source for di/dt measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:DIDT:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:DIDT:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:DIDT:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:DIDT:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:DIDT:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemDidtSourceedgetype
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:DIDT:SOURCEEDGEType command.
Description
- This command sets or queries the edge type for di/dt measurement of the specified power
measurement number.
specifies the number of the power measurement badge.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:DIDT:SOURCEEDGEType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:DIDT:SOURCEEDGEType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:DIDT:SOURCEEDGEType valuecommand.
SCPI Syntax
- POWer:POWer<x>:DIDT:SOURCEEDGEType {RISE|FALL}
- POWer:POWer<x>:DIDT:SOURCEEDGEType?
PowerPowerItemDvdt
¶
Bases: SCPICmdRead
The POWer:POWer<x>:DVDT command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:DVDT?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:DVDT?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Properties
.inputsource: ThePOWer:POWer<x>:DVDT:INPUTSOurcecommand..sourceedgetype: ThePOWer:POWer<x>:DVDT:SOURCEEDGETypecommand.
inputsource
property
¶
inputsource: PowerPowerItemDvdtInputsource
Return the POWer:POWer<x>:DVDT:INPUTSOurce command.
Description
- This command sets or queries the input source for dv/dt measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:DVDT:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:DVDT:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:DVDT:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:DVDT:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:DVDT:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
sourceedgetype
property
¶
sourceedgetype: PowerPowerItemDvdtSourceedgetype
Return the POWer:POWer<x>:DVDT:SOURCEEDGEType command.
Description
- This command sets or queries the edge type for dv/dt measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:DVDT:SOURCEEDGEType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:DVDT:SOURCEEDGEType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:DVDT:SOURCEEDGEType valuecommand.
SCPI Syntax
- POWer:POWer<x>:DVDT:SOURCEEDGEType {RISE|FALL}
- POWer:POWer<x>:DVDT:SOURCEEDGEType?
PowerPowerItemDvdtInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:DVDT:INPUTSOurce command.
Description
- This command sets or queries the input source for dv/dt measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:DVDT:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:DVDT:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:DVDT:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:DVDT:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:DVDT:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemDvdtSourceedgetype
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:DVDT:SOURCEEDGEType command.
Description
- This command sets or queries the edge type for dv/dt measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:DVDT:SOURCEEDGEType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:DVDT:SOURCEEDGEType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:DVDT:SOURCEEDGEType valuecommand.
SCPI Syntax
- POWer:POWer<x>:DVDT:SOURCEEDGEType {RISE|FALL}
- POWer:POWer<x>:DVDT:SOURCEEDGEType?
PowerPowerItemEfficiency
¶
Bases: SCPICmdRead
The POWer:POWer<x>:EFFICIENCY command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown on the power measurement badge of the UI.
Properties
.inputtype: ThePOWer:POWer<x>:EFFICIENCY:INPUTTypecommand..iout1source: ThePOWer:POWer<x>:EFFICIENCY:IOUT1SOUrcecommand..iout2source: ThePOWer:POWer<x>:EFFICIENCY:IOUT2SOUrcecommand..iout3source: ThePOWer:POWer<x>:EFFICIENCY:IOUT3SOUrcecommand..isource: ThePOWer:POWer<x>:EFFICIENCY:ISOUrcecommand..numofoutputs: ThePOWer:POWer<x>:EFFICIENCY:NUMOFOutputscommand..output1type: ThePOWer:POWer<x>:EFFICIENCY:OUTPUT1Typecommand..output2type: ThePOWer:POWer<x>:EFFICIENCY:OUTPUT2Typecommand..output3type: ThePOWer:POWer<x>:EFFICIENCY:OUTPUT3Typecommand..outputtype: ThePOWer:POWer<x>:EFFICIENCY:OUTPUTTypecommand..vout1source: ThePOWer:POWer<x>:EFFICIENCY:VOUT1SOUrcecommand..vout2source: ThePOWer:POWer<x>:EFFICIENCY:VOUT2SOUrcecommand..vout3source: ThePOWer:POWer<x>:EFFICIENCY:VOUT3SOUrcecommand..vsource: ThePOWer:POWer<x>:EFFICIENCY:VSOUrcecommand.
inputtype
property
¶
inputtype: PowerPowerItemEfficiencyInputtype
Return the POWer:POWer<x>:EFFICIENCY:INPUTType command.
Description
- This command sets or queries the input type (AC or DC) for power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:INPUTType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:INPUTType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:INPUTType valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:INPUTType {AC|DC}
- POWer:POWer<x>:EFFICIENCY:INPUTType?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown on the power measurement badge of the UI.ACsets the input voltage type to AC.DCsets the input voltage type to DC.
iout1source
property
¶
iout1source: PowerPowerItemEfficiencyIout1source
Return the POWer:POWer<x>:EFFICIENCY:IOUT1SOUrce command.
Description
- This command sets or queries the output 1 current source for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:IOUT1SOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:IOUT1SOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:IOUT1SOUrce valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:IOUT1SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:IOUT1SOUrce?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
iout2source
property
¶
iout2source: PowerPowerItemEfficiencyIout2source
Return the POWer:POWer<x>:EFFICIENCY:IOUT2SOUrce command.
Description
- This command sets or queries the output 2 current source for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:IOUT2SOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:IOUT2SOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:IOUT2SOUrce valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:IOUT2SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:IOUT2SOUrce?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
iout3source
property
¶
iout3source: PowerPowerItemEfficiencyIout3source
Return the POWer:POWer<x>:EFFICIENCY:IOUT3SOUrce command.
Description
- This command sets or queries the output 3 current source for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:IOUT3SOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:IOUT3SOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:IOUT3SOUrce valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:IOUT3SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:IOUT3SOUrce?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
isource
property
¶
isource: PowerPowerItemEfficiencyIsource
Return the POWer:POWer<x>:EFFICIENCY:ISOUrce command.
Description
- This command sets or queries the current source for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:ISOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:ISOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:EFFICIENCY:ISOUrcecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:ISOUrce
- POWer:POWer<x>:EFFICIENCY:ISOUrce?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
numofoutputs
property
¶
numofoutputs: PowerPowerItemEfficiencyNumofoutputs
Return the POWer:POWer<x>:EFFICIENCY:NUMOFOutputs command.
Description
- This command sets or queries the number of outputs for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:NUMOFOutputs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:NUMOFOutputs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:NUMOFOutputs valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:NUMOFOutputs {ONE|TWO|THREE}
- POWer:POWer<x>:EFFICIENCY:NUMOFOutputs?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.ONE, TWO, THREE sets the number of outputs to test in the power efficiency measurement.
output1type
property
¶
output1type: PowerPowerItemEfficiencyOutput1type
Return the POWer:POWer<x>:EFFICIENCY:OUTPUT1Type command.
Description
- This command sets or queries the Output1 type (AC or DC) for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT1Type?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT1Type?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT1Type valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:OUTPUT1Type {AC|DC}
- POWer:POWer<x>:EFFICIENCY:OUTPUT1Type?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown on the power measurement badge of the UI.ACsets the output1 voltage type to AC.DCsets the output1 voltage type to DC.
output2type
property
¶
output2type: PowerPowerItemEfficiencyOutput2type
Return the POWer:POWer<x>:EFFICIENCY:OUTPUT2Type command.
Description
- This command sets or queries the Output2 type (AC or DC) for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT2Type?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT2Type?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT2Type valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:OUTPUT2Type {AC|DC}
- POWer:POWer<x>:EFFICIENCY:OUTPUT2Type?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown on the power measurement badge of the UI.ACsets the output2 voltage type to AC.DCsets the output2 voltage type to DC.
output3type
property
¶
output3type: PowerPowerItemEfficiencyOutput3type
Return the POWer:POWer<x>:EFFICIENCY:OUTPUT3Type command.
Description
- This command sets or queries the Output3 type (AC or DC) for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT3Type?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT3Type?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT3Type valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:OUTPUT3Type {AC|DC}
- POWer:POWer<x>:EFFICIENCY:OUTPUT3Type?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown on the power measurement badge of the UI.ACsets the output3 voltage type to AC.DCsets the output3 voltage type to DC.
outputtype
property
¶
outputtype: PowerPowerItemEfficiencyOutputtype
Return the POWer:POWer<x>:EFFICIENCY:OUTPUTType command.
Description
- This command sets or queries the output type (AC or DC) for power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUTType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUTType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUTType valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:OUTPUTType {AC|DC}
- POWer:POWer<x>:EFFICIENCY:OUTPUTType?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown on the power measurement badge of the UI.ACsets the output3 voltage type to AC.DCsets the output3 voltage type to DC.
vout1source
property
¶
vout1source: PowerPowerItemEfficiencyVout1source
Return the POWer:POWer<x>:EFFICIENCY:VOUT1SOUrce command.
Description
- This command sets or queries the output 1 voltage source for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:VOUT1SOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:VOUT1SOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:VOUT1SOUrce valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:VOUT1SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:VOUT1SOUrce?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
vout2source
property
¶
vout2source: PowerPowerItemEfficiencyVout2source
Return the POWer:POWer<x>:EFFICIENCY:VOUT2SOUrce command.
Description
- This command sets or queries the output 2 voltage source for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:VOUT2SOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:VOUT2SOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:VOUT2SOUrce valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:VOUT2SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:VOUT2SOUrce?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
vout3source
property
¶
vout3source: PowerPowerItemEfficiencyVout3source
Return the POWer:POWer<x>:EFFICIENCY:VOUT3SOUrce command.
Description
- This command sets or queries the output 3 voltage source for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:VOUT3SOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:VOUT3SOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:VOUT3SOUrce valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:VOUT3SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:VOUT3SOUrce?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
vsource
property
¶
vsource: PowerPowerItemEfficiencyVsource
Return the POWer:POWer<x>:EFFICIENCY:VSOUrce command.
Description
- This command sets or queries the voltage source for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:VSOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:VSOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:EFFICIENCY:VSOUrcecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:VSOUrce
- POWer:POWer<x>:EFFICIENCY:VSOUrce?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemEfficiencyInputtype
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:EFFICIENCY:INPUTType command.
Description
- This command sets or queries the input type (AC or DC) for power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:INPUTType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:INPUTType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:INPUTType valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:INPUTType {AC|DC}
- POWer:POWer<x>:EFFICIENCY:INPUTType?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown on the power measurement badge of the UI.ACsets the input voltage type to AC.DCsets the input voltage type to DC.
PowerPowerItemEfficiencyIout1source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:EFFICIENCY:IOUT1SOUrce command.
Description
- This command sets or queries the output 1 current source for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:IOUT1SOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:IOUT1SOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:IOUT1SOUrce valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:IOUT1SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:IOUT1SOUrce?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemEfficiencyIout2source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:EFFICIENCY:IOUT2SOUrce command.
Description
- This command sets or queries the output 2 current source for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:IOUT2SOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:IOUT2SOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:IOUT2SOUrce valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:IOUT2SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:IOUT2SOUrce?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemEfficiencyIout3source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:EFFICIENCY:IOUT3SOUrce command.
Description
- This command sets or queries the output 3 current source for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:IOUT3SOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:IOUT3SOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:IOUT3SOUrce valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:IOUT3SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:IOUT3SOUrce?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemEfficiencyIsource
¶
Bases: SCPICmdWriteNoArguments, SCPICmdRead
The POWer:POWer<x>:EFFICIENCY:ISOUrce command.
Description
- This command sets or queries the current source for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:ISOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:ISOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:EFFICIENCY:ISOUrcecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:ISOUrce
- POWer:POWer<x>:EFFICIENCY:ISOUrce?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemEfficiencyNumofoutputs
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:EFFICIENCY:NUMOFOutputs command.
Description
- This command sets or queries the number of outputs for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:NUMOFOutputs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:NUMOFOutputs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:NUMOFOutputs valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:NUMOFOutputs {ONE|TWO|THREE}
- POWer:POWer<x>:EFFICIENCY:NUMOFOutputs?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.ONE, TWO, THREE sets the number of outputs to test in the power efficiency measurement.
PowerPowerItemEfficiencyOutput1type
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:EFFICIENCY:OUTPUT1Type command.
Description
- This command sets or queries the Output1 type (AC or DC) for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT1Type?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT1Type?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT1Type valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:OUTPUT1Type {AC|DC}
- POWer:POWer<x>:EFFICIENCY:OUTPUT1Type?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown on the power measurement badge of the UI.ACsets the output1 voltage type to AC.DCsets the output1 voltage type to DC.
PowerPowerItemEfficiencyOutput2type
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:EFFICIENCY:OUTPUT2Type command.
Description
- This command sets or queries the Output2 type (AC or DC) for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT2Type?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT2Type?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT2Type valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:OUTPUT2Type {AC|DC}
- POWer:POWer<x>:EFFICIENCY:OUTPUT2Type?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown on the power measurement badge of the UI.ACsets the output2 voltage type to AC.DCsets the output2 voltage type to DC.
PowerPowerItemEfficiencyOutput3type
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:EFFICIENCY:OUTPUT3Type command.
Description
- This command sets or queries the Output3 type (AC or DC) for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT3Type?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT3Type?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUT3Type valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:OUTPUT3Type {AC|DC}
- POWer:POWer<x>:EFFICIENCY:OUTPUT3Type?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown on the power measurement badge of the UI.ACsets the output3 voltage type to AC.DCsets the output3 voltage type to DC.
PowerPowerItemEfficiencyOutputtype
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:EFFICIENCY:OUTPUTType command.
Description
- This command sets or queries the output type (AC or DC) for power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUTType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUTType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:OUTPUTType valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:OUTPUTType {AC|DC}
- POWer:POWer<x>:EFFICIENCY:OUTPUTType?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown on the power measurement badge of the UI.ACsets the output3 voltage type to AC.DCsets the output3 voltage type to DC.
PowerPowerItemEfficiencyVout1source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:EFFICIENCY:VOUT1SOUrce command.
Description
- This command sets or queries the output 1 voltage source for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:VOUT1SOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:VOUT1SOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:VOUT1SOUrce valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:VOUT1SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:VOUT1SOUrce?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemEfficiencyVout2source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:EFFICIENCY:VOUT2SOUrce command.
Description
- This command sets or queries the output 2 voltage source for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:VOUT2SOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:VOUT2SOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:VOUT2SOUrce valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:VOUT2SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:VOUT2SOUrce?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemEfficiencyVout3source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:EFFICIENCY:VOUT3SOUrce command.
Description
- This command sets or queries the output 3 voltage source for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:VOUT3SOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:VOUT3SOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:EFFICIENCY:VOUT3SOUrce valuecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:VOUT3SOUrce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:EFFICIENCY:VOUT3SOUrce?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemEfficiencyVsource
¶
Bases: SCPICmdWriteNoArguments, SCPICmdRead
The POWer:POWer<x>:EFFICIENCY:VSOUrce command.
Description
- This command sets or queries the voltage source for the power Efficiency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:EFFICIENCY:VSOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:EFFICIENCY:VSOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:EFFICIENCY:VSOUrcecommand.
SCPI Syntax
- POWer:POWer<x>:EFFICIENCY:VSOUrce
- POWer:POWer<x>:EFFICIENCY:VSOUrce?
Info
POWer<x>is the number of a power efficiency measurement. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemFrequency
¶
Bases: SCPICmdRead
The POWer:POWer<x>:FREQUENCY command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:FREQUENCY?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:FREQUENCY?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Properties
.edge: ThePOWer:POWer<x>:FREQUENCY:EDGecommand..inputsource: ThePOWer:POWer<x>:FREQUENCY:INPUTSOurcecommand.
edge
property
¶
Return the POWer:POWer<x>:FREQUENCY:EDGe command.
Description
- This command sets or queries the edge type for frequency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:FREQUENCY:EDGe?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:FREQUENCY:EDGe?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:FREQUENCY:EDGe valuecommand.
SCPI Syntax
- POWer:POWer<x>:FREQUENCY:EDGe {RISE|FALL}
- POWer:POWer<x>:FREQUENCY:EDGe?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
inputsource
property
¶
inputsource: PowerPowerItemFrequencyInputsource
Return the POWer:POWer<x>:FREQUENCY:INPUTSOurce command.
Description
- This command sets or queries the input source for frequency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:FREQUENCY:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:FREQUENCY:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:FREQUENCY:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:FREQUENCY:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:FREQUENCY:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemFrequencyEdge
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:FREQUENCY:EDGe command.
Description
- This command sets or queries the edge type for frequency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:FREQUENCY:EDGe?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:FREQUENCY:EDGe?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:FREQUENCY:EDGe valuecommand.
SCPI Syntax
- POWer:POWer<x>:FREQUENCY:EDGe {RISE|FALL}
- POWer:POWer<x>:FREQUENCY:EDGe?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
PowerPowerItemFrequencyInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:FREQUENCY:INPUTSOurce command.
Description
- This command sets or queries the input source for frequency measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:FREQUENCY:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:FREQUENCY:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:FREQUENCY:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:FREQUENCY:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:FREQUENCY:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemGating
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:GATing command.
Description
- This command sets or queries the gating type for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:GATing?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:GATing?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:GATing valuecommand.
SCPI Syntax
- POWer:POWer<x>:GATing {NONE|CURSOR|SCREEN|LOGIC}
- POWer:POWer<x>:GATing?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.NONEmakes measurement across the entire waveform record.CURSORmakes measurements on that portion of the waveform between the cursors. Selecting Cursors opens cursors on the measurement source. Set the cursors so that the waveform area of interest is in between the cursors.SCREENtakes measurements on that portion of the waveform shown in the display. When Zoom is on, the display is the zoom window.LOGICtakes measurements only when the logical state of a specified waveform is true.
Properties
.global: ThePOWer:POWer<x>:GATing:GLOBalcommand.
global_
property
¶
global_: PowerPowerItemGatingGlobal
Return the POWer:POWer<x>:GATing:GLOBal command.
Description
- This command sets or queries the gating settings for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:GATing:GLOBal?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:GATing:GLOBal?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:GATing:GLOBal valuecommand.
SCPI Syntax
- POWer:POWer<x>:GATing:GLOBal {ON|OFF|1|0}
- POWer:POWer<x>:GATing:GLOBal?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.1selects the gating settings as Global.ONselects the gating settings as Global.0selects the gating settings as Local.OFFselects the gating settings as Local.
PowerPowerItemGatingGlobal
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:GATing:GLOBal command.
Description
- This command sets or queries the gating settings for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:GATing:GLOBal?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:GATing:GLOBal?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:GATing:GLOBal valuecommand.
SCPI Syntax
- POWer:POWer<x>:GATing:GLOBal {ON|OFF|1|0}
- POWer:POWer<x>:GATing:GLOBal?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.1selects the gating settings as Global.ONselects the gating settings as Global.0selects the gating settings as Local.OFFselects the gating settings as Local.
PowerPowerItemHarmonics
¶
Bases: SCPICmdRead
The POWer:POWer<x>:HARMONICS command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS?query and raise an AssertionError if the returned value does not matchvalue.
Properties
.class: ThePOWer:POWer<x>:HARMONICS:CLASscommand..clfile: ThePOWer:POWer<x>:HARMONICS:CLFilecommand..cmethod: ThePOWer:POWer<x>:HARMONICS:CMEThodcommand..fundcurrent: ThePOWer:POWer<x>:HARMONICS:FUNDCURRentcommand..horder: ThePOWer:POWer<x>:HARMONICS:HORDercommand..hsource: ThePOWer:POWer<x>:HARMONICS:HSOURcecommand..ipower: ThePOWer:POWer<x>:HARMONICS:IPOWercommand..isource: ThePOWer:POWer<x>:HARMONICS:ISOURcecommand..linefrequency: ThePOWer:POWer<x>:HARMONICS:LINEFREQUEncycommand..oddeven: ThePOWer:POWer<x>:HARMONICS:ODDEVencommand..pfactor: ThePOWer:POWer<x>:HARMONICS:PFACtorcommand..powerrating: ThePOWer:POWer<x>:HARMONICS:POWERRatingcommand..rcurrent: ThePOWer:POWer<x>:HARMONICS:RCURRentcommand..standard: ThePOWer:POWer<x>:HARMONICS:STANDardcommand..startfrequency: ThePOWer:POWer<x>:HARMONICS:STARTFREQUEncycommand..units: ThePOWer:POWer<x>:HARMONICS:UNITscommand..vsource: ThePOWer:POWer<x>:HARMONICS:VSOURcecommand.
class_
property
¶
class_: PowerPowerItemHarmonicsClass
Return the POWer:POWer<x>:HARMONICS:CLASs command.
Description
- This command sets or queries the class type for the harmonics measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:CLASs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:CLASs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:CLASs valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:CLASs {CLASSA|CLASSB|CLASSC|CLASSD}
- POWer:POWer<x>:HARMONICS:CLASs?
clfile
property
¶
clfile: PowerPowerItemHarmonicsClfile
Return the POWer:POWer<x>:HARMONICS:CLFile command.
Description
- This command sets or queries the custom limits file path for the harmonics measurement. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:CLFile?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:CLFile?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:CLFile valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:CLFile <QString>
- POWer:POWer<x>:HARMONICS:CLFile?
Info
<QString>specifies the custom limits file path that can be loaded when the harmonics standard type selected is CUSTom.
cmethod
property
¶
cmethod: PowerPowerItemHarmonicsCmethod
Return the POWer:POWer<x>:HARMONICS:CMEThod command.
Description
- This command sets or queries the fundamental current method for the harmonics measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:CMEThod?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:CMEThod?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:CMEThod valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:CMEThod {RATed|MEASured}
- POWer:POWer<x>:HARMONICS:CMEThod?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.RATed:select to use the standard input current values in the measurement.MEASured:select to use the measured the input current values in the measurement.
fundcurrent
property
¶
fundcurrent: PowerPowerItemHarmonicsFundcurrent
Return the POWer:POWer<x>:HARMONICS:FUNDCURRent command.
Description
- This command sets or queries the fundamental current value for the harmonics measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:FUNDCURRent?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:FUNDCURRent?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:FUNDCURRent valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:FUNDCURRent <NR1>
- POWer:POWer<x>:HARMONICS:FUNDCURRent?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR1>ranges from 0 to 16.
horder
property
¶
horder: PowerPowerItemHarmonicsHorder
Return the POWer:POWer<x>:HARMONICS:HORDer command.
Description
- This command sets or queries the order value for the harmonics measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:HORDer?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:HORDer?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:HORDer valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:HORDer <NR1>
- POWer:POWer<x>:HARMONICS:HORDer?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR1>ranges from 40 to 100.
hsource
property
¶
hsource: PowerPowerItemHarmonicsHsource
Return the POWer:POWer<x>:HARMONICS:HSOURce command.
Description
- This command sets or queries the source type for the harmonics measurement of the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:HSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:HSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:HSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:HSOURce {CURRent|VOLTage}
- POWer:POWer<x>:HARMONICS:HSOURce?
ipower
property
¶
ipower: PowerPowerItemHarmonicsIpower
Return the POWer:POWer<x>:HARMONICS:IPOWer command.
Description
- This command sets or queries the input power value for the harmonics measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:IPOWer?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:IPOWer?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:IPOWer valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:IPOWer <NR1>
- POWer:POWer<x>:HARMONICS:IPOWer?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR1>ranges from 0 to 600.
isource
property
¶
isource: PowerPowerItemHarmonicsIsource
Return the POWer:POWer<x>:HARMONICS:ISOURce command.
Description
- This command sets or queries the current source for SOA measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:HARMONICS:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
linefrequency
property
¶
linefrequency: PowerPowerItemHarmonicsLinefrequency
Return the POWer:POWer<x>:HARMONICS:LINEFREQUEncy command.
Description
- This command sets or queries the value for the line frequency for the Harmonics measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:LINEFREQUEncy?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:LINEFREQUEncy?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:LINEFREQUEncy valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:LINEFREQUEncy {Auto|FIFTyhz|SIXTyhz|THREESIXTyhz|FOURHUNDREdhz|SIXFIFTyhz|EIGHTHUNDREdhz|CUSTom}
- POWer:POWer<x>:HARMONICS:LINEFREQUEncy?
Info
POWer<x>is the Power measurement identifier number.Autoautomatically detects and sets the line frequency value.FIFTyhzsets the line frequency value to 50 Hz.SIXTyhzsets the line frequency value to 60 Hz.THREESIXTyhzsets the line frequency value to 360 Hz.FOURHUNDREdhzsets the line frequency value to 400 Hz.SIXFIFTyhzsets the line frequency value to 650 Hz.EIGHTHUNDREdhzsets the line frequency value to 800 Hz.CUSTomsets the line frequency value to Custom. The default value for custom is 100 Hz. Use thePOWer:POWer<x>:HARMONICS:LINEFREQUEncycommand to set a custom line frequency value.
oddeven
property
¶
oddeven: PowerPowerItemHarmonicsOddeven
Return the POWer:POWer<x>:HARMONICS:ODDEVen command.
Description
- This command sets or queries the harmonics value analysis format of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:ODDEVen?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:ODDEVen?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:ODDEVen valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:ODDEVen {ALL|ODD|EVEN}
- POWer:POWer<x>:HARMONICS:ODDEVen?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.ALLto display all harmonics values.ODDto display only the odd values of harmonics.EVENto display only the even values of harmonics.
pfactor
property
¶
pfactor: PowerPowerItemHarmonicsPfactor
Return the POWer:POWer<x>:HARMONICS:PFACtor command.
Description
- This command sets or queries the value of power factor for the harmonics measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:PFACtor?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:PFACtor?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:PFACtor valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:PFACtor <NR1>
- POWer:POWer<x>:HARMONICS:PFACtor?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR1>ranges from 0 to 1.
powerrating
property
¶
powerrating: PowerPowerItemHarmonicsPowerrating
Return the POWer:POWer<x>:HARMONICS:POWERRating command.
Description
- This command sets or queries the power level for the harmonics measurement of the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:POWERRating?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:POWERRating?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:POWERRating valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:POWERRating {HIGH|LOW}
- POWer:POWer<x>:HARMONICS:POWERRating?
rcurrent
property
¶
rcurrent: PowerPowerItemHarmonicsRcurrent
Return the POWer:POWer<x>:HARMONICS:RCURRent command.
Description
- This command sets or queries the rated current for the harmonics measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:RCURRent?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:RCURRent?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:RCURRent valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:RCURRent <NR1>
- POWer:POWer<x>:HARMONICS:RCURRent?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR1>ranges from 0 to 100.
standard
property
¶
standard: PowerPowerItemHarmonicsStandard
Return the POWer:POWer<x>:HARMONICS:STANDard command.
Description
- This command sets or queries the test mode for harmonics measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:STANDard?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:STANDard?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:STANDard valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:STANDard {NONe|IEC|MIL|AM14|DO160|CUSTOM}
- POWer:POWer<x>:HARMONICS:STANDard?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.NONe= No standard.IEC= IEC 61000-3-2 standard.MIL= MIL-STD-1399 standard.AM14= AM14 standard.DO160= DO160 standard.CUSTOM= CUSTOM standard.
startfrequency
property
¶
startfrequency: PowerPowerItemHarmonicsStartfrequency
Return the POWer:POWer<x>:HARMONICS:STARTFREQUEncy command.
Description
- This command sets or queries the value for the start frequency for the Harmonics measurement. in the range of 1 Hz to 1 GHz.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:STARTFREQUEncy?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:STARTFREQUEncy?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:STARTFREQUEncy valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:STARTFREQUEncy <NR2>
- POWer:POWer<x>:HARMONICS:STARTFREQUEncy?
Info
POWer<x>is the Power measurement.<NR2>sets the starting frequency, in hertz.
units
property
¶
units: PowerPowerItemHarmonicsUnits
Return the POWer:POWer<x>:HARMONICS:UNITs command.
