Commands and Queries
- DIAGnostic:DATA?
- DIAGnostic:IMMediate
- DIAGnostic:IMMediate?
- DIAGnostic:SELect {ALL|FPANel|AREGister|DTIMing|A1Memory|A2Memory|A3Memory|A4Memory|CREGister|CPLock|O1Register|O1ALevel|O1MLevel|O2Register|O2ALevel|O2MLevel}
- DIAGnostic:SELect?

Diagnostic

Bases: SCPICmdRead

The DIAGnostic command tree.

Usage
  • Using the .query() method will send the DIAGnostic? query.
  • Using the .verify(value) method will send the DIAGnostic? query and raise an AssertionError if the returned value does not match value.
Properties
  • .data: The DIAGnostic:DATA command.
  • .select: The DIAGnostic:SELect command.
  • .immediate: The DIAGnostic:IMMediate command.

data property

Return the DIAGnostic:DATA command.

Description
  • This command returns the results of last executed tests for the NORMal diagnostic type in the form of a numeric value of 0 for no errors or -330 for one or more tests failed. Additional error details can be found by using the subsystem, area, and test queries such as DIAGnostic:RESult? [,[,]].
Usage
  • Using the .query() method will send the DIAGnostic:DATA? query.
  • Using the .verify(value) method will send the DIAGnostic:DATA? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- DIAGnostic:DATA?

immediate property

Return the DIAGnostic:IMMediate command.

Description
  • This command executes all of the NORMal diagnostic tests. The query form of this command executes all of the NORMal diagnostics and returns the results in the form of numeric of values of 0 for no errors or -330 for one or more tests failed. This changes the active mode to DIAGnostic, if necessary, and returns back to the original active mode when done. This makes a single pass of all of the NORMal diagnostics.
Usage
  • Using the .query() method will send the DIAGnostic:IMMediate? query.
  • Using the .verify(value) method will send the DIAGnostic:IMMediate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write() method will send the DIAGnostic:IMMediate command.
SCPI Syntax
- DIAGnostic:IMMediate
- DIAGnostic:IMMediate?

select property

Return the DIAGnostic:SELect command.

Description
  • This command selects the self-test routines. The query form of this command returns the selected test routine. The following selections are available: ALL FPANel - Front panel read/write access test DTIMing - Data timing measurement (for AWG5000 series only) AREGister - AWG register read back A1Memory - CH1 waveform memory test A2Memory
  • CH2 waveform memory test A3Memory - CH3 waveform memory test (for AWG5000 series only) A4Memory - CH4 Waveform memory test (for AWG5000 series only) CREGister - Clock register read back CPLock - PLL Lock/unlock O1Register - Output1 register read back O1ALevel - Output1 analog level O1MLevel - Output1 Marker level (for AWG7000 series only) O2Register - Output2 register read back O2ALevel - Output2 analog level O2MLevel
  • Output2 marker level (for AWG7000 series only)
Usage
  • Using the .query() method will send the DIAGnostic:SELect? query.
  • Using the .verify(value) method will send the DIAGnostic:SELect? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the DIAGnostic:SELect value command.
SCPI Syntax
- DIAGnostic:SELect {ALL|FPANel|AREGister|DTIMing|A1Memory|A2Memory|A3Memory|A4Memory|CREGister|CPLock|O1Register|O1ALevel|O1MLevel|O2Register|O2ALevel|O2MLevel}
- DIAGnostic:SELect?

DiagnosticData

Bases: SCPICmdRead

The DIAGnostic:DATA command.

Description
  • This command returns the results of last executed tests for the NORMal diagnostic type in the form of a numeric value of 0 for no errors or -330 for one or more tests failed. Additional error details can be found by using the subsystem, area, and test queries such as DIAGnostic:RESult? [,[,]].
Usage
  • Using the .query() method will send the DIAGnostic:DATA? query.
  • Using the .verify(value) method will send the DIAGnostic:DATA? query and raise an AssertionError if the returned value does not match value.
SCPI Syntax
- DIAGnostic:DATA?

DiagnosticImmediate

Bases: SCPICmdWriteNoArguments, SCPICmdRead

The DIAGnostic:IMMediate command.

Description
  • This command executes all of the NORMal diagnostic tests. The query form of this command executes all of the NORMal diagnostics and returns the results in the form of numeric of values of 0 for no errors or -330 for one or more tests failed. This changes the active mode to DIAGnostic, if necessary, and returns back to the original active mode when done. This makes a single pass of all of the NORMal diagnostics.
Usage
  • Using the .query() method will send the DIAGnostic:IMMediate? query.
  • Using the .verify(value) method will send the DIAGnostic:IMMediate? query and raise an AssertionError if the returned value does not match value.
  • Using the .write() method will send the DIAGnostic:IMMediate command.
SCPI Syntax
- DIAGnostic:IMMediate
- DIAGnostic:IMMediate?

DiagnosticSelect

Bases: SCPICmdWrite, SCPICmdRead

The DIAGnostic:SELect command.

Description
  • This command selects the self-test routines. The query form of this command returns the selected test routine. The following selections are available: ALL FPANel - Front panel read/write access test DTIMing - Data timing measurement (for AWG5000 series only) AREGister - AWG register read back A1Memory - CH1 waveform memory test A2Memory - CH2 waveform memory test A3Memory - CH3 waveform memory test (for AWG5000 series only) A4Memory - CH4 Waveform memory test (for AWG5000 series only) CREGister - Clock register read back CPLock - PLL Lock/unlock O1Register - Output1 register read back O1ALevel - Output1 analog level O1MLevel - Output1 Marker level (for AWG7000 series only) O2Register
  • Output2 register read back O2ALevel - Output2 analog level O2MLevel - Output2 marker level (for AWG7000 series only)
Usage
  • Using the .query() method will send the DIAGnostic:SELect? query.
  • Using the .verify(value) method will send the DIAGnostic:SELect? query and raise an AssertionError if the returned value does not match value.
  • Using the .write(value) method will send the DIAGnostic:SELect value command.
SCPI Syntax
- DIAGnostic:SELect {ALL|FPANel|AREGister|DTIMing|A1Memory|A2Memory|A3Memory|A4Memory|CREGister|CPLock|O1Register|O1ALevel|O1MLevel|O2Register|O2ALevel|O2MLevel}
- DIAGnostic:SELect?