save_and_recall
¶
The save_and_recall commands module.
These commands are used in the following models: DPO2K, DPO2KB, DPO4K, DPO4KB, DPO5K, DPO5KB, DPO70KC, DPO70KD, DPO70KDX, DPO70KSX, DPO7K, DPO7KC, DSA70KC, DSA70KD, MDO3, MDO3K, MDO4K, MDO4KB, MDO4KC, MSO2K, MSO2KB, MSO4K, MSO4KB, MSO5K, MSO5KB, MSO70KC, MSO70KDX
THIS FILE IS AUTO-GENERATED, IT SHOULD NOT BE MANUALLY MODIFIED.
Please report an issue if one is found.
Commands and Queries
- *RCL <NR1>
- *SAV <NR1>
Rcl
¶
Bases: SCPICmdWrite
The *RCL command.
Description
- This command restores the state of the oscilloscope from a copy of the settings stored in
memory (The settings are stored using the
*SAVcommand).
Usage
- Using the
.write(value)method will send the*RCL valuecommand.
SCPI Syntax
- *RCL <NR1>
Info
<NR1>is a value in the range from 1 to 10, which specifies a saved setup storage location.
Sav
¶
Bases: SCPICmdWrite
The *SAV command.
Description
- Stores the state of the oscilloscope to a specified memory location. You can use the
*RCLcommand to restore the oscilloscope to this saved state at a later time.
Usage
- Using the
.write(value)method will send the*SAV valuecommand.
SCPI Syntax
- *SAV <NR1>
Info
<NR1>specifies a location in which to save the state of the oscilloscope. Location values range from 1 through 10. Using an out-of-range location value causes an execution error. Any settings that have been stored previously at this location will be overwritten.