Description
- This command sets or queries the harmonics results units of the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:UNITs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:UNITs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:UNITs valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:UNITs {LOG|LINear}
- POWer:POWer<x>:HARMONICS:UNITs?
vsource
property
¶
vsource: PowerPowerItemHarmonicsVsource
Return the POWer:POWer<x>:HARMONICS:VSOURce command.
Description
- This command sets or queries the voltage source for SOA measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:HARMONICS:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemHarmonicsClass
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:HARMONICS:CLASs command.
Description
- This command sets or queries the class type for the harmonics measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:CLASs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:CLASs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:CLASs valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:CLASs {CLASSA|CLASSB|CLASSC|CLASSD}
- POWer:POWer<x>:HARMONICS:CLASs?
PowerPowerItemHarmonicsClfile
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:HARMONICS:CLFile command.
Description
- This command sets or queries the custom limits file path for the harmonics measurement. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:CLFile?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:CLFile?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:CLFile valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:CLFile <QString>
- POWer:POWer<x>:HARMONICS:CLFile?
Info
<QString>specifies the custom limits file path that can be loaded when the harmonics standard type selected is CUSTom.
PowerPowerItemHarmonicsCmethod
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:HARMONICS:CMEThod command.
Description
- This command sets or queries the fundamental current method for the harmonics measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:CMEThod?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:CMEThod?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:CMEThod valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:CMEThod {RATed|MEASured}
- POWer:POWer<x>:HARMONICS:CMEThod?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.RATed:select to use the standard input current values in the measurement.MEASured:select to use the measured the input current values in the measurement.
PowerPowerItemHarmonicsFundcurrent
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:HARMONICS:FUNDCURRent command.
Description
- This command sets or queries the fundamental current value for the harmonics measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:FUNDCURRent?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:FUNDCURRent?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:FUNDCURRent valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:FUNDCURRent <NR1>
- POWer:POWer<x>:HARMONICS:FUNDCURRent?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR1>ranges from 0 to 16.
PowerPowerItemHarmonicsHorder
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:HARMONICS:HORDer command.
Description
- This command sets or queries the order value for the harmonics measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:HORDer?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:HORDer?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:HORDer valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:HORDer <NR1>
- POWer:POWer<x>:HARMONICS:HORDer?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR1>ranges from 40 to 100.
PowerPowerItemHarmonicsHsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:HARMONICS:HSOURce command.
Description
- This command sets or queries the source type for the harmonics measurement of the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:HSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:HSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:HSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:HSOURce {CURRent|VOLTage}
- POWer:POWer<x>:HARMONICS:HSOURce?
PowerPowerItemHarmonicsIpower
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:HARMONICS:IPOWer command.
Description
- This command sets or queries the input power value for the harmonics measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:IPOWer?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:IPOWer?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:IPOWer valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:IPOWer <NR1>
- POWer:POWer<x>:HARMONICS:IPOWer?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR1>ranges from 0 to 600.
PowerPowerItemHarmonicsIsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:HARMONICS:ISOURce command.
Description
- This command sets or queries the current source for SOA measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:HARMONICS:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemHarmonicsLinefrequency
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:HARMONICS:LINEFREQUEncy command.
Description
- This command sets or queries the value for the line frequency for the Harmonics measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:LINEFREQUEncy?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:LINEFREQUEncy?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:LINEFREQUEncy valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:LINEFREQUEncy {Auto|FIFTyhz|SIXTyhz|THREESIXTyhz|FOURHUNDREdhz|SIXFIFTyhz|EIGHTHUNDREdhz|CUSTom}
- POWer:POWer<x>:HARMONICS:LINEFREQUEncy?
Info
POWer<x>is the Power measurement identifier number.Autoautomatically detects and sets the line frequency value.FIFTyhzsets the line frequency value to 50 Hz.SIXTyhzsets the line frequency value to 60 Hz.THREESIXTyhzsets the line frequency value to 360 Hz.FOURHUNDREdhzsets the line frequency value to 400 Hz.SIXFIFTyhzsets the line frequency value to 650 Hz.EIGHTHUNDREdhzsets the line frequency value to 800 Hz.CUSTomsets the line frequency value to Custom. The default value for custom is 100 Hz. Use thePOWer:POWer<x>:HARMONICS:LINEFREQUEncycommand to set a custom line frequency value.
PowerPowerItemHarmonicsOddeven
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:HARMONICS:ODDEVen command.
Description
- This command sets or queries the harmonics value analysis format of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:ODDEVen?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:ODDEVen?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:ODDEVen valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:ODDEVen {ALL|ODD|EVEN}
- POWer:POWer<x>:HARMONICS:ODDEVen?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.ALLto display all harmonics values.ODDto display only the odd values of harmonics.EVENto display only the even values of harmonics.
PowerPowerItemHarmonicsPfactor
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:HARMONICS:PFACtor command.
Description
- This command sets or queries the value of power factor for the harmonics measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:PFACtor?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:PFACtor?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:PFACtor valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:PFACtor <NR1>
- POWer:POWer<x>:HARMONICS:PFACtor?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR1>ranges from 0 to 1.
PowerPowerItemHarmonicsPowerrating
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:HARMONICS:POWERRating command.
Description
- This command sets or queries the power level for the harmonics measurement of the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:POWERRating?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:POWERRating?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:POWERRating valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:POWERRating {HIGH|LOW}
- POWer:POWer<x>:HARMONICS:POWERRating?
PowerPowerItemHarmonicsRcurrent
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:HARMONICS:RCURRent command.
Description
- This command sets or queries the rated current for the harmonics measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:RCURRent?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:RCURRent?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:RCURRent valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:RCURRent <NR1>
- POWer:POWer<x>:HARMONICS:RCURRent?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR1>ranges from 0 to 100.
PowerPowerItemHarmonicsStandard
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:HARMONICS:STANDard command.
Description
- This command sets or queries the test mode for harmonics measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:STANDard?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:STANDard?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:STANDard valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:STANDard {NONe|IEC|MIL|AM14|DO160|CUSTOM}
- POWer:POWer<x>:HARMONICS:STANDard?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.NONe= No standard.IEC= IEC 61000-3-2 standard.MIL= MIL-STD-1399 standard.AM14= AM14 standard.DO160= DO160 standard.CUSTOM= CUSTOM standard.
PowerPowerItemHarmonicsStartfrequency
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:HARMONICS:STARTFREQUEncy command.
Description
- This command sets or queries the value for the start frequency for the Harmonics measurement. in the range of 1 Hz to 1 GHz.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:STARTFREQUEncy?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:STARTFREQUEncy?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:STARTFREQUEncy valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:STARTFREQUEncy <NR2>
- POWer:POWer<x>:HARMONICS:STARTFREQUEncy?
Info
POWer<x>is the Power measurement.<NR2>sets the starting frequency, in hertz.
PowerPowerItemHarmonicsUnits
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:HARMONICS:UNITs command.
Description
- This command sets or queries the harmonics results units of the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:UNITs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:UNITs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:UNITs valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:UNITs {LOG|LINear}
- POWer:POWer<x>:HARMONICS:UNITs?
PowerPowerItemHarmonicsVsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:HARMONICS:VSOURce command.
Description
- This command sets or queries the voltage source for SOA measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:HARMONICS:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:HARMONICS:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:HARMONICS:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:HARMONICS:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:HARMONICS:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemImpedance
¶
Bases: SCPICmdRead
The POWer:POWer<x>:IMPEDANCE command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the number of the Impedance power measurement.
Properties
.ampval: ThePOWer:POWer<x>:IMPEDANCE:AMP<x>Valcommand..ampmode: ThePOWer:POWer<x>:IMPEDANCE:AMPModecommand..analysismethod: ThePOWer:POWer<x>:IMPEDANCE:ANALYSISMethodcommand..autorbw: ThePOWer:POWer<x>:IMPEDANCE:AUTORbwcommand..connectstatus: ThePOWer:POWer<x>:IMPEDANCE:CONNECTSTATuscommand..constamplitude: ThePOWer:POWer<x>:IMPEDANCE:CONSTAMPlitudecommand..freqval: ThePOWer:POWer<x>:IMPEDANCE:FREQ<x>Valcommand..genipaddress: ThePOWer:POWer<x>:IMPEDANCE:GENIPADDresscommand..generator: ThePOWer:POWer<x>:IMPEDANCE:GENeratorcommand..impedance: ThePOWer:POWer<x>:IMPEDANCE:IMPEDANCEcommand..inputsource: ThePOWer:POWer<x>:IMPEDANCE:INPUTSOurcecommand..outputsource: ThePOWer:POWer<x>:IMPEDANCE:OUTPUTSOurcecommand..ppd: ThePOWer:POWer<x>:IMPEDANCE:PPDcommand..startfrequency: ThePOWer:POWer<x>:IMPEDANCE:STARTFREQuencycommand..stopfrequency: ThePOWer:POWer<x>:IMPEDANCE:STOPFREQuencycommand..testconnection: ThePOWer:POWer<x>:IMPEDANCE:TESTCONNectioncommand.
ampmode
property
¶
ampmode: PowerPowerItemImpedanceAmpmode
Return the POWer:POWer<x>:IMPEDANCE:AMPMode command.
Description
- Sets or queries the power amplitude mode for the Impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:AMPMode?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:AMPMode?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:AMPMode valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:AMPMode {CONSTant|PROFile}
- POWer:POWer<x>:IMPEDANCE:AMPMode?
Info
POWer<x>is the number of the Impedance power measurement.CONSTantsets the generator to output a constant level signal.PROFileuses related commands to set the generator output signal profile (Start frequency, Stop frequency, and Amplitude for each profile step).
ampval
property
¶
ampval: dict[int, PowerPowerItemImpedanceAmpvalItem]
Return the POWer:POWer<x>:IMPEDANCE:AMP<x>Val command.
Description
- Sets or queries the signal generator amplitude setting of the specified profile step, for the specified Impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:AMP<x>Val?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:AMP<x>Val?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:AMP<x>Val valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:AMP<x>Val <NR3>
- POWer:POWer<x>:IMPEDANCE:AMP<x>Val?
Info
POWer<x>is the number of the Impedance power measurement.AMP<x>specifies the number of the profile step. The valid range is 1 to 10.<NR3>sets the generator output amplitude for the specified profile step, in the range -100 V to 100 V.
analysismethod
property
¶
analysismethod: PowerPowerItemImpedanceAnalysismethod
Return the POWer:POWer<x>:IMPEDANCE:ANALYSISMethod command.
Description
- This command sets or queries the Analysis Method for Impedance measurements.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:ANALYSISMethod?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:ANALYSISMethod?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:ANALYSISMethod valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:ANALYSISMethod {SV|FFT}
- POWer:POWer<x>:IMPEDANCE:ANALYSISMethod?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.SVsets the Analysis Method as Spectrum View.FFTsets the Analysis Method as FFT.
autorbw
property
¶
autorbw: PowerPowerItemImpedanceAutorbw
Return the POWer:POWer<x>:IMPEDANCE:AUTORbw command.
Description
- This command enables Auto RBW computation.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:AUTORbw?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:AUTORbw?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:AUTORbw valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:AUTORbw {True|False}
- POWer:POWer<x>:IMPEDANCE:AUTORbw?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.Trueenables Auto RBW computation.Falsedisables Auto RBW computation.
connectstatus
property
¶
connectstatus: PowerPowerItemImpedanceConnectstatus
Return the POWer:POWer<x>:IMPEDANCE:CONNECTSTATus command.
Description
- Queries the instrument’s connection status to the external generator, for the
specified Impedance measurement. The Impedance measurement generator IP address (for
external generators) is set with
POWer:POWer<x>:IMPEDANCE: GENIPADDress.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:CONNECTSTATus?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:CONNECTSTATus?query and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:CONNECTSTATus?
Info
POWer<x>is the number of the Impedance power measurement.
constamplitude
property
¶
constamplitude: PowerPowerItemImpedanceConstamplitude
Return the POWer:POWer<x>:IMPEDANCE:CONSTAMPlitude command.
Description
- Sets or queries the constant amplitude value for the specified Impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:CONSTAMPlitude?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:CONSTAMPlitude?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:CONSTAMPlitude valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:CONSTAMPlitude <NR3>
- POWer:POWer<x>:IMPEDANCE:CONSTAMPlitude?
Info
POWer<x>is the number of the Impedance power measurement.<NR3>specifies the constant amplitude value, in the range of -100 V to 100 V.
freqval
property
¶
freqval: dict[int, PowerPowerItemImpedanceFreqvalItem]
Return the POWer:POWer<x>:IMPEDANCE:FREQ<x>Val command.
Description
- Sets or queries the signal generator start frequency of the specified profile step, for the specified Impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:FREQ<x>Val?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:FREQ<x>Val?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:FREQ<x>Val valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:FREQ<x>Val <NR3>
- POWer:POWer<x>:IMPEDANCE:FREQ<x>Val?
Info
POWer<x>is the number of the Impedance power measurement.FREQ<x>specifies the number of the profile step. The valid range is 1 to 10.<NR3>sets the start frequency, in the range of 10 Hz to 50 MHz, for the specified profile step. You can only set the starting frequency for each profile step; the stop frequency is automatically set to same value as the start frequency of the next profile step. For example, if Step one is set to 1MHz, and Step two is set to 2MHz, then the Step one stop frequency is 2`MHz.
generator
property
¶
generator: PowerPowerItemImpedanceGenerator
Return the POWer:POWer<x>:IMPEDANCE:GENerator command.
Description
- Sets or queries the generator source for the Impedance power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:GENerator?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:GENerator?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:GENerator valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:GENerator {INTernal|EXTernal}
- POWer:POWer<x>:IMPEDANCE:GENerator?
Info
POWer<x>is the number of the Impedance power measurement.INTernalsets the internal generator as the source for the Impedance power measurement.EXTernalsets the external generator as the source for the Impedance power measurement. Supported external generators are the Tektronix AFG31000 and AFG3000 series.
genipaddress
property
¶
genipaddress: PowerPowerItemImpedanceGenipaddress
Return the POWer:POWer<x>:IMPEDANCE:GENIPADDress command.
Description
- Sets or queries the external generator IP Address associated with the specified Impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:GENIPADDress?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:GENIPADDress?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:GENIPADDress valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:GENIPADDress <QString>
- POWer:POWer<x>:IMPEDANCE:GENIPADDress?
Info
POWer<x>is the number of the Impedance power measurement.<QString>is the IP address of the generator.
impedance
property
¶
impedance: PowerPowerItemImpedanceImpedance
Return the POWer:POWer<x>:IMPEDANCE:IMPEDANCE command.
Description
- Sets or queries the output impedance of the generator for the specified Impedance power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:IMPEDANCE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:IMPEDANCE?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:IMPEDANCE valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:IMPEDANCE {FIFTy|HIGHZ}
- POWer:POWer<x>:IMPEDANCE:IMPEDANCE?
Info
POWer<x>is the number of the Impedance power measurement.FIFTysets the measurement impedance to 50`Ω.HIGHZsets the measurement impedance to 1`MΩ.
inputsource
property
¶
inputsource: PowerPowerItemImpedanceInputsource
Return the POWer:POWer<x>:IMPEDANCE:INPUTSOurce command.
Description
- Sets or queries the source for the Impedance input measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:IMPEDANCE:INPUTSOurce?
Info
POWer<x>is the number of the Impedance power measurement.CH<x>specifies the instrument input channel number.MATH<x>specifies the instrument math waveform number.REF<x>specifies the instrument reference waveform number.
outputsource
property
¶
outputsource: PowerPowerItemImpedanceOutputsource
Return the POWer:POWer<x>:IMPEDANCE:OUTPUTSOurce command.
Description
- Sets or queries the source for the Impedance output measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:OUTPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:OUTPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:OUTPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:OUTPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:IMPEDANCE:OUTPUTSOurce?
Info
POWer<x>is the number of the Impedance power measurement.CH<x>specifies the instrument input channel number.MATH<x>specifies the instrument math waveform number.REF<x>specifies the instrument reference waveform number.
ppd
property
¶
Return the POWer:POWer<x>:IMPEDANCE:PPD command.
Description
- Sets or queries the value for points per decade for the specified Impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:PPD?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:PPD?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:PPD valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:PPD <NR1>
- POWer:POWer<x>:IMPEDANCE:PPD?
Info
POWer<x>is the number of the Impedance power measurement.<NR1>specifies the number of frequency points between the start and stop frequency in terms of log scale, in the range of 10 to 100 points.
startfrequency
property
¶
startfrequency: PowerPowerItemImpedanceStartfrequency
Return the POWer:POWer<x>:IMPEDANCE:STARTFREQuency command.
Description
- Sets or queries the value for the start frequency of the specified Impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:STARTFREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:STARTFREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:STARTFREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:STARTFREQuency <NR3>
- POWer:POWer<x>:IMPEDANCE:STARTFREQuency?
Info
POWer<x>is the number of the Impedance power measurement.<NR3>is a floating point number representing the start frequency, in the range of 10 Hz to 50 MHz.
stopfrequency
property
¶
stopfrequency: PowerPowerItemImpedanceStopfrequency
Return the POWer:POWer<x>:IMPEDANCE:STOPFREQuency command.
Description
- Sets or queries the value of the specified Impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:STOPFREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:STOPFREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:STOPFREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:STOPFREQuency <NR3>
- POWer:POWer<x>:IMPEDANCE:STOPFREQuency?
Info
POWer<x>is the number of the Impedance power measurement.<NR3>specifies the start frequency, in the range of 10 Hz to 50 MHz.
testconnection
property
¶
testconnection: PowerPowerItemImpedanceTestconnection
Return the POWer:POWer<x>:IMPEDANCE:TESTCONNection command.
Description
- This command tests the connection with the external instrument for the specified Impedance measurement.
Usage
- Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:TESTCONNection valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:TESTCONNection {EXECute}
Info
POWer<x>is the number of the Impedance power measurement.EXECuteruns the test connection function.
PowerPowerItemImpedanceAmpmode
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:IMPEDANCE:AMPMode command.
Description
- Sets or queries the power amplitude mode for the Impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:AMPMode?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:AMPMode?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:AMPMode valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:AMPMode {CONSTant|PROFile}
- POWer:POWer<x>:IMPEDANCE:AMPMode?
Info
POWer<x>is the number of the Impedance power measurement.CONSTantsets the generator to output a constant level signal.PROFileuses related commands to set the generator output signal profile (Start frequency, Stop frequency, and Amplitude for each profile step).
PowerPowerItemImpedanceAmpvalItem
¶
Bases: ValidatedDynamicNumberCmd, SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:IMPEDANCE:AMP<x>Val command.
Description
- Sets or queries the signal generator amplitude setting of the specified profile step, for the specified Impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:AMP<x>Val?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:AMP<x>Val?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:AMP<x>Val valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:AMP<x>Val <NR3>
- POWer:POWer<x>:IMPEDANCE:AMP<x>Val?
Info
POWer<x>is the number of the Impedance power measurement.AMP<x>specifies the number of the profile step. The valid range is 1 to 10.<NR3>sets the generator output amplitude for the specified profile step, in the range -100 V to 100 V.
PowerPowerItemImpedanceAnalysismethod
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:IMPEDANCE:ANALYSISMethod command.
Description
- This command sets or queries the Analysis Method for Impedance measurements.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:ANALYSISMethod?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:ANALYSISMethod?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:ANALYSISMethod valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:ANALYSISMethod {SV|FFT}
- POWer:POWer<x>:IMPEDANCE:ANALYSISMethod?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.SVsets the Analysis Method as Spectrum View.FFTsets the Analysis Method as FFT.
PowerPowerItemImpedanceAutorbw
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:IMPEDANCE:AUTORbw command.
Description
- This command enables Auto RBW computation.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:AUTORbw?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:AUTORbw?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:AUTORbw valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:AUTORbw {True|False}
- POWer:POWer<x>:IMPEDANCE:AUTORbw?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.Trueenables Auto RBW computation.Falsedisables Auto RBW computation.
PowerPowerItemImpedanceConnectstatus
¶
Bases: SCPICmdRead
The POWer:POWer<x>:IMPEDANCE:CONNECTSTATus command.
Description
- Queries the instrument’s connection status to the external generator, for the specified
Impedance measurement. The Impedance measurement generator IP address (for external
generators) is set with
POWer:POWer<x>:IMPEDANCE: GENIPADDress.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:CONNECTSTATus?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:CONNECTSTATus?query and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:CONNECTSTATus?
Info
POWer<x>is the number of the Impedance power measurement.
PowerPowerItemImpedanceConstamplitude
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:IMPEDANCE:CONSTAMPlitude command.
Description
- Sets or queries the constant amplitude value for the specified Impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:CONSTAMPlitude?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:CONSTAMPlitude?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:CONSTAMPlitude valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:CONSTAMPlitude <NR3>
- POWer:POWer<x>:IMPEDANCE:CONSTAMPlitude?
Info
POWer<x>is the number of the Impedance power measurement.<NR3>specifies the constant amplitude value, in the range of -100 V to 100 V.
PowerPowerItemImpedanceFreqvalItem
¶
Bases: ValidatedDynamicNumberCmd, SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:IMPEDANCE:FREQ<x>Val command.
Description
- Sets or queries the signal generator start frequency of the specified profile step, for the specified Impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:FREQ<x>Val?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:FREQ<x>Val?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:FREQ<x>Val valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:FREQ<x>Val <NR3>
- POWer:POWer<x>:IMPEDANCE:FREQ<x>Val?
Info
POWer<x>is the number of the Impedance power measurement.FREQ<x>specifies the number of the profile step. The valid range is 1 to 10.<NR3>sets the start frequency, in the range of 10 Hz to 50 MHz, for the specified profile step. You can only set the starting frequency for each profile step; the stop frequency is automatically set to same value as the start frequency of the next profile step. For example, if Step one is set to 1MHz, and Step two is set to 2MHz, then the Step one stop frequency is 2`MHz.
PowerPowerItemImpedanceGenerator
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:IMPEDANCE:GENerator command.
Description
- Sets or queries the generator source for the Impedance power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:GENerator?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:GENerator?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:GENerator valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:GENerator {INTernal|EXTernal}
- POWer:POWer<x>:IMPEDANCE:GENerator?
Info
POWer<x>is the number of the Impedance power measurement.INTernalsets the internal generator as the source for the Impedance power measurement.EXTernalsets the external generator as the source for the Impedance power measurement. Supported external generators are the Tektronix AFG31000 and AFG3000 series.
PowerPowerItemImpedanceGenipaddress
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:IMPEDANCE:GENIPADDress command.
Description
- Sets or queries the external generator IP Address associated with the specified Impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:GENIPADDress?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:GENIPADDress?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:GENIPADDress valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:GENIPADDress <QString>
- POWer:POWer<x>:IMPEDANCE:GENIPADDress?
Info
POWer<x>is the number of the Impedance power measurement.<QString>is the IP address of the generator.
PowerPowerItemImpedanceImpedance
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:IMPEDANCE:IMPEDANCE command.
Description
- Sets or queries the output impedance of the generator for the specified Impedance power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:IMPEDANCE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:IMPEDANCE?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:IMPEDANCE valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:IMPEDANCE {FIFTy|HIGHZ}
- POWer:POWer<x>:IMPEDANCE:IMPEDANCE?
Info
POWer<x>is the number of the Impedance power measurement.FIFTysets the measurement impedance to 50`Ω.HIGHZsets the measurement impedance to 1`MΩ.
PowerPowerItemImpedanceInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:IMPEDANCE:INPUTSOurce command.
Description
- Sets or queries the source for the Impedance input measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:IMPEDANCE:INPUTSOurce?
Info
POWer<x>is the number of the Impedance power measurement.CH<x>specifies the instrument input channel number.MATH<x>specifies the instrument math waveform number.REF<x>specifies the instrument reference waveform number.
PowerPowerItemImpedanceOutputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:IMPEDANCE:OUTPUTSOurce command.
Description
- Sets or queries the source for the Impedance output measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:OUTPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:OUTPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:OUTPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:OUTPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:IMPEDANCE:OUTPUTSOurce?
Info
POWer<x>is the number of the Impedance power measurement.CH<x>specifies the instrument input channel number.MATH<x>specifies the instrument math waveform number.REF<x>specifies the instrument reference waveform number.
PowerPowerItemImpedancePpd
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:IMPEDANCE:PPD command.
Description
- Sets or queries the value for points per decade for the specified Impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:PPD?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:PPD?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:PPD valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:PPD <NR1>
- POWer:POWer<x>:IMPEDANCE:PPD?
Info
POWer<x>is the number of the Impedance power measurement.<NR1>specifies the number of frequency points between the start and stop frequency in terms of log scale, in the range of 10 to 100 points.
PowerPowerItemImpedanceStartfrequency
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:IMPEDANCE:STARTFREQuency command.
Description
- Sets or queries the value for the start frequency of the specified Impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:STARTFREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:STARTFREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:STARTFREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:STARTFREQuency <NR3>
- POWer:POWer<x>:IMPEDANCE:STARTFREQuency?
Info
POWer<x>is the number of the Impedance power measurement.<NR3>is a floating point number representing the start frequency, in the range of 10 Hz to 50 MHz.
PowerPowerItemImpedanceStopfrequency
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:IMPEDANCE:STOPFREQuency command.
Description
- Sets or queries the value of the specified Impedance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IMPEDANCE:STOPFREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IMPEDANCE:STOPFREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:STOPFREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:STOPFREQuency <NR3>
- POWer:POWer<x>:IMPEDANCE:STOPFREQuency?
Info
POWer<x>is the number of the Impedance power measurement.<NR3>specifies the start frequency, in the range of 10 Hz to 50 MHz.
PowerPowerItemImpedanceTestconnection
¶
Bases: SCPICmdWrite
The POWer:POWer<x>:IMPEDANCE:TESTCONNection command.
Description
- This command tests the connection with the external instrument for the specified Impedance measurement.
Usage
- Using the
.write(value)method will send thePOWer:POWer<x>:IMPEDANCE:TESTCONNection valuecommand.
SCPI Syntax
- POWer:POWer<x>:IMPEDANCE:TESTCONNection {EXECute}
Info
POWer<x>is the number of the Impedance power measurement.EXECuteruns the test connection function.
PowerPowerItemInductance
¶
Bases: SCPICmdRead
The POWer:POWer<x>:INDUCTANCE command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INDUCTANCE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INDUCTANCE?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Properties
.edgesource: ThePOWer:POWer<x>:INDUCTANCE:EDGESourcecommand..isource: ThePOWer:POWer<x>:INDUCTANCE:ISOUrcecommand..vsource: ThePOWer:POWer<x>:INDUCTANCE:VSOURcecommand.
edgesource
property
¶
edgesource: PowerPowerItemInductanceEdgesource
Return the POWer:POWer<x>:INDUCTANCE:EDGESource command.
Description
- This command sets or queries the edge source for the power inductance measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INDUCTANCE:EDGESource?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INDUCTANCE:EDGESource?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:INDUCTANCE:EDGESource valuecommand.
SCPI Syntax
- POWer:POWer<x>:INDUCTANCE:EDGESource {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:INDUCTANCE:EDGESource?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
isource
property
¶
isource: PowerPowerItemInductanceIsource
Return the POWer:POWer<x>:INDUCTANCE:ISOUrce command.
Description
- This command sets or queries the current signal source for the inductance measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INDUCTANCE:ISOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INDUCTANCE:ISOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:INDUCTANCE:ISOUrcecommand.
SCPI Syntax
- POWer:POWer<x>:INDUCTANCE:ISOUrce
- POWer:POWer<x>:INDUCTANCE:ISOUrce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
vsource
property
¶
vsource: PowerPowerItemInductanceVsource
Return the POWer:POWer<x>:INDUCTANCE:VSOURce command.
Description
- This command sets or queries the voltage source for inductance measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INDUCTANCE:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INDUCTANCE:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:INDUCTANCE:VSOURcecommand.
SCPI Syntax
- POWer:POWer<x>:INDUCTANCE:VSOURce
- POWer:POWer<x>:INDUCTANCE:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemInductanceEdgesource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:INDUCTANCE:EDGESource command.
Description
- This command sets or queries the edge source for the power inductance measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INDUCTANCE:EDGESource?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INDUCTANCE:EDGESource?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:INDUCTANCE:EDGESource valuecommand.
SCPI Syntax
- POWer:POWer<x>:INDUCTANCE:EDGESource {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:INDUCTANCE:EDGESource?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemInductanceIsource
¶
Bases: SCPICmdWriteNoArguments, SCPICmdRead
The POWer:POWer<x>:INDUCTANCE:ISOUrce command.
Description
- This command sets or queries the current signal source for the inductance measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INDUCTANCE:ISOUrce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INDUCTANCE:ISOUrce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:INDUCTANCE:ISOUrcecommand.
SCPI Syntax
- POWer:POWer<x>:INDUCTANCE:ISOUrce
- POWer:POWer<x>:INDUCTANCE:ISOUrce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemInductanceVsource
¶
Bases: SCPICmdWriteNoArguments, SCPICmdRead
The POWer:POWer<x>:INDUCTANCE:VSOURce command.
Description
- This command sets or queries the voltage source for inductance measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INDUCTANCE:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INDUCTANCE:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:INDUCTANCE:VSOURcecommand.
SCPI Syntax
- POWer:POWer<x>:INDUCTANCE:VSOURce
- POWer:POWer<x>:INDUCTANCE:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemInputcap
¶
Bases: SCPICmdRead
The POWer:POWer<x>:INPUTCAP command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INPUTCAP?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INPUTCAP?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Properties
.isource: ThePOWer:POWer<x>:INPUTCAP:ISOURcecommand..peakcurrent: ThePOWer:POWer<x>:INPUTCAP:PEAKCURRentcommand..peakvoltage: ThePOWer:POWer<x>:INPUTCAP:PEAKVOLTagecommand..vsource: ThePOWer:POWer<x>:INPUTCAP:VSOURcecommand.
isource
property
¶
isource: PowerPowerItemInputcapIsource
Return the POWer:POWer<x>:INPUTCAP:ISOURce command.
Description
- This command sets or queries the inrush current input source of the specified Input Capacitance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INPUTCAP:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INPUTCAP:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:INPUTCAP:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:INPUTCAP:ISOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:INPUTCAP:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>is the channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>is the Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>is the Math waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
peakcurrent
property
¶
peakcurrent: PowerPowerItemInputcapPeakcurrent
Return the POWer:POWer<x>:INPUTCAP:PEAKCURRent command.
Description
- This command sets or queries the peak current value of the specified Input Capacitance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INPUTCAP:PEAKCURRent?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INPUTCAP:PEAKCURRent?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:INPUTCAP:PEAKCURRent valuecommand.
SCPI Syntax
- POWer:POWer<x>:INPUTCAP:PEAKCURRent <NR3>
- POWer:POWer<x>:INPUTCAP:PEAKCURRent?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the peak current value, in amps, in the range -100 A to 100 A.
peakvoltage
property
¶
peakvoltage: PowerPowerItemInputcapPeakvoltage
Return the POWer:POWer<x>:INPUTCAP:PEAKVOLTage command.
Description
- This command sets or queries the peak voltage value of the specified Input Capacitance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INPUTCAP:PEAKVOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INPUTCAP:PEAKVOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:INPUTCAP:PEAKVOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:INPUTCAP:PEAKVOLTage <NR3>
- POWer:POWer<x>:INPUTCAP:PEAKVOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the peak voltage value in the range -100 V to 100 V.
vsource
property
¶
vsource: PowerPowerItemInputcapVsource
Return the POWer:POWer<x>:INPUTCAP:VSOURce command.
Description
- This command sets or queries the input voltage source of the specified Input Capacitance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INPUTCAP:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INPUTCAP:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:INPUTCAP:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:INPUTCAP:VSOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:INPUTCAP:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>is the channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>is the Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>is the Math waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
PowerPowerItemInputcapIsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:INPUTCAP:ISOURce command.
Description
- This command sets or queries the inrush current input source of the specified Input Capacitance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INPUTCAP:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INPUTCAP:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:INPUTCAP:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:INPUTCAP:ISOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:INPUTCAP:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>is the channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>is the Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>is the Math waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
PowerPowerItemInputcapPeakcurrent
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:INPUTCAP:PEAKCURRent command.
Description
- This command sets or queries the peak current value of the specified Input Capacitance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INPUTCAP:PEAKCURRent?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INPUTCAP:PEAKCURRent?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:INPUTCAP:PEAKCURRent valuecommand.
SCPI Syntax
- POWer:POWer<x>:INPUTCAP:PEAKCURRent <NR3>
- POWer:POWer<x>:INPUTCAP:PEAKCURRent?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the peak current value, in amps, in the range -100 A to 100 A.
PowerPowerItemInputcapPeakvoltage
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:INPUTCAP:PEAKVOLTage command.
Description
- This command sets or queries the peak voltage value of the specified Input Capacitance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INPUTCAP:PEAKVOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INPUTCAP:PEAKVOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:INPUTCAP:PEAKVOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:INPUTCAP:PEAKVOLTage <NR3>
- POWer:POWer<x>:INPUTCAP:PEAKVOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the peak voltage value in the range -100 V to 100 V.
PowerPowerItemInputcapVsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:INPUTCAP:VSOURce command.
Description
- This command sets or queries the input voltage source of the specified Input Capacitance measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INPUTCAP:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INPUTCAP:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:INPUTCAP:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:INPUTCAP:VSOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:INPUTCAP:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>is the channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>is the Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>is the Math waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
PowerPowerItemInrushcurrent
¶
Bases: SCPICmdRead
The POWer:POWer<x>:INRUSHcurrent command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INRUSHcurrent?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INRUSHcurrent?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Properties
.inputsource: ThePOWer:POWer<x>:INRUSHcurrent:INPUTSOurcecommand..peakcurrent: ThePOWer:POWer<x>:INRUSHcurrent:PEAKCURRentcommand.
inputsource
property
¶
inputsource: PowerPowerItemInrushcurrentInputsource
Return the POWer:POWer<x>:INRUSHcurrent:INPUTSOurce command.
Description
- This command sets or returns the input source of the specified Inrush Current measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INRUSHcurrent:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INRUSHcurrent:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:INRUSHcurrent:INPUTSOurcecommand.
SCPI Syntax
- POWer:POWer<x>:INRUSHcurrent:INPUTSOurce
- POWer:POWer<x>:INRUSHcurrent:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>is the channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>is the Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>is the Math waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
peakcurrent
property
¶
peakcurrent: PowerPowerItemInrushcurrentPeakcurrent
Return the POWer:POWer<x>:INRUSHcurrent:PEAKCURRent command.
Description
- This command sets or returns the peak current value of the specified Inrush Current measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INRUSHcurrent:PEAKCURRent?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INRUSHcurrent:PEAKCURRent?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:INRUSHcurrent:PEAKCURRent valuecommand.
SCPI Syntax
- POWer:POWer<x>:INRUSHcurrent:PEAKCURRent <NR3>
- POWer:POWer<x>:INRUSHcurrent:PEAKCURRent?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the peak current value, in amps, in the range -100 A to 100 A.
PowerPowerItemInrushcurrentInputsource
¶
Bases: SCPICmdWriteNoArguments, SCPICmdRead
The POWer:POWer<x>:INRUSHcurrent:INPUTSOurce command.
Description
- This command sets or returns the input source of the specified Inrush Current measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INRUSHcurrent:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INRUSHcurrent:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:INRUSHcurrent:INPUTSOurcecommand.
SCPI Syntax
- POWer:POWer<x>:INRUSHcurrent:INPUTSOurce
- POWer:POWer<x>:INRUSHcurrent:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>is the channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>is the Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>is the Math waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
PowerPowerItemInrushcurrentPeakcurrent
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:INRUSHcurrent:PEAKCURRent command.
Description
- This command sets or returns the peak current value of the specified Inrush Current measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:INRUSHcurrent:PEAKCURRent?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:INRUSHcurrent:PEAKCURRent?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:INRUSHcurrent:PEAKCURRent valuecommand.
SCPI Syntax
- POWer:POWer<x>:INRUSHcurrent:PEAKCURRent <NR3>
- POWer:POWer<x>:INRUSHcurrent:PEAKCURRent?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the peak current value, in amps, in the range -100 A to 100 A.
PowerPowerItemIvsintegralv
¶
Bases: SCPICmdRead
The POWer:POWer<x>:IVSINTEGRALV command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IVSINTEGRALV?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IVSINTEGRALV?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Properties
.isource: ThePOWer:POWer<x>:IVSINTEGRALV:ISOURcecommand..vsource: ThePOWer:POWer<x>:IVSINTEGRALV:VSOURcecommand.
isource
property
¶
isource: PowerPowerItemIvsintegralvIsource
Return the POWer:POWer<x>:IVSINTEGRALV:ISOURce command.
Description
- This command sets or queries the current source for I vs Integral V measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IVSINTEGRALV:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IVSINTEGRALV:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IVSINTEGRALV:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:IVSINTEGRALV:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:IVSINTEGRALV:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
vsource
property
¶
vsource: PowerPowerItemIvsintegralvVsource
Return the POWer:POWer<x>:IVSINTEGRALV:VSOURce command.
Description
- This command sets or queries the voltage source for I vs Integral V measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IVSINTEGRALV:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IVSINTEGRALV:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IVSINTEGRALV:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:IVSINTEGRALV:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:IVSINTEGRALV:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemIvsintegralvIsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:IVSINTEGRALV:ISOURce command.
Description
- This command sets or queries the current source for I vs Integral V measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IVSINTEGRALV:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IVSINTEGRALV:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IVSINTEGRALV:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:IVSINTEGRALV:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:IVSINTEGRALV:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemIvsintegralvVsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:IVSINTEGRALV:VSOURce command.
Description
- This command sets or queries the voltage source for I vs Integral V measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:IVSINTEGRALV:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:IVSINTEGRALV:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:IVSINTEGRALV:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:IVSINTEGRALV:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:IVSINTEGRALV:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemLabel
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:LABel command.
Description
- This command sets or queries the label for the specified power measurement. As the label can contain non 7-bit ASCII text, it is stored in Percent Encoding format. The power measurement badge is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:LABel?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:LABel?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:LABel valuecommand.
SCPI Syntax
- POWer:POWer<x>:LABel <QString>
- POWer:POWer<x>:LABel?
PowerPowerItemLineripple
¶
Bases: SCPICmdRead
The POWer:POWer<x>:LINERIPPLE command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:LINERIPPLE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:LINERIPPLE?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Properties
.inputsource: ThePOWer:POWer<x>:LINERIPPLE:INPUTSOurcecommand..lfrequency: ThePOWer:POWer<x>:LINERIPPLE:LFREQuencycommand.
inputsource
property
¶
inputsource: PowerPowerItemLinerippleInputsource
Return the POWer:POWer<x>:LINERIPPLE:INPUTSOurce command.
Description
- This command sets or queries the input source for line ripple measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:LINERIPPLE:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:LINERIPPLE:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:LINERIPPLE:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:LINERIPPLE:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:LINERIPPLE:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
lfrequency
property
¶
lfrequency: PowerPowerItemLinerippleLfrequency
Return the POWer:POWer<x>:LINERIPPLE:LFREQuency command.
Description
- This command sets or queries the frequency present for line ripple measurement of the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:LINERIPPLE:LFREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:LINERIPPLE:LFREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:LINERIPPLE:LFREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:LINERIPPLE:LFREQuency {FIFty|SIXty|FOURHundred}
- POWer:POWer<x>:LINERIPPLE:LFREQuency?
PowerPowerItemLinerippleInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:LINERIPPLE:INPUTSOurce command.
Description
- This command sets or queries the input source for line ripple measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:LINERIPPLE:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:LINERIPPLE:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:LINERIPPLE:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:LINERIPPLE:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:LINERIPPLE:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemLinerippleLfrequency
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:LINERIPPLE:LFREQuency command.
Description
- This command sets or queries the frequency present for line ripple measurement of the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:LINERIPPLE:LFREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:LINERIPPLE:LFREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:LINERIPPLE:LFREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:LINERIPPLE:LFREQuency {FIFty|SIXty|FOURHundred}
- POWer:POWer<x>:LINERIPPLE:LFREQuency?
PowerPowerItemMagneticloss
¶
Bases: SCPICmdRead
The POWer:POWer<x>:MAGNETICLOSS command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGNETICLOSS?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGNETICLOSS?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Properties
.isource: ThePOWer:POWer<x>:MAGNETICLOSS:ISOURcecommand..vsource: ThePOWer:POWer<x>:MAGNETICLOSS:VSOURcecommand.
isource
property
¶
isource: PowerPowerItemMagneticlossIsource
Return the POWer:POWer<x>:MAGNETICLOSS:ISOURce command.
Description
- This command sets or queries the current source for the magnetic loss measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGNETICLOSS:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGNETICLOSS:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGNETICLOSS:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGNETICLOSS:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGNETICLOSS:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
vsource
property
¶
vsource: PowerPowerItemMagneticlossVsource
Return the POWer:POWer<x>:MAGNETICLOSS:VSOURce command.
Description
- This command sets or queries the voltage source for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGNETICLOSS:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGNETICLOSS:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGNETICLOSS:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGNETICLOSS:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGNETICLOSS:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemMagneticlossIsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGNETICLOSS:ISOURce command.
Description
- This command sets or queries the current source for the magnetic loss measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGNETICLOSS:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGNETICLOSS:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGNETICLOSS:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGNETICLOSS:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGNETICLOSS:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemMagneticlossVsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGNETICLOSS:VSOURce command.
Description
- This command sets or queries the voltage source for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGNETICLOSS:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGNETICLOSS:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGNETICLOSS:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGNETICLOSS:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGNETICLOSS:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemMagproperty
¶
Bases: SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the magnetic property power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Properties
.areaofcrosssection: ThePOWer:POWer<x>:MAGPROPERTY:AREAofcrosssectioncommand..edgesource: ThePOWer:POWer<x>:MAGPROPERTY:EDGESOURcecommand..isource: ThePOWer:POWer<x>:MAGPROPERTY:ISOURcecommand..length: ThePOWer:POWer<x>:MAGPROPERTY:LENgthcommand..primaryturns: ThePOWer:POWer<x>:MAGPROPERTY:PRIMARYTURNscommand..sec1source: ThePOWer:POWer<x>:MAGPROPERTY:SEC1SOURcecommand..sec1turns: ThePOWer:POWer<x>:MAGPROPERTY:SEC1TURNscommand..sec2source: ThePOWer:POWer<x>:MAGPROPERTY:SEC2SOURcecommand..sec2turns: ThePOWer:POWer<x>:MAGPROPERTY:SEC2TURNscommand..sec3source: ThePOWer:POWer<x>:MAGPROPERTY:SEC3SOURcecommand..sec3turns: ThePOWer:POWer<x>:MAGPROPERTY:SEC3TURNscommand..sec4source: ThePOWer:POWer<x>:MAGPROPERTY:SEC4SOURcecommand..sec4turns: ThePOWer:POWer<x>:MAGPROPERTY:SEC4TURNscommand..sec5source: ThePOWer:POWer<x>:MAGPROPERTY:SEC5SOURcecommand..sec5turns: ThePOWer:POWer<x>:MAGPROPERTY:SEC5TURNscommand..sec6source: ThePOWer:POWer<x>:MAGPROPERTY:SEC6SOURcecommand..sec6turns: ThePOWer:POWer<x>:MAGPROPERTY:SEC6TURNscommand..secphase: ThePOWer:POWer<x>:MAGPROPERTY:SECPhasecommand..secvolt: ThePOWer:POWer<x>:MAGPROPERTY:SECVoltcommand..secwindings: ThePOWer:POWer<x>:MAGPROPERTY:SECWINDingscommand..units: ThePOWer:POWer<x>:MAGPROPERTY:UNITscommand..vsource: ThePOWer:POWer<x>:MAGPROPERTY:VSOURcecommand.
areaofcrosssection
property
¶
areaofcrosssection: PowerPowerItemMagpropertyAreaofcrosssection
Return the POWer:POWer<x>:MAGPROPERTY:AREAofcrosssection command.
Description
- This command sets or queries the coil cross section area for magnetic measurement of
the specified power measurement number. Note: The coil Cross Section Area unit depends
on the setting of the Units value (
seePOWer:POWer<x>:MAGPROPERTY:UNITs). The command returns the same numeric value regardless of the units setting.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:AREAofcrosssection?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:AREAofcrosssection?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:AREAofcrosssection valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:AREAofcrosssection <NR2>
- POWer:POWer<x>:MAGPROPERTY:AREAofcrosssection?
Info
POWer<x>is the magnetic property power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR2>is the cross section area in square meters, in the range of 1 nanometer.
edgesource
property
¶
edgesource: PowerPowerItemMagpropertyEdgesource
Return the POWer:POWer<x>:MAGPROPERTY:EDGESOURce command.
Description
- This command sets or queries the edge source type for the magnetic property measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:EDGESOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:EDGESOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:EDGESOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:EDGESOURce {Current|VOLTAGE}
- POWer:POWer<x>:MAGPROPERTY:EDGESOURce?
Info
POWer<x>is the magnetic property power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.Currentsets the measurement to use the primary voltage source as the signal edge for the magnetic property measurement.VOLTAGEsets the measurement to use the primary current source as the signal edge for the magnetic property measurement.
isource
property
¶
isource: PowerPowerItemMagpropertyIsource
Return the POWer:POWer<x>:MAGPROPERTY:ISOURce command.
Description
- This command sets or queries the current source for the magnetic property measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
length
property
¶
length: PowerPowerItemMagpropertyLength
Return the POWer:POWer<x>:MAGPROPERTY:LENgth command.
Description
- This command sets or queries the conductor length of the primary winding for magnetic
measurement of the specified power measurement number. Note: The Magnetic Length unit
depends on the setting of the Units value (
seePOWer:POWer<x>:MAGPROPERTY:UNITs). The command returns the same numeric value regardless of the units setting.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:LENgth?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:LENgth?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:LENgth valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:LENgth <NR2>
- POWer:POWer<x>:MAGPROPERTY:LENgth?
Info
POWer<x>is the magnetic property power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR2>is the magnetic length, in the range of 1.00E-09 through 1,000,000.
primaryturns
property
¶
primaryturns: PowerPowerItemMagpropertyPrimaryturns
Return the POWer:POWer<x>:MAGPROPERTY:PRIMARYTURNs command.
Description
- This command sets or queries the number of primary turns for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:PRIMARYTURNs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:PRIMARYTURNs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:PRIMARYTURNs valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:PRIMARYTURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:PRIMARYTURNs?
Info
POWer<x>is the magnetic property power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR1>is the integer number of turns in the primary winding, in the range of 1 to 1 M.
sec1source
property
¶
sec1source: PowerPowerItemMagpropertySec1source
Return the POWer:POWer<x>:MAGPROPERTY:SEC1SOURce command.
Description
- This command sets or queries the current source channel for secondary winding 1 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC1SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC1SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC1SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC1SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC1SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
sec1turns
property
¶
sec1turns: PowerPowerItemMagpropertySec1turns
Return the POWer:POWer<x>:MAGPROPERTY:SEC1TURNs command.
Description
- This command sets or queries the number of turns of secondary winding 1 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC1TURNs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC1TURNs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC1TURNs valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC1TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC1TURNs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR1>is the number of turns on the secondary winding, and ranges from 0 to 1,000,000.
sec2source
property
¶
sec2source: PowerPowerItemMagpropertySec2source
Return the POWer:POWer<x>:MAGPROPERTY:SEC2SOURce command.
Description
- This command sets or queries the current source for secondary winding2 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC2SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC2SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC2SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC2SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC2SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
sec2turns
property
¶
sec2turns: PowerPowerItemMagpropertySec2turns
Return the POWer:POWer<x>:MAGPROPERTY:SEC2TURNs command.
Description
- This command sets or queries the number of turns of secondary winding 2 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC2TURNs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC2TURNs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC2TURNs valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC2TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC2TURNs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR1>is the number of turns on the secondary winding, and ranges from 0 to 1,000,000.
sec3source
property
¶
sec3source: PowerPowerItemMagpropertySec3source
Return the POWer:POWer<x>:MAGPROPERTY:SEC3SOURce command.
Description
- This command sets or queries the current source channel for secondary winding 3 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC3SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC3SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC3SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC3SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC3SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
sec3turns
property
¶
sec3turns: PowerPowerItemMagpropertySec3turns
Return the POWer:POWer<x>:MAGPROPERTY:SEC3TURNs command.
Description
- This command sets or queries the number of turns of secondary winding 3 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC3TURNs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC3TURNs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC3TURNs valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC3TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC3TURNs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR1>is the number of turns on the secondary winding, and ranges from 0 to 1,000,000.
sec4source
property
¶
sec4source: PowerPowerItemMagpropertySec4source
Return the POWer:POWer<x>:MAGPROPERTY:SEC4SOURce command.
Description
- This command sets or queries the current source for secondary winding 4 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC4SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC4SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC4SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC4SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC4SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
sec4turns
property
¶
sec4turns: PowerPowerItemMagpropertySec4turns
Return the POWer:POWer<x>:MAGPROPERTY:SEC4TURNs command.
Description
- This command sets or queries the number of turns of secondary winding 4 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC4TURNs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC4TURNs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC4TURNs valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC4TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC4TURNs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR1>is the number of turns on the secondary winding, and ranges from 0 to 1,000,000.
sec5source
property
¶
sec5source: PowerPowerItemMagpropertySec5source
Return the POWer:POWer<x>:MAGPROPERTY:SEC5SOURce command.
Description
- This command sets or queries the current source for secondary winding 5 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC5SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC5SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC5SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC5SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC5SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
sec5turns
property
¶
sec5turns: PowerPowerItemMagpropertySec5turns
Return the POWer:POWer<x>:MAGPROPERTY:SEC5TURNs command.
Description
- This command sets or queries the number of turns of secondary winding 5 for magnetic measurement of the specified power measurement badge.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC5TURNs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC5TURNs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC5TURNs valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC5TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC5TURNs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR1>is the number of turns on the secondary winding, and ranges from 0 to 1,000,000.
sec6source
property
¶
sec6source: PowerPowerItemMagpropertySec6source
Return the POWer:POWer<x>:MAGPROPERTY:SEC6SOURce command.
Description
- This command sets or queries the current source for secondary winding 6 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC6SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC6SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC6SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC6SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC6SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
sec6turns
property
¶
sec6turns: PowerPowerItemMagpropertySec6turns
Return the POWer:POWer<x>:MAGPROPERTY:SEC6TURNs command.
Description
- This command sets or queries the number of turns of secondary winding 6 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC6TURNs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC6TURNs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC6TURNs valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC6TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC6TURNs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR1>is the number of turns on the secondary winding, and ranges from 0 to 1,000,000.
secphase
property
¶
secphase: PowerPowerItemMagpropertySecphase
Return the POWer:POWer<x>:MAGPROPERTY:SECPhase command.
Description
- This command sets or returns the value for the phase difference between secondary and primary voltage.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SECPhase?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SECPhase?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SECPhase valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SECPhase <NR3>
- POWer:POWer<x>:MAGPROPERTY:SECPhase?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>sets the value for the phase difference between secondary and primary voltage, in the range of -180 to 180 degrees.
secvolt
property
¶
secvolt: PowerPowerItemMagpropertySecvolt
Return the POWer:POWer<x>:MAGPROPERTY:SECVolt command.
Description
- This command enables or disables secondary voltage input for measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SECVolt?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SECVolt?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SECVolt valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SECVolt {True|False}
- POWer:POWer<x>:MAGPROPERTY:SECVolt?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.Trueenables secondary voltage source.Falsedisables secondary voltage source.
secwindings
property
¶
secwindings: PowerPowerItemMagpropertySecwindings
Return the POWer:POWer<x>:MAGPROPERTY:SECWINDings command.
Description
- This command sets or queries the number of secondary windings for the magnetic property measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SECWINDings?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SECWINDings?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SECWINDings valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SECWINDings {None|ONE|TWO|THREE|FOUR|FIVE|SIX}
- POWer:POWer<x>:MAGPROPERTY:SECWINDings?
Info
POWer<x>is the magnetic property power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.None, ONE, TWO, THREE, FOUR, FIVE, SIX sets the number of secondary windings to the specified value.
units
property
¶
Return the POWer:POWer<x>:MAGPROPERTY:UNITs command.
Description
- This command sets or queries the units for magnetic measurements of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:UNITs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:UNITs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:UNITs valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:UNITs {SI|CGS}
- POWer:POWer<x>:MAGPROPERTY:UNITs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.SIsets the measurement to International System of Units.CGSsets the measurement to Gaussian units.
vsource
property
¶
vsource: PowerPowerItemMagpropertyVsource
Return the POWer:POWer<x>:MAGPROPERTY:VSOURce command.
Description
- This command sets or queries the primary winding voltage source for the magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemMagpropertyAreaofcrosssection
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:AREAofcrosssection command.
Description
- This command sets or queries the coil cross section area for magnetic measurement of the
specified power measurement number. Note: The coil Cross Section Area unit depends on the
setting of the Units value (
seePOWer:POWer<x>:MAGPROPERTY:UNITs). The command returns the same numeric value regardless of the units setting.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:AREAofcrosssection?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:AREAofcrosssection?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:AREAofcrosssection valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:AREAofcrosssection <NR2>
- POWer:POWer<x>:MAGPROPERTY:AREAofcrosssection?
Info
POWer<x>is the magnetic property power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR2>is the cross section area in square meters, in the range of 1 nanometer.
PowerPowerItemMagpropertyEdgesource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:EDGESOURce command.
Description
- This command sets or queries the edge source type for the magnetic property measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:EDGESOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:EDGESOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:EDGESOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:EDGESOURce {Current|VOLTAGE}
- POWer:POWer<x>:MAGPROPERTY:EDGESOURce?
Info
POWer<x>is the magnetic property power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.Currentsets the measurement to use the primary voltage source as the signal edge for the magnetic property measurement.VOLTAGEsets the measurement to use the primary current source as the signal edge for the magnetic property measurement.
PowerPowerItemMagpropertyIsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:ISOURce command.
Description
- This command sets or queries the current source for the magnetic property measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemMagpropertyLength
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:LENgth command.
Description
- This command sets or queries the conductor length of the primary winding for magnetic
measurement of the specified power measurement number. Note: The Magnetic Length unit
depends on the setting of the Units value (
seePOWer:POWer<x>:MAGPROPERTY:UNITs). The command returns the same numeric value regardless of the units setting.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:LENgth?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:LENgth?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:LENgth valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:LENgth <NR2>
- POWer:POWer<x>:MAGPROPERTY:LENgth?
Info
POWer<x>is the magnetic property power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR2>is the magnetic length, in the range of 1.00E-09 through 1,000,000.
PowerPowerItemMagpropertyPrimaryturns
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:PRIMARYTURNs command.
Description
- This command sets or queries the number of primary turns for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:PRIMARYTURNs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:PRIMARYTURNs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:PRIMARYTURNs valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:PRIMARYTURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:PRIMARYTURNs?
Info
POWer<x>is the magnetic property power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR1>is the integer number of turns in the primary winding, in the range of 1 to 1 M.
PowerPowerItemMagpropertySec1source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:SEC1SOURce command.
Description
- This command sets or queries the current source channel for secondary winding 1 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC1SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC1SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC1SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC1SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC1SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemMagpropertySec1turns
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:SEC1TURNs command.
Description
- This command sets or queries the number of turns of secondary winding 1 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC1TURNs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC1TURNs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC1TURNs valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC1TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC1TURNs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR1>is the number of turns on the secondary winding, and ranges from 0 to 1,000,000.
PowerPowerItemMagpropertySec2source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:SEC2SOURce command.
Description
- This command sets or queries the current source for secondary winding2 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC2SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC2SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC2SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC2SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC2SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemMagpropertySec2turns
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:SEC2TURNs command.
Description
- This command sets or queries the number of turns of secondary winding 2 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC2TURNs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC2TURNs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC2TURNs valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC2TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC2TURNs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR1>is the number of turns on the secondary winding, and ranges from 0 to 1,000,000.
PowerPowerItemMagpropertySec3source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:SEC3SOURce command.
Description
- This command sets or queries the current source channel for secondary winding 3 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC3SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC3SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC3SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC3SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC3SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemMagpropertySec3turns
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:SEC3TURNs command.
Description
- This command sets or queries the number of turns of secondary winding 3 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC3TURNs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC3TURNs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC3TURNs valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC3TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC3TURNs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR1>is the number of turns on the secondary winding, and ranges from 0 to 1,000,000.
PowerPowerItemMagpropertySec4source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:SEC4SOURce command.
Description
- This command sets or queries the current source for secondary winding 4 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC4SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC4SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC4SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC4SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC4SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemMagpropertySec4turns
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:SEC4TURNs command.
Description
- This command sets or queries the number of turns of secondary winding 4 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC4TURNs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC4TURNs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC4TURNs valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC4TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC4TURNs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR1>is the number of turns on the secondary winding, and ranges from 0 to 1,000,000.
PowerPowerItemMagpropertySec5source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:SEC5SOURce command.
Description
- This command sets or queries the current source for secondary winding 5 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC5SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC5SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC5SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC5SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC5SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemMagpropertySec5turns
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:SEC5TURNs command.
Description
- This command sets or queries the number of turns of secondary winding 5 for magnetic measurement of the specified power measurement badge.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC5TURNs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC5TURNs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC5TURNs valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC5TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC5TURNs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR1>is the number of turns on the secondary winding, and ranges from 0 to 1,000,000.
PowerPowerItemMagpropertySec6source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:SEC6SOURce command.
Description
- This command sets or queries the current source for secondary winding 6 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC6SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC6SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC6SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC6SOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:SEC6SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemMagpropertySec6turns
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:SEC6TURNs command.
Description
- This command sets or queries the number of turns of secondary winding 6 for magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SEC6TURNs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC6TURNs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SEC6TURNs valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SEC6TURNs <NR1>
- POWer:POWer<x>:MAGPROPERTY:SEC6TURNs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.<NR1>is the number of turns on the secondary winding, and ranges from 0 to 1,000,000.
PowerPowerItemMagpropertySecphase
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:SECPhase command.
Description
- This command sets or returns the value for the phase difference between secondary and primary voltage.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SECPhase?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SECPhase?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SECPhase valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SECPhase <NR3>
- POWer:POWer<x>:MAGPROPERTY:SECPhase?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>sets the value for the phase difference between secondary and primary voltage, in the range of -180 to 180 degrees.
PowerPowerItemMagpropertySecvolt
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:SECVolt command.
Description
- This command enables or disables secondary voltage input for measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SECVolt?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SECVolt?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SECVolt valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SECVolt {True|False}
- POWer:POWer<x>:MAGPROPERTY:SECVolt?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.Trueenables secondary voltage source.Falsedisables secondary voltage source.
PowerPowerItemMagpropertySecwindings
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:SECWINDings command.
Description
- This command sets or queries the number of secondary windings for the magnetic property measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:SECWINDings?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SECWINDings?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:SECWINDings valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:SECWINDings {None|ONE|TWO|THREE|FOUR|FIVE|SIX}
- POWer:POWer<x>:MAGPROPERTY:SECWINDings?
Info
POWer<x>is the magnetic property power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.None, ONE, TWO, THREE, FOUR, FIVE, SIX sets the number of secondary windings to the specified value.
PowerPowerItemMagpropertyUnits
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:UNITs command.
Description
- This command sets or queries the units for magnetic measurements of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:UNITs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:UNITs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:UNITs valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:UNITs {SI|CGS}
- POWer:POWer<x>:MAGPROPERTY:UNITs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.SIsets the measurement to International System of Units.CGSsets the measurement to Gaussian units.
PowerPowerItemMagpropertyVsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:MAGPROPERTY:VSOURce command.
Description
- This command sets or queries the primary winding voltage source for the magnetic measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:MAGPROPERTY:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:MAGPROPERTY:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:MAGPROPERTY:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:MAGPROPERTY:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:MAGPROPERTY:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemNdutycycle
¶
Bases: SCPICmdRead
The POWer:POWer<x>:NDUTYCYCLE command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:NDUTYCYCLE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:NDUTYCYCLE?query and raise an AssertionError if the returned value does not matchvalue.
Properties
.edgetype: ThePOWer:POWer<x>:NDUTYCYCLE:EDGETypecommand..inputsource: ThePOWer:POWer<x>:NDUTYCYCLE:INPUTSOurcecommand.
edgetype
property
¶
edgetype: PowerPowerItemNdutycycleEdgetype
Return the POWer:POWer<x>:NDUTYCYCLE:EDGEType command.
Description
- This command sets or queries the clock edge type for negative duty cycle measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:NDUTYCYCLE:EDGEType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:NDUTYCYCLE:EDGEType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:NDUTYCYCLE:EDGEType valuecommand.
SCPI Syntax
- POWer:POWer<x>:NDUTYCYCLE:EDGEType {RISE|FALL|BOTH}
- POWer:POWer<x>:NDUTYCYCLE:EDGEType?
inputsource
property
¶
inputsource: PowerPowerItemNdutycycleInputsource
Return the POWer:POWer<x>:NDUTYCYCLE:INPUTSOurce command.
Description
- This command sets or queries the input source for negative duty cycle measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:NDUTYCYCLE:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:NDUTYCYCLE:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:NDUTYCYCLE:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:NDUTYCYCLE:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:NDUTYCYCLE:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemNdutycycleEdgetype
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:NDUTYCYCLE:EDGEType command.
Description
- This command sets or queries the clock edge type for negative duty cycle measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:NDUTYCYCLE:EDGEType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:NDUTYCYCLE:EDGEType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:NDUTYCYCLE:EDGEType valuecommand.
SCPI Syntax
- POWer:POWer<x>:NDUTYCYCLE:EDGEType {RISE|FALL|BOTH}
- POWer:POWer<x>:NDUTYCYCLE:EDGEType?
PowerPowerItemNdutycycleInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:NDUTYCYCLE:INPUTSOurce command.
Description
- This command sets or queries the input source for negative duty cycle measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:NDUTYCYCLE:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:NDUTYCYCLE:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:NDUTYCYCLE:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:NDUTYCYCLE:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:NDUTYCYCLE:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemNpulsewidth
¶
Bases: SCPICmdRead
The POWer:POWer<x>:NPULSEWIDTH command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:NPULSEWIDTH?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:NPULSEWIDTH?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Properties
.inputsource: ThePOWer:POWer<x>:NPULSEWIDTH:INPUTSOurcecommand.
inputsource
property
¶
inputsource: PowerPowerItemNpulsewidthInputsource
Return the POWer:POWer<x>:NPULSEWIDTH:INPUTSOurce command.
Description
- This command sets or queries the input source for negative pulse width measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:NPULSEWIDTH:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:NPULSEWIDTH:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:NPULSEWIDTH:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:NPULSEWIDTH:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:NPULSEWIDTH:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemNpulsewidthInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:NPULSEWIDTH:INPUTSOurce command.
Description
- This command sets or queries the input source for negative pulse width measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:NPULSEWIDTH:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:NPULSEWIDTH:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:NPULSEWIDTH:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:NPULSEWIDTH:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:NPULSEWIDTH:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>= A channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemPdutycycle
¶
Bases: SCPICmdRead
The POWer:POWer<x>:PDUTYCYCLE command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PDUTYCYCLE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PDUTYCYCLE?query and raise an AssertionError if the returned value does not matchvalue.
Properties
.edgetype: ThePOWer:POWer<x>:PDUTYCYCLE:EDGETypecommand..inputsource: ThePOWer:POWer<x>:PDUTYCYCLE:INPUTSOurcecommand.
edgetype
property
¶
edgetype: PowerPowerItemPdutycycleEdgetype
Return the POWer:POWer<x>:PDUTYCYCLE:EDGEType command.
Description
- This command sets or queries the clock edge type for positive duty cycle measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PDUTYCYCLE:EDGEType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PDUTYCYCLE:EDGEType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PDUTYCYCLE:EDGEType valuecommand.
SCPI Syntax
- POWer:POWer<x>:PDUTYCYCLE:EDGEType {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:PDUTYCYCLE:EDGEType?
Info
CH<x>specifies an analog channel to use source.MATH<x>specifies a math waveform to use as the source.REF<x>specifies a reference waveform to use as the source.
inputsource
property
¶
inputsource: PowerPowerItemPdutycycleInputsource
Return the POWer:POWer<x>:PDUTYCYCLE:INPUTSOurce command.
Description
- This command sets or queries the input source for positive duty cycle measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PDUTYCYCLE:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PDUTYCYCLE:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PDUTYCYCLE:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:PDUTYCYCLE:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:PDUTYCYCLE:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemPdutycycleEdgetype
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PDUTYCYCLE:EDGEType command.
Description
- This command sets or queries the clock edge type for positive duty cycle measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PDUTYCYCLE:EDGEType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PDUTYCYCLE:EDGEType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PDUTYCYCLE:EDGEType valuecommand.
SCPI Syntax
- POWer:POWer<x>:PDUTYCYCLE:EDGEType {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:PDUTYCYCLE:EDGEType?
Info
CH<x>specifies an analog channel to use source.MATH<x>specifies a math waveform to use as the source.REF<x>specifies a reference waveform to use as the source.
PowerPowerItemPdutycycleInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PDUTYCYCLE:INPUTSOurce command.
Description
- This command sets or queries the input source for positive duty cycle measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PDUTYCYCLE:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PDUTYCYCLE:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PDUTYCYCLE:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:PDUTYCYCLE:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:PDUTYCYCLE:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemPeriod
¶
Bases: SCPICmdRead
The POWer:POWer<x>:PERIOD command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PERIOD?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PERIOD?query and raise an AssertionError if the returned value does not matchvalue.
Properties
.edge: ThePOWer:POWer<x>:PERIOD:EDGecommand..inputsource: ThePOWer:POWer<x>:PERIOD:INPUTSOurcecommand.
edge
property
¶
edge: PowerPowerItemPeriodEdge
Return the POWer:POWer<x>:PERIOD:EDGe command.
Description
- This command sets or queries the edge type for period measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PERIOD:EDGe?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PERIOD:EDGe?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PERIOD:EDGe valuecommand.
SCPI Syntax
- POWer:POWer<x>:PERIOD:EDGe {RISE|FALL}
- POWer:POWer<x>:PERIOD:EDGe?
inputsource
property
¶
inputsource: PowerPowerItemPeriodInputsource
Return the POWer:POWer<x>:PERIOD:INPUTSOurce command.
Description
- This command sets or queries the input source for period measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PERIOD:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PERIOD:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PERIOD:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:PERIOD:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:PERIOD:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemPeriodEdge
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PERIOD:EDGe command.
Description
- This command sets or queries the edge type for period measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PERIOD:EDGe?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PERIOD:EDGe?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PERIOD:EDGe valuecommand.
SCPI Syntax
- POWer:POWer<x>:PERIOD:EDGe {RISE|FALL}
- POWer:POWer<x>:PERIOD:EDGe?
PowerPowerItemPeriodInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PERIOD:INPUTSOurce command.
Description
- This command sets or queries the input source for period measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PERIOD:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PERIOD:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PERIOD:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:PERIOD:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:PERIOD:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemPowerquality
¶
Bases: SCPICmdRead
The POWer:POWer<x>:POWERQUALITY command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:POWERQUALITY?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:POWERQUALITY?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Properties
.ccycles: ThePOWer:POWer<x>:POWERQUALITY:CCYClescommand..freference: ThePOWer:POWer<x>:POWERQUALITY:FREFerencecommand..isource: ThePOWer:POWer<x>:POWERQUALITY:ISOURcecommand..stype: ThePOWer:POWer<x>:POWERQUALITY:STYPecommand..vsource: ThePOWer:POWer<x>:POWERQUALITY:VSOURcecommand.
ccycles
property
¶
ccycles: PowerPowerItemPowerqualityCcycles
Return the POWer:POWer<x>:POWERQUALITY:CCYCles command.
Description
- This command sets or queries the calculate cycles over full cycles settings for the specified power quality measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:POWERQUALITY:CCYCles?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:POWERQUALITY:CCYCles?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:POWERQUALITY:CCYCles valuecommand.
SCPI Syntax
- POWer:POWer<x>:POWERQUALITY:CCYCles {ON|OFF|1|0}
- POWer:POWer<x>:POWERQUALITY:CCYCles?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.1selects the calculate cycles over full cycles.ONselects the calculate cycles over full cycles.0deselects the calculate cycles over full cycles.OFFdeselects the calculate cycles over full cycles.
freference
property
¶
freference: PowerPowerItemPowerqualityFreference
Return the POWer:POWer<x>:POWERQUALITY:FREFerence command.
Description
- This command sets or queries the frequency reference type for power quality measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:POWERQUALITY:FREFerence?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:POWERQUALITY:FREFerence?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:POWERQUALITY:FREFerence valuecommand.
SCPI Syntax
- POWer:POWer<x>:POWERQUALITY:FREFerence {VOLTage|CURRent}
- POWer:POWer<x>:POWERQUALITY:FREFerence?
isource
property
¶
isource: PowerPowerItemPowerqualityIsource
Return the POWer:POWer<x>:POWERQUALITY:ISOURce command.
Description
- This command sets or queries the current source for power quality measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:POWERQUALITY:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:POWERQUALITY:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:POWERQUALITY:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:POWERQUALITY:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:POWERQUALITY:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
stype
property
¶
Return the POWer:POWer<x>:POWERQUALITY:STYPe command.
Description
- This command sets or queries the source type.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:POWERQUALITY:STYPe?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:POWERQUALITY:STYPe?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:POWERQUALITY:STYPe valuecommand.
SCPI Syntax
- POWer:POWer<x>:POWERQUALITY:STYPe {AC|DC}
- POWer:POWer<x>:POWERQUALITY:STYPe?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.ACsets the signal type as AC.DCsets the signal type as DC.
vsource
property
¶
vsource: PowerPowerItemPowerqualityVsource
Return the POWer:POWer<x>:POWERQUALITY:VSOURce command.
Description
- This command sets or queries the voltage source for power quality measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:POWERQUALITY:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:POWERQUALITY:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:POWERQUALITY:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:POWERQUALITY:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:POWERQUALITY:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemPowerqualityCcycles
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:POWERQUALITY:CCYCles command.
Description
- This command sets or queries the calculate cycles over full cycles settings for the specified power quality measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:POWERQUALITY:CCYCles?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:POWERQUALITY:CCYCles?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:POWERQUALITY:CCYCles valuecommand.
SCPI Syntax
- POWer:POWer<x>:POWERQUALITY:CCYCles {ON|OFF|1|0}
- POWer:POWer<x>:POWERQUALITY:CCYCles?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.1selects the calculate cycles over full cycles.ONselects the calculate cycles over full cycles.0deselects the calculate cycles over full cycles.OFFdeselects the calculate cycles over full cycles.
PowerPowerItemPowerqualityFreference
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:POWERQUALITY:FREFerence command.
Description
- This command sets or queries the frequency reference type for power quality measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:POWERQUALITY:FREFerence?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:POWERQUALITY:FREFerence?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:POWERQUALITY:FREFerence valuecommand.
SCPI Syntax
- POWer:POWer<x>:POWERQUALITY:FREFerence {VOLTage|CURRent}
- POWer:POWer<x>:POWERQUALITY:FREFerence?
PowerPowerItemPowerqualityIsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:POWERQUALITY:ISOURce command.
Description
- This command sets or queries the current source for power quality measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:POWERQUALITY:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:POWERQUALITY:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:POWERQUALITY:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:POWERQUALITY:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:POWERQUALITY:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemPowerqualityStype
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:POWERQUALITY:STYPe command.
Description
- This command sets or queries the source type.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:POWERQUALITY:STYPe?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:POWERQUALITY:STYPe?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:POWERQUALITY:STYPe valuecommand.
SCPI Syntax
- POWer:POWer<x>:POWERQUALITY:STYPe {AC|DC}
- POWer:POWer<x>:POWERQUALITY:STYPe?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.ACsets the signal type as AC.DCsets the signal type as DC.
PowerPowerItemPowerqualityVsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:POWERQUALITY:VSOURce command.
Description
- This command sets or queries the voltage source for power quality measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:POWERQUALITY:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:POWERQUALITY:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:POWERQUALITY:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:POWERQUALITY:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:POWERQUALITY:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemPpulsewidth
¶
Bases: SCPICmdRead
The POWer:POWer<x>:PPULSEWIDTH command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PPULSEWIDTH?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PPULSEWIDTH?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Properties
.inputsource: ThePOWer:POWer<x>:PPULSEWIDTH:INPUTSOurcecommand.
inputsource
property
¶
inputsource: PowerPowerItemPpulsewidthInputsource
Return the POWer:POWer<x>:PPULSEWIDTH:INPUTSOurce command.
Description
- This command sets or queries the input source for positive pulse width measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PPULSEWIDTH:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PPULSEWIDTH:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PPULSEWIDTH:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:PPULSEWIDTH:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:PPULSEWIDTH:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemPpulsewidthInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PPULSEWIDTH:INPUTSOurce command.
Description
- This command sets or queries the input source for positive pulse width measurement in the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PPULSEWIDTH:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PPULSEWIDTH:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PPULSEWIDTH:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:PPULSEWIDTH:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:PPULSEWIDTH:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemPreset
¶
Bases: SCPICmdWrite
The POWer:POWer<x>:PRESET command.
Description
- This command runs a power preset action for the specified power measurement number.
Usage
- Using the
.write(value)method will send thePOWer:POWer<x>:PRESET valuecommand.
SCPI Syntax
- POWer:POWer<x>:PRESET {EXECute}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.EXECuteruns the power preset action.
PowerPowerItemPsrr
¶
Bases: SCPICmdRead
The POWer:POWer<x>:PSRR command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Properties
.ampval: ThePOWer:POWer<x>:PSRR:AMP<x>Valcommand..ampmode: ThePOWer:POWer<x>:PSRR:AMPModecommand..analysismethod: ThePOWer:POWer<x>:PSRR:ANALYSISMethodcommand..autorbw: ThePOWer:POWer<x>:PSRR:AUTORbwcommand..connectstatus: ThePOWer:POWer<x>:PSRR:CONNECTSTATuscommand..constamplitude: ThePOWer:POWer<x>:PSRR:CONSTAMPlitudecommand..freqval: ThePOWer:POWer<x>:PSRR:FREQ<x>Valcommand..genipaddress: ThePOWer:POWer<x>:PSRR:GENIPADDresscommand..generator: ThePOWer:POWer<x>:PSRR:GENeratorcommand..impedance: ThePOWer:POWer<x>:PSRR:IMPEDancecommand..inputsource: ThePOWer:POWer<x>:PSRR:INPUTSOurcecommand..outputsource: ThePOWer:POWer<x>:PSRR:OUTPUTSOurcecommand..ppd: ThePOWer:POWer<x>:PSRR:PPDcommand..startfrequency: ThePOWer:POWer<x>:PSRR:STARTFREQuencycommand..stopfrequency: ThePOWer:POWer<x>:PSRR:STOPFREQuencycommand..testconnection: ThePOWer:POWer<x>:PSRR:TESTCONNectioncommand.
ampmode
property
¶
ampmode: PowerPowerItemPsrrAmpmode
Return the POWer:POWer<x>:PSRR:AMPMode command.
Description
- This command sets or queries the amplitude mode for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:AMPMode?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:AMPMode?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:AMPMode valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:AMPMode {CONSTant|PROFile}
- POWer:POWer<x>:PSRR:AMPMode?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.CONSTantsets the amplitude mode to output a constant amplitude signal from the DUT stimulus generator for all frequency bands.PROFileenables configuring the generator to set amplitude values for each frequency band.
ampval
property
¶
ampval: dict[int, PowerPowerItemPsrrAmpvalItem]
Return the POWer:POWer<x>:PSRR:AMP<x>Val command.
Description
- This command sets or queries the generator amplitude value of the specified configuration step for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:AMP<x>Val?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:AMP<x>Val?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:AMP<x>Val valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:AMP<x>Val <NR3>
- POWer:POWer<x>:PSRR:AMP<x>Val?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
analysismethod
property
¶
analysismethod: PowerPowerItemPsrrAnalysismethod
Return the POWer:POWer<x>:PSRR:ANALYSISMethod command.
Description
- This command sets or queries the Analysis Method for PSRR measurements.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:ANALYSISMethod?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:ANALYSISMethod?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:ANALYSISMethod valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:ANALYSISMethod {SV|FFT}
- POWer:POWer<x>:PSRR:ANALYSISMethod?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.SVsets the Analysis Method as Spectrum View.FFTsets the Analysis Method as FFT.
autorbw
property
¶
autorbw: PowerPowerItemPsrrAutorbw
Return the POWer:POWer<x>:PSRR:AUTORbw command.
Description
- This command enables Auto RBW computation.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:AUTORbw?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:AUTORbw?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:AUTORbw valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:AUTORbw {True|False}
- POWer:POWer<x>:PSRR:AUTORbw?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.Trueenables Auto RBW computation.Falsedisables Auto RBW computation.
connectstatus
property
¶
connectstatus: PowerPowerItemPsrrConnectstatus
Return the POWer:POWer<x>:PSRR:CONNECTSTATus command.
Description
- Queries the external instrument’s connection status for the specified Power Supply Rejection Ratio (PSRR) measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:CONNECTSTATus?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:CONNECTSTATus?query and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:PSRR:CONNECTSTATus?
Info
POWer<x>is the number of the PSRR power measurement.
constamplitude
property
¶
constamplitude: PowerPowerItemPsrrConstamplitude
Return the POWer:POWer<x>:PSRR:CONSTAMPlitude command.
Description
- This command sets or queries the constant amplitude voltage for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:CONSTAMPlitude?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:CONSTAMPlitude?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:CONSTAMPlitude valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:CONSTAMPlitude <NR3>
- POWer:POWer<x>:PSRR:CONSTAMPlitude?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>is the constant amplitude voltage value for the measurement, in the range of -100 V to 100 V.
freqval
property
¶
freqval: dict[int, PowerPowerItemPsrrFreqvalItem]
Return the POWer:POWer<x>:PSRR:FREQ<x>Val command.
Description
- This command sets or queries the generator frequency value of the specified configuration step for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:FREQ<x>Val?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:FREQ<x>Val?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:FREQ<x>Val valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:FREQ<x>Val <NR3>
- POWer:POWer<x>:PSRR:FREQ<x>Val?
Info
POWer<x>sets the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.FREQ<x>sets the configuration step number, in the range of 1 to 11. Values outside this range will report an error.<NR3>sets the frequency of the specified configuration step number, in the range of 10 Hz to 50 MHz.
generator
property
¶
generator: PowerPowerItemPsrrGenerator
Return the POWer:POWer<x>:PSRR:GENerator command.
Description
- Sets or queries the generator source for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:GENerator?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:GENerator?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:GENerator valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:GENerator {INTernal|EXTernal}
- POWer:POWer<x>:PSRR:GENerator?
Info
POWer<x>is the number of the PSRR power measurement.INTernalsets the internal generator as the source for the Power Supply Rejection Ratio (PSRR) power measurement.EXTernalsets the external generator as the source for the Power Supply Rejection Ratio (PSRR) power measurement.
genipaddress
property
¶
genipaddress: PowerPowerItemPsrrGenipaddress
Return the POWer:POWer<x>:PSRR:GENIPADDress command.
Description
- Sets or queries the instrument’s IP Address associated with the specified Power Supply Rejection Ratio (PSRR) measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:GENIPADDress?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:GENIPADDress?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:GENIPADDress valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:GENIPADDress <QString>
- POWer:POWer<x>:PSRR:GENIPADDress?
Info
POWer<x>is the number of the PSRR power measurement.<QString>is the IP address of the generator.
impedance
property
¶
impedance: PowerPowerItemPsrrImpedance
Return the POWer:POWer<x>:PSRR:IMPEDance command.
Description
- This command sets or queries the vertical termination impedance for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:IMPEDance?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:IMPEDance?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:IMPEDance valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:IMPEDance {FIFTy|HIGHZ}
- POWer:POWer<x>:PSRR:IMPEDance?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.FIFTysets the impedance to be 50 Ω.HIGHZsets the impedance to be 1 MΩ.
inputsource
property
¶
inputsource: PowerPowerItemPsrrInputsource
Return the POWer:POWer<x>:PSRR:INPUTSOurce command.
Description
- This command sets or queries the input source for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:INPUTSOurce CH<x>
- POWer:POWer<x>:PSRR:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.CH<x>sets the channel to use for the input source.
outputsource
property
¶
outputsource: PowerPowerItemPsrrOutputsource
Return the POWer:POWer<x>:PSRR:OUTPUTSOurce command.
Description
- This command sets or queries the output source for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:OUTPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:OUTPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:OUTPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:OUTPUTSOurce CH<x>
- POWer:POWer<x>:PSRR:OUTPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.CH<x>sets the channel to use for the output source.
ppd
property
¶
Return the POWer:POWer<x>:PSRR:PPD command.
Description
- This command sets or queries the points per decade (PPD) value for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:PPD?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:PPD?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:PPD valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:PPD <NR3>
- POWer:POWer<x>:PSRR:PPD?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>is the PPD value for the measurement, in the range of 10 to 100 points.
startfrequency
property
¶
startfrequency: PowerPowerItemPsrrStartfrequency
Return the POWer:POWer<x>:PSRR:STARTFREQuency command.
Description
- This command sets or queries the start frequency value for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:STARTFREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:STARTFREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:STARTFREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:STARTFREQuency <NR3>
- POWer:POWer<x>:PSRR:STARTFREQuency?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>is the starting frequency for the measurement, in the range of 10 Hz to 50 MHz.
stopfrequency
property
¶
stopfrequency: PowerPowerItemPsrrStopfrequency
Return the POWer:POWer<x>:PSRR:STOPFREQuency command.
Description
- This command sets or queries the stop frequency value for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:STOPFREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:STOPFREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:STOPFREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:STOPFREQuency <NR3>
- POWer:POWer<x>:PSRR:STOPFREQuency?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>is the stop frequency for the measurement, in the range of 10 Hz to 50 MHz.
testconnection
property
¶
testconnection: PowerPowerItemPsrrTestconnection
Return the POWer:POWer<x>:PSRR:TESTCONNection command.
Description
- This command tests the connection with the external instrument for the specified Power Supply Rejection Ratio (PSRR) measurement.
Usage
- Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:TESTCONNection valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:TESTCONNection {EXECute}
Info
POWer<x>is the number of the PSRR power measurement.EXECuteruns the test connection function.
PowerPowerItemPsrrAmpmode
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PSRR:AMPMode command.
Description
- This command sets or queries the amplitude mode for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:AMPMode?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:AMPMode?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:AMPMode valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:AMPMode {CONSTant|PROFile}
- POWer:POWer<x>:PSRR:AMPMode?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.CONSTantsets the amplitude mode to output a constant amplitude signal from the DUT stimulus generator for all frequency bands.PROFileenables configuring the generator to set amplitude values for each frequency band.
PowerPowerItemPsrrAmpvalItem
¶
Bases: ValidatedDynamicNumberCmd, SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PSRR:AMP<x>Val command.
Description
- This command sets or queries the generator amplitude value of the specified configuration step for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:AMP<x>Val?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:AMP<x>Val?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:AMP<x>Val valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:AMP<x>Val <NR3>
- POWer:POWer<x>:PSRR:AMP<x>Val?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemPsrrAnalysismethod
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PSRR:ANALYSISMethod command.
Description
- This command sets or queries the Analysis Method for PSRR measurements.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:ANALYSISMethod?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:ANALYSISMethod?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:ANALYSISMethod valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:ANALYSISMethod {SV|FFT}
- POWer:POWer<x>:PSRR:ANALYSISMethod?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.SVsets the Analysis Method as Spectrum View.FFTsets the Analysis Method as FFT.
PowerPowerItemPsrrAutorbw
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PSRR:AUTORbw command.
Description
- This command enables Auto RBW computation.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:AUTORbw?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:AUTORbw?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:AUTORbw valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:AUTORbw {True|False}
- POWer:POWer<x>:PSRR:AUTORbw?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.Trueenables Auto RBW computation.Falsedisables Auto RBW computation.
PowerPowerItemPsrrConnectstatus
¶
Bases: SCPICmdRead
The POWer:POWer<x>:PSRR:CONNECTSTATus command.
Description
- Queries the external instrument’s connection status for the specified Power Supply Rejection Ratio (PSRR) measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:CONNECTSTATus?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:CONNECTSTATus?query and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:PSRR:CONNECTSTATus?
Info
POWer<x>is the number of the PSRR power measurement.
PowerPowerItemPsrrConstamplitude
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PSRR:CONSTAMPlitude command.
Description
- This command sets or queries the constant amplitude voltage for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:CONSTAMPlitude?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:CONSTAMPlitude?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:CONSTAMPlitude valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:CONSTAMPlitude <NR3>
- POWer:POWer<x>:PSRR:CONSTAMPlitude?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>is the constant amplitude voltage value for the measurement, in the range of -100 V to 100 V.
PowerPowerItemPsrrFreqvalItem
¶
Bases: ValidatedDynamicNumberCmd, SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PSRR:FREQ<x>Val command.
Description
- This command sets or queries the generator frequency value of the specified configuration step for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:FREQ<x>Val?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:FREQ<x>Val?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:FREQ<x>Val valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:FREQ<x>Val <NR3>
- POWer:POWer<x>:PSRR:FREQ<x>Val?
Info
POWer<x>sets the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.FREQ<x>sets the configuration step number, in the range of 1 to 11. Values outside this range will report an error.<NR3>sets the frequency of the specified configuration step number, in the range of 10 Hz to 50 MHz.
PowerPowerItemPsrrGenerator
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PSRR:GENerator command.
Description
- Sets or queries the generator source for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:GENerator?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:GENerator?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:GENerator valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:GENerator {INTernal|EXTernal}
- POWer:POWer<x>:PSRR:GENerator?
Info
POWer<x>is the number of the PSRR power measurement.INTernalsets the internal generator as the source for the Power Supply Rejection Ratio (PSRR) power measurement.EXTernalsets the external generator as the source for the Power Supply Rejection Ratio (PSRR) power measurement.
PowerPowerItemPsrrGenipaddress
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PSRR:GENIPADDress command.
Description
- Sets or queries the instrument’s IP Address associated with the specified Power Supply Rejection Ratio (PSRR) measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:GENIPADDress?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:GENIPADDress?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:GENIPADDress valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:GENIPADDress <QString>
- POWer:POWer<x>:PSRR:GENIPADDress?
Info
POWer<x>is the number of the PSRR power measurement.<QString>is the IP address of the generator.
PowerPowerItemPsrrImpedance
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PSRR:IMPEDance command.
Description
- This command sets or queries the vertical termination impedance for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:IMPEDance?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:IMPEDance?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:IMPEDance valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:IMPEDance {FIFTy|HIGHZ}
- POWer:POWer<x>:PSRR:IMPEDance?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.FIFTysets the impedance to be 50 Ω.HIGHZsets the impedance to be 1 MΩ.
PowerPowerItemPsrrInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PSRR:INPUTSOurce command.
Description
- This command sets or queries the input source for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:INPUTSOurce CH<x>
- POWer:POWer<x>:PSRR:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.CH<x>sets the channel to use for the input source.
PowerPowerItemPsrrOutputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PSRR:OUTPUTSOurce command.
Description
- This command sets or queries the output source for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:OUTPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:OUTPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:OUTPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:OUTPUTSOurce CH<x>
- POWer:POWer<x>:PSRR:OUTPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.CH<x>sets the channel to use for the output source.
PowerPowerItemPsrrPpd
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PSRR:PPD command.
Description
- This command sets or queries the points per decade (PPD) value for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:PPD?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:PPD?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:PPD valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:PPD <NR3>
- POWer:POWer<x>:PSRR:PPD?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>is the PPD value for the measurement, in the range of 10 to 100 points.
PowerPowerItemPsrrStartfrequency
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PSRR:STARTFREQuency command.
Description
- This command sets or queries the start frequency value for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:STARTFREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:STARTFREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:STARTFREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:STARTFREQuency <NR3>
- POWer:POWer<x>:PSRR:STARTFREQuency?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>is the starting frequency for the measurement, in the range of 10 Hz to 50 MHz.
PowerPowerItemPsrrStopfrequency
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:PSRR:STOPFREQuency command.
Description
- This command sets or queries the stop frequency value for the Power Supply Rejection Ratio (PSRR) power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:PSRR:STOPFREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:PSRR:STOPFREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:STOPFREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:STOPFREQuency <NR3>
- POWer:POWer<x>:PSRR:STOPFREQuency?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>is the stop frequency for the measurement, in the range of 10 Hz to 50 MHz.
PowerPowerItemPsrrTestconnection
¶
Bases: SCPICmdWrite
The POWer:POWer<x>:PSRR:TESTCONNection command.
Description
- This command tests the connection with the external instrument for the specified Power Supply Rejection Ratio (PSRR) measurement.
Usage
- Using the
.write(value)method will send thePOWer:POWer<x>:PSRR:TESTCONNection valuecommand.
SCPI Syntax
- POWer:POWer<x>:PSRR:TESTCONNection {EXECute}
Info
POWer<x>is the number of the PSRR power measurement.EXECuteruns the test connection function.
PowerPowerItemRdson
¶
Bases: SCPICmdRead
The POWer:POWer<x>:RDSON command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:RDSON?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:RDSON?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Properties
.devicetype: ThePOWer:POWer<x>:RDSON:DEVICETypecommand..isource: ThePOWer:POWer<x>:RDSON:ISOURcecommand..vsource: ThePOWer:POWer<x>:RDSON:VSOURcecommand.
devicetype
property
¶
devicetype: PowerPowerItemRdsonDevicetype
Return the POWer:POWer<x>:RDSON:DEVICEType command.
Description
- This command sets or queries the device type for the power drain source on resistance measurement for RDSon measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:RDSON:DEVICEType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:RDSON:DEVICEType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:RDSON:DEVICEType valuecommand.
SCPI Syntax
- POWer:POWer<x>:RDSON:DEVICEType {SWITCHING|PNJUNCTION}
- POWer:POWer<x>:RDSON:DEVICEType?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.SWITCHINGsets the Device Type to a switching device (v/i).PNJUNCTIONsets the Device Type to a PN Junction device (dv/di).
isource
property
¶
isource: PowerPowerItemRdsonIsource
Return the POWer:POWer<x>:RDSON:ISOURce command.
Description
- This command sets or queries the current source for RDSon measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:RDSON:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:RDSON:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:RDSON:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:RDSON:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:RDSON:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
vsource
property
¶
vsource: PowerPowerItemRdsonVsource
Return the POWer:POWer<x>:RDSON:VSOURce command.
Description
- This command sets or queries the voltage source for RDSon measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:RDSON:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:RDSON:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:RDSON:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:RDSON:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:RDSON:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemRdsonDevicetype
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:RDSON:DEVICEType command.
Description
- This command sets or queries the device type for the power drain source on resistance measurement for RDSon measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:RDSON:DEVICEType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:RDSON:DEVICEType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:RDSON:DEVICEType valuecommand.
SCPI Syntax
- POWer:POWer<x>:RDSON:DEVICEType {SWITCHING|PNJUNCTION}
- POWer:POWer<x>:RDSON:DEVICEType?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.SWITCHINGsets the Device Type to a switching device (v/i).PNJUNCTIONsets the Device Type to a PN Junction device (dv/di).
PowerPowerItemRdsonIsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:RDSON:ISOURce command.
Description
- This command sets or queries the current source for RDSon measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:RDSON:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:RDSON:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:RDSON:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:RDSON:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:RDSON:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemRdsonVsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:RDSON:VSOURce command.
Description
- This command sets or queries the voltage source for RDSon measurement of the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:RDSON:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:RDSON:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:RDSON:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:RDSON:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:RDSON:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemReflevels
¶
Bases: SCPICmdRead
The POWer:POWer<x>:REFLevels command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Properties
.absolute: ThePOWer:POWer<x>:REFLevels:ABSolutecommand tree..basetop: ThePOWer:POWer<x>:REFLevels:BASETopcommand..method: ThePOWer:POWer<x>:REFLevels:METHodcommand..percent: ThePOWer:POWer<x>:REFLevels:PERCentcommand tree.
absolute
property
¶
absolute: PowerPowerItemReflevelsAbsolute
Return the POWer:POWer<x>:REFLevels:ABSolute command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Sub-properties
.fallhigh: ThePOWer:POWer<x>:REFLevels:ABSolute:FALLHighcommand..falllow: ThePOWer:POWer<x>:REFLevels:ABSolute:FALLLowcommand..fallmid: ThePOWer:POWer<x>:REFLevels:ABSolute:FALLMidcommand..hysteresis: ThePOWer:POWer<x>:REFLevels:ABSolute:HYSTeresiscommand..risehigh: ThePOWer:POWer<x>:REFLevels:ABSolute:RISEHighcommand..riselow: ThePOWer:POWer<x>:REFLevels:ABSolute:RISELowcommand..risemid: ThePOWer:POWer<x>:REFLevels:ABSolute:RISEMidcommand..type: ThePOWer:POWer<x>:REFLevels:ABSolute:TYPEcommand.
basetop
property
¶
basetop: PowerPowerItemReflevelsBasetop
Return the POWer:POWer<x>:REFLevels:BASETop command.
Description
- This command sets or queries the reference level base top method for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:BASETop?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:BASETop?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:BASETop valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:BASETop {AUTO|MINMax|MEANhistogram|MODEhistogram|EYEhistogram}
- POWer:POWer<x>:REFLevels:BASETop?
method
property
¶
method: PowerPowerItemReflevelsMethod
Return the POWer:POWer<x>:REFLevels:METHod command.
Description
- This command sets or queries the method to configure reference level values for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:METHod?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:METHod?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:METHod valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:METHod {PERCent|ABSolute}
- POWer:POWer<x>:REFLevels:METHod?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.PERCentsets the power measurement to use absolute values to configure reference level values.ABSolutesets the power measurement to use percentage to configure reference level values.
percent
property
¶
percent: PowerPowerItemReflevelsPercent
Return the POWer:POWer<x>:REFLevels:PERCent command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Sub-properties
.fallhigh: ThePOWer:POWer<x>:REFLevels:PERCent:FALLHighcommand..falllow: ThePOWer:POWer<x>:REFLevels:PERCent:FALLLowcommand..fallmid: ThePOWer:POWer<x>:REFLevels:PERCent:FALLMidcommand..hysteresis: ThePOWer:POWer<x>:REFLevels:PERCent:HYSTeresiscommand..risehigh: ThePOWer:POWer<x>:REFLevels:PERCent:RISEHighcommand..riselow: ThePOWer:POWer<x>:REFLevels:PERCent:RISELowcommand..risemid: ThePOWer:POWer<x>:REFLevels:PERCent:RISEMidcommand..type: ThePOWer:POWer<x>:REFLevels:PERCent:TYPEcommand.
PowerPowerItemReflevelsAbsolute
¶
Bases: SCPICmdRead
The POWer:POWer<x>:REFLevels:ABSolute command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Properties
.fallhigh: ThePOWer:POWer<x>:REFLevels:ABSolute:FALLHighcommand..falllow: ThePOWer:POWer<x>:REFLevels:ABSolute:FALLLowcommand..fallmid: ThePOWer:POWer<x>:REFLevels:ABSolute:FALLMidcommand..hysteresis: ThePOWer:POWer<x>:REFLevels:ABSolute:HYSTeresiscommand..risehigh: ThePOWer:POWer<x>:REFLevels:ABSolute:RISEHighcommand..riselow: ThePOWer:POWer<x>:REFLevels:ABSolute:RISELowcommand..risemid: ThePOWer:POWer<x>:REFLevels:ABSolute:RISEMidcommand..type: ThePOWer:POWer<x>:REFLevels:ABSolute:TYPEcommand.
fallhigh
property
¶
fallhigh: PowerPowerItemReflevelsAbsoluteFallhigh
Return the POWer:POWer<x>:REFLevels:ABSolute:FALLHigh command.
Description
- This command sets or queries the falling edge for high reference level in absolute units for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLHigh?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLHigh?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLHigh valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:ABSolute:FALLHigh <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:FALLHigh?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -40000 to 40000.
falllow
property
¶
Return the POWer:POWer<x>:REFLevels:ABSolute:FALLLow command.
Description
- This command sets or queries the falling edge for low reference level in absolute units for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLLow?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLLow?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLLow valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:ABSolute:FALLLow <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:FALLLow?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -40000 to 40000.
fallmid
property
¶
Return the POWer:POWer<x>:REFLevels:ABSolute:FALLMid command.
Description
- This command sets or queries the falling edge for mid reference level in absolute units for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLMid?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLMid?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLMid valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:ABSolute:FALLMid <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:FALLMid?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -40000 to 40000.
hysteresis
property
¶
hysteresis: PowerPowerItemReflevelsAbsoluteHysteresis
Return the POWer:POWer<x>:REFLevels:ABSolute:HYSTeresis command.
Description
- This command sets or queries the absolute hysteresis value for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute:HYSTeresis?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:HYSTeresis?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:HYSTeresis valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:ABSolute:HYSTeresis <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:HYSTeresis?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 0.0000005 to 10.
risehigh
property
¶
risehigh: PowerPowerItemReflevelsAbsoluteRisehigh
Return the POWer:POWer<x>:REFLevels:ABSolute:RISEHigh command.
Description
- This command sets or queries the rising edge for high reference level in absolute units for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISEHigh?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISEHigh?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISEHigh valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:ABSolute:RISEHigh <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:RISEHigh?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -40000 to 40000.
riselow
property
¶
Return the POWer:POWer<x>:REFLevels:ABSolute:RISELow command.
Description
- This command sets or queries the rising edge for low reference level in absolute units for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISELow?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISELow?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISELow valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:ABSolute:RISELow <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:RISELow?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -40000 to 40000.
risemid
property
¶
Return the POWer:POWer<x>:REFLevels:ABSolute:RISEMid command.
Description
- This command sets or queries the rising edge for mid reference level in absolute units for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISEMid?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISEMid?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISEMid valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:ABSolute:RISEMid <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:RISEMid?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -40000 to 40000.
type
property
¶
Return the POWer:POWer<x>:REFLevels:ABSolute:TYPE command.
Description
- This command sets or queries the type of measurement levels when reference level is set to absolute for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute:TYPE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:TYPE?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:TYPE valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:ABSolute:TYPE {SAME|UNIQue}
- POWer:POWer<x>:REFLevels:ABSolute:TYPE?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.SAME: select when the rising edge and falling edge reference levels are same.UNIQue: select when the rising edge and falling edge reference levels are different.
PowerPowerItemReflevelsAbsoluteFallhigh
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:ABSolute:FALLHigh command.
Description
- This command sets or queries the falling edge for high reference level in absolute units for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLHigh?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLHigh?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLHigh valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:ABSolute:FALLHigh <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:FALLHigh?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -40000 to 40000.
PowerPowerItemReflevelsAbsoluteFalllow
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:ABSolute:FALLLow command.
Description
- This command sets or queries the falling edge for low reference level in absolute units for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLLow?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLLow?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLLow valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:ABSolute:FALLLow <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:FALLLow?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -40000 to 40000.
PowerPowerItemReflevelsAbsoluteFallmid
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:ABSolute:FALLMid command.
Description
- This command sets or queries the falling edge for mid reference level in absolute units for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLMid?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLMid?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:FALLMid valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:ABSolute:FALLMid <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:FALLMid?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -40000 to 40000.
PowerPowerItemReflevelsAbsoluteHysteresis
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:ABSolute:HYSTeresis command.
Description
- This command sets or queries the absolute hysteresis value for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute:HYSTeresis?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:HYSTeresis?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:HYSTeresis valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:ABSolute:HYSTeresis <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:HYSTeresis?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 0.0000005 to 10.
PowerPowerItemReflevelsAbsoluteRisehigh
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:ABSolute:RISEHigh command.
Description
- This command sets or queries the rising edge for high reference level in absolute units for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISEHigh?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISEHigh?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISEHigh valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:ABSolute:RISEHigh <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:RISEHigh?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -40000 to 40000.
PowerPowerItemReflevelsAbsoluteRiselow
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:ABSolute:RISELow command.
Description
- This command sets or queries the rising edge for low reference level in absolute units for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISELow?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISELow?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISELow valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:ABSolute:RISELow <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:RISELow?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -40000 to 40000.
PowerPowerItemReflevelsAbsoluteRisemid
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:ABSolute:RISEMid command.
Description
- This command sets or queries the rising edge for mid reference level in absolute units for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISEMid?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISEMid?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:RISEMid valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:ABSolute:RISEMid <NR1>
- POWer:POWer<x>:REFLevels:ABSolute:RISEMid?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -40000 to 40000.
PowerPowerItemReflevelsAbsoluteType
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:ABSolute:TYPE command.
Description
- This command sets or queries the type of measurement levels when reference level is set to absolute for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:ABSolute:TYPE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:TYPE?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:ABSolute:TYPE valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:ABSolute:TYPE {SAME|UNIQue}
- POWer:POWer<x>:REFLevels:ABSolute:TYPE?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.SAME: select when the rising edge and falling edge reference levels are same.UNIQue: select when the rising edge and falling edge reference levels are different.
PowerPowerItemReflevelsBasetop
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:BASETop command.
Description
- This command sets or queries the reference level base top method for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:BASETop?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:BASETop?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:BASETop valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:BASETop {AUTO|MINMax|MEANhistogram|MODEhistogram|EYEhistogram}
- POWer:POWer<x>:REFLevels:BASETop?
PowerPowerItemReflevelsMethod
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:METHod command.
Description
- This command sets or queries the method to configure reference level values for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:METHod?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:METHod?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:METHod valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:METHod {PERCent|ABSolute}
- POWer:POWer<x>:REFLevels:METHod?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.PERCentsets the power measurement to use absolute values to configure reference level values.ABSolutesets the power measurement to use percentage to configure reference level values.
PowerPowerItemReflevelsPercent
¶
Bases: SCPICmdRead
The POWer:POWer<x>:REFLevels:PERCent command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Properties
.fallhigh: ThePOWer:POWer<x>:REFLevels:PERCent:FALLHighcommand..falllow: ThePOWer:POWer<x>:REFLevels:PERCent:FALLLowcommand..fallmid: ThePOWer:POWer<x>:REFLevels:PERCent:FALLMidcommand..hysteresis: ThePOWer:POWer<x>:REFLevels:PERCent:HYSTeresiscommand..risehigh: ThePOWer:POWer<x>:REFLevels:PERCent:RISEHighcommand..riselow: ThePOWer:POWer<x>:REFLevels:PERCent:RISELowcommand..risemid: ThePOWer:POWer<x>:REFLevels:PERCent:RISEMidcommand..type: ThePOWer:POWer<x>:REFLevels:PERCent:TYPEcommand.
fallhigh
property
¶
fallhigh: PowerPowerItemReflevelsPercentFallhigh
Return the POWer:POWer<x>:REFLevels:PERCent:FALLHigh command.
Description
- This command sets or queries the falling edge for high reference level in percentage for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLHigh?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLHigh?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLHigh valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:PERCent:FALLHigh <NR1>
- POWer:POWer<x>:REFLevels:PERCent:FALLHigh?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 1 to 99.
falllow
property
¶
Return the POWer:POWer<x>:REFLevels:PERCent:FALLLow command.
Description
- This command sets or queries the falling edge for low reference level in percentage for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLLow?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLLow?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLLow valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:PERCent:FALLLow <NR1>
- POWer:POWer<x>:REFLevels:PERCent:FALLLow?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 1 to 99.
fallmid
property
¶
Return the POWer:POWer<x>:REFLevels:PERCent:FALLMid command.
Description
- This command sets or queries the falling edge for mid reference level in percentage for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLMid?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLMid?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLMid valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:PERCent:FALLMid <NR1>
- POWer:POWer<x>:REFLevels:PERCent:FALLMid?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 1 to 99.
hysteresis
property
¶
hysteresis: PowerPowerItemReflevelsPercentHysteresis
Return the POWer:POWer<x>:REFLevels:PERCent:HYSTeresis command.
Description
- This command sets or queries the hysteresis in percentage for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent:HYSTeresis?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:HYSTeresis?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:HYSTeresis valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:PERCent:HYSTeresis <NR1>
- POWer:POWer<x>:REFLevels:PERCent:HYSTeresis?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 1 to 99.
risehigh
property
¶
risehigh: PowerPowerItemReflevelsPercentRisehigh
Return the POWer:POWer<x>:REFLevels:PERCent:RISEHigh command.
Description
- This command sets or queries the rising edge for high reference level in percentage for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent:RISEHigh?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:RISEHigh?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:RISEHigh valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:PERCent:RISEHigh <NR1>
- POWer:POWer<x>:REFLevels:PERCent:RISEHigh?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 1 to 99.
riselow
property
¶
Return the POWer:POWer<x>:REFLevels:PERCent:RISELow command.
Description
- This command sets or queries the rising edge for low reference level in percentage for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent:RISELow?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:RISELow?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:RISELow valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:PERCent:RISELow <NR1>
- POWer:POWer<x>:REFLevels:PERCent:RISELow?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 1 to 99.
risemid
property
¶
Return the POWer:POWer<x>:REFLevels:PERCent:RISEMid command.
Description
- This command sets or queries the rising edge for mid reference level in percentage for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent:RISEMid?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:RISEMid?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:RISEMid valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:PERCent:RISEMid <NR1>
- POWer:POWer<x>:REFLevels:PERCent:RISEMid?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 1 to 99.
type
property
¶
Return the POWer:POWer<x>:REFLevels:PERCent:TYPE command.
Description
- This command sets or queries the reference levels for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent:TYPE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:TYPE?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:TYPE valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:PERCent:TYPE {TENNinety|TWENtyeighty|CUSTom}
- POWer:POWer<x>:REFLevels:PERCent:TYPE?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.TENNinetyto set the low reference levels as 10% and high reference levels as 90%.TWENtyeightyto set the low reference levels as 20% and high reference levels as 80%.CUSTomto set the custom low, high, and mid reference levels for rising and falling edges.
PowerPowerItemReflevelsPercentFallhigh
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:PERCent:FALLHigh command.
Description
- This command sets or queries the falling edge for high reference level in percentage for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLHigh?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLHigh?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLHigh valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:PERCent:FALLHigh <NR1>
- POWer:POWer<x>:REFLevels:PERCent:FALLHigh?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 1 to 99.
PowerPowerItemReflevelsPercentFalllow
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:PERCent:FALLLow command.
Description
- This command sets or queries the falling edge for low reference level in percentage for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLLow?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLLow?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLLow valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:PERCent:FALLLow <NR1>
- POWer:POWer<x>:REFLevels:PERCent:FALLLow?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 1 to 99.
PowerPowerItemReflevelsPercentFallmid
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:PERCent:FALLMid command.
Description
- This command sets or queries the falling edge for mid reference level in percentage for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLMid?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLMid?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:FALLMid valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:PERCent:FALLMid <NR1>
- POWer:POWer<x>:REFLevels:PERCent:FALLMid?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 1 to 99.
PowerPowerItemReflevelsPercentHysteresis
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:PERCent:HYSTeresis command.
Description
- This command sets or queries the hysteresis in percentage for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent:HYSTeresis?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:HYSTeresis?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:HYSTeresis valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:PERCent:HYSTeresis <NR1>
- POWer:POWer<x>:REFLevels:PERCent:HYSTeresis?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 1 to 99.
PowerPowerItemReflevelsPercentRisehigh
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:PERCent:RISEHigh command.
Description
- This command sets or queries the rising edge for high reference level in percentage for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent:RISEHigh?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:RISEHigh?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:RISEHigh valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:PERCent:RISEHigh <NR1>
- POWer:POWer<x>:REFLevels:PERCent:RISEHigh?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 1 to 99.
PowerPowerItemReflevelsPercentRiselow
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:PERCent:RISELow command.
Description
- This command sets or queries the rising edge for low reference level in percentage for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent:RISELow?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:RISELow?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:RISELow valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:PERCent:RISELow <NR1>
- POWer:POWer<x>:REFLevels:PERCent:RISELow?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 1 to 99.
PowerPowerItemReflevelsPercentRisemid
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:PERCent:RISEMid command.
Description
- This command sets or queries the rising edge for mid reference level in percentage for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent:RISEMid?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:RISEMid?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:RISEMid valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:PERCent:RISEMid <NR1>
- POWer:POWer<x>:REFLevels:PERCent:RISEMid?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 1 to 99.
PowerPowerItemReflevelsPercentType
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:REFLevels:PERCent:TYPE command.
Description
- This command sets or queries the reference levels for the specified power measurement number.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:REFLevels:PERCent:TYPE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:TYPE?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:REFLevels:PERCent:TYPE valuecommand.
SCPI Syntax
- POWer:POWer<x>:REFLevels:PERCent:TYPE {TENNinety|TWENtyeighty|CUSTom}
- POWer:POWer<x>:REFLevels:PERCent:TYPE?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.TENNinetyto set the low reference levels as 10% and high reference levels as 90%.TWENtyeightyto set the low reference levels as 20% and high reference levels as 80%.CUSTomto set the custom low, high, and mid reference levels for rising and falling edges.
PowerPowerItemResults
¶
Bases: SCPICmdRead
The POWer:POWer<x>:RESUlts command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:RESUlts?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:RESUlts?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Properties
.allacqs: ThePOWer:POWer<x>:RESUlts:ALLAcqscommand tree..currentacq: ThePOWer:POWer<x>:RESUlts:CURRentacqcommand tree.
allacqs
property
¶
allacqs: PowerPowerItemResultsAllacqs
Return the POWer:POWer<x>:RESUlts:ALLAcqs command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:RESUlts:ALLAcqs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Sub-properties
.maximum: ThePOWer:POWer<x>:RESUlts:ALLAcqs:MAXimumcommand..mean: ThePOWer:POWer<x>:RESUlts:ALLAcqs:MEANcommand..minimum: ThePOWer:POWer<x>:RESUlts:ALLAcqs:MINimumcommand..pk2pk: ThePOWer:POWer<x>:RESUlts:ALLAcqs:PK2PKcommand..population: ThePOWer:POWer<x>:RESUlts:ALLAcqs:POPUlationcommand..stddev: ThePOWer:POWer<x>:RESUlts:ALLAcqs:STDDevcommand.
currentacq
property
¶
currentacq: PowerPowerItemResultsCurrentacq
Return the POWer:POWer<x>:RESUlts:CURRentacq command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:RESUlts:CURRentacq?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Sub-properties
.f1mag: ThePOWer:POWer<x>:RESUlts:CURRentacq:F1MAGcommand..f3mag: ThePOWer:POWer<x>:RESUlts:CURRentacq:F3MAGcommand..frequency: ThePOWer:POWer<x>:RESUlts:CURRentacq:FREQUENCYcommand..irms: ThePOWer:POWer<x>:RESUlts:CURRentacq:IRMScommand..maximum: ThePOWer:POWer<x>:RESUlts:CURRentacq:MAXimumcommand..mean: ThePOWer:POWer<x>:RESUlts:CURRentacq:MEANcommand..minimum: ThePOWer:POWer<x>:RESUlts:CURRentacq:MINimumcommand..pk2pk: ThePOWer:POWer<x>:RESUlts:CURRentacq:PK2PKcommand..pohcl: ThePOWer:POWer<x>:RESUlts:CURRentacq:POHCLcommand..pohcm: ThePOWer:POWer<x>:RESUlts:CURRentacq:POHCMcommand..pohcs: ThePOWer:POWer<x>:RESUlts:CURRentacq:POHCScommand..population: ThePOWer:POWer<x>:RESUlts:CURRentacq:POPUlationcommand..rms: ThePOWer:POWer<x>:RESUlts:CURRentacq:RMScommand..status: ThePOWer:POWer<x>:RESUlts:CURRentacq:STATUScommand..stddev: ThePOWer:POWer<x>:RESUlts:CURRentacq:STDDevcommand..thdf: ThePOWer:POWer<x>:RESUlts:CURRentacq:THDFcommand..thdr: ThePOWer:POWer<x>:RESUlts:CURRentacq:THDRcommand..trpwr: ThePOWer:POWer<x>:RESUlts:CURRentacq:TRPWRcommand..vrms: ThePOWer:POWer<x>:RESUlts:CURRentacq:VRMScommand.
PowerPowerItemResultsAllacqs
¶
Bases: SCPICmdRead
The POWer:POWer<x>:RESUlts:ALLAcqs command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:RESUlts:ALLAcqs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Properties
.maximum: ThePOWer:POWer<x>:RESUlts:ALLAcqs:MAXimumcommand..mean: ThePOWer:POWer<x>:RESUlts:ALLAcqs:MEANcommand..minimum: ThePOWer:POWer<x>:RESUlts:ALLAcqs:MINimumcommand..pk2pk: ThePOWer:POWer<x>:RESUlts:ALLAcqs:PK2PKcommand..population: ThePOWer:POWer<x>:RESUlts:ALLAcqs:POPUlationcommand..stddev: ThePOWer:POWer<x>:RESUlts:ALLAcqs:STDDevcommand.
maximum
property
¶
maximum: PowerPowerItemResultsAllacqsMaximum
Return the POWer:POWer<x>:RESUlts:ALLAcqs:MAXimum command.
Description
- This command queries the maximum value of all acquisitions for the measurement
parameter in the specified power measurement number. The power measurement number is
specified by x. Note: Above entries are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:MAXimum? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:MAXimum? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:ALLAcqs:MAXimum? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
mean
property
¶
Return the POWer:POWer<x>:RESUlts:ALLAcqs:MEAN command.
Description
- This command queries the mean value of all acquisitions for the measurement parameter
in the specified power measurement number
. Note: Above entries are entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:MEAN? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:MEAN? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:ALLAcqs:MEAN? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
minimum
property
¶
minimum: PowerPowerItemResultsAllacqsMinimum
Return the POWer:POWer<x>:RESUlts:ALLAcqs:MINimum command.
Description
- This command queries the minimum value of all acquisitions for the measurement
parameter of the specified power measurement
. Note: Above entries are entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:MINimum? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:MINimum? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:ALLAcqs:MINimum? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
pk2pk
property
¶
Return the POWer:POWer<x>:RESUlts:ALLAcqs:PK2PK command.
Description
- This command queries the peak-to-peak value of all acquisitions for the measurement
parameter in the specified power measurement number. Note: Above entries are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:PK2PK? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:PK2PK? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:ALLAcqs:PK2PK? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
population
property
¶
population: PowerPowerItemResultsAllacqsPopulation
Return the POWer:POWer<x>:RESUlts:ALLAcqs:POPUlation command.
Description
- This command queries the population (number of complete cycles) of all acquisitions
for the measurement parameter in the specified power measurement number. Note: Above
entries are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:POPUlation? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:POPUlation? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:ALLAcqs:POPUlation? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
stddev
property
¶
Return the POWer:POWer<x>:RESUlts:ALLAcqs:STDDev command.
Description
- This command queries the standard deviation value of all acquisitions for the
measurement parameter in the specified power measurement number. Note: Above entries
are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:STDDev? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:STDDev? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:ALLAcqs:STDDev? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
PowerPowerItemResultsAllacqsMaximum
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:ALLAcqs:MAXimum command.
Description
- This command queries the maximum value of all acquisitions for the measurement parameter
in the specified power measurement number. The power measurement number is specified by x.
Note: Above entries are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:MAXimum? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:MAXimum? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:ALLAcqs:MAXimum? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
PowerPowerItemResultsAllacqsMean
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:ALLAcqs:MEAN command.
Description
- This command queries the mean value of all acquisitions for the measurement parameter in
the specified power measurement number
. Note: Above entries are entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:MEAN? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:MEAN? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:ALLAcqs:MEAN? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
PowerPowerItemResultsAllacqsMinimum
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:ALLAcqs:MINimum command.
Description
- This command queries the minimum value of all acquisitions for the measurement parameter
of the specified power measurement
. Note: Above entries are entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:MINimum? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:MINimum? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:ALLAcqs:MINimum? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
PowerPowerItemResultsAllacqsPk2pk
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:ALLAcqs:PK2PK command.
Description
- This command queries the peak-to-peak value of all acquisitions for the measurement
parameter in the specified power measurement number. Note: Above entries are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:PK2PK? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:PK2PK? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:ALLAcqs:PK2PK? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
PowerPowerItemResultsAllacqsPopulation
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:ALLAcqs:POPUlation command.
Description
- This command queries the population (number of complete cycles) of all acquisitions for
the measurement parameter in the specified power measurement number. Note: Above entries
are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:POPUlation? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:POPUlation? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:ALLAcqs:POPUlation? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
PowerPowerItemResultsAllacqsStddev
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:ALLAcqs:STDDev command.
Description
- This command queries the standard deviation value of all acquisitions for the measurement
parameter in the specified power measurement number. Note: Above entries are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:STDDev? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:ALLAcqs:STDDev? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:ALLAcqs:STDDev? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
PowerPowerItemResultsCurrentacq
¶
Bases: SCPICmdRead
The POWer:POWer<x>:RESUlts:CURRentacq command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:RESUlts:CURRentacq?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Properties
.f1mag: ThePOWer:POWer<x>:RESUlts:CURRentacq:F1MAGcommand..f3mag: ThePOWer:POWer<x>:RESUlts:CURRentacq:F3MAGcommand..frequency: ThePOWer:POWer<x>:RESUlts:CURRentacq:FREQUENCYcommand..irms: ThePOWer:POWer<x>:RESUlts:CURRentacq:IRMScommand..maximum: ThePOWer:POWer<x>:RESUlts:CURRentacq:MAXimumcommand..mean: ThePOWer:POWer<x>:RESUlts:CURRentacq:MEANcommand..minimum: ThePOWer:POWer<x>:RESUlts:CURRentacq:MINimumcommand..pk2pk: ThePOWer:POWer<x>:RESUlts:CURRentacq:PK2PKcommand..pohcl: ThePOWer:POWer<x>:RESUlts:CURRentacq:POHCLcommand..pohcm: ThePOWer:POWer<x>:RESUlts:CURRentacq:POHCMcommand..pohcs: ThePOWer:POWer<x>:RESUlts:CURRentacq:POHCScommand..population: ThePOWer:POWer<x>:RESUlts:CURRentacq:POPUlationcommand..rms: ThePOWer:POWer<x>:RESUlts:CURRentacq:RMScommand..status: ThePOWer:POWer<x>:RESUlts:CURRentacq:STATUScommand..stddev: ThePOWer:POWer<x>:RESUlts:CURRentacq:STDDevcommand..thdf: ThePOWer:POWer<x>:RESUlts:CURRentacq:THDFcommand..thdr: ThePOWer:POWer<x>:RESUlts:CURRentacq:THDRcommand..trpwr: ThePOWer:POWer<x>:RESUlts:CURRentacq:TRPWRcommand..vrms: ThePOWer:POWer<x>:RESUlts:CURRentacq:VRMScommand.
f1mag
property
¶
Return the POWer:POWer<x>:RESUlts:CURRentacq:F1MAG command.
Description
- This command queries the first harmonics magnitude value for the specified power measurement number.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:F1MAG? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:F1MAG? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:F1MAG? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
f3mag
property
¶
Return the POWer:POWer<x>:RESUlts:CURRentacq:F3MAG command.
Description
- This command queries the third harmonics magnitude value for the specified power measurement number.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:F3MAG? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:F3MAG? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:F3MAG? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
frequency
property
¶
frequency: PowerPowerItemResultsCurrentacqFrequency
Return the POWer:POWer<x>:RESUlts:CURRentacq:FREQUENCY command.
Description
- This command queries the fundamental frequency for the specified power measurement number.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:FREQUENCY? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:FREQUENCY? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:FREQUENCY? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
irms
property
¶
Return the POWer:POWer<x>:RESUlts:CURRentacq:IRMS command.
Description
- This command queries the RMS current value for the specified power measurement number.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:IRMS? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:IRMS? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:IRMS? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
maximum
property
¶
Return the POWer:POWer<x>:RESUlts:CURRentacq:MAXimum command.
Description
- This command queries the maximum value of the current acquisition for the measurement
parameter in the specified power measurement number. Note: Above entries are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:MAXimum? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:MAXimum? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:MAXimum? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
mean
property
¶
Return the POWer:POWer<x>:RESUlts:CURRentacq:MEAN command.
Description
- This command queries the mean value of the current acquisition for the measurement
parameter of the specified power measurement
. Note: Above entries are entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:MEAN? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:MEAN? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:MEAN? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
minimum
property
¶
Return the POWer:POWer<x>:RESUlts:CURRentacq:MINimum command.
Description
- This command queries the minimum value of the current acquisition for the measurement
parameter in the specified power measurement number. Note: Above entries are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:MINimum? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:MINimum? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:MINimum? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.'OUTPUT1' - 'OUTPUT7'are the parameters for the Turn On Time and Turn Off Time measurements.'GAINCROSSOVERFREQ', 'PHASECROSSOVERFREQ', 'GM', 'PM'are the parameters for the Control Loop Response measurement.'MAXPSRR', 'MAXPSRRFREQ', 'MINPSRR', 'MINPSRRFREQ'are the parameters for the PSRR measurement.
pk2pk
property
¶
Return the POWer:POWer<x>:RESUlts:CURRentacq:PK2PK command.
Description
- This command queries the peak-to-peak value of the current acquisition for the
measurement parameter in the specified power measurement number. Note: Above entries
are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:RESUlts:CURRentacq:PK2PK?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:PK2PK?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:PK2PK valuecommand.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:PK2PK ? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
- POWer:POWer<x>:RESUlts:CURRentacq:PK2PK?
pohcl
property
¶
Return the POWer:POWer<x>:RESUlts:CURRentacq:POHCL command.
Description
- This command queries the limit of partial odd harmonic current for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:POHCL? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:POHCL? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:POHCL? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
pohcm
property
¶
Return the POWer:POWer<x>:RESUlts:CURRentacq:POHCM command.
Description
- This command queries the measured value of partial odd harmonic current for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:POHCM? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:POHCM? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:POHCM? harmonics
pohcs
property
¶
Return the POWer:POWer<x>:RESUlts:CURRentacq:POHCS command.
Description
- This command queries the status of partial odd harmonic current for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:POHCS? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:POHCS? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:POHCS? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
population
property
¶
population: PowerPowerItemResultsCurrentacqPopulation
Return the POWer:POWer<x>:RESUlts:CURRentacq:POPUlation command.
Description
- This command queries the population (number of complete cycles) of the current
acquisition for the measurement parameter in the specified power measurement number.
The power measurement number is specified by x. Note: Above entries are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:POPUlation? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:POPUlation? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:POPUlation? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
rms
property
¶
Return the POWer:POWer<x>:RESUlts:CURRentacq:RMS command.
Description
- This command queries the RMS value of the source selected for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:RMS? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:RMS? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:RMS? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
status
property
¶
Return the POWer:POWer<x>:RESUlts:CURRentacq:STATUS command.
Description
- This command queries the status of the measurement for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:STATUS? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:STATUS? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:STATUS? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
stddev
property
¶
Return the POWer:POWer<x>:RESUlts:CURRentacq:STDDev command.
Description
- This command queries the standard deviation value of the current acquisition for the
measurement parameter in the specified power measurement number. The power measurement
number is specified by x. Note: Above entries are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:STDDev? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:STDDev? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:STDDev? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString>= the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
thdf
property
¶
Return the POWer:POWer<x>:RESUlts:CURRentacq:THDF command.
Description
- This command queries the total harmonic distortion (fundamental) value for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:THDF? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:THDF? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:THDF? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
thdr
property
¶
Return the POWer:POWer<x>:RESUlts:CURRentacq:THDR command.
Description
- This command queries the total harmonic distortion (RMS) value for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:THDR? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:THDR? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:THDR? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
trpwr
property
¶
Return the POWer:POWer<x>:RESUlts:CURRentacq:TRPWR command.
Description
- This command queries the true power value for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:TRPWR? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:TRPWR? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:TRPWR? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
vrms
property
¶
Return the POWer:POWer<x>:RESUlts:CURRentacq:VRMS command.
Description
- This command queries the RMS voltage value for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:VRMS? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:VRMS? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:VRMS? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
PowerPowerItemResultsCurrentacqF1mag
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:F1MAG command.
Description
- This command queries the first harmonics magnitude value for the specified power measurement number.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:F1MAG? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:F1MAG? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:F1MAG? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
PowerPowerItemResultsCurrentacqF3mag
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:F3MAG command.
Description
- This command queries the third harmonics magnitude value for the specified power measurement number.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:F3MAG? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:F3MAG? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:F3MAG? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
PowerPowerItemResultsCurrentacqFrequency
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:FREQUENCY command.
Description
- This command queries the fundamental frequency for the specified power measurement number.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:FREQUENCY? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:FREQUENCY? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:FREQUENCY? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
PowerPowerItemResultsCurrentacqIrms
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:IRMS command.
Description
- This command queries the RMS current value for the specified power measurement number.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:IRMS? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:IRMS? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:IRMS? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
PowerPowerItemResultsCurrentacqMaximum
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:MAXimum command.
Description
- This command queries the maximum value of the current acquisition for the measurement
parameter in the specified power measurement number. Note: Above entries are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:MAXimum? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:MAXimum? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:MAXimum? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
PowerPowerItemResultsCurrentacqMean
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:MEAN command.
Description
- This command queries the mean value of the current acquisition for the measurement
parameter of the specified power measurement
. Note: Above entries are entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:MEAN? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:MEAN? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:MEAN? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
PowerPowerItemResultsCurrentacqMinimum
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:MINimum command.
Description
- This command queries the minimum value of the current acquisition for the measurement
parameter in the specified power measurement number. Note: Above entries are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:MINimum? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:MINimum? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:MINimum? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.'OUTPUT1' - 'OUTPUT7'are the parameters for the Turn On Time and Turn Off Time measurements.'GAINCROSSOVERFREQ', 'PHASECROSSOVERFREQ', 'GM', 'PM'are the parameters for the Control Loop Response measurement.'MAXPSRR', 'MAXPSRRFREQ', 'MINPSRR', 'MINPSRRFREQ'are the parameters for the PSRR measurement.
PowerPowerItemResultsCurrentacqPk2pk
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:RESUlts:CURRentacq:PK2PK command.
Description
- This command queries the peak-to-peak value of the current acquisition for the measurement
parameter in the specified power measurement number. Note: Above entries are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:RESUlts:CURRentacq:PK2PK?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:PK2PK?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:PK2PK valuecommand.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:PK2PK ? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
- POWer:POWer<x>:RESUlts:CURRentacq:PK2PK?
PowerPowerItemResultsCurrentacqPohcl
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:POHCL command.
Description
- This command queries the limit of partial odd harmonic current for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:POHCL? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:POHCL? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:POHCL? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
PowerPowerItemResultsCurrentacqPohcm
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:POHCM command.
Description
- This command queries the measured value of partial odd harmonic current for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:POHCM? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:POHCM? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:POHCM? harmonics
PowerPowerItemResultsCurrentacqPohcs
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:POHCS command.
Description
- This command queries the status of partial odd harmonic current for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:POHCS? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:POHCS? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:POHCS? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
PowerPowerItemResultsCurrentacqPopulation
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:POPUlation command.
Description
- This command queries the population (number of complete cycles) of the current acquisition
for the measurement parameter in the specified power measurement number. The power
measurement number is specified by x. Note: Above entries are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:POPUlation? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:POPUlation? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:POPUlation? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString> =the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
PowerPowerItemResultsCurrentacqRms
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:RMS command.
Description
- This command queries the RMS value of the source selected for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:RMS? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:RMS? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:RMS? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
PowerPowerItemResultsCurrentacqStatus
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:STATUS command.
Description
- This command queries the status of the measurement for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:STATUS? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:STATUS? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:STATUS? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
PowerPowerItemResultsCurrentacqStddev
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:STDDev command.
Description
- This command queries the standard deviation value of the current acquisition for the
measurement parameter in the specified power measurement number. The power measurement
number is specified by x. Note: Above entries are
entries, and must be entered in enclosing quotes.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:STDDev? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:STDDev? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:STDDev? {InputPwr|Output1Pwr|Output2Pwr|Output3Pwr|Efficiency<x>|TotalEfficiency|INDUCT|IVSINTV|MAGLOSS|Bpeak|Br|Hc|Hmax|IRipple|DeltaB|DeltaH|Permeability|RDS|TRUEPWR|APPPWR|REPWR|PWRFACTOR|PHASE|PWRFREQ|ICFACTOR|VCFACTOR|IRMS|VRMS|TONENRG|TONLOSS|TOFFENRG|TOFFLOSS|CONDENRG|CONDLOSS|TTLLOSS|TTLENRG|DVBYDT|DIBYDT|SOAHITSCNT|LRIPRMS|LRIPPKPK|SWRIPRMS|SWRIPPKPK|PRIOD|FREQ|PDUTY|NDUTY|PPULSE|NPULSE|AMPL|PKPK|HIGH|LOW|MAX|MIN|INRUSH|CAPACITANCE|OUTPUT<x>|GAINCROSSOVERFREQ|PHASECROSSOVERFREQ|GM|PM|MAXPSRR|MAXPSRRFREQ|MINPSRR|MINPSRRFREQ}
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<QString>= the measurement result that you want to return from the specified power measurement number. Available results depend on the power measurement being taken in the specified measurement number. The validarguments are. InputPwr‘, ’ Output1Pwr ‘, ’ Output2Pwr ‘, ’ Output3Pwr ‘, ’ Efficiency1 ‘, ’ Efficiency2 ‘, ’ Efficiency3 ‘, ’ TotalEfficiency ‘.INDUCT‘.IVSINTV‘.MAGLOSS‘.Bpeak‘, ’ Br ‘, ’ Hc ‘, ’ Hmax ‘, ’ IRipple ‘, ’ DeltaB ‘, ’ DeltaH ‘, ’ Permeability ‘.RDS‘.TRUEPWR‘, ’ APPPWR ‘, ’ REPWR ‘, ’ PWRFACTOR ‘, ’ PHASE ‘, ’ PWRFREQ ‘, ’ ICFACTOR ‘, ’ VCFACTOR ‘, ’ IRMS ‘, ’ VRMS ‘.TONENRG‘, ’ TONLOSS ‘, ’ TOFFENRG ‘, ’ TOFFLOSS ‘, ’ CONDENRG ‘, ’ CONDLOSS ‘, ’ TTLLOSS ‘, ’ TTLENRG ‘.DVBYDT’ is the parameter for the dV by dt measurement.DIBYDT’ is the parameter for the dI by dt measurement.SOAHITSCNT’ is the parameter for the SOA measurement.LRIPRMS’ and ’ LRIPPKPK ’ are the parameters for the Line Ripple measurement.SWRIPRMS’ and ’ SWRIPPKPK ’ are the parameters for the Switching Ripple measurement.PRIOD’ is the parameter for the Cycle Period measurement.FREQ’ is the parameter for the Cycle Frequency measurement.PDUTY’ is the parameter for the Positive Duty Cycle measurement.NDUTY’ is the parameter for the Negative Duty Cycle measurement.PPULSE’ is the parameter for the Positive Pulse Width measurement.NPULSE’ is the parameter for the Negative Pulse Width measurement.AMPL’ is the parameter for the Cycle Amplitude measurement.PKPK’ is the parameter for the Cycle Peak-Peak measurement.HIGH’ is the parameter for the Cycle Top measurement.LOW’ is the parameter for the Cycle Base measurement.Max’ is the parameter for the Cycle Max measurement.MIN’ is the parameter for the Cycle Min measurement.INRUSH’ is the parameter for the Inrush Current measurement.CAPACITANCE’ is the parameter for the Input Capacitance measurement.OUTPUT1’ - ’ OUTPUT7 ’ are the parameters for the Turn On Time and Turn Off Time measurements.GAINCROSSOVERFREQ‘, ’ PHASECROSSOVERFREQ ‘, ’ GM ‘, ’ PM ’ are the parameters for the Control Loop Response measurement.MAXPSRR‘, ’ MAXPSRRFREQ ‘, ’ MINPSRR ‘, ’ MINPSRRFREQ ’ are the parameters for the PSRR measurement.
PowerPowerItemResultsCurrentacqThdf
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:THDF command.
Description
- This command queries the total harmonic distortion (fundamental) value for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:THDF? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:THDF? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:THDF? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
PowerPowerItemResultsCurrentacqThdr
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:THDR command.
Description
- This command queries the total harmonic distortion (RMS) value for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:THDR? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:THDR? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:THDR? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
PowerPowerItemResultsCurrentacqTrpwr
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:TRPWR command.
Description
- This command queries the true power value for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:TRPWR? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:TRPWR? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:TRPWR? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
PowerPowerItemResultsCurrentacqVrms
¶
Bases: SCPICmdReadWithArguments
The POWer:POWer<x>:RESUlts:CURRentacq:VRMS command.
Description
- This command queries the RMS voltage value for the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query(argument)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:VRMS? argumentquery. - Using the
.verify(argument, value)method will send thePOWer:POWer<x>:RESUlts:CURRentacq:VRMS? argumentquery and raise an AssertionError if the returned value does not matchvalue.
SCPI Syntax
- POWer:POWer<x>:RESUlts:CURRentacq:VRMS? harmonics
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
PowerPowerItemSeqsetup
¶
Bases: SCPICmdWrite
The POWer:POWer<x>:SEQSETup command.
Description
- This command sets up the instrument’s horizontal, vertical, and trigger parameters to optimize for taking the specified power measurement.
Usage
- Using the
.write(value)method will send thePOWer:POWer<x>:SEQSETup valuecommand.
SCPI Syntax
- POWer:POWer<x>:SEQSETup Execute
Info
POWer<x>is the Power measurement identifier number. The number must be for a power measurement that requires a single sequence acquisition.Executesets the measurement to run an acquisition and acquire data for the specified single sequence power measurement.
PowerPowerItemSequence
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SEQuence command.
Description
- This command sets or queries the run state of a single sequence power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SEQuence?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SEQuence?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SEQuence valuecommand.
SCPI Syntax
- POWer:POWer<x>:SEQuence {RUN|RERUN}
- POWer:POWer<x>:SEQuence?
Info
POWer<x>is the Power measurement identifier number. The number must be for a power measurement that requires a single sequence acquisition.RUNsets the measurement to run an acquisition and acquire data for the specified single sequence power measurement.RERUNsets the measurement to rerun an acquisition and acquire data for the specified single sequence power measurement.
PowerPowerItemSoa
¶
Bases: SCPICmdRead
The POWer:POWer<x>:SOA command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Properties
.isource: ThePOWer:POWer<x>:SOA:ISOURcecommand..point: ThePOWer:POWer<x>:SOA:POINT<x>command..recallmask: ThePOWer:POWer<x>:SOA:RECAllmaskcommand..savemask: ThePOWer:POWer<x>:SOA:SAVemaskcommand..vsource: ThePOWer:POWer<x>:SOA:VSOURcecommand.
isource
property
¶
isource: PowerPowerItemSoaIsource
Return the POWer:POWer<x>:SOA:ISOURce command.
Description
- This command sets or queries the current source for SOA measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SOA:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:SOA:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SOA:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
point
property
¶
point: dict[int, PowerPowerItemSoaPointItem]
Return the POWer:POWer<x>:SOA:POINT<x> command.
Description
- This command sets or queries the X or Y coordinate value for an SOA mask of a specified power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:POINT<x>?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:POINT<x>?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SOA:POINT<x> valuecommand.
SCPI Syntax
- POWer:POWer<x>:SOA:POINT<x> <NR1>
- POWer:POWer<x>:SOA:POINT<x>?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.POINT<x> = 1X | 2X | 3X | 4X | 5X | 6X | 7X | 8X | 9X | 10X | 11X | 12X | 13X |14X | 15X | 16X | 17X | 18X | 19X | 20X | 21X | 22X | 23X | 24X | 25X | 26X | 27X| 28X | 29X | 30X | 31X | 32X | 1Y | 2Y | 3Y | 4Y | 5Y | 6Y | 7Y | 8Y | 9Y | 10Y |11Y | 12Y | 13Y | 14Y | 15Y | 16Y | 17Y | 18Y | 19Y | 20Y | 21Y | 22Y | 23Y |24Y | 25Y | 26Y | 27Y | 28Y | 29Y | 30Y | 31Y | 32Y<NR1>sets the specified SOA mask X or Y point value, as a floating number, in the range from -5000 to 5000.
recallmask
property
¶
recallmask: PowerPowerItemSoaRecallmask
Return the POWer:POWer<x>:SOA:RECAllmask command.
Description
- This command recalls or queries the recall mask file name in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:RECAllmask?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:RECAllmask?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:SOA:RECAllmaskcommand.
SCPI Syntax
- POWer:POWer<x>:SOA:RECAllmask
- POWer:POWer<x>:SOA:RECAllmask?
Sub-properties
.filename: ThePOWer:POWer<x>:SOA:RECAllmask:FILENamecommand.
savemask
property
¶
savemask: PowerPowerItemSoaSavemask
Return the POWer:POWer<x>:SOA:SAVemask command.
Description
- This command saves the mask file as per the name configured and at the configured path or queries the mask file name, path, and file type for the SOA measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:SAVemask?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:SAVemask?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:SOA:SAVemaskcommand.
SCPI Syntax
- POWer:POWer<x>:SOA:SAVemask
- POWer:POWer<x>:SOA:SAVemask?
Sub-properties
.autoincrement: ThePOWer:POWer<x>:SOA:SAVemask:AUTOINCrementcommand..filename: ThePOWer:POWer<x>:SOA:SAVemask:FILENamecommand..folder: ThePOWer:POWer<x>:SOA:SAVemask:FOLDercommand.
vsource
property
¶
vsource: PowerPowerItemSoaVsource
Return the POWer:POWer<x>:SOA:VSOURce command.
Description
- This command sets or queries the voltage source for SOA measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SOA:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:SOA:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SOA:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemSoaIsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SOA:ISOURce command.
Description
- This command sets or queries the current source for SOA measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SOA:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:SOA:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SOA:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemSoaPointItem
¶
Bases: ValidatedDynamicNumberCmd, SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SOA:POINT<x> command.
Description
- This command sets or queries the X or Y coordinate value for an SOA mask of a specified power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:POINT<x>?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:POINT<x>?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SOA:POINT<x> valuecommand.
SCPI Syntax
- POWer:POWer<x>:SOA:POINT<x> <NR1>
- POWer:POWer<x>:SOA:POINT<x>?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.POINT<x> = 1X | 2X | 3X | 4X | 5X | 6X | 7X | 8X | 9X | 10X | 11X | 12X | 13X |14X | 15X | 16X | 17X | 18X | 19X | 20X | 21X | 22X | 23X | 24X | 25X | 26X | 27X| 28X | 29X | 30X | 31X | 32X | 1Y | 2Y | 3Y | 4Y | 5Y | 6Y | 7Y | 8Y | 9Y | 10Y |11Y | 12Y | 13Y | 14Y | 15Y | 16Y | 17Y | 18Y | 19Y | 20Y | 21Y | 22Y | 23Y |24Y | 25Y | 26Y | 27Y | 28Y | 29Y | 30Y | 31Y | 32Y<NR1>sets the specified SOA mask X or Y point value, as a floating number, in the range from -5000 to 5000.
PowerPowerItemSoaRecallmask
¶
Bases: SCPICmdWriteNoArguments, SCPICmdRead
The POWer:POWer<x>:SOA:RECAllmask command.
Description
- This command recalls or queries the recall mask file name in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:RECAllmask?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:RECAllmask?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:SOA:RECAllmaskcommand.
SCPI Syntax
- POWer:POWer<x>:SOA:RECAllmask
- POWer:POWer<x>:SOA:RECAllmask?
Properties
.filename: ThePOWer:POWer<x>:SOA:RECAllmask:FILENamecommand.
filename
property
¶
filename: PowerPowerItemSoaRecallmaskFilename
Return the POWer:POWer<x>:SOA:RECAllmask:FILEName command.
Description
- This command sets or queries the file name for saving SOA mask file name in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:RECAllmask:FILEName?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:RECAllmask:FILEName?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:SOA:RECAllmask:FILENamecommand.
SCPI Syntax
- POWer:POWer<x>:SOA:RECAllmask:FILEName
- POWer:POWer<x>:SOA:RECAllmask:FILEName?
PowerPowerItemSoaRecallmaskFilename
¶
Bases: SCPICmdWriteNoArguments, SCPICmdRead
The POWer:POWer<x>:SOA:RECAllmask:FILEName command.
Description
- This command sets or queries the file name for saving SOA mask file name in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:RECAllmask:FILEName?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:RECAllmask:FILEName?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:SOA:RECAllmask:FILENamecommand.
SCPI Syntax
- POWer:POWer<x>:SOA:RECAllmask:FILEName
- POWer:POWer<x>:SOA:RECAllmask:FILEName?
PowerPowerItemSoaSavemask
¶
Bases: SCPICmdWriteNoArguments, SCPICmdRead
The POWer:POWer<x>:SOA:SAVemask command.
Description
- This command saves the mask file as per the name configured and at the configured path or queries the mask file name, path, and file type for the SOA measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:SAVemask?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:SAVemask?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:SOA:SAVemaskcommand.
SCPI Syntax
- POWer:POWer<x>:SOA:SAVemask
- POWer:POWer<x>:SOA:SAVemask?
Properties
.autoincrement: ThePOWer:POWer<x>:SOA:SAVemask:AUTOINCrementcommand..filename: ThePOWer:POWer<x>:SOA:SAVemask:FILENamecommand..folder: ThePOWer:POWer<x>:SOA:SAVemask:FOLDercommand.
autoincrement
property
¶
autoincrement: PowerPowerItemSoaSavemaskAutoincrement
Return the POWer:POWer<x>:SOA:SAVemask:AUTOINCrement command.
Description
- This command sets or queries the state of auto-increment for saved SOA mask file names in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:SAVemask:AUTOINCrement?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:SAVemask:AUTOINCrement?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:SOA:SAVemask:AUTOINCrementcommand.
SCPI Syntax
- POWer:POWer<x>:SOA:SAVemask:AUTOINCrement
- POWer:POWer<x>:SOA:SAVemask:AUTOINCrement?
filename
property
¶
filename: PowerPowerItemSoaSavemaskFilename
Return the POWer:POWer<x>:SOA:SAVemask:FILEName command.
Description
- This command sets or queries the mask file name for SOA measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:SAVemask:FILEName?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:SAVemask:FILEName?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:SOA:SAVemask:FILENamecommand.
SCPI Syntax
- POWer:POWer<x>:SOA:SAVemask:FILEName
- POWer:POWer<x>:SOA:SAVemask:FILEName?
folder
property
¶
folder: PowerPowerItemSoaSavemaskFolder
Return the POWer:POWer<x>:SOA:SAVemask:FOLDer command.
Description
- This command sets or queries the mask file folder path for SOA measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:SAVemask:FOLDer?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:SAVemask:FOLDer?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:SOA:SAVemask:FOLDercommand.
SCPI Syntax
- POWer:POWer<x>:SOA:SAVemask:FOLDer
- POWer:POWer<x>:SOA:SAVemask:FOLDer?
PowerPowerItemSoaSavemaskAutoincrement
¶
Bases: SCPICmdWriteNoArguments, SCPICmdRead
The POWer:POWer<x>:SOA:SAVemask:AUTOINCrement command.
Description
- This command sets or queries the state of auto-increment for saved SOA mask file names in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:SAVemask:AUTOINCrement?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:SAVemask:AUTOINCrement?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:SOA:SAVemask:AUTOINCrementcommand.
SCPI Syntax
- POWer:POWer<x>:SOA:SAVemask:AUTOINCrement
- POWer:POWer<x>:SOA:SAVemask:AUTOINCrement?
PowerPowerItemSoaSavemaskFilename
¶
Bases: SCPICmdWriteNoArguments, SCPICmdRead
The POWer:POWer<x>:SOA:SAVemask:FILEName command.
Description
- This command sets or queries the mask file name for SOA measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:SAVemask:FILEName?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:SAVemask:FILEName?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:SOA:SAVemask:FILENamecommand.
SCPI Syntax
- POWer:POWer<x>:SOA:SAVemask:FILEName
- POWer:POWer<x>:SOA:SAVemask:FILEName?
PowerPowerItemSoaSavemaskFolder
¶
Bases: SCPICmdWriteNoArguments, SCPICmdRead
The POWer:POWer<x>:SOA:SAVemask:FOLDer command.
Description
- This command sets or queries the mask file folder path for SOA measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:SAVemask:FOLDer?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:SAVemask:FOLDer?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:SOA:SAVemask:FOLDercommand.
SCPI Syntax
- POWer:POWer<x>:SOA:SAVemask:FOLDer
- POWer:POWer<x>:SOA:SAVemask:FOLDer?
PowerPowerItemSoaVsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SOA:VSOURce command.
Description
- This command sets or queries the voltage source for SOA measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SOA:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SOA:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SOA:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:SOA:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SOA:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemSwitchingloss
¶
Bases: SCPICmdRead
The POWer:POWer<x>:SWITCHINGLOSS command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS?query and raise an AssertionError if the returned value does not matchvalue.
Properties
.devicetype: ThePOWer:POWer<x>:SWITCHINGLOSS:DEVICETypecommand..gatesource: ThePOWer:POWer<x>:SWITCHINGLOSS:GATESOurcecommand..ilevelabs: ThePOWer:POWer<x>:SWITCHINGLOSS:ILEVELAbscommand..ilevelpct: ThePOWer:POWer<x>:SWITCHINGLOSS:ILEVELPctcommand..isource: ThePOWer:POWer<x>:SWITCHINGLOSS:ISOURcecommand..levelunits: ThePOWer:POWer<x>:SWITCHINGLOSS:LEVELUNItscommand..rdson: ThePOWer:POWer<x>:SWITCHINGLOSS:RDSOncommand..swlconfigtype: ThePOWer:POWer<x>:SWITCHINGLOSS:SWLCONFIGTypecommand..vcesat: ThePOWer:POWer<x>:SWITCHINGLOSS:VCESatcommand..vglevel: ThePOWer:POWer<x>:SWITCHINGLOSS:VGLevelcommand..vlevelabs: ThePOWer:POWer<x>:SWITCHINGLOSS:VLEVELAbscommand..vlevelpct: ThePOWer:POWer<x>:SWITCHINGLOSS:VLEVELPctcommand..vsource: ThePOWer:POWer<x>:SWITCHINGLOSS:VSOURcecommand.
devicetype
property
¶
devicetype: PowerPowerItemSwitchinglossDevicetype
Return the POWer:POWer<x>:SWITCHINGLOSS:DEVICEType command.
Description
- This command sets or queries the conduction calculation method for switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:DEVICEType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:DEVICEType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:DEVICEType valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:DEVICEType {MOSFET|BJT}
- POWer:POWer<x>:SWITCHINGLOSS:DEVICEType?
gatesource
property
¶
gatesource: PowerPowerItemSwitchinglossGatesource
Return the POWer:POWer<x>:SWITCHINGLOSS:GATESOurce command.
Description
- This command sets or queries the gate voltage (Vg) for the switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:GATESOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:GATESOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:GATESOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:GATESOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SWITCHINGLOSS:GATESOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
ilevelabs
property
¶
ilevelabs: PowerPowerItemSwitchinglossIlevelabs
Return the POWer:POWer<x>:SWITCHINGLOSS:ILEVELAbs command.
Description
- This command sets or queries the current level (Ton-Start & Stop) in absolute units for switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:ILEVELAbs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:ILEVELAbs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:ILEVELAbs valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:ILEVELAbs <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:ILEVELAbs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -100 to 100.
ilevelpct
property
¶
ilevelpct: PowerPowerItemSwitchinglossIlevelpct
Return the POWer:POWer<x>:SWITCHINGLOSS:ILEVELPct command.
Description
- This command sets or queries the current level (Ton-Start & Stop) in percentage for switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:ILEVELPct?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:ILEVELPct?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:ILEVELPct valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:ILEVELPct <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:ILEVELPct?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.ranges from 0.0001 to 90.
isource
property
¶
isource: PowerPowerItemSwitchinglossIsource
Return the POWer:POWer<x>:SWITCHINGLOSS:ISOURce command.
Description
- This command sets or queries the current source for the switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SWITCHINGLOSS:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
levelunits
property
¶
levelunits: PowerPowerItemSwitchinglossLevelunits
Return the POWer:POWer<x>:SWITCHINGLOSS:LEVELUNIts command.
Description
- This command sets or queries the level units for switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:LEVELUNIts?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:LEVELUNIts?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:LEVELUNIts valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:LEVELUNIts {PERCent|ABSolute}
- POWer:POWer<x>:SWITCHINGLOSS:LEVELUNIts?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.PERCentto set the High, Mid, and Low reference levels in percentage.ABSoluteto set the High, Mid, and Low reference levels to specific signal levels.
rdson
property
¶
Return the POWer:POWer<x>:SWITCHINGLOSS:RDSOn command.
Description
- This command sets or queries the RDS(on) value for switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:RDSOn?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:RDSOn?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:RDSOn valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:RDSOn <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:RDSOn?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 0 to 100.
swlconfigtype
property
¶
swlconfigtype: PowerPowerItemSwitchinglossSwlconfigtype
Return the POWer:POWer<x>:SWITCHINGLOSS:SWLCONFIGType command.
Description
- This command sets or queries the configuration type for the switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:SWLCONFIGType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:SWLCONFIGType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:SWLCONFIGType valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:SWLCONFIGType {SMPS|PFC|FLYBACK}
- POWer:POWer<x>:SWITCHINGLOSS:SWLCONFIGType?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.SMPS: Select this option in case of signals without noise and ringing. The V.PFC: Select this option when input DUT signals are from Power Factor Correction Circuit. For this case, V.FLYBACK: Select this option when input signals are ringing. This option does not require a V.
vcesat
property
¶
Return the POWer:POWer<x>:SWITCHINGLOSS:VCESat command.
Description
- This command sets or queries the value for the VCE(sat) value for switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:VCESat?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VCESat?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VCESat valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:VCESat <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:VCESat?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 0.001 to 100.
vglevel
property
¶
vglevel: PowerPowerItemSwitchinglossVglevel
Return the POWer:POWer<x>:SWITCHINGLOSS:VGLevel command.
Description
- This command sets or queries the gate voltage value (V g Level Ton-Start) for the switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:VGLevel?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VGLevel?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VGLevel valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:VGLevel <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:VGLevel?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -100 to 100.
vlevelabs
property
¶
vlevelabs: PowerPowerItemSwitchinglossVlevelabs
Return the POWer:POWer<x>:SWITCHINGLOSS:VLEVELAbs command.
Description
- This command sets or queries the voltage level (Ton-Start & Stop) in absolute units for switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:VLEVELAbs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VLEVELAbs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VLEVELAbs valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:VLEVELAbs <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:VLEVELAbs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -100 to 100.
vlevelpct
property
¶
vlevelpct: PowerPowerItemSwitchinglossVlevelpct
Return the POWer:POWer<x>:SWITCHINGLOSS:VLEVELPct command.
Description
- This command sets or queries the voltage level (Ton-Start & Stop) in percentage for switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:VLEVELPct?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VLEVELPct?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VLEVELPct valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:VLEVELPct <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:VLEVELPct?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 0.0001 to 90.
vsource
property
¶
vsource: PowerPowerItemSwitchinglossVsource
Return the POWer:POWer<x>:SWITCHINGLOSS:VSOURce command.
Description
- This command sets or queries the voltage source for the switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SWITCHINGLOSS:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemSwitchinglossDevicetype
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SWITCHINGLOSS:DEVICEType command.
Description
- This command sets or queries the conduction calculation method for switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:DEVICEType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:DEVICEType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:DEVICEType valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:DEVICEType {MOSFET|BJT}
- POWer:POWer<x>:SWITCHINGLOSS:DEVICEType?
PowerPowerItemSwitchinglossGatesource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SWITCHINGLOSS:GATESOurce command.
Description
- This command sets or queries the gate voltage (Vg) for the switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:GATESOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:GATESOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:GATESOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:GATESOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SWITCHINGLOSS:GATESOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemSwitchinglossIlevelabs
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SWITCHINGLOSS:ILEVELAbs command.
Description
- This command sets or queries the current level (Ton-Start & Stop) in absolute units for switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:ILEVELAbs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:ILEVELAbs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:ILEVELAbs valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:ILEVELAbs <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:ILEVELAbs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -100 to 100.
PowerPowerItemSwitchinglossIlevelpct
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SWITCHINGLOSS:ILEVELPct command.
Description
- This command sets or queries the current level (Ton-Start & Stop) in percentage for switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:ILEVELPct?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:ILEVELPct?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:ILEVELPct valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:ILEVELPct <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:ILEVELPct?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.ranges from 0.0001 to 90.
PowerPowerItemSwitchinglossIsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SWITCHINGLOSS:ISOURce command.
Description
- This command sets or queries the current source for the switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:ISOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:ISOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:ISOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:ISOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SWITCHINGLOSS:ISOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemSwitchinglossLevelunits
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SWITCHINGLOSS:LEVELUNIts command.
Description
- This command sets or queries the level units for switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:LEVELUNIts?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:LEVELUNIts?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:LEVELUNIts valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:LEVELUNIts {PERCent|ABSolute}
- POWer:POWer<x>:SWITCHINGLOSS:LEVELUNIts?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.PERCentto set the High, Mid, and Low reference levels in percentage.ABSoluteto set the High, Mid, and Low reference levels to specific signal levels.
PowerPowerItemSwitchinglossRdson
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SWITCHINGLOSS:RDSOn command.
Description
- This command sets or queries the RDS(on) value for switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:RDSOn?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:RDSOn?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:RDSOn valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:RDSOn <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:RDSOn?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 0 to 100.
PowerPowerItemSwitchinglossSwlconfigtype
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SWITCHINGLOSS:SWLCONFIGType command.
Description
- This command sets or queries the configuration type for the switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:SWLCONFIGType?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:SWLCONFIGType?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:SWLCONFIGType valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:SWLCONFIGType {SMPS|PFC|FLYBACK}
- POWer:POWer<x>:SWITCHINGLOSS:SWLCONFIGType?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.SMPS: Select this option in case of signals without noise and ringing. The V.PFC: Select this option when input DUT signals are from Power Factor Correction Circuit. For this case, V.FLYBACK: Select this option when input signals are ringing. This option does not require a V.
PowerPowerItemSwitchinglossVcesat
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SWITCHINGLOSS:VCESat command.
Description
- This command sets or queries the value for the VCE(sat) value for switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:VCESat?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VCESat?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VCESat valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:VCESat <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:VCESat?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 0.001 to 100.
PowerPowerItemSwitchinglossVglevel
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SWITCHINGLOSS:VGLevel command.
Description
- This command sets or queries the gate voltage value (V g Level Ton-Start) for the switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:VGLevel?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VGLevel?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VGLevel valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:VGLevel <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:VGLevel?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -100 to 100.
PowerPowerItemSwitchinglossVlevelabs
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SWITCHINGLOSS:VLEVELAbs command.
Description
- This command sets or queries the voltage level (Ton-Start & Stop) in absolute units for switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:VLEVELAbs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VLEVELAbs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VLEVELAbs valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:VLEVELAbs <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:VLEVELAbs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from -100 to 100.
PowerPowerItemSwitchinglossVlevelpct
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SWITCHINGLOSS:VLEVELPct command.
Description
- This command sets or queries the voltage level (Ton-Start & Stop) in percentage for switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:VLEVELPct?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VLEVELPct?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VLEVELPct valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:VLEVELPct <NR1>
- POWer:POWer<x>:SWITCHINGLOSS:VLEVELPct?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 0.0001 to 90.
PowerPowerItemSwitchinglossVsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SWITCHINGLOSS:VSOURce command.
Description
- This command sets or queries the voltage source for the switching loss measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGLOSS:VSOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VSOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGLOSS:VSOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGLOSS:VSOURce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SWITCHINGLOSS:VSOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemSwitchingripple
¶
Bases: SCPICmdRead
The POWer:POWer<x>:SWITCHINGRIPPLE command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGRIPPLE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGRIPPLE?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.
Properties
.inputsource: ThePOWer:POWer<x>:SWITCHINGRIPPLE:INPUTSOurcecommand..lfrequency: ThePOWer:POWer<x>:SWITCHINGRIPPLE:LFREQuencycommand.
inputsource
property
¶
inputsource: PowerPowerItemSwitchingrippleInputsource
Return the POWer:POWer<x>:SWITCHINGRIPPLE:INPUTSOurce command.
Description
- This command sets or queries the input source for switching ripple measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGRIPPLE:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGRIPPLE:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGRIPPLE:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGRIPPLE:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SWITCHINGRIPPLE:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
lfrequency
property
¶
lfrequency: PowerPowerItemSwitchingrippleLfrequency
Return the POWer:POWer<x>:SWITCHINGRIPPLE:LFREQuency command.
Description
- This command sets or queries the switching frequency for switching ripple measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGRIPPLE:LFREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGRIPPLE:LFREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGRIPPLE:LFREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGRIPPLE:LFREQuency <NR1>
- POWer:POWer<x>:SWITCHINGRIPPLE:LFREQuency?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 50 to 1000000.
PowerPowerItemSwitchingrippleInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SWITCHINGRIPPLE:INPUTSOurce command.
Description
- This command sets or queries the input source for switching ripple measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGRIPPLE:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGRIPPLE:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGRIPPLE:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGRIPPLE:INPUTSOurce {CH<x>|MATH<x>|REF<x>}
- POWer:POWer<x>:SWITCHINGRIPPLE:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemSwitchingrippleLfrequency
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:SWITCHINGRIPPLE:LFREQuency command.
Description
- This command sets or queries the switching frequency for switching ripple measurement in the specified power measurement number. The power measurement number is specified by x.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:SWITCHINGRIPPLE:LFREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:SWITCHINGRIPPLE:LFREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:SWITCHINGRIPPLE:LFREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:SWITCHINGRIPPLE:LFREQuency <NR1>
- POWer:POWer<x>:SWITCHINGRIPPLE:LFREQuency?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR1>ranges from 50 to 1000000.
PowerPowerItemTurnofftime
¶
Bases: SCPICmdRead
The POWer:POWer<x>:TURNOFFtime command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Properties
.frequency: ThePOWer:POWer<x>:TURNOFFtime:FREQuencycommand..inputlevel: ThePOWer:POWer<x>:TURNOFFtime:INPUTLEVelcommand..inputsource: ThePOWer:POWer<x>:TURNOFFtime:INPUTSOurcecommand..maxtime: ThePOWer:POWer<x>:TURNOFFtime:MAXTIMecommand..maxvoltage: ThePOWer:POWer<x>:TURNOFFtime:MAXVoltagecommand..numoutputs: ThePOWer:POWer<x>:TURNOFFtime:NUMOUTputscommand..output1source: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT1SOURcecommand..output1voltage: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT1VOLTagecommand..output2source: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT2SOURcecommand..output2voltage: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT2VOLTagecommand..output3source: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT3SOURcecommand..output3voltage: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT3VOLTagecommand..output4source: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT4SOURcecommand..output4voltage: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT4VOLTagecommand..output5source: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT5SOURcecommand..output5voltage: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT5VOLTagecommand..output6source: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT6SOURcecommand..output6voltage: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT6VOLTagecommand..output7source: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT7SOURcecommand..output7voltage: ThePOWer:POWer<x>:TURNOFFtime:OUTPUT7VOLTagecommand..type: ThePOWer:POWer<x>:TURNOFFtime:TYPEcommand.
frequency
property
¶
frequency: PowerPowerItemTurnofftimeFrequency
Return the POWer:POWer<x>:TURNOFFtime:FREQuency command.
Description
- This command sets or queries the input frequency used by the AC or DC converter of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:FREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:FREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:FREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:FREQuency <NR3>
- POWer:POWer<x>:TURNOFFtime:FREQuency?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the frequency, in Hertz, from 1 Hz to 500 Hz.
inputlevel
property
¶
inputlevel: PowerPowerItemTurnofftimeInputlevel
Return the POWer:POWer<x>:TURNOFFtime:INPUTLEVel command.
Description
- This command sets or queries the input voltage level of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:INPUTLEVel?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:INPUTLEVel?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:INPUTLEVel valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:INPUTLEVel <NR3>
- POWer:POWer<x>:TURNOFFtime:INPUTLEVel?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the voltage level, in volts, from -500 V to 500 V.
inputsource
property
¶
inputsource: PowerPowerItemTurnofftimeInputsource
Return the POWer:POWer<x>:TURNOFFtime:INPUTSOurce command.
Description
- This command sets or queries the input source of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:INPUTSOurce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
maxtime
property
¶
maxtime: PowerPowerItemTurnofftimeMaxtime
Return the POWer:POWer<x>:TURNOFFtime:MAXTIMe command.
Description
- This command sets or queries the maximum turn off time of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:MAXTIMe?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:MAXTIMe?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:MAXTIMe valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:MAXTIMe <NR3>
- POWer:POWer<x>:TURNOFFtime:MAXTIMe?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the maximum time value, in seconds, in the range 1 second to 500 seconds.
maxvoltage
property
¶
maxvoltage: PowerPowerItemTurnofftimeMaxvoltage
Return the POWer:POWer<x>:TURNOFFtime:MAXVoltage command.
Description
- This command sets or queries the maximum voltage of the specified Turn OffTime measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:MAXVoltage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:MAXVoltage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:MAXVoltage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:MAXVoltage <NR3>
- POWer:POWer<x>:TURNOFFtime:MAXVoltage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the maximum voltage in the range 1 V to 500 V.
numoutputs
property
¶
numoutputs: PowerPowerItemTurnofftimeNumoutputs
Return the POWer:POWer<x>:TURNOFFtime:NUMOUTputs command.
Description
- This command sets or queries the number of outputs of the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:NUMOUTputs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:NUMOUTputs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:NUMOUTputs valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:NUMOUTputs {ONE|TWO|THREE|FOUR|FIVE|SIX|SEVEN}
- POWer:POWer<x>:TURNOFFtime:NUMOUTputs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.ONEthrough SEVEN sets the number of outputs for the Turn Off Time measurement.
output1source
property
¶
output1source: PowerPowerItemTurnofftimeOutput1source
Return the POWer:POWer<x>:TURNOFFtime:OUTPUT1SOURce command.
Description
- This command sets or queries the output 1 source of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT1SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT1SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT1SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT1SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT1SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
output1voltage
property
¶
output1voltage: PowerPowerItemTurnofftimeOutput1voltage
Return the POWer:POWer<x>:TURNOFFtime:OUTPUT1VOLTage command.
Description
- This command sets or queries the output 1 voltage level of the of the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT1VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT1VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT1VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT1VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT1VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
output2source
property
¶
output2source: PowerPowerItemTurnofftimeOutput2source
Return the POWer:POWer<x>:TURNOFFtime:OUTPUT2SOURce command.
Description
- This command sets or queries the output 2 source of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT2SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT2SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT2SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT2SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT2SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
output2voltage
property
¶
output2voltage: PowerPowerItemTurnofftimeOutput2voltage
Return the POWer:POWer<x>:TURNOFFtime:OUTPUT2VOLTage command.
Description
- This command sets or queries the output 2 voltage level of the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT2VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT2VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT2VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT2VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT2VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
output3source
property
¶
output3source: PowerPowerItemTurnofftimeOutput3source
Return the POWer:POWer<x>:TURNOFFtime:OUTPUT3SOURce command.
Description
- This command sets or queries the output 3 source of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT3SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT3SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT3SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT3SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT3SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>A channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>A Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>A Math waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
output3voltage
property
¶
output3voltage: PowerPowerItemTurnofftimeOutput3voltage
Return the POWer:POWer<x>:TURNOFFtime:OUTPUT3VOLTage command.
Description
- This command sets or queries the output 3 voltage level of the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT3VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT3VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT3VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT3VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT3VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
output4source
property
¶
output4source: PowerPowerItemTurnofftimeOutput4source
Return the POWer:POWer<x>:TURNOFFtime:OUTPUT4SOURce command.
Description
- This command sets or queries the output 4 source of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT4SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT4SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT4SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT4SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT4SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
output4voltage
property
¶
output4voltage: PowerPowerItemTurnofftimeOutput4voltage
Return the POWer:POWer<x>:TURNOFFtime:OUTPUT4VOLTage command.
Description
- This command sets or queries the output 4 voltage level of the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT4VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT4VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT4VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT4VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT4VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
output5source
property
¶
output5source: PowerPowerItemTurnofftimeOutput5source
Return the POWer:POWer<x>:TURNOFFtime:OUTPUT5SOURce command.
Description
- This command sets or queries the output 5 source of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT5SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT5SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT5SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT5SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT5SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
output5voltage
property
¶
output5voltage: PowerPowerItemTurnofftimeOutput5voltage
Return the POWer:POWer<x>:TURNOFFtime:OUTPUT5VOLTage command.
Description
- This command sets or queries the output 5 voltage level of the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT5VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT5VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT5VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT5VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT5VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
output6source
property
¶
output6source: PowerPowerItemTurnofftimeOutput6source
Return the POWer:POWer<x>:TURNOFFtime:OUTPUT6SOURce command.
Description
- This command sets or queries the output 6 source of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT6SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT6SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT6SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT6SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT6SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
output6voltage
property
¶
output6voltage: PowerPowerItemTurnofftimeOutput6voltage
Return the POWer:POWer<x>:TURNOFFtime:OUTPUT6VOLTage command.
Description
- This command sets or queries the output 6 voltage level of the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT6VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT6VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT6VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT6VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT6VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
output7source
property
¶
output7source: PowerPowerItemTurnofftimeOutput7source
Return the POWer:POWer<x>:TURNOFFtime:OUTPUT7SOURce command.
Description
- This command sets or queries the output 7 source of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT7SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT7SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT7SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT7SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT7SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
output7voltage
property
¶
output7voltage: PowerPowerItemTurnofftimeOutput7voltage
Return the POWer:POWer<x>:TURNOFFtime:OUTPUT7VOLTage command.
Description
- This command sets or queries the output 7 voltage level of the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT7VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT7VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT7VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT7VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT7VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
type
property
¶
Return the POWer:POWer<x>:TURNOFFtime:TYPE command.
Description
- This command sets or queries the type of AC/DC converter used in the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:TYPE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:TYPE?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:TYPE valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:TYPE {DCDC|ACDC}
- POWer:POWer<x>:TURNOFFtime:TYPE?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.DCDCsets the measurement to use a DC to DC converter.ACDCsets the measurement to use an AC to DC converter.
PowerPowerItemTurnofftimeFrequency
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:FREQuency command.
Description
- This command sets or queries the input frequency used by the AC or DC converter of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:FREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:FREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:FREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:FREQuency <NR3>
- POWer:POWer<x>:TURNOFFtime:FREQuency?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the frequency, in Hertz, from 1 Hz to 500 Hz.
PowerPowerItemTurnofftimeInputlevel
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:INPUTLEVel command.
Description
- This command sets or queries the input voltage level of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:INPUTLEVel?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:INPUTLEVel?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:INPUTLEVel valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:INPUTLEVel <NR3>
- POWer:POWer<x>:TURNOFFtime:INPUTLEVel?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the voltage level, in volts, from -500 V to 500 V.
PowerPowerItemTurnofftimeInputsource
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:INPUTSOurce command.
Description
- This command sets or queries the input source of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:INPUTSOurce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:INPUTSOurce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemTurnofftimeMaxtime
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:MAXTIMe command.
Description
- This command sets or queries the maximum turn off time of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:MAXTIMe?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:MAXTIMe?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:MAXTIMe valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:MAXTIMe <NR3>
- POWer:POWer<x>:TURNOFFtime:MAXTIMe?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the maximum time value, in seconds, in the range 1 second to 500 seconds.
PowerPowerItemTurnofftimeMaxvoltage
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:MAXVoltage command.
Description
- This command sets or queries the maximum voltage of the specified Turn OffTime measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:MAXVoltage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:MAXVoltage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:MAXVoltage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:MAXVoltage <NR3>
- POWer:POWer<x>:TURNOFFtime:MAXVoltage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the maximum voltage in the range 1 V to 500 V.
PowerPowerItemTurnofftimeNumoutputs
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:NUMOUTputs command.
Description
- This command sets or queries the number of outputs of the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:NUMOUTputs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:NUMOUTputs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:NUMOUTputs valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:NUMOUTputs {ONE|TWO|THREE|FOUR|FIVE|SIX|SEVEN}
- POWer:POWer<x>:TURNOFFtime:NUMOUTputs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.ONEthrough SEVEN sets the number of outputs for the Turn Off Time measurement.
PowerPowerItemTurnofftimeOutput1source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:OUTPUT1SOURce command.
Description
- This command sets or queries the output 1 source of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT1SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT1SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT1SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT1SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT1SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemTurnofftimeOutput1voltage
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:OUTPUT1VOLTage command.
Description
- This command sets or queries the output 1 voltage level of the of the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT1VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT1VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT1VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT1VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT1VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
PowerPowerItemTurnofftimeOutput2source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:OUTPUT2SOURce command.
Description
- This command sets or queries the output 2 source of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT2SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT2SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT2SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT2SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT2SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemTurnofftimeOutput2voltage
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:OUTPUT2VOLTage command.
Description
- This command sets or queries the output 2 voltage level of the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT2VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT2VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT2VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT2VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT2VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
PowerPowerItemTurnofftimeOutput3source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:OUTPUT3SOURce command.
Description
- This command sets or queries the output 3 source of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT3SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT3SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT3SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT3SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT3SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>A channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>A Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>A Math waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
PowerPowerItemTurnofftimeOutput3voltage
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:OUTPUT3VOLTage command.
Description
- This command sets or queries the output 3 voltage level of the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT3VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT3VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT3VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT3VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT3VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
PowerPowerItemTurnofftimeOutput4source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:OUTPUT4SOURce command.
Description
- This command sets or queries the output 4 source of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT4SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT4SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT4SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT4SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT4SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemTurnofftimeOutput4voltage
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:OUTPUT4VOLTage command.
Description
- This command sets or queries the output 4 voltage level of the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT4VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT4VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT4VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT4VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT4VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
PowerPowerItemTurnofftimeOutput5source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:OUTPUT5SOURce command.
Description
- This command sets or queries the output 5 source of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT5SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT5SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT5SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT5SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT5SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemTurnofftimeOutput5voltage
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:OUTPUT5VOLTage command.
Description
- This command sets or queries the output 5 voltage level of the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT5VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT5VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT5VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT5VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT5VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
PowerPowerItemTurnofftimeOutput6source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:OUTPUT6SOURce command.
Description
- This command sets or queries the output 6 source of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT6SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT6SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT6SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT6SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT6SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemTurnofftimeOutput6voltage
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:OUTPUT6VOLTage command.
Description
- This command sets or queries the output 6 voltage level of the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT6VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT6VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT6VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT6VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT6VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
PowerPowerItemTurnofftimeOutput7source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:OUTPUT7SOURce command.
Description
- This command sets or queries the output 7 source of the specified Turn Off Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT7SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT7SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT7SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT7SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNOFFtime:OUTPUT7SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
PowerPowerItemTurnofftimeOutput7voltage
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:OUTPUT7VOLTage command.
Description
- This command sets or queries the output 7 voltage level of the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT7VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT7VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:OUTPUT7VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:OUTPUT7VOLTage <NR2>
- POWer:POWer<x>:TURNOFFtime:OUTPUT7VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
PowerPowerItemTurnofftimeType
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNOFFtime:TYPE command.
Description
- This command sets or queries the type of AC/DC converter used in the specified Turn Off Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNOFFtime:TYPE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNOFFtime:TYPE?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNOFFtime:TYPE valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNOFFtime:TYPE {DCDC|ACDC}
- POWer:POWer<x>:TURNOFFtime:TYPE?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.DCDCsets the measurement to use a DC to DC converter.ACDCsets the measurement to use an AC to DC converter.
PowerPowerItemTurnontime
¶
Bases: SCPICmdRead
The POWer:POWer<x>:TURNONtime command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.
Properties
.frequency: ThePOWer:POWer<x>:TURNONtime:FREQuencycommand..inputlevel: ThePOWer:POWer<x>:TURNONtime:INPUTLEVelcommand..inputsource: ThePOWer:POWer<x>:TURNONtime:INPUTSOurcecommand..maxtime: ThePOWer:POWer<x>:TURNONtime:MAXTIMecommand..maxvoltage: ThePOWer:POWer<x>:TURNONtime:MAXVoltagecommand..numoutputs: ThePOWer:POWer<x>:TURNONtime:NUMOUTputscommand..output1source: ThePOWer:POWer<x>:TURNONtime:OUTPUT1SOURcecommand..output1voltage: ThePOWer:POWer<x>:TURNONtime:OUTPUT1VOLTagecommand..output2source: ThePOWer:POWer<x>:TURNONtime:OUTPUT2SOURcecommand..output2voltage: ThePOWer:POWer<x>:TURNONtime:OUTPUT2VOLTagecommand..output3source: ThePOWer:POWer<x>:TURNONtime:OUTPUT3SOURcecommand..output3voltage: ThePOWer:POWer<x>:TURNONtime:OUTPUT3VOLTagecommand..output4source: ThePOWer:POWer<x>:TURNONtime:OUTPUT4SOURcecommand..output4voltage: ThePOWer:POWer<x>:TURNONtime:OUTPUT4VOLTagecommand..output5source: ThePOWer:POWer<x>:TURNONtime:OUTPUT5SOURcecommand..output5voltage: ThePOWer:POWer<x>:TURNONtime:OUTPUT5VOLTagecommand..output6source: ThePOWer:POWer<x>:TURNONtime:OUTPUT6SOURcecommand..output6voltage: ThePOWer:POWer<x>:TURNONtime:OUTPUT6VOLTagecommand..output7source: ThePOWer:POWer<x>:TURNONtime:OUTPUT7SOURcecommand..output7voltage: ThePOWer:POWer<x>:TURNONtime:OUTPUT7VOLTagecommand..type: ThePOWer:POWer<x>:TURNONtime:TYPEcommand.
frequency
property
¶
frequency: PowerPowerItemTurnontimeFrequency
Return the POWer:POWer<x>:TURNONtime:FREQuency command.
Description
- This command sets or queries the input frequency used by the AC or DC converter of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:FREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:FREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:FREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:FREQuency <NR3>
- POWer:POWer<x>:TURNONtime:FREQuency?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the frequency, in Hertz, from 1 Hz to 500 Hz.
inputlevel
property
¶
inputlevel: PowerPowerItemTurnontimeInputlevel
Return the POWer:POWer<x>:TURNONtime:INPUTLEVel command.
Description
- This command sets or returns the input voltage level of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:INPUTLEVel?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:INPUTLEVel?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:INPUTLEVel valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:INPUTLEVel <NR3>
- POWer:POWer<x>:TURNONtime:INPUTLEVel?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the voltage level, in volts, from -500 V to 500 V.
inputsource
property
¶
inputsource: PowerPowerItemTurnontimeInputsource
Return the POWer:POWer<x>:TURNONtime:INPUTSOurce command.
Description
- This command sets or queries the input source of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:TURNONtime:INPUTSOurcecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:INPUTSOurce
- POWer:POWer<x>:TURNONtime:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
maxtime
property
¶
maxtime: PowerPowerItemTurnontimeMaxtime
Return the POWer:POWer<x>:TURNONtime:MAXTIMe command.
Description
- This command sets or returns the maximum turn on time of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:MAXTIMe?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:MAXTIMe?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:MAXTIMe valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:MAXTIMe <NR3>
- POWer:POWer<x>:TURNONtime:MAXTIMe?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the maximum time value, in seconds, in the range 1 second to 500 seconds.
maxvoltage
property
¶
maxvoltage: PowerPowerItemTurnontimeMaxvoltage
Return the POWer:POWer<x>:TURNONtime:MAXVoltage command.
Description
- This command sets or returns the maximum voltage setting of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:MAXVoltage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:MAXVoltage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:MAXVoltage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:MAXVoltage <NR3>
- POWer:POWer<x>:TURNONtime:MAXVoltage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the maximum voltage in the range 1 V to 500 V.
numoutputs
property
¶
numoutputs: PowerPowerItemTurnontimeNumoutputs
Return the POWer:POWer<x>:TURNONtime:NUMOUTputs command.
Description
- This command sets or queries the number of outputs for the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:NUMOUTputs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:NUMOUTputs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:NUMOUTputs valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:NUMOUTputs {ONE|TWO|THREE|FOUR|FIVE|SIX|SEVEN}
- POWer:POWer<x>:TURNONtime:NUMOUTputs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.ONEthrough SEVEN sets the number of outputs for the specified Turn On Time power measurement.
output1source
property
¶
output1source: PowerPowerItemTurnontimeOutput1source
Return the POWer:POWer<x>:TURNONtime:OUTPUT1SOURce command.
Description
- This command sets or queries the output 1 source of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT1SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT1SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT1SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT1SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT1SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
output1voltage
property
¶
output1voltage: PowerPowerItemTurnontimeOutput1voltage
Return the POWer:POWer<x>:TURNONtime:OUTPUT1VOLTage command.
Description
- This command sets or queries the output 1 voltage level of the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT1VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT1VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT1VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT1VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT1VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
output2source
property
¶
output2source: PowerPowerItemTurnontimeOutput2source
Return the POWer:POWer<x>:TURNONtime:OUTPUT2SOURce command.
Description
- This command sets or queries the output 2 source of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT2SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT2SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT2SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT2SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT2SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>A channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>A Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>A Math waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
output2voltage
property
¶
output2voltage: PowerPowerItemTurnontimeOutput2voltage
Return the POWer:POWer<x>:TURNONtime:OUTPUT2VOLTage command.
Description
- This command sets or queries the output 2 voltage level of the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT2VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT2VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT2VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT2VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT2VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
output3source
property
¶
output3source: PowerPowerItemTurnontimeOutput3source
Return the POWer:POWer<x>:TURNONtime:OUTPUT3SOURce command.
Description
- This command sets or queries the output 3 source of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT3SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT3SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT3SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT3SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT3SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>A channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>A Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>AMath waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
output3voltage
property
¶
output3voltage: PowerPowerItemTurnontimeOutput3voltage
Return the POWer:POWer<x>:TURNONtime:OUTPUT3VOLTage command.
Description
- This command sets or queries the output 3 voltage level of the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT3VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT3VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT3VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT3VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT3VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
output4source
property
¶
output4source: PowerPowerItemTurnontimeOutput4source
Return the POWer:POWer<x>:TURNONtime:OUTPUT4SOURce command.
Description
- This command sets or queries the output 4 source of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT4SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT4SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT4SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT4SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT4SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>A channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>A Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>A Math waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
output4voltage
property
¶
output4voltage: PowerPowerItemTurnontimeOutput4voltage
Return the POWer:POWer<x>:TURNONtime:OUTPUT4VOLTage command.
Description
- This command sets or queries the output 4 voltage level of the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT4VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT4VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT4VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT4VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT4VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
output5source
property
¶
output5source: PowerPowerItemTurnontimeOutput5source
Return the POWer:POWer<x>:TURNONtime:OUTPUT5SOURce command.
Description
- This command sets or queries the output 5 source of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT5SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT5SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT5SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT5SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT5SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>A channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>A Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>A Math waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
output5voltage
property
¶
output5voltage: PowerPowerItemTurnontimeOutput5voltage
Return the POWer:POWer<x>:TURNONtime:OUTPUT5VOLTage command.
Description
- This command sets or queries the output 5 voltage level of the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT5VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT5VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT5VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT5VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT5VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
output6source
property
¶
output6source: PowerPowerItemTurnontimeOutput6source
Return the POWer:POWer<x>:TURNONtime:OUTPUT6SOURce command.
Description
- This command sets or queries the output 6 source of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT6SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT6SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT6SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT6SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT6SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>A channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>A Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>AMath waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
output6voltage
property
¶
output6voltage: PowerPowerItemTurnontimeOutput6voltage
Return the POWer:POWer<x>:TURNONtime:OUTPUT6VOLTage command.
Description
- This command sets or queries the output 6 voltage level of the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT6VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT6VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT6VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT6VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT6VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
output7source
property
¶
output7source: PowerPowerItemTurnontimeOutput7source
Return the POWer:POWer<x>:TURNONtime:OUTPUT7SOURce command.
Description
- This command sets or queries the output 7 source of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT7SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT7SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT7SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT7SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT7SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>A channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>A Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>A Math waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
output7voltage
property
¶
output7voltage: PowerPowerItemTurnontimeOutput7voltage
Return the POWer:POWer<x>:TURNONtime:OUTPUT7VOLTage command.
Description
- This command sets or queries the output 7 voltage level of the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT7VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT7VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT7VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT7VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT7VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
type
property
¶
Return the POWer:POWer<x>:TURNONtime:TYPE command.
Description
- This command sets or queries the type of AC/DC converter used in the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:TYPE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:TYPE?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:TYPE valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:TYPE {DCDC|ACDC}
- POWer:POWer<x>:TURNONtime:TYPE?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.DCDCsets the measurement to use a DC to DC converter.ACDCsets the measurement to use an AC to DC converter.
PowerPowerItemTurnontimeFrequency
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:FREQuency command.
Description
- This command sets or queries the input frequency used by the AC or DC converter of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:FREQuency?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:FREQuency?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:FREQuency valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:FREQuency <NR3>
- POWer:POWer<x>:TURNONtime:FREQuency?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the frequency, in Hertz, from 1 Hz to 500 Hz.
PowerPowerItemTurnontimeInputlevel
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:INPUTLEVel command.
Description
- This command sets or returns the input voltage level of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:INPUTLEVel?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:INPUTLEVel?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:INPUTLEVel valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:INPUTLEVel <NR3>
- POWer:POWer<x>:TURNONtime:INPUTLEVel?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the voltage level, in volts, from -500 V to 500 V.
PowerPowerItemTurnontimeInputsource
¶
Bases: SCPICmdWriteNoArguments, SCPICmdRead
The POWer:POWer<x>:TURNONtime:INPUTSOurce command.
Description
- This command sets or queries the input source of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:INPUTSOurce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:INPUTSOurce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write()method will send thePOWer:POWer<x>:TURNONtime:INPUTSOurcecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:INPUTSOurce
- POWer:POWer<x>:TURNONtime:INPUTSOurce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemTurnontimeMaxtime
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:MAXTIMe command.
Description
- This command sets or returns the maximum turn on time of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:MAXTIMe?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:MAXTIMe?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:MAXTIMe valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:MAXTIMe <NR3>
- POWer:POWer<x>:TURNONtime:MAXTIMe?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the maximum time value, in seconds, in the range 1 second to 500 seconds.
PowerPowerItemTurnontimeMaxvoltage
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:MAXVoltage command.
Description
- This command sets or returns the maximum voltage setting of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:MAXVoltage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:MAXVoltage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:MAXVoltage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:MAXVoltage <NR3>
- POWer:POWer<x>:TURNONtime:MAXVoltage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR3>is a floating point number that represents the maximum voltage in the range 1 V to 500 V.
PowerPowerItemTurnontimeNumoutputs
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:NUMOUTputs command.
Description
- This command sets or queries the number of outputs for the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:NUMOUTputs?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:NUMOUTputs?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:NUMOUTputs valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:NUMOUTputs {ONE|TWO|THREE|FOUR|FIVE|SIX|SEVEN}
- POWer:POWer<x>:TURNONtime:NUMOUTputs?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.ONEthrough SEVEN sets the number of outputs for the specified Turn On Time power measurement.
PowerPowerItemTurnontimeOutput1source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:OUTPUT1SOURce command.
Description
- This command sets or queries the output 1 source of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT1SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT1SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT1SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT1SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT1SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the power measurement badge on the UI.CH<x>sets the channel specifier;is 1 through 8 and is limited by the number of FlexChannels in your instrument. MATH<x>= A math waveform specifier;is ≥1. REF<x>= A reference waveform specifier;is ≥1.
PowerPowerItemTurnontimeOutput1voltage
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:OUTPUT1VOLTage command.
Description
- This command sets or queries the output 1 voltage level of the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT1VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT1VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT1VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT1VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT1VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
PowerPowerItemTurnontimeOutput2source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:OUTPUT2SOURce command.
Description
- This command sets or queries the output 2 source of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT2SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT2SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT2SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT2SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT2SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>A channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>A Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>A Math waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
PowerPowerItemTurnontimeOutput2voltage
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:OUTPUT2VOLTage command.
Description
- This command sets or queries the output 2 voltage level of the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT2VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT2VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT2VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT2VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT2VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
PowerPowerItemTurnontimeOutput3source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:OUTPUT3SOURce command.
Description
- This command sets or queries the output 3 source of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT3SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT3SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT3SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT3SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT3SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>A channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>A Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>AMath waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
PowerPowerItemTurnontimeOutput3voltage
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:OUTPUT3VOLTage command.
Description
- This command sets or queries the output 3 voltage level of the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT3VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT3VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT3VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT3VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT3VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
PowerPowerItemTurnontimeOutput4source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:OUTPUT4SOURce command.
Description
- This command sets or queries the output 4 source of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT4SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT4SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT4SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT4SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT4SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>A channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>A Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>A Math waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
PowerPowerItemTurnontimeOutput4voltage
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:OUTPUT4VOLTage command.
Description
- This command sets or queries the output 4 voltage level of the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT4VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT4VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT4VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT4VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT4VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
PowerPowerItemTurnontimeOutput5source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:OUTPUT5SOURce command.
Description
- This command sets or queries the output 5 source of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT5SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT5SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT5SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT5SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT5SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>A channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>A Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>A Math waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
PowerPowerItemTurnontimeOutput5voltage
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:OUTPUT5VOLTage command.
Description
- This command sets or queries the output 5 voltage level of the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT5VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT5VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT5VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT5VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT5VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
PowerPowerItemTurnontimeOutput6source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:OUTPUT6SOURce command.
Description
- This command sets or queries the output 6 source of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT6SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT6SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT6SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT6SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT6SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>A channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>A Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>AMath waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
PowerPowerItemTurnontimeOutput6voltage
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:OUTPUT6VOLTage command.
Description
- This command sets or queries the output 6 voltage level of the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT6VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT6VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT6VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT6VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT6VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
PowerPowerItemTurnontimeOutput7source
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:OUTPUT7SOURce command.
Description
- This command sets or queries the output 7 source of the specified Turn On Time measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT7SOURce?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT7SOURce?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT7SOURce valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT7SOURce {CH<x>|REF<x>|MATH<x>}
- POWer:POWer<x>:TURNONtime:OUTPUT7SOURce?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.CH<x>A channel specifier in the range of 1 through 8 and is limited by the number of instrument input channels.REF<x>A Reference waveform specifier ≥1. This is the equivalent of the number shown on a Reference waveform badge in the UI.MATH<x>A Math waveform specifier ≥1. This is the equivalent of the number shown on a Math waveform badge in the UI.
PowerPowerItemTurnontimeOutput7voltage
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:OUTPUT7VOLTage command.
Description
- This command sets or queries the output 7 voltage level of the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:OUTPUT7VOLTage?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT7VOLTage?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:OUTPUT7VOLTage valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:OUTPUT7VOLTage <NR2>
- POWer:POWer<x>:TURNONtime:OUTPUT7VOLTage?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.<NR2>sets the output voltage value, in the range of -6,000 volts to +6,000 volts.
PowerPowerItemTurnontimeType
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TURNONtime:TYPE command.
Description
- This command sets or queries the type of AC/DC converter used in the specified Turn On Time power measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TURNONtime:TYPE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TURNONtime:TYPE?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TURNONtime:TYPE valuecommand.
SCPI Syntax
- POWer:POWer<x>:TURNONtime:TYPE {DCDC|ACDC}
- POWer:POWer<x>:TURNONtime:TYPE?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown on a power measurement badge in the UI.DCDCsets the measurement to use a DC to DC converter.ACDCsets the measurement to use an AC to DC converter.
PowerPowerItemType
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:TYPe command.
Description
- This command sets or queries the measurement type of the specified power measurement number. If the measurement number does not exist, this command creates a new power measurement, assigns the specified measurement number to the new measurement, and then assigns the measurement type to the new measurement.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:TYPe?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:TYPe?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:TYPe valuecommand.
SCPI Syntax
- POWer:POWer<x>:TYPe P <Measurement Type>
- POWer:POWer<x>:TYPe?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<Measurement Type> = CYCLEAmp | CYCLEBase | CYCLEMAx |CYCLEMIn | CYCLEPKpk | CYCLETop | DIDT | DVDT | EFFICIENCY| FREQuency | HARMonics |IMPEDANCE| INDUCTANCE | INPUTCAP| INRUSHcurrent| | IVSINTEGRALV | LINERIpple | MAGNETICLOSS |MAGPROPERTY | NDUTYCycle | NPULSEWidth | PDUTYCycle | PERIod| POWERQUALity | PPULSEWidth | RDSON | SOA | SWITCHINGLOss |SWITCHINGRIpple | TURNOFFtime | TURNONtime| CLRESPONSE | PSRR
PowerPowerItemWrap
¶
Bases: SCPICmdRead
The POWer:POWer<x>:WRAP command tree.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:WRAP?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:WRAP?query and raise an AssertionError if the returned value does not matchvalue.
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.
Properties
.degrees: ThePOWer:POWer<x>:WRAP:DEGreescommand..state: ThePOWer:POWer<x>:WRAP:STATEcommand.
degrees
property
¶
degrees: PowerPowerItemWrapDegrees
Return the POWer:POWer<x>:WRAP:DEGrees command.
Description
- This command sets or returns the phase wrap value for FRA measurements.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:WRAP:DEGrees?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:WRAP:DEGrees?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:WRAP:DEGrees valuecommand.
SCPI Syntax
- POWer:POWer<x>:WRAP:DEGrees NR3
- POWer:POWer<x>:WRAP:DEGrees?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>specifies the phase wrap value for FRA measurements.
state
property
¶
state: PowerPowerItemWrapState
Return the POWer:POWer<x>:WRAP:STATE command.
Description
- This command sets or returns the phase wrap status for FRA measurements.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:WRAP:STATE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:WRAP:STATE?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:WRAP:STATE valuecommand.
SCPI Syntax
- POWer:POWer<x>:WRAP:STATE {ON|OFF}
- POWer:POWer<x>:WRAP:STATE?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.ONspecifies that phase wrap has been turned on for FRA measurements.OFFspecifies that phase wrap has been turned off for FRA measurements.
PowerPowerItemWrapDegrees
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:WRAP:DEGrees command.
Description
- This command sets or returns the phase wrap value for FRA measurements.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:WRAP:DEGrees?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:WRAP:DEGrees?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:WRAP:DEGrees valuecommand.
SCPI Syntax
- POWer:POWer<x>:WRAP:DEGrees NR3
- POWer:POWer<x>:WRAP:DEGrees?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.<NR3>specifies the phase wrap value for FRA measurements.
PowerPowerItemWrapState
¶
Bases: SCPICmdWrite, SCPICmdRead
The POWer:POWer<x>:WRAP:STATE command.
Description
- This command sets or returns the phase wrap status for FRA measurements.
Usage
- Using the
.query()method will send thePOWer:POWer<x>:WRAP:STATE?query. - Using the
.verify(value)method will send thePOWer:POWer<x>:WRAP:STATE?query and raise an AssertionError if the returned value does not matchvalue. - Using the
.write(value)method will send thePOWer:POWer<x>:WRAP:STATE valuecommand.
SCPI Syntax
- POWer:POWer<x>:WRAP:STATE {ON|OFF}
- POWer:POWer<x>:WRAP:STATE?
Info
POWer<x>is the power measurement number. This is the equivalent of the number shown in the UI for a power measurement badge.ONspecifies that phase wrap has been turned on for FRA measurements.OFFspecifies that phase wrap has been turned off for FRA measurements